ASTM E2767-13(2018)
(Practice)Standard Practice for Digital Imaging and Communication in Nondestructive Evaluation (DICONDE) for X-ray Computed Tomography (CT) Test Methods
Standard Practice for Digital Imaging and Communication in Nondestructive Evaluation (DICONDE) for X-ray Computed Tomography (CT) Test Methods
SIGNIFICANCE AND USE
5.1 Personnel that are responsible for the creation, transfer, and storage of X-ray tomographic NDE data will use this standard. This practice defines a set of information modules that along with Practice E2339 and the DICOM standard provide a standard means to organize X-ray tomography test parameters and results. The X-ray CT test results may be displayed and analyzed on any device that conforms to this standard. Personnel wishing to view any tomographic inspection data stored according to Practice E2339 may use this document to help them decode and display the data contained in the DICONDE-compliant inspection record.
SCOPE
1.1 This practice facilitates the interoperability of X-ray computed tomography (CT) imaging equipment by specifying image data transfer and archival storage methods in commonly accepted terms. This document is intended to be used in conjunction with Practice E2339 on Digital Imaging and Communication in Nondestructive Evaluation (DICONDE). Practice E2339 defines an industrial adaptation of the NEMA Standards Publication titled Digital Imaging and Communications in Medicine (DICOM, see http://medical.nema.org), an international standard for image data acquisition, review, storage and archival storage. The goal of Practice E2339, commonly referred to as DICONDE, is to provide a standard that facilitates the display and analysis of NDE test results on any system conforming to the DICONDE standard. Toward that end, Practice E2339 provides a data dictionary and a set of information modules that are applicable to all NDE modalities. This practice supplements Practice E2339 by providing information object definitions, information modules and a data dictionary that are specific to X-ray CT test methods.
1.2 This practice has been developed to overcome the issues that arise when analyzing or archiving data from tomographic test equipment using proprietary data transfer and storage methods. As digital technologies evolve, data must remain decipherable through the use of open, industry-wide methods for data transfer and archival storage. This practice defines a method where all the X-ray CT technique parameters and test results are communicated and stored in a standard manner regardless of changes in digital technology.
1.3 This practice does not specify:
1.3.1 A testing or validation procedure to assess an implementation's conformance to the standard.
1.3.2 The implementation details of any features of the standard on a device claiming conformance.
1.3.3 The overall set of features and functions to be expected from a system implemented by integrating a group of devices each claiming DICONDE conformance.
1.4 Although this practice contains no values that require units, it does describe methods to store and communicate data that do require units to be properly interpreted. The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E2767 −13 (Reapproved 2018)
Standard Practice for
Digital Imaging and Communication in Nondestructive
Evaluation (DICONDE) for X-ray Computed Tomography (CT)
Test Methods
This standard is issued under the fixed designation E2767; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.
1. Scope* 1.3.2 The implementation details of any features of the
standard on a device claiming conformance.
1.1 This practice facilitates the interoperability of X-ray
1.3.3 The overall set of features and functions to be ex-
computed tomography (CT) imaging equipment by specifying
pected from a system implemented by integrating a group of
image data transfer and archival storage methods in commonly
devices each claiming DICONDE conformance.
accepted terms. This document is intended to be used in
conjunction with Practice E2339 on Digital Imaging and 1.4 Although this practice contains no values that require
Communication in Nondestructive Evaluation (DICONDE). units, it does describe methods to store and communicate data
Practice E2339 defines an industrial adaptation of the NEMA that do require units to be properly interpreted. The values
Standards Publication titled Digital Imaging and Communica- stated in SI units are to be regarded as standard. No other units
tions in Medicine (DICOM, see http://medical.nema.org), an of measurement are included in this standard.
international standard for image data acquisition, review,
1.5 This standard does not purport to address all of the
storage and archival storage. The goal of Practice E2339,
safety concerns, if any, associated with its use. It is the
commonly referred to as DICONDE, is to provide a standard
responsibility of the user of this standard to establish appro-
that facilitates the display and analysis of NDE test results on
priate safety, health, and environmental practices and deter-
any system conforming to the DICONDE standard. Toward
mine the applicability of regulatory limitations prior to use.
