IEC 60789:2005/COR1:2009
(Corrigendum)Corrigendum 1 - Medical electrical equipment - Characteristics and test conditions of radionuclide imaging devices - Anger type gamma cameras
Corrigendum 1 - Medical electrical equipment - Characteristics and test conditions of radionuclide imaging devices - Anger type gamma cameras
Corrigendum 1 - Appareils électromédicaux - Caractéristiques et conditions d'essai des dispositifs d'imagerie par radionucléides - Gamma caméras de type Anger
General Information
Relations
Standards Content (Sample)
IEC 60789
(Third edition – 2005)
Medical electrical equipment –
Characteristics and test conditions of radionuclide imaging devices –
Anger type gamma cameras
CORRIGENDUM 1
Table 1 – Radionuclides and energy windows to be used for performance measurements
Instead of:
Table 1 – Radionuclides and energy windows
to be used for performance measurements
Energy window
Radionuclide
keV
99m
Tc
141, with a tolerance of ± 7,5 %
I
364, with a tolerance of ± 10 %
Ga 93, 184, 300, with a tolerance of ± 20 %
Ga 511, with a tolerance of ± 10 %
NOTE Because the characteristics of a gamma camera may change noticeably
between 122 keV (57Co) and 141 keV (99mTc), the former is not included as a suitable
radionuclide. However, it may be useful in some circumstances, e.g. for quality control.
read:
Table 1 – Radionuclides and energy windows
to be used for performance measurements
Energy window
Radionuclide
keV
99m
Tc 141 (± 7,5 %)
I
364 (± 10 %)
Ga
93, 184, 300 (± 20 %)
Ga 511 (± 10 %)
NOTE Because the characteristics of a gamma camera may change noticeably
57 99m
between 122 keV ( Co) and 141 keV ( Tc), the former is not included as a suitable
radionuclide. However, it may be useful in some circumstances, e.g. for quality control.