Standard Test Method for Calibration of Pyrheliometers by Comparison to Reference Pyrheliometers

SCOPE
1.1 This test method has been harmonized with, and is technically equivalent to, ISO 9059.
1.2 Two types of calibrations are covered by this test method. One is the calibration of a secondary reference pyrheliometer using an absolute cavity pyrheliometer as the primary standard pyrheliometer, and the other is the transfer of calibration from a secondary reference to one or more field pyrheliometers. This test method proscribes the calibration procedures and the calibration hierarchy, or traceability, for transfer of the calibrations.Note 1
It is not uncommon, and is indeed desirable, for both the reference and field pyrheliometers to be of the same manufacturer and model designation.
1.3 This test method is relevant primarily for the calibration of reference pyrheliometers with field angles of 5 to 6, using as the primary reference instrument a self-calibrating absolute cavity pyrheliometer having field angles of about 5. Pyrheliometers with field angles greater than 6.5 shall not be designated as reference pyrheliometers.
1.4 When this test method is used to transfer calibration to field pyrheliometers having field angles both less than 5 or greater than 6.5, it will be necessary to employ the procedure defined by Angstrom and Rodhe.
1.5 This test method requires that the spectral response of the absolute cavity chosen as the primary standard pyrheliometer be nonselective over the range from 0.3 to 10-m wavelength. Both reference and field pyrheliometers covered by this test method shall be nonselective over a range from 0.3 to 4-m wavelength.
1.6 The primary and secondary reference pyrheliometers shall not be field instruments and their exposure to sunlight shall be limited to calibration or intercomparisons. These reference instruments shall be stored in an isolated cabinet or room equipped with standard laboratory temperature and humidity control.Note 2
At a laboratory where calibrations are performed regularly, it is advisable to maintain a group of two or three secondary reference pyrheliometers that are included in every calibration. These serve as controls to detect any instability or irregularity in the standard reference pyrheliometer.
1.7 This test method is applicable to calibration procedures using natural sunshine only.

General Information

Status
Historical
Publication Date
30-Sep-2005
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E816-05 - Standard Test Method for Calibration of Pyrheliometers by Comparison to Reference Pyrheliometers
English language
11 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E816–05
Standard Test Method for
Calibration of Pyrheliometers by Comparison to Reference
1
Pyrheliometers
This standard is issued under the fixed designation E816; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
Accurateandprecisemeasurementofthedirect(beam)radiationcomponentofsunlightarerequired
in (1) the calibration of reference pyranometers by the shading disk or optical occluding methods, (2)
determination of the energy collected by concentrating solar collectors, including exposure levels
achievedinuseofPracticeG90dealingwithFresnel-reflectingconcentratortestmachines,and (3)the
assessment of the direct beam for energy budget analyses, geographic mapping of solar energy, and
asanaidinthedeterminationoftheconcentrationofaerosolandparticulatepollution,andwatervapor
effects.
ThistestmethodrequirescalibrationtotheWorldRadiometricReference(WRR),maintainedbythe
World Meteorological Organization (WMO), Geneva. The Intercomparison of Absolute Cavity
Pyrheliometers, also calledAbsolute Cavity Radiometers, on which the WRR depends, is covered by
procedures adopted by WMO and by various U.S. Organizations who occasionally convene such
intercomparisonsforthepurposeoftransferringtheWRRtotheUnitedStates,andtomaintainingthe
WRR in the United States. These procedures are not covered by this test method.
1. Scope 1.4 When this test method is used to transfer calibration to
field pyrheliometers having field angles both less than 5° or
1.1 This test method has been harmonized with, and is
greater than 6.5°, it will be necessary to employ the procedure
technically equivalent to, ISO9059.
2
defined by Angstrom and Rodhe.
1.2 Two types of calibrations are covered by this test
1.5 This test method requires that the spectral response of
method. One is the calibration of a secondary reference
the absolute cavity chosen as the primary standard pyrheliom-
pyrheliometer using an absolute cavity pyrheliometer as the
eter be nonselective over the range from 0.3 to 10-µm
primarystandardpyrheliometer,andtheotheristhetransferof
wavelength. Both reference and field pyrheliometers covered
calibration from a secondary reference to one or more field
by this test method shall be nonselective over a range from 0.3
pyrheliometers. This test method proscribes the calibration
to 4-µm wavelength.
procedures and the calibration hierarchy, or traceability, for
1.6 The primary and secondary reference pyrheliometers
transfer of the calibrations.
shall not be field instruments and their exposure to sunlight
NOTE 1—It is not uncommon, and is indeed desirable, for both the
shall be limited to calibration or intercomparisons. These
reference and field pyrheliometers to be of the same manufacturer and
reference instruments shall be stored in an isolated cabinet or
model designation.
room equipped with standard laboratory temperature and
1.3 Thistestmethodisrelevantprimarilyforthecalibration
humidity control.
of reference pyrheliometers with field angles of 5 to 6°, using
NOTE 2—At a laboratory where calibrations are performed regularly, it
as the primary reference instrument a self-calibrating absolute
is advisable to maintain a group of two or three secondary reference
cavity pyrheliometer having field angles of about 5°. Pyrheli-
pyrheliometers that are included in every calibration. These serve as
ometers with field angles greater than 6.5° shall not be
controls to detect any instability or irregularity in the standard reference
designated as reference pyrheliometers.
pyrheliometer.
1.7 This test method is applicable to calibration procedures
1
This test method is under the jurisdiction of ASTM Committee G03 on
using natural sunshine only.
Weathering and Durability and is the direct responsibility of Subcommittee G03.09
on Radiometry.
Current edition approved Oct. 1, 2005. Published November 2005. Originally
2
Angstrom, A., and Rodhe, B., “Pyrheliometric Measurements with Special
approved in 1981. Last previous edition approved in 2002 as E816–95(2002).
DOI: 10.1520/E0816-05. Regard to the Circumsolar Sky Radiation,” Tellus, Vol 18, 1966, pp. 25–33.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
E816–05
2. Referenced Documents 3.1.7 reference pyrheliometer—pyrheliometers of any cat-
3
egory serving as a reference in calibration transfer procedures.
2.1 ASTM Standards:
They are selected and well-tested instruments (see Table2 of
E772 Terminology Relating to Solar Energy Conversion
ISO9060), that have a
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.