ASTM E1125-16
(Test Method)Standard Test Method for Calibration of Primary Non-Concentrator Terrestrial Photovoltaic Reference Cells Using a Tabular Spectrum
Standard Test Method for Calibration of Primary Non-Concentrator Terrestrial Photovoltaic Reference Cells Using a Tabular Spectrum
SIGNIFICANCE AND USE
5.1 The electrical output of a photovoltaic device is dependent on the spectral content of the illumination source, its intensity, and the device temperature. To make standardized, accurate measurements of the performance of photovoltaic devices under a variety of light sources when the intensity is measured with a calibrated reference cell, it is necessary to account for the error in the short-circuit current that occurs if the relative quantum efficiency of the reference cell is not identical to the quantum efficiency of the device to be tested. A similar error occurs if the spectral irradiance distribution of the test light source is not identical to the desired reference spectral irradiance distribution. These errors are accounted for by the spectral mismatch parameter (described in Test Method E973), which is a quantitative measure of the error in the short-circuit current measurement. It is the intent of this test method to provide a recognized procedure for calibrating, characterizing, and reporting the calibration data for primary photovoltaic reference cells using a tabular reference spectrum.
5.2 The calibration of a reference cell is specific to a particular spectral irradiance distribution. It is the responsibility of the user to specify the applicable irradiance distribution, for example Tables G173. This test method allows calibration with respect to any tabular spectrum.
5.2.1 Tables G173 do not provide spectral irradiance data for wavelengths longer than 4 μm, yet pyrheliometers (see 6.1) typically have response in the 4–10 μm region. To mitigate this discrepancy, the Tables G173 spectra must be extended with the data provided in Annex A2.
5.3 A reference cell should be recalibrated at yearly intervals, or every six months if the cell is in continuous use outdoors.
5.4 Recommended physical characteristics of reference cells can be found in Specification E1040.
5.5 High-quality silicon primary reference cells are expected to be stable d...
SCOPE
1.1 This test method is intended for calibration and characterization of primary terrestrial photovoltaic reference cells to a desired reference spectral irradiance distribution, such as Tables G173. The recommended physical requirements for these reference cells are described in Specification E1040. Reference cells are principally used in the determination of the electrical performance of photovoltaic devices.
1.2 Primary photovoltaic reference cells are calibrated in natural sunlight using the relative quantum efficiency of the cell, the relative spectral distribution of the sunlight, and a tabulated reference spectral irradiance distribution. Selection of the reference spectral irradiance distribution is left to the user.
1.3 This test method requires the use of a pyrheliometer that is calibrated according to Test Method E816, which requires the use of a pyrheliometer that is traceable to the World Radiometric Reference (WRR). Therefore, reference cells calibrated according to this test method are traceable to the WRR.
1.4 This test method is used to calibrate primary reference cells; Test Method E1362 may be used to calibrate secondary and non-primary reference cells (these terms are defined in Terminology E772).
1.5 This test method applies only to the calibration of a photovoltaic cell that shows a linear dependence of its short-circuit current on irradiance over its intended range of use, as defined in Test Method E1143.
1.6 This test method applies only to the calibration of a reference cell fabricated with a single photovoltaic junction.
1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the a...
General Information
Relations
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1125 − 16 An American National Standard
Standard Test Method for
Calibration of Primary Non-Concentrator Terrestrial
1
Photovoltaic Reference Cells Using a Tabular Spectrum
This standard is issued under the fixed designation E1125; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.8 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.1 This test method is intended for calibration and charac-
responsibility of the user of this standard to establish appro-
terization of primary terrestrial photovoltaic reference cells to
priate safety, health, and environmental practices and deter-
a desired reference spectral irradiance distribution, such as
mine the applicability of regulatory limitations prior to use.
Tables G173. The recommended physical requirements for
1.9 This international standard was developed in accor-
these reference cells are described in Specification E1040.
dance with internationally recognized principles on standard-
Reference cells are principally used in the determination of the
ization established in the Decision on Principles for the
electrical performance of photovoltaic devices.
Development of International Standards, Guides and Recom-
1.2 Primary photovoltaic reference cells are calibrated in
mendations issued by the World Trade Organization Technical
natural sunlight using the relative quantum efficiency of the
Barriers to Trade (TBT) Committee.
cell, the relative spectral distribution of the sunlight, and a
tabulated reference spectral irradiance distribution. Selection
2. Referenced Documents
of the reference spectral irradiance distribution is left to the
2
user.
2.1 ASTM Standards:
E490Standard Solar Constant and Zero Air Mass Solar
1.3 Thistestmethodrequirestheuseofapyrheliometerthat
Spectral Irradiance Tables
is calibrated according to Test Method E816, which requires
E772Terminology of Solar Energy Conversion
the use of a pyrheliometer that is traceable to the World
E816Test Method for Calibration of Pyrheliometers by
Radiometric Reference (WRR). Therefore, reference cells
Comparison to Reference Pyrheliometers
calibrated according to this test method are traceable to the
E927Specification for Solar Simulation for Photovoltaic
WRR.
