Standard Specification for Disposable Glass Culture Tubes

ABSTRACT
This specification covers the requirements for disposable glass culture tubes for use in culturing applications. Disposable glass culture tubes shall be made of type I (borosilicate) or type II (soda-lime) glass. Tube design shall be of one-piece construction. Other requirements shall include top or open end finish requirement, bottom or closed end requirement, residual thermal stress requirement, workmanship, and tube dimensions.
SCOPE
1.1 This specification covers the requirements for disposable glass tubes suitable for general testing and culturing applications in blood banks, hematology, bacteriology, virology, and tissue culture laboratories.  
1.2 For practical purposes, the word “disposable” according to this specification and expected product performance expressed in this specification describes those disposable glass culture tubes that are to be used one time only. Any institution or individual who reuses a disposable glass culture tube must bear full responsibility for its safety and effectiveness.
1.3 For packaging standards, choose among the following: Specifications E920 or E921 or Practice E1133.

General Information

Status
Historical
Publication Date
31-Oct-2009
Current Stage
Ref Project

Relations

Buy Standard

Technical specification
ASTM E890-94(2009) - Standard Specification for Disposable Glass Culture Tubes
English language
2 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E890 −94(Reapproved 2009)
Standard Specification for
Disposable Glass Culture Tubes
This standard is issued under the fixed designation E890; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope E1133 Practice for Performance Testing of Packaged Labo-
ratory Apparatus for United States Government Procure-
1.1 This specification covers the requirements for dispos-
ments
able glass tubes suitable for general testing and culturing
applications in blood banks, hematology, bacteriology,
3. Materials
virology, and tissue culture laboratories.
3.1 Thedisposableglassculturetubesshallbemadeofglass
1.2 For practical purposes, the word “disposable” according
in accordance with the requirements of Type I, Class A or B
to this specification and expected product performance ex-
(borosilicate glass) or Type II (soda-lime glass) of Specifica-
pressed in this specification describes those disposable glass
tion E438.
culture tubes that are to be used one time only. Any institution
4. Design
or individual who reuses a disposable glass culture tube must
bear full responsibility for its safety and effectiveness. 4.1 The tubes shall be of one-piece construction in accor-
dance with Fig. 1 for shape, and any cross section of the tube,
1.3 For packaging standards, choose among the following:
taken in a plane perpendicular to the longitudinal axis, shall
Specifications E920 or E921 or Practice E1133.
preferably be circular.
2. Referenced Documents
4.2 The top or open end of the tube shall be smoothly
2.1 ASTM Standards: fire-polished at right angles to the horizontal axis of the tube.
C148 Test Methods for Polariscopic Examination of Glass
4.3 Thebottomorclosedendofthetubeshallbecompletely
Containers
closed and shall have a spherical radius inside and outside in
E438 Specification for Glasses in Laboratory Apparatus
accordance with Fig. 1. The wall thickness in the bottom shall
E671 Specification for Maximum Permissible Thermal Re-
be at least 66 ⁄3 % of the side wall thickness, but not more than
sidual Stress in Annealed Glass Laboratory Apparatus
210 % of the side wall thickness.
E920 Specification for Commercially Packaged Laboratory
4.4 Residual thermal stress shall not exceed 750 psi when
Apparatus
determined in accordance with Specification E671.
E921 Specification for Export Packaged Laboratory Appa-
ratus 4.5 Workmanship—The glass tube shall be free of defects
that impair serviceability. The glass tube shall be transparent,
clean, dry, and reasonably free of foreign material, loose or
This specification is under the jurisdiction of ASTM Committee E41 on
Laboratory Apparatus and is the direct responsibility of Subcommittee E41.01 on
embedded,
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.