Medical electrical equipment - Part 2: Particular requirements for the safety of therapeutic X-ray equipment operating in the range 10 kV to 1 MV

EN following parallel vote

Medizinische elektrische Geräte - Teil 2: Besondere Festlegungen für die Sicherheit von Therapie-Röntgeneinrichtungen im Betriebsbereich von 10 kV bis 1 MV

Appareils électromédicaux - Partie 2: Règles particulières de sécurité pour les équipements à rayonnement X de thérapie fonctionnant dans la gamme de 10 kV à 1 MV

Medicinska električna oprema - 2. del: Posebne varnostne zahteve za terapevtsko rentgensko opremo, ki deluje v območju od 10 kV do 1 MV - Dopolnilo A1 (IEC 60601-2-8:1987/A1:1997)

General Information

Status
Withdrawn
Publication Date
08-Sep-1997
Withdrawal Date
31-May-1998
Current Stage
9960 - Withdrawal effective - Withdrawal
Start Date
14-Apr-2018
Completion Date
14-Apr-2018

Relations

Buy Standard

Amendment
EN 60601-2-8:1998/A1:1998
English language
26 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-september-1998
0HGLFLQVNDHOHNWULþQDRSUHPDGHO3RVHEQHYDUQRVWQH]DKWHYH]DWHUDSHYWVNR
UHQWJHQVNRRSUHPRNLGHOXMHYREPRþMXRGN9GR09'RSROQLOR$ ,(&
$
Medical electrical equipment - Part 2: Particular requirements for the safety of
therapeutic X-ray equipment operating in the range 10 kV to 1 MV - Amendment A1 (IEC
60601-2-8:1987/A1:1997)
Medizinische elektrische Geräte - Teil 2: Besondere Festlegungen für die Sicherheit von
Therapie-Röntgeneinrichtungen im Betriebsbereich von 10 kV bis 1 MV (IEC 60601-2-
8:1987/A1:1997)
Appareils électromédicaux - Partie 2: Règles particulières de sécurité pour les
équipements à rayonnement X de thérapie fonctionnant dans la gamme de 10 kV à 1 MV
(CEI 60601-2-8:1987/A1:1997)
Ta slovenski standard je istoveten z: EN 60601-2-8:1997/A1:1997
ICS:
11.040.50 Radiografska oprema Radiographic equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

NORME
CEI
INTERNATIONALE
IEC
60601-2-8
INTERNATIONAL
STANDARD
AMENDEMENT 1
AMENDMENT 1
1997-08
Amendement 1
Appareils électromédicaux –
Partie 2:
Règles particulières de sécurité pour
les équipements à rayonnement X de thérapie
fonctionnant dans la gamme de 10 kV à 1 MV
Amendment 1
Medical electrical equipment –
Part 2:
Particular requirements for the safety of
therapeutic X-ray equipment operating in
the range 10 kV to 1 MV
 IEC 1997 Droits de reproduction réservés  Copyright - all rights reserved
International Electrotechnical Commission 3, rue de Varembé Geneva, Switzerland
Telefax: +41 22 919 0300 e-mail: inmail@iec.ch IEC web site http: //www.iec.ch
CODE PRIX
Commission Electrotechnique Internationale
PRICE CODE S
International Electrotechnical Commission
Pour prix, voir catalogue en vigueur
For price, see current catalogue

60601-2-8 Amend. 1 © IEC:1997 – 3 –
FOREWORD
This amendment has been prepared by subcommittee 62C: Equipment for radiotherapy,
nuclear medicine and radiation dosimetry, of IEC technical committee 62: Electrical equipment
in medical practice.
The text of this amendment is based on the following documents:
FDIS Report on voting
62C/186/FDIS 62C/193/RVD
Full information on the voting for the approval of this amendment can be found in the report on
voting indicated in the above table.
––––––––
Page 3
CONTENTS
Add, after PREFACE, the following title:
INTRODUCTION
Add the titles of the following new subclauses:
1.3.103 IEC 61217
1.5 Collateral Standards
6.3 Marking of controls and instruments
Replace the title of SECTION TWO by the following:
SECTION TWO – ENVIRONMENTAL CONDITIONS
Replace the title of clause 18 by the following:
18 Protective earthing, functional earthing and potential equalization
Page 5
Add the titles of the following new subclauses:
29.103 Indication of X-RADIATION output
29.104 Agreement between indicated values and effective values
29.105 General test conditions
29.106 Settings for measurements
29.107 Number of measurements
29.108 Measurements and evaluation

60601-2-8 Amend. 1 © IEC:1997 – 5 –
Replace the title of SECTION SIX by the following:
SECTION SIX – PROTECTION AGAINST HAZARDS OF IGNITION
OF FLAMMABLE ANAESTHETIC MIXTURES
Replace the title of SECTION SEVEN by the following:
SECTION SEVEN – PROTECTION AGAINST EXCESSIVE TEMPERATURES
AND OTHER SAFETY HAZARDS
Delete the title of clause 48 and replace it by “Biocompatibility”.