that end, Practice E2339 provides a data dictionary and a set of
1.6 This international standard was developed in accor-
information modules that are applicable to all NDE modalities.
dance with internationally recognized principles on standard-
This practice supplements Practice E2339 by providing infor-
ization established in the Decision on Principles for the
mation object definitions, information modules and a data
Development of International Standards, Guides and Recom-
dictionary that are specific to X-ray CT test methods.
mendations issued by the World Trade Organization Technical
1.2 This practice has been developed to overcome the issues
Barriers to Trade (TBT) Committee.
that arise when analyzing or archiving data from tomographic
test equipment using proprietary data transfer and storage
2. Referenced Documents
methods. As digital technologies evolve, data must remain
2.1 ASTM Standards:
decipherable through the use of open, industry-wide methods
E1316 Terminology for Nondestructive Examinations
for data transfer and archival storage. This practice defines a
E1441 Guide for Computed Tomography (CT) Imaging
method where all the X-ray CT technique parameters and test
E2339 Practice for Digital Imaging and Communication in
results are communicated and stored in a standard manner
Nondestructive Evaluation (DICONDE)
regardless of changes in digital technology.
2.2 Other Standard:
1.3 This practice does not specify:
DICOM National Electrical Manufacturers Association
1.3.1 A testing or validation procedure to assess an imple-
Standard for Digital Imaging and Communication in
mentation’s conformance to the standard.
Medicine (DICOM), 2011.
1 2
This practice is under the jurisdiction of ASTM Committee E07 on Nonde- For referenced ASTM standards, visit the ASTM website, www.astm.org, or
structive Testing and is the direct responsibility of Subcommittee E07.11 on Digital contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Imaging and Communication in Nondestructive Evaluation (DICONDE). Standards volume information, refer to the standard’s Document Summary page on
CurrenteditionapprovedJune1,2018.PublishedJuly2018.Originallyapproved the ASTM website.
in 2010. Last previous edition approved in 2013 as E2767 - 13. DOI:10.1520/ Available from National Electrical Manufacturers Association (NEMA), 1300
E2767-13R18. N. 17th St., Suite 1752, Rosslyn, VA 22209, http://www.nema.org.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E2767−13 (2018)
3. Terminology 5. Significance and Use
5.1 Personnel that are responsible for the creation, transfer,
3.1 Definitions:
and storage of X-ray tomographic NDE data will use this
3.1.1 Nondestructive evaluation terms used in this practice
standard. This practice defines a set of information modules
can be found in Standard Terminology for Nondestructive
that along with Practice E2339 and the DICOM standard
Examinations, E1316.
provide a standard means to organize X-ray tomography test
3.1.2 ComputedTomography terms used in this practice can
parameters and results. The X-ray CT test results may be
be found in Standard Guide for Computed Tomography (CT)
displayed and analyzed on any device that conforms to this
Imaging, E1441.
standard. Personnel wishing to view any tomographic inspec-
3.1.3 DICONDE terms used in this practice are defined in
tion data stored according to Practice E2339 may use this
Practice E2339.
document to help them decode and display the data contained
in the DICONDE-compliant inspection record.
4. Summary of Practice
4.1 A fundamental principle of DICONDE is the use of
6. Information Object Definitions
standard definitions and attribute formats for data communica-
6.1 X-ray CT Image IOD Description:
tion and storage. This means all systems that are DICONDE
6.1.1 The X-ray CT Image Information Object Definition
compliant use a common data dictionary and common com-
specifies an image that has been created by an X-ray CT
munication protocols. To further standardization, the elements
imaging device for NDE purposes. To avoid duplication of
in the data dictionary are organized into common groups
relevant material from the DICOM standard, the IOD defini-
referred to as information modules. The data dictionary and
tion will follow that for CTImages found in Part 3, SectionA.3
information modules common to all NDE modalities are
of the DICOM standard except as noted below in Table 1.
defined in Practice E2339.
