SIGNIFICANCE AND USE
5.1 Intervertebral body fusion devices are generally simple geometric-shaped devices, which are often porous or hollow in nature. Their function is to support the anterior column of the spine to facilitate arthrodesis of the motion segment.  
5.2 This test method is designed to quantify the subsidence characteristics of different designs of intervertebral body fusion devices since this is a potential clinical failure mode. These tests are conducted in vitro in order to simplify the comparison of simulated vertebral body subsidence induced by the intervertebral body fusion devices.  
5.3 The static axial compressive loads that will be applied to the intervertebral body fusion devices and test blocks will differ from the complex loading seen in vivo, and therefore, the results from this test method may not be used to directly predict in vivo performance. The results, however, can be used to compare the varying degrees of subsidence between different intervertebral body fusion device designs for a given density of simulated bone.  
5.4 The location within the simulated vertebral bodies and position of the intervertebral body fusion device with respect to the loading axis will be dependent upon the design and manufacturer's recommendation for implant placement.
SCOPE
1.1 This test method specifies the materials and methods for the axial compressive subsidence testing of non-biologic intervertebral body fusion devices, spinal implants designed to promote arthrodesis at a given spinal motion segment.  
1.2 This test method is intended to provide a basis for the mechanical comparison among past, present, and future non-biologic intervertebral body fusion devices. This test method is intended to enable the user to mechanically compare intervertebral body fusion devices and does not purport to provide performance standards for intervertebral body fusion devices.  
1.3 This test method describes a static test method by specifying a load type and a specific method of applying this load. This test method is designed to allow for the comparative evaluation of intervertebral body fusion devices.  
1.4 Guidelines are established for measuring test block deformation and determining the subsidence of intervertebral body fusion devices.  
1.5 Since some intervertebral body fusion devices require the use of additional implants for stabilization, the testing of these types of implants may not be in accordance with the manufacturer's recommended usage.  
1.6 Units—The values stated in SI units are to be regarded as the standard with the exception of angular measurements, which may be reported in terms of either degrees or radians.  
1.7 The use of this standard may involve the operation of potentially hazardous equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language
  • Standard
    8 pages
    English language

ABSTRACT
This specification covers the material requirements and characterization techniques for glass and glass-ceramic biomaterials intended for use as bulk porous or powdered surgical implants, or as coatings on surgical devices, but not including drug delivery systems. Glass and glass-ceramic biomaterials should be evaluated thoroughly for biocompatibility before human use. Tests shall be performed to determine the properties of the biomaterials, in accordance with the following test methods: bulk composition; density; flexural strength; Young's modulus; hardness; surface area; bond strength of glass or glass ceramic coating; crystallinity; thermal expansion; and particle size.
SCOPE
1.1 This specification covers the material requirements and characterization techniques for glass and glass-ceramic biomaterials intended for use as bulk porous or powdered surgical implants, or as coatings on surgical devices, but not including drug delivery systems.  
1.2 The biological response to glass and glass-ceramic biomaterials in bone and soft tissue has been demonstrated in clinical use (1-12)2 and laboratory studies (13-17).  
1.3 This specification excludes synthetic hydroxylapatite, hydroxylapatite coatings, aluminum oxide ceramics, alpha- and beta-tricalcium phosphate, and whitlockite.  
1.4 Warning—Mercury has been designated by EPA and many state agencies as a hazardous material that can cause central nervous system, kidney, and liver damage. Mercury, or its vapor, may be hazardous to health and corrosive to materials. Caution should be taken when handling mercury and mercury-containing products. See the applicable product Material Safety Data Sheet (MSDS) for details and EPA’s website (http://www.epa.gov/mercury/faq.htm) for additional information. Users should be aware that selling mercury or mercury-containing products, or both, in your state may be prohibited by state law.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    4 pages
    English language
  • Technical specification
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 IBFDs can be single-piece or multicomponent designs and can be porous or hollow in nature. Their function is to support the anterior column of the spine to facilitate arthrodesis of the motion segment.  
5.2 Intra-operative IBFD assembly failures can result in significant clinical consequences.4 This test method outlines materials and methods for the comparative characterization and evaluation of the intra-operative impact performance of IBFD assemblies.  
5.3 The impact forces applied during a surgical procedure may be highly variable and, therefore, the results from these tests may not directly predict in vivo performance. The results, however, can be used to compare mechanical performance of different IBFD assemblies. The tests may also identify the weakest, most likely to fail points in particular IBFD-inserter combinations, thus enabling design improvements.  
5.4 Intra-operative clinical failures may be due to several factors, some of which may not be simulated in the current method. For example, off-axis impact loads applied to the IBFD assembly are not simulated in the current method yet may contribute to intra-operative clinical failures in some designs. The user of this standard should consider incorporating such factors into their evaluations.
SCOPE
1.1 This test method covers the materials and methods for impact testing of lumbar intervertebral body fusion devices (IBFD).  
1.2 This test method is intended to provide a basis for the mechanical comparison among nonbiologic IBFD assemblies (the IBFD and associated inserter tool). This test method is intended to enable the user to compare these IBFD assemblies under impact loads to simulate the intra-operative surgical technique used to insert the IBFD.  
1.3 The test method describes the impact test by specifying impact energies and specific methods for applying these energies. The tests are designed to allow for the comparative evaluation of IBFD assemblies.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard, with the exception of angular measurements, which may be reported in terms of either degrees or radians.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    12 pages
    English language

ABSTRACT
This specification covers the chemical, mechanical, and metallurgical requirements for wrought annealed, cold worked, or hot rolled titanium-6aluminum-7niobium alloy (UNS R56700) bar and wire to be used in the manufacture of surgical implants. Titanium mill products covered in this specification shall be formed with the conventional forging and rolling equipment found in primary ferrous and nonferrous plants, and may be furnished as descaled or pickled, sandblasted, chemically milled, ground, machined, peeled, polished, or cold drawn. The alloy shall be multiple melted in arc furnaces (including furnaces such as plasma arc and electron beam) of a type conventionally used for reactive metals. Heat analysis shall conform to the chemical composition requirements prescribed for aluminum, niobium, tantalum, iron, oxygen, carbon, nitrogen, hydrogen, and titanium. The material shall conform to the specified requirements for mechanical properties such as ultimate tensile strength, yield strength, and elongation. A minimum of two tension tests from each lot shall be performed. Special requirements for the microstructure are detailed as well.
SCOPE
1.1 This specification covers the chemical, mechanical, and metallurgical requirements for wrought annealed, cold-worked, or hot-worked titanium-6aluminum-7niobium alloy bar, wire, sheet, strip, and plate to be used in the manufacture of surgical implants (1-7).2  
1.2 The SI units in this standard are the primary units. The values stated in either primary SI units or secondary inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    6 pages
    English language
  • Technical specification
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 The crystallinity of UHMWPE will influence its mechanical properties, such as creep and stiffness. The reported crystallinity will depend on the integration range used to determine the heat of fusion, and the theoretical heat of fusion of 100 % crystalline polyethylene used to calculate the percent crystallinity in an unknown specimen. Differential scanning calorimetry is an effective means of accurately measuring both heat of fusion and melting temperature.  
5.2 This test method is useful for both process control and research.
SCOPE
1.1 This quantitative test method discusses the measurement of the heat of fusion and the melting point of ultra-high molecular weight polyethylene (UHMWPE), and the subsequent calculation of the percentage of crystallinity. The method uses a differential scanning calorimeter and can be performed in the laboratory or in the field.  
1.2 This test method can be used for UHMWPE in powder form, consolidated form, finished product, or a used product. It can also be used for irradiated or chemically crosslinked UHMWPE.  
1.3 This test method does not suggest a desired range of crystallinity or melting points for specific applications.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language
  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
4.1 Standard Structure—This standard has been organized according to a typical bioprinting operational workflow. In this way the standard has three main sections: pre-printing considerations, printing considerations, and post-printing considerations. There is also a section on bioprinting modalities and additional considerations of product release, containers, and transport. Certain processes will appear across multiple sections, for example cytocompatibility or crosslinking, as these issues have considerations that take place prior to printing, during the printing process, and following the printing process. Contents of main sections are listed below.    
Scope  
Section 1  
Referenced Documents  
Section 2  
Terminology  
Section 3  
Significance and Use  
Section 4  
Bioprinting Modalities  
Section 5  
Pre-Printing Considerations  
Section 6  
Printing Considerations  
Section 7  
Post-Printing Considerations  
Section 8  
Additional Considerations  
Section 9  
Keywords  
Section 10  
References  
4.1.1 Pre-Printing Considerations—Pre-printing considerations include: bioink common applications, support material, and bioink selection considerations. Bioink selection considerations include: formulation of bioinks, bioink properties, changes in properties resulting from formulations, sterility, cellular component, and fugitive element considerations. Within the bioink properties there are considerations related to the viscoelastic properties, chemical properties, structure of polymer and functional groups, purity of material, mechanism of crosslinking, and degradation considerations. Contents of the section on pre-printing considerations are listed below.    
Bioink Common Applications  
6.2  
Support Material Considerations  
6.3  
Bioink Selection  
6.4  
Formulation of Bioinks  
6.4.1  
Concentration of Components  
6.4.1.1  
  
...
SCOPE
1.1 This guide is a resource for bioprinting tissue-engineered medical products (TEMPs) with bioinks and biomaterial inks. There are existing standards that cover biomaterials and scaffolds in a more general fashion (Guide F2150, Guide F2027, ISO 10993 series). This guide focuses specifically on extrusion bioprinting utilizing bioinks and biomaterial inks with inherent or inducible fluidic properties with or without encapsulated cells used to construct TEMPs. For the remainder of this guide, both bioinks and biomaterial inks will be collectively referred to as bioinks.  
1.2 For the purposes of this guide, bioprinting is defined as the three-dimensional printing of materials (bioinks) to fabricate structured constructs for use in biological or medical applications.  
1.3 TEMPs may be produced by many different bioprinting modalities, including but not limited to the following: electrospinning, electrospray, extrusion-based, droplet-based, inkjet-based, and laser-assisted bioprinting. Extrusion-based bioprinting is the primary focus of this document since it is currently the most well-understood modality used to construct TEMPs, but other bioprinting modalities are also addressed.  
1.4 This guide will focus on bioinks and biomaterials used as inks with inherent or inducible fluidic properties. These inks may or may not contain encapsulated cells. Chemical properties of the inks and other factors that affect printability are addressed.  
1.5 Pre-printing and printing considerations are the focus of this guide, but considerations regarding post-printing product stabilization are also addressed.  
1.6 This guide will address assessments regarding the sterility and cytocompatibility of bioinks, including chemical and physical benchtop tests, as well as measures of post-printing cell viability.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsi...

