SIGNIFICANCE AND USE
5.1 The principle use of this procedure is for the comparison of the brightness between batches of fluorescent penetrants compared to a specified standard, as a batch quality control test.  
5.2 The procedure is also utilized in monitoring the brightness of an in-use penetrant against the brightness of the unused sample of the same material.  
5.3 The significance of the results are not absolute values but rather relative comparisons at a point in time, by a particular laboratory or operator on the specified fluorometer.
SCOPE
1.1 This test method describes the techniques for comparing the brightness of the penetrants used in the fluorescent dye penetrant process. This comparison is performed under controlled conditions that eliminate most of the variables present in actual penetrant examination. Thus, the brightness factor is isolated and is measured independently of the other factors which affect the performance of a penetrant system.  
1.2 The brightness of a penetrant indication is affected by the developer with which it is used. This test method, however, measures the brightness of a penetrant on a convenient filter paper substrate which serves as a substitute for the developer.  
1.3 The brightness measurement obtained is color-corrected to approximate the color response of the average human eye. Since most examinations are done by human eyes, this number has more practical value than a measurement in units of energy emitted. Also, the comparisons are expressed as a percentage of some chosen standard penetrant because no absolute system of measurement exists at this time.  
1.4 The measurements made by this standard compare the brightness of a candidate penetrant to that of a standard penetrant when tested according to the technique. There is no known correlation between the results obtained and the brightness of actual flaw indications obtained using the penetrant in inspection.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 This practice is intended primarily for the testing of flat panel composites and sandwich core panels to an acceptance criteria most typically specified in a purchase order or other contractual document.  
5.2 Basis of Application—There are areas in this practice that require agreement between the cognizant engineering organization and the supplier, or specific direction from the cognizant engineering organization.
SCOPE
1.1 This practice covers two procedures for ultrasonic testing (UT) of flat panel (parallel surfaces) composites and flat sandwich core panels. Typical as-fabricated lay-ups include uniaxial, cross ply, and angle ply laminates, as well as honeycomb sandwich core materials. These procedures can be used throughout the life cycle of the materials: product and process design optimization, on line process control, after manufacture inspection, and in-service inspection. Contact methods, such as angle-beam techniques using shear waves, are not discussed.  
1.2 Ultrasonic testing is a common subsurface method for detection of laminar discontinuities. Two techniques can be considered based on panel surface accessibility; pulse echo for one sided and through transmission (bubblers/squirters) for two sided. As used in this practice, both require the use of a pulsed straight-beam ultrasonic longitudinal wave followed by observing indications of either the reflected (pulse-echo) or received (through transmission) wave. The general types of anomalies detected by both techniques include foreign materials, delamination, disbond/un-bond, fiber de-bonding, inclusions, porosity, and voids.  
1.3 This practice provides two ultrasonic test procedures. These test procedures can be applied to small area manual scanning and large area automated scanning. Each has its own merits and requirements for inspection and shall be selected as agreed upon in a contractual document.  
1.3.1 Test Procedure A, Pulse Echo (Non-contacting and Contacting), is at a minimum a single transducer transmitting and receiving a longitudinal wave in the range of 0.5 MHz to 20 MHz (see Fig. 1). This procedure requires access to only one side of the specimen. This procedure can be conducted by automated or manual means. Automated and manual test results may be imaged or recorded.
FIG. 1 Test Procedure A, Example Pulse Echo Apparatus Set-ups  
1.3.2 Test Procedure B, Through Transmission,  is a combination of two transducers. One transmits a longitudinal wave and the other receives the longitudinal wave in the range of 0.5 MHz to 20 MHz (see Fig. 2 for an example set-up using squirters). This procedure requires access to both sides of the specimen. This procedure is automated and the examination results are recorded.
FIG. 2 Test Procedure B, Through Transmission Apparatus Set-up Using Squirters  
1.4 This practice does not specify accept-reject criteria.  
1.5 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language
  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
3.1 The terms found in this standard are intended to be used uniformly and consistently in all nondestructive testing standards. The purpose of this standard is to promote a clear understanding and interpretation of the NDT standards in which they are used.
SCOPE
1.1 This standard covers the terminology used in the standards prepared by the E07 Committee on Nondestructive Testing. These nondestructive testing (NDT) methods include: acoustic emission, electromagnetic testing, gamma- and X-radiology, leak testing, liquid penetrant testing, magnetic particle testing, neutron radiology and gauging, ultrasonic testing, and other technical methods.  
1.2 Committee E07 recognizes that the terms examination, testing, and inspection are commonly used as synonyms in nondestructive testing. For uniformity and consistency in E07 nondestructive testing standards, Committee E07 encourages the use of the terms examination or inspection and their derivatives when describing the application of nondestructive test methods. In a specific standard, either examination or inspection shall be used consistently throughout the document. Similarly, E07 encourages the use of the term test and its derivatives when referring to the body of knowledge of a nondestructive testing method. There are, however, appropriate exceptions when the term test and its derivatives may be used to describe the application of a nondestructive test, such as measurements which produce a numeric result (for example, when using the leak testing method to perform a leak test on a component, or an ultrasonic measurement of velocity). Additionally, the term test should be used when referring to the NDT method, that is, Radiologic Testing (RT), Ultrasonic Testing (UT), and so forth. (Example: Radiologic Testing (RT) is often used to examine material to detect internal discontinuities.)
Note 1: The following sentences clarify this policy and illustrate its use:
(a) Nondestructive testing methods are used extensively for the examination or inspection of materials and components.
(b) The E07 Committee on Nondestructive Testing has prepared many documents to promote uniform usage of the nondestructive testing methods that are applied to examine or inspect materials and components.  
(c) Radiologic Testing (RT) is often used to inspect material to detect internal discontinuities.
(d) Magnetic Particle Testing (MT), Liquid Penetrant Testing (PT), and Visual Testing (VT) are often used to examine the surface of a component.
(e) The Bubble Leak Testing (BLT) method is sometimes used to leak test a pressure containing component to detect leaks.
(f) A guide for Nondestructive Testing of additively manufactured materials will describe several methods but a practice will focus on a single inspection method.  
1.3 Section A defines terms that are common to multiple NDT methods, whereas the subsequent sections define terms pertaining to specific NDT methods.  
1.4 As shown on the chart below, when a nondestructive examination or inspection produces an indication, the indication is subject to interpretation as false, nonrelevant, or relevant. If it has been interpreted as relevant, the necessary subsequent evaluation will result in the decision to accept or reject the material. With the exception of accept and reject, which retain the meaning found in most dictionaries, all the words used in the chart are defined in Section A.    
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    42 pages
    English language
  • Standard
    42 pages
    English language

SIGNIFICANCE AND USE
4.1 This guide can be used by an end user to specify infrared examinations of electrical and mechanical equipment and an infrared thermographer to perform them.  
4.2 This guide lists the joint responsibilities of the end user and the infrared thermographer when using infrared thermography.  
4.3 The purpose of an infrared examination is to identify and document exceptions in the end user's electrical or mechanical systems, or both.  
4.3.1 In electrical equipment, warm exceptions are usually created by an increase in resistance caused by loose or deteriorated connections, short circuits, overloads, load imbalances or faulty, mismatched or improperly installed components. Cool exceptions are usually caused by failed components.  
4.3.2 In mechanical equipment, warm exceptions are usually created by friction caused by improper lubrication, misalignment, worn components, or mechanical loading anomalies. Cool exceptions are usually caused by failed components.  
4.3.3 Exceptions in insulation systems are usually caused by missing or deteriorated materials, improper installation, or insufficient amounts of material.  
4.4 Providing opinions about the causes of exceptions, the integrity of the equipment, or recommendations for corrective actions require knowledge and skills beyond those of infrared thermography.  
4.5 Infrared examinations provide data about equipment at the time of examination only.  
4.6 Infrared examinations are not remedial.  
4.7 An infrared examination of electrical and mechanical equipment does not assure its proper operation. Other tests and proper maintenance are necessary to ensure their reliable performance.
SCOPE
1.1 This guide lists the responsibilities of the end user and the infrared thermographer when examining electrical and mechanical systems.  
1.2 This guide outlines the specific content required to document qualitative and quantitative infrared examinations of electrical and mechanical equipment.2  
1.3 This guide may involve use of equipment and materials in the presence of heated, moving or all of these or electrically energized equipment.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 Personnel that are responsible for the creation, transfer, and storage of X-ray tomographic NDE data will use this standard. This practice defines a set of information modules that along with Practice E2339 and the DICOM standard provide a standard means to organize X-ray tomography test parameters and results. The X-ray CT test results may be displayed and analyzed on any device that conforms to this standard. Personnel wishing to view any tomographic inspection data stored according to Practice E2339 may use this document to help them decode and display the data contained in the DICONDE-compliant inspection record.
SCOPE
1.1 This practice facilitates the interoperability of X-ray computed tomography (CT) imaging equipment by specifying image data transfer and archival storage methods in commonly accepted terms. This document is intended to be used in conjunction with Practice E2339 on Digital Imaging and Communication in Nondestructive Evaluation (DICONDE). Practice E2339 defines an industrial adaptation of the NEMA Standards Publication titled Digital Imaging and Communications in Medicine (DICOM, see http://medical.nema.org), an international standard for image data acquisition, review, storage, and archival storage. The goal of Practice E2339, commonly referred to as DICONDE, is to provide a standard that facilitates the display and analysis of NDE test results on any system conforming to the DICONDE standard. Toward that end, Practice E2339 provides a data dictionary and a set of information modules that are applicable to all NDE modalities. This practice supplements Practice E2339 by providing information object definitions, information modules and a data dictionary that are specific to X-ray CT test methods.  
1.2 This practice has been developed to overcome the issues that arise when analyzing or archiving data from tomographic test equipment using proprietary data transfer and storage methods. As digital technologies evolve, data must remain decipherable through the use of open, industry-wide methods for data transfer and archival storage. This practice defines a method where all the X-ray CT technique parameters and test results are communicated and stored in a standard manner regardless of changes in digital technology.  
1.3 This practice does not specify:  
1.3.1 A testing or validation procedure to assess an implementation's conformance to the standard.  
1.3.2 The implementation details of any features of the standard on a device claiming conformance.  
1.3.3 The overall set of features and functions to be expected from a system implemented by integrating a group of devices each claiming DICONDE conformance.  
1.4 Units—Although this practice contains no values that require units, it does describe methods to store and communicate data that do require units to be properly interpreted. The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    11 pages
    English language
  • Standard
    11 pages
    English language

SIGNIFICANCE AND USE
4.1 This practice establishes the basic parameters for the application and control of the radiographic method. This practice is written so it can be specified on the engineering drawing, specification, or contract. It is not a detailed how-to procedure to be used by the NDT facility and, therefore, must be supplemented by a detailed procedure (see 6.1). Practices E1030/E1030M, E1032, and E1416 contain information to help develop detailed technique/procedure requirements.
SCOPE
1.1 This practice2 covers the minimum requirements for radiographic examination for metallic and nonmetallic materials.  
1.2 Applicability—The criteria for the radiographic examination in this practice are applicable to all types of metallic and nonmetallic materials. When specified, it may be used for radiographic inspection of metallic or non-metallic materials, weldments, castings, and brazed materials. The requirements expressed in this practice are intended to control the quality of the radiographic images and are not intended to establish acceptance criteria for parts and materials.  
1.3 Basis of Application—There are areas in this practice that may require agreement between the cognizant engineering organization and the supplier, or specific direction from the cognizant engineering organization. These items should be addressed in the purchase order or the contract.  
1.3.1 DoD contracts.  
1.3.2 Personnel qualification, 5.1.1.  
1.3.3 Agency qualification, 5.1.2.  
1.3.4 Digitizing techniques, 5.4.5.  
1.3.5 Alternate image quality indicator (IQI) types, 5.5.3.  
1.3.6 Examination sequence, 6.6.  
1.3.7 Non-film techniques, 6.7.  
1.3.8 Radiographic quality levels, 6.9.  
1.3.9 Optical density, 6.10.  
1.3.10 IQI qualification exposure, 6.13.3.  
1.3.11 Non-requirement for IQI, 6.18.  
1.3.12 Examination coverage for welds, A2.2.2.  
1.3.13 Electron beam welds, A2.3.  
1.3.14 Geometric unsharpness, 6.23.  
1.3.15 Responsibility for examination, 6.27.1.  
1.3.16 Examination report, 6.27.2.  
1.3.17 Retention of radiographs, 6.27.8.  
1.3.18 Storage of radiographs, 6.27.9.  
1.3.19 Reproduction of radiographs, 6.27.10 and 6.27.10.1.  
1.3.20 Acceptable parts, 6.28.1.  
1.4 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    17 pages
    English language
  • Standard
    17 pages
    English language

