Iron ores — Determination of various elements by X-ray fluorescence spectrometry — Part 4: Performance-based method using fusion preparation method

This document specifies a performance-based method for the chemical analysis of natural and processed iron ores by fused bead wavelength and energy dispersive X-ray fluorescence (XRF). It is applicable to all elements of interest when adequate calibrations have been established.

Titre manque — Partie 4: Titre manque

General Information

Status
Published
Publication Date
29-Nov-2021
Current Stage
6060 - International Standard published
Start Date
30-Nov-2021
Due Date
10-Mar-2022
Completion Date
30-Nov-2021
Ref Project

Relations

Buy Standard

Technical specification
ISO/TS 9516-4:2021 - Iron ores -- Determination of various elements by X-ray fluorescence spectrometry
English language
12 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

TECHNICAL ISO/TS
SPECIFICATION 9516-4
First edition
2021-11
Iron ores — Determination of various
elements by X-ray fluorescence
spectrometry —
Part 4:
Performance-based method using
fusion preparation method
Reference number
ISO/TS 9516-4:2021(E)
© ISO 2021
---------------------- Page: 1 ----------------------
ISO/TS 9516-4:2021(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2021

All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may

be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on

the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below

or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
© ISO 2021 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/TS 9516-4:2021(E)
Contents Page

Foreword ........................................................................................................................................................................................................................................iv

Introduction .................................................................................................................................................................................................................................v

1 Scope ................................................................................................................................................................................................................................. 1

2 Normative references ..................................................................................................................................................................................... 1

3 Terms and definitions .................................................................................................................................................................................... 1

4 Reagents and materials ................................................................................................................................................................................ 2

4.1 Pure reagents ........................................................................................................................................................................................... 2

4.2 Flux ................................................................................................................................................................................................................... 3

4.3 Releasing agent ...................................................................................................................................................................................... 3

4.4 Oxidizing agents .................................................................................................................................................................................... 3

4.5 Certified reference materials .................................................................................................................................................... 3

4.6 Reference materials ........................................................................................................................................................................... 3

5 Sampling and samples ........................................................................................................................................... .........................................3

5.1 Laboratory sample .............................................................................................................................................................................. 3

5.2 Preparation of test samples, CRMs and RMs ............................................................................................................... 4

5.2.1 General ........................................................................................................................................................................................ 4

5.2.2 Ores having significant contents of combined water or easily oxidizable

compounds .............................................................................................................................................................................. 4

5.2.3 Ores outside the scope of 5.2.2 .............................................................................................................................. 4

5.3 Test portion ............................................................................................................................................................................................... 4

6 Apparatus .................................................................................................................................................................................................................... 4

7 Measurements ........................................................................................................................................................................................................5

7.1 General ........................................................................................................................................................................................................... 5

7.1.1 Analytical line ....................................................................................................................................................................... 5

7.1.2 Voltage and current ......................................................................................................................................................... 6

7.1.3 Measuring times ................................................................................................................................................................. 6

7.1.4 Background measurements...................................................................................................................................... 6

8 Calibration and validation .........................................................................................................................................................................6

8.1 Principles ..................................................................................................................................................................................................... 6

8.2 Preparation of fusion beads ........................................................................................................................................................ 6

8.3 Calibration and validation samples ...................................................................................................................................... 6

8.4 Validation of the calibration ....................................................................................................................................................... 7

8.4.1 Specimens ................................................................................................................................................................................. 7

8.4.2 Trueness validation ........................................................................................................................................... ............... 7

8.5 Calibration maintenance ............................................................................................................................................................... 8

8.5.1 Monitor disc ............................................................................................................................................................................ 8

8.5.2 Duplicate monitor discs ............................................................................................................................................... 8

8.5.3 Monitor measurements ................................................................................................................................................ 8

8.5.4 Quality control measurements .............................................................................................................................. 9

9 Reporting ..................................................................................................................................................................................................................10

9.1 General ........................................................................................................................................................................................................ 10

9.2 Calculation of results ..................................................................................................................................................................... 10

9.3 Number of decimals ........................................................................................................................................................................ 11

10 Test report ...............................................................................................................................................................................................................11

Bibliography .............................................................................................................................................................................................................................12

iii
© ISO 2021 – All rights reserved
---------------------- Page: 3 ----------------------
ISO/TS 9516-4:2021(E)
Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards

bodies (ISO member bodies). The work of preparing International Standards is normally carried out

through ISO technical committees. Each member body interested in a subject for which a technical

committee has been established has the right to be represented on that committee. International

organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.

ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of

electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are

described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the

different types of ISO documents should be noted. This document was drafted in accordance with the

editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of

any patent rights identified during the development of the document will be in the Introduction and/or

on the ISO list of patent declarations received (see www.iso.org/patents).

Any trade name used in this document is information given for the convenience of users and does not

constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and

expressions related to conformity assessment, as well as information about ISO’s adherence to

the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see

www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 102, Iron ore and direct reduced iron,

Subcommittee SC 2 Chemical analysis.
A list of all parts in the ISO 9516 series can be found on the ISO website.

Any feedback or questions on this document should be directed to the user’s national standards body. A

complete listing of these bodies can be found at www.iso.org/members.html.
© ISO 2021 – All rights reserved
---------------------- Page: 4 ----------------------
ISO/TS 9516-4:2021(E)
Introduction

X-ray fluorescence (XRF) spectrometry is a multi-elemental non-destructive analytical methodology

used for quantitative and qualitative determinations. It is element specific covering an elemental range

from boron (B) to uranium (U).

Once the sample has been dissolved into a borate glass it may be introduced to the spectrometer for

analysis. The sample is then irradiated by intense radiation from an X-ray tube. Analysis of fused glass

beads offers advantages over pressed powder techniques as it eliminates particle size effects, and it

produces a homogeneous specimen for each element.

In some instances, the relationship between intensity (or intensity ratios) and concentration can

be linear. For most analytes there is no direct straightforward relationship between intensity and

concentration. With samples of differing compositions, the X-rays are absorbed differently in the

different samples giving rise to what are generally referred to as matrix effects. These inter-element

effects can be corrected using mathematical models derived from the known physics of X-rays.

Calibration can be based on binary standards (prepared from pure oxides or liquid solutions), reference

materials, secondary standards, or combinations therewith.
© ISO 2021 – All rights reserved
---------------------- Page: 5 ----------------------
TECHNICAL SPECIFICATION ISO/TS 9516-4:2021(E)
Iron ores — Determination of various elements by X-ray
fluorescence spectrometry —
Part 4:
Performance-based method using fusion preparation
method

WARNING — This document can involve hazardous materials, operations and equipment. This

document does not purport to address all of the safety problems associated with its use. It is the

responsibility of the user of this document to establish appropriate health and safety practices

and determine the applicability of regulatory limitations prior to use.
1 Scope

This document specifies a performance-based method for the chemical analysis of natural and

processed iron ores by fused bead wavelength and energy dispersive X-ray fluorescence (XRF).

It is applicable to all elements of interest when adequate calibrations have been established.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content

constitutes requirements of this document. For dated references, only the edition cited applies. For

undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 2596, Iron ores — Determination of hygroscopic moisture in analytical samples — Gravimetric, Karl

Fischer and mass-loss methods
ISO 3082, Iron ores — Sampling and sample preparation procedures
ISO 7764, Iron ores — Preparation of predried test samples for chemical analysis

ISO Guide 31, Reference materials — Contents of certificates, labels and accompanying documentation

ISO Guide 35, Reference materials — Guidance for characterization and assessment of homogeneity and

stability
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at http:// www .electropedia .org/
3.1
primary standard

standard that is designated or widely acknowledged as having the highest metrological qualities and

whose value is accepted without reference to other standards of the same quantity

Note 1 to entry: The concept of a primary standard is equally valid for base quantities and derived quantities.

