Fire tests — Calibration and use of radiometers and heat flux meters — Part 1: General principles

ISO/TS 14934-1:2002 gives guidelines for calibration and use of radiometers and heat flux meters in fire testing and for correction of the sensitivity function due to convection effects. It briefly describes the calibration methods, the most commonly used types of radiometers and heat flux meters, and the fire tests in which these transducers are used. ISO/TS 14934-1:2002 is applicable to total hemispherical radiometers, total heat flux meters of Schmidt-Boelter (thermopile) and Gardon (foil) type. It applies only to instruments having plane receivers and does not apply to receivers in the form of wires, spheres, etc.

Essais au feu — Étalonnage des appareils de mesure du flux rayonné et du flux thermique — Partie 1: Principes généraux

General Information

Status
Withdrawn
Publication Date
10-Dec-2002
Withdrawal Date
10-Dec-2002
Current Stage
9599 - Withdrawal of International Standard
Completion Date
10-Jun-2010
Ref Project

Relations

Buy Standard

Technical specification
ISO/TS 14934-1:2002 - Fire tests -- Calibration and use of radiometers and heat flux meters
English language
20 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

TECHNICAL ISO/TS
SPECIFICATION 14934-1
First edition
2002-12-15

Reaction-to-fire tests — Calibration and
use of radiometers and heat flux
meters —
Part 1:
General principles
Essais de réaction au feu — Étalonnage des appareils de mesure du
flux rayonné et du flux thermique —
Partie 1: Principes généraux




Reference number
ISO/TS 14934-1:2002(E)
©
ISO 2002

---------------------- Page: 1 ----------------------
ISO/TS 14934-1:2002(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


©  ISO 2002
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2002 — All rights reserved

---------------------- Page: 2 ----------------------
ISO/TS 14934-1:2002(E)
Contents Page
Foreword. iv
Introduction . v
1 Scope. 1
2 Normative references . 1
3 Terms and definitions. 2
4 Principle . 3
4.1 General. 3
4.2 Principles of measuring radiant heat flux . 3
4.3 Principles of primary calibration of a heat flux meter. 4
4.3.1 General. 4
4.3.2 Principles of primary calibration apparatus “VBBC” of BNM-LNE. 4
4.3.3 Principles of primary calibration apparatus NT FIRE 050 at SP. 5
4.3.4 Principles of primary calibration apparatus VTBB at NIST . 5
4.4 Principles of secondary calibration of a heat flux meter. 6
4.5 Principles of using total heat flux meters to set the radiant heat flux in a fire test method . 7
5 Primary calibration methods for radiometers and total heat flux meters . 7
5.1 Requirements of a primary radiation calibration. 7
5.2 Primary calibration apparatus “VBBC” of BNM-LNE — France [1] . 8
5.2.1 General. 8
5.2.2 Calibration procedure. 8
5.3 Primary calibration apparatus NT FIRE 050 at SP — Sweden [2] . 8
5.4 Primary calibration apparatus “VTBB” at NIST — USA [3]. 8
6 Secondary calibration method for radiometers and total heat flux meters . 9
7 Use of total heat flux meters to set/measure the radiant heat flux in fire test methods . 10
7.1 General. 10
7.2 ISO 5657 ignitability test. 10
7.3 ISO 5659-2 smoke density chamber and ISO 5660-1 cone calorimeter test . 11
7.4 ISO 5658-2 and IMO Resolution A.653 spread of flame test and EN ISO 9239-1 radiant
panel test for floorings . 11
7.5 ISO/TR 14696 intermediate scale calorimeter . 13
Annex A (informative) Description of radiometers and heat flux meters . 14
Annex B (informative) Heat flux measurements in fire test methods . 17
Bibliography . 20

© ISO 2002 — All rights reserved iii

---------------------- Page: 3 ----------------------
ISO/TS 14934-1:2002(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
In other circumstances, particularly when there is an urgent market requirement for such documents, a
technical committee may decide to publish other types of normative document:
 an ISO Publicly Available Specification (ISO/PAS) represents an agreement between technical experts in
an ISO working group and is accepted for publication if it is approved by more than 50 % of the members
of the parent committee casting a vote;
 an ISO Technical Specification (ISO/TS) represents an agreement between the members of a technical
committee and is accepted for publication if it is approved by 2/3 of the members of the committee casting
a vote.
An ISO/PAS or ISO/TS is reviewed after three years in order to decide whether it will be confirmed for a
further three years, revised to become an International Standard, or withdrawn. If the ISO/PAS or ISO/TS is
confirmed, it is reviewed again after a further three years, at which time it must either be transformed into an
International Standard or be withdrawn.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO/TS 14934-1 was prepared by Technical Committee ISO/TC 92, Fire safety, Subcommittee SC 1, Fire
initiation and growth.
ISO/TS 14934 consists of the following parts, under the general title Reaction-to-fire tests — Calibration and
use of radiometers and heat flux meters:
 Part 1: General principles
 Part 2: Primary calibration
 Part 3: Secondary calibration
 Part 4: Guidance on the use of heat flux meters in fire tests
iv © ISO 2002 — All rights reserved

