This document specifies strategies for risk estimation and evaluation of biological harms with respect to:
genotoxicity;
carcinogenicity;
reproductive toxicity; and
developmental toxicity.
This document is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity has been established.
This document is not applicable to active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products which are covered by regulations.
NOTE            Guidance on selecting relevant biological effects for medical devices is covered in ISO 10993-1.

  • Standard
    53 pages
    English language
    e-Library read for
    ×1 day

This document specifies strategies for risk estimation and evaluation of biological harms with respect to:
genotoxicity;
carcinogenicity;
reproductive toxicity; and
developmental toxicity.
This document is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity has been established.
This document is not applicable to active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products which are covered by regulations.
NOTE            Guidance on selecting relevant biological effects for medical devices is covered in ISO 10993-1.

  • Standard
    53 pages
    English language
    e-Library read for
    ×1 day

This document provides requirements and recommendations on the conditions (means of delivery) and minimum required information to deliver when providing electronic instructions for use (eIFU) for in vitro diagnostic medical devices (IVDs). The content of the eIFU is not in the scope of this standard, as it is covered by the documents in the ISO 18113 series.

  • Standard
    6 pages
    English language
    sale 15% off

This document specifies requirements for reference measurement procedures (RMP) for measurands used in laboratory medicine.
This document applies to:
RMPs providing values of differential or rational quantities where each quantity value is a numerical value multiplied by a measurement unit. Annex A provides information on ordinal quantities and nominal properties;
any person, body or institution developing RMPs for measurands used in laboratory medicine.

  • Standard
    37 pages
    English language
    e-Library read for
    ×1 day

This document specifies requirements for certified reference materials (CRMs) of higher metrological order and the content of the supporting documentation and the calibration hierarchies as described in ISO 17511:2020, 5.2.1, 5.3.1, 5.4.1, 5.5.1, 5.6.1, 5.7.1. It is applicable to CRMs intended for use as either primary reference materials (PRMs), secondary calibrators or international conventional calibrators within calibration hierarchies appropriate for measurands used in laboratory medicine, or for applications as trueness controls. It also specifies requirements for determining the certified value of a CRM, including evaluation, and reporting of the assigned uncertainty.
This document is applicable primarily to CRMs with assigned property values where the property has a magnitude that can be expressed as a quantitative scalar number or ratio to a reference or refers to a counting scale as also described in ISO 17511:2020, Clause 1.
When a CRM includes multiple measurands, this document is applied to each of the certified quantity values present in the CRM.
Although intended to be applicable to producers of CRMs, this document is also useful for reference materials (RMs) that are not in conformity with the full metrological requirements of CRMs. For example, this document does not apply to an RM created by an in vitro diagnostic medical device (IVD MD) manufacturer for use as working calibrator or end-user calibrator within a calibration hierarchy traceable to a CRM, although some content can be useful in assessing its performance.

  • Standard
    32 pages
    English language
    e-Library read for
    ×1 day

This document specifies strategies for risk estimation and evaluation of biological harms with respect to: genotoxicity; carcinogenicity; reproductive toxicity; and developmental toxicity. This document is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity has been established. This document is not applicable to active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products which are covered by regulations. NOTE Guidance on selecting relevant biological effects for medical devices is covered in ISO 10993-1.

  • Standard
    41 pages
    English language
    sale 15% off
  • Standard
    45 pages
    French language
    sale 15% off

This document specifies requirements for reference measurement procedures (RMP) for measurands used in laboratory medicine.
This document applies to:
RMPs providing values of differential or rational quantities where each quantity value is a numerical value multiplied by a measurement unit. Annex A provides information on ordinal quantities and nominal properties;
any person, body or institution developing RMPs for measurands used in laboratory medicine.