© ISO 2021 – All rights reserved
---------------------- Page: 6 ----------------------
ISO/TS 9516-4:2021(E)

Note 2 to entry: A primary standard is never used directly for measurement other than for comparison with

other primary standards or reference standards.
3.2
reference material

standard, generally having the highest metrological quality available at a given location or in a given

organization, from which measurements made there are derived
3.3
certified reference material
CRM

material, e.g. iron ores, supplied by an organization conforming to the requirements for the competence

of reference material (3.2) producers

Note 1 to entry: The requirements for the competence of reference material producers are given in ISO 17034.

Note 2 to entry: CRMs shall be supplied with a certificate of analysis giving information on the average value

and standard deviation (with and between laboratory precision statistics) and measurement of uncertainty in

accordance with ISO Guide 31.
3.4
accepted reference value
ARV

value that serves as an agreed upon reference for comparison, and which is derived as:

a) a theoretical or established value, based on scientific principles;

b) an assigned or certified value, based on experimental work of some national or international

organization;

c) a consensus or certified value, based on collaborative experimental work under the auspices of a

scientific or engineering group.

Note 1 to entry: When none of the above are available, the ARV is the expectation of the (measurable) quantity,

i.e. the mean of a specified population of measurements.
3.5
referee method

method that is independent of other methods (i.e. calibrated

with primary standards or able to arrive at the final results using direct measurements or calculations

from known physical/chemical laws) and is in the initial stages of the traceability chain

Note 1 to entry: Primary methods include gravimetry, titrimetry, coulometry and isotope dilution mass

spectrometry.
4 Reagents and materials
4.1 Pure reagents

Reagents shall be of analytical quality and, wherever possible, be pure oxides or carbonates, except for

the calibration of such elements as sulfur, chlorine, bromine, or phosphorus, which do not form stable

oxides or carbonates, where some guarantee of stoichiometry is required.

Reagents shall be free of (or corrected for) the presence of moisture (and, in the case of oxides, carbon

dioxide) when weighed out for fusion.

Reagents shall be used in a known stoichiometry in terms of content. In order to achieve this, they

can be treated before use. Generally, the oxides of iron, silicon, manganese, aluminium, titanium and

magnesium shall be heated to 1 000 °C. More information can be found in ISO 9516-1. The procedures

specified in ISO 9516-1 ensure that the correct oxidation state is obtained.
© ISO 2021 – All rights reserved
---------------------- Page: 7 ----------------------
ISO/TS 9516-4:2021(E)
All laboratory reagents used for referee methods shall be traceable.

Where reagents have been ignited, they should be covered during cooling in the desiccator and weighed

as soon as possible.
4.2 Flux

High purity lithium borate or sodium borate fluxes should be used. Prior to using, the levels of

contamination shall be checked. Good results have been obtained with a mixture that contains 6 parts

of Li B O and 11 parts of LiBO by mass. This composition is a eutectic and can be used at relatively

2 4 7 2
low temperatures.
NOTE 1 This mixture is commercially available under the designation 12:22.
NOTE 2 A typical sample to flux ratio is 1 unit of sample for 10 units of flux.
4.3 Releasing agent

To facilitate the casting of the sample-flux mixture and the releasing of the resultant bead after

cooling, a releasing agent such as ammonium iodide (NH I) or lithium iodide (LiI) is recommended.

Alternatively, lithium bromide (LiBr) can be used.

If a bromide-containing releasing agent is used, the line overlap of the Br L-lines with Al Kα shall be

taken into account.
NOTE Fluxes with an integrated releasing agent are commercially available.
4.4 Oxidizing agents

Generally, no oxidizing agents are required if the iron is present as hematite (Fe O ). However, if the

2 3

samples to be analysed contain, for example, magnetite (Fe O ) or if a gas burner is used without

3 4

oxygen supplement, then the addition of an oxidizing agent to the flux can be required. In those cases,

the use of lithium nitrate (LiNO ) is recommended.

NOTE 1 Sodium nitrate can also be used as an oxidizing agent. This will render the analysis of sodium

impossible.

