Standard Test Method for Spectral Bandwidth and Wavelength Accuracy of Fluorescence Spectrometers

SCOPE
1.1 This test method covers the testing of the spectral bandwidth and wavelength accuracy of fluorescence spectrometers.

General Information

Status
Historical
Publication Date
09-Oct-1998
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E388-72(1998) - Standard Test Method for Spectral Bandwidth and Wavelength Accuracy of Fluorescence Spectrometers
English language
2 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E388–72 (Reapproved 1998)
Standard Test Method for
Spectral Bandwidth and Wavelength Accuracy of
Fluorescence Spectrometers
This standard is issued under the fixed designation E 388; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope 576.96, and 579.07 nm. These are listed in the Mercury Arc
Emission Spectrum in the Ultraviolet and Visible Regions fig.
1.1 This test method covers the testing of the spectral
of Practice E 275.
bandwidth and wavelength accuracy of fluorescence spectrom-
6.1.1 Other lines are suitable for calibration such as some of
eters.
the weaker lines included in the Mercury Arc Emission
2. Referenced Documents Spectrum in the Ultraviolet andVisible Regions fig. of Practice
E 275, but not included in the above list. The comparatively
2.1 ASTM Standards:
low resolution monochromators often used in fluorescence
E 275 Practice for Describing and Measuring Performance
equipment may not resolve pairs of lines such as at 404.66 and
of Ultraviolet, Visible, and Near Infrared Spectrophotom-
407.78, or at 576.96 and 579.07 nm.
eters
6.1.2 In instruments using grating monochromators, addi-
3. Summary of Test Method
tional weaker lines are found due to second order diffraction of
mercury lines. Thus, lines appear at 507.30, 593.46, 668.30
3.1 The difference between the apparent wavelength and the
nm, arising from the 253.65, 296.73, and 334.15-nm lines,
known wavelength for a series of mercury emission lines is
respectively.
used as a test for wavelength accuracy. The apparent width of
6.2 Calibration and Adjustment of Emission Monochroma-
some of these lines is used as a test for spectral bandwidth.
tor:
4. Apparatus
6.2.1 With the mercury arc source properly aligned in the
sample cell compartment, adjust the position of the wavelength
4.1 Fluorescence Spectrometer to be tested.
dial to give maximum signal for each of the mercury lines and
4.2 Mercury Arc, Low-pressure , sufficiently small to be
record the wavelength reading. The difference between the
placed in the sample cell holder of the instrument.
observed value and the corresponding value in 6.1 represents
5. Reagent
the correction that must be subtracted algebraically from the
reading on the dial. The corrections may be recorded or the
5.1 Glycogen Suspension—Dissolve1gof glycogen per
monochromator adjusted to give the proper values. Since there
litre of water, or use a Ludox suspension containing 1 mL of
is some backlash in the wavelength drive, always adjust the
Ludox per litre of water.
dial to the peak reading from the same direction.
6. Procedure
6.2.2 When calibrating scanning-type instruments, turn the
dial to give the peak reading in the same direction that the dial
6.1 Lines suitable for calibration are the nine mercury lines
at 253.65, 296.73, 334.15, 404.66, 407.78, 435.84, 546.07, is turned by the scan motor. Check the dial reading against the
value recorded while scanning and, if necessary, correct as in
6.2.1.
This test method is under the jurisdiction of ASTM Committee E-13 on
6.3 In cases where the monochromator is designed so that a
Molecular Spectroscopy and is the direct responsibility of Subcommittee E13.06 on
lateral displacement of the calibration source from a position
Molecular Luminescence.
directly in front of the entrance slit appears as a wavelength
Current edition approved June 29, 1972. Published September 1972. Originally
published as E 388 – 69 T. Last previous edition E 388 – 69 T.
shift, proceed as follows:
Annual Book of ASTM Standards, Vol 14.01.
The Pen Ray Quartz Lamp, manufactured by Ultraviolet Products, Inc., San
Gabriel, CA, and available from apparatus distributors, has been found satisfactory.
Ludox is an aqueous suspension of colloidal silica, manufactured by E. I.
DuPont De Nemours and Co., Industrial and Biochemicals Dept., Wilmington, DE.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
E388
6.3.1 Instead of placing the mercury lamp in front of the 6.4.6 The match of the excitation monochromator with the
entrance slit of the monochromator, fill a sample cell with a emission monochromator may
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.