Iron ores — Determination of various elements by X-ray fluorescence spectrometry — Part 4: Performance-based method using fusion preparation method

This document specifies a performance-based method for the chemical analysis of natural and processed iron ores by fused bead wavelength and energy dispersive X-ray fluorescence (XRF). It is applicable to all elements of interest when adequate calibrations have been established.

Titre manque — Partie 4: Titre manque

General Information

Status
Published
Publication Date
29-Nov-2021
Current Stage
6060 - International Standard published
Start Date
30-Nov-2021
Due Date
10-Mar-2022
Completion Date
30-Nov-2021
Ref Project

Relations

Buy Standard

Technical specification
ISO/TS 9516-4:2021 - Iron ores -- Determination of various elements by X-ray fluorescence spectrometry
English language
12 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

TECHNICAL ISO/TS
SPECIFICATION 9516-4
First edition
2021-11
Iron ores — Determination of various
elements by X-ray fluorescence
spectrometry —
Part 4:
Performance-based method using
fusion preparation method
Reference number
ISO/TS 9516-4:2021(E)
© ISO 2021

---------------------- Page: 1 ----------------------
ISO/TS 9516-4:2021(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2021
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
  © ISO 2021 – All rights reserved

---------------------- Page: 2 ----------------------
ISO/TS 9516-4:2021(E)
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Reagents and materials . 2
4.1 Pure reagents . 2
4.2 Flux . 3
4.3 Releasing agent . 3
4.4 Oxidizing agents . 3
4.5 Certified reference materials . 3
4.6 Reference materials . 3
5 Sampling and samples . .3
5.1 Laboratory sample . 3
5.2 Preparation of test samples, CRMs and RMs . 4
5.2.1 General . 4
5.2.2 Ores having significant contents of combined water or easily oxidizable
compounds . 4
5.2.3 Ores outside the scope of 5.2.2 . 4
5.3 Test portion . 4
6 Apparatus . 4
7 Measurements .5
7.1 General . 5
7.1.1 Analytical line . 5
7.1.2 Voltage and current . 6
7.1.3 Measuring times . 6
7.1.4 Background measurements. 6
8 Calibration and validation .6
8.1 Principles . 6
8.2 Preparation of fusion beads . 6
8.3 Calibration and validation samples . 6
8.4 Validation of the calibration . 7
8.4.1 Specimens . 7
8.4.2 Trueness validation . . 7
8.5 Calibration maintenance . 8
8.5.1 Monitor disc . 8
8.5.2 Duplicate monitor discs . 8
8.5.3 Monitor measurements . 8
8.5.4 Quality control measurements . 9
9 Reporting .10
9.1 General . 10
9.2 Calculation of results . 10
9.3 Number of decimals . 11
10 Test report .11
Bibliography .12
iii
© ISO 2021 – All rights reserved

---------------------- Page: 3 ----------------------
ISO/TS 9516-4:2021(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO’s adherence to
the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 102, Iron ore and direct reduced iron,
Subcommittee SC 2 Chemical analysis.
A list of all parts in the ISO 9516 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
  © ISO 2021 – All rights reserved

---------------------- Page: 4 ----------------------
ISO/TS 9516-4:2021(E)
Introduction
X-ray fluorescence (XRF) spectrometry is a multi-elemental non-destructive analytical methodology
used for quantitative and qualitative determinations. It is element specific covering an elemental range
from boron (B) to uranium (U).
Once the sample has been dissolved into a borate glass it may be introduced to the spectrometer for
analysis. The sample is then irradiated by intense radiation from an X-ray tube. Analysis of fused glass
beads offers advantages over pressed powder techniques as it eliminates particle size effects, and it
produces a homogeneous specimen for each element.
In some instances, the relationship between intensity (or intensity ratios) and concentration can
be linear. For most analytes there is no direct straightforward relationship between intensity and
concentration. With samples of differing compositions, the X-rays are absorbed differently in the
different samples giving rise to what are generally referred to as matrix effects. These inter-element
effects can be corrected using mathematical models derived from the known physics of X-rays.
Calibration can be based on binary standards (prepared from pure oxides or liquid solutions), reference
materials, secondary standards, or combinations therewith.
v
© ISO 2021 – All rights reserved

