ASTM ISO/ASTM51650-21
(Practice)Standard Practice for Use of a Cellulose Triacetate Dosimetry System
Standard Practice for Use of a Cellulose Triacetate Dosimetry System
SIGNIFICANCE AND USE
4.1 The CTA dosimetry system provides a means for measuring absorbed dose based on a change in optical absorbance in the CTA dosimeter following exposure to ionizing radiation (2-10).
4.2 CTA dosimetry systems are commonly used in industrial radiation processing, for example in the modification of polymers and sterilization of health care products.
4.3 CTA dosimeter film can be particularly useful in absorbed dose mapping because it is available in a reel of 100 m whereby the user can cut any length of strip for use. When the CTA film is measured using a strip measurement device with a narrow distance interval (for example, 2 mm), it can provide high resolution results in a linear direction.
4.4 CTA is used to measure relative dose such as depth dose profiles in electron beam and reference phantom tests to assess irradiator changes in gamma.
4.5 When CTA is used as a routine monitoring dosimeter the user must take into consideration the effects of the multiple influence quantities that can affect the result and use appropriate techniques, as discussed herein, for characterizing and mitigating such influences and understanding their contribution to measurement uncertainty. Without such effort the dosimetry system may not meet the user’s requirements for dosimetric release of some types of products (for example, health care products).
SCOPE
1.1 This is a practice for using a cellulose triacetate (CTA) dosimetry system to measure absorbed dose in materials irradiated by photons or electrons in terms of absorbed dose to water. CTA is used as a routine dosimetry system or used for relative dose measurements (that is, non-traceable dose measurements).
1.2 The CTA dosimeter is classified as a type II dosimeter on the basis of the complex effect of influence quantities on its response (see ISO/ASTM Practice 52628).
1.3 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of ISO/ASTM 52628 “Practice for Dosimetry in Radiation Processing” for a CTA dosimetry system. It is intended to be read in conjunction with ISO/ASTM 52628.
1.4 This practice covers the use of CTA dosimetry systems under the following conditions:
1.4.1 The absorbed dose range is 10 kGy to 300 kGy.
Note 1: The dosimeter film irradiated to doses exceeding 200 kGy becomes brittle to some degree and must be handled with care. This may limit the practical dose range depending on the type of testing and handling required.
1.4.2 The absorbed-dose rate range is 3 Gy/s to 4 ×1010 Gy·s (1).2
1.4.3 The photon energy range is 0.1 to 50 MeV.
1.4.4 The electron energy range is 0.2 to 50 MeV.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Buy Standard
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: 51650 − 21
Standard Practice for
1
Use of a Cellulose Triacetate Dosimetry System
This standard is issued under the fixed designation 51650; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.6 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.1 This is a practice for using a cellulose triacetate (CTA)
responsibility of the user of this standard to establish appro-
dosimetry system to measure absorbed dose in materials
priate safety, health, and environmental practices and deter-
irradiated by photons or electrons in terms of absorbed dose to
mine the applicability of regulatory limitations prior to use.
water. CTA is used as a routine dosimetry system or used for
1.7 This international standard was developed in accor-
relative dose measurements (that is, non-traceable dose mea-
dance with internationally recognized principles on standard-
surements).
ization established in the Decision on Principles for the
1.2 The CTA dosimeter is classified as a type II dosimeter
Development of International Standards, Guides and Recom-
on the basis of the complex effect of influence quantities on its
mendations issued by the World Trade Organization Technical
response (see ISO/ASTM Practice 52628).
Barriers to Trade (TBT) Committee.
1.3 This document is one of a set of standards that provides
2. Referenced documents
recommendations for properly implementing dosimetry in
3
radiation processing, and describes a means of achieving 2.1 ASTM Standards:
compliance with the requirements of ISO/ASTM 52628 “Prac-
E275 Practice for Describing and Measuring Performance of
tice for Dosimetry in Radiation Processing” for a CTA dosim- Ultraviolet and Visible Spectrophotometers
etry system. It is intended to be read in conjunction with E3083 Terminology Relating to Radiation Processing: Do-
ISO/ASTM 52628. simetry and Applications
3
2.2 ISO/ASTM Standards:
1.4 This practice covers the use of CTA dosimetry systems
51261 Practice for Calibration of Routine Dosimetry Sys-
under the following conditions:
tems for Radiation Processing
1.4.1 The absorbed dose range is 10 kGy to 300 kGy.
51707 Guide for Estimating Uncertainties in Dosimetry for
NOTE 1—The dosimeter film irradiated to doses exceeding 200 kGy
Radiation Processing
becomes brittle to some degree and must be handled with care. This may
51818 Standard Specification for Synthetic Fiber Reinforced
limit the practical dose range depending on the type of testing and
Concrete Culvert, Storm Drain, and Sewer Pipe
handling required.