  • Standard
    37 pages
    English language
    e-Library read for
    ×1 day

This document specifies requirements for certified reference materials (CRMs) of higher metrological order and the content of the supporting documentation and the calibration hierarchies as described in ISO 17511:2020, 5.2.1, 5.3.1, 5.4.1, 5.5.1, 5.6.1, 5.7.1. It is applicable to CRMs intended for use as either primary reference materials (PRMs), secondary calibrators or international conventional calibrators within calibration hierarchies appropriate for measurands used in laboratory medicine, or for applications as trueness controls. It also specifies requirements for determining the certified value of a CRM, including evaluation, and reporting of the assigned uncertainty.
This document is applicable primarily to CRMs with assigned property values where the property has a magnitude that can be expressed as a quantitative scalar number or ratio to a reference or refers to a counting scale as also described in ISO 17511:2020, Clause 1.
When a CRM includes multiple measurands, this document is applied to each of the certified quantity values present in the CRM.
Although intended to be applicable to producers of CRMs, this document is also useful for reference materials (RMs) that are not in conformity with the full metrological requirements of CRMs. For example, this document does not apply to an RM created by an in vitro diagnostic medical device (IVD MD) manufacturer for use as working calibrator or end-user calibrator within a calibration hierarchy traceable to a CRM, although some content can be useful in assessing its performance.

  • Standard
    32 pages
    English language
    e-Library read for
    ×1 day

This document specifies requirements for certified reference materials (CRMs) of higher metrological order and the content of the supporting documentation and the calibration hierarchies as described in ISO 17511:2020, 5.2.1, 5.3.1, 5.4.1, 5.5.1, 5.6.1, 5.7.1. It is applicable to CRMs intended for use as either primary reference materials (PRMs), secondary calibrators or international conventional calibrators within calibration hierarchies appropriate for measurands used in laboratory medicine, or for applications as trueness controls. It also specifies requirements for determining the certified value of a CRM, including evaluation, and reporting of the assigned uncertainty. This document is applicable primarily to CRMs with assigned property values where the property has a magnitude that can be expressed as a quantitative scalar number or ratio to a reference or refers to a counting scale as also described in ISO 17511:2020, Clause 1. When a CRM includes multiple measurands, this document is applied to each of the certified quantity values present in the CRM. Although intended to be applicable to producers of CRMs, this document is also useful for reference materials (RMs) that are not in conformity with the full metrological requirements of CRMs. For example, this document does not apply to an RM created by an in vitro diagnostic medical device (IVD MD) manufacturer for use as working calibrator or end-user calibrator within a calibration hierarchy traceable to a CRM, although some content can be useful in assessing its performance.

  • Standard
    20 pages
    English language
    sale 15% off
  • Standard
    22 pages
    French language
    sale 15% off

This document specifies requirements for reference measurement procedures (RMP) for measurands used in laboratory medicine. This document applies to: RMPs providing values of differential or rational quantities where each quantity value is a numerical value multiplied by a measurement unit. Annex A provides information on ordinal quantities and nominal properties; any person, body or institution developing RMPs for measurands used in laboratory medicine.

  • Standard
    25 pages
    English language
    sale 15% off
  • Standard
    27 pages
    French language
    sale 15% off

This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required.
This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs).
These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.

  • Standard
    53 pages
    English language
    e-Library read for
    ×1 day

This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K.
EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228].
NOTE            This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.

  • Standard
    135 pages
    English language
    e-Library read for
    ×1 day

This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required.
This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs).
These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.

  • Standard
    53 pages
    English language
    e-Library read for
    ×1 day

This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K.
EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228].
NOTE            This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.

  • Standard
    135 pages
    English language
    e-Library read for
    ×1 day

This document specifies good clinical practice (GCP) for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies the general requirements intended to
protect the rights, safety and well-being of human subjects, users or other persons,
ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
define the responsibilities of the sponsor and principal investigator, and
assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.
Other standards or national requirements can also apply to the investigational device(s) under consideration or the clinical investigation(s).
NOTE            For Software as a Medical Device (SaMD), where appropriate, demonstration of the analytical validity (the SaMD’s output is accurate for a given input), the scientific validity (the SaMD’s output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD’s output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [5]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document can consider whether either specific sections or requirements of this document, or both, can be applicable.

  • Standard
    106 pages
    English language
    e-Library read for
    ×1 day

This document specifies a process for a medical laboratory to identify and manage the risks to patients, laboratory workers and service providers that are associated with medical laboratory examinations. The process includes identifying, estimating, evaluating, controlling and monitoring the risks.
The requirements of this document are applicable to all aspects of the examinations and services of a medical laboratory, including the pre-examination, examination, and post-examination aspects including accurate transmission of examination results into the electronic medical record, as well as other technical and management processes described in ISO 15189.
This document does not specify acceptable levels of risk.
This document does not apply to risks from post-examination clinical decisions made by healthcare providers.
This document complements the management of risks affecting medical laboratory enterprises that are addressed by ISO 31000, such as business, economic, legal, and regulatory risks.

