Standard Practice for Manufacturing Characterization of Digital Detector Arrays

SIGNIFICANCE AND USE
4.1 This practice provides a means to compare DDAs on a common set of technical measurements, realizing that in practice, adjustments can be made to achieve similar results even with disparate DDAs, given geometric magnification, or other industrial radiologic settings that may compensate for one shortcoming of a device.  
4.2 A user should understand the definitions and corresponding performance parameters used in this practice in order to make an informed decision on how a given DDA can be used in the target application.  
4.3 The factors that will be evaluated for each DDA are: interpolated basic spatial resolution (iSRbdetector), efficiency (normalized Detector  SNR (SNRN) at 1 mGy, for different energies and beam qualities), achievable contrast sensitivity (CSa), specific material thickness range (SMTR) and ISO-MTL , image lag, burn-in, bad pixels distribution and statistics and internal scatter ratio (ISR).  
4.4 Given that each of these parameters are discussed together in many of the following sections, the following list will be helpful in selecting the key sections for a given test as follows. It should be noted that other sections of the document are needed to establish the appropriate technique for the parameter under test. Note that for each parameter (test), the first section listed is typically an apparatus or gauge (if required), the second section listed are the standardized measurements, and the third section listed involves the analysis or computations:  
4.4.1 For iSRbdetector, see 5.1, 7.7, 8.2.  
4.4.2 For Detector Efficiency, see 5.3, 7.8, 8.3.  
4.4.3 For CSa, see 5.2, 7.9, 8.4.  
4.4.4 For SMTR, see 5.2 (or 7.9, if already completed), 7.10, 8.5.  
4.4.5 For ISO-MTL, see 5.2 (or 7.9, if already completed), 7.10, 8.6.  
4.4.6 For Image Lag, see 7.11.1, 8.7.1.  
4.4.7 For Burn-in, see 5.4, 7.11.2, 8.7.2.  
4.4.8 For Bad Pixel Tests, see 6.2, 7.12, 8.8.  
4.4.9 For ISR, see 5.4, 7.13, 8.9.
SCOPE
1.1 This practice describes the evaluation of Digital Detector Arrays (DDAs), and assures that one common standard exists for quantitative comparison of DDAs so that an appropriate DDA is selected to meet NDT requirements.  
1.2 This practice is intended for use by manufacturers or integrators of DDAs to provide quantitative results of DDA characteristics for NDT user or purchaser consumption. Some of these tests require specialized test phantoms to assure consistency among results among suppliers or manufacturers. These tests are not intended for users to complete, nor are they intended for long term stability tracking and lifetime measurements. However, they may be used for this purpose, if so desired.
Note 1: Further information on application of DDAs is contained in Guide E2736 and Practices E2698 and E2737.  
1.3 The results reported based on this practice should be based on a group of at least three individual detectors for a particular model number.  
1.4 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of the other, and values from the two systems shall not be combined.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Dec-2021
Current Stage
Ref Project