---------------------- Page: 4 ----------------------
ISO/TS 14934-1:2002(E)
Introduction
Radiant heat transfer is an important mode of fire spread, particularly in large fires with flames and hot gas
layer thickness larger than 1 m. To represent optimally realistic scenarios, many fire test methods specify the
radiation level. Therefore, it is of great importance in fire safety engineering and in fire testing that the radiant
heat flux be measured with sufficient accuracy (see 5.1).
In practice, radiant heat flux is usually measured with total heat flux meters of the Schmidt-Boelter
(thermopile) or Gardon (foil) type. Such meters register the combined heat flux from radiation and convection.
This introduces an uncertainty, as the measured heat flux will contain an unknown contribution from the
convection heat transfer. The actual contribution due to convection, in calibrations and fire tests, will depend
upon a number of factors such as the design of the heat flux meter, the orientation of the meter, the cooling
water temperature, the temperature and flow conditions close to the meter, and the calibration method. In
many practical situations, the uncertainty in the convection can amount to 25 % of the total heat flux measured.
To overcome the difficulties with the convection influence, a calibration procedure is outlined where primary
calibration is performed on two different types of heat flux meters:
a) total hemispherical radiometer or a cavity radiometer which is sensitive only to radiation; and
b) total heat flux meter, as is typically used, which detects both modes of heat transfer.
Where possible, an effort should be made to minimize the convective influence. In all calibrations and
measurements of radiative heat flux, the uncertainty calculations should include the uncertainty due to the
residual convective component. For secondary calibration methods, a combined use of hemispherical
radiometers and total heat flux meters makes it possible to estimate the convection contribution. The same
arrangement can be used in calibration of fire test methods.
Primary calibration is performed in fully characterized blackbody facilities, with total combined expanded
uncertainty of less than ± 3,0 % with a 95 % confidence level, in the measured heat flux. One such facility is
an evacuated blackbody with the unique characteristic of negligible convection and conduction effects on
calibration. Other non-evacuated blackbody facilities are also suitable to be primary radiative flux calibration
sources, provided that they are fully characterized, including any convection effects, and the combined
expanded uncertainty is less than ± 3,0 %.
It should be noted that the wavelength spectrum and angular distribution of the radiation from a fire may be
different from that of a blackbody source. This may introduce extra sources of error to the combined expanded
uncertainty when a heat flux meter is used.
In this Technical Specification, three different methods of calibrations using blackbody radiation sources are
proposed for provisional evaluation. The objective of this evaluation phase, expected to last about three years,
is to determine the relative merits and limitations of the methods and the associated total combined
uncertainty. The results and the operational experience gained during the evaluation phase will be reviewed to
recommend a suitable test standard.
Within the ongoing European project “Improving heat flux meter calibration for fire testing laboratories HFCAL”
SMT4-CT98-2266, total heat flux meters of the Schmidt-Boelter or Gardon type and a total hemispherical
radiometer of the Gunners type will be characterized with respect to wavelength, geometry and convection.
Different types of emissivity coatings will be investigated. Calibration results of two of the primary calibration
methods described in this Technical Specification, the LNE vacuum blackbody cavity (VBBC) [1], and the
NT FIRE 050 [2], and of secondary calibration methods will be compared in a round robin test.