  • Standard
    92 pages
    English language
    e-Library read for
    ×1 day

This document specifies good clinical practice (GCP) for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies the general requirements intended to
protect the rights, safety and well-being of human subjects, users or other persons,
ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
define the responsibilities of the sponsor and principal investigator, and
assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.
Other standards or national requirements can also apply to the investigational device(s) under consideration or the clinical investigation(s).
NOTE            For Software as a Medical Device (SaMD), where appropriate, demonstration of the analytical validity (the SaMD’s output is accurate for a given input), the scientific validity (the SaMD’s output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD’s output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [5]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document can consider whether either specific sections or requirements of this document, or both, can be applicable.

  • Standard
    106 pages
    English language
    e-Library read for
    ×1 day

This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K. EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228]. NOTE This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.

  • Standard
    122 pages
    English language
    sale 15% off
  • Standard
    131 pages
    French language
    sale 15% off

This document specifies a process for a medical laboratory to identify and manage the risks to patients, laboratory workers and service providers that are associated with medical laboratory examinations. The process includes identifying, estimating, evaluating, controlling and monitoring the risks.
The requirements of this document are applicable to all aspects of the examinations and services of a medical laboratory, including the pre-examination, examination, and post-examination aspects including accurate transmission of examination results into the electronic medical record, as well as other technical and management processes described in ISO 15189.
This document does not specify acceptable levels of risk.
This document does not apply to risks from post-examination clinical decisions made by healthcare providers.
This document complements the management of risks affecting medical laboratory enterprises that are addressed by ISO 31000, such as business, economic, legal, and regulatory risks.

  • Standard
    92 pages
    English language
    e-Library read for
    ×1 day

This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required. This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs). These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.

  • Standard
    41 pages
    English language
    sale 15% off
  • Standard
    44 pages
    French language
    sale 15% off

This document specifies a process for a medical laboratory to identify and manage the risks to patients, laboratory workers and service providers that are associated with medical laboratory examinations. The process includes identifying, estimating, evaluating, controlling and monitoring the risks. The requirements of this document are applicable to all aspects of the examinations and services of a medical laboratory, including the pre-examination, examination, and post-examination aspects including accurate transmission of examination results into the electronic medical record, as well as other technical and management processes described in ISO 15189. This document does not specify acceptable levels of risk. This document does not apply to risks from post-examination clinical decisions made by healthcare providers. This document complements the management of risks affecting medical laboratory enterprises that are addressed by ISO 31000, such as business, economic, legal, and regulatory risks.

  • Standard
    83 pages
    English language
    sale 15% off
  • Standard
    92 pages
    French language
    sale 15% off

This document specifies good clinical practice (GCP) for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices. For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I). This document specifies the general requirements intended to protect the rights, safety and well-being of human subjects, users or other persons, ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results, define the responsibilities of the sponsor and principal investigator, and assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices. Other standards or national requirements can also apply to the investigational device(s) under consideration or the clinical investigation(s). NOTE For Software as a Medical Device (SaMD), where appropriate, demonstration of the analytical validity (the SaMD’s output is accurate for a given input), the scientific validity (the SaMD’s output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD’s output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [5]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD. This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document can consider whether either specific sections or requirements of this document, or both, can be applicable.