Table1isnotstand-aloneandmustbeusedinconjunctionwith
4.2 The data dictionary and information modules specified Part 3, SectionA.3 of the DICOM standard to have a complete
in Practice E2339 do not cover the information storage
definition of the DICONDE CT information object. In addition
requirements for each individual modality (CT, DR, CR, UT,
to the DICONDE information modules specified in Practice
etc.).Additions to the data dictionary and information modules
E2339 and this standard, the DICONDE CT IOD utilizes
are required to support the individual modalities. This practice
additional DICOM standard information modules as noted in
contains the additions to the DICONDE data dictionary and
Table 1.
information modules necessary for X-ray CT inspection.
6.1.2 This IOD will use the Service-Object Pair (SOP)
Classes for the CT IOD as defined in Part 4, Section B.5 of the
4.3 The highest organizational level in the DICONDE
DICOM standard.
information model is the information object definition (IOD).
An information object definition is a collection of the infor- 6.2 X-ray CT Multi-Frame Image IOD Description:
mation modules necessary to represent a set of test results from 6.2.1 The X-ray CT Multi-Frame (CT-MF) Image Informa-
a specific modality. This practice contains information object tion Object Definition specifies a multi-frame image that has
definitions for X-ray CT inspection. been created by an X-ray CT imaging device for NDE
TABLE 1 X-ray CT-Image Information Object Definition
A
DICOM Module DICONDE Module Reference Usage
Patient Component Practice E2339, Section 7 M
Clinical Trial Subject Not Applicable
General Study Component Study Practice E2339, Section 7 M
Patient Study Not Applicable
Clinical Trial Study Not Applicable
General Series Component Series Practice E2339, Section 7 M
Clinical Trial Series Not Applicable
General Equipment NDE Equipment Practice E2339, Section 7 M
Contrast/bolus Not Applicable
CT Image NDE CT Image 7.1 M
NDE CT Detector 7.2 U
X-Ray Collimator X-Ray Collimator DICOM Part 3, U
Section C.8.19.6.12
X-Ray Filtration X-Ray Filtration DICOM Part 3, U
Section C.8.7.10
X-Ray Grid X-Ray Grid DICOM Part 3, U
Section C.8.7.11
NDE Indication Practice E2339, Section 7 U
NDE Geometry Practice E2339, Section 7 U
NDE CT Calibration 7.3 U
Data
A
Definition of usage codes can be found in Part 3, Section A.1.3 of the DICOM standard.
E2767−13 (2018)
purposes. To avoid duplication of relevant material from the DIRECT
DICOM standard, the IOD definition will follow that for SCINTILLATOR
Enhanced CT Images found in Part 3, Section A.38 of the 7.2.1.2 For X-ray CT Detector module, Detector Configu-
DICOM standard except as noted below in Table 2. Table 2 is ration (0018,7005) is specified to use the following defined
not stand-alone and must be used in conjunction with Part 3, terms:
Section A.38 of the DICOM standard to have a complete AREA
definition of the DICONDE CT-MF information object. LINEAR
6.2.2 This IOD will use the Service-Object Pair (SOP)
7.3 NDE X-ray CT Calibration Data Module:
Classes for the Enhanced CT IOD as defined in Part 4, Section
7.3.1 Table 5 specifies the Attributes that describe NDE
B.5of the DICOM standard.
X-ray CT calibration data.
7.4 NDE Enhanced CT Image Module.:
7. Information Modules
7.4.1 Table 6 specifies the attributes that describe the NDE
7.1 NDE CT Image Module:
Enhanced CT Image Module.
7.1.1 Table 3 specifies the Attributes that describe the NDE
7.4.2 Table 7 specifies the attributes that describe the NDE
X-ray CT image module.
CT X-ray Details Functional Group macro.
7.2 NDE X-ray CT Detector Module:
8. Keywords
7.2.1 Table 4 specifies the Attributes that describe NDE
X-ray CT Detectors. 8.1 DICOM; DICONDE; X-ray; computed tomography;
7.2.1.1 For X-ray CT Detector module, Detector Type CT;digitaldatatransmission;digitaldatastorage;database;file
(0018,7004) is specified to use the following defined terms: format
TABLE 2 X-ray CT-MF Image Information Object Definition
A
DICOM Module DICONDE Module Reference Usage
Patient Component Practice E2339, Section 7 M
Specimen Identification Not Applicable
Clinical Trial Subject Not Applicable
General Study Component Study Practice E2339, Section 7 M
Patient Study Not Applicable
Clinical Trial Study Not Applicable
General Series Component Series Practice E2339, Section 7 M
Clinical Trial Series Not Applicable
Synchronization Not Applicable
General Equipment NDE Equipment Practice E2339, Section 7 M
Enhanced Contrast/Bolus Not Applicable
Cardiac Synchronization Not Applicable
Respiratory Synchronization Not Applicable
Enhanced CT Image NDE Enhanced CT Image 7.4 M
NDE Indication Practice E2339, Section 7 U
NDE Geometry Practice E2339, Section 7 U
NDE CT Calibration Data 7.3 U
Acquisition Context Needed for DICOM compatibility
A
Definition of usage codes can be found in Part 3, Section A.1.3 of the DICOM standard.
E2767−13 (2018)
TABLE 3 NDE CT Image Module Attributes
A
Attribute Name Tag VR VM Type Description
Image Type (0008,0008) CS 1-n 1 Image identification characteristics.
Samples Per Pixel (0028,0002) US 1 1 Number of samples per pixel (planes) in this image.
Photometric Interpretation (0028,0004) CS 1 1 Specifies the intended interpretation of the pixel data.
Bits Allocated (0028,0100) US 1 1 Number of bits allocated for each pixel data.
Bits Stored (0028,0101) US 1 1 Number of bits stored for each pixel data.
High Bit (0028,0102) US 1 1 Most significant bit for pixel data.
Rescale Intercept (0028,1052) DS 1 1 The value b in relationship between stored values (SV) and the
output units. Output units = m*SV+b
Rescale Slope (0028,1053) DS 1 1 m in the equation specified in Rescale Intercept (0028,1052).
Rescale Type (0028,1054) LO 1 1 Specifies the output units of Rescale Slope (0028,1053) and
Rescale Intercept (0028,1052).
KVP (0018,0060) DS 1 2 Peak kilo voltage output of the X-ray generator used.
Acquisition Number (0020,0012) IS 1 2 A number identifying the single continuous gathering of data over a
period of time which resulted in this image.
Scan Options (0018,0022) CS 1-n 3 Parameters of the scanning sequence.
Data Collection Diameter (0018,0090) DS 1 3 The diameter in mm of the region over which data were collected.
Reconstruction Diameter (0018,1100) DS 1 3 Diameter in mm of the region from within which data were used in
creating the reconstruction of the image. Data may exist outside of
this portion of the region and portions of the component may exist
outside this region.
Distance Source to (0018,1110) DS 1 3 Distance in mm from source to detector center.
Detector
Distance Source to (0018,1111) DS 1 3 Distance in mm from source to isocenter (center of field of view).
Component
Gantry/Detector Tilt (0018,1120) DS 1 3 Nominal angle of tilt in degrees of the scanning gantry. Not
intended for mathematical calculations.
Table Height (0018,1130) DS 1 3 The distance in mm of the top of the part positioner table to the
center of rotation; below the center is positive.
Rotation Direction (0018,1140) DS 1 3 Direction of rotation of the source when relevant, about nearest
principal axis of equipment. Enumerated Values: CW = clockwise
CC = counterclockwise
Exposure Time (0018,1150) DS 1 3 Time of X-ray exposure in msec.
X-ray Tube Current (0018,1151) DS 1 3 X-ray Tube Current in mA.
Exposure (0018,1152) DS 1 3 The exposure expressed in mA, for example, calculated from
Exposure Time and X-ray Tube Current.
Exposure in µAs (0018,1153) DS 1 3 The exposure expressed in mµ, for example, calculated from
Exposure Time and X-ray Tube Current.
Filter Type (0018,1160) DS 1 3 Label for the type of filter inserted into the X-ray beam.
Generator Power (0018,1170) DS 1 3 Power in kW to the X-ray generator.