Replace the title of SECTION EIGHT by the following:
SECTION EIGHT – ACCURACY OF OPERATING DATA AND
PROTECTION AGAINST HAZARDOUS OUTPUT
Delete the titles of subclauses 50.1 to 50.104.
Page 7
Delete the title of clause 59.4.
Replace “APPENDIX AA” by “ANNEX AA”.
Add, after ANNEX AA, the following new ANNEX BB:
ANNEX BB – List of standards mentioned in this Particular Standard
Page 9
FOREWORD
Replace the title by the following:
MEDICAL ELECTRICAL EQUIPMENT –
Part 2: Particular requirements for the safety of
therapeutic X-ray equipment operating
in the range 10 kV to 1 MV
60601-2-8 Amend. 1 © IEC:1997 – 7 –
Page 9
Replace the foreword and the preface by the following new foreword:
MEDICAL ELECTRICAL EQUIPMENT –
Part 2: Particular requirements for the safety of
therapeutic X-ray equipment operating
in the range 10 kV to 1 MV
FOREWORD
1) The IEC (International Electrotechnical Commission) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of the IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, the IEC publishes International Standards. Their preparation is
entrusted to technical committees; any IEC National Committee interested in the subject dealt with may
participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. The IEC collaborates closely with the International Organization
for Standardization (ISO) in accordance with conditions determined by agreement between the two
organizations.
2) The formal decisions or agreements of the IEC on technical matters express, as nearly as possible, an
international consensus of opinion on the relevant subjects since each technical committee has representation
from all interested National Committees.
3) The documents produced have the form of recommendations for international use and are published in the form
of standards, technical reports or guides and they are accepted by the National Committees in that sense.
4) In order to promote international unification, IEC National Committees undertake to apply IEC International
Standards transparently to the maximum extent possible in their national and regional standards. Any
divergence between the IEC Standard and the corresponding national or regional standard shall be clearly
indicated in the latter.
5) The IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with one of its standards.
6) Attention is drawn to the possibility that some of the elements of this International Standard may be the subject
of patent rights. The IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60601-2-8 has been prepared by subcommittee 62B: Diagnostic
imaging equipment, of IEC technical committee 62: Electrical equipment in medical practice.
The text of this standard is based on the following documents:
Six Months’ Rule Report on voting
62B(CO)49 62B(CO)64
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
Subclauses, figures or tables that are additional to those of the General Standard are
numbered starting from 101; additional annexes are lettered AA, etc., and additional items aa),
bb), etc.
Annex AA is for information only.
Annex BB forms an integral part of this standard.