---------------------- Page: 5 ----------------------
TECHNICAL SPECIFICATION ISO/TS 9516-4:2021(E)
Iron ores — Determination of various elements by X-ray
fluorescence spectrometry —
Part 4:
Performance-based method using fusion preparation
method
WARNING — This document can involve hazardous materials, operations and equipment. This
document does not purport to address all of the safety problems associated with its use. It is the
responsibility of the user of this document to establish appropriate health and safety practices
and determine the applicability of regulatory limitations prior to use.
1 Scope
This document specifies a performance-based method for the chemical analysis of natural and
processed iron ores by fused bead wavelength and energy dispersive X-ray fluorescence (XRF).
It is applicable to all elements of interest when adequate calibrations have been established.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 2596, Iron ores — Determination of hygroscopic moisture in analytical samples — Gravimetric, Karl
Fischer and mass-loss methods
ISO 3082, Iron ores — Sampling and sample preparation procedures
ISO 7764, Iron ores — Preparation of predried test samples for chemical analysis
ISO Guide 31, Reference materials — Contents of certificates, labels and accompanying documentation
ISO Guide 35, Reference materials — Guidance for characterization and assessment of homogeneity and
stability
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at http:// www .electropedia .org/
3.1
primary standard
standard that is designated or widely acknowledged as having the highest metrological qualities and
whose value is accepted without reference to other standards of the same quantity
Note 1 to entry: The concept of a primary standard is equally valid for base quantities and derived quantities.
1
© ISO 2021 – All rights reserved

---------------------- Page: 6 ----------------------
ISO/TS 9516-4:2021(E)
Note 2 to entry: A primary standard is never used directly for measurement other than for comparison with
other primary standards or reference standards.
3.2
reference material
RM
standard, generally having the highest metrological quality available at a given location or in a given
organization, from which measurements made there are derived
3.3
certified reference material
CRM
material, e.g. iron ores, supplied by an organization conforming to the requirements for the competence
of reference material (3.2) producers
Note 1 to entry: The requirements for the competence of reference material producers are given in ISO 17034.
Note 2 to entry: CRMs shall be supplied with a certificate of analysis giving information on the average value
and standard deviation (with and between laboratory precision statistics) and measurement of uncertainty in
accordance with ISO Guide 31.
3.4
accepted reference value
ARV
value that serves as an agreed upon reference for comparison, and which is derived as:
a) a theoretical or established value, based on scientific principles;
b) an assigned or certified value, based on experimental work of some national or international
organization;
c) a consensus or certified value, based on collaborative experimental work under the auspices of a
scientific or engineering group.
Note 1 to entry: When none of the above are available, the ARV is the expectation of the (measurable) quantity,
i.e. the mean of a specified population of measurements.
3.5
referee method
method that is independent of other methods (i.e. calibrated
with primary standards or able to arrive at the final results using direct measurements or calculations
from known physical/chemical laws) and is in the initial stages of the traceability chain
Note 1 to entry: Primary methods include gravimetry, titrimetry, coulometry and isotope dilution mass
spectrometry.
4 Reagents and materials
4.1 Pure reagents
Reagents shall be of analytical quality and, wherever possible, be pure oxides or carbonates, except for
the calibration of such elements as sulfur, chlorine, bromine, or phosphorus, which do not form stable
oxides or carbonates, where some guarantee of stoichiometry is required.
Reagents shall be free of (or corrected for) the presence of moisture (and, in the case of oxides, carbon
dioxide) when weighed out for fusion.
Reagents shall be used in a known stoichiometry in terms of content. In order to achieve this, they
can be treated before use. Generally, the oxides of iron, silicon, manganese, aluminium, titanium and
magnesium shall be heated to 1 000 °C. More information can be found in ISO 9516-1. The procedures
specified in ISO 9516-1 ensure that the correct oxidation state is obtained.
2
  © ISO 2021 – All rights reserved