NOTE 2 Typical mass of the oxidizing agent added is 0,5 units to 1 unit by mass for every unit of sample.

4.5 Certified reference materials

CRMs prepared in accordance with 5.2 may be used to establish calibration and to validate calibration.

4.6 Reference materials

RMs can be materials, e.g. iron ores, homogenized and prepared by a laboratory. The reference analysis

of a RM shall be the average result from interlaboratory co-operative testing involving at least four

laboratories able to meet the performance criteria.
5 Sampling and samples
5.1 Laboratory sample

For analysis, use a laboratory sample of less than 100 μm particle size which has been taken and

prepared in accordance with ISO 3082. In the case of ores having significant contents of combined

water or easily oxidizable compounds, use a particle size of less than 160 μm.
© ISO 2021 – All rights reserved
---------------------- Page: 8 ----------------------
ISO/TS 9516-4:2021(E)
5.2 Preparation of test samples, CRMs and RMs
5.2.1 General
Depending on the ore type, proceed in accordance with either 5.2.2 or 5.2.3.

5.2.2 Ores having significant contents of combined water or easily oxidizable compounds

Prepare an air-equilibrated test sample in accordance with ISO 2596 with the following types of ore:

a) processed ores containing metallic iron;

b) natural or processed ores in which the non-oxidized sulfur content is higher than 0,2 % mass

fraction;

c) natural or processed ores in which the content of combined water is higher than 2,5 % mass

fraction.
NOTE Loss on ignition can be used as an estimate of combined water.
5.2.3 Ores outside the scope of 5.2.2

Prepare a pre-dried test sample by thoroughly mixing the laboratory sample and, taking multiple

increments, extracting a test sample in such a manner that it is representative of the whole contents of

the container. The pre-dried test portion shall be prepared in accordance with ISO 7764.

5.3 Test portion

Thoroughly mix the laboratory sample and, taking multiple increments, extract a test sample in such a

manner that it is representative of the whole contents of the container. One disc from each test sample

shall be prepared.

At least one CRM, of the same type (mineralogy and chemistry) as the ore used in the test discs, should

be prepared and analysed.
6 Apparatus
The usual laboratory apparatus and, in particular, the following shall be used.
6.1 Analytical balance, capable of weighing to four decimal places.

6.2 Crucibles and moulds. Articles made of platinum alloy with certified purity. The recommended

alloys for making the pieces are platinum, platinum/gold, rhodium or platinum/rhodium.

The amount of sample and flux used should respect the volume of the crucible and mould available. It

is important to take into account possible projections caused by the movement of the fusion machines

during the fusion process.

The surface of the moulds should be perfectly flat throughout the life of the article. Regular polishing

and checking the bottom alignment are recommended.
6.3 Fusion equipment.

Any gas heating, electric heating or induction heating equipment may be used. Equipment shall be

capable of maintaining a temperature of at least 1 050 °C with capacity to vary temperature, time and

agitation intensity.
© ISO 2021 – All rights reserved
---------------------- Page: 9 ----------------------
ISO/TS 9516-4:2021(E)

The temperature shall be checked regularly with an optical pyrometer. It is important that during these

measurements, the crucible contains a mass of flux similar to the mass that is used in the analysis.

In machines with more than one independent melting position, it is necessary to check their temperature

homogeneity. Burners on gas machines and chambers in electrical machines shall maintain temperature

stability during the melting process.

For gas burners, to minimize the loss of sulfur during the fusion, provisions shall be made to add oxygen

to the flame. Alternatively, oxidizing aids can be added to the flux, see 4.4.
6.4 Electric furnace, capable of maintaining a temperature of at least 1 050 °C.
6.5 Gas burner.

When gas burners are used the temperature of the melt shall be in the range of 1 000 °C to 1 050 °C.

To minimize the loss of sulfur during the fusion, provisions shall be made to add oxygen to the flame.

Alternatively, oxidising aids can be added to the flux, see 4.4.

The temperature shall be checked regularly with an optical pyrometer. It is important that during these

measurements, the crucible contains a mass of flux similar to the mass that is used in the analysis.