Buy Standard

Standard
ASTM E2597/E2597M-22 - Standard Practice for Manufacturing Characterization of Digital Detector Arrays
English language
17 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E2597/E2597M-22 - Standard Practice for Manufacturing Characterization of Digital Detector Arrays
English language
17 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E2597/E2597M − 22
Standard Practice for
1
Manufacturing Characterization of Digital Detector Arrays
ThisstandardisissuedunderthefixeddesignationE2597/E2597M;thenumberimmediatelyfollowingthedesignationindicatestheyear
of original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.
A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2
1.1 This practice describes the evaluation of Digital Detec- 2.1 ASTM Standards:
tor Arrays (DDAs), and assures that one common standard E1316Terminology for Nondestructive Examinations
exists for quantitative comparison of DDAs so that an appro- E1815Test Method for Classification of Film Systems for
priate DDA is selected to meet NDT requirements. Industrial Radiography
E2002Practice for Determining Total Image Unsharpness
1.2 This practice is intended for use by manufacturers or
and Basic Spatial Resolution in Radiography and Radios-
integrators of DDAs to provide quantitative results of DDA
copy
characteristics for NDT user or purchaser consumption. Some
E2445Practice for Performance Evaluation and Long-Term
of these tests require specialized test phantoms to assure
Stability of Computed Radiography Systems
consistency among results among suppliers or manufacturers.
E2446Practice for Manufacturing Characterization of Com-
These tests are not intended for users to complete, nor are they
puted Radiography Systems
intended for long term stability tracking and lifetime measure-
E2698Practice for Radiographic Examination Using Digital
ments. However, they may be used for this purpose, if so
Detector Arrays
desired.
E2736Guide for Digital Detector Array Radiography
NOTE 1—Further information on application of DDAs is contained in
E2737Practice for Digital Detector Array Performance
Guide E2736 and Practices E2698 and E2737.
Evaluation and Long-Term Stability
3
1.3 The results reported based on this practice should be
2.2 ISO Standards:
based on a group of at least three individual detectors for a
ISO7004 Photography—Industrial Radiographic Films—
particular model number.
Determination of ISO Speed, ISO Average Gradient and
ISO Gradients G2 and G4 When Exposed to X- and
1.4 Units—The values stated in either SI units or inch-
Gamma-Radiation
pound units are to be regarded separately as standard. The
ISO 17636-2Non-destructive Testing of Welds — Radio-
values stated in each system are not necessarily exact equiva-
lents; therefore, to ensure conformance with the standard, each graphic Testing — Part 2: X- and Gamma-ray Techniques
With Digital detectors
system shall be used independently of the other, and values
from the two systems shall not be combined. ISO 10893-7Non-destructiveTesting of SteelTubes — Part
7: Digital Radiographic Testing of the Weld Seam of
1.5 This standard does not purport to address all of the
Welded Steel Tubes for the Detection of Imperfections
safety concerns, if any, associated with its use. It is the
2.3 Other Standards:
responsibility of the user of this standard to establish appro-
EN 12681-2Founding — Radiographic Testing — Part 2:
priate safety, health, and environmental practices and deter-
4
Techniques With Digital Detectors
mine the applicability of regulatory limitations prior to use.
IEC 62220-1Medical Electrical Equipment Characteristics
1.6 This international standard was developed in accor-
ofDigitalX-rayImagingDevicesPart1:Determinationof
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
1
This practice is under the jurisdiction of ASTM Committee E07 on Nonde- Standards volume information, refer to the standard’s Document Summary page on
structive Testing and is the direct responsibility of Subcommittee E07.01 on the ASTM website.
3
Radiology (X and Gamma) Method. Available from International Organization for Standardization (ISO), ISO
Current edition approved Jan. 1, 2022. Published February 2022. Originally Central Secretariat, Chemin de Blandonnet 8, CP 401, 1214 Vernier, Geneva,
approved in 2007. Last previous edition approved in 2014 as E2597/E2597M– 14. Switzerland, https://www.iso.org.
4
DOI: 10.1520/E2597_E2597M-22. Available from https://www.en-standard.eu/.
Copyright © ASTM International, 100 Barr
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E2597/E2597M − 14 E2597/E2597M − 22
Standard Practice for
1
Manufacturing Characterization of Digital Detector Arrays
This standard is issued under the fixed designation E2597/E2597M; the number immediately following the designation indicates the year
of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval.
A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice describes the evaluation of Digital Detector Arrays (DDAs), and assures that one common standard exists for
quantitative comparison of DDAs so that an appropriate DDA is selected to meet NDT requirements.
1.2 This practice is intended for use by manufacturers or integrators of DDAs to provide quantitative results of DDA
characteristics for NDT user or purchaser consumption. Some of these tests require specialized test phantoms to assure consistency
among results among suppliers or manufacturers. These tests are not intended for users to complete, nor are they intended for long
term stability tracking and lifetime measurements. However, they may be used for this purpose, if so desired.
NOTE 1—Further information on application of DDAs is contained in Guide E2736 and Practices E2698 and E2737.
1.3 The results reported based on this standardpractice should be based on a group of at least three individual detectors for a
particular model number.
1.4 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in
each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used
independently of the other, and values from the two systems shall not be combined.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2
2.1 ASTM Standards:
E1316 Terminology for Nondestructive Examinations
E1815 Test Method for Classification of Film Systems for Industrial Radiography
E2002 Practice for Determining Total Image Unsharpness and Basic Spatial Resolution in Radiography and Radioscopy
E2445 Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems
1
This practice is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.01 on Radiology (X and
Gamma) Method.
Current edition approved Jan. 1, 2014Jan. 1, 2022. Published February 2014February 2022. Originally approved in 2007. Last previous edition approved in 20072014 as
ε1
E2597E2597/E2597M- 07 – 14. . DOI: 10.1520/E2597_E2597M-14.10.1520/E2597_E2597M-22.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E2597/E2597M − 22
E2446 Practice for Manufacturing Characterization of Computed Radiography Systems
E2698 Practice for Radiographic Examination Using Digital Detector Arrays
E2736 Guide for Digital Detector Array Radiography
E2737 Practice for Digital Detector Array Performance Evaluation and Long-Term Stability
3
2.2 OtherISO Standards:
ISO 7004 Photography—Industrial Radiographic Films—Determination of ISO Speed, ISO Average Gradient and ISO
Gradients G2 and G4 When Exposed to X- and Gamma-Radiation
IEC 62220-1ISO 17636-2 Medical Electrical Equipment Characteristics of Digital X-ray Imaging Devices Part 1: Determination
of the Detective Quantum EfficiencyNon-destructive Testing of Welds — Radiographic Testing — Part 2: X- and Gamma-ray
Techniques With Digital detectors
ISO 10893-7 Non-destructive Testing of Steel Tubes — Part 7: Digital Rad
...

Questions, Comments and Discussion