Ultrasonics - Pulse-echo scanners - Part 1: Techniques for calibrating spatial measurement systems and measurement of system point-spread function response

Ultraschall - Impuls-Echo-Scanner - Teil 1: Verfahren für die Kalibrierung von räumlichen Messsystemen und Messung der Charakteristik der Punktverwaschungsfunktion des Systems

Ultrasons - Scanners à impulsion et écho - Partie 1: Techniques pour l'étalonnage des systèmes de mesure spatiaux et des mesures de la réponse de la fonction de dispersion ponctuelle du système

Ultrazvok - Impulzno-odmevni skenerji - 1. del: Tehnike za kalibriranje prostorskih merilnih sistemov in meritve karakteristike funkcije razpršitve točk - Dopolnilo A1 (IEC 61391-1:2006/A1:2017)

General Information

Status
Published
Publication Date
30-Nov-2017
Withdrawal Date
30-Nov-2020
Technical Committee
Drafting Committee
Current Stage
6060 - Document made available - Publishing
Start Date
01-Dec-2017
Completion Date
01-Dec-2017

Relations

Buy Standard

Amendment
EN 61391-1:2008/A1:2018 - BARVE
English language
28 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-februar-2018
8OWUD]YRN,PSXO]QRRGPHYQLVNHQHUMLGHO7HKQLNH]DNDOLEULUDQMHSURVWRUVNLK
PHULOQLKVLVWHPRYLQPHULWYHNDUDNWHULVWLNHIXQNFLMHUD]SUãLWYHWRþN'RSROQLOR$
,(&$
Ultrasonics - Pulse-echo scanners - Part 1: Techniques for calibrating spatial
measurement systems and measurement of point-spread function response (IEC 61391-
1:2006/A1:2017)
Ultraschall - Impuls-Echo-Scanner - Teil 1: Verfahren für die Kalibrierung von räumlichen
Messsystemen und Messung der Charakteristik der Punktverwaschungsfunktion des
Systems (IEC 61391-1:2006/A1:2017)
Spécifications pour types particuliers de fils de bobinage - Partie 0-7: Exigences
générales - Fil de section circulaire, isolé en continu, en cuivre émaillé, sans défaut
électrique (IEC 61391-1:2006/A1:2017)
Ta slovenski standard je istoveten z: EN 61391-1:2006/A1:2017
ICS:
11.040.55 'LDJQRVWLþQDRSUHPD Diagnostic equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD EN 61391-1:2006/A1

NORME EUROPÉENNE
EUROPÄISCHE NORM
December 2017
ICS 17.140.50
English Version
Ultrasonics - Pulse-echo scanners -
Part 1: Techniques for calibrating spatial measurement systems
and measurement of system point-spread function response
(IEC 61391-1:2006/A1:2017)
Ultrasons - Scanners à impulsion et écho -  Ultraschall - Impuls-Echo-Scanner -
Partie 1: Techniques pour l'étalonnage des systèmes de Teil 1: Verfahren für die Kalibrierung von räumlichen
mesure spatiaux et des mesures de la réponse de la Messsystemen und Messung der Charakteristik der
fonction de dispersion ponctuelle du système Punktverwaschungsfunktion des Systems
(IEC 61391-1:2006/A1:2017) (IEC 61391-1:2006/A1:2017)
This amendment A1 modifies the European Standard EN 61391-1:2006; it was approved by CENELEC on 2017-08-14. CENELEC
members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the
status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden,
Switzerland, Turkey and the United Kingdom.

European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2017 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 61391-1:2006/A1:2017 E

European foreword
The text of document 87/650/FDIS, future IEC 61391-1:2006/A1, prepared by IEC/TC 87 "Ultrasonics"
was submitted to the IEC-CENELEC parallel vote and approved by CENELEC as
The following dates are fixed:
(dop) 2018-06-01
• latest date by which the document has to be
implemented at national level by
publication of an identical national
standard or by endorsement
• latest date by which the national (dow) 2020-12-01
standards conflicting with the
document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.