  • Guide
    20 pages
    English language

ABSTRACT
This specification covers UNS S31673 chromium-nickel-molybdenum wrought annealed stainless steel surgical fixation wires. The wires should conform to the required values of tensile strength and elongation. Wire surfaces should be processed to minimize tool marks, nicks, scratches, cracks, cavities, spurs, and other imperfections that would impair wire serviceability.
SCOPE
1.1 This specification covers the chemical, mechanical, and metallurgical requirements for the manufacture of wrought 18chromium-14nickel-2.5molybdenum stainless steel in the form of surgical fixation wire.  
1.2 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of the other, and values from the two systems shall not be combined.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    4 pages
    English language
  • Technical specification
    4 pages
    English language

SIGNIFICANCE AND USE
4.1 The term “surface texture” is used to describe the local deviations of a surface from an ideal shape. Surface texture usually consists of long wavelength repetitive features that occur as results of chatter, vibration, or heat treatments during the manufacture of implants. Short wavelength features superimposed on the long wavelength features of the surface, which may arise from polishing or etching of the implant, are referred to as roughness.  
4.2 This guide provides an overview of techniques that are available for measuring the surface in terms of Cartesian coordinates and the parameters used to describe surface texture. It is important to appreciate that it is not possible to measure surface texture per se, but to derive values for parameters that can be used to describe it. ISO has published a series of standards on surface texture measurements that may be consulted for more information (ISO 3274, ISO 4287, ISO 4288, ISO 5436-2, ISO 10993-19, ISO 12179, ISO 13565-1, ISO 19606, ISO 21920-1, ISO 21920-2, ISO 21920-3, ISO 25178-1, ISO 25178-2, ISO 25178-3, ISO 25178-6, ISO 25178-70, ISO 25178-71, ISO 25178-72, ISO 25178-73, ISO 25178-600, ISO 25178-601, ISO 25178-602, ISO 25178-603, ISO 25178-604, ISO 25178-605, ISO 25178-606, ISO 25178-607, ISO 25178-700, ISO 25178-701).
SCOPE
1.1 This guide describes some of the more common methods that are available for measuring the topographical features of a surface and provides an overview of the parameters that are used to quantify them. Being able to reliably derive a set of parameters that describe the texture of biomaterial surfaces is a key aspect in the manufacture of safe and effective implantable medical devices that have the potential to trigger an adverse biological reaction in situ.  
1.2 This guide is not intended to apply to porous structures with average pore dimensions in excess of approximately 50 nm (0.05 μm).  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    10 pages
    English language
  • Guide
    10 pages
    English language

SIGNIFICANCE AND USE
5.1 This standard guide may be used by medical device manufacturers as part of their design plan and implementation of the validation of the cleaning instructions of their reusable medical devices.  
5.2 This guide helps medical device manufacturers to identify the appropriate method(s) for detecting and quantifying markers for the simulated-use test soil (see Guide F3208), thereby evaluating whether the medical device can be adequately cleaned.  
5.3 This guide describes various test methods for the different analytes.  
5.4 This guide specifies the validation criteria for analyte detection methods.
SCOPE
1.1 This standard guide provides methods and considerations for detecting and quantifying test soil(s) from reusable medical device(s) that result from simulated-use testing of medical devices during validation of the cleaning procedures as described in the instructions for use (IFU) provided by the medical device manufacturer.  
1.2 The methods described are for detecting and measuring markers (analytes) that are components within the most common test soils and are relevant to the clinical use of the device. Appropriate test soils without protein, carbon, or carbohydrates (for example, bone) will require other methods.  
1.3 This is a part of a series of ASTM standard guides for validating cleaning instructions. The scope of the first guide in the series is selecting appropriate test soils (Guide F3208). The second in the series (Guide F3293) describes methods for inoculating medical devices with test soil. The third in the series (Guide F3321) describes methods for extracting soils for measuring residual soil on medical devices after the performance of cleaning process. This is the fourth guide in the series and describes the methods of detecting and quantifying residual analytes on the device.  
1.4 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    8 pages
    English language
  • Guide
    8 pages
    English language

SIGNIFICANCE AND USE
5.1 The evaluation of particulates from endovascular devices is a critical parameter to help ensure that the devices can be delivered to and from the treatment site with minimal unintended effects due to particulate generation.  
5.2 This guide is intended for the development and design of a simulated use model for the evaluation of particulate generation by those manufacturing and/or evaluating endovascular devices.  
5.3 This guide may be useful for establishment of credibility of a simulated use model as part of development testing and regulatory submission testing and filings.
SCOPE
1.1 This standard provides recommendations regarding the development of a simulated use model for particulate generation testing of endovascular devices.  
1.2 Corresponding recommendations will be provided for the coronary, peripheral, and neurovascular anatomy respectively, as these different environments each have unique challenges and considerations that should be incorporated into the model development.  
1.3 Considerations include, but are not limited to: anatomical considerations (for example, tortuosity, dimensions, disease state), working path length of the human vasculature, model/device interactions (for example, material properties, vessel straightening due to device placement), and device/device interactions (for example, use of delivery catheters and/or guide wires).  
1.4 Standardizing the model development allows for better interpretation of the test results and comparison of performance of similar medical devices.  
1.5 Explicit models for devices (or groups of devices) are not provided in this document as specific device attributes (for example, device flexibility, device material), anatomical target locations, and intended patient population can strongly impact model design. This document enables the development of a model that appropriately and sufficiently challenges the subject device with respect to particulate generation.  
1.6 While this document is intended to aid in the development and design of the simulated use model for the assessment of particulate generation, additional information regarding the evaluation of the particulate matter and particulate measurements can be found in AAMI TIR 42. Similarly, this document is not intended to address testing associated with the use of the model.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    5 pages
    English language

SCOPE
1.1 This guide provides a general guide to transportation, including packaging and shipping, of possibly infectious materials, tissues, and fluids that have been removed from patients during revision surgery, at postmortem, or as part of animal studies, including packaging and shipping.  
1.2 This guide does not address any materials, tissues, or fluids that may contain prions.  
1.3 Individuals must be properly trained prior to shipping possibly infectious materials.  
1.4 This guide is a compilation of national and international regulations and guidelines that apply to the packaging and shipment of possibly infectious materials.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    9 pages
    English language
  • Guide
    9 pages
    English language

SIGNIFICANCE AND USE
3.1 The purpose of this guide is to provide a procedure for determining the appropriate attributes to evaluate in a shelf-life study for an endovascular device.
SCOPE
1.1 This guide addresses the determination of appropriate device attributes for testing as part of a shelf-life study for endovascular devices. Combination and biodegradable devices (for example, drug devices, biologic devices, or drug biologics) may require additional considerations, depending on their nature.  
1.2 This guide does not directly provide any test methods for conducting shelf-life testing.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    6 pages
    English language

ABSTRACT
This specification covers the requirements for wrought seamless and welded and drawn cobalt alloy small diameter tubing used for the manufacture of surgical implants. Product variables that differentiate small diameter medical tubing from the bar, wire, sheet, and strip product forms are addressed. This specification applies to straight length tubing of specified diameters and thickness. Seamless tubing shall be made from bar, hollow bar, rod, or hollow rod raw material forms through a prescribed process. Welded and drawn tubing shall be made from strip or sheet raw material forms that meet the specified chemical requirements. The tubing shall be subject to tensile testing.
SCOPE
1.1 This specification covers the requirements for wrought seamless and welded and drawn cobalt alloy small diameter tubing used for the manufacture of surgical implants. Material shall conform to the applicable requirements of Specifications F90, F562, F688, F1058 or F1537, Alloy 1. This specification addresses those product variables that differentiate small diameter medical tubing from the bar, wire, sheet, and strip product forms covered in these specifications.  
1.2 This specification applies to straight length tubing with 6.3 mm [0.250 in.] and smaller nominal outside diameter (OD) and 0.76 mm [0.030 in.] and thinner nominal wall thickness.  
1.3 The specifications in 2.1 are referred to as the ASTM material standard(s) in this specification.  
1.4 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of the other, and values from the two systems shall not be combined.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    6 pages
    English language
  • Technical specification
    6 pages
    English language

ABSTRACT
This specification covers the material requirements and functional dimensions for fixation pins and wires. Factors such as bending strength, fatigue strength, break strength (Knowles type only), torsion strength, and ductility are considered important but the values and associated test methods for which have not been established yet.
SCOPE
1.1 This specification covers functional dimensions for fixation pins and wires.  
1.2 In recognition of many broad and varied uses of such pins and wires, many options are included. A variety, but not necessarily all, of the options are illustrated in Figs. 1-3.
FIG. 1 Fixation Pins and Wires
Note 1: Pins and wires may be smooth shank or threaded.
Note 2: Point angle and helix angle, where applicable, is as specified by the manufacturer.
Note 3: On square or triangular shanks, flats are equal and corners are on the same circumference as the pin diameter. Shank diameters on pins larger than 3.2 mm may be reduced.
Note 4: Optional designs, both ends pointed or point with suture hole.
FIG. 2 Fixation Pin—Knowles Type
Note 1: Shall be made of one piece of material; that is, not an assembly of two or more pieces.
Note 2: Thread major diameter may be the same or larger than the shaft diameter.
Note 3: Thread length shall be optional to suit the intended medical application.
Note 4: Optional marking in this location.
FIG. 3 Fixation Pin—Hagie Type
Note 1: Thread major diameter may be the same or larger than the shaft diameter.
Note 2: Thread length shall be optional to suit the intended medical application.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    5 pages
    English language
  • Technical specification
    5 pages
    English language