Testing
1.4 This test method is used to calibrate primary reference
E948Test Method for Electrical Performance of Photovol-
cells; Test Method E1362 may be used to calibrate secondary
taic Cells Using Reference Cells Under Simulated Sun-
and non-primary reference cells (these terms are defined in
light
Terminology E772).
E973Test Method for Determination of the Spectral Mis-
1.5 This test method applies only to the calibration of a
match Parameter Between a Photovoltaic Device and a
photovoltaic cell that shows a linear dependence of its short-
Photovoltaic Reference Cell
circuit current on irradiance over its intended range of use, as
E1021TestMethodforSpectralResponsivityMeasurements
defined in Test Method E1143.
of Photovoltaic Devices
E1040Specification for Physical Characteristics of Noncon-
1.6 This test method applies only to the calibration of a
centrator Terrestrial Photovoltaic Reference Cells
reference cell fabricated with a single photovoltaic junction.
E1143Test Method for Determining the Linearity of a
1.7 The values stated in SI units are to be regarded as
Photovoltaic Device Parameter with Respect To a Test
standard. No other units of measurement are included in this
Parameter
standard.
E1362Test Methods for Calibration of Non-Concentrator
Photovoltaic Non-Primary Reference Cells
1
This test method is under the jurisdiction of ASTM Committee E44 on Solar,
GeothermalandOtherAlternativeEnergySourcesandisthedirectresponsibilityof
2
Subcommittee E44.09 on Photovoltaic Electric Power Conversion. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved July 1, 2016. Published October 2016. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1986. Last previous edition approved in 2015 as E1125–10 (2015). Standards volume information, refer to the standard’s Document Summary page on
DOI: 10.1520/E1125-16. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1125 − 16
E2554Practice for Estimating and Monitoring the Uncer- 3.3.20 O (λ,T)—quantum efficiency, reference cell (%).
D
tainty of Test Results of a Test Method Using Control
3.3
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1125 − 10 (Reapproved 2015) E1125 − 16
Standard Test Method for
Calibration of Primary Non-Concentrator Terrestrial
1
Photovoltaic Reference Cells Using a Tabular Spectrum
This standard is issued under the fixed designation E1125; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method is intended to be used for calibration and characterization of primary terrestrial photovoltaic reference cells
to a desired reference spectral irradiance distribution, such as Tables G173. The recommended physical requirements for these
reference cells are described in Specification E1040. Reference cells are principally used in the determination of the electrical
performance of photovoltaic devices.
1.2 Primary photovoltaic reference cells are calibrated in natural sunlight using the relative spectral responsequantum efficiency
of the cell, the relative spectral distribution of the sunlight, and a tabulated reference spectral irradiance distribution. Selection of
the reference spectral irradiance distribution is left to the user.
1.3 This test method requires the use of a pyrheliometer that is calibrated according to Test Method E816, which requires the
use of a pyrheliometer that is traceable to the World Radiometric Reference (WRR). Therefore, reference cells calibrated according
to this test method are traceable to the WRR.
1.4 This test method is a technique that may be used instead of the procedures found in used to calibrate primary reference cells;
Test Method E1362. This test method offers convenience in its ability to characterize a reference cell under any spectrum for which
tabulated data are available. The selection may be used to calibrate secondary and non-primary reference cells (these terms are
defined in Terminology E772of the specific reference spectrum is left to the user.).
1.5 This test method applies only to the calibration of a photovoltaic cell that shows a linear dependence of its short-circuit
current on irradiance over its intended range of use, as defined in Test Method E1143.
1.6 This test method applies only to the calibration of a reference cell fabricated with a single photovoltaic junction.
1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2
2.1 ASTM Standards:
E490 Standard Solar Constant and Zero Air Mass Solar Spectral Irradiance Tables
E772 Terminology of Solar Energy Conversion
E816 Test Method for Calibration of Pyrheliometers by Comparison to Reference Pyrheliometers
E927 Specification for Solar Simulation for Photovoltaic Testing
E948 Test Method for Electrical Performance of Photovoltaic Cells Using Reference Cells Under Simulated Sunlight
E973 Test Method for Determination of the Spectral Mismatch Parameter Between a Photovoltaic Device and a Photovoltaic
Reference Cell
E1021 Test Method for Spectral Responsivity Measurements of Photovoltaic Devices
E1040 Specification for Physical Characteristics of Nonconcentrator Terrestrial Photovoltaic Reference Cells
1
This test method is under the jurisdiction of ASTM Committee E44 on Solar, Geothermal and Other Alternative Energy Sources and is the direct responsibility of
Subcommittee E44.09 on Photovoltaic Electric Power Conversion.
Current edition approved March 1, 2015July 1, 2016. Published April 2015October 2016. Originally approved in 1986. Last previous edition approved in 20102015 as
E1125 – 10.E1125 – 10 (2015). DOI: 10.1520/E1125-10R15.10.1520/E1125-16.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1125 − 16
E1328E1143 Terminology Relating to Photovoltaic Solar Energy
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.