---------------------- Page: 7 ----------------------
ISO/TS 9516-4:2021(E)
All laboratory reagents used for referee methods shall be traceable.
Where reagents have been ignited, they should be covered during cooling in the desiccator and weighed
as soon as possible.
4.2 Flux
High purity lithium borate or sodium borate fluxes should be used. Prior to using, the levels of
contamination shall be checked. Good results have been obtained with a mixture that contains 6 parts
of Li B O and 11 parts of LiBO by mass. This composition is a eutectic and can be used at relatively
2 4 7 2
low temperatures.
NOTE 1 This mixture is commercially available under the designation 12:22.
NOTE 2 A typical sample to flux ratio is 1 unit of sample for 10 units of flux.
4.3 Releasing agent
To facilitate the casting of the sample-flux mixture and the releasing of the resultant bead after
cooling, a releasing agent such as ammonium iodide (NH I) or lithium iodide (LiI) is recommended.
4
Alternatively, lithium bromide (LiBr) can be used.
If a bromide-containing releasing agent is used, the line overlap of the Br L-lines with Al Kα shall be
taken into account.
NOTE Fluxes with an integrated releasing agent are commercially available.
4.4 Oxidizing agents
Generally, no oxidizing agents are required if the iron is present as hematite (Fe O ). However, if the
2 3
samples to be analysed contain, for example, magnetite (Fe O ) or if a gas burner is used without
3 4
oxygen supplement, then the addition of an oxidizing agent to the flux can be required. In those cases,
the use of lithium nitrate (LiNO ) is recommended.
3
NOTE 1 Sodium nitrate can also be used as an oxidizing agent. This will render the analysis of sodium
impossible.
NOTE 2 Typical mass of the oxidizing agent added is 0,5 units to 1 unit by mass for every unit of sample.
4.5 Certified reference materials
CRMs prepared in accordance with 5.2 may be used to establish calibration and to validate calibration.
4.6 Reference materials
RMs can be materials, e.g. iron ores, homogenized and prepared by a laboratory. The reference analysis
of a RM shall be the average result from interlaboratory co-operative testing involving at least four
laboratories able to meet the performance criteria.
5 Sampling and samples
5.1 Laboratory sample
For analysis, use a laboratory sample of less than 100 μm particle size which has been taken and
prepared in accordance with ISO 3082. In the case of ores having significant contents of combined
water or easily oxidizable compounds, use a particle size of less than 160 μm.
3
© ISO 2021 – All rights reserved

---------------------- Page: 8 ----------------------
ISO/TS 9516-4:2021(E)
5.2 Preparation of test samples, CRMs and RMs
5.2.1 General
Depending on the ore type, proceed in accordance with either 5.2.2 or 5.2.3.
5.2.2 Ores having significant contents of combined water or easily oxidizable compounds
Prepare an air-equilibrated test sample in accordance with ISO 2596 with the following types of ore:
a) processed ores containing metallic iron;
b) natural or processed ores in which the non-oxidized sulfur content is higher than 0,2 % mass
fraction;
c) natural or processed ores in which the content of combined water is higher than 2,5 % mass
fraction.
NOTE Loss on ignition can be used as an estimate of combined water.
5.2.3 Ores outside the scope of 5.2.2
Prepare a pre-dried test sample by thoroughly mixing the laboratory sample and, taking multiple
increments, extracting a test sample in such a manner that it is representative of the whole contents of
the container. The pre-dried test portion shall be prepared in accordance with ISO 7764.
5.3 Test portion
Thoroughly mix the laboratory sample and, taking multiple increments, extract a test sample in such a
manner that it is representative of the whole contents of the container. One disc from each test sample
shall be prepared.
At least one CRM, of the same type (mineralogy and chemistry) as the ore used in the test discs, should
be prepared and analysed.
6 Apparatus
The usual laboratory apparatus and, in particular, the following shall be used.
6.1 Analytical balance, capable of weighing to four decimal places.
6.2 Crucibles and moulds. Articles made of platinum alloy with certified purity. The recommended
alloys for making the pieces are platinum, platinum/gold, rhodium or platinum/rhodium.
The amount of sample and flux used should respect the volume of the crucible and mould available. It
is important to take into account possible projections caused by the movement of the fusion machines
during the fusion process.
The surface of the moulds should be perfectly flat throughout the life of the article. Regular polishing
and checking the bottom alignment are recommended.
6.3 Fusion equipment.
Any gas heating, electric heating or induction heating equipment may be used. Equipment shall be
capable of maintaining a temperature of at least 1 050 °C with capacity to vary temperature, time and
agitation intensity.
4
  © ISO 2021 – All rights reserved