© ISO 2002 — All rights reserved v

---------------------- Page: 5 ----------------------
TECHNICAL SPECIFICATION ISO/TS 14934-1:2002(E)

Reaction-to-fire tests — Calibration and use of radiometers and
heat flux meters —
Part 1:
General principles
1 Scope
This Technical Specification gives guidelines for calibration and use of radiometers and heat flux meters in fire
testing and for correction of the sensitivity function due to convection effects.
It briefly describes the calibration methods, the most commonly used types of radiometers and heat flux
meters, and the fire tests in which these transducers are used.
This Technical Specification is applicable to total hemispherical radiometers, total heat flux meters of Schmidt-
Boelter (thermopile) and Gardon (foil) type. It applies only to instruments having plane receivers and does not
apply to receivers in the form of wires, spheres, etc.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 5657:1997, Reaction to fire tests — Ignitability of building products using a radiant heat source
ISO 5658-2:1996, Reaction to fire tests — Spread of flame — Part 2: Lateral spread on building products in
vertical configuration
ISO 5659-2:1994, Plastics — Smoke generation — Part 2: Determination of optical density by a single-
chamber test
ISO 5660-1:2002, Reaction-to-fire tests — Heat release, smoke production and mass loss rate — Part 1: Heat
release rate (cone calorimeter method)
EN ISO 9239-1:2002, Reaction to fire tests for floorings — Part 1: Determination of the burning behaviour
using a radiant heat source
EN ISO 13943:2000, Fire safety — Vocabulary
ISO/TR 14696:1999, Reaction to fire tests — Determination of fire parameters of materials, products and
assemblies using an intermediate-scale heat release calorimeter (ICAL)
VIM, International vocabulary of basic and general terms in metrology, BIPM/IEC/IFCC/ISO/IUPAC/IUPAP/OIML,
ISBN 92-67-01075-1
GUM, Guide to the expression of uncertainty in measurement, BIPM/IEC/IFCC/ISO/IUPAC/IUPAP/OIML,
ISBN 92-67-10188-9
© ISO 2002 — All rights reserved 1

---------------------- Page: 6 ----------------------
ISO/TS 14934-1:2002(E)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 13943, VIM, GUM and the
following apply.
3.1
blackbody radiant source
radiating source which is designed to absorb all radiation incident upon it and not reflect any radiation
NOTE The emissivity of an ideal blackbody radiant source is unity.
3.2
calorimeter
apparatus that measures heat by detecting the change in its body temperature over time
3.3
convection
transmission of heat by a surrounding fluid involving movement of the fluid
3.4
emissivity
ratio of the radiation emitted by a surface to the radiation emitted by a perfect blackbody radiator at the same
temperature
3.5
heat flow rate
energy per unit time
3.6
irradiance (at a point on a surface)
amount of radiant power per unit area that flows across or onto a surface
3.7
primary standard
absolute standard to which all other calibrated measuring instruments can be traced
3.8
radiant heat flux
power (energy per unit time) per unit area emitted, transferred or received in the form of heat radiation
3.9
radiative heat transfer
transmission of heat by electromagnetic radiation
3.10
radiation
emission and propagation of electromagnetic waves through space or some medium
3.11
radiometer
transducer (instrument) that converts radiant heat flux into an electrical signal
3.12
radiosity
rate at which radiant energy leaves a surface by combined emission and reflection of radiation
3.13
secondary standard
standard instrument with a calibration traceable to the primary standard
2 © ISO 2002 — All rights reserved

---------------------- Page: 7 ----------------------
ISO/TS 14934-1:2002(E)
3.14
sensing surface
active part of the transducer which detects the heat flux through the surface
3.15
sensitivity (of a radiometer or a total heat flux meter)
ratio of the output voltage to the irradiance in the plane of the receiver
3.16
total heat flux
total amount of heat flow rate per unit area incident on a surface which includes both radiative and convective
heat transfer
3.17
total heat flux meter
transducer (instrument) that responds to both radiative and convective heat transfer
3.18
total hemispherical radiometer
radiometer that responds only to radiative heat transfer with an acceptance angle approaching 180°
4 Principle
4.1 General
All heat flux meters for daily use in fire testing can be calibrated either by using a blackbody radiant source
(primary calibration) or by using a transfer calibration method (secondary calibration), whichever provides the
desired accuracy for the user.
4.2 Principles of measuring radiant heat flux
Either a total hemispherical radiometer or a total heat flux meter can be used to measure the radiant flux
during a fire test.
Although the use of total hemispherical radiometers is not currently wide spread, their use in daily
measurement or during calibration in fire test methods has the advantage of requiring no correction for
convective effects to the measured radiant heat flux. Only the radiant heat flux is measured and a total
hemispherical radiometer may be used in any of the methods mentioned below, without the need to apply a
correction for any convective heat transfer. Caution is recommended to ensure that the angular response of
the radiometer has near-true cosine dependence.
When a total heat flux meter is used, assessment of the convective contribution shall be documented in all
stages of calibration and use. This assessment may result in modification of the procedure so as to reduce the
convective component and/or increase the uncertainty of the incident radiant flux for the calibration or
measurement. Documented characterization of the convective component using additional measurements
(e.g. sensing element temperature, local velocities and ambient gas temperatures near the sensing element)
and/or modelling may allow the isolation of the radiant component leading to reduced uncertainty in
measurements. Total heat flux meters respond to any heat that is transferred to or from the sensing surface,
and cannot distinguish between radiant or convective components of heat transfer. Hence, other means are
necessary to quantify their relative importance.
However, it may not be necessary to measure the sensitivity of the meter to convection for every single
specimen of heat flux meter. It is possible that corrections can be established for each separate type of meter
for use in a particular calibration or fire test method. Thus, the output signal of a meter to total heat flux could
be corrected to apply to only the radiant heat flux for every total heat flux meter used in that specific method.
© ISO 2002 — All rights reserved 3