Endorsement notice
The text of the International Standard IEC 61391-1:2006/A1:2017 was approved by CENELEC as a
European Standard without any modification.
In the Bibliography of EN 61391-1:2006, the following notes have to be added for the standards
indicated:
IEC 62563-1:2009 NOTE Harmonized as EN 62563-1:2010 (not modified).
IEC 61391-2:2010 NOTE Harmonized as EN 61391-2:2010 (not modified).
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications

The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1  Where an International Publication has been modified by common modifications, indicated by (mod), the relevant
EN/HD applies.
NOTE 2  Up-to-date information on the latest versions of the European Standards listed in this annex is available here:
www.cenelec.eu.
Publication Year Title EN/HD Year
In Annex ZA of EN 61391-1:2006 replace IEC 61102:1991 by the following:
IEC 62127-1 2007 Ultrasonics - Hydrophones - EN 62127-1 2007
Part 1: Measurement and characterization
of medical ultrasonic fields up to 40 MHz
Add to Annex ZA of EN 61391-1:2006 the following new references:
IEC 60050-801 1994 International Electrotechnical Vocabulary - -
(IEV) -
Chapter 801: Acoustics and
electroacoustics
IEC 60050-802 2011 International Electrotechnical Vocabulary - - -
Part 802: Ultrasonics
IEC 61391-1 ®
Edition 1.0 2017-07
INTERNATIONAL
STANDARD
colour
inside
AMENDMENT 1
Ultrasonics – Pulse-echo scanners –

Part 1: Techniques for calibrating spatial measurement systems and

measurement of system point-spread function response

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 17.140.50 ISBN 978-2-8322-4289-6

– 2 – IEC 61391-1:2006/AMD1:2017
© IEC 2017
FOREWORD
This amendment has been prepared by IEC technical committee 87: Ultrasonics.
The text of this amendment is based on the following documents:
FDIS Report on voting
87/650/FDIS 87/653/RVD
Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC website under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
A bilingual version of this publication may be issued at a later date.

IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
____________
2 Normative references
Replace:
IEC 61102:1991, Measurement and characterisation of ultrasonic fields using hydrophones in
the frequency range 0,5 MHz to 15 MHz
with:
IEC 62127-1:2007, Ultrasonics – Hydrophones – Part 1: Measurement and characterization of
medical ultrasonic fields up to 40 MHz
Insert the following new normative references in proper numerical sequence:
IEC 60050-801:1994, International Electrotechnical Vocabulary – Chapter 801: Acoustics and
electroacoustics
IEC 61391-1:2006/AMD1:2017 – 3 –
© IEC 2017
IEC 60050-802:2011, International Electrotechnical Vocabulary – Part 802: Ultrasonics
3 Terms and definitions
Replace the first two paragraphs with the following new paragraph:
For the purposes of this document, the terms and definitions given in IEC 60050-801:1994,
IEC 60050-802:2011, IEC 62127-1:2007 and the following apply. See also related
International Standards, Technical Specifications and Technical Reports for definitions and
explanations [1] [2] [3] [4] [34] [35] [36] [37] [38] [39].
3.25
point-spread function
PSF
Add the following new sentence at the end of the NOTE:
The problem is solved by PSF mapping – see Annex D.
Add the following new terms and definitions to Clause 3, starting with 3.45.
3.45
accuracy
closeness of agreement between a test result and the accepted reference value
[SOURCE: ISO 5725-1:1994, 3.6]
3.46
axial resolution in a PSF-map
twice the Half-Width-at-Half-Maximum (HWHM) of a function’s trace created from a set of
increasing pixel values, commencing near zero and terminating at the first maximum value
(centre of the PSF) and representing the leading edge of the echo signal from a point reflector
located on the main beam axis
Note 1 to entry: The axial resolution in a PSF map differs from the axial resolution specified by 3.5. It is used
for the PSF-mapping only to simplify the data acquisition.
Note 2 to entry: A detailed explanation of the axial resolution in the PSF-map measuring method is in D.6.1.4.
Note 3 to entry: The axial resolution mainly depends on the ultrasound frequency used, not on sonograph
construction.
Note 4 to entry: Axial resolution in a PSF-map is expressed in metres.
3.47
brightness
luminance as perceived by the human visual system
[SOURCE: IEC 62563-1:2009, 3.1.2]
3.48
contrast
C
ratio of the difference of the luminance of two image areas, L − L , divided by the average of
1 2
the two luminance values:
C = 2 (L − L )/(L + L )
1 2 1 2
[SOURCE: IEC 62563-1:2009, 3.1.6]