SIGNIFICANCE AND USE
5.1 This practice is useful for characterizing material microstructure or measuring variations in microstructure that occur because of material processing conditions and thermal, mechanical, or chemical exposure (3). When applied to monolithic or composite ceramics, the procedure should reveal microstructural gradients due to density, porosity, and grain variations. This practice may also be applied to polycrystalline metals to assess variations in grain size, porosity, and multiphase constituents.  
5.2 This practice is useful for measuring and comparing microstructural variations among different samples of the same material or for sensing and measuring subtle microstructural variations within a given sample.  
5.3 This practice is useful for mapping variations in the attenuation coefficient and the attenuation spectrum as they pertain to variations in the microstructure and associated properties of monolithic ceramics, ceramic composites and metals.  
5.4 This practice is useful for establishing a reference database for comparing materials and for calibrating ultrasonic attenuation measurement equipment.  
5.5 This practice is not recommended for highly attenuating monolithics or composites that are thick, highly porous, or that have rough or highly textured surfaces. For these materials Practice E664/E664M may be appropriate. Guide E1495/E1495M is recommended for assessing attenuation differences among composite plates and laminates that may exhibit, for example, pervasive matrix porosity or matrix crazing in addition to having complex fiber architectures or thermomechanical degradation (3). The proposed ASTM Standard Practice for Measuring Ultrasonic Velocity in Advanced Ceramics (C1331) is recommended for characterizing monolithic ceramics with significant porosity or porosity variations (4).
SCOPE
1.1 This practice describes a procedure for measurement of ultrasonic attenuation coefficients for advanced structural ceramic materials. The procedure is based on a broadband buffered piezoelectric probe used in the pulse-echo contact mode and emitting either longitudinal or shear waves. The primary objective of this practice is materials characterization.  
1.2 The procedure requires coupling an ultrasonic probe to the surface of a plate-like sample and the recovery of successive front surface and back surface echoes (refer to Fig. 3). Power spectra of the echoes are used to calculate the attenuation spectrum (attenuation coefficient as a function of ultrasonic frequency) for the sample material. The transducer bandwidth and spectral response are selected to cover a range of frequencies and corresponding wavelengths that interact with microstructural features of interest in solid test samples.  
1.3 The purpose of this practice is to establish fundamental procedures for measurement of ultrasonic attenuation coefficients. These measurements should distinguish and quantify microstructural differences among solid samples and therefore help establish a reference database for comparing materials and calibrating ultrasonic attenuation measurement equipment.  
1.4 This practice applies to monolithic ceramics and also polycrystalline metals. This practice may be applied to whisker reinforced ceramics, particulate toughened ceramics, and ceramic composites provided that similar constraints on sample size, shape, and finish are met as described herein for monolithic ceramics.  
1.5 This practice sets forth the constraints on sample size, shape, and finish that will assure valid attenuation coefficient measurements. This practice also describes the instrumentat- ion, methods, and data processing procedures for accomplishing the measurements.  
1.6 This practice is not recommended for highly attenuating materials such as very thick, very porous, rough-surfaced monolithics or composites. This practice is not recommended for highly nonuniform, heterogeneous, cracked, defective, or otherwise flaw...

  • Standard
    11 pages
    English language

SIGNIFICANCE AND USE
5.1 As with conventional radiography, radioscopic examination is broadly applicable to any material or examination object through which a beam of penetrating radiation may be passed and detected including metals, plastics, ceramics, composites, and other nonmetallic materials. In addition to the benefits normally associated with radiography, radioscopic examination may be either a dynamic, filmless technique allowing the examination part to be manipulated and imaging parameters optimized while the object is undergoing examination, or a static, filmless technique wherein the examination part is stationary with respect to the X-ray beam. Systems with digital detector arrays (DDAs) or an analog component such as an electro-optic device or an analog camera may be used in dynamic mode. If achievable video rates are not adequate to examine features of interest in dynamic mode then averaging techniques with no movement of the test object shall be used – in this case, if using a DDA, Practice E2698 shall be used. If used with a high speed camera system, the user must be aware of the various image conversion materials decay time such that the converter signal can change as fast or faster than the frame rate. Linear Detector Arrays (LDAs) and flying spot systems may be considered radioscopic configurations as they are included in as shown in Guide E1000.  
5.2 This practice establishes the basic parameters for the application and control of the radioscopic examination method. This practice is written so it can be specified on the engineering drawing, specification, or contract.  
5.3 Weld Examination—Additional information on radioscopic weld examination may be found in Practice E1416.  
5.4 Casting Examination—Additional information on radioscopic casting examination may be found in Practice E1734.  
5.5 Electronic Components—Radioscopic examination of electronic components shall comply with Practice E1161.  
5.6 Explosives and Propellants—Radioscopic examination of exp...
SCOPE
1.1 This practice2 covers application details for radioscopic examination using penetrating radiation using an analog component such as an electro-optic device (for example, X-ray image intensifier (XRII) or analog camera, or both) or a Digital Detector Array (DDA) used in dynamic mode radioscopy. Radioscopy is a radiographic technique that can be used in (1) dynamic mode radioscopy to track motion or optimize radiographic parameters in real-time, or both (25 to 30 frames per second), near real-time (a few frames per second), or high speed (hundreds to thousands of frames per second) or (2) static mode radioscopy where there is no motion of the object during exposure as a filmless recording medium. This practice is not to be used for static mode radioscopy using DDAs. If static radioscopy using a DDA (that is, DDA radiography) is being performed, use Practice E2698.  
1.1.1 This practice also may be used for Linear Detector Array (LDA) applications where an LDA uses relative perpendicular motion of either the detector or component under examination to build an image line by line.  
1.1.2 This practice may also be used for “flying spot” applications where a pencil beam of X-rays rasters over an area to build an image point by point.  
1.2 This practice establishes the minimum requirements for radioscopic examination of metallic and non-metallic materials using X-ray or gamma radiation. Since the techniques involved and the applications for radioscopic examination are diverse, this practice is not intended to be limiting or restrictive, but rather to address the general applications of the technology and thereby facilitate its use. Refer to Guides E94 and E1000, and Terminology E1316, provide additional information and guidance.  
1.3 Basis of Application:  
1.3.1 The requirements of this practice and Practice E1411 shall be used together. The requirements of Practice E1411 will provide the performance qualification ...

  • Standard
    12 pages
    English language
  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
5.1 Shearography is commonly used during product process design and optimization, process control, after manufacture inspection, and in service inspection, and can be used to measure static and dynamic axial (tensile and compressive) strain, as well as shearing, Poisson, bending, and torsional strains. The general types of defects detected by shearography include delamination, deformation under load, disbond/unbond, microcracks, and thickness variation.  
5.2 Additional information is given in Guide E2533 about the advantages and limitations of the shearography technique, use of related ASTM documents, specimen geometry and size considerations, calibration and standardization, and physical reference standards.  
5.3 For procedures for shearography of filament-wound pressure vessels, otherwise known as composite overwrapped pressure vessels, consult Guide E2982.  
5.4 Factors that influence shearography and therefore shall be reported include but are not limited to the following: laminate (matrix and fiber) material, lay-up geometry, fiber volume fraction (flat panels); facing material, core material, facing stack sequence, core geometry (cell size); core density, facing void content, and facing volume percent reinforcement (sandwich core materials); processing and fabrication methods, overall thickness, specimen alignment, specimen conditioning, specimen geometry, and test environment (flat panels and sandwich core materials). Shearography has been used with excellent results for composite and metal face sheet sandwich panels with both honeycomb and foam cores, solid monolithic composite laminates, foam cryogenic fuel tank insulation, bonded cork insulation, aircraft tires, elastomeric and plastic coatings. Frequently, defects at multiple and far side bond lines can be detected.
SCOPE
1.1 This practice describes procedures for shearography of polymer matrix composites and sandwich core materials made entirely or in part from fiber-reinforced polymer matrix composites. The composite materials under consideration typically contain continuous high modulus (greater than 20 GPa (3 × 106 psi)) fibers, but may also contain discontinuous fiber, fabric, or particulate reinforcement.  
1.2 This practice describes established shearography procedures that are currently used by industry and federal agencies that have demonstrated utility in quality assurance of polymer matrix composites and sandwich core materials during product process design and optimization, manufacturing process control, after manufacture inspection, and in service inspection.  
1.3 This practice has utility for testing of polymer matrix composites and sandwich core materials containing but not limited to bismaleimide, epoxy, phenolic, poly(amideimide), polybenzimidazole, polyester (thermosetting and thermoplas- tic), poly(ether ether ketone), poly(ether imide), polyimide (thermosetting and thermoplastic), poly(phenylene sulfide), or polysulfone matrices; and alumina, aramid, boron, carbon, glass, quartz, or silicon carbide fibers. Typical as-fabricated geometries include uniaxial, cross-ply and angle-ply laminates; as well as honeycomb and foam core sandwich materials and structures.  
1.4 This practice does not specify accept-reject criteria and is not intended to be used as a means for approving polymer matrix composites or sandwich core materials for service.  
1.5 To ensure proper use of the referenced standards, there are recognized nondestructive testing (NDT) specialists that are certified according to industry and company NDT specifications. It is recommended that an NDT specialist be a part of any composite component design, quality assurance, in-service maintenance, or damage examination activity.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental...

  • Standard
    10 pages
    English language

SIGNIFICANCE AND USE
4.1 This standard provides a practice for RIQR evaluations of film and non-film imaging systems when exposed through an absorber material. Three alternate data evaluation methods are provided in Section 9. Determining RIQR requires the comparison of at least two radiographs or radiographic processes whereby the relative degree of image quality difference may be determined using the EPS plaque arrangement of Fig. 1 as a relative image quality indicator (RIQI). In conjunction with the RIQI, a specified radiographic technique or method must be established and carefully controlled for each radiographic process. This practice is designed to allow the determination of subtle changes in EPS that may arise to radiographic imaging system performance levels resultant from process improvements/changes or change of equipment attributes. This practice does not address relative unsharpness of a radiographic imaging system as provided in Practice E2002. The common element with any relative comparison is the use of the same RIQI arrangement for both processes under evaluation.  
4.2 In addition to the standard evaluation method described in Section 9, there may be other techniques/methods in which the basic RIQR arrangement of Fig. 1 might be utilized to perform specialized assessments of relative image quality performance. For example, other radiographic variables can be altered to facilitate evaluations provided these differences are known and documented for both processes. Where multiple radiographic process variables are evaluated, it is incumbent upon the user of this practice to control those normal process attributes to the degree suitable for the application. Specialized RIQR techniques may also be useful with micro focus X-ray, isotope sources of radiation or with the use of non-film radiographic imaging systems. RIQR may also be useful in evaluating imaging systems with alternate materials (RIQI and base plate) such as plastic, copper-nickel, or aluminum. When using any ...
SCOPE
1.1 This standard covers a practice whereby industrial radiographic imaging systems or techniques may be comparatively assessed using the concept of relative image quality response (RIQR). Changes within a radiographic technique such as film/detector types, distances, or filtering/collimation can be comparatively assessed using this standard. The RIQR method presented within this practice is based upon the use of equivalent penetrameter sensitivity (EPS) described within Practice E1025 and subsection 5.4 of this practice. Fig. 1 illustrates a relative image quality indicator (RIQI) that has four different plaque thicknesses (0.38 mm, 0.25 mm, 0.20 mm, and 0.13 mm (0.015 in., 0.010 in., 0.008 in., and 0.005 in.)) sequentially positioned (from top to bottom) on an absorber plate of a specified material and thickness. The four plaques contain a total of 14 different arrays of penetrameter-type hole sizes designed to render varied conditions of threshold visibility when exposed to the appropriate radiation. Each “EPS” array consists of 30 identical holes; thus, providing the user with a quantity of threshold sensitivity levels suitable for relative image qualitative response comparisons. There are two standard materials (steel and plastic) specified herein for the RIQI and absorber. For special applications the user may design a non-standard RIQI-absorber configuration; however the RIQI configuration shall be controlled by a drawing similar to Fig. 1. Use of a non-standard RIQI-absorber configuration shall be described in the user’s written technique and approved by the CEO.  
1.2 This practice is not intended to qualify the performance of a specific radiographic technique nor for assurance that a radiographic technique will detect specific discontinuities in a specimen undergoing radiographic examination.  
1.3 This practice is not intended to be used to classify or derive performance classification categories for radiog...

  • Standard
    8 pages
    English language
  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
5.1 This practice is intended primarily for the manual ultrasonic scanning of electrofusion joints used in the construction and maintenance of polyethylene piping systems.  
5.2 Polyethylene piping has been used instead of steel alloys in the petrochemical, power, water, gas distribution, and mining industries due to its reliability and resistance to corrosion and erosion.  
5.3 This practice is not intended to provide 100 % joint examination. This practice specifies a minimum scanning grid that represents only a portion of the welded interface. As such, there exists a possibility of omitting flaws. In addition, selected areas of the welded interface may not be accessible. The extent of examination shall be specified in the contractual agreement.  
5.4 The joining process can be subject to a variety of flaws including, but not limited to, lack of fusion, particulate contamination, short-stab depth, inclusions, and voids.  
5.5 Polyethylene material can have a range of acoustic characteristics that make electrofusion joint examination difficult. Polyethylene materials are highly attenuative, which often limits the use of higher ultrasonic frequencies. It also exhibits a natural high frequency filtering effect. An example of the range of acoustic characteristics is provided in Table 1.6 The table notes the wide range of acoustic velocities reported in the literature. This makes it essential that the reference blocks are made from pipes with the same Specification D3350 density cell classification as the electrofusion fitting examined. (A) A range of velocity and attenuation values have been noted in the literature (1-9).  
5.6 Polyethylene is reported to have a shear velocity of 987 m/s. However, due to extremely high attenuation in shear mode (on the order of 5 dB/mm [127 dB/inch] at 2 MHz) no practical examinations can be carried out using shear mode (6).  
5.7 Due to the wide range of applications, joint acceptance criteria for polyethylene pipe are usually ...
SCOPE
1.1 This practice establishes a procedure for ultrasonic testing (UT) of electrofusion joints in polyethylene pipe systems. This practice provides one ultrasonic examination procedure for ultrasonic pulse-echo straight beam contact testing, using straight-beam longitudinal waves introduced by direct contact of the search unit with the material being examined.  
1.2 The practice is intended to be used on polyethylene electrofusion socket (for example, couplings) and saddle (for example, tees) fittings for use on polyethylene pipe ranging in diameters from nominal 0.5 in. to 12 in. [12 mm to 300 mm] with pipe dimension ratios (DR) ranging from 6.3 to 17. Greater and lesser thicknesses and greater and lesser diameters may be tested using this practice if the technique can be demonstrated to provide adequate detection on mockups of the same wall thickness and geometry.  
1.3 This practice does not address ultrasonic examination of butt fusions. Ultrasonic testing of polyethylene butt fusion joints is addressed in Practice E3044/E3044M.
Note 1: The notes in this practice are for information only and shall not be considered part of this practice.
Note 2: This standard references HDPE and MDPE materials for pipe applications defined by Specification D3350.  
1.4 This practice does not specify acceptance criteria. Refer to Specification F1055 and Practice F1290 for destructive acceptance criteria.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of the other, and values from the two systems shall not be combined.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safet...

