Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-012: Particular requirements for climatic and environmental testing and other temperature conditioning equipment

1 Scope and object This clause of Part 1 is applicable except as follows: 1.1.1 Equipment included in scope Replacement: Replace the first paragraph by the following: This group safety publication is primarily intended to be used as a product safety standard for the products mentioned in the scope, but shall also be used by technical committees in the preparation of their publications for products similar to those mentioned in the scope of this standard, in accordance with the principles laid down in IEC Guide 104 and lSO/lEC Guide 51. This Part 2 of IEC 61010 specifies safety requirements for electrical equipment and their accessories within the categories a) through c), wherever they are intended to be used, whenever that equipment incorporates one or more of the following characteristics: – A REFRIGERATING SYSTEM that is acted on or impacted by an integral heating function such that the combined heating and cooling system generates additional and/or more severe HAZARDS than those for the two systems if treated separately. – The materials being treated in the intended application introduce significant heat into the REFRIGERATING SYSTEM that the cooling system in the application yield additional and/or more severe HAZARDS than those for the cooling system if operated at the maximum RATED ambient alone. – An irradiation function for the materials being treated presenting additional HAZARDS. – A function to expose the materials being treated to excessive humidity, carbon dioxide, salt mist, or other substances which may result in additional HAZARDS. – A function of MECHANICAL MOVEMENT presenting additional HAZARDS. – Provision for an OPERATOR to walk-in to the operating area to load or unload the materials being treated.

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2-012: Besondere Anforderungen an Klima- und Umwelttestgeräte und andere Temperatur-Konditionierungsgeräte

Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-012: Exigences particulières pour les appareils d'essais climatiques et d'environnement, et autres appareils de conditionnement de température

Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 2-012. del: Posebne zahteve za opremo za klimatska in okoljska preskušanja ter drugo opremo za uravnavanje temperature

General Information

Status
Not Published
Current Stage
5020 - Vote circulated
Due Date
24-Dec-2021
Completion Date
24-Dec-2021

RELATIONS

Buy Standard

Draft
FprEN IEC 61010-2-012:2019/oprAA:2020
English language
12 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (sample)

SLOVENSKI STANDARD
SIST FprEN IEC 61010-2-
012:2019/oprAA:2020
01-november-2020
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-012. del: Posebne zahteve za opremo za klimatska in okoljska
preskušanja ter drugo opremo za uravnavanje temperature

Safety requirements for electrical equipment for measurement, control and laboratory

use - Part 2-012: Particular requirements for climatic and environmental testing and other

temperature conditioning equipment

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2

-012: Besondere Anforderungen an Klima- und Umwelttestgeräte und andere
Temperatur-Konditionierungsgeräte
Règles de sécurité pour appareils électriques de mesurage, de régulation et de

laboratoire - Partie 2-012: Exigences particulières pour les appareils d'essais climatiques

et d'environnement, et autres appareils de conditionnement de température
Ta slovenski standard je istoveten z: FprEN IEC 61010-2-012:2019/prAA
ICS:
19.040 Preskušanje v zvezi z Environmental testing
okoljem
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST FprEN IEC 61010-2- en,fr,de
012:2019/oprAA:2020

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST FprEN IEC 61010-2-012:2019/oprAA:2020
---------------------- Page: 2 ----------------------
SIST FprEN IEC 61010-2-012:2019/oprAA:2020
EUROPEAN STANDARD DRAFT
FprEN IEC 61010-2-
NORME EUROPÉENNE
012:2019
EUROPÄISCHE NORM
prAA
September 2020
ICS
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-012: Particular requirements
for climatic and environmental testing and other temperature
conditioning equipment

Règles de sécurité pour appareils électriques de mesurage, Sicherheitsbestimmungen für elektrische Mess-, Steuer-,

de régulation et de laboratoire - Partie 2-012: Exigences Regel- und Laborgeräte - Teil 2-012: Besondere

particulières pour les appareils d'essais climatiques et Anforderungen an Klima- und Umwelttestgeräte und andere

d'environnement, et autres appareils de conditionnement de Temperatur-Konditionierungsgeräte

température

This draft amendment prAA, if approved, will modify the European Standard FprEN IEC 61010-2-012:2019; it is submitted to CENELEC

members for enquiry.
Deadline for CENELEC: 2020-11-27.
It has been drawn up by CLC/TC 66X.

If this draft becomes an amendment, CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which

stipulate the conditions for giving this amendment the status of a national standard without any alteration.

This draft amendment was established by CENELEC in three official versions (English, French, German).

A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to

the CEN-CENELEC Management Centre has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,

Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the

Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,

Turkey and the United Kingdom.

Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to

provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and

shall not be referred to as a European Standard.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels

© 2020 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.

---------------------- Page: 3 ----------------------
SIST FprEN IEC 61010-2-012:2019/oprAA:2020
EN IEC 61010-2-012:2019/prAA:2020 (E)
European foreword

This document (FprEN 61010-2-012:2019/prAA:2020) has been prepared by CLC/TC 66X “Safety of

measuring, control, and laboratory equipment”.
This document is currently submitted to the Enquiry.
The following dates are proposed:
• latest date by which the existence of this (doa) dor + 6 months
document has to be announced at national
level
• latest date by which this document has to be (dop) dor + 12 months
implemented at national level by publication of
an identical national standard or by
endorsement
• latest date by which the national standards (dow) dor + 36 months
conflicting with this document have to be (to be confirmed or
withdrawn modified when voting)
This document amends FprEN 61010-2-012:2019.

This document has been prepared under a mandate given to CENELEC by the European Commission

and the European Free Trade Association, and supports essential requirements of EU Directive(s).

For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this

document.
---------------------- Page: 4 ----------------------
SIST FprEN IEC 61010-2-012:2019/oprAA:2020
EN IEC 61010-2-012:2019/prAA:2020 (E)
1 Addition to Clause 2, “Normative references”
Add the following normative references to Clause 2:
"EN 809, Pumps and pump units for liquids - Common safety requirements

EN 837-1, Pressure gauges - Part 1: Bourdon tube pressure gauges - Dimensions, metrology,

requirements and testing

EN 837-2, Pressure gauges - Part 2: Selection and installation recommendations for pressure gauges

EN 837-3, Pressure gauges - Part 3: Diaphragm and capsule pressure gauges - Dimensions,

metrology, requirements and testing

EN 1736, Refrigerating systems and heat pumps - Flexible pipe elements, vibration isolators,

expansion joints and non-metallic tubes - Requirements, design and installation

EN 1779:1999, Non-destructive testing - Leak testing - Criteria for method and technique selection

EN 12178, Refrigerating systems and heat pumps - Liquid level indicating devices - Requirements,

testing and marking

EN 12263, Refrigerating systems and heat pumps - Safety switching devices for limiting the pressure -

Requirements and tests

EN 12284, Refrigerating systems and heat pumps - Valves - Requirements, testing and marking

EN 12693, Refrigerating systems and heat pumps - Safety and environmental requirements - Positive

displacement refrigerant compressors

EN 13136, Refrigerating systems and heat pumps - Pressure relief devices and their associated piping

- Methods for calculation
EN 13445 (series), Unfired pressure vessels
EN 13480 (series), Metallic industrial piping

EN 14276-1, Pressure equipment for refrigerating systems and heat pumps - Part 1: Vessels -

General requirements

EN 14276-2, Pressure equipment for refrigerating systems and heat pumps - Part 2: Piping - General

requirements

EN 61770, Electric appliances connected to the water mains - Avoidance of backsiphonage and failure

of hose-sets

IEC 60204-1, Safety of machinery - Electrical equipment of machines - Part 1: General requirements

ISO 4126-1, Safety devices for protection against excessive pressure — Part 1: Safety valves

ISO 4126-2, Safety devices for protection against excessive pressure — Part 2: Bursting disc safety

devices"
2 Modification to 5.4.3, “Equipment installation”
Add the following item after n):

“o) Instructions on the prevention of back-siphonage into potable water systems (see 11.101)”

---------------------- Page: 5 ----------------------
SIST FprEN IEC 61010-2-012:2019/oprAA:2020
EN IEC 61010-2-012:2019/prAA:2020 (E)
3 Modification to 6.8.3.1, “The a.c. voltage test”
Replace the first sentence by the following sentence:

“The voltage tester shall be capable of maintaining the test voltage throughout the test within ± 5 % of

the specified value.”
4 Modifications to Clause 11, “Protection against HAZARDS from fluids and
solid foreign objects”
In 11.1, “General”, delete the added text after the first conformity statement.
Add the following subclause:
“11.101 Protection of hot and cold water services

Back-siphonage from the equipment to the potable water services shall be prevented by means

meeting the relevant requirements of EN 61770. Attention is drawn to the existence of national and

local regulations. If the means are to be provided by the RESPONSIBLE BODY, this shall be stated in the

manufacturer’s installation instructions.

Conformity is checked by inspection and by examination of the manufacturer’s instructions.”

5 Modification to Annex CC, “Safety requirements for components and
piping”
Replace Annex CC by the following text:
"Annex CC
(normative)
Safety requirements for components and piping

The sealed system components can be considered pressure vessels as stated in the Pressure

Equipment Directive (PED) 2014/68/EU according to the classification in Table CC.1 and Table CC.2.

