ABSTRACT
This specification applies to heat meters used to measure heat in heat exchange circuits in which energy is absorbed (cooling) or given up (heating) by a flowing liquid. For this specification, the necessary elements of a heat meter consist of a sensor to measure flow of the heat-conveying liquid, a pair of temperature sensors that measure the temperature differential across the heat exchange circuit, and a device that receives input from the flow and temperature sensors and calculates energy. This specification does not cover electrical safety or mechanical safety (including pressure safety).
This specification defines heat meters as complete or combined instruments. A complete instrument refers to a heat meter that does not have separable subassemblies, while a combined meter is a heat meter that has separable subassemblies. It specifies the allowable flow rate maximum permissible error (MPE) by accuracy class and turndown; the combined allowable error percentages for the temperature sensor pair and the heat calculator for measured difference temperatures in –17°C [2°F] increments; maximum lead cross section and length requirements; types of instruments; metrological characteristics of flow sensors of heat meters and complete instruments; data exchange and communications protocols; heat meter testing methods; operating conditions during testing; type approval tests and measurements; surge transients for signal and dc lines; surge transients for ac power lines; carrier frequencies; field strength; initial verification tests; test temperature ranges; product marking and inscriptions; and installation and operation instructions.
SCOPE
1.1 This specification defines general specifications for heat meters. Heat meters are instruments that measure heat in heat exchange circuits in which energy is absorbed (cooling) or given up (heating) by a flowing liquid.  
1.2 For this specification, the necessary elements of a heat meter consist of a sensor to measure flow of the heat-conveying liquid, a pair of temperature sensors that measure the temperature differential across the heat exchange circuit, and a device that receives input from the flow and temperature sensors and calculates energy.  
1.3 Electrical safety is not a part of this specification.  
1.4 Mechanical safety (including pressure safety) is not a part of this specification.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    24 pages
    English language
  • Technical specification
    24 pages
    English language

SCOPE
1.1 This specification defines general specifications for heat meters. Heat meters are instruments that measure heat in heat exchange circuits in which energy is absorbed (cooling) or given up (heating) by a flowing liquid.  
1.2 For this specification, the necessary elements of a heat meter consist of a sensor to measure flow of the heat-conveying liquid, a pair of temperature sensors that measure the temperature differential across the heat exchange circuit, and a device that receives input from the flow and temperature sensors and calculates energy.  
1.3 Electrical safety is not a part of this specification.  
1.4 Mechanical safety (including pressure safety) is not a part of this specification.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    24 pages
    English language

Frequently Asked Questions

E44.25 is a Technical Committee within ASTM International. It is named "Heat Metering". This committee has published 2 standards.

E44.25 develops ASTM standards in the area of Information technology. Currently, there are 2 published standards from this technical committee.

ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.

A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.

Loading...