– 4 – IEC 61391-1:2006/AMD1:2017
© IEC 2017
3.49
dynamic imaging
real-time imaging
imaging with a frame rate that is high enough to observe moving structures in apparently
continuous motion
3.50
elevational resolution in a PSF-map
difference of point-reflector displacements in passing through the scanning plane in an
elevational direction, which result in decreases of MER of −6 dB compared to the MER-value
in the beam centre
Note 1 to entry: The elevational resolution in a PSF-map differs from the elevational resolution specified by
3.12. It is used for the PSF-mapping only to simplify the data acquisition.
Note 2 to entry: Detailed explanation of the method is in D.6.1.3.
Note 3 to entry: Elevational resolution in a PSF-map is expressed in metres.
3.51
overall gain
G
o
basic level of gain that is uniform for the whole scan area but modified by TGC relative to the
depth of the scan
3.52
profile line
set of pixel values ordered along an abscissa according to the sequence during their
acquisition
3.53
lateral resolution in a PSF-map
Full-Width at Half-Maximum (FWHM) of the PSF, measured in a lateral direction
Note 1 to entry: The lateral resolution in a PSF-map differs from the lateral resolution specified by 3.17. It is
used for the PSF-mapping only to simplify the data acquisition.
Note 2 to entry: Detailed explanation of the method is in D.6.1.2.
Note 3 to entry: Lateral resolution in a PSF-map is expressed in metres.
3.54
measuring grid
matrix of points specified by Cartesian coordinates x and z defined in a plane parallel to the
i j
scanning plane
Note 1 to entry: Each point determines the position (x ,z ) in which individual measurement of PSF is performed.
i j
Note 2 to entry: The step ∆x is defined as an increment x – x in the lateral direction. The step ∆z is defined as
i+1 i
an increment z – z in the axial direction.
j+1 j
3.55
performance evaluation
tests performed to assess specific absolute performance of the object tested
Note 1 to entry: Typical times for ultrasound-system performance evaluation are at pre-purchase evaluation,
new- and repaired-system acceptance testing, at time of performance difficulties, and at end-of-useful-life
evaluations.
[SOURCE: IEC TS 62736:2016, 3.5]

IEC 61391-1:2006/AMD1:2017 – 5 –
© IEC 2017
3.56
precision
closeness of agreement between independent test results obtained under stipulated
conditions
[SOURCE: ISO 5725-1:1994, 3.12]
3.57
scanning window
area on the surface of the test tank dedicated for transducer application to obtain a suitable
sonogram of the target
Note 1 to entry: It is important that the scanning window be covered by flexible foil made of material with similar
acoustic properties to the working liquid to avoid ultrasound field reflections and absorption.
Note 2 to entry: The foil flexibility should assure proper acoustical contact of any type of curved transducer.
Note 3 to entry: It is important that the foil covering the scanning-window be tough enough to prevent its damage
during coupling the measured transducer to the scanning window, to prevent resultant leakage of working liquid
from the measuring tank.
Note 4 to entry: The scanning window has the identical function as the test object scanning surface in the
case of tissue-mimicking test objects (see 3.34).
3.58
side-lobe signal
echo signal generated by ultrasound signal transmitted/received in a direction different from
the central axis of the transducer
3.59
test tank
tank designed to be suitable for providing specified kind of tests, which is filled with a
working liquid and equipped with scanning window(s)
Replace the title of Clause 4 with the following new title:
4 Symbols and abbreviated terms
Add the following symbols and abbreviated terms
...