60601-2-8 Amend. 1 © IEC:1997 – 9 –
In this Particular Standard, the following print types are used:
– requirements, compliance with which can be tested, and definitions, in roman type;
– explanations, advice, general statements and exceptions: in small roman type;
– TERMS USED THROUGHOUT THIS PARTICULAR STANDARD WHICH HAVE BEEN LISTED IN ANNEX AA
AND DEFINED IN CLAUSE 2, OR IN THE GENERAL STANDARD 60601-1 AND ITS COLLATERAL
STANDARDS OR IN IEC 60788: SMALL CAPITALS.
Replace, throughout the Standard, the term “X-RAY GENERATOR” by “X-RAY EQUIPMENT”.
Add, after the foreword, the following new introduction:
INTRODUCTION
The use of X-RAY EQUIPMENT for RADIOTHERAPY purposes may expose the PATIENT to danger if
the EQUIPMENT fails to deliver the required dose to the PATIENT, or if the EQUIPMENT design does
not satisfy standards of electrical and mechanical safety. The EQUIPMENT may also cause
danger to persons in the vicinity if the EQUIPMENT itself fails to contain the RADIATION
adequately and/or if there are inadequacies in the design of the TREATMENT ROOM.
This Particular Standard establishes requirements to be complied with by the MANUFACTURERS
in the design and construction of therapeutic X-RAY EQUIPMENT. Clause 29 contains limits
beyond which INTERLOCKS prevent, INTERRUPT or TERMINATE IRRADIATION in order to avoid an
unsafe condition.
Clause 29 does not attempt to define optimum performance requirements. Its purpose is to
identify those features of design that are regarded, at the present time, as essential for the
safe operation of such EQUIPMENT. It places limits on the degradation of EQUIPMENT
performance beyond which it can be presumed that a fault condition exists, e.g. a component
failure, and where an INTERLOCK then operates to prevent continued operation of the
EQUIPMENT.
It should be understood that, before installation, a MANUFACTURER can provide a Compliance
Certificate relating only to TYPE TESTS: data available from SITE TESTS should be incorporated in
the ACCOMPANYING DOCUMENTS, in the form of a SITE TEST report, by those who test the
EQUIPMENT at installation.
Page 11
Replace the title by the following new title:
MEDICAL ELECTRICAL EQUIPMENT –
Part 2: Particular requirements for the safety of
therapeutic X-ray equipment operating in
the range 10 kV to 1 MV
60601-2-8 Amend. 1 © IEC:1997 – 11 –
1.1 Scope
Replace the text of this subclause by the following:
Addition:
This Particular Standard applies to therapeutic X-RAY EQUIPMENT with NOMINAL X-RAY TUBE
VOLTAGES in the range 10 kV to 1 MV when connected to alternating current SUPPLY MAINS.
1.2 Object
Replacement:
Replace the first paragraph of this subclause by the following:
This Particular Standard establishes requirements for the safety of therapeutic X-RAY
EQUIPMENT including the requirement for accuracy and reproducibility of performance to the
extent that these are related to RADIATION QUALITY and the quantity of IONIZING RADIATION
produced and thus must be considered as aspects of safety.
1.3.101 Relation to the General Standard
Replace the first two paragraphs by the following:
This Particular Standard shall be read in conjunction with IEC 60601-1 (1988) and its
amendments 1 (1991) and 2 (1995).
As in the General Standard, the requirements of this Particular Standard are followed by
compliance tests.
The numbering of sections, clauses and subclauses of this Particular Standard corresponds
with that of the General Standard. The changes to the text of the General Standard are
SPECIFIED by the use of the following words:
"Replacement" means that the clause or subclause of the General Standard is replaced
completely by the text of this Particular Standard.
"Addition" means that the text of this Particular Standard is additional to the requirements of
the General Standard.
"Amendment" means that the clause or subclause of the General Standard is amended as
indicated by the text of this Particular Standard.
Replace the last paragraph on page 11 and the first paragraph on page 13, by the following
paragraph:
Where there is no corresponding section, clause or subclause in this Particular Standard, the
section, clause or subclause of the General Standard, although possibly not relevant, applies
without modification. Where it is intended that any part of the General Standard, although
possibly relevant, is not to be applied, a statement to that effect is given in this Particular
Standard. Unless otherwise stated, all clauses of the General Standard shall apply.

60601-2-8 Amend. 1 © IEC:1997 – 13 –
Page 13
1.3.102 Superseded IEC Standard
Delete this subclause and add the following new subclause 1.3.103:
1.3.103 IEC 61217
This standard defines the designation of EQUIPMENT movements, the marking of scales, their
zero positions, and the direction of movements with increasing values.
1.4 Environmental conditions
Delete this subclause and add the following new subclauses 1.5 to 1.5.104:
1.5 Collateral Standards
1.5.101 IEC 60601-1-1
This Collateral Standard does not apply.
1.5.102 IEC 60601-1-2
ELECTROMAGNETIC COMPATIBILITY (EMC) – Requirements and tests: see clause 36.
1.5.103 IEC 60601-1-3
This Collateral Standard does not apply.
1.5.104 IEC 60601-1-4
NOTE – The Collateral Standard IEC 60601-1-4 and/or a future amendment to the General Standard, with a
consequent amendment to this Particular Standard, will provide a comprehensive treatment of all safety aspects
of PROGRAMMABLE ELECTRONIC SYSTEMS used in therapeutic X-RAY EQUIPMENT.
Page 15
4 General requirements for tests
4.1 Type tests and routine tests
Replace this subclause by the following new subclauses 4.1 and 4.1.101:
4.1 Tests
4.1.101 Test grades
TYPE TEST SITE TEST SPECIFIED
Three grades of and two of procedures are in clause 29 of this
Particular Standard; their requirements are as follows:
• TYPE TEST grade A: An analysis of EQUIPMENT design, as related to the SPECIFIED
RADIATION safety provisions, which shall result in a statement included in the technical
description, regarding the working principles or constructional means by which the
requirement is fulfilled.
60601-2-8 Amend. 1 © IEC:1997 – 15 –
YPE TEST SITE TEST
• T / grade B: Visual inspection or functional test or measurement of the
EQUIPMENT. The test shall be in accordance with the procedure SPECIFIED in this Particular
Standard and shall be based on operating states, including fault condition states, which are
achievable only without interference with the circuitry or construction of the EQUIPMENT.
• TYPE TEST/SITE TEST grade C: Functional test or measurement of the EQUIPMENT. The test
SPECIFIED SITE TEST
shall be in accordance with the prin
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.