---------------------- Page: 9 ----------------------
ISO/TS 9516-4:2021(E)
The temperature shall be checked regularly with an optical pyrometer. It is important that during these
measurements, the crucible contains a mass of flux similar to the mass that is used in the analysis.
In machines with more than one independent melting position, it is necessary to check their temperature
homogeneity. Burners on gas machines and chambers in electrical machines shall maintain temperature
stability during the melting process.
For gas burners, to minimize the loss of sulfur during the fusion, provisions shall be made to add oxygen
to the flame. Alternatively, oxidizing aids can be added to the flux, see 4.4.
6.4 Electric furnace, capable of maintaining a temperature of at least 1 050 °C.
6.5 Gas burner.
When gas burners are used the temperature of the melt shall be in the range of 1 000 °C to 1 050 °C.
To minimize the loss of sulfur during the fusion, provisions shall be made to add oxygen to the flame.
Alternatively, oxidising aids can be added to the flux, see 4.4.
The temperature shall be checked regularly with an optical pyrometer. It is important that during these
measurements, the crucible contains a mass of flux similar to the mass that is used in the analysis.
6.6 Fusion machine, with gas heating station(s) or induction furnaces, normally with agitation
capacity and which allows the creation of fusion programmes with variation of time, agitation intensity
and temperature. In machines with more than one independent melting position, it is necessary to check
their temperature homogeneity. Burners, on gas machines and chamber in resistances on electrical
machines, shall maintain temperature stability during the melting process.
6.7 XRF spectrometer.
Any wavelength dispersive XRF spectrometer equipped with a vacuum path, provided that the
instrument has been checked. Performance checks can be carried out in accordance with ISO 9516-1,
7
accumulating at least 4 × 10 counts for each measurement. At this number of counts, the counting
statistical error will be limited to about 0,016 %. This can depend on the count rate. In case of doubt,
check with the instrument manufacturer. See also 7.1.3.
Although this method is written for WDXRF equipment, its principles may also be applied to the use of
EDXRF instrumentation.
Modern instruments generally have some form of dead time correction, and although the correction
is not always perfect, it should be acceptable for this method, which is tolerant of small errors in dead
time correction.
7 Measurements
7.1 General
7.1.1 Analytical line
For the analyte elements Fe, Si, Ca, Mn, Al, Ti, Mg, P, S and K, the Kα (K-L ) is recommended. For iron,
1,2 2,3
the Kα is the preferred line, provided the count rate is well within the dynamic range of the detector
1,2
system used. If the count rate is too high, then the Kβ (K-M ) can be used. If analytes with an atomic
1,3 2,3
number higher than 52 are analysed, the preferred line is the Lα (L -M ). For the determination of
1 3 5
lead (if included), the Lβ (L -M ) should be used. Furthermore, if both As and Pb are analysed, then the
1 2 4
interference of the Pb Lα (L -M ) on As Kα (K-L ) shall be taken into account or arsenic shall be
1 3 5 1,2 2,3
determined using the As Kβ (K-M ) line. If barium is analysed, then either the Kα (or K-L ), the
1,3 2,3 1,2 2,3
Lα (L -M ) or the Lβ (L -M ) can be used, depending upon which gives the best results.
1 3 5 1 2 4
5
© ISO 2021 – All rights reserved

---------------------- Page: 10 ----------------------
ISO/TS 9516-4:2021(E)
7.1.2 Voltage and current
Voltages of 50 kV to 60 kV are recommended for the measurements of all the analytes with atomic
number 22 or higher. For the other analytes, voltages between 25 kV and 40 kV can be used. It is
highly recommended to adjust the current for each of these voltage settings in such a way that all the
measurements are done at constant power of the generator.
For the best possible stability, it is advisable to work at the same constant voltage and constant current
for all analytes. Tube voltages of around 40 kV to 50 kV are adequate.
7.1.3 Measuring times
The measuring time in combination with the count rate determine the counting statistical error. The
measuring time should be chosen in such a way that the counting statistical error is lower than 0,5 %
for Si, Ca, Ti and Al. This requirement can typically be met by accumulating at least 40 000 counts. For
7
Fe, the counting statistical error should be at most 0,016 %. At low count rates, 4 × 10 counts need to
be accumulated.
The counting statistical error for a detection system with a given dead time and for a given number
of counts accumulated increases with increasing count rate. The effect becomes noticeable when the
product of the dead time (expressed in seconds) and the count rate (expressed in counts per second)
is 0,15 or higher. In that case, longer measuring times shall be used. At count rates below 300 000
counts per second, the effect can be neglected in most modern instrumentation. Consult the instrument
manufacturer for more detail and guidance.
7.1.4 Background measurements
Measurements of the background are not required. However, if elements are not measured at trace
levels, background measurements can improve the accuracy of determination at trace levels.
8 Calibration and validation
8.1 Principles
The calibration equations and inter-element corrections are established using calibration standards
produced using fused beads made from pure reagents (or combinations thereof), CRMs, RMs or any
combination of these. The validity of the calibration is confirmed by analysing one or more CRMs,
representative of the range of an
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.