  • Standard
    85 pages
    English language
    sale 15% off
  • Standard
    85 pages
    English language
    sale 15% off
  • Standard
    91 pages
    French language
    sale 15% off
  • Standard
    91 pages
    French language
    sale 15% off

This document specifies requirements and provides recommendations for the pre-examination process of cell-free DNA (cfDNA) from body fluid specimens other than blood, including but not limited to the collection, handling, storage, transport, processing and documentation of human body fluids, such as urine, pleural effusions, ascites, cerebrospinal fluid (CSF), and saliva, intended for cfDNA examination. Processing includes multiple steps, such as centrifugation for specimen purification and isolation of cfDNA.
This document does not cover dedicated measures for cytohistological analysis of nucleated cells derived from body fluid, nor measures for preserving and handling of pathogens, and other bacterial or whole microbiome DNA in body fluids described.
Dedicated measures for preserving circulating cell-free DNA (ccfDNA) from blood are covered in ISO 20186-3.
This document is applicable to medical laboratories, health institutions including facilities collecting and handling specimens, laboratory customers, in vitro diagnostic examination developers and manufacturers, biobanks, institutions and organizations performing biomedical research, and regulatory authorities.
NOTE            International, national or regional regulations or requirements can also apply to specific topics covered in this document.

  • Standard
    33 pages
    English language
    e-Library read for
    ×1 day

This document specifies requirements and provides recommendations for the pre-examination process of cell-free DNA (cfDNA) from body fluid specimens other than blood, including but not limited to the collection, handling, storage, transport, processing and documentation of human body fluids, such as urine, pleural effusions, ascites, cerebrospinal fluid (CSF), and saliva, intended for cfDNA examination. Processing includes multiple steps, such as centrifugation for specimen purification and isolation of cfDNA.
This document does not cover dedicated measures for cytohistological analysis of nucleated cells derived from body fluid, nor measures for preserving and handling of pathogens, and other bacterial or whole microbiome DNA in body fluids described.
Dedicated measures for preserving circulating cell-free DNA (ccfDNA) from blood are covered in ISO 20186-3.
This document is applicable to medical laboratories, health institutions including facilities collecting and handling specimens, laboratory customers, in vitro diagnostic examination developers and manufacturers, biobanks, institutions and organizations performing biomedical research, and regulatory authorities.
NOTE            International, national or regional regulations or requirements can also apply to specific topics covered in this document.

  • Standard
    33 pages
    English language
    e-Library read for
    ×1 day
  • Standard
    6 pages
    English language
    sale 15% off
  • Standard
    6 pages
    French language
    sale 15% off
  • Standard
    6 pages
    French language
    sale 15% off

This document specifies the control and approval of in vitro diagnostic reagents used in animal health for immunological analyses with a qualitative expression of test results.
This document is applicable to diagnostic reagents, as a priority for infectious (bacterial, viral, fungal or parasitic) or prion diseases and associated animal species for which harmonization of practices in this area is needed, i.e. those for which the national, regional or international regulatory framework provides for the control of trade in animals and/or animal products and/or the definition of a health status (absence of infection) of areas, establishments or individuals. While all reagents designated by the competent authorities fall under the scope of this document, the authorities or any other animal health stakeholder can choose to derogate in specific and exceptional situations such as emerging, exotic or rare diseases.
This document is not applicable to all existing diagnostic reagents, in particular those for which certain parameters described in this document cannot be validly evaluated in accordance with international requirements, due, e.g. to the absence of a specific reference method and/or accessible and duly validated reference materials (RMs).
This document does not cover the step in which the user verifies a reagent (analysis method adoption).

  • Standard
    15 pages
    English language
    e-Library read for
    ×1 day

This document specifies terms and definitions applicable to the EN 18000 series and requirements concerning information to be provided by applicants submitting animal health in vitro diagnostic reagents to control.
This document is applicable to diagnostic reagents, as a priority for infectious (bacterial, viral, fungal or parasitic) or prion diseases and associated animal species for which harmonization of practices in this area is necessary, i.e. those for which the national, regional or international regulatory framework provides for the control of trade in animals and/or animal products and/or the definition of a health status (absence of infection) of areas, establishments or individuals. While all reagents designated by the competent authorities fall under the scope of this document, the authorities or any other animal health stakeholder can choose to derogate in specific and exceptional situations such as emerging, exotic or rare diseases.
This document is not applicable to all existing diagnostic reagents, in particular those for which certain parameters described in this document cannot be validly evaluated in accordance with international requirements due, e.g. to the absence of a specific reference method and/or accessible and duly validated reference materials (RMs).
This document does not cover the step in which the user verifies a reagent (analysis method adoption).