ABSTRACT
This specification covers the chemical requirements for wrought stainless steels used for the manufacture of surgical instruments. Classes of stainless steels covered here are Class 3 (austenitic stainless steel), Class 4 (martensitic stainless steel), Class 5 (precipitation hardening stainless steel), and Class 6 (ferritic stainless steel). The data contained in this specification, such as typical hardness values, common heat treating cycles, and examples of selected stainless steels that have been used for surgical instruments, are provided for reference only. Mechanical property, heat treatment, hardness, and all other requirements except for chemical composition, are governed by the appropriate material standards as specified or as agreed upon between purchaser and supplier.
SCOPE
1.1 This specification covers the chemistry requirements for wrought stainless steels used for the manufacture of surgical instruments. The data contained in Tables 1-4 of this specification, including typical hardness values, common heat treating cycles, and examples of selected stainless steels that have been used for surgical instruments, is provided for reference only. Mechanical property requirements, heat treating requirements, hardness requirements, and all other requirements except chemistry are governed by the appropriate material standards as referenced below or as agreed upon between the purchaser and supplier.      
1.2 The SI units in this standard are the primary units. The values stated in either primary SI units or secondary inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of each other. Combining values from the two systems may result in nonconformance with the standard.  
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    8 pages
    English language
  • Technical specification
    8 pages
    English language

SIGNIFICANCE AND USE
5.1 This test method provides a rapid, economical method for determination of transformation temperatures.  
5.2 Measurement of the specimen motion closely parallels many shape memory applications and provides a result that is applicable to the function of the material.  
5.3 This test method uses a wire, tube, strip specimen, or a wire, tube, or strip specimen extracted from a component; thus, it provides an assessment of a nickel titanium product in its semifinished or finished form.  
5.4 This test method may be used on annealed samples to determine the transformation temperatures and ensure the alloy formulation, since chemical analysis is not precise enough to adequately determine the nickel-to-titanium ratio of shape memory alloys.  
5.5 In general, the transformation temperatures measured by this method will not be the same as those measured by the DSC method defined in Test Method F2004. Therefore, the results of DSC and BFR cannot be compared directly.  
5.5.1 The BFR method measures the transformation temperatures by tracking shape recovery of stress-induced martensite deformed below the R′s temperature or the As temperature. In contrast, the DSC method measures the start, peak, and finish temperatures of the thermal transformation of martensite to R-phase or to austenite. See Refs (1-4).  
5.6 The test method is applicable to shape memory alloys with Af temperatures in the range of approximately –25 to 90 °C.
SCOPE
1.1 This test method describes a procedure for quantitatively determining the martensite-to-austenite or the martensite to R-phase transformation temperature of annealed, aged, shape-set, or tempered nickel-titanium alloy specimens by deforming the specimen in bending and measuring the deformation recovered during heating through the thermal transformation (BFR method). See 3.1.1.
Note 1: For aged, shape-set, or tempered specimens the transformation may be from martensite to austenite or from martensite to R-phase. See Reference (1)2 for details.  
1.2 The test specimen may be wire, tube, or strip or a specimen extracted from a semifinished or finished component.  
1.2.1 For specimens not in the form of a wire, tube, or strip that are extracted from semifinished or finished components, a wire, tube, or strip shaped test specimen shall be made from the component such that the deformation mode in the test specimen is pure bending.  
1.2.2 Other specimen geometries or displacements resulting in a more complex strain state, such as bending with torsion or buckling, are beyond the scope of this standard.  
1.3 Ruggedness tests have demonstrated that sample Af must be limited to obtain good test results. See 5.6 for details. Ruggedness tests have demonstrated that deformation strain, deformation temperature, and equilibration time at the deformation temperature must be controlled to obtain good test results. See 9.1, 9.2, and 9.4 for details.  
1.4 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language
  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
4.1 This test method can be used to describe the effects of materials, manufacturing, and design variables on the fatigue/cyclic creep performance of UHMWPE bearing components subject to substantial rotation in the transverse plane (relative to the tibial tray) for a relatively large number of cycles.  
4.2 The loading and kinematics of bearing component designs in vivo will, in general, differ from the loading and kinematics defined in this test method. The results obtained here cannot be used to directly predict in vivo performance. However, this test method is designed to enable comparisons between the fatigue performance of different bearing component designs when tested under similar conditions.  
4.3 The test described is applicable to any bicompartmental knee design, including mobile bearing knees that have mechanisms in the tibial articulating component to constrain the posterior movement of the femoral component and a built-in retention mechanism to keep the articulating component on the tibial plate.
SCOPE
1.1 This standard specifies a test method for determining the endurance properties and deformation, under specified laboratory conditions, of ultra high molecular weight polyethylene (UHMWPE) tibial bearing components used in bicompartmental or tricompartmental knee prosthesis designs.  
1.2 This test method is intended to simulate near posterior edge loading similar to the type of loading that would occur during high flexion motions such as squatting or kneeling.  
1.3 Although the methodology described attempts to identify physiological orientations and loading conditions, the interpretation of results is limited to an in vitro comparison between knee prosthesis designs and their ability to resist deformation and fracture under stated test conditions.  
1.4 This test method applies to bearing components manufactured from UHMWPE.  
1.5 This test method could be adapted to address unicompartmental total knee replacement (TKR) systems, provided that the designs of the unicompartmental systems have sufficient constraint to allow use of this test method. This test method does not include instructions for testing two unicompartmental knees as a bicompartmental system.  
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language
  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
4.1 This practice is a guideline for short-term and long-term assessment of skeletal muscle and bone tissue responses to long-term implant materials. For testing of final finished medical devices, the test article for implantation shall be as for intended use, including packaging and sterilization. The tissue responses to the test article are compared to the skeletal muscle and/or bone tissue response(s) elicited by control materials. The controls consistently demonstrate known cellular reaction and wound healing.
SCOPE
1.1 This practice provides guidelines for biological assessment of tissue responses to nonabsorbable for medical device implants. It assesses the effects of the material that is implanted intramuscularly or intraosseously. The experimental protocol is not designed to provide a comprehensive assessment of the systemic toxicity, immune response, carcinogenicity, or mutagenicity of the material since other standards address these issues. It applies only to materials with projected applications in humans where the materials will reside in bone or skeletal muscle tissue in excess of 30 days. Applications in other organ systems or tissues may be inappropriate and are therefore excluded. Control materials are well recognized with a well-characterized long-term response and can include metals and any one of the metal alloys in Specification F67, F75, F90, F136, F138, or F562, high purity dense aluminum oxide as described in Specification F603, ultra high molecular weight polyethylene as stated in Specification F648, or USP polyethylene negative control.  
1.2 The values stated in SI units, including units officially accepted for use with SI, are to be regarded as standard. No other systems of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language
  • Standard
    5 pages
    English language

ABSTRACT
This specification covers chemical, mechanical, and metallurgical general requirements for metal injection molded (MIM) cobalt-28chromium-6molybddenum components to be used in manufacturing surgical implants. In this specification, the MIM components covered may have been densified beyond their as-sintered density by post-sinter processing. For the chemical requirements, the components supplied in this specification must conform in accordance to the chemical requirements specified herein in Table 1. The product analysis tolerances must also conform to the product tolerances presented in Table 2. The specification also enumerates the mechanical requirements for MIM components wherein the tensile properties of the MIM must conform to the mechanical properties in Table 3. The microstructural requirements and specimen preparation shall be in accordance with Guide E3 and Practice E407.
SCOPE
1.1 This specification covers chemical, mechanical, and metallurgical requirements for metal injection molded (MIM) cobalt-28chromium-6molybdenum components to be used in the manufacture of surgical implants  
1.2 The MIM components covered by this specification may have been densified beyond their as-sintered density by post-sinter processing.  
1.3 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the standard.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    5 pages
    English language