---------------------- Page: 8 ----------------------
ISO/TS 14934-1:2002(E)
4.3 Principles of primary calibration of a heat flux meter
4.3.1 General
Primary calibration methods use a blackbody radiant source, such as the LNE vacuum blackbody cavity
(VBBC) [1], and the NT FIRE 050 [2], shown in Figures 1 and 2, respectively. These are defined as primary
calibration methods because the heat flux they emit can be calculated directly from their temperature, surface
characteristics and geometry, and they do not require a calibrated heat flux meter to provide a transfer
measurement. Another source used at the National Institute of Standard and Technology (NIST) is a variable
temperature blackbody (VTBB) which is a heated graphite tube [3]. Figure 3 shows a schematic layout of the
VTBB. The VTBB is used in a transfer mode to calibrate heat flux meters against an electrical substitution
radiometer (ESR).
4.3.2 Principles of primary calibration apparatus “VBBC” of BNM-LNE
The primary standard consists of a cavity giving nearly blackbody radiation (see Figure 1). The cavity is a
horizontally orientated cylinder with a diameter of 160 mm and a length of 410 mm. The blackbody cavity can
be evacuated to about 0,5 Pa by a primary roughing pump and a molecular turbo-pump. The temperature at
several positions in the cavity is measured and recorded continuously during calibration.
The blackbody cavity is electrically heated through the cylindrical wall. Four regulators, whose thermocouples
are localized close to each heater, control the heating of the blackbody. Moreover, three reflecting diaphragms
surround the heat flux meter in order to limit the losses generated by this opening. A water circuit cools the
external wall to ensure safety. The moving enclosure fits into the blackbody enclosure and is used to convey
all the measuring instruments to the blackbody cavity.
The LNE VBBC is described in 5.2.

Key
1 heat flux meter
2 reflecting diaphragms
3 evacuated blackbody cavity
Figure 1 — Schematic drawing of inner part of LNE vacuum blackbody cavity (VBBC)
4 © ISO 2002 — All rights reserved

---------------------- Page: 9 ----------------------
ISO/TS 14934-1:2002(E)
4.3.3 Principles of primary calibration apparatus NT FIRE 050 at SP
In NT FIRE 050, natural convection from the surface of the sensing element is minimized by mounting the
meter vertically in an aperture at the bottom of the spherical furnace. This design encourages the cooler air
just above the sensing element to remain stationary below the heated air in the furnace. The cooler insert has
a number of shields, which protect the gauge from receiving radiation reflected from the cooler wall. They also
help to conserve the stratification of air, which reduces the convective heat transfer between the gauge and
the surroundings.
Although the convection is minimized in NT FIRE 050, it is necessary to quantify the uncertainty as a function
of flux level due to any remaining convection and other factors for each type of total heat flux meter that is
calibrated in the furnace (Gardon or Schmidt-Boelter). This uncertainty or correction is determined by
measuring the radiant heat flux in the NT FIRE 050 using a total hemispherical radiometer and the typical total
heat flux meter, that were both calibrated earlier in the LNE vacuum blackbody (VBBC). For the total heat flux
meter, it is important that cooling water flow rate and temperature are maintained at the same level during
tests in both the LNE and NT blackbody facilities. The difference in the radiant heat flux in NT FIRE 050
between the two meters, as a function of flux level, may be used to correct for the convective component
when other heat flux meters of the same type are calibrated in NT FIRE 050. It should be noted that effects of
the different view angles of the different shaped blackbodies and the different acceptance angles of the two
types of meters also need to be accounted for. NT FIRE 050 is further described in 5.3.

