M/023 - Medical devices
Standardization mandate to CEN/CENELEC concerning the development of European standards relating to medical devices
Mandate M/023 authorizes CEN and CENELEC to develop European standards related to medical devices. The mandate aims to establish harmonized standards that support the safety, performance, and regulatory compliance of medical devices within the European market. Through this standardization work, the mandate facilitates conformity assessment, enhances product reliability, and promotes innovation in the medical device sector. The decision regarding this mandate was taken under resolution CEN/BTS3 58/1993.
Purpose
This mandate concerns the development of European standards related to medical devices. Its primary purpose is to request the European standardisation organisations CEN and CENELEC to create harmonised standards that support the regulation, safety, performance, and conformity assessment of medical devices within the European Union.
Standardisation request
The mandate tasks CEN and CENELEC with the elaboration of European standards covering various aspects of medical devices. These standards are intended to provide technical specifications and criteria that manufacturers and other stakeholders can follow to demonstrate compliance with relevant EU legislation governing medical devices.
Expected deliverables
The deliverables expected from this mandate include a comprehensive set of harmonised European standards for medical devices. These standards should address essential requirements for device safety, performance, quality assurance, risk management, and possibly testing methods. They aim to facilitate compliance assessment processes and ensure a high level of protection for patients and users within the EU.
Context
The mandate (M/023) was issued in 1993 and is documented under resolution reference CEN/BTS3 58/1993. It reflects the EU’s approach to harmonising medical device regulations by relying on European standardisation bodies to develop technical standards. These standards play a key role in implementing the EU medical device directives and later regulations by providing commonly accepted norms that support market surveillance and regulatory enforcement.
This mandate covers the development of European standards related to medical devices. It focuses on standardisation work for products and technologies within the medical devices sector.
General Information
IEC 60601-2-57:2023 applies to basic safety and essential performance of equipment incorporating one or more sources of optical radiation in the wavelength range 200 nm to 3 000 nm, with the exception of laser radiation, and intended to create photobiological effects in humans for therapeutic, diagnostic, monitoring, and cosmetic or aesthetic applications; hereafter referred to as light source equipment (ls equipment).
- Standard31 pagesEnglish languagee-Library read for1 day
IEC 60601-2-57:2023 applies to basic safety and essential performance of equipment incorporating one or more sources of optical radiation in the wavelength range 200 nm to 3 000 nm, with the exception of laser radiation, and intended to create photobiological effects in humans for therapeutic, diagnostic, monitoring, and cosmetic or aesthetic applications; hereafter referred to as light source equipment (ls equipment).
- Standard31 pagesEnglish languagee-Library read for1 day
This document specifies a test method for the measurement of residual chemicals used in product manufacture, particularly potentially Type IV allergenic substances employed and remaining in medical gloves. It also provides information on the extraction medium, the method of extraction and quantitative assay of residual chemicals.
This document does not provide information on the allergenic potential, biological evaluation, or safety to the user of any product.
- Standard23 pagesEnglish languagee-Library read for1 day
This document specifies a test method for the measurement of residual chemicals used in product manufacture, particularly potentially Type IV allergenic substances employed and remaining in medical gloves. It also provides information on the extraction medium, the method of extraction and quantitative assay of residual chemicals.
This document does not provide information on the allergenic potential, biological evaluation, or safety to the user of any product.
- Standard23 pagesEnglish languagee-Library read for1 day
2022-05-21-JO- to ensure continuity, HAS assessment should be prepared by Tony Wilkes
- Amendment14 pagesEnglish languagee-Library read for1 day
2022-05-21-JO- to ensure continuity, HAS assessment should be prepared by Tony Wilkes
- Amendment14 pagesEnglish languagee-Library read for1 day
This European Standard specifies requirements for the design, testing, performance and equipping of road ambulances used for the transport, monitoring, treatment and care of patients. It contains requirements for the patient’s compartment in terms of the working environment, ergonomic design and the safety of the crew and patients.This European Standard does not cover the training of the staff which is the responsibility of the authority/authorities in the country where the ambulance is to be registered.
This European Standard is applicable to road ambulances capable of transporting at least one person on a stretcher and excludes the transportation of hospital beds.
This standard also specifies requirements for ambulances intended to carry transport incubator systems.
The European Standard covers the specific requirements of each type of road ambulance which are
designated according to the patient condition e.g. patient transport road ambulance types A1, A2, B and C.
This European Standard gives general requirements for medical devices carried in road ambulances and used therein and outside hospitals and clinics in situations where the ambient conditions can differ from normal indoor conditions.
- Standard58 pagesEnglish languagee-Library read for1 day
- Standard61 pagesGerman languagee-Library read for1 day
IEC 60601-2-66:2019 is available as IEC 60601-2-66:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-66:2019 applies to the BASIC SAFETY of HEARING AIDS and HEARING AID SYSTEMS, hereafter also referred to as ME EQUIPMENT or ME SYSTEM. IEC 60601-2-66:2019 cancels and replaces the second edition published in 2015. It constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) revision of the definition about ESSENTIAL PERFORMANCE; b) revision of the application of IEC 60601-1-2:2014 for electromagnetic disturbances; c) correction of the used voltage for HEARING AIDS from 1,6 V to 4,5 V; d) correction of the drop test level from 1,5 m to 1,0 m; e) correction of the wording of IEC 60601-2-66:2015.