SIGNIFICANCE AND USE
4.1 Linear displacement sensor systems play an important role in orthopedic applications to measure micromotion during simulated use of joint prostheses.  
4.2 Linear displacement sensor systems must be calibrated for use in the laboratory to ensure reliable conversions of the system’s electrical output to engineering units.  
4.3 Linear displacement sensor systems should be calibrated before initial use, at least annually thereafter, after any change in the electronic configuration that employs the sensor, after any significant change in test conditions using the sensor that differ from conditions during the last calibration, and after any physical action on the sensor that might affect its response.  
4.4 Verification of sensor performance in accordance with calibration should be performed on a per use basis both before and after testing. Such verification can be done with a less accurate standard than that used for calibration, and may be done with only a few points.  
4.5 Linear displacement sensor systems generally have a working range within which voltage output is linearly proportional to displacement of the sensor. This procedure is applicable to the linear range of the sensor. Recommended practice is to use the linear displacement sensor system only within its linear working range.
SCOPE
1.1 This practice covers the procedures for calibration of linear displacement sensors and their corresponding power supply, signal conditioner, and data acquisition systems (linear displacement sensor systems) for use in measuring micromotion. It covers any sensor used to measure displacement that gives an electrical voltage output that is linearly proportional to displacement. This includes, but is not limited to, linear variable differential transformers (LVDTs) and differential variable reluctance transducers (DVRTs).  
1.2 This calibration procedure is used to determine the relationship between output of the linear displacement sensor system and displacement. This relationship is used to convert readings from the linear displacement sensor system into engineering units.  
1.3 This calibration procedure is also used to determine the error of the linear displacement sensor system over the range of its use.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    4 pages
    English language
  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 This test method is designed to produce data indicative of the degree of crosslinking in ultra high molecular weight polyethylene that has been crosslinked chemically or by ionizing radiation.  
5.2 The results are sensitive to the test temperature, solvent, and method used. For the comparison of data between institutions, care must be taken to have the same test conditions and reagents.  
5.3 The data can be used for dose uniformity analysis, fundamental research, and quality assurance testing.
SCOPE
1.1 This test method describes how the crosslink density, molecular weight between crosslinks, and number of repeat units between crosslinks in ultra high molecular weight polyethylene (UHMWPE) crosslinked by ionizing radiation or by chemical means can be determined by measuring the swelling ratio of samples immersed in o-xylene. Examples of experimental techniques used to make these measurements are discussed herein.  
1.2 The test method reported here measures the change in height of a sample specimen while it is immersed in the solvent. Volumetric swell ratios assume that the sample is crosslinked isotropically, and that the change in dimension will be uniform in all directions. This technique avoids uncertainty induced by solvent evaporation or temperature change.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 This test method can be used to describe the effects of materials, manufacturing, and design variables on the fatigue performance of metallic tibial trays subject to cyclic loading for relatively large numbers of cycles.  
4.2 The loading of tibial tray designs in vivo will, in general, differ from the loading defined in this practice. The results obtained here cannot be used to directly predict in vivo performance. However, this practice is designed to allow for comparisons between the fatigue performance of different metallic tibial tray designs, when tested under similar conditions.  
4.3 In order for fatigue data on tibial trays to be comparable, reproducible, and capable of being correlated among laboratories, it is essential that uniform procedures be established.
SCOPE
1.1 This test method covers a procedure for the fatigue testing of metallic tibial trays used in partial knee joint replacements.  
1.2 This test method covers the procedures for the performance of fatigue tests on metallic tibial components using a cyclic, constant-amplitude force. It applies to tibial trays which cover either the medial or the lateral plateau of the tibia.  
1.3 This test method may require modifications to accommodate other tibial tray designs.  
1.4 This test method is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic tibial trays with unsupported mid-section of the condyle.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
3.1 This practice is intended to confirm the method of obtaining and evaluating the fluorescent penetrant indications on metallic surgical implants.
SCOPE
1.1 This practice is intended as a standard for fluorescent penetrant inspection of metallic surgical implants.  
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    2 pages
    English language
  • Standard
    2 pages
    English language

ABSTRACT
This specification covers polyetheretherketone (PEEK) polymer in virgin forms as pellets, powder, and so forth. It provides requirements and associated test methods for these thermoplastics when they are to be used in the manufacture of intracorporeal devices such as surgical implants or components of surgical or dental devices. The PEEK polymer in the scope of this specification is a pure semicrystalline homopolymer consisting of phenylene rings connected by ether (E) and carbonyl (or ketone, K) groups along the polymer chain. Its polymeric structure is defined by the repeating unit EEK. As with any material, some characteristics may be altered by the processing techniques (molding, extrusion, machining, assembly, sterilization, and so forth) required for the production of a specific part or device. Therefore, properties of fabricated forms of these polymers shall be evaluated using appropriate test methods such as US Pharmacopeia test (infrared spectroscopy, viscosity measurements, and total heavy metal content (as lead)), and differential scanning calorimetry. Fabricated forms, material or forms containing colorants, fillers, processing aids, or other additives, as well as polymer blends which contain PEEK, or reclaimed materials, are not covered by this specification. The properties which shall be evaluated are the following: density, tensile strength, elongation, flexural strength, flexural modulus, and impact strength. The biocompatibility of PEEK polymers is also detailed.
SCOPE
1.1 This specification covers polyetheretherketone (PEEK) polymer in virgin forms (for example, pellets, granules, powder, filaments used in additive manufacturing) and fabricated forms. It provides requirements and associated test methods for these thermoplastics when they are to be used in the manufacture of intracorporeal devices such as surgical implants or components of surgical or dental devices.  
1.2 The properties included in this specification are those applicable for PEEK polymers only. Indicated properties are for fabricated forms. Materials or forms containing colorants, fillers, processing aids, or other additives, as well as polymer blends which contain PEEK, or reclaimed materials, are not covered by this specification.  
1.3 This specification is designed to recommend physical, chemical, and biological test methods to establish a reasonable level of confidence concerning the performance of virgin PEEK polymers for use in medical implant devices.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 When evaluating material in accordance with this specification, hazardous materials, operations, and equipment may be involved. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    5 pages
    English language
  • Technical specification
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 These test methods are intended to determine the ultimate failure load of a ceramic femoral knee component. This information can be used for evaluation of different ceramic component designs or different ceramic materials, or for series production control.  
5.2 Although the test methodology described attempts to identify physiologically relevant intraoperative and in vivo loading conditions, the interpretation of results is limited to an in vitro comparison between ceramic femoral component designs and materials regarding their static ultimate failure load under the stated test conditions.
SCOPE
1.1 The test methods included in this standard cover two procedures for static burst testing of a ceramic femoral component used in total knee replacement (TKR). The two procedures are used to determine the static ultimate failure load of a ceramic femoral knee component. Both procedures are simulating in vivo loading conditions. One of the procedures additionally simulates intraoperative loading conditions. The standard applies to cruciate retaining (CR) femoral components which cover both the medial and lateral condyles and the patellar surface of the femur. These test methods may require modifications to accommodate other femoral component designs.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    11 pages
    English language

ABSTRACT
These test methods cover the establishment of performance requirements for the utilization of a single-use, enteral feeding device with a retention balloon, used by medical professionals for providing a means of nutrition and/or administration of medication to patients by means of natural orifice (nasal, oral, transluminal) and or a surgically created stoma. The product is manufactured in various sizes and materials such as silicone, urethane, and various polymers (as well as combinations of these) and is provided nonsterile for sterilization and sterile for single use only. The following test methods are: Flow rate through feeding lumen test method which covers the determination of flow rates through the drainage lumen of the enteral feeding device with retention balloon, balloon burst volume test method which covers the determination of balloon integrity of enteral feeding devices with retention balloon, balloon volume maintenance test method which is applicable enteral feeding devices with retention balloon to test the integrity of the inflation system to maintain balloon volume, balloon concentricity test method which is applicable enteral feeding devices with retention balloon to test the concentricy of the balloon, balloon size and shaft size test method which evaluates the retention balloon shaft size, balloon integrity test method which evaluates the integrtity of the retention balloon of the enteral feeding device, and balloon integrity in simulated gastric fluid test method which assesses the ability of the retention balloon to withstand gastric acidity levels without rupture, therefore, maintaining its functional purpose of retention.
SCOPE
1.1 These test methods cover the establishment of performance requirements for the utilization of a single-use, enteral feeding device with a retention balloon, used by medical professionals for providing a means of nutrition and/or administration of medication to patients by means of natural orifice (nasal, oral, transluminal) and or a surgically created stoma. The product is manufactured in various sizes and materials such as silicone, urethane, and various polymers (as well as combinations of these) and is provided nonsterile for sterilization and sterile for single use only. Rationale for these test methods can be found in Appendix X1.  
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    8 pages
    English language

ABSTRACT
This specification covers the material requirements for calcium phosphate coatings for surgical implant applications. In particulate and monolithic form, the calcium phosphate materials system has been well-characterized regarding biological response and laboratory characterization. This specification includes hydroxylapatite coatings, tricalcium phosphate coatings, or combinations thereof, with or without intentional minor additions of other ceramic or metallic, and applied by methods including, but not limited to, the following: mechanical capture, plasma spray deposition, dipping/sintering, electrophoretic deposition, porcelainizing, and sputtering. Substrates may include smooth, porous, textured, and other implantable topographical forms. This specification excludes organic coatings that may contain calcium and phosphate ionic species. Materials shall be tested and the individual grades shall conform to chemical requirements such as elemental analysis for calcium and phosphates, and intentional additions, trace element analysis for hydroxylapatite and beta tricalcium phosphate; crystallographic characterization such as Fourier Transform infrared spectroscopy, and environmental stability; physical characterization such as coverage of substrate, thickness, porosity, color, surface topography, and density; and mechanical characterization such as tensile bond strength, shear strength, and fatigue strength. The test specimen fabrication and contact with calcium phosphate coatings are also detailed.
SCOPE
1.1 This specification covers the material requirements for calcium phosphate coatings for surgical implant applications.  
1.2 In particulate and monolithic form, the calcium phosphate materials system has been well characterized regarding biological response (1, 2)2 and laboratory characterization (2-4). Several publications (5-10) have documented the in vitro and in vivo properties of selected calcium phosphate coating systems.  
1.3 This specification covers hydroxylapatite coatings, other calcium phosphate (for example, octacalcium calcium phosphate, amorphous calcium phosphate, dicalcium phosphate dihydrate) coatings, or a coating containing a combination of two or more calcium phosphate phases, with or without intentional minor additions of other elements or compounds (for example, fluorine, manganese, magnesium, carbonate),3 and applied by methods including, but not limited to, the following: (1) plasma spray deposition, (2) solution precipitation, (3) dipping/sintering, (4) electrophoretic deposition, and (5) sputtering.  
1.4 For a coating containing two or more calcium phosphate phases, one or more of which will be a major phase or major phases in the coating, while the other phase(s) may occur as a second or minor phases, the phase composition(s) of the coating should be determined against each corresponding crystalline phase, respectively. See X1.2.  
1.5 Substrates may include smooth, porous, textured, and other implantable topographical forms.  
1.6 This specification excludes organic coatings that may contain calcium and phosphate ionic species.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    5 pages
    English language
  • Technical specification
    5 pages
    English language