Key
1 blackbody cavity
2 cooler insert with flanges
3 heat flux meter
Figure 2 — Schematic drawing of NT FIRE 050
4.3.4 Principles of primary calibration apparatus VTBB at NIST
The 25 mm Variable-Temperature Blackbody is a primary facility used in radiance temperature calibrations. It
has a large aperture and is particularly suitable for calibrating heat-flux sensors. The 25 mm VTBB (Figure 3)
has been extensively used to calibrate sensors and to study problems related to calibration using blackbody
radiation. It is a thermally insulated and electrically heated graphite tube cavity. The heated tube cavity
diameter is 25 mm and the heated section is 28,2 cm long with a centre 3 mm thick partition.
© ISO 2002 — All rights reserved 5

---------------------- Page: 10 ----------------------
ISO/TS 14934-1:2002(E)
The tube end caps are water-cooled and are directly connected to the heating electrodes. The design
provides a sharp temperature gradient between the end cap and the graphite heater element. This helps in
achieving a uniform temperature distribution along the cavity length of the graphite tube. Different lengths of
graphite extension tubes can be attached to the end caps. The sensors are placed close to the blackbody exit
to achieve the highest possible heat-flux levels.
An optical pyrometer measures the blackbody temperature by sensing radiation from one end of the furnace.
A proportional-integral-differential (PID) controller regulates the power supply to maintain the furnace
temperature to within ± 0,1 K of the set value. The maximum recommended operating temperature for the
furnace is 2 973 K. The heat-flux sensors to be calibrated and the reference radiometer (ESR) are located at a
fixed distance away from the exit of the blackbody. At a distance of 12 mm from the exit, the maximum heat-
2 2
flux is approximately 50 kW/m to 60 kW/m . When calibrating at lower heat-flux levels of up to about
2
10 kW/m , the sensor and the radiometer are located at a distance of about 60 mm from the exit.

Key
1 heated graphite tube dual-cavity 6 power supply
2 water-cooled copper end caps 7 purge gas lines (argon)
3 graphite extension (not cooled) 8 transfer standard radiometer
4 control pyrometer 9 test heat flux meter
5 temperature controller/computer
Figure 3 — Schematic layout of the NIST 25 mm Variable-Temperature Blackbody (VTBB)
4.4 Principles of secondary calibration of a heat flux meter
One transfer calibration method, shown schematically in Figure 4, is given in BS 6809 [4]. Also fire test
methods such as ISO 5660 or ISO 5657 can be used as transfer calibration methods. In these methods, the
heat flux meter under calibration is compared to a total hemispherical radiometer or total heat flux meter.
Again, it is critical that convection effects be documented and minimized where possible.
In all transfer calibration methods, the total heat flux meters also measure convective heat transfer, which can
be a substantial fraction of the total heat flux. This convective component comes from two principal sources.
The first is the natural convection from the sensing surface of the total heat flux meter. This depends on the
temperature of the meter relative to the local environment. When the meter is cooled below the room
temperature, the convective component can be positive (heating). When the sensing surface of the meter is
locally heated by the incident flux, the convective component can be negative (cooling). A second convective
source can be crosscurrents that occur during the test. The local temperature of the face of the meter relative
to the local air (as well as the velocities, mounting and orientation) will determine the direction and magnitude
of the contribution. A comparison with a total hemispherical radiometer can help to establish the significance
of the convection component for each type of total heat flux meter during the transfer calibration.
6 © ISO 2002 — All rights reserved

---------------------- Page: 11 ----------------------
ISO/TS 14934-1:2002(E)
When calibration is performed, the total heat flux meter is replaced by the primary or transfer calibrated meter.
The position of the heat flux meter and the transfer meter must be identical when the record is taken.

Key
1 radiant panel
2 heat flux meter or transfer meter
Figure 4 — Schematic drawing of a transfer calibration method (mounting rack is excluded)
4.5 Principles of using total heat flux meters to set the radiant heat flux in a fire test method
When a total heat flux meter is used in a fire test method, documented characterization of the convection
component is always needed. The documented characterization shall be established for each fire test method
used by the fire testing laboratory, and for the type of total heat flux meter that is to be used for that specific
method. Further, it shall be re-established if the test conditions change due to use of a new type meter,
rebuilding of the apparatus or changes in the test environment.
The significance of the convection component may be determined by comparing the heat flux recorded by a
total hemispherical radiometer with the heat
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.