LINAC Energy (0014,5002) IS 1 3 The energy of the LINA
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E2767 − 13 E2767 − 13 (Reapproved 2018)
Standard Practice for
Digital Imaging and Communication in Nondestructive
Evaluation (DICONDE) for X-ray Computed Tomography (CT)
Test Methods
This standard is issued under the fixed designation E2767; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
1.1 This practice facilitates the interoperability of X-ray computed tomography (CT) imaging equipment by specifying image
data transfer and archival storage methods in commonly accepted terms. This document is intended to be used in conjunction with
Practice E2339 on Digital Imaging and Communication in Nondestructive Evaluation (DICONDE). Practice E2339 defines an
industrial adaptation of the NEMA Standards Publication titled Digital Imaging and Communications in Medicine (DICOM, see
http://medical.nema.org), an international standard for image data acquisition, review, storage and archival storage. The goal of
Practice E2339, commonly referred to as DICONDE, is to provide a standard that facilitates the display and analysis of NDE test
results on any system conforming to the DICONDE standard. Toward that end, Practice E2339 provides a data dictionary and a
set of information modules that are applicable to all NDE modalities. This practice supplements Practice E2339 by providing
information object definitions, information modules and a data dictionary that are specific to X-ray CT test methods.
1.2 This practice has been developed to overcome the issues that arise when analyzing or archiving data from tomographic test
equipment using proprietary data transfer and storage methods. As digital technologies evolve, data must remain decipherable
through the use of open, industry-wide methods for data transfer and archival storage. This practice defines a method where all
the X-ray CT technique parameters and test results are communicated and stored in a standard manner regardless of changes in
digital technology.
1.3 This practice does not specify:
1.3.1 A testing or validation procedure to assess an implementation’s conformance to the standard.
1.3.2 The implementation details of any features of the standard on a device claiming conformance.
1.3.3 The overall set of features and functions to be expected from a system implemented by integrating a group of devices each
claiming DICONDE conformance.
1.4 Although this practice contains no values that require units, it does describe methods to store and communicate data that
do require units to be properly interpreted. The values stated in SI units are to be regarded as standard. No other units of
measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
E1316 Terminology for Nondestructive Examinations
E1441 Guide for Computed Tomography (CT) Imaging
This practice is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.11 on Digital Imaging
and Communication in Nondestructive Evaluation (DICONDE).
Current edition approved Dec. 15, 2013June 1, 2018. Published January 2014July 2018. Originally approved in 2010. Last previous edition approved in 20112013 as
E2767 - 11.E2767 - 13. DOI:10.1520/E2767-13.DOI:10.1520/E2767-13R18.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E2767 − 13 (2018)
E2339 Practice for Digital Imaging and Communication in Nondestructive Evaluation (DICONDE)
2.2 Other Standard:
DICOM National Electrical Manufacturers Association Standard for Digital Imaging and Communication in Medicine
(DICOM), 2011.
3. Terminology
3.1 Definitions:
3.1.1 Nondestructive evaluation terms used in this practice can be found in Standard Terminology for Nondestructive
Examinations, E1316.
3.1.2 Computed Tomography terms used in this practice can be found in Standard Guide for Computed Tomography (CT)
Imaging, E1441.
3.1.3 DICONDE terms used in this practice are defined in Practice E2339.
4. Summary of Practice
4.1 A fundamental principle of DICONDE is the use of standard definitions and attribute formats for data communication and
storage. This means all systems that are DICONDE compliant use a common data dictionary and common communication
protocols. To further standardization, the elements in the data dictionary are organized into common groups referred to as
information modules. The data dictionary and information modules common to all NDE modalities are defined in Practice E2339.
4.2 The data dictionary and information modules specified in Practice E2339 do not cover the information storage requirements
for each individual modality (CT, DR, CR, UT, etc.). Additions to the data dictionary and information modules are required to
support the individual modalities. This practice contains the additions to the DICONDE data dictionary and information modules
necessary for X-ray CT inspection.
4.3 The highest organizational level in the DICONDE information model is the information object definition (IOD). An
information object definition is a collection of the information modules necessary to represent a set of test results from a specific
modality. This practice contains information object definitions for X-ray CT inspection.
5. Significance and Use
5.1 Personnel that are responsible for the creation, transfer, and storage of X-ray tomographic NDE data will use this standard.
