Standard Guide for Assessing the Service Life of a Brush Part Intended to Clean a Medical Device

SIGNIFICANCE AND USE
5.1 This guide describes the use of test methods in Guides F3275 and F3276 to assess the service life of a brush part intended to clean a medical device.  
5.2 In the case of a brush part intended to clean a lumen, the force required to move a brush part within a tube, an indicator of the friction a brush exerts on a surface, is a measurable parameter that can change over time and will decrease as the brush part loses integrity.  
5.3 In the case of a brush part intended to clean the external surface, the force required to move the brush across a surface and the pressure the brush exerts on that surface are measurable parameters that can change over time and will decrease as the brush part loses integrity.  
5.4 By providing objective, repeatable methods for evaluating performance under test conditions, this guide can improve the ability to assess the effectiveness of various brush part designs.
SCOPE
1.1 This guide describes methods for assessing the service life, under prescribed laboratory conditions, of a brush part designed to clean a medical device. The method utilizes force testers to mechanically actuate a brush part at a constant rate. This action continues until the brush part demonstrates a significant reduction in cleaning power as measured by the force exerted during testing.  
1.2 The test methods utilized in this guide are those described in Guides F3275 and F3276. In this guide, the number of repetitions is open-ended and determined by the measurable fatigue of the brush part as measured by a reduction in force, as well as any observation of wear or damage to the brush part.  
1.3 Brushes designed to clean medical devices after clinical use play an important role in the effective reprocessing of those medical devices. Instructions for use from the brush manufacturer should supply information related to the service life of the brush. This may be stated in terms of (1) a time period; (2) the number of uses; (3) inspection of the brush for wear and damage.  
1.4 Inspection for wear should always be a part of the instructions for use of a brush. Application of this guide can help to determine like mode(s) of observable failure of a brush part.  
1.5 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
14-Oct-2020
Current Stage
Ref Project

Buy Standard

Guide
ASTM F3487-20 - Standard Guide for Assessing the Service Life of a Brush Part Intended to Clean a Medical Device
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F3487 − 20
Standard Guide for
Assessing the Service Life of a Brush Part Intended to
1
Clean a Medical Device
This standard is issued under the fixed designation F3487; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
1.1 This guide describes methods for assessing the service
Barriers to Trade (TBT) Committee.
life, under prescribed laboratory conditions, of a brush part
designed to clean a medical device. The method utilizes force
2. Referenced Documents
testers to mechanically actuate a brush part at a constant rate.
2
2.1 ASTM Standards:
This action continues until the brush part demonstrates a
F3275 Guide for Using a Force Tester to Evaluate Perfor-
significant reduction in cleaning power as measured by the
mance of a Brush Part Designed to Clean the Internal
force exerted during testing.
Channel of a Medical Device
1.2 The test methods utilized in this guide are those de-
F3276 Guide for Using a Force Tester to Evaluate the
scribed in Guides F3275 and F3276. In this guide, the number
Performance of a Brush Part Designed to Clean the
of repetitions is open-ended and determined by the measurable
External Surface of a Medical Device
fatigue of the brush part as measured by a reduction in force,
as well as any observation of wear or damage to the brush part.
3. Terminology
1.3 Brushes designed to clean medical devices after clinical
3.1 Definitions:
useplayanimportantroleintheeffectivereprocessingofthose
3.1.1 brush set—working end of the brush that comes in
medical devices. Instructions for use from the brush manufac-
contact with the targeted surface of the substrate.
turershouldsupplyinformationrelatedtotheservicelifeofthe
4. Summary of Guide
brush. This may be stated in terms of (1) a time period; (2) the
number of uses; (3) inspection of the brush for wear and
4.1 This guide describes the application of test methods
damage.
described in Guides F3275 and F3276 to assess the service life
1.4 Inspection for wear should always be a part of the of a brush part intended to clean a medical device.
instructions for use of a brush. Application of this guide can
4.2 Through exhaustive repetitions of the applicable
help to determine like mode(s) of observable failure of a brush
method, the service life of the brush part can be estimated and
part.
described.
1.5 Units—The values stated in SI units are to be regarded
4.3 Determination of end-of-life:
as standard. No other units of measurement are included in this
4.3.1 The end-of-life for a brush part may be determined by
standard.
observable physical wear of the brush part. This may include
1.6 This standard does not purport to address all of the (see Fig. 1):
safety concerns, if any, associated with its use. It is the
4.3.1.1 Missing bristles,
responsibility of the user of this standard to establish appro- 4.3.1.2 Fallen bristles,
priate safety, health, and environmental practices and deter- 4.3.1.3 Twisted bristles,
mine the applicability of regulatory limitations prior to use. 4.3.1.4 Bent bristles,
1.7 This international standard was developed in accor- 4.3.1.5 Bentshaftofthebrush(typicallybrushesintendedto
dance with internationally recognized principles on standard- clean lumens),
ization established in the Decision on Principles for the 4.3.1.6 Damage to the block of the brush part (typically
brushes intended to clean surfaces), and
1
This guide is under the jurisdiction of ASTM Committee F04 on Medical and
2
Surgical Materials and Devices and is the direct responsibility of Subcommittee For referenced ASTM standards, visit the ASTM website, www.astm.org, or
F04.15 on Material Test Methods. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Current edition approved Oct. 15, 2020. Published November 2020. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/F3487-20. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
F3487 − 20
FIG. 1 Observable Physical Wear of Brush Part
4.3.1.7 Untwistingofthewires(inthecaseofatwistedwire 6.2.1.2 For this guide, test soil is not required (Section 7.2
brush). of F3276).
4.3.2 Theend-of-lifemayalsobeindicatedbyareductionin
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.