SIGNIFICANCE AND USE
4.1 The purpose of this test guide is to provide load profile information on how one could test a total knee replacement in order to evaluate in vitro its function and wear during several types of knee motions as described in 4.2 and 4.3.  
4.2 This test guide may help characterize the magnitude and location of implant wear as an implant is repetitively moved according to specified load and displacement waveforms.  
4.3 This test guide may also help characterize the functional limitations of a total knee replacement as its motion is guided by these waveforms. These limitations may be observed as impingement, subluxation, or high loading in the soft tissue constraints, whether they are represented physically or virtually.  
4.4 The motions and load conditions in vivo will, in general, differ from the load and motions defined in this guide. The results obtained from this guide cannot be used to directly predict in vivo performance. However, this guide is designed to allow for comparisons in performance of different knee designs, when tested under similar conditions.
SCOPE
1.1 Motion path, load history, and loading modalities all contribute to the wear, degradation, and damage of implanted prosthetics. Simulating a variety of functional activities promises more realistic testing for wear and damage mode evaluation. Such activities are often called activities of daily living (ADLs). ADLs identified in the literature include walking, stair ascent and descent, sit-to-stand, stand-to-sit, squatting, kneeling, cross-legged sitting, into bath, out of bath, turning, and cutting motions (1-7).2 Activities other than walking gait often involve an extended range of motion and higher imposed loading conditions, which have the ability to cause damage and modes of failure other than normal wear (8-10).  
1.2 This document provides guidance for functional simulation that could be used to evaluate in vitro the durability of knee prosthetic devices under force control.  
1.3 Function simulation is defined as the reproduction of loads and motions that might be encountered in activities of daily living, but it does not necessarily cover every possible type of loading. Functional simulation differs from typical wear testing in that it attempts to exercise the prosthetic device through a variety of loading and motion conditions such as might be encountered in situ in the human body in order to reveal various damage modes and damage mechanisms that might be encountered throughout the life of the prosthetic device.  
1.4 Force control is defined as the mode of control of the test machine that accepts a force level as the set point input and which utilizes a force feedback signal in a control loop to achieve that set point input. For knee simulation, the flexion motion is placed under angular displacement control, internal and external rotation is placed under torque control, and axial load, anterior-posterior shear, and medial-lateral shear are placed under force control.  
1.5 This document establishes kinetic and kinematic test conditions for several activities of daily living, including walking, turning navigational movements, stair climbing, stair descent, and squatting. The kinetic and kinematic test conditions are expressed as reference waveforms used to drive the relevant simulator machine actuators. These waveforms represent motion, as in the case of flexion extension, or kinetic signals representing the forces and moments resulting from body dynamics, gravitation, and the active musculature acting across the knee.  
1.6 This document does not address the assessment or measurement of damage modes, or wear or failure of the prosthetic device.  
1.7 This document is a guide. As defined by ASTM in their “Form and Style for ASTM Standards” book in section C15.2, “A standard guide is a compendium of information or series of options that does not recommend a specific course of action. Guides are intended ...

  • Guide
    34 pages
    English language
  • Guide
    34 pages
    English language

ABSTRACT
This specification provides requirements for materials, finish and marking, care and handling, and the acceptable dimensions and tolerances for metallic bone screws that are implanted into bone. There are a large variety of medical bone screws currently in use, the following type of screws are used: type HA - spherical undersurface of head, shallow, asymmetrical buttress thread, and deep screw head, type HB - spherical undersurface of head, deep, asymmetrical buttress thread, and shallow screw head, type HC - conical undersurface of head, symmetrical thread, and type HD - conical undersurface of head, symmetrical thread. The torsional strength, breaking angle, axial pullout strength, insertion torque, self-tapping force, and removal torque shall be tested to meet the requirements prescribed.
SIGNIFICANCE AND USE
A1.1 Significance and Use
A1.1.1 This test method is used to measure the torsional yield strength, maximum torque, and breaking angle of the bone screw under standard conditions. The results obtained in this test method are not intended to predict the torque encountered while inserting or removing a bone screw in human or animal bone. This test method is intended only to measure the uniformity of the product tested or to compare the mechanical properties of different, yet similarly sized, products.
SCOPE
1.1 This specification provides requirements for materials, finish and marking, care and handling, and the acceptable dimensions and tolerances for metallic bone screws that are implanted into bone. The dimensions and tolerances in this specification are applicable only to metallic bone screws described in this specification.  
1.2 This specification provides performance considerations and standard test methods for measuring mechanical properties in torsion of metallic bone screws that are implanted into bone. These test methods may also be applicable to other screws besides those whose dimensions and tolerances are specified here. The following annexes are included:  
1.2.1 Annex A1—Test Method for Determining the Torsional Properties of Metallic Bone Screws.  
1.2.2 Annex A2—Test Method for Driving Torque of Medical Bone Screws.  
1.2.3 Annex A3—Test Method for Determining the Axial Pullout Load of Medical Bone Screws.  
1.2.4 Annex A4—Test Method for Determining the Self-Tapping Performance of Self-Tapping Medical Bone Screws.  
1.2.5 Annex A5—Specifications for Type HA and Type HB Metallic Bone Screws.  
1.2.6 Annex A6—Specifications for Type HC and Type HD Metallic Bone Screws.  
1.2.7 Annex A7—Specifications for Metallic Bone Screw Drive Connections.  
1.3 This specification is based, in part, upon ISO 5835, ISO 6475, and ISO 9268.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 Multiple test methods are included in this standard. However, the user is not necessarily obligated to test using all of the described methods. Instead, the user should only select, with justification, test methods that are appropriate for a particular device design. This may only be a subset of the herein described test methods.  
1.6 This standard may involve the use of hazardous materials, operations, and equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    22 pages
    English language
  • Technical specification
    22 pages
    English language

SIGNIFICANCE AND USE
5.1 The current hip simulator wear test standards (ISO 14242-1 or ISO 14242-3) stipulate only one load waveform and one set of articulation motions. There is a need for more versatile and rigorous wear test regimes, but the knowledge of what represents realistic high demand wear test features is limited. More research is clearly needed before a standard that defines what a representative high demand wear test should include can be written. The objective of this guide is to advise researchers on the possible high demand wear test features that should be included in evaluation of hard-on-hard articulations.  
5.2 This guide makes suggestions of what high demand test features may need to be added to an overall high demand wear test regime. The features described here are not meant to be all inclusive. Based on current knowledge they appear to be relevant to adverse conditions that can occur in clinical use.  
5.3 All the test features, both conventional and high demand, could have interactive effects on the wear of the components.
SCOPE
1.1 The objective of this guide is to advise researchers on the possible high demand wear test features that should be included in evaluation of hard-on-hard articulations. This guide makes suggestions for high demand test features that may need to be added to an overall wear test regime. Device articulating components manufactured from other metallic alloys, ceramics, or with coated or elementally modified surfaces without significant clinical use could possibly be evaluated with this guide. However, such materials may include risks and failure mechanisms that are not addressed in this guide.  
1.2 Hard-on-hard hip bearing systems include metal-on-metal (for example, Specifications F75, F799, and F1537; ISO 5832-4, ISO 5832-12), ceramic-on-ceramic (for example, ISO 6474-1, ISO 6474-2, ISO 13356), ceramic-on-metal, or any other bearing systems where both the head and cup components have high surface hardness. An argument has been made that the hard-on-hard THR articulation may be better for younger, more active patients. These younger patients may be more physically fit and expect to be able to perform more energetic activities. Consequently, new designs of hard-on-hard THR articulations may have some implantations subjected to more demanding and longer wear performance requirements.  
1.3 Total Hip Replacement (THR) with metal-on-metal articulations have been used clinically for more than 50 years (1, 2).2 Early designs had mixed clinical results. Eventually they were eclipsed by THR systems using metal-on-polyethylene articulations. In the 1990s the metal-on-metal articulation again became popular with more modern designs (3), including surface replacement.  
1.4 In the 1970s the first ceramic-on-ceramic THR articulations were used. In general, the early results were not satisfactory (4, 5). Improvement in alumina, and new designs in the 1990s improved the results for ceramic-on-ceramic articulations (6).  
1.5 The values stated in SI units are to be regarded as the standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    8 pages
    English language
  • Guide
    8 pages
    English language

SIGNIFICANCE AND USE
5.1 This guide describes the use of test methods in Guide F3276 to assess the compatibility of a brush part with the material substrates used in the construction of medical devices.  
5.2 By providing objective, repeatable methods for evaluating performance under test conditions, this standard can improve the ability to assess the effectiveness of various brush part designs.
SCOPE
1.1 This guide describes methods for assessing the compatibility, under prescribed laboratory conditions, of a cleaning brush part with substrate materials used in the construction of medical devices. The method utilizes a force tester to mechanically actuate a brush part at a constant rate. This action continues until there is any level of visible degradation, including but not limited to scratching or shaving of the substrate material.  
1.2 The test methods utilized in this guide are those described in Guide F3276. In this guide, the number of repetitions is determined by the demonstrable degradation, if any, of the substrate, up to a specified maximum number of repetitions.  
1.3 Brushes designed to clean medical devices after clinical use play an important role in the effective reprocessing of those medical devices. Instructions for use from the brush manufacturer should supply information related to the compatibility, or more importantly, incompatibility with materials that make up the composition of a medical device. This may be stated in terms of being suitable for specific materials, not suitable for specific materials, or suitable for a limited number of uses for specific materials.  
1.4 Selecting the correct brush for the medical device to be cleaned is always a key factor to achieve effective cleaning. One of the significant factors when selecting a brush is selecting one that will not cause damage to the medical device, including the material the medical device is constructed of. Assessing if a brush part could damage a medical device because of the material the device is made of is an important step in determining the appropriate (and inappropriate) use of a brush.  
1.5 Units—The values stated in SI units are to be regarded as the standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    3 pages
    English language
  • Guide
    3 pages
    English language

ABSTRACT
This specification covers the requirements for unfinished cobalt-28chromium-6molybdenum (UNS R30075) investment product alloy castings for surgical implant applications, and casting alloys of the same in the form of shot, bar, or ingots to be used in the manufacture of surgical implants. This specification does not apply to completed surgical implants made from castings. Both product castings and casting alloys shall conform to specified chemical composition and mechanical requirements including ultimate tensile strength, yield strength, elongation, and reduction of area. Product castings shall additional undergo liquid penetrant, radiographic, metallographic, and hardness examination.
SCOPE
1.1 This specification covers the chemical, mechanical, and metallurgical requirements for cobalt-28 chromium-6 molybdenum alloy unfinished investment product castings for surgical implant applications and casting alloy in the form of shot, bar, or ingots to be used in the manufacture of surgical implants. This specification does not apply to completed surgical implants made from castings.  
1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    5 pages
    English language
  • Technical specification
    5 pages
    English language

ABSTRACT
This specification covers chemical and crystallographic requirements for biocompatible beta-tricalcium phosphate for surgical implant applications. Elemental analysis for calcium and phosphorus will be consistent with the expected stoichiometry of beta-tricalcium phosphate. The calcium and phosphorus content shall be determined using a suitable method such X-ray fluorescence. A quantitative X-ray diffraction analysis shall indicate a minimum beta-tricalcium phosphate content of 95 % as determined using powder diffraction method. The analysis of other trace elements may be required, based on the conditions, apparatus, or environment. It is recommended that all metals or oxides present in concentrations equal or greater than 0.1 % be noted in material descriptions.
SCOPE
1.1 This specification covers chemical and crystallographic requirements for beta-tricalcium phosphate (β-TCP) raw materials intended for use in medical device applications. For a material to be identified as medical-grade beta-tricalcium phosphate, it must conform to this specification (see Appendix X1).  
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    4 pages
    English language
  • Technical specification
    4 pages
    English language

ABSTRACT
This specification covers chemical and crystallographic requirements for hydroxylapatite intended for surgical implants. Elemental analysis for calcium and phosphorus will be consistent with the expected stoichiometry of hydroxylapatite. The calcium and phosphorus contents shall be determined using a suitable method such as ion chromatography. A quantitative X-ray diffraction analysis shall indicate a minimum hydroxylapatite content of 95 %. The concentration of trace elements such as arsenic, cadmium, mercury, and lead shall be determined for hydroxylapatite derived from natural resources. The analysis of other trace elements may be required, based on the conditions, apparatus, or environments specific to the manufacturing techniques and raw materials. Either inductively coupled plasma/mass spectroscopy (ICP/MS), atomic absorption (AAS), or the methods mentioned shall be used.
SCOPE
1.1 This specification covers chemical and crystallographic requirements for hydroxylapatite intended for surgical implants. For a material to be called medical-grade hydroxylapatite, it must conform to this specification. (See Appendix X1.)  
1.2 The biological response to hydroxylapatite in soft tissue and bone has been characterized by a history of clinical use (1-3)2 and by laboratory studies  (4-6).  
1.3 This specification includes powder, particulate, and forms intended for use as surgical implants, components of surgical implants, or as raw materials for manufacturing processes such as thermal spray coating, electrophoretic deposition, physical vapor deposition, and so forth.  
1.4 This specification specifically excludes hydroxylapatite coatings, amorphous calcium phosphate, ceramic-glasses, tribasic calcium phosphate, whitlockite, and alpha- and beta-tricalcium phosphate (see Specification F1088).  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    4 pages
    English language
  • Technical specification
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 This standard guide is to be used to help assess the biocompatibility of materials used in medical devices (for example, externally communicating, implants, and other body contact medical devices). It is designed to test the effect of particles and other wear debris and/or degradation products on the generation of FBR and other (local and systemic) host responses of immune/inflammatory origin.  
5.2 The appropriateness of the selected testing methods should be carefully considered by the user since not all materials or applications need to be tested by this guide. Existing biocompatibility screening methods may not be fully predictive of the human response, and testing approaches such as those described here are needed for continuous improvement of the predictability of biocompatibility testing. The effectiveness of animal testing in terms of its predictability of human outcomes is dependent on the study design. If possible, study endpoints should be chosen to minimize interspecies variability and to investigate clinically relevant biological responses. While testing approaches should remain at the user’s discretion, the following should be taken into consideration when selecting most appropriate tests and study endpoints.  
5.2.1 Device-induced responses usually involve both innate and adaptive immunities, which raises possible need for specific testing for each of these immune response types.
5.2.1.1 Device-related adaptive immune responses are mostly due to lymphocyte-mediated delayed-type hypersensitivity. In vivo allergenicity to a test material (which can be introduced via different routes) should be assessed by monitoring for any signs of allergic and acute toxicity reactions, for example, scratch, tremor, and dyspnea. In addition, ex vivo analysis on immunophenotyping of the isolated splenocytes/lymphocytes from the same studies should be considered.
5.2.1.2 Device-related innate immune responses are mostly mediated by macrophages and can be assess...
SCOPE
1.1 The purpose of this standard guide is to describe the principles and approaches to testing of medical device debris and degradation products from device materials (for example, particles from wear) for their potential to activate a cascade of biological responses at local and systemic levels in the body. In order to ascertain the role of device debris and degradation products in stimulating such responses, the nature of the responses and the consequences of the responses should be evaluated. This is an emerging area. The continuously updated information gained from the testing results and related published literature is necessary to improve the study designs, as well as predictive value and interpretation of the test results regarding debris/degradation product related responses. Some of the procedures listed here may, on further testing, not prove to be predictive of clinical responses to device-related debris and degradation products. However, only the continuing use of standard protocols will establish the most useful testing approaches with reliable study endpoints and measurement techniques. Since there are many possible and established ways of determining the debris/degradation product related responses in vivo, a single standard protocol is not stated. However, this recommended guide indicates which testing approaches are most applicable per expected biological responses and which necessary information should be supplied with the test results. To address the general role of chronic inflammation in exaggerating device-related foreign body response (FBR), the recommendations in this standard include the assessment of device-related pro-inflammatory responses and subsequent tissue remodeling potential.  
1.2 This document is to provide the users with updated scientific knowledge that may help better characterize medical device debris related responses. It is to help the users to optimize their plans for particle...

  • Guide
    10 pages
    English language
  • Guide
    10 pages
    English language

ABSTRACT
This specification covers the chemical, mechanical, and metallurgical requirements for wrought cobalt-20chromium-15tungsten-10nickel alloy bars, rods, wires, sheets, and strips (except surgical fixation wires) for use in surgical implants. All alloys shall be furnished to the purchaser's specifications, in the annealed or cold-worked condition. Then, bars and wires shall be finished bright annealed, cold drawn, pickled, ground, or ground and polished, as specified by the purchaser. While, sheets shall be furnished bright annealed, pickled, cold-rolled, or polished, as specified by the purchaser. The alloys shall adhere to specified values of ultimate tensile strength, yield strength, elongation, and microcleanliness.
SCOPE
1.1 This specification covers the chemical, mechanical, and metallurgical requirements for wrought cobalt-20chromium-15tungsten-10nickel alloy used for surgical implants. The properties specified apply specifically to wrought bar, rod, wire, sheet, and strip, but do not apply to surgical fixation wire (see Specification F1091).  
1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of the other, and values from the two systems shall not be combined.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    5 pages
    English language
  • Technical specification
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 This practice is a guideline for a screening test of candidate materials or assessment of local tissue response to absorbable medical devices which are expected to undergo complete absorption within three years.  
5.2 This practice is similar to those for studies on candidate materials or medical devices that are not absorbable, such as those specified in Practices F763, F981, and F1408; however, analysis of the host response must take into account the effect of degradation and degradation products on the inflammatory response at the local tissue site and on subsequent healing of the implantation site, as well as the potential for adverse distal tissue effects.  
5.3 For testing of absorbable medical devices, the test article for implantation should be in the final finished form as for intended use, including packaging and sterilization (if applicable). Configurations specific to the animal study may be needed. The test article’s surface-area-to-body mass or mass-to-body mass ratios within the animal model should be established by calculating based on surface-area-to-body mass or mass-to-body mass ratios in humans during the device’s intended clinical use. Worst-case clinical dose should be considered in the study design. For implantation studies incorporating evaluation of both local tissue responses and systemic toxicity, exaggerated material surface area or mass-to-body mass ratios (for example, a 2X to 10X safety factor to assess implant safety for regulatory submissions) compared to clinical use (for example, largest device size, maximum number of devices) should be considered, unless otherwise justified. For example, implantation of exaggerated doses may not be feasible in the selected animal model. For some devices, additional animal group(s) for exaggerated conditions should be considered if dose response information is needed. Additionally, for some devices, exaggerated dose at a specific implantation site can also be used to evaluate local tissue res...
SCOPE
1.1 This practice provides experimental protocols for biological assays of tissue reactions to absorbable biomaterials for implant applications. This practice applies only to absorbable materials with projected clinical applications in which the materials will reside in bone or soft tissue longer than 30 days and less than three years. Other standards with designated implantation times are available to address shorter time periods. Careful consideration should be given to the appropriateness of this practice for slowly degrading materials that will remain for longer than three years. It is anticipated that the tissue response to degrading biomaterials will be different from the response to nonabsorbable materials. In many cases, a chronic inflammatory response may be observed during the degradation phase, but the local histology should return to normal after absorption; therefore, the minimal tissue response usually equated with biocompatibility may require long implantations.  
1.2 The time period for implant absorption can depend on variables of chemical composition, implant size, implant location, and animal models. Therefore, the selected time points for assessing tissue effects may be selected based on the rate of absorption.  
1.3 These protocols assess the effects of the material on the animal tissue in which it is implanted. They do not fully assess systemic toxicity, carcinogenicity, reproductive and development toxicity, or mutagenicity of the material. Other standards are available to address these issues.  
1.4 To maximize use of the animals in the study protocol, some aspects of systemic toxicity, including effects of degradation products on different organs and tissues downstream of or surrounding the target site, can be addressed with this practice.  
1.5 Because animal models are not identical to human biology, this practice cannot account for all potential biological hazards, for example the effect o...

  • Standard
    7 pages
    English language
  • Standard
    7 pages
    English language

SIGNIFICANCE AND USE
4.1 Interactions of items with the MR environment have resulted in serious injuries and death of patients and other individuals. This practice lists hazards that may be present in the MR environment. It specifies marking of items anticipated to enter the MR environment and recommends information that should be included in the associated labeling.  
4.2 This practice provides a uniform system of visual icons and terms for marking items for use in the MR environment.
SCOPE
1.1 This practice applies to medical devices and other items that are anticipated to enter the magnetic resonance (MR) environment.
Note 1: “Medical devices and other items” will be referred to as “items” for the remainder of this practice.  
1.2 The practice specifies the marking of items anticipated to enter the MR environment by means of terms and icons, and recommends information that should be included in the labeling.  
1.3 MR image artifacts are not in the scope of the mandatory portions of this practice because they do not present a direct safety issue resulting from specific characteristics of the MR examination (see X1.12).  
1.4 The values stated in SI units are to be regarded as standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    15 pages
    English language

SIGNIFICANCE AND USE
5.1 This test method is intended to assess a combination of inherent chemical irritation and mechanical irritation for products and materials expected to come into contact with the skin. It is a comparative approach whereby the potential irritation of a test material is compared to that of a reference material similar in form and composition. The reference material should have a known safety and irritation profile.
SCOPE
1.1 The behind-the-knee (BTK) method, using the popliteal fossa of human volunteers as a test site, simultaneously evaluates the inherent chemical irritation and the potential for mechanical irritation of substrates and products that are designed to come into repeated or extended close contact with the skin (see validation references (1-7)).2 This is a bilateral test comparing a test material to a reference material with a known safety profile.  
1.2 This test method shall be used by qualified health care professionals experienced in good clinical practice (GCP) procedures.  
1.3 This test method can be performed using human subjects on either intact or compromised skin. Testing should be performed on intact skin for test substrates or products expected to have contact with normal, intact skin, or for direct comparison to products with a known skin irritation profile. Testing can be performed on compromised skin for test substrates or products that may commonly come into contact with damaged skin (for example, skin with diaper rash, or chapped skin) or skin that is expected to be hydrated.  
1.4 Visual scoring of erythema and dryness is performed by a trained skin grader on a predefined scale.  
1.5 Prior to use in this test, materials shall undergo overall favorable biocompatibility testing consistent with the approach outlined in protocol Practice F748 or ISO 10993-1:2009. As a part of this series of testing, irritation per Practice F719 or ISO 10993-10 shall be conducted.  
1.6 The values stated in inch-pound units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    24 pages
    English language
  • Standard
    24 pages
    English language

ABSTRACT
This specification covers the chemical, mechanical, and metallurgical requirements for wrought titanium-15 molybdenum alloy to be used in the manufacture of surgical implants. The product shall be classified as: strip; sheet; plate; bar; or wire. The heat analysis shall conform to the chemical composition prescribed. Material shall be furnished in the beta annealed condition. Bend tests and tension tests shall be performed to conform to the specified requirements.
SCOPE
1.1 This specification covers the chemical, mechanical, and metallurgical requirements for wrought titanium-15 molybdenum alloy to be used in the manufacture of surgical implants (1).2  
1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of each other, and values from the two systems shall not be combined.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    6 pages
    English language
  • Technical specification
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 Due to the variety of small bone fractures, plates used for the fixation of these fractures come in a variety of shapes and configurations. Table 1 categorizes the plate types for each anatomical area. Flat plates are the simplest; see Fig. 2 for an example of a basic flat plate. Many other plates have features to accommodate specific anatomies, such as condylar, complex (such as cuneiform), pre-contoured (such as metatarsophalangeal joint (MPJ)), step, orbital, orthognathic step, and wedge plates. Other plates, such as mesh-based and burr hole plates, are generally flat but are designed to be used in specific anatomical regions, so their designs are not the same as conventional straight plates. If test data is used from one type of plate for justification of the mechanical properties of another type of plate, this justification shall be described in the final report.  
4.2 Most of the testing described herein is focused on a “functional unit,” which can be described as a single-line fracture being spanned by a plate with one screw hole on each side of the fracture. This configuration allows for the simplest determination of worst-case size if the strut geometry is the determining factor for the worst case. If a worst-case size cannot be isolated to a functional unit/strut geometry, perhaps due to irregular screw hole patterns or the shape of the plate, it is understandable that some tests would need to be modified, or possibly removed from test consideration, to accommodate the shape of the plate or the screw hole. Any test modifications or omissions shall be described in the final report with a rationale related to the plate’s anatomical use, indications, and functional requirements.
SCOPE
1.1 This standard is intended to provide guidance for the static testing of small bone metallic plates used for fracture fixation. Small bone plates referred to in this standard would be used in minimally load-bearing anatomical areas of the far extremities, such as the fingers and toes, and in the cranium and upper face. Lower face/mandible, wrist, and ankle fixation plates would generally be larger and carry a substantial amount of load and should not be evaluated under this standard.  
1.2 ASTM Specification F382 and ISO 9585 are currently available for the testing of metallic bone plates as well, so the user can choose to use any of the tests in these standards for small bone plates. However, due to plate size, Specification F382 and ISO 9585 test setup and execution difficulty can be increased for small bone plates. Thus, this standard offers alternative test methods that are more appropriate for metallic bone plates used in small bone fracture fixation.  
1.3 This standard is not intended to address the mechanical performance of the plating construct or accessory components (for example, screws and wires).  
1.4 This standard is intended to provide a basis for the mechanical comparison of small bone plates. Due to the complex and varying biomechanics found in the areas of the body where these plates are used, this standard should only be used to compare the in vitro mechanical performance of small bone plates and not used to infer in vivo performance characteristics.  
1.5 This standard describes static tests by specifying load types and specific methods of applying these loads. Tests for evaluating and characterizing these loads include the following: static torsion, static cantilever beam bending, static lateral bending, and static three-point bending.  
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 Multiple tests are cited in this standard. However, it must be noted that the user is not obligated to test using all of the described methods. Instead, the user should only select test methods that are appropriate for a particular device design.  
1.8 This standard does not purport to address all of th...

  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
4.1 This test method can be used to describe the effects of materials, manufacturing, and design variables on the performance of metal or composite-backed anatomic glenoid prostheses’ locking mechanisms to resist static shear loading.  
4.2 The glenoid component is used in shoulder replacements and should conform to the criteria specified in Specification F1378.  
4.3 The loading of metal or composite-backed anatomic glenoid prostheses in vivo will, in general, differ from the loading defined in this test method. The results obtained here cannot be used to directly predict in vivo performance. However, this test method is designed to allow for comparisons between different metal or composite-backed anatomic glenoid locking mechanism designs, when tested under the same testing conditions.  
4.4 This test method may not be appropriate for all types of implant applications. The user is cautioned to consider the appropriateness of the method in view of the materials being tested and their potential application.  
4.5 In order for the test data on metal or composite-backed anatomic glenoid components to be comparable, reproducible, and capable of being correlated among laboratories, it is essential that uniform procedures be established.
SCOPE
1.1 This test method covers a method for determining the static shear disassembly force of modular anatomic glenoid components used in anatomic total shoulder arthroplasty prostheses.  
1.2 Although the methodology described does not replicate all physiological force conditions, it is a means of in vitro comparison of modular anatomic glenoid component designs and the strength of the retention mechanism between the articular insert and glenoid backing under the stated test conditions.  
1.3 This test method covers modular glenoid components comprised of a separate articular insert and backing. The insert and backing can be fabricated from any combination of the following materials: metal alloys, polymeric materials, composite materials.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    4 pages
    English language
  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 Multi-axis force measuring platforms are used to measure the ground reaction forces produced at the interface between a subject's foot or shoe and the supporting ground surface. These platforms are used in various settings ranging from research laboratories to healthcare facilities. The use of force platforms has become particularly important in gait analysis where clinical evaluations have become a billable clinical service.  
5.2 Of particular importance is the application of force platforms in the treatment of cerebral palsy (CP) (1, 2).3 An estimated 8000 to 10 000 infants born each year will develop CP (3) while today’s affected population is over 764 000 patients (4). Quantitative gait analysis, using force platforms and motion capture systems, provides a valuable tool in evaluating the pathomechanics of children with CP. This type of mechanical evaluation provides a quantitative basis for treating neuromuscular conditions. In other words, surgical decisions are in part guided by information gained from the use of force platform measurements (5, 6).  
5.3 Another application is treatment of spina bifida. According to the Gait and Clinical Movement Analysis Society (GCMAS) (7), an instrumented gait analysis is the standard of expert care for children with gait abnormalities secondary to spina bifida. The main objective of diagnostic gait analysis is to define the pathological consequences of neural tube defects as they relate to gait. The use of instrumented gait analysis allows physicians to determine which surgical or non-surgical interventions would provide the best outcome.  
5.4 More recently, force platforms have been used for pre- and post-surgical evaluation of TKA (total knee arthroplasty) and THA (total hip arthroplasty) patients. Such data provides an objective measure of the mechanical outcome of the surgical procedure.  
5.5 In addition to the clinical applications there are numerous medical and human performance research activities which r...
SCOPE
1.1 This standard recommends practices for performance verification of multi-axis force platforms commonly used for measuring ground reaction forces during gait, balance, and other activities.  
1.1.1 This standard provides a method to quantify the relationship between applied input force and force platform output signals across the manufacturer’s defined spatial working surface and specified force operating range.  
1.1.2 This standard provides definitions of the critical parameters necessary to quantify the behavior of multi-axis force measuring platforms and the methods to measure the parameters.  
1.1.3 This standard presents methods for the quantification of spatially distributed errors and absolute measuring performance of the force platform at discrete spatial intervals and discrete force levels on the working surface of the platform.  
1.1.4 This standard further defines certain important derived parameters, notably COP (center of pressure) and methods to quantify and report the measuring performance of such derived parameters at spatial intervals and force levels across the working range of the force platform.  
1.1.5 This standard defines the requirements for a report suitable to characterize the force platform’s performance and provide traceable documentation to be distributed by the manufacturer or calibration facility to the users of such platforms.  
1.1.6 Dynamic characteristics and applications where the force platform is incorporated in other equipment, such as instrumented treadmills and stairs, are beyond the scope of this standard.  
1.1.7 This standard is written for purposes of multi-axis force platform verification. However, the methods and procedures are applicable to calibration of force platforms by manufacturers.  
1.2 The values stated in SI units are to be regarded as the standard. Other metric and inch-pound values are regarded as equivalent when required.  
1.3 This standard does n...