This practice defines a set of information modules that along with Practice E2339 and the DICOM standard provide a standard
means to organize X-ray tomography test parameters and results. The X-ray CT test results may be displayed and analyzed on any
device that conforms to this standard. Personnel wishing to view any tomographic inspection data stored according to Practice
E2339 may use this document to help them decode and display the data contained in the DICONDE-compliant inspection record.
6. Information Object Definitions
6.1 X-ray CT Image IOD Description:
6.1.1 The X-ray CT Image Information Object Definition specifies an image that has been created by an X-ray CT imaging
device for NDE purposes. To avoid duplication of relevant material from the DICOM standard, the IOD definition will follow that
for CT Images found in Part 3, Section A.3 of the DICOM standard except as noted below in Table 1. Table 1 is not stand-alone
and must be used in conjunction with Part 3, Section A.3 of the DICOM standard to have a complete definition of the DICONDE
CT information object. In addition to the DICONDE information modules specified in Practice E2339 and this standard, the
DICONDE CT IOD utilizes additional DICOM standard information modules as noted in Table 1.
6.1.2 This IOD will use the Service-Object Pair (SOP) Classes for the CT IOD as defined in Part 4, Section B.5 of the DICOM
standard.
6.2 X-ray CT Multi-Frame Image IOD Description:
6.2.1 The X-ray CT Multi-Frame (CT-MF) Image Information Object Definition specifies a multi-frame image that has been
created by an X-ray CT imaging device for NDE purposes. To avoid duplication of relevant material from the DICOM standard,
the IOD definition will follow that for Enhanced CT Images found in Part 3, Section A.38 of the DICOM standard except as noted
below in Table 2. Table 2 is not stand-alone and must be used in conjunction with Part 3, Section A.38 of the DICOM standard
to have a complete definition of the DICONDE CT-MF information object.
6.2.2 This IOD will use the Service-Object Pair (SOP) Classes for the Enhanced CT IOD as defined in Part 4, Section B.5of
the DICOM standard.
7. Information Modules
7.1 NDE CT Image Module:
7.1.1 Table 3 specifies the Attributes that describe the NDE X-ray CT image module.
Available from National Electrical Manufacturers Association (NEMA), 1300 N. 17th St., Suite 1752, Rosslyn, VA 22209, http://www.nema.org.
E2767 − 13 (2018)
TABLE 1 X-ray CT-Image Information Object Definition
A
DICOM Module DICONDE Module Reference Usage
Patient Component Practice E2339, Section 7 M
Clinical Trial Subject Not Applicable
General Study Component Study Practice E2339, Section 7 M
Patient Study Not Applicable
Clinical Trial Study Not Applicable
General Series Component Series Practice E2339, Section 7 M
Clinical Trial Series Not Applicable
General Equipment NDE Equipment Practice E2339, Section 7 M
Contrast/bolus Not Applicable
CT Image NDE CT Image 7.1 M
NDE CT Detector 7.2 U
X-Ray Collimator X-Ray Collimator DICOM Part 3, U
Section C.8.19.6.12
X-Ray Filtration X-Ray Filtration DICOM Part 3, U
Section C.8.7.10
X-Ray Grid X-Ray Grid DICOM Part 3, U
Section C.8.7.11
NDE Indication Practice E2339, Section 7 U
NDE Geometry Practice E2339, Section 7 U
NDE CT Calibration 7.3 U
Data
A
Definition of usage codes can be found in Part 3, Section A.1.3 of the DICOM standard.
TABLE 2 X-ray CT-MF Image Information Object Definition
A
DICOM Module DICONDE Module Reference Usage
Patient Component Practice E2339, Section 7 M
Specimen Identification Not Applicable
Clinical Trial Subject Not Applicable
General Study Component Study Practice E2339, Section 7 M
Patient Study Not Applicable
Clinical Trial Study Not Applicable
General Series Component Series Practice E2339, Section 7 M
Clinical Trial Series Not Applicable
Synchronization Not Applicable
General Equipment NDE Equipment Practice E2339, Section 7 M
Enhanced Contrast/Bolus Not Applicable
Cardiac Synchronization Not Applicable
Respiratory Synchronization Not Applicable
Enhanced CT Image NDE Enhanced CT Image 7.4 M
NDE Indication Practice E2339, Section 7 U
NDE Geometry Practice E2339, Section 7 U
NDE CT Calibration Data 7.3 U
Acquisition Context Needed for DICOM compatibility
A
Definition of usage codes can be found in Part 3, Section A.1.3 of the DICOM standard.
