Corrigendum 1 - Medical electrical equipment - Part 2-20: Particular requirements for the basic safety and essential performance of infant transport incubators

Corrigendum 1 - Appareils électromédicaux - Partie 2-20: Exigences particulières pour la sécurité de base et les performances essentielles des incubateurs de transport pour nouveau-nés

General Information

Status
Published
Publication Date
28-Feb-2012
Current Stage
DELPUB - Deleted Publication
Completion Date
02-Sep-2020
Ref Project

Relations

Buy Standard

Standard
IEC 60601-2-20:2009/COR1:2012 - Corrigendum 1 - Medical electrical equipment - Part 2-20: Particular requirements for the basic safety and essential performance of infant transport incubators Released:2/29/2012
English and French language
10 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


IEC 60601-2-20
(Second edition – 2009)
Medical electrical equipment –

Part 2-20: Particular requirements for the basic safety

and essential performance of infant transport incubators

CO RRI G ENDU M 1
201.4.3.101 ESSENTIAL PERFORMANCE of INFANT TRANSPORT INCUBATORS
Replace the existing Table 201.101 by the following:

Table 201.101 – Additional ESSENTIAL PERFORMANCE requirements
Requirement Subclause
ESSENTIAL PERFORMANCE requirement 1 201.12.1.104 or generation of a
visual and audible alarm in
compliance with 201.9.6.2.1.102
ESSENTIAL PERFORMANCE requirement 2 201.12.1.106 or generation of a
visual and audible alarm in
compliance with 201.9.6.2.1.102

201.9.6.2.1.101 Sound level within the COMPARTMENT
In the first paragraph, replace the reference to "201.12.3.107" by "201.9.6.2.1.103".

February 2012
CEI 60601-2-20
(Deuxième édition – 2009
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.