  • Standard
    10 pages
    English language
  • Standard
    10 pages
    English language

SCOPE
1.1 This specification covers total elbow replacement (TER) prostheses and hemi-elbow replacement (“hemi”) prostheses used to provide functioning articulation by employing humeral, ulnar, and/or radial components that allow for the restoration of motion of the human elbow joint complex.  
1.2 Included within the scope of this specification are elbow prosthesis components for primary and revision surgery with linked and non-linked designs and components implanted with or without use of bone cement.  
1.3 This specification is intended to provide basic descriptions of material and prosthesis geometry. In addition, those characteristics determined to be important to the in vivo performance of the prosthesis are defined. However, compliance with this specification does not itself mean that a device will provide satisfactory clinical performance.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    9 pages
    English language
  • Technical specification
    9 pages
    English language

SIGNIFICANCE AND USE
5.1 These methods are intended to determine whether a material, product, or part of a product has the degree of radiopacity desired for its application as a medical device in the human body. This method allows for comparison with or without the use of a body mimic. Comparisons without the use of a body mimic should be used with caution as the relative radiopacity can be affected when imaging through the human body.  
5.2 These methods allow for both qualitative and quantitative evaluation in different comparative situations.
SCOPE
1.1 These test methods cover the determination of the radiopacity of materials and products utilizing X-ray based techniques, including fluoroscopy, angiography, CT (computed tomography), and DEXA (dual energy X-ray absorptiometry), also known as DXA, The results of these measurements are an indication of the likelihood of locating the product within the human body.  
1.2 Radiopacity is determined by (a) qualitatively comparing image(s) of a test specimen and a user-defined standard, with or without the use of a body mimic; or (b) quantitatively determining the specific difference in optical density or pixel intensity between the image of a test specimen and the image of a user-defined standard, with or without the use of a body mimic.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 This guide is aimed at providing a range of  in vivo models to aid in preclinical research and development of tissue-engineered medical products (TEMPs) intended for the clinical repair or regeneration of bone.  
4.2 This guide includes a description of the animal models, surgical considerations, and tissue processing as well as the qualitative and quantitative analysis of tissue specimens.  
4.3 The user is encouraged to use appropriate ASTM and other guidelines to conduct cytotoxicity and biocompatibility tests on materials, TEMPs, or both, prior to assessment of the in vivo models described herein.  
4.4 It is recommended that safety testing be in accordance with the provisions of the FDA Good Laboratory Practices Regulations 21 CFR 58.  
4.5 Safety and effectiveness studies to support regulatory submissions (for example, Investigational Device Exemption (IDE), Premarket Approval (PMA), 510K, Investigational New Drug (IND), or Biologics License Application (BLA) submissions in the U.S.) should conform to appropriate guidelines of the regulatory bodies for development of medical devices, biologics, or drugs, respectively.  
4.6 Animal model outcomes are not necessarily predictive of human results and should, therefore, be interpreted cautiously with respect to potential applicability to human conditions.
SCOPE
1.1 This guide covers general guidelines for the in vivo assessment of tissue-engineered medical products (TEMPs) intended to repair or regenerate bone. TEMPs included in this guide may be composed of natural or synthetic biomaterials (biocompatible and biodegradable) or composites thereof, and may contain cells or biologically active agents such as growth factors, synthetic peptides, plasmids, or cDNA. The models described in this guide are segmental critical size defects which, by definition, will not fill with viable tissue without treatment. Thus, these models represent a stringent test of a material’s ability to induce or augment bone growth.  
1.2 Guidelines include a description and rationale of various animal models including rat (murine), rabbit (leporine), dog (canine), goat (caprine), and sheep (ovine). Outcome measures based on radiographic, histologic, and mechanical analyses are described briefly and referenced. The user should refer to specific test methods for additional detail.  
1.3 This guide is not intended to include the testing of raw materials, preparation of biomaterials, sterilization, or packaging of the product. ASTM standards for these steps are available in the Referenced Documents (Section 2).  
1.4 The use of any of the methods included in this guide may not produce a result that is consistent with clinical performance in one or more specific applications.  
1.5 Other preclinical methods may also be appropriate and this guide is not meant to exclude such methods. The material must be suitable for its intended purpose. Additional biological testing in this regard would be required.  
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    14 pages
    English language

SIGNIFICANCE AND USE
5.1 The purpose of this practice is to provide data that can be used for comparison and evaluation of the accuracy of different CAS systems.  
5.2 The use of CAS systems and robotic tracking systems is becoming increasingly common and requires a degree of trust by the user that the data provided by the system meets necessary accuracy requirements. In order to evaluate the potential use of these systems, and to make informed decisions about suitability of a system for a given procedure, objective performance data of such systems are necessary. While the end user will ultimately want to know the accuracy parameters of a system under clinical application, the first step must be to characterize the digitization accuracy of the tracking subsystem in a controlled environment under controlled conditions.  
5.3 In order to make comparisons within and between systems, a standardized way of measuring and reporting point accuracy is needed. Parameters such as coordinate system, units of measure, terminology, and operational conditions must be standardized.
SCOPE
1.1 This standard will measure the effects on the accuracy of computer assisted surgery (CAS) systems of the environmental influences caused by equipment utilized for bone preparation during the intended clinical application for the system. The environmental vibration effect covered in this standard will include mechanical vibration from: cutting saw (sagittal or reciprocating), burrs, drills, and impact loading. The change in accuracy from detaching and re-attaching or disturbing a restrained connection that does not by design require repeating the registration process of a reference base will also be measured.  
1.2 It should be noted that one system may need to undergo multiple iterations (one for each clinical application) of this standard to document its accuracy during different clinical applications since each procedure may have different exposure to outside forces given the surgical procedure variability from one procedure to the next.  
1.3 All units of measure will be reported as millimeters for this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language

ABSTRACT
This specification covers total wrist implants that provide articulation function through radial carpal components. However, this specification does not include implants with ceramic coated or porous coated surfaces, one-piece elastomeric implants, and those used for custom applications. Implants covered by this specification are of the constrained, partially constrained, or unconstrained joint prosthesis type. The implants shall be manufactured from biocompatible materials made of unalloyed titanium, cobalt-28chromium-6molybdenum alloy, wrought cobalt-20chromium-15tungsten-10nickel alloy, wrought titanium-6aluminum-4vanadium extra low interstitial alloy, wrought 35cobalt-35nickel-20chromium-10molybdenum alloy, wrought cobalt-20nickel-20chromium-3.5molybdenum-3.5tungsten-5iron alloy, cobalt-28chromium-6molybdenum alloy, titanium-6aluminum-4vanadium alloy, or wrought cobalt-28chromium-6molybdenum alloy with dimensions in conformity to the requirements specified. Tests for polymeric creep (cold flow), wear, and range of motion of the device as well as in vitro laboratory testing shall be performed and shall conform to the requirements specified. Per service requirement, metallic implants shall undergo fluorescent penetrant inspection, while cast metallic implants shall undergo radiography.
SCOPE
1.1 This specification describes total wrist implants used to provide functioning articulation by employing radial and carpal components.  
1.2 This specification excludes those implants with ceramic-coated or porous-coated surfaces, one-piece elastomeric implants (with or without grommets), and those devices used for custom applications.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    6 pages
    English language
  • Technical specification
    6 pages
    English language

SCOPE
1.1 This specification covers chemical, mechanical, and metallurgical requirements for two types of metal injection molded (MIM) titanium-6aluminum-4vanadium components to be used in the manufacture of surgical implants.  
1.2 The Type 1 MIM components covered by this specification may have been densified beyond their as-sintered density by post sinter processing.  
1.3 Units—The values in either inch-pound or SI are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independent of the other. Combining values from the two systems may result in nonconformance with the specification.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    5 pages
    English language

ABSTRACT
This specification covers all single-use clinical thermometer probe covers and sheaths intended for use with any clinical thermometer. Requirements are given for safety, toxicity, handling, labeling, and physical integrity. Testing procedures for appropriate requirements and a glossary of terms used are provided. Toxicity test, leakage test, and compatibility test shall be performed to meet the requirements prescribed.
SCOPE
1.1 This specification covers all single-use clinical thermometer probe covers and sheaths intended for use with any clinical thermometer. Requirements are given for safety, toxicity, handling, labeling, and physical integrity. Testing procedures for appropriate requirements and a glossary of terms used within the standards are provided.  
1.2 The requirements contained herein are intended to ensure adequate isolation of the patient from the temperature-measuring device. In addition, the safety and health of the patient shall not be adversely affected. When used in accordance with the manufacturer’s instructions, the probe cover, sheath, and temperature-measuring device shall remit correct temperature readings as required in Specifications E667 and E1112.  
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Technical specification
    3 pages
    English language

Frequently Asked Questions

F04 is a Technical Committee within ASTM International. It is named "Medical and Surgical Materials and Devices" and is responsible for: The scope of the Committee shall be the development of standardized nomenclature and definitions of terms, test methods, recommended practices, guides, specifications and performance standards for medical and surgical materials and devices. The Committee will encourage research in this field and sponsor symposia, workshops and publications to facilitate the development of such standards. The Committee will promote liaison with other ASTM Committees and other organizations with mutual interests. This committee has published 1604 standards.

F04 develops ASTM standards in the area of Information technology. The scope of work includes: The scope of the Committee shall be the development of standardized nomenclature and definitions of terms, test methods, recommended practices, guides, specifications and performance standards for medical and surgical materials and devices. The Committee will encourage research in this field and sponsor symposia, workshops and publications to facilitate the development of such standards. The Committee will promote liaison with other ASTM Committees and other organizations with mutual interests. Currently, there are 1604 published standards from this technical committee.

ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.

A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.

Loading...