7.2 NDE X-ray CT Detector Module:
7.2.1 Table 4 specifies the Attributes that describe NDE X-ray CT Detectors.
7.2.1.1 For X-ray CT Detector module, Detector Type (0018,7004) is specified to use the following defined terms:
DIRECT
SCINTILLATOR
7.2.1.2 For X-ray CT Detector module, Detector Configuration (0018,7005) is specified to use the following defined terms:
AREA
LINEAR
7.3 NDE X-ray CT Calibration Data Module:
7.3.1 Table 5 specifies the Attributes that describe NDE X-ray CT calibration data.
7.4 NDE Enhanced CT Image Module.:
7.4.1 Table 6 specifies the attributes that describe the NDE Enhanced CT Image Module.
7.4.2 Table 7 specifies the attributes that describe the NDE CT X-ray Details Functional Group macro.
8. Keywords
8.1 DICOM; DICONDE; X-ray; computed tomography; CT; digital data transmission; digital data storage; database; file format
E2767 − 13 (2018)
TABLE 3 NDE CT Image Module Attributes
A
Attribute Name Tag VR VM Type Description
Image Type (0008,0008) CS 1-n 1 Image identification characteristics.
Samples Per Pixel (0028,0002) US 1 1 Number of samples per pixel (planes) in this image.
Photometric Interpretation (0028,0004) CS 1 1 Specifies the intended interpretation of the pixel data.
Bits Allocated (0028,0100) US 1 1 Number of bits allocated for each pixel data.
Bits Stored (0028,0101) US 1 1 Number of bits stored for each pixel data.
High Bit (0028,0102) US 1 1 Most significant bit for pixel data.
Rescale Intercept (0028,1052) DS 1 1 The value b in relationship between stored values (SV) and the
output units. Output units = m*SV+b
Rescale Slope (0028,1053) DS 1 1 m in the equation specified in Rescale Intercept (0028,1052).
Rescale Type (0028,1054) LO 1 1 Specifies the output units of Rescale Slope (0028,1053) and
Rescale Intercept (0028,1052).
KVP (0018,0060) DS 1 2 Peak kilo voltage output of the X-ray generator used.
Acquisition Number (0020,0012) IS 1 2 A number identifying the single continuous gathering of data over a
period of time which resulted in this image.
Scan Options (0018,0022) CS 1-n 3 Parameters of the scanning sequence.
Data Collection Diameter (0018,0090) DS 1 3 The diameter in mm of the region over which data were collected.
Reconstruction Diameter (0018,1100) DS 1 3 Diameter in mm of the region from within which data were used in
creating the reconstruction of the image. Data may exist outside of
this portion of the region and portions of the component may exist
outside this region.
Distance Source to (0018,1110) DS 1 3 Distance in mm from source to detector center.
Detector
Distance Source to (0018,1111) DS 1 3 Distance in mm from source to isocenter (center of field of view).
Component
Gantry/Detector Tilt (0018,1120) DS 1 3 Nominal angle of tilt in degrees of the scanning gantry. Not
intended for mathematical calculations.
Table Height (0018,1130) DS 1 3 The distance in mm of the top of the part positioner table to the
center of rotation; below the center is positive.
Rotation Direction (0018,1140) DS 1 3 Direction of rotation of the source when relevant, about nearest
principal axis of equipment. Enumerated Values: CW = clockwise
CC = counterclockwise
Exposure Time (0018,1150) DS 1 3 Time of X-ray exposure in msec.
X-ray Tube Current (0018,1151) DS 1 3 X-ray Tube Current in mA.
Exposure (0018,1152) DS 1 3 The exposure expressed in mA, for example, calculated from
Exposure Time and X-ray Tube Current.
Exposure in μAs (0018,1153) DS 1 3 The exposure expressed in m
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.