  • Standard
    32 pages
    English language
    e-Library read for
    ×1 day

This document specifies requirements and provides recommendations for the pre-examination process of cell-free DNA (cfDNA) from body fluid specimens other than blood, including but not limited to the collection, handling, storage, transport, processing and documentation of human body fluids, such as urine, pleural effusions, ascites, cerebrospinal fluid (CSF), and saliva, intended for cfDNA examination. Processing includes multiple steps, such as centrifugation for specimen purification and isolation of cfDNA. This document does not cover dedicated measures for cytohistological analysis of nucleated cells derived from body fluid, nor measures for preserving and handling of pathogens, and other bacterial or whole microbiome DNA in body fluids described. Dedicated measures for preserving circulating cell-free DNA (ccfDNA) from blood are covered in ISO 20186-3. This document is applicable to medical laboratories, health institutions including facilities collecting and handling specimens, laboratory customers, in vitro diagnostic examination developers and manufacturers, biobanks, institutions and organizations performing biomedical research, and regulatory authorities. NOTE International, national or regional regulations or requirements can also apply to specific topics covered in this document.

  • Standard
    25 pages
    English language
    sale 15% off
  • Standard
    27 pages
    French language
    sale 15% off

This document provides supplemental information and guidance on how to implement the requirements listed in ISO 35001 [1]. This document does not add requirements to those in ISO 35001 [1].

  • Technical specification
    139 pages
    English language
    sale 15% off

This document specifies the requirements and general principles governing the biological evaluation of medical devices within a risk management process according to ISO 14971.
This document applies to the biological evaluation of medical devices that have direct contact or indirect contact with either:
—     a patient's body during intended use or reasonably foreseeable misuse; or
—     the body of other users who are not patients, if the medical device is intended for personal protection (e.g. medical gloves, surgical masks).
Biological evaluation assesses the biological safety of the medical device by considering the biological risks associated with:
—     constituents of a medical device; and
—     tissue-device interactions (including physical effects).
The biological evaluation specified in this document can address the biological safety of the medical device, considering the life cycle from design and development through initial use of the finished medical device to final decommissioning or withdrawal from use. The biological evaluation considers both the biological safety of the finished device in first use, and the significance of any changes to the medical device which can occur throughout the life cycle. However, the evaluation of risks related to environmental impacts of decommissioning of medical devices are not within the scope of this document. This document does not mandate re-testing of medical devices that are already on the market and have established and acceptable safety profiles (see 6.6.2).
This document can be useful to support clinical or usability evaluations of medical devices. For example, a biological evaluation is a pre-requisite for conducting a clinical trial. This means that principles outlined in this document can be applied to the evaluation of prototype or development stage devices, as well as to finished medical devices.
Other parts of the ISO 10993 series cover specific aspects of biological evaluation, such as chemical characterization, biological testing, sample preparation, animal welfare and toxicological risk assessment.
For some types of medical devices, specific requirements from other standards (outside the ISO 10993 series) can be considered with a justification for the approach taken if there are differences between the requirements of the ISO 10993 series and those provided in other standards. For example, the ISO 18562 series provides specific requirements for biological evaluation of breathing gas pathway medical devices and ISO 7405 provides specific requirements for biological evaluation of dental devices.
The evaluation of risks related to infectious agents [e.g. bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents] is not within the scope of this document.
NOTE 1        The evaluation of bacterial endotoxins is addressed by ISO 11737-3.
NOTE 2        The evaluation of risks related to viruses, TSE agents and other pathogens originating from materials of animal origin is addressed by the ISO 22442 series.

  • Standard
    59 pages
    English language
    e-Library read for
    ×1 day

This document specifies the control and approval of in vitro diagnostic reagents used in animal health for immunological analyses with a qualitative expression of test results.
This document is applicable to diagnostic reagents, as a priority for infectious (bacterial, viral, fungal or parasitic) or prion diseases and associated animal species for which harmonization of practices in this area is needed, i.e. those for which the national, regional or international regulatory framework provides for the control of trade in animals and/or animal products and/or the definition of a health status (absence of infection) of areas, establishments or individuals. While all reagents designated by the competent authorities fall under the scope of this document, the authorities or any other animal health stakeholder can choose to derogate in specific and exceptional situations such as emerging, exotic or rare diseases.
This document is not applicable to all existing diagnostic reagents, in particular those for which certain parameters described in this document cannot be validly evaluated in accordance with international requirements, due, e.g. to the absence of a specific reference method and/or accessible and duly validated reference materials (RMs).
This document does not cover the step in which the user verifies a reagent (analysis method adoption).