If the components or piping are classified as a Category II or a higher pressure vessel as stated in the

PED, then the requirements of Table CC.3 shall apply including the use of a Notified Body to the PED.

Table CC.1 — Parameters of pressure vessels according to EN 14276-1
PS × V
Fluid Nature V (L) Category/Article
PS (bar)
(bar × L)
if and and and and then
Not subjected to
≤ 0,5 - -
PED
≤ 1 -
Art. 4.3
≤ 25
Art. 4.3
> 0,5 and ≤ 200
> 25 and ≤ 50 I
> 1
Group 1 Gas
> 50 and ≤ 200 II
> 200 and ≤ 1
≤ 1 - III
000
> 200 and
III
≤ 1 000
≤ 1 000 > 1
> 1 000 IV
---------------------- Page: 6 ----------------------
SIST FprEN IEC 61010-2-012:2019/oprAA:2020
EN IEC 61010-2-012:2019/prAA:2020 (E)
PS × V
Fluid Nature V (L) Category/Article
PS (bar)
(bar × L)
if and and and and then
> 1 000 - - IV
Not subjected to
≤ 0,5 - -
PED
≤ 1 -
Art. 4.3
> 0,5 and ≤ 500
≤ 200
Art. 4.3
Liquid
> 0,5 and ≤ 10 I
> 1 > 200
> 10 and ≤ 500 II
> 500 < 1 - II
> 500 > 1 - III
Not subjected to
≤ 0,5 - -
PED
≤ 1 -
Art. 4.3
≤ 50
Art. 4.3
> 0,5 and ≤ 1 000
> 50 and ≤ 200 I
> 1
> 200 and
Gas
≤ 1 000
≤ 1 - III
> 1 000 and
> 1 000 and
≤ 3 000
III
≤ 3 000
> 0,5 and ≤ 4 > 1 000 III
Group 2
> 1
> 4 > 3 000 IV
> 3 000 - - IV
Not subjected to
≤ 0,5 - -
PED
> 0,5 and ≤ 10 - -
Art. 4.3
> 10 and ≤ 1 000 ≤ 10 -
Art. 4.3
Liquid
> 10 and ≤ 1 000 > 10 ≤ 10 000
Art. 4.3
> 10 and ≤ 500 - > 10 000 I
> 1 000 < 10 - I
> 500 > 10 > 10 000 II
1 bar = 0,1 Mpa.
PED = Pressure Equipment Directive
Art. 4.3 = reference to Article 4.3 of the Pressure Equipment Directive

Liquids are considered to be fluids having a vapour pressure of not more than 0,5 bar above the normal

atmospheric pressure (1 013 mbar)
---------------------- Page: 7 ----------------------
SIST FprEN IEC 61010-2-012:2019/oprAA:2020
EN IEC 61010-2-012:2019/prAA:2020 (E)
Table CC.2 — Parameters of piping according to EN 14276-2
PS × DN (bar)
Fluid Nature DN Category/Article
PS (bar)
if and and And and then
Not subjected to
≤ 0,5 - -
PED
≤ 25 -
Art. 4.3
Gas
> 25 and ≤ 100 ≤ 1 000 I
> 0,5
> 1 000 and
> 100 and ≤ 350 II
≤ 3 500
> 350 > 3 500 III
Group 1
Not subjected to
≤ 0,5 - -
PED
≤ 25 -
Art. 4.3
> 0,5
d c
- ≤ 2 000
Liquid Art. 4.3
> 0,5 and ≤ 10 - I
> 2 000
> 10 and ≤ 500 II
> 25
> 500 - III
Not subjected to
≤ 0,5 - -
PED
≤ 32 -
Art. 4.3
- ≤ 1 000
Art. 4.3
Gas
> 1 000 and
> 32 and ≤ 100 I
> 0,5
≤ 3 500
> 3 500 and
> 100 and ≤ 250 II
≤ 5 000
Group 2
> 250 > 5 000 III
Not subjected to
≤ 0,5 - -
PED
> 0,5 and ≤ 10 - -
Art. 4.3
d - - ≤ 5 000
Art. 4.3
Liquid
- ≤ 200 -
Art. 4.3
> 10 and ≤ 500 > 5 000 I
> 200
> 500 - II
1 bar = 0,1 Mpa.
PED = Pressure Equipment Directive.
Art. 4.3 = reference to Article 4.3 of the Pressure Equipment Directive.

Liquids are considered to be fluids having a vapour pressure of not more than 0,5 bar above the normal

atmospheric pressure (1 013 mbar).
...

Questions, Comments and Discussion