  • Standard
    9 pages
    English language

ABSTRACT
This practice covers the procedure for sorting electrically conductive materials using the thermoelectric method, which is based on the seebeck effect. The procedure relates to the use of direct- and comparator-type thermoelectric instruments for distinguishing variations in materials which affect the thermoelectric properties of those materials. The two techniques that are primarily used in thermoelectric sorting are direct and comparative instrumentation. In the direct instruments, equipment is standardized by placing materials with known chemistry and metallurgical structure in the test system. In the comparative instruments, the thermoelectric response of the test piece is compared with that of a known standard(s) and the response indicates whether the piece is within the acceptance limits. The electronic apparatus shall be capable of maintaining a sufficient temperature differential across the electrodes to produce a suitable thermoelectric voltage. The different procedures for sorting electrically conductive materials are presented in details.
SCOPE
1.1 This practice covers the procedure for sorting materials using the thermoelectric method, which is based on the Seebeck effect. The procedure relates to the use of direct- and comparator-type thermoelectric instruments for distinguishing variations in materials which affect the thermoelectric properties of those materials.  
1.2 While the practice is most commonly applied to the sorting of metals, it may be applied to other electrically conductive materials.  
1.3 Thermoelectric sorting may also be applied to the sorting of materials on the basis of plating thickness, case depth, and hardness.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 As with conventional radiography, radioscopic examination is broadly applicable to the many materials and object configurations which may be penetrated with X-rays or gamma rays. The high degree of variation in architecture and performance among radioscopic systems due to component selection, physical arrangement, and object variables makes it necessary to establish the performance that the selected radioscopic system is capable of achieving in specific applications. The manufacturer or integrator of the radioscopic system, as well as the user, require a common basis for determining the performance level of the radioscopic system.  
5.2 This practice does not purport to provide a method to measure the performance of individual radioscopic system components that are manufactured according to a variety of industry standards. This practice covers measurement of the combined performance of the radioscopic system elements when operated together as a functional radioscopic system.  
5.3 This practice addresses the performance of radioscopic systems in the static mode or dynamic mode, that can allow relative test-part motion between source, part, and detector, and may or may not have the ability to effect parameter changes during the radioscopic examination process. Users of radioscopy are cautioned that the dynamic aspects of radioscopy can have beneficial as well as detrimental effects upon system performance.  
5.4 Radioscopic system performance measured pursuant to this practice does not guarantee the level of performance which may be realized in actual operation but does provide a baseline against which periodic performance evaluations can be compared to ensure the system is operating within established limits. The effects of object-geometry and orientation-generated scattered radiation cannot be reliably predicted by a standardized examination. All radioscopic systems age and degrade in performance as a function of time. Maintenance and operator adjustments, i...
SCOPE
1.1 This practice covers test and measurement details for measuring the performance of X-ray and gamma ray radioscopic systems. Radioscopy is a radiographic technique that can be used in (1) dynamic mode radioscopy to track motion or optimize radiographic parameters in real-time (25 to 30 frames per second), or both, near real-time (a few frames per second), or high speed (hundreds to thousands of frames per second) or (2) static mode radioscopy where there is no motion of the object during exposure as a filmless recording medium. This practice2 provides application details for radioscopic examination using penetrating radiation using an analog component such as an electro-optic device (for example, X-ray image intensifier (XRII) or analog camera, or both) or a Digital Detector Array (DDA) used in dynamic mode radioscopy. This practice is not to be used for static mode radioscopy using DDAs. If static radioscopy using a DDA (that is, DDA radiography) is being performed, use Practice E2698.  
1.1.1 This practice also may be used for Linear Detector Array (LDA) applications where an LDA uses relative perpendicular motion between the detector and component to build an image line by line.  
1.1.2 This practice may also be used for “flying spot” applications where a pencil beam of X-rays rasters over an object to build an image point by point.  
1.2 Basis of Application:  
1.2.1 The requirements of this practice and Practice E1255 shall be used together. The requirements of Practice E1255 provide the minimum requirements for radioscopic examination of materials. This practice is intended as a means of initially qualifying and re-qualifying a radioscopic system for a specified application by determining its performance when operated in a static or dynamic mode. Re-qualification may require agreement between the cognizant engineering organization and the supplier, or specific direction from the cognizant engineering organi...

  • Standard
    13 pages
    English language
  • Standard
    13 pages
    English language

SIGNIFICANCE AND USE
5.1 This practice is intended for the semi-automated or automated ultrasonic examination of electrofusion joints used in the construction and maintenance of polyethylene piping systems.  
5.2 Polyethylene piping has been used instead of steel alloys in the petrochemical, power, water, gas distribution, and mining industries due to its reliability and resistance to corrosion and erosion.  
5.3 The joining process can be subject to a variety of flaws including, but not limited to: lack of fusion, cold fusion, particulate contamination, inclusions, short stab depth, and voids.  
5.4 Polyethylene material can have a range of acoustic characteristics that make electrofusion joint examination difficult. Polyethylene materials are highly attenuative, which often limits the use of higher ultrasonic frequencies. It also exhibits a natural high frequency filtering effect. An example of the range of acoustic characteristics is provided in Table 1.6 The table notes the wide range of acoustic velocities reported in the literature. This makes it essential that the reference blocks are made from pipe grade polyethylene with the same density cell class as the electrofusion fitting examined. (A) A range of velocity and attenuation values have been noted in the literature (1-9).  
5.5 Polyethylene is reported to have a shear velocity of 987 m/s. However, due to extremely high attenuation in shear mode (on the order of 5 dB/mm (127 dB/in.) at 2 MHz) no practical examinations can be carried out using shear mode (6).  
5.6 Due to the wide range of applications, joint acceptance criteria for polyethylene pipe are usually project-specific.  
5.7 A cross-sectional view of a typical joint between polyethylene pipe and an electrofusion coupling is illustrated in Fig. 1.
FIG. 1 Typical Cross-Sectional View of an Electrofusion Coupling Joint
SCOPE
1.1 This practice covers procedures for phased array ultrasonic testing (PAUT) of electrofusion joints in polyethylene pipe systems. Although high density polyethylene (HDPE) and medium density polyethylene (MDPE) materials are most commonly used, the procedures described may apply to other types of polyethylene.  
Note 1: The notes in this practice are for information only and shall not be considered part of this practice.
Note 2: This standard references HDPE and MDPE for pipe applications defined by Specification D3350.  
1.2 This practice does not address ultrasonic examination of butt fusions. Ultrasonic testing of polyethylene butt fusion joints is addressed in Practice E3044/E3044M.  
1.3 Phased array ultrasonic testing (PAUT) of polyethylene electrofusion joints uses longitudinal waves introduced by an array probe mounted on a zero degree wedge. This practice is intended to be used on polyethylene electrofusion couplings for use on polyethylene pipe ranging in diameters from nominal 4 in. to 28 in. (100 mm to 710 mm) and for coupling wall thicknesses from 0.3 in. to 2 in. (8 mm to 50 mm). Greater and lesser thicknesses and diameters may be tested using this standard practice if the technique can be demonstrated to provide adequate detection on mockups of the same geometry.  
1.4 This practice does not specify acceptance criteria.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of the other, and values from the two systems shall not be combined.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recogniz...

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 Absolute probe coil methods, when used in conjunction with reference standards of known value, provide a means for determining the electrical conductivity of nonmagnetic materials.  
4.2 Electrical conductivity of a specimen, when used in conjunction with another method listed and compared to reference charts, can be used as a means of determining: (1) type of metal or alloy, (2) type of heat treatment (for aluminum this evaluation should be used in conjunction with a hardness examination), (3) aging of the alloy, (4) effects of corrosion, (5) heat damage, (6) temper, and (7) hardness.
SCOPE
1.1 This test method covers a procedure for determining the electrical conductivity of nonmagnetic materials, typically nonmagnetic metals, using the electromagnetic (eddy current) method. The procedure has been written primarily for use with commercially available direct reading electrical conductivity instruments. General purpose eddy current instruments may also be used for electrical conductivity measurements but will not be addressed in this test method.  
1.2 This test method is applicable to nonmagnetic materials that have either a flat or slightly curved surface and includes metals with or without a thin nonconductive coating.  
1.3 Eddy current determinations of electrical conductivity may be used in the sorting of nonmagnetic materials with respect to variables such as type of alloy, aging, cold deformation, heat treatment, effects associated with non-uniform heating or overheating, and effects of corrosion. The usefulness of the examinations of these properties is dependent on the amount of electrical conductivity change caused by a change in the specific variable.  
1.4 Electrical conductivity, when evaluated with eddy current instruments, is usually expressed as a percentage of the conductivity of the International Annealed Copper Standard (% IACS) or Siemens/meter (S/m). The conductivity of the Annealed Copper Standard is defined to be 0.58 × 108 S/m (100 % IACS) at 20 °C.  
1.5 The values stated in SI units are regarded as standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 The velocity measurements described in this practice may be used to characterize material variations that affect mechanical or physical properties. This procedure is useful for measuring variations in microstructural features such as grain structure, pore fractions, and density variations in monolithic ceramics.  
4.2 Velocity measurements described herein can assess subtle variations in porosity within a given material or component, as, for example, in ceramic superconductors and structural ceramic specimens (2, 3).  
4.3 In addition to ceramics and ceramic composites, the velocity measurements described herein may be applied to polycrystalline and single crystal metals, metal matrix composites, and polymer matrix composites.  
4.4 An alternative technique for velocity measurement is given in Practice E494.
SCOPE
1.1 This practice covers a procedure for measurement of ultrasonic velocity in structural engineering solids such as monolithic ceramics, toughened ceramics, and ceramic matrix composites.  
1.2 This practice is based on the broadband pulse-echo contact ultrasonic method. The procedure involves a computer-implemented, frequency-domain method for precise measurement of time delays between pairs of echoes returned by the back surface of a test sample or part.  
1.3 This practice describes a procedure for using a digital cross-correlation algorithm for velocity measurement. The cross-correlation function yields a time delay between any two echo waveforms (1).2  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    7 pages
    English language

SIGNIFICANCE AND USE
5.1 The purpose of this practice is to outline a procedure for using GWT to locate areas in metal pipes in which wall loss has occurred due to corrosion or erosion.  
5.2 GWT does not provide a direct measurement of wall thickness, but is sensitive to a combination of the CSC and circumferential extent and axial extent of any metal loss. Based on this information, a classification of the severity can be assigned.  
5.3 The GWT method provides a screening tool to quickly identify any discontinuity along the pipe. Where a possible defect is found, follow-up inspection of suspected areas with ultrasonic testing or other NDT methods is normally required to obtain detailed thickness information, nature, and extent of damage.  
5.4 GWT also provides some information on the axial length of a discontinuity, provided that the axial length is longer than roughly a quarter of the wavelength of the excitation signal.  
5.5 The identification and severity assessment of any possible defects is qualitative only. An interpretation process to differentiate between relevant and non-relevant signals is necessary.  
5.6 This practice only covers the application specified in the scope. The GWT method has the capability and can be used for applications where the pipe is insulated, buried, in road crossings, and where access is limited.  
5.7 GWT shall be performed by qualified and certified personnel, as specified in the contract or purchase order. Qualifications shall include training specific to the use of the equipment employed, interpretation of the test results and guided wave technology.  
5.8 A documented program that includes training, examination and experience for the GWT personnel certification shall be maintained by the supplying party.
SCOPE
1.1 This practice provides a procedure for the use of guided wave testing (GWT), also previously known as long range ultrasonic testing (LRUT) or guided wave ultrasonic testing (GWUT).  
1.2 GWT utilizes ultrasonic guided waves, sent in the axial direction of the pipe, to non-destructively test pipes for defects or other features by detecting changes in the cross-section or stiffness of the pipe, or both.  
1.3 GWT is a screening tool. The method does not provide a direct measurement of wall thickness or the exact dimensions of defects/defected area; an estimate of the defect severity however can be provided.  
1.4 This practice is intended for use with tubular carbon steel or low-alloy steel products having Nominal Pipe size (NPS) 2 to 48 corresponding to 60.3 mm to 1219.2 mm (2.375 in. to 48 in.) outer diameter, and wall thickness between 3.81 mm and 25.4 mm (0.15 in. and 1 in.).  
1.5 This practice covers GWT using piezoelectric transduction technology.  
1.6 This practice only applies to GWT of basic pipe configuration. This includes pipes that are straight, constructed of a single pipe size and schedules, fully accessible at the test location, jointed by girth welds, supported by simple contact supports and free of internal, or external coatings, or both; the pipe may be insulated or painted.  
1.7 This practice provides a general procedure for performing the examination and identifying various aspects of particular importance to ensure valid results, but actual interpretation of the data is excluded.  
1.8 This practice does not establish an acceptance criterion. Specific acceptance criteria shall be specified in the contractual agreement by the responsible system user or engineering entity.  
1.9 Units—The values stated in SI units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.  
1.10 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and de...

  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
5.1 The POD analysis method described herein is based on a well-known and well established statistical regression method. It shall be used to quantify the demonstrated POD for a specific set of examination parameters and known range of discontinuity sizes under the following conditions.  
5.1.1 The initial response from a nondestructive evaluation inspection system is ultimately binary in nature (that is, hit or miss).  
5.1.2 Discontinuity size is the predictor variable and can be accurately quantified.  
5.1.3 A relationship between discontinuity size and POD exists and is best described by a generalized linear model with the appropriate link function for binary outcomes.  
5.2 This practice does not limit the use of a generalized linear model with more than one predictor variable or other types of statistical models if justified as more appropriate for the hit/miss data.  
5.3 If the initial response from a nondestructive evaluation inspection system is measurable and can be classified as a continuous variable (for example, data collected from an Eddy Current inspection system), then Practice E3023 may be more appropriate.  
5.4 Prior to performing the analysis it is assumed that the discontinuity of interest is clearly defined; the number and distribution of induced discontinuity sizes in the POD specimen set is known and well-documented; discontinuities in the POD specimen set are unobstructed; and the POD examination administration procedure (including data collection method) is well-designed, well-defined, under control, and unbiased. The analysis results are only valid if convergence is achieved and the model adequately represents the data.  
5.5 The POD analysis method described herein is consistent with the analysis method for binary data described in MIL-HDBK-1823A, and is included in several widely utilized POD software packages to perform a POD analysis on hit/miss data. It is also found in statistical software packages that have generalized line...
SCOPE
1.1 This practice covers the procedure for performing a statistical analysis on nondestructive testing hit/miss data to determine the demonstrated probability of detection (POD) for a specific set of examination parameters. Topics covered include the standard hit/miss POD curve formulation, validation techniques, and correct interpretation of results.  
1.2 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    14 pages
    English language
  • Standard
    14 pages
    English language

SIGNIFICANCE AND USE
5.1 Liquid penetrant testing methods indicate the presence, location, and to a limited extent, the nature and magnitude of the detected discontinuities. Each of the various penetrant methods has been designed for specific uses such as critical service items, volume of parts, portability, or localized areas of examination. The method selected will depend accordingly on the design and service requirements of the parts or materials being tested.
SCOPE
1.1 This practice2 covers procedures for penetrant examination of materials. Penetrant testing is a nondestructive testing method for detecting discontinuities that are open to the surface such as cracks, seams, laps, cold shuts, shrinkage, laminations, through leaks, or lack of fusion and is applicable to in-process, final, and maintenance examinations. It can be effectively used in the examination of nonporous, metallic materials, ferrous and nonferrous metals, and of nonmetallic materials such as nonporous glazed or fully densified ceramics, as well as certain nonporous plastics, and glass.  
1.2 This practice also provides a reference:  
1.2.1 By which a liquid penetrant examination process recommended or required by individual organizations can be reviewed to ascertain its applicability and completeness.  
1.2.2 For use in the preparation of process specifications and procedures dealing with the liquid penetrant testing of parts and materials. Agreement by the customer requesting penetrant testing is strongly recommended. All areas of this practice may be open to agreement between the cognizant engineering organization and the supplier, or specific direction from the cognizant engineering organization.  
1.2.3 For use in the organization of facilities and personnel concerned with liquid penetrant testing.  
1.3 This practice does not indicate or suggest criteria for evaluation of the indications obtained by penetrant testing. It should be pointed out, however, that after indications have been found, they must be interpreted or classified and then evaluated. For this purpose there must be a separate code, standard, or a specific agreement to define the type, size, location, and direction of indications considered acceptable, and those considered unacceptable.  
1.4 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    19 pages
    English language
  • Standard
    19 pages
    English language

SIGNIFICANCE AND USE
4.1 Purpose—Practices to be employed for the radiographic examination of materials and components with neutrons using digital neutron detectors are outlined herein. They are intended as a guide for the assessment of a digital neutron radiograph’s characteristics. For information on neutron beam lines for imaging and film neutron radiography, refer to Guide E748.  
4.2 Limitations—Acceptance standards have not been established for any material or production process. Neutron radiography, whether performed by means of a reactor, an accelerator, subcritical assembly, or radioactive source, will be consistent in sensitivity and spatial resolution only if the consistency of all details of the technique, such as neutron source, collimation, geometry, imaging system, etc., are maintained. This guide is limited to the use of digital neutron detectors in combination with neutron conversion materials for image recording. This guide is intended for use with thermal and cold neutron spectrums. The production of thermal neutron radiographs by employing the use of film and appropriate conversion screens is covered in Guide E748.  
4.3 Interpretation and Acceptance Standards—Interpretat- ion and acceptance standards are not covered by this guide. Designation of accept-reject standards is recognized to be within the cognizance of product specifications.  
4.4 Other Aspects of the Neutron Radiographic Process—For many important aspects of neutron radiography such as technique, files, viewing of radiographs, storage of radiographs, film processing, and record keeping, refer to Guide E94, which covers these aspects for X-ray radiography. (See Section 2.)
SCOPE
1.1 This guide covers the evaluation, qualification, and quantification of digital neutron images. These images can be acquired by many methods, including: neutron sensitive imaging plates (Computed Radiography – CR), Digital Detector Arrays – DDA’s (amorphous silicon, CMOS, CCD, etc.), micro-channel plates, neutron sensitive fluoroscopes, neutron sensitive scintillators coupled to optical cameras, digitized radiographic films, and linear diode arrays.  
1.2 This guide does not purport to establish what is considered an acceptable image but is intended to only give guidance on digital neutron imaging, as well as image quality metrics of importance, and how they can be measured and reported.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    7 pages
    English language

SIGNIFICANCE AND USE
4.1 IEM Applications and Capabilities—IEM has been successfully applied to a wide range of NDT applications in the manufacture, maintenance, and repair of metallic and non-metallic parts. Examples of anomalies detected are discussed in 1.1 and 6.2. IEM has been proven to provide fast, cost-effective, and accurate NDT solutions in nearly all manufacturing, maintenance, or repair modalities. Examples of the successful application focuses include, but are not limited to: sintered powder metals, castings, forgings, stampings, ceramics, glass, wood, weldments, heat treatment, composites, additive manufacturing, machined products, and brazed products.  
4.2 General Approach and Equipment Requirements for IEM:  
4.2.1 IEM systems are comprised of hardware and software capable of inducing vibrations, recording the component response to the induced vibrations, and executing analysis of the data collected.  
4.2.2 Hardware Requirements—Examples of a tabletop impact excitation system and a production-grade drop excitation system are shown in Fig. 1 and Fig. 2, respectively. IEM systems include: an excitation device (for example, modal hammer / impact device / dropping system) providing an impulse excitation to the object, a vibration detector (for example., microphone), a signal amplifier, an Analog-to-Digital Converter (ADC), an embedded logic, and a data User Interface (UI). Tested parts can typically be on any surface type, but they can also be supported (for example, foam support, held with an elastic) in consideration of possible damping influences. The following schematics show the basic parts for an impact excitation approach (Fig. 3) and a drop excitation approach (Fig. 4).
FIG. 1 IEM Tabletop Testing System Using a Non-Instrumented Impactor
FIG. 2 Production-Grade Drop Excitation System
FIG. 3 Schematic of Impact Excitation Approach
FIG. 4 Schematic of Drop Excitation Approach  
4.3 Constraints and Limitations:  
4.3.1 IEM needs a change in structural i...
SCOPE
1.1 This practice covers a general procedure for using the Impulse Excitation Method (IEM) to facilitate natural frequency measurement and detection of defects and material variations in metallic and non-metallic parts. This test method is also known as Impulse Excitation Technique (IET), Acoustic Resonance Testing (ART), ping testing, tap testing, and other names. IEM is listed as a Resonance Ultrasound Spectroscopy (RUS) method. The method applies an impulse load to excite and then record resonance frequencies of a part. These recorded resonance frequencies are compared to a reference population or within subgroups/families of examples of the same part, or modeled frequencies, or both.  
1.2 Absolute frequency shifting, resonance damping, and resonance pattern differences can be used to distinguish acceptable parts from parts with material differences and defects. These defects and material differences include, cracks, voids, porosity, material elastic property differences, and residual stress. IEM can be applied to parts made with manufacturing processes including, but not limited to, powdered metal sintering, casting, forging, machining, composite layup, and additive manufacturing (AM).  
1.3 This practice is intended for use with instruments capable of exciting, measuring, recording, and analyzing multiple whole body, mechanical vibration resonance frequencies in acoustic or ultrasonic frequency ranges, or both. This practice does not provide inspection acceptance criteria for parts. However, it does discuss the processes for establishing acceptance criteria specific to impulse testing. These criteria include frequency acceptability windows for absolute frequency shifting, scoring criteria for statistical analysis methods (Z-score), Gage Repeatability & Reproducibility (R&R) for diagnostic resonance modes, and inspection criteria adjustment (compensation) for manufacturing process and environmental variations.  
...

  • Standard
    14 pages
    English language

SIGNIFICANCE AND USE
6.1 This guide provides procedures for the application of contact straight-beam examination for the detection and quantitative evaluation of discontinuities in materials.  
6.2 Although not all requirements of this guide can be applied universally to all examinations, situations, and materials, it does provide basis for establishing contractual criteria between the users, and may be used as a general guide for preparing detailed specifications for a particular application.  
6.3 This guide is directed towards the evaluation of discontinuities detectable with the beam normal to the entry surface. If discontinuities or other orientations are of concern, alternate scanning techniques are required.
SCOPE
1.1 This guide covers procedures for the contact ultrasonic examination of bulk materials or parts by transmitting pulsed ultrasonic waves into the material and observing the indications of reflected waves. This guide covers only examinations in which one search unit is used as both transmitter and receiver (pulse-echo). This guide includes general requirements and procedures that may be used for detecting discontinuities, locating depth and distance from a point of reference and for making a relative or approximate evaluation of the size of discontinuities as compared to a reference standard.  
1.2 This guide complements Practice E114 by providing more detailed procedures for the selection and standardization of the examination system and for evaluation of the indications obtained.  
1.3 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.  
1.4 This guide does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this guide to establish the appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    7 pages
    English language
  • Guide
    7 pages
    English language

SIGNIFICANCE AND USE
5.1 The procedures described in this practice have proven utility in the inspecting (1) monolithic polymer matrix composites (laminates) for bulk defects, (2) metals for corrosion during the service life of the part of interest, (3) thickness checks, (4) adhesive bonding of metals, composites, and sandwich core constructions, (5) coatings, and (6) composite filament windings. Both unpressurized, and with suitable precautions, pressurized materials and components are inspected.  
5.2 This practice provides guidelines for the application of longitudinal wave examination to the detection and quantitative evaluation of damage, discontinuities, and thickness variations in materials.  
5.3 This practice is intended primarily for the testing of parts to acceptance criteria most typically specified in a purchase order or other contractual document, and for testing of parts in-service to detect and evaluate damage.  
5.4 MAUT search units provide near-surface resolution and detection of small discontinuities comparable to phased array transducers. They may or may not be capable of beam steering. The advantage of MAUT for straight-beam longitudinal wave inspections is the ability to provide real-time C-scan data, which facilitates data interpretation and shortens inspection time. Depending on inspection needs, data can be displayed as A-, B- or C-scans, or three-dimensional renderings. Toggling between pulse-echo and through transmission ultrasonic (TTU) modes without having to use another system or changing transducers is also possible.  
5.5 The MAUT technique has proven utility in the inspection of multi-ply carbon-fiber reinforced laminates used in primary aircraft structures.11  
5.6 For ultrasonic testing of laminate composites and sandwich core materials using conventional UT equipment consult Practice E2580. Consult Practice E114 for ultrasonic testing of materials by the pulse-echo method using straightbeam longitudinal waves introduced by a piezoelectric elemen...
SCOPE
1.1 This practice covers procedures for matrix array ultrasonic testing (MAUT) of monolithic composites, composite sandwich constructions, and metallic test articles. These procedures can be used throughout the life cycle of a part during product and process design optimization, on line process control, post-manufacturing inspection, and in-service inspection.  
1.2 In general, ultrasonic testing is a common volumetric method for detection of embedded or subsurface discontinuities. This practice includes general requirements and procedures which may be used for detecting flaws and for making a relative or approximate evaluation of the size of discontinuities and part anomalies. The types of flaws or discontinuities detected include interply delaminations, foreign object debris (FOD), inclusions, disbond/un-bond, fiber debonding, fiber fracture, porosity, voids, impact damage, thickness variation, and corrosion.  
1.3 Typical test articles include monolithic composite layups such as uniaxial, cross ply and angle ply laminates, sandwich constructions, bonded structures, and filament windings, as well as forged, wrought and cast metallic parts. Two techniques can be considered based on accessibility of the inspection surface: namely, pulse echo inspection for one-sided access and through-transmission for two-sided access. As used in this practice, both require the use of a pulsed straight-beam ultrasonic longitudinal wave followed by observing indications of either the reflected (pulse-echo) or received (through transmission) wave.  
1.4 This practice provides two ultrasonic test procedures. Each has its own merits and requirements for inspection and shall be selected as agreed upon in a contractual document.  
1.4.1 Test Procedure A, Pulse Echo (non-contacting and contacting) is at a minimum a single matrix array transducer transmitting and receiving longitudinal waves in the range of 0.5 MHz to 20 MHz (see Fig. 1). Th...

  • Standard
    14 pages
    English language
  • Standard
    14 pages
    English language

SIGNIFICANCE AND USE
4.1 This practice is intended to be used by the DDA user to measure and record the baseline performance of an acquired DDA in order to monitor its performance throughout its service as an imaging system. This practice is not intended to be used as an “acceptance test” of a DDA.  
4.2 This practice defines the tests to be performed and their required intervals. Also defined are the methods of tabulating results that DDA users will complete following initial baselining of the DDA system. These tests will also be performed periodically at the stated required intervals to evaluate the DDA system to determine if the system remains within acceptable operational limits as established in this practice and defined between the user and CEO.  
4.3 There are several factors that affect the quality of a DDA image including the basic spatial resolution, geometric unsharpness, scatter, signal to noise ratio, contrast sensitivity, contrast/noise ratio, image lag, and for some types of DDAs, burn-in. There are several additional factors and settings which can affect these results (for example, integration time, detector parameters, imaging software, and even X-ray radiation quality). Additionally, detector correction techniques may have an impact on the quality of the image. This practice delineates tests for each of the properties listed herein and establishes standard techniques for assuring repeatability throughout the lifecycle testing of the DDA.
SCOPE
1.1 This practice covers the baseline and periodic performance evaluation of Digital Detector Array (DDA) systems used for industrial radiography. It is intended to ensure that the evaluation of image quality, as far as this is influenced by the DDA system, meets the needs of users, and their customers, and enables process control to monitor long-term stability of the DDA system.  
1.2 This practice specifies the fundamental parameters of DDA systems to be measured to determine baseline performance, and to track the long-term stability of the DDA system.  
1.3 The DDA system tests specified in this practice shall be completed upon acceptance of the system from the manufacturer to baseline the performance of the DDA. Periodic performance testing shall then be used to monitor long-term stability of the system in order to identify when an action needs to be taken due to system degradation beyond a certain defined level.  
1.4 Two types of phantoms, the duplex plate and the five-groove wedge, are used for testing as specified herein. The use of these two types of phantoms is not intended to exclude the use of other phantom configurations. In the event the tests or phantoms specified herein are not sufficient or appropriate, the user, in coordination with the cognizant engineering organization (CEO) may develop additional or modified tests, test objects, phantoms, or image quality indicators to evaluate the DDA system performance. Acceptance levels for these ALTERNATE test methods shall be determined by agreement between the user and CEO.  
1.5 The user of this practice shall consider that higher energies than 450 keV may require different test methods or modifications to the test methods described here. This practice is not intended for usage with isotopes.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    18 pages
    English language
  • Standard
    18 pages
    English language

SIGNIFICANCE AND USE
5.1 The gauge is intended to provide a means for measuring image or detector unsharpness and basic spatial image or detector resolution as independently as practicable from the imaging system and contrast sensitivity limitations. When the duplex plate gauge is positioned directly on the film or the digital detector instead on the test object, then the determined image unsharpness UIm corresponds to the inherent film or detector unsharpness (Udetector) and the determined basic spatial image resolution SRbimage corresponds to the basic spatial detector resolution SRbdetector. SRbimage provides a value which depends on the combined effect of detector unsharpness and geometric unsharpness.  
5.2 Basis of Application:  
5.2.1 The following items are subject to contractual agreement between the parties using or referencing this standard.
5.2.1.1 Personnel Qualification—Personnel performing examinations to this practice shall be qualified in accordance with NAS410, EN 4179, ANSI/ASNT CP 189, ISO 9712, or SNT-TC-1A and certified by the employer or certifying agency as applicable. Other equivalent qualification documents may be used when specified on the contract or purchase order. The applicable revision shall be the latest unless otherwise specified in the contractual agreement between parties.
5.2.1.2 If specified in the contractual agreement, NDT agencies shall be qualified and evaluated as described in Specification E543. The applicable edition of Specification E543 shall be specified in the contract.
SCOPE
1.1 This practice covers the design and basic use of a gauge used to determine the image unsharpness and the basic spatial resolution of film radiographs or of digital images taken with CR imaging plates, digital detector arrays (DDA), or radioscopic systems (for example, with X-ray image intensifiers) for applications with Se-75, Ir-192, Co-60 and high energy sources with tube potentials ≥ 300 kV. The IQI may be used at lower tube potentials too, if the expected unsharpness is > 200 µm (SRbimage or SRbdetector > 100 µm). For the measurement of lower unsharpness values, the usage of the gauge described in Practice E2002 is recommended.  
1.2 This practice is applicable to radiographic and radioscopic imaging systems utilizing X-ray and gamma ray radiation sources.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 The gauge described can be used effectively if the duplex wire IQI of Practice E2002 does not provide sufficient contrast resolution.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
5.1 Personnel that are responsible for the creation, transfer, and storage of eddy current NDE test results will use this standard. This practice defines a set of information modules that, along with Practice E2339 and the DICOM standard, provide a standard means to organize eddy current test parameters and results. The eddy current examination results may be displayed or analyzed on any device that conforms to the standard. Personnel wishing to view any eddy current examination data stored according to Practice E2339 may use this document to help them decode and display the data contained in the DICONDE compliant inspection record.
SCOPE
1.1 This practice covers the interoperability of eddy current imaging and data acquisition equipment by specifying the image data transfer and archival storage in commonly accepted terms. This document is intended to be used in conjunction with Practice E2339 on Digital Imaging and Communication in Nondestructive Evaluation (DICONDE). Practice E2339 defines an industrial adaptation of NEMA PS3 / ISO 12052, an international standard for image data acquisition, review, storage, and archival storage. The goal of Practice E2339, commonly referred to as DICONDE, is to provide a standard that facilitates the display and analysis of NDE results on any system conforming to the DICONDE standard. Toward that end, Practice E2339 provides a data dictionary and a set of information modules that are applicable to all NDE modalities. This practice supplements Practice E2339 by providing information object definitions, information modules, and a data dictionary that are specific to eddy current test methods.  
1.2 This practice has been developed to overcome the issues that arise when analyzing or archiving data from eddy current test equipment using proprietary data transfer and storage methods. As digital technologies evolve, data must remain decipherable through the use of open, industry-wide methods for data transfer and archival storage. This practice defines a method where all the eddy current technique parameters and inspection data are communicated and stored in a standard manner regardless of changes in digital technology.  
1.3 This practice does not specify:  
1.3.1 A testing or validation procedure to assess an implementation's conformance to the standard,  
1.3.2 The implementation details of any features of the standard on a device claiming conformance, or  
1.3.3 The overall set of features and functions to be expected from a system implemented by integrating a group of devices each claiming DICONDE conformance.  
1.4 Units—Although this practice contains no values that require units, it does describe methods to store and communicate data that do require units to be properly interpreted. The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    12 pages
    English language
  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
5.1 Personnel that are responsible for the creation, transfer, and storage of computed radiography NDE data will use this practice. This practice will define a set of information modules that along with the Practice E2339 and the DICOM standard will provide a standard means to organize CR inspection data. The CR inspection data may be displayed and analyzed on any device that conforms to the standard. Personnel wishing to view any CR inspection data stored in accordance with Practice E2339 may use this practice to help them decode and display the data contained in the DICONDE compliant inspection record.
SCOPE
1.1 This practice covers the interoperability of computed radiography (CR) imaging and data acquisition equipment by specifying image data transfer and archival storage methods in commonly accepted terms. This practice is intended to be used in conjunction with Practice E2339 on Digital Imaging and Communication in Nondestructive Evaluation (DICONDE). Practice E2339 defines an industrial adaptation of NEMA PS3 / ISO 12052 (DICOM, see http://medical.nema.org), an international standard for image data acquisition, review, storage, and archival storage. The goal of Practice E2339, commonly referred to as DICONDE, is to provide a standard that facilitates the display and analysis of NDE results on any system conforming to the DICONDE standard. Toward that end, Practice E2339 provides a data dictionary and a set of information modules that are applicable to all NDE modalities. This practice supplements Practice E2339 by providing information object definitions, information modules and a data dictionary that are specific to computed radiography test methods.  
1.2 This practice has been developed to overcome the issues that arise when analyzing or archiving data from CR test equipment using proprietary data transfer and storage methods. As digital technologies evolve, data must remain decipherable through the use of open, industry-wide methods for data transfer and archival storage. This practice defines a method where all standard CR technique parameters and test results are communicated and stored in a standard manner regardless of changes in digital technology.  
1.3 This practice does not specify:  
1.3.1 A testing or validation procedure to assess an implementation's conformance to the standard.  
1.3.2 The implementation details of any features of the standard on a device claiming conformance.  
1.3.3 The overall set of features and functions to be expected from a system implemented by integrating a group of devices each claiming DICONDE conformance.  
1.4 Although this practice contains no values that require units, it does describe methods to store and communicate data that do require units to be properly interpreted. The SI units required by this practice are to be regarded as standard. No other units of measurement are included in this practice.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language
  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 Personnel that are responsible for the creation, transfer, and storage of digital X-ray test results will use this standard. This practice defines a set of information modules that, along with Practice E2339 and the DICOM standard, provides a standard means to organize digital X-ray test parameters and results. The digital X-ray test results may be displayed and analyzed on any device that conforms to this standard. Personnel wishing to view any digital X-ray inspection data stored according to Practice E2339 may use this document to help them decode and display the data contained in the DICONDE-compliant inspection record.
SCOPE
1.1 This practice covers the interoperability of digital X-ray imaging equipment using Digital Detector Arrays (DDA) as described in Practice E2698 by specifying image data transfer and archival methods in commonly accepted terms. A separate practice, Practice E2738, addresses this topic for digital X-ray imaging equipment using Computed Radiography (CR). This document is intended to be used in conjunction with Practice E2339 on Digital Imaging and Communication in Nondestructive Evaluation (DICONDE). Practice E2339 defines an industrial adaptation of DICOM NEMA PS3 / ISO 12052 (DICOM, see http://medical.nema.org), an international standard for image data acquisition, review, storage, and archival storage. The goal of Practice E2339, commonly referred to as DICONDE, is to provide a standard that facilitates the display and analysis of NDE results on any system conforming to the DICONDE standard. Toward that end, Practice E2339 provides a data dictionary and a set of information modules that are applicable to all NDE modalities. This practice supplements Practice E2339 by providing information object definitions, information modules, and a data dictionary that are specific to digital X-ray test methods.  
1.2 This practice has been developed to overcome the issues that arise when analyzing or archiving data from digital X-ray test equipment using proprietary data transfer and storage methods. As digital technologies evolve, data must remain decipherable through the use of open, industry-wide methods for data transfer and archival storage. This practice defines a method where all the digital X-ray technique parameters and test results are communicated and stored in a standard manner regardless of changes in digital technology.  
1.3 This practice does not specify:  
1.3.1 A testing or validation procedure to assess an implementation's conformance to the standard.  
1.3.2 The implementation details of any features of the standard on a device claiming conformance.  
1.3.3 The overall set of features and functions to be expected from a system implemented by integrating a group of devices each claiming DICONDE conformance.  
1.4 Units—Although this practice contains no values that require units, it does describe methods to store and communicate data that do require units to be properly interpreted. The SI units required by this practice are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 Personnel that are responsible for the creation, transfer, and storage of ultrasonic test results will use this standard. This practice defines a set of information modules that, along with Practice E2339 and the DICOM standard, provides a standard means to organize ultrasonic test parameters and results. The ultrasonic test results may be displayed and analyzed on any device that conforms to this standard. Personnel wishing to view any ultrasonic inspection data stored in DICONDE format may use this document to help them decode and display the data contained in the DICONDE compliant inspection record.
SCOPE
1.1 This practice covers the interoperability of ultrasonic imaging equipment by specifying image data transfer and archival storage methods in commonly accepted terms. This document is intended to be used in conjunction with Practice E2339. Practice E2339 defines an industrial adaptation of NEMA PS3 / ISO 12052 (DICOM, see http://medical.nema.org), an international standard for image data acquisition, review, transfer, and archival storage. The goal of Practice E2339, commonly referred to as DICONDE, is to provide a standard that facilitates the display and analysis of NDE test results on any system conforming to the DICONDE standard. Toward that end, Practice E2339 provides a data dictionary and set of information modules that are applicable to all NDE modalities. This practice supplements Practice E2339 by providing information object definitions, information modules, and data dictionary that are specific to ultrasonic test methods.  
1.2 This practice has been developed to overcome the issues that arise when analyzing or archiving data from ultrasonic test equipment using proprietary data transfer and storage methods. As digital technologies evolve, data must remain decipherable through the use of open, industry-wide methods for data transfer and archival storage. This practice defines a method where all the ultrasonic technique parameters and test results are communicated and stored in a standard format regardless of changes in digital technology.  
1.3 This practice does not specify:  
1.3.1 A testing or validation procedure to assess an implementation's conformance to the standard.  
1.3.2 The implementation details of any features of the standard on a device claiming conformance.  
1.3.3 The overall set of features and functions to be expected from a system implemented by integrating a group of devices each claiming DICONDE conformance.  
1.4 Although this practice contains no values that require units, it does describe methods to store and communicate data that do require units to be properly interpreted. The SI units required by this practice are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    9 pages
    English language
  • Standard
    9 pages
    English language

SIGNIFICANCE AND USE
4.1 There are several factors affecting the quality of a CR image including the basic spatial resolution of the IP system, geometrical unsharpness, scatter and contrast sensitivity. There are several additional factors (for example, software and scanning parameters) that affect the accurate reading of images on exposed IPs using an optical scanner.  
4.2 This practice is to be used to establish a characterization of CR system by performance levels on the basis of a normalized SNR, interpolated basic spatial detector resolution and EPS. The CR system performance levels in this practice do not refer to any particular manufacturers’ imaging plates. A CR system performance level results from the use of a particular imaging plate together with the exposure conditions, standardized phantom, the scanner type, and software and the scanning parameters. This characterization system provides a means to compare differing CR technologies, as is common practice with film systems, which guides the user to the appropriate configuration, IP, and technique for the application at hand. The performance level selected may not match the imaging performance of a corresponding film class because of the difference in the spatial resolution and scatter sensitivity. Therefore, the user should always use IQIs for proof of contrast sensitivity and basic spatial resolution.  
4.3 The measured performance parameters are presented in a characterization chart. This enables users to select specific CR systems by the different characterization data to find the best system for his specific application.  
4.4 The quality factors can be determined most accurately by the tests described in this practice. Some of the system tests require special tools, which may not be available in user laboratories. Simpler tests are described for quality assurance and long term stability tests in Practice E2445.  
4.5 Manufacturers of industrial CR systems or certification agencies will use this practice. Users of i...
SCOPE
1.1 This practice covers the manufacturing characterization of computed radiography (CR) systems, consisting of a particular phosphor imaging plate (IP), scanner, software, scanner operational parameters, and an image display monitor, in combination with specified metal screens for industrial radiography.  
1.2 The practice defines system tests to be used to characterize the systems of different suppliers and make them comparable for users.  
1.3 This practice is intended for use by manufacturers of CR systems or certification agencies to provide quantitative results of CR system characteristics for nondestructive testing (NDT) user or purchaser consumption. Some of these tests require specialized test phantoms to ensure consistency of results among suppliers or manufacturers. These tests are not intended for users to complete, nor are they intended for long term stability tracking and lifetime measurements. However, they may be used for this purpose, if so desired. Practice E2445 describes tests which are intended for users to observe the CR performance and test the long term stability.  
1.4 The CR system performance is described by the basic spatial resolution, contrast, signal and noise parameters, and the equivalent penetrameter sensitivity (EPS). Some of these parameters are used to compare with DDA characterization and film characterization data (see Practice E2597 and Test Method E1815).
Note 1: For film system characterization, the signal is represented by the optical density of 2 (above fog and base) and the noise as granularity. The signal-to-noise ratio is normalized by the aperture (similar to the basic spatial resolution) of the system and is part of characterization. This normalization is given by the scanning circular aperture of 100 µm of the micro-photometer, which is defined in Test Method E1815 for film system characterization.  
1.5 The measurement of CR systems in this practice is restricted ...

  • Standard
    18 pages
    English language
  • Standard
    18 pages
    English language

SIGNIFICANCE AND USE
5.1 This guide is intended to evaluate performance assessment of combinations of phased-array probes and instruments. It is not intended to define performance and acceptance criteria, but rather to provide data from which such criteria may be established.  
5.2 Recommended procedures described in this guide are intended to provide performance-related measurements that can be reproduced under the specified test conditions using simple targets and the phased-array test system itself. It is intended for phased-array flaw detection instruments operating in the nominal frequency range of 1 MHz to 20 MHz, but the procedures are applicable to measurements on instruments utilizing significantly higher frequency components.  
5.3 This guide is not intended for service calibration, or maintenance of circuitry for which the manufacturer’s instructions are available.  
5.4 Implementation of specific assessments may require more detailed procedural instructions in a format of the using facility.  
5.5 The measurement data obtained may be employed by users of this guide to specify, describe, or provide performance criteria for procurement and quality assurance, or service evaluation of the operating characteristics of phased-array systems.  
5.6 Not all assessments described in this guide are applicable to all systems. All or portions of the guide may be used as determined by the user.
SCOPE
1.1 This guide covers procedures for evaluating some performance characteristics of phased-array ultrasonic examination instruments and systems.  
1.2 Evaluation of these characteristics is intended to be used for either comparing instruments and systems or, by periodic repetition, for detecting long-term changes in the characteristics of a given instrument or system. Significant changes may be indicative of impending failure, and, if beyond certain limits, will require corrective maintenance. Some electronic instrument characteristics in phased-array units are similar to non-phased-array units and may be measured as described in Practice E1065 or Guide E1324.  
1.3 Ultrasonic examination systems using pulsed-wave trains and A-scan presentation (rf or video) may be evaluated.  
1.4 This guide establishes no performance limits for examination systems; if such acceptance criteria are required, these shall be specified by the using parties. Where acceptance criteria are implied herein, they are for example only and are subject to more or less restrictive limits imposed by customer’s and end user’s controlling documents.  
1.5 The specific parameters to be evaluated, conditions, frequency of test, and report data required shall be determined by the user.  
1.6 This guide may be used for the evaluation of a complete examination system, including search unit, instrument, interconnections, scanner fixtures, connected alarm, and auxiliary devices. This guide is not intended to be used as a substitute for calibration or standardization of an instrument or system to inspect any given material.  
1.7 Required test apparatus includes selected test blocks and position encoders in addition to the instrument or system to be evaluated.  
1.8 Alternate procedures, such as examples described in this document, or others, may only be used with customer approval.  
1.9 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.10 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.11 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendatio...

  • Guide
    17 pages
    English language
  • Guide
    17 pages
    English language

SIGNIFICANCE AND USE
3.1 The primary use of this guide is to provide a standardized approach for the data file to be used for the transfer of digital radiological data from one user to another where the two users are working with dissimilar systems. This guide describes the contents, both required and optional for an intermediate data file that can be created from the native format of the radiological system on which the data was collected and that can be converted into the native format of the receiving radiological data analysis system. This guide will also be useful in the archival storage and retrieval of radiological data as either a data format specifier or as a guide to the data elements which should be included in the archival file.  
3.2 Although the recommended field listing includes more than 100 field numbers, only about half of those are regarded as essential and are marked Footnote C in Table 1. Fields so marked must be included in the data base. The other fields recommended provide additional information that a user will find helpful in understanding the radiological image and examination result. These header field items will, in most cases, make up only a very small part of a radiological examination file. The actual stream of radiological data that make up the image will take up the largest part of the data base. Since a radiological image file will normally be large, the concept of data compression will be considered in many cases. Compressed data should be noted, along with a description of the compression method, as indicated in Field No. 92 (see Table 1). (A) Field numbers are for reference only. They do not imply a necessity to include all these fields in any specific data base nor imply a requirement that fields used be in this particular order.(B) Units listed first are SI; those in parentheses are inch-pound (English).(C) Denotes essential field for computerization of examination results, regardless of examination method.(D) Denotes essential field for radiogra...
SCOPE
1.1 This guide provides a listing and description of the fields that are recommended for inclusion in a digital radiological examination data base to facilitate the transfer of such data. This guide sets guidelines for the format of data fields for computerized transfer of digital image files obtained from radiographic, radioscopic, computed radiographic, or other radiological examination systems. The field listing includes those fields regarded as necessary for inclusion in the data base: (1) regardless of the radiological examination method (as indicated by Footnote C in Table 1), (2) for radioscopic examination (as indicated by Footnote F in Table 1), and (3) for radiographic examination (as indicated by Footnote D in Table 1). In addition, other optional fields are listed as a reminder of the types of information that may be useful for additional understanding of the data or applicable to a limited number of applications.  
1.2 It is recognized that organizations may have in place an internal format for the storage and retrieval of radiological examination data. This guide should not impede the use of such formats since it is probable that the necessary fields are already included in such internal data bases, or that the few additions can easily be made. The numerical listing and its order indicated in this guide is only for convenience; the specific numbers and their order carry no inherent significance and are not part of the data file.  
1.3 Current users of Guide E1475 do not have to change their software. First time users should use the XML structure of Table A1.1 for their data.  
1.4 The types of radiological examination systems that appear useful in relation to this guide include radioscopic systems as described in Guide E1000, Practices E1255, E1411, E2597, E2698 and E2737, and radiographic systems as described in Guide E94 and Practices E748, E1742, E2033, E2445, and E2446. Many of the terms used are def...

  • Guide
    7 pages
    English language

SIGNIFICANCE AND USE
3.1 The provisions of this guide are intended to control the quality of industrial radiographs and unexposed films only and are not intended for controlling the acceptability of the materials or products radiographed. It is further intended that this guide be used as an adjunct to Guide E94.  
3.2 The necessity for applying specific control procedures such as those described in this guide is dependent to a certain extent, on the degree to which a user adheres to good processing and storage practices as a matter of routine procedure.
SCOPE
1.1 This guide may be used for the control and maintenance of industrial radiographs and unexposed films used for industrial radiography.  
1.2 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.
Note 1: For information purposes, refer to Terminology E1316. The terms stated therein, however, are not specifically referenced in the text of this document.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    3 pages
    English language

SIGNIFICANCE AND USE
4.1 The requirements in this practice are intended to control the quality of the radioscopic images to produce satisfactory and consistent results. This practice is not intended for controlling the acceptability of the casting. The radioscopic method may be used for detecting volumetric discontinuities and density variations that are within the sensitivity range of this practice. The dynamic aspects of radioscopy are useful for maximizing defect response.
SCOPE
1.1 This practice covers a uniform procedure for radioscopic examination of castings. Radioscopic examination of weldments can be found in E1416.  
1.2 This practice applies only to radioscopic examination in which an image is finally presented on a display screen (monitor) for evaluation. Test part acceptance may be based on a static or dynamic image. The examination results may be recorded for later review. This practice does not apply to fully automated systems in which evaluation is performed automatically by a computer.  
1.3 Due to the many complex geometries and part configurations inherent with castings, it is necessary to recognize the potential limitations associated with obtaining complete radioscopic coverage. Consideration shall be given to areas where geometry or part configuration does not allow for complete radioscopic coverage.  
1.4 The values stated in inch-pound units are to be regarded as the standard. The SI units given in parentheses are for information only.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language
  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
4.1 This test method provides a relative means for classification of film systems used for industrial radiography. The film system consists of the film and associated processing system (the type of processing and processing chemistry). Section 9 describes specific parameters used for this test method. In general, the classification for hard X-rays, as described in Section 9, can be transferred to other radiation energies and metallic screen types, as well as screens without films. The usage of film system parameters outside the energy ranges specified may result in changes to a film/system performance classification.  
4.1.1 The film performance is described by contrast and noise parameters. The contrast is represented by gradient and the noise by granularity.  
4.1.2 A film system is assigned a particular class if it meets the minimum performance parameters: for Gradient G at  D – D0 = 2.0 and D – D0 = 4.0, and gradient/noise ratio at D – D0 = 2.0, and the maximum performance parameter: granularity σD at  D = 2.0.  
4.2 This test method describes how the parameters shall be measured and demonstrates how a classification table can be constructed.  
4.3 Manufacturers of industrial radiographic film systems and developer chemistry will be the users of this test method. The result is a classification table as shown by the example given in Table 2. Another table also includes speed data for user information. Users of industrial radiographic film systems may also perform the tests and measurements outlined in this test method, provided that the required test equipment is used and the methodology is followed strictly. (A) Family of films ranging in speed and image quality.  
4.4 The publication of classes for industrial radiography film systems will enable specifying bodies and contracting parties to agree to particular system classes, which are capable of providing known image qualities. See 8.2.  
4.5 ISO 11699–1 and European standard EN 584-1 describe the same m...
SCOPE
1.1 This test method covers a procedure for determination of the performance of film systems used for industrial radiography. This test method establishes minimum requirements that correspond to system classes.  
1.2 This test method is to be used only for direct exposure-type film exposed with lead intensifying screens. The performance of films exposed with fluorescent (light-emitting) intensifying screens cannot be determined accurately by this test method.  
1.3 The values stated in SI units are to be regarded as standard. The values given in parentheses are for information only.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
4.1 This guide is intended as a source of tutorial and reference information that can be used during establishment of computed radiography techniques and procedures by qualified CR personnel for specific applications. All materials presented within this guide may not be suited for all levels of computed radiographic personnel.  
4.2 This guide is intended to build upon an established basic knowledge of radiographic fundamentals (that is, film systems) as may be found in Guide E94. Similarly, materials presented within this guide are not intended as “all-inclusive” but are intended to address basic CR topics and issues that complement a general knowledge of computed radiography as described in 1.2 and 3.2.28.  
4.3 Materials presented within this guide may be useful in the development of end-user training programs designed by qualified CR personnel or activities that perform similar functions. Computed radiography is considered a rapidly advancing inspection technology that will require the user maintain knowledge of the latest CR apparatus and technique innovations. Section 11 of this guide contains technical reference materials that may be useful in further advancement of knowledge associated with computed radiography.
SCOPE
1.1 This guide provides general tutorial information regarding the fundamental and physical principles of computed radiography (CR), definitions and terminology required to understand the basic CR process. An introduction to some of the limitations that are typically encountered during the establishment of techniques and basic image processing methods are also provided. This guide does not provide specific techniques or acceptance criteria for specific end-user inspection applications. Information presented within this guide may be useful in conjunction with those standards of 1.2.  
1.2 CR techniques for general inspection applications may be found in Practice E2033. Technical qualification attributes for CR systems may be found in Practice E2445. Criteria for classification of CR system technical performance levels may be found in Practice E2446. Reference Images Standards E2422, E2660, and E2669 contain digital reference acceptance illustrations.  
1.3 The values stated in SI units are to be regarded as the standard. The inch-pound units given in parentheses are for information only.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    23 pages
    English language

SIGNIFICANCE AND USE
5.1 The purpose of the alternating current field measurement method is to evaluate threads for surface breaking discontinuities such as fatigue cracks running along the thread root. The examination results may then be used to determine the fate of the test piece. This may involve re-examination by an alternative technique, immediate scrapping of the test piece, or reworking to remove discontinuities (beyond the scope of this practice). This practice is not intended for the examination of threads for non-surface breaking discontinuities.
SCOPE
1.1 This practice describes procedures to be followed during alternating current field measurement examination of drillstring threads on tubulars used for oil and gas exploration and production for detection and, if required, sizing of service-induced surface breaking discontinuities transverse to the pipe.  
1.2 This practice is intended for use on threads in any metallic material.  
1.3 This practice does not establish acceptance criteria. Typical industry practice is to reject these connections on detection of a confirmed crack.  
1.4 While the alternating current field measurement technique is capable of detecting discontinuities in these connections, supplemental surface NDT methods such as magnetic particle testing for ferrous metals and penetrant testing for non-ferrous metals may detect additional discontinuities.  
1.5 Units—The values stated in either inch-pound units or SI units are to be regarded separately as standard. The values stated in each system might not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from both systems may result in nonconformance with the standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    10 pages
    English language
  • Standard
    10 pages
    English language

SIGNIFICANCE AND USE
4.1 The purpose of this practice is to establish instruction, procedures, and equipment qualification to inspect a component for specified discontinuities. This practice is written so it can be specified on the engineering drawing, specification, or contract. This practice requires a detailed procedure delineating the technique or procedure requirements and shall be approved by the Cognizant Engineering Organization (CEO).  
4.2 The requirements in this practice shall be used when placing a CT system into NDT service and establishing a baseline of system performance measures. Monitoring the system performance over time shall be performed, including detector correction procedures, performance measurements, and system maintenance.  
4.3 For CT examinations where specified discontinuities are not known/identified/requested/available (that is, engineering analysis, engineering information only), portions of Section 7 – 9 requirements do not necessarily apply. For these types of examinations, the purchase order or contract shall specify the development of a specific procedure, approved by purchaser and supplier, to specify the requirements of this practice to be incorporated into the examination.
SCOPE
1.1 This practice establishes the minimum requirements for computed tomography (CT) examination of components using cone beam systems. The purpose of this practice is to establish instruction, procedures, and equipment qualification to inspect a component for specified discontinuities.  
1.2 This practice applies to systems with a Digital Detector Array (DDA) and an X-ray source. It does not cover the use of isotope sources.  
1.3 There are sections in this practice that may require agreement between the purchaser and the supplier, or specific direction from the cognizant engineering organization (CEO). These items should be addressed in the purchase order or the contract. Generally, the items are application-specific, performance related, or both.  
1.4 Applications for cone-beam CT are diverse. This practice is not intended to be limiting or restrictive. Refer to Guide E1441 for guidance on CT fundamentals and Guide E1672 for tradeoffs in selection of a CT system.  
1.5 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. NCRP 144 or NIST Handbook 114, or both, may be used as guides to ensure that radiographic procedures are performed so that personnel shall not receive a radiation dosage exceeding the maximum permitted by the city, state, or national codes.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    7 pages
    English language

ABSTRACT
This test method covers a uniform procedure for radioscopic examination of weldments. Requirements expressed in this test method are intended to control the quality of the radioscopic images and are not intended for controlling acceptability or quality of welds. It applies only to the use of equipment for radioscopic examination in which the image is finally presented on a television monitor for operator evaluation. The examination may be recorded for later review. It does not apply to fully automated systems where evaluation is automatically performed by computer. Unless otherwise specified by the applicable job order or contract, radioscopic examination shall be performed in accordance with a written procedure which includes: material and thickness range to be examined; equipment to be used, including specifications of source parameters and imaging equipment parameters; examination geometry, including source-to-object distance, object-to-detector-distance and orientation; image quality indicator designation and placement; test-object scan plan, indicating the range of motions and manipulation speeds through which the test object shall be manipulated in order to ensure satisfactory results; image-processing parameters; image-display parameters; and image storage.
SCOPE
1.1 This practice covers a uniform procedure for radioscopic examination of weldments. Requirements expressed in this practice are intended to control the quality of the radioscopic images and are not intended for controlling acceptability or quality of welds.  
1.2 This practice applies only to the use of equipment for radioscopic examination in which the image is finally presented on a display screen (monitor) for operator evaluation. The examination may be recorded for later review. It does not apply to fully automated systems where evaluation is automatically performed by computer.  
1.3 The radioscopic extent, the quality level, and the acceptance criteria to be applied shall be specified in the contract, purchase order, product specification, or drawings.  
1.4 This practice can be used for the detection of discontinuities. This practice also facilitates the examination of a weld from several directions, such as perpendicular to the weld surface and along both weld bevel angles. The radioscopic techniques described in this practice provide adequate assurance for defect detectability; however, it is recognized that, for special applications, specific techniques using more stringent requirements may be needed to provide additional detection capability. The use of specific radioscopic techniques shall be agreed upon between purchaser and supplier.  
1.5 Units—The values stated in inch-pound units are to be regarded as the standard. The SI units given in parentheses are for information only.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 7.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 UV-A and visible light sources are used to provide adequate illumination levels for liquid penetrant and magnetic particle examination. UV-A sources, UV-A radiometers, visible light sources, and illuminance photometers are used to verify specified viewing conditions.  
5.2 Fluorescence is typically produced by irradiating the fluorescent dyes/pigments with UV-A radiation. The fluorescent dyes/pigments absorb the UV-A radiation and re-emit light energy in the visible spectrum. This process allows fluorescence to be observed by the human eye.  
5.3 UV-A sources may emit visible light above 400 nm (4000 Å), which may reduce the visibility of fluorescent indications. High intensity UV-A sources may cause UV fade, causing fluorescent indications to degrade or disappear.
SCOPE
1.1 This guide addresses the use of UV-A/Visible light sources and meters used for the examination of materials by the liquid penetrant and magnetic particle processes. This guide may be used to establish practices and procedures to measure irradiance and illuminance levels.  
1.2 This guide also acts as a reference:  
1.2.1 To assist in the selection of irradiance and illumination sources and meters that meet the applicable specifications or standards.  
1.2.2 For use in the preparation of internal documentation dealing with liquid penetrant or magnetic particle examination of materials and parts.  
1.3 Units—The values stated in SI units are to be regarded as standard. The values given in parentheses are mathematical conversions to inch-pound units that are provided for information only and are not considered standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    6 pages
    English language
  • Guide
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 Real-time detection and assessment of cracks and other flaws in concrete structures is of great importance. A number of methods have been developed and standardized in recent decades for non-destructive evaluation of concrete structures as well as methods for in-place evaluation of concrete properties. Review of some of these methods can be found in ACI 228.2R-13, ACI 228.1R-03, and ACI 437R-03. They include visual inspection, stress-wave methods such as impact echo, pulse velocity, impulse response, nuclear methods, active and passive infrared thermography, ground-penetrating radar and others. These methods in most of the cases are not used for overall inspection of the concrete structure due to limited accessibility, significant thickness of concrete components, or other reasons and are not applied for continuous long-term monitoring. Further, these methods cannot be utilized for estimation of flaw propagation rate or evaluation of flaw sensitivity to operational level loads or environmental changes, or both.  
5.2 In addition to the previously mentioned non-destructive tests methods, vibration, displacement, tilt, shock, strain monitoring, and other methods have been applied to monitor, periodically or continuously, various factors that can affect the integrity of concrete structures during operation. However, these methods monitor risk factors that are not necessarily associated with actual damage accumulation in the monitored structures.  
5.3 Monitoring the opening or elongation of existing cracks can be performed as well using different technologies. These may include moving scales (Fig. 1), vibrating wire, draw wire, or other crack opening displacement meters, optical and digital microscopes, strain gages, or visual assessment. However, this type of monitoring is only applicable to surface cracks and requires long monitoring periods.
FIG. 1 Moving Scale Crack Opening Monitor  
5.4 This guide is meant to be used for development of acoustic emission a...
SCOPE
1.1 This guide describes the application of acoustic emission (AE) technology for examination of concrete and reinforced concrete structures during or after construction, or in service.  
1.2 Structures under consideration include but are not limited to buildings, bridges, hydraulic structures, tunnels, decks, pre/post-tensioned (PT) structures, piers, nuclear containment units, storage tanks, and associated structural elements.  
1.3 AE examinations may be conducted periodically (short-term) or monitored continuously (long-term), under normal service conditions or under specially designed loading procedures. Examples of typical examinations are the detection of growing cracks in structures or their elements under normal service conditions or during controlled load testing, long term monitoring of pre-stressed cables, and establishing safe operational loads.  
1.4 AE examination results are achieved through detection, location, and characterization of active AE sources within concrete and reinforced concrete. Such sources include micro- and macro-crack development in concrete due to loading scenarios such as fatigue, overload, settlement, impact, seismicity, fire and explosion, and also environmental effects such as temperature gradients and internal or external chemical attack (such as sulfate attack and alkali-silica reaction) or radiation. Other AE source mechanisms include corrosion of rebar or other metal parts, corrosion and rupture of cables in pre-stressed concrete, as well as friction due to structural movement or instability, or both.  
1.5 This guide discusses selection of the AE apparatus, setup, system performance verification, detection and processing of concrete damage related AE activity. The guide also provides approaches that may be used in analysis and interpretation of acoustic emission data, assessment of examination results and establishing accept/reject criteria.  
1.6 Units—The values stated ...

  • Guide
    8 pages
    English language
  • Guide
    8 pages
    English language

SIGNIFICANCE AND USE
5.1 The damage-based design approach will permit an additional method of design for GFRP materials. This is a very useful technique to determine the performance of different types of resins and composition of GFRP materials in order to develop a damage tolerant and reliable design. This AE-based method is not unique, other damage-sensitive evaluation methods can also be used.  
5.2 This practice involves the use of acoustic emission instrumentation and examination techniques as a means of damage detection to support a destructive test, in order to derive the damage-based design stress.  
5.3 This practice is not intended as a definitive predictor of long-term performance of GFRP materials (such as those used in vessels). For this reason, codes and standards require cyclic proof testing of prototypes (for example, vessels) which are not a part of this practice.  
5.4 Other design methods exist and are permitted.
SCOPE
1.1 This practice details procedures for establishing the direct stress and shear stress damage-based design values for use in the damage-based design criterion for materials to be used in GFRP vessels and other GFRP structures. The practice uses data derived from acoustic emission examination of four-point beam bending tests and in-plane shear tests (see ASME Section X, Article RT-8).  
1.2 The onset of lamina damage is indicated by the presence of significant acoustic emission during the reload portion of load/reload cycles. “Significant emission” is defined with historic index.  
1.3 Units—The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units which are provided for information only and are not considered standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 APR technology is used for detection, location and identification of internal diameter (ID) flaws-indications and blockages in tube bundles.  
5.2 Reliable and accurate examination of tube bundles is of great importance in different industries. On-time detection of flaws reduces a risk of catastrophic failure and minimizes unplanned shutdowns of plant equipment. Fast examination capability is of great importance due to reduction of maintenance time.  
5.3 APR examinations are performed for quality control of newly manufactured tube bundles as well as for in-service inspection.  
5.4 Performing an APR examination requires access to an open end of each tube to be examined.  
5.5 Flaws that can be readily detected and identified include but are not limited to through-wall holes, ID pitting, erosion, blockages, bulging due to creep and plastic deformation due to bending.  
5.6 APR can be applied to tube bundles made of metal, graphite, plastic or other solid materials with straight and curved sections. The APR technology has been found effective on tubes with diameters between 12.7 mm [1/2 in.] to 101.6 mm [4 in.] and lengths up to 18 metres [60 feet].  
5.7 Closed cracks on ID surface, without significant geometrical alternation on ID surface, may not be detected by APR.  
5.8 APR technology can be used for flaw sizing when special signal and data analysis methods are developed and applied.  
5.9 In addition to detection of flaws and blockages, APR technology can be applied for assessing tube ID surface cleanliness, providing valuable information for equipment maintenance and improving its performance.  
5.10 Other nondestructive test methods may be used to verify and evaluate the significance of APR indications, their exact position, depth, dimension and orientation. These include remote visual inspection, eddy current and ultrasonic testing.  
5.11 Procedures for using other NDT methods are beyond the scope of this practice.  
5.12 Acceptable flaw size...
SCOPE
1.1 This practice describes use of Acoustic Pulse Reflectometry (APR) technology for examination of the internal surface of typical tube bundles found in heat exchangers, boilers, tubular air heaters and reactors, during shutdown periods.  
1.2 The purpose of APR examination is to detect, locate and identify flaws such as through-wall holes, ID wall loss due to pitting and/or erosion as well as full or partial tube blockages. APR may not be effective in detecting cracks with tight boundaries.  
1.3 APR technology utilizes generation of sound waves through the air in the examined tube, then detecting reflections created by discontinuities and/or blockages. Analysis of the initial phase (positive or negative) and the shape of the reflected acoustic wave are used to identify the type of flaw causing the reflection.  
1.4 When proper methods of signal and data analysis are developed, APR technology can be applied for sizing of flaw/blockage indications.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standards.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 The emissivity of a specimen can cause surface temperature measurement errors. Two procedures are provided for measuring and compensating for this error source.  
5.2 These procedures can be used in the field or laboratory, using commonly available materials.  
5.3 These procedures can be used with any infrared radiometers that have the required computer capabilities.  
5.4 The values of emissivity are defined only in terms of the procedure for the purpose of process control and nondestructive evaluation of materials.
SCOPE
1.1 This practice covers procedures for measuring and compensating for emissivity when measuring the surface temperature of a specimen with an infrared imaging radiometer.2  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 These procedures may involve use of equipment and materials in the presence of heated or electrically-energized equipment, or both.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 The purpose of this practice is to outline a procedure for using GWT to locate areas in metal pipes in which wall loss has occurred due to corrosion or erosion.  
5.2 GWT does not provide a direct measurement of wall thickness, but is sensitive to a combination of the CSC (or reflection coefficient) and circumferential extent and axial extent of any metal loss. Based on this information, a classification of the severity can be assigned.  
5.3 The GWT method provides a screening tool to quickly identify any discontinuity along the pipe. Where a possible defect is found, a follow-up inspection of suspected areas with ultrasonic testing or other NDT methods is normally required to obtain detailed thickness information, nature, and extent of damage.  
5.4 GWT also provides some information on the axial length of a discontinuity, provided that the axial length is longer than roughly a quarter of the wavelength.  
5.5 The identification and severity assessment of any possible defects is qualitative only. An interpretation process to differentiate between relevant and non-relevant signals is necessary.  
5.6 This practice only covers the application specified in the scope. The GWT method has the capability and can be used for applications where the pipe is insulated, buried, in road crossings, and where access is limited.  
5.7 GWT shall be performed by qualified and certified personnel, as specified in the contract or purchase order. Qualifications shall include training specific to the use of the equipment employed, interpretation of the test results, and guided wave technology.  
5.8 A documented program which includes training, examination, and experience for the GWT personnel certification shall be maintained by the supplying party.
SCOPE
1.1 This practice provides a guide for the use of waves generated using magnetostrictive transduction for guided wave testing (GWT) welded tubulars. Magnetostrictive materials transduce or convert time varying magnetic fields into mechanical energy. As a magnetostrictive material is magnetized, it strains. Conversely, if an external force produces a strain in a magnetostrictive material, the material’s magnetic state will change. This bi-directional coupling between the magnetic and mechanical states of a magnetostrictive material provides a transduction capability that can be used for both actuation and sensing devices.  
1.2 GWT utilizes ultrasonic guided waves in the 10 to approximately 250 kHz range, sent in the axial direction of the pipe, to non-destructively test pipes for discontinuities or other features by detecting changes in the cross-section or stiffness of the pipe, or both.  
1.3 GWT is a screening tool. The method does not provide a direct measurement of wall thickness or the exact dimensions of discontinuities. However, an estimate of the severity of the discontinuity can be obtained.  
1.4 This practice is intended for use with tubular carbon steel products having nominal pipe size (NPS) 2 to 48 corresponding to 60.3 to 1219.2 mm (2.375 to 48 in.) outer diameter, and wall thickness between 3.81 and 25.4 mm (0.15 and 1 in.).  
1.5 This practice only applies to GWT of basic pipe configuration. This includes pipes that are straight, constructed of a single pipe size and schedules, fully accessible at the test location, jointed by girth welds, supported by simple contact supports and free of internal, or external coatings, or both; the pipe may be insulated or painted.  
1.6 This practice provides a general practice for performing the examination. The interpretation of the guided wave data obtained is complex and training is required to properly perform data interpretation.  
1.7 This practice does not establish an acceptance criterion. Specific acceptance criteria shall be specified in the contractual agreement by the cognizant engineer.  
1.8 Units—The values stated in SI units are to be regarded as standard. The values given...

  • Standard
    11 pages
    English language

SIGNIFICANCE AND USE
5.1 The practices referenced in this document are applicable to measuring the height of planar flaws open to the surface that originate on the far-surface or near-surface of the component. These practices are applicable to through-wall sizing of mechanical or thermal fatigue flaws, stress corrosion flaws, or any other surface-connected planar flaws.  
5.2 The techniques outlined describe proven ultrasonic flaw sizing practices and their associated limitations, using refracted longitudinal wave and shear wave techniques as applied to ferritic or austenitic components. Other materials may be examined using this guide with appropriate standardization reference blocks. The practices described are applicable to both manual and automated examinations.  
5.3 The techniques recommended in this standard guide use Time of Flight (TOF) or Delta Time of Flight (ΔTOF) methods to accurately measure the flaw size. This guide does not include the use of signal amplitude methods to determine flaw size.  
5.4 Generally, with these sizing methods the volume of material (or component thickness) to be sized is divided into thirds; the inner 1/3 , the middle 1/3 and the near 1/3. Using the far-surface Creeping Wave Method the user can qualitatively segregate the flaw into the approximate 1/3 zone.  
5.5 The sizing methods are used in 1/3 zones to quantitatively size the crack, that is, Tip-diffraction for the far 1/3 , Bi-Modal method for the middle 1/3 , and the Focused Longitudinal Wave or Focused Shear Wave Methods for the near 1/3 . These 1/3 zones are generally applicable to most sizing applications, however, the various sizing methods have applications outside these 1/3 zones provided a proper reference block and technique is demonstrated.
SCOPE
1.1 This guide provides tutorial information and a description of the principles and ultrasonic examination techniques for measuring the height of planar flaws which are open to the surface. The practices and technology described in this standard guide are intended as a reference to be used when selecting a specific ultrasonic flaw sizing technique as well as establishing a means for instrument standardization.2  
1.2 This standard guide does not provide or suggest accuracy or tolerances of the techniques described. Parameters such as search units, examination surface conditions, material composition, etc. can all have a bearing on the accuracy of results. It is recommended that users assess accuracy and tolerances applicable for each application.  
1.3 This guide does not purport to provide instruction to measure flaw length.  
1.4 This standard guide does not provide, suggest, or specify acceptance standards. After flaw-sizing evaluation has been made, the results should be applied to an appropriate code or standard that specifies acceptance criteria.  
1.5 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory requirements prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    23 pages
    English language

SIGNIFICANCE AND USE
4.1 A major concern of metals producers, warehouses, and users is to establish and maintain the identity of metals from melting to their final application. This involves the use of standard quality assurance practices and procedures throughout the various stages of manufacturing and processing, at warehouses and materials receiving, and during fabrication and final installation of the product. These practices typically involve standard chemical analyses and physical tests to meet product acceptance standards, which are slow. Several pieces from a production run are usually destroyed or rendered unusable through mechanical and chemical testing, and the results are used to assess the entire lot using statistical methods. Statistical quality assurance methods are usually effective; however, mixed grades, off-chemistry, and nonstandard physical properties remain the primary causes for claims in the metals industry. A more comprehensive verification of product properties is necessary. Nondestructive means are available to supplement conventional metals grade verification techniques, and to monitor chemical and physical properties at selected production stages, in order to assist in maintaining the identities of metals and their consistency in mechanical properties.  
4.2 Nondestructive methods have the potential for monitoring grade during production on a continuous or statistical basis, for monitoring properties such as hardness and case depth, and for verifying the effectiveness of heat treatment, cold-working, and the like. They are quite often used in the field for solving problems involving off-grade and mixed-grade materials.  
4.3 The nondestructive methods covered in this guide provide both direct and indirect responses to the sample being evaluated. Spectrometric analysis instruments respond to the presence and percents of alloying constituents. The electromagnetic (eddy current) and thermoelectric methods, on the other hand, are among those that respond to pr...
SCOPE
1.1 This guide is intended for tutorial purposes only. It describes the general requirements, methods, and procedures for the nondestructive identification and sorting of metals.  
1.2 It provides guidelines for the selection and use of methods suited to the requirements of particular metals sorting or identification problems.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. For specific precautionary statements, see Section 10.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    13 pages
    English language

SIGNIFICANCE AND USE
4.1 This practice is intended primarily for the examination of wrought metals, forged, rolled, machined parts or components to an ultrasonic class most typically specified in the purchase order or other contract document.
SCOPE
1.1 Purpose—This practice establishes the minimum requirements for ultrasonic examination of wrought products.
Note 1: This practice was adopted to replace MIL-STD-2154, 30 Sept. 1982. This practice is intended to be used for the same applications as the document which it replaced. Users should carefully review its requirements when considering its use for new, or different applications, or both.  
1.2 Application—This practice is applicable for examination of materials such as, wrought metals and wrought metal products having a thickness or cross section equal to 0.250 in. (6.35 mm) or greater.  
1.2.1 Wrought Aluminum Alloy Products—Examination shall be in accordance with Practice B594. Angle beam scans of wrought aluminum alloy products shall be performed in accordance with this practice as agreed upon by the purchaser and supplier.  
1.3 Acceptance Class—When examination is performed in accordance with this practice, engineering drawings, specifications, or other applicable documents shall indicate the acceptance criteria. Five ultrasonic acceptance classes are defined in Table 1. One or more of these classes may be used to establish the acceptance criteria or additional or alternate criteria may be specified.  
1.4 Order of Precedence—Contractual requirements and authorized direction from the cognizant engineering organization may add to or modify the requirements of this practice. Otherwise, in the event of conflict between the text of this practice and the references cited herein, the text of this practice takes precedence. Nothing in this practice, however, supersedes applicable laws and regulations unless a specific exemption has been obtained.  
1.5 Measurement Values—The values stated in inch-pounds are to be regarded as standard. The metric equivalents are in parentheses.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    23 pages
    English language
  • Standard
    23 pages
    English language

SIGNIFICANCE AND USE
5.1 This practice gives a measure of a thermal imaging system's effectiveness for detecting a small spot within a large background. Thus, it relates to the detection of small material defects such as voids, pits, cracks, inclusions, and occlusions.  
5.2 MDTD values provide estimates of detection capability that may be used to compare one system with another. (Lower MDTD values indicate better detection capability.)  
5.3 Due to the partially subjective nature of the procedure, repeatability and reproducibility are apt to be poor and MDTD differences less than 0.2 °C are considered to be insignificant.
Note 2: Values obtained under idealized laboratory conditions may or may not correlate directly with service performance.
SCOPE
1.1 This practice covers the determination of the minimum detectable temperature difference (MDTD) capability of a compound observer-thermal imaging system as a function of the angle subtended by the target.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 Radiographic examination may be used during product and process design optimization, on line process control, after manufacture inspection, and in service inspection. In addition to verifying structural placement, radiographic examination can be used in the case of honeycomb core materials to detect node bonds, core-to-core splices, and core-to-structure splices. Radiographic examination is especially well suited for detecting sub-surface flaws. The general types of defects detected by radiographic examination include blown core, core corrosion, damaged filaments, density variation, entrapped fluid, fiber debonding, fiber misalignment, foreign material, fractures, inclusions, micro-cracks, node bond failure, porosity/voids, and thickness variation.  
5.2 Factors that influence image formation and X-ray attenuation in radiographic examination, and which are relevant to interpreting the images for the conditions of interest, should be included in the examination request. Examples include, but not limited to, the following: laminate (matrix and fiber) material, lay-up geometry, fiber volume fraction (flat panels); facing material, core material, facing stack sequence, core geometry (cell size); core density, facing void content, adhesive void content, and facing volume percent reinforcement (sandwich core materials); overall thickness, specimen alignment, and specimen geometry relative to the beam (flat panels and sandwich core materials).  
5.3 Information regarding discontinuities that are detectable using radiographic examination methods can be found in Guide E2533.
SCOPE
1.1 This practice is intended to be used as a supplement to Practices E1742, E1255, E2033, and E2698.  
1.2 This practice describes procedures for radiographic examination of flat panel composites and sandwich core materials made entirely or in part from fiber-reinforced polymer matrix composites. Radiographic examination is: a) Film Radiography (RT), b) Computed Radiography (CR) with Imaging Plate, c) Digital Radiography (DR) with Digital Detector Array’s (DDA), and d) Radioscopic (RTR) Real Time Radiography with a detection system such as an Image Intensifier. The composite materials under consideration typically contain continuous high modulus fibers (> 20 GPa), such as those listed in 1.4.  
1.3 This practice describes established radiographic examination methods that are currently used by industry that have demonstrated utility in quality assurance of flat panel composites and sandwich core materials during product process design and optimization, process control, after manufacture inspection, in service examination, and health monitoring. Additional guidance can be found in E2533, Guide for Nondestructive Testing of Polymer Matrix Composites Used in Aerospace.  
1.4 This practice has utility for examination of flat panel composites and sandwich constructions containing, but not limited to, bismaleimide, epoxy, phenolic, poly(amide imide), polybenzimidazole, polyester (thermosetting and thermoplastic), poly(ether ether ketone), poly(ether imide), polyimide (thermosetting and thermoplastic), poly(phenylene sulfide), or polysulfone matrices; and alumina, aramid, boron, carbon, glass, quartz, or silicon carbide fibers. Typical as-fabricated geometries include uniaxial, cross ply and angle ply laminates; as well as honeycomb core sandwich constructions.  
1.5 This practice does not specify accept-reject criteria and is not intended to be used as a means for approving flat panel composites or sandwich core materials for service.  
1.6 To ensure proper use of the referenced standards, there are recognized nondestructive testing (NDT) specialists that are certified according to industry and company NDT specifications. It is recommended that a NDT specialist be a part of any composite component design, quality assurance, in service maintenance or damage examination.  
1.7 This standard does not purport to address a...

  • Standard
    5 pages
    English language

Frequently Asked Questions

E07 is a Technical Committee within ASTM International. It is named "Nondestructive Testing" and is responsible for: The scope of the Committee includes the promotion of knowledge, stimulation of research and the development of standard specifications, test methods, practices, guides, reference radiographs, terminology, and classifications relating to nondestructive testing (NDT) methods, equipment and applications. NDT methods addressed include X-and gamma radiology, neutron radiology, liquid penetrant, magnetic particle, acoustic emission, ultrasonic, electromagnetic, leak testing, visual/optical examination, and other emerging NDT methods as may be considered appropriate by the Committee. The Committee also addresses requirements for NDT laboratory or agency quality systems, but does not develop standards for the certification of personnel performing nondestructive testing. The work of this Committee will be coordinated with other ASTM technical committees and other national and international organizations having mutual or related interests. This committee has published 1155 standards.

E07 develops ASTM standards in the area of Information technology. The scope of work includes: The scope of the Committee includes the promotion of knowledge, stimulation of research and the development of standard specifications, test methods, practices, guides, reference radiographs, terminology, and classifications relating to nondestructive testing (NDT) methods, equipment and applications. NDT methods addressed include X-and gamma radiology, neutron radiology, liquid penetrant, magnetic particle, acoustic emission, ultrasonic, electromagnetic, leak testing, visual/optical examination, and other emerging NDT methods as may be considered appropriate by the Committee. The Committee also addresses requirements for NDT laboratory or agency quality systems, but does not develop standards for the certification of personnel performing nondestructive testing. The work of this Committee will be coordinated with other ASTM technical committees and other national and international organizations having mutual or related interests. Currently, there are 1155 published standards from this technical committee.

ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.

A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.

Loading...