6.6 Fusion machine, with gas heating station(s) or induction furnaces, normally with agitation

capacity and which allows the creation of fusion programmes with variation of time, agitation intensity

and temperature. In machines with more than one independent melting position, it is necessary to check

their temperature homogeneity. Burners, on gas machines and chamber in resistances on electrical

machines, shall maintain temperature stability during the melting process.
6.7 XRF spectrometer.

Any wavelength dispersive XRF spectrometer equipped with a vacuum path, provided that the

instrument has been checked. Performance checks can be carried out in accordance with ISO 9516-1,

accumulating at least 4 × 10 counts for each measurement. At this number of counts, the counting

statistical error will be limited to about 0,016 %. This can depend on the count rate. In case of doubt,

check with the instrument manufacturer. See also 7.1.3.

Although this method is written for WDXRF equipment, its principles may also be applied to the use of

EDXRF instrumentation.

Modern instruments generally have some form of dead time correction, and although the correction

is not always perfect, it should be acceptable for this method, which is tolerant of small errors in dead

time correction.
7 Measurements
7.1 General
7.1.1 Analytical line

For the analyte elements Fe, Si, Ca, Mn, Al, Ti, Mg, P, S and K, the Kα (K-L ) is recommended. For iron,

1,2 2,3

the Kα is the preferred line, provided the count rate is well within the dynamic range of the detector

1,2

system used. If the count rate is too high, then the Kβ (K-M ) can be used. If analytes with an atomic

1,3 2,3

number higher than 52 are analysed, the preferred line is the Lα (L -M ). For the determination of

1 3 5

lead (if included), the Lβ (L -M ) should be used. Furthermore, if both As and Pb are analysed, then the

1 2 4

interference of the Pb Lα (L -M ) on As Kα (K-L ) shall be taken into account or arsenic shall be

1 3 5 1,2 2,3

determined using the As Kβ (K-M ) line. If barium is analysed, then either the Kα (or K-L ), the

1,3 2,3 1,2 2,3

Lα (L -M ) or the Lβ (L -M ) can be used, depending upon which gives the best results.

1 3 5 1 2 4
© ISO 2021 – All rights reserved
---------------------- Page: 10 ----------------------
ISO/TS 9516-4:2021(E)
7.1.2 Voltage and current

Voltages of 50 kV to 60 kV are recommended for the measurements of all the analytes with atomic

number 22 or higher. For the other analytes, voltages between 25 kV and 40 kV can be used. It is

highly recommended to adjust the current for each of these voltage settings in such a way that all the

measurements are done at constant power of the generator.

For the best possible stability, it is advisable to work at the same constant voltage and constant current

for all analytes. Tube voltages of around 40 kV to 50 kV are adequate.
7.1.3 Measuring times

The measuring time in combination with the count rate determine the counting statistical error. The

measuring time should be chosen in such a way that the counting statistical error is lower than 0,5 %

for Si, Ca, Ti and Al. This requirement can typically be met by accumulating at least 40 000 counts. For

Fe, the counting statistical error should be at most 0,016 %. At low count rates, 4 × 10 counts need to

be accumulated.

The counting statistical error for a detection system with a given dead time and for a given number

of counts accumulated increases with increasing count rate. The effect becomes noticeable when the

product of the dead time (expressed in seconds) and the count rate (expressed in counts per second)

is 0,15 or higher. In that case, longer measuring times shall be used. At count rates below 300 000

counts per second, the effect can be neglected in most modern instrumentation. Consult the instrument

manufacturer for more detail and guidance.
7.1.4 Background measurements

Measurements of the background are not required. However, if elements are not measured at trace

levels, background measurements can improve the accuracy of determination at trace levels.

8 Calibration and validation
8.1 Principles

The calibration equations and inter-element corrections are established using calibration standards

produced using fused beads made from pure reagents (or combinations thereof), CRMs, RMs or any

combination of these. The validity of the calibration is confirmed by analysing one or more CRMs,

representative of the range of an
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.