- Standard64 pagesEnglish languagee-Library read for1 day
IEC 60118-13:2019 covers the relevant EMC phenomena for hearing aids. Hearing aid immunity to high frequency fields originating from digital wireless devices such as mobile phones was identified as one of the most relevant EMC phenomena impacting hearing aids. IEC 60118-13:2019 cancels and replaces the fourth edition published in 2016 and constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) it introduces a new measurement method and set of EMC requirements for hearing aids immunity to mobile digital wireless devices; b) generic EMC requirements for hearing aids are no longer included – should be covered by other standards as appropriate.
- Standard25 pagesEnglish languagee-Library read for1 day
This document specifies requirements for safety and performance of active implantable circulatory support devices, including type tests, animal studies and clinical evaluation requirements.
NOTE The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify main requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses and are not intended to be used for the routine testing of manufactured products.
Included in the scope of this document are:
— ventricular assist devices (VAD), left or right heart support;
— total artificial hearts (TAH);
— biventricular assist devices (biVAD);
— percutaneous assist devices;
— paediatric assist devices.
- Standard79 pagesEnglish languagee-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices that are intended to treat hearing impairment via electrical stimulation of the auditory pathways. Devices which treat hearing impairment via means other than electrical stimulation are not covered by this document.
The tests that are specified in this document are type tests and are to be carried out on samples of a device to show compliance.
This document is also applicable to non-implantable parts and accessories of the devices (see NOTE).
The electrical characteristics of the implantable part are determined by either the appropriate method detailed in this document or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document applies.
NOTE A device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, this document specifies those requirements of non-implantable parts and accessories which could affect the safety or performance of the implantable part.
- Standard77 pagesEnglish languagee-Library read for1 day
This document specifies particular requirements for active implantable medical devices intended to deliver a medicinal substance to site-specific locations within the human body, to provide basic assurance of safety for both patients and users. It amends and supplements ISO 14708-1:2014. The requirements of this document take priority over those of ISO 14708-1.
This document is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body.
This document is also applicable to some non-implantable parts and accessories of the devices defined in Clause 3.
The tests that are specified in this document are type tests intended to be carried out on a sample of a device to show compliance and are not intended to be used for the routine testing of manufactured products.
NOTE This document is not intended to apply to non-implantable infusion systems.
- Standard68 pagesEnglish languagee-Library read for1 day
This document specifies requirements that are applicable to implantable cardioverter defibrillators and CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this document or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac resynchronization is covered by ISO 14708-2.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
- Standard75 pagesEnglish languagee-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization.
The tests that are specified in this document are type tests, and are to be carried out on samples of a device to show compliance.
This document was designed for bradyarrhythmia pulse generators used with endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of leadless technologies for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The electrical characteristics of the implantable pulse generator or lead are determined either by the appropriate method detailed in this particular standard or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in this particular standard applies.
Any features of an active implantable medical device intended to treat tachyarrhythmias are covered by ISO 14708-6.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
- Standard80 pagesEnglish languagee-Library read for1 day
This document specifies the performance requirements and test methods for hollow devices and porous devices as well as the chemical indicators and biological indicators that are utilized within these devices for testing a specific steam penetration performance of type B cycles and some type S cycles of small steam sterilizers according to EN 13060.
NOTE The hollow and porous devices described in this document are not intended for use as surrogate devices for hollow and porous medical devices used in health care facilities.
a) Chemical indicators used with a porous device specified in this document are designed to demonstrate the adequacy of steam penetration into a porous device in small steam sterilizers (see EN 13060).
This document specifies the requirements for:
— a reference porous device (RPD) as a reference device by which alternative porous indicator systems (APISs) can be shown to be equivalent in performance according to this document, i.e. a textile test pack in which steam penetration is judged by thermometric means;
— an alternative porous chemical indicator system equivalent in performance to the RPD, i.e. an APIS, usually commercially manufactured, of any design.
b) Chemical indicators used with a hollow load device specified in this document are designed to demonstrate the adequacy of steam penetration into a narrow lumen (previously known as hollow load A) in small steam sterilizers (see EN 13060).
This document specifies the requirements for:
— a reference hollow device (RHD) used as a reference device in this document, i.e. a lumened device with attached capsule in which steam penetration is judged by inactivation or survival of a specified biological indicator;
— an alternative hollow device:
— employing the same specific test load as defined for the RHD and a chemical indicator designed specifically for use in the reference hollow test load, i.e. a lumened device with an attached capsule in which steam penetration is judged by visual examination of a chemical indicator;
— equivalent in performance to the RHD, i.e. an alternative hollow device, usually commercially manufactured, of any design.
- Standard48 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.
- Amendment11 pagesEnglish languagee-Library read for1 day
This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
- Amendment11 pagesEnglish languagee-Library read for1 day
This document specifies the performance requirements and test methods for hollow devