Standard Test Method for Measuring Recoil of Balloon-Expandable Stents

SCOPE
1.1 The purpose of this test is to quantify the percentage by which the diameter of a stent decreases from its expanded diameter while still on the delivery balloon to its relaxed diameter after deflating the balloon. This test is appropriate for stents manufactured from a material that is plastically deformed when the stent's diameter is increased from its predeployed size to its postdeployed size by mechanical means. This test may be performed in air at room temperature unless there is a known temperature dependence of the material, in which case, the temperature at which the test is conducted shall be stated in the report.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-Jan-2001
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM F2079-01 - Standard Test Method for Measuring Recoil of Balloon-Expandable Stents
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn. Contact ASTM
International (www.astm.org) for the latest information.
Designation: F 2079 – 01
Standard Test Method for
Measuring Recoil of Balloon-Expandable Stents
This standard is issued under the fixed designation F 2079; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope 4. Significance and Use
1.1 The purpose of this test is to quantify the percentage by 4.1 Minimal stent recoil is a desirable feature of a stent
which the diameter of a stent decreases from its expanded because it minimizes the maximum diameter to which a stent
diameter while still on the delivery balloon to its relaxed must be expanded to achieve its final relaxed diameter. A stent
diameter after deflating the balloon. This test is appropriate for having a high recoil must be expanded to a greater diameter to
stents manufactured from a material that is plastically de- achieve its final relaxed diameter than a stent having low
formed when the stent’s diameter is increased from its prede- recoil. Practically, excessive expansion of the vessel into which
ployed size to its postdeployed size by mechanical means. This the stent is to be implanted may cause tissue damage resulting
test may be performed in air at room temperature unless there in a poor immediate result or poor long-term outcome. Stent
is a known temperature dependence of the material, in which recoil is affected by intrinsic properties of the material used to
case, the temperature at which the test is conducted shall be construct the stent and the specific geometric design of the
stated in the report. stent; therefore, measuring stent recoil is an essential part of
1.2 This standard does not purport to address all of the evaluating the design.
safety concerns, if any, associated with its use. It is the
5. Apparatus
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica- 5.1 A means to inflate with noncompressible fluid, typically
water, the delivery balloon on which the stent is mounted. The
bility of regulatory limitations prior to use.
means used must be capable of achieving the pressure required
2. Terminology
to maintain the expanded diameter of the stent until it can be
2.1 Definitions: measured and may include a device to monitor pressure.
2.1.1 labeled diameter, n—desired final stent (inner/outer) 5.2 A means to measure the outer diameter of the stent
diameter after deflation of the delivery balloon. without deforming the stent. Typically, a calibrated optical
2.1.2 stent recoil, n—the amount, expressed as a percentage, system, which does not require contact with the stent, is used.
by which the diameter of a stent changes from the expanded The resolution of the measurement system shall be 0.01 mm or
diameter measured with the stent on the inflated delivery better. The accuracy of the system shall be 2 % of reading or
balloon to the final value measured after deflating the balloon. better.
3. Summary of Test Method 6. Sampling, Test Specimens, and Test Units
3.1 The delivery balloon bearing a sample device represen- 6.1 Unless otherwise justified, all samples selected for
tative of product that will be marketed either premounted, or testing should be taken from fully processed, clinical quality
mounted on the delivery balloon at the time of use, is inflated product. It is not required that these devices undergo terminal
to expand the stent to slightly greater than its labeled diameter. sterilization. Cosmetic rejects or other nonclinical samples may
The outer diameter of the stent is measured in a minimum of be used if the cause for rejection has been shown not to affect
two approximately orthogonal projections while the stent is stent recoil.
still on the inflated delivery balloon. The balloon is deflated 6.2 The number of specimens tested for each unique stent
and the outer diameter of the stent is remeasured in the same geometry should be sufficient to meet sampling requirements
projections at approximately the same location. for desired specification limits. If a single stent geometry is
intended to be used for more than one labeled diameter, recoil
shall be evaluated for test specimens expanded to the smallest
and largest diameters intended for that geometry.
This test method is under the jurisdiction of ASTM Committee F04 on Medical
6.3 Most stents are comprised of a repetitious continuous
and Surgical Materials and Devices and is the direct responsibility of Subcommittee
pattern or of repeating subunits. If stents are made longer by
F04.30 on Cardiovascular Standards.
Current edition approved Jan. 10, 2001. Published March 2001. additional repetition of the basic geometric pattern or by
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn. Contact ASTM
International (www.astm.org) for the latest information.
F2079–01
adding identical subunits, then recoil need be measured on only
wh
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.