52628 Practice for Dosimetry in Radiation Processing
10
1.4.2 The absorbed-dose rate range is 3 Gy/s to 4 ×10
52701 Guide for Performance Characterization of Dosim-
2
Gy·s (1).
eters and Dosimetry Systems for Use in Radiation Pro-
1.4.3 The photon energy range is 0.1 to 50 MeV.
cessing
1.4.4 The electron energy range is 0.2 to 50 MeV.
2.3 International Commission on Radiation Units and Mea-
4
1.5 The values stated in SI units are to be regarded as
surements (ICRU) Reports:
standard. No other units of measurement are included in this
ICRU Report 85a Fundamental Quantities and Units for
standard.
Ionizing Radiation
ICRU Report 80 Dosimetry Systems for Use in Radiation
1
Processing
This practice is under the jurisdiction of ASTM Committee E61 on Radiation
Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry
Systems. Originally developed as a joint ASTM/ISO standard in conjunction with
3
ISO/TC 85/WG 3. For referenced ASTM and ISO/ASTM standards, visit the ASTM website,
Current edition approved Oct. 1, 2021. Published May 2024. Originally approved www.astm.org, or contact ASTM Customer Service at service@astm.org. For
in 1994. Last previous edition approved in 2013 as ISO/ASTM 51650:2013(E). Annual Book of ASTM Standards volume information, refer to the standard’s
DOI: 10.1520/51650-21. Document Summary page on the ASTM website.
2 4
The boldface numbers in parentheses refer to the bibliography at the end of this Available from the International Commission on Radiation Units and
standard. Measurements, 7910 Woodmont Ave., suite 800, Bethesda, MD 20814, USA.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
51650 − 21
2.4 ISO Standard: 4.2 CTA dosimetry systems are common
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: ISO/ASTM 51650 − 2013(E) 51650 − 21
Standard Practice for
1
Use of a Cellulose Triacetate Dosimetry System
This standard is issued under the fixed designation ISO/ASTM 51650; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last
reapproval. A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This is a practice for using a cellulose triacetate (CTA) dosimetry system to measure absorbed dose in materials irradiated by
photons or electrons in terms of absorbed dose to water. The CTA dosimetry system is classifiedis used as a routine dosimetry
system.system or used for relative dose measurements (that is, non-traceable dose measurements).
1.2 The CTA dosimeter is classified as a type II dosimeter on the basis of the complex effect of influence quantities on its response
(see ASTMISO/ASTM Practice E262852628).
1.3 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation
processing, and describes a means of achieving compliance with the requirements of ASTMISO/ASTM E262852628 “Practice for
Dosimetry in Radiation Processing” for a CTA dosimetry system. It is intended to be read in conjunction with ASTMISO/ASTM
E262852628.
1.4 This practice covers the use of CTA dosimetry systems under the following conditions:
1.4.1 The absorbed dose range is 10 kGy to 300 kGy.
NOTE 1—The dosimeter film irradiated to doses exceeding 200 kGy becomes brittle to some degree and must be handled with care. This may limit the
practical dose range depending on the type of testing and handling required.
10 2
1.4.2 The absorbed-dose rate range is 3 Gy/s to 4×104 ×10 Gy/sGy·s (1).
1.4.3 The photon energy range is 0.1 to 50 MeV.
1.4.4 The electron energy range is 0.2 to 50 MeV.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and healthsafety, health, and environmental practices and determine
the applicability of regulatory limitations prior to use.
1
This practice is under the jurisdiction of ASTM Committee E61 on Radiation Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry Systems,
and is also under the jurisdiction of . Originally developed as a joint ASTM/ISO standard in conjunction with ISO/TC 85/WG 3.
Current edition approved April 9, 2013Oct. 1, 2021. Published June 2013May 2024. Originally published as ASTM E 1650–94 with title: Practice for Use of Cellulose
Acetate Dosimetry Systems. ASTM E 1650–94 was adopted by ISO in 1998 with the intermediate designation ISO 15570:1998(E). The present Third Edition of International
Standard ISO/ASTM 51650:2013(E) is a major revision of the Second Edition of ISO/ASTM 51650:2005(E).approved in 1994. Last previous edition approved in 2013 as
ISO/ASTM 51650:2013(E). DOI: 10.1520/51650-21.
2
The boldface numbers in parentheses refer to the bibliography at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
51650 − 21
1.7 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced documents
3
2.1 ASTM Standards:
E170 Terminology Relating to Radiation Measurements and Dosimetry
E275 Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers
E2628E3083 Practice for Dosimetry in Radiation ProcessingTerminology Relating to Radiation Processing: Dosimetry and
Applications
E2701 Guide for Performance Characterization of Dosimeters and Dosimetry Systems for Use in Radiation Processing
3
2.2 ISO/ASTM Standards:
51261 Practice for Calibration of Routine Dosimetry Systems for Radiation Processing
51707 Guide for Estimating Uncertainties in Dosimetry for Radiation Processing
51818 Standard Specification for Synthetic Fiber Reinforced Concrete Culvert, Storm Drain, and
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.