  • Standard
    15 pages
    English language
    e-Library read for
    ×1 day

This document specifies terms and definitions applicable to the EN 18000 series and requirements concerning information to be provided by applicants submitting animal health in vitro diagnostic reagents to control.
This document is applicable to diagnostic reagents, as a priority for infectious (bacterial, viral, fungal or parasitic) or prion diseases and associated animal species for which harmonization of practices in this area is necessary, i.e. those for which the national, regional or international regulatory framework provides for the control of trade in animals and/or animal products and/or the definition of a health status (absence of infection) of areas, establishments or individuals. While all reagents designated by the competent authorities fall under the scope of this document, the authorities or any other animal health stakeholder can choose to derogate in specific and exceptional situations such as emerging, exotic or rare diseases.
This document is not applicable to all existing diagnostic reagents, in particular those for which certain parameters described in this document cannot be validly evaluated in accordance with international requirements due, e.g. to the absence of a specific reference method and/or accessible and duly validated reference materials (RMs).
This document does not cover the step in which the user verifies a reagent (analysis method adoption).

  • Standard
    32 pages
    English language
    e-Library read for
    ×1 day

This document specifies the requirements and general principles governing the biological evaluation of medical devices within a risk management process according to ISO 14971.
This document applies to the biological evaluation of medical devices that have direct contact or indirect contact with either:
—     a patient's body during intended use or reasonably foreseeable misuse; or
—     the body of other users who are not patients, if the medical device is intended for personal protection (e.g. medical gloves, surgical masks).
Biological evaluation assesses the biological safety of the medical device by considering the biological risks associated with:
—     constituents of a medical device; and
—     tissue-device interactions (including physical effects).
The biological evaluation specified in this document can address the biological safety of the medical device, considering the life cycle from design and development through initial use of the finished medical device to final decommissioning or withdrawal from use. The biological evaluation considers both the biological safety of the finished device in first use, and the significance of any changes to the medical device which can occur throughout the life cycle. However, the evaluation of risks related to environmental impacts of decommissioning of medical devices are not within the scope of this document. This document does not mandate re-testing of medical devices that are already on the market and have established and acceptable safety profiles (see 6.6.2).
This document can be useful to support clinical or usability evaluations of medical devices. For example, a biological evaluation is a pre-requisite for conducting a clinical trial. This means that principles outlined in this document can be applied to the evaluation of prototype or development stage devices, as well as to finished medical devices.
Other parts of the ISO 10993 series cover specific aspects of biological evaluation, such as chemical characterization, biological testing, sample preparation, animal welfare and toxicological risk assessment.
For some types of medical devices, specific requirements from other standards (outside the ISO 10993 series) can be considered with a justification for the approach taken if there are differences between the requirements of the ISO 10993 series and those provided in other standards. For example, the ISO 18562 series provides specific requirements for biological evaluation of breathing gas pathway medical devices and ISO 7405 provides specific requirements for biological evaluation of dental devices.
The evaluation of risks related to infectious agents [e.g. bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents] is not within the scope of this document.
NOTE 1        The evaluation of bacterial endotoxins is addressed by ISO 11737-3.
NOTE 2        The evaluation of risks related to viruses, TSE agents and other pathogens originating from materials of animal origin is addressed by the ISO 22442 series.

  • Standard
    59 pages
    English language
    e-Library read for
    ×1 day

This document defines the core terms and definitions in the field of laboratory design.

  • Standard
    30 pages
    English language
    sale 15% off
  • Standard
    33 pages
    French language
    sale 15% off
  • Standard
    33 pages
    French language
    sale 15% off

This document specifies the requirements and general principles governing the biological evaluation of medical devices within a risk management process according to ISO 14971. This document applies to the biological evaluation of medical devices that have direct contact or indirect contact with either: — a patient's body during intended use or reasonably foreseeable misuse; or — the body of other users who are not patients, if the medical device is intended for personal protection (e.g. medical gloves, surgical masks). Biological evaluation assesses the biological safety of the medical device by considering the biological risks associated with: — constituents of a medical device; and — tissue-device interactions (including physical effects). The biological evaluation specified in this document can address the biological safety of the medical device, considering the life cycle from design and development through initial use of the finished medical device to final decommissioning or withdrawal from use. The biological evaluation considers both the biological safety of the finished device in first use, and the significance of any changes to the medical device which can occur throughout the life cycle. However, the evaluation of risks related to environmental impacts of decommissioning of medical devices are not within the scope of this document. This document does not mandate re-testing of medical devices that are already on the market and have established and acceptable safety profiles (see REF Section_sec_6.6.2 \r \h 6.6.2 08D0C9EA79F9BACE118C8200AA004BA90B020000000800000012000000530065006300740069006F006E005F007300650063005F0036002E0036002E0032000000 ). This document can be useful to support clinical or usability evaluations of medical devices. For example, a biological evaluation is a pre-requisite for conducting a clinical trial. This means that principles outlined in this document can be applied to the evaluation of prototype or development stage devices, as well as to finished medical devices. Other parts of the ISO 10993 series cover specific aspects of biological evaluation, such as chemical characterization, biological testing, sample preparation, animal welfare and toxicological risk assessment. For some types of medical devices, specific requirements from other standards (outside the ISO 10993 series) can be considered with a justification for the approach taken if there are differences between the requirements of the ISO 10993 series and those provided in other standards. For example, the ISO 18562 series provides specific requirements for biological evaluation of breathing gas pathway medical devices and ISO 7405 provides specific requirements for biological evaluation of dental devices. The evaluation of risks related to infectious agents [e.g. bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents] is not within the scope of this document. NOTE 1 The evaluation of bacterial endotoxins is addressed by ISO 11737-3. NOTE 2 The evaluation of risks related to viruses, TSE agents and other pathogens originating from materials of animal origin is addressed by the ISO 22442 series.

  • Standard
    42 pages
    English language
    sale 15% off
  • Standard
    46 pages
    French language
    sale 15% off
  • Standard
    46 pages
    French language
    sale 15% off

IEC 61326-2-6:2025 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of IN VITRO DIAGNOSTIC MEDICAL ELECTRICAL EQUIPMENT (IVD MEE). This part of IEC 61326 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of IVD MEE in the presence of electromagnetic disturbances and to electromagnetic disturbances emitted by IVD MEE.
BASIC SAFETY with regard to electromagnetic disturbances is applicable to all IVD MEE.
NOTE 1 Performance with respect to electromagnetic disturbances other than ESSENTIAL PERFORMANCE is the subject of IEC 61326-1:2020
NOTE 2 IT equipment can be a part of an IVD MEE, if it is required to maintain BASIC SAFETY or ESSENTIAL PERFORMANCE.
This edition includes the following significant technical changes with respect to the previous edition:
- Update of the document with respect to test levels and documentation.

  • Standard
    36 pages
    English language
    e-Library read for
    ×1 day

This document specifies test methods for the evaluation of biological effects of medical devices used in dentistry. It includes testing of pharmacological agents that are an integral part of the device under test.
This document does not cover testing of materials and devices that do not come into direct or indirect contact with the patient's body.

  • Standard
    69 pages
    English language
    e-Library read for
    ×1 day

This document specifies test methods for the evaluation of biological effects of medical devices used in dentistry. It includes testing of pharmacological agents that are an integral part of the device under test.
This document does not cover testing of materials and devices that do not come into direct or indirect contact with the patient's body.

  • Standard
    69 pages
    English language
    e-Library read for
    ×1 day

This document specifies test methods for the evaluation of biological effects of medical devices used in dentistry. It includes testing of pharmacological agents that are an integral part of the device under test. This document does not cover testing of materials and devices that do not come into direct or indirect contact with the patient's body.

  • Standard
    60 pages
    English language
    sale 15% off
  • Standard
    65 pages
    French language
    sale 15% off
  • Standard
    5 pages
    English language
    sale 15% off
  • Standard
    6 pages
    French language
    sale 15% off
  • Standard
    6 pages
    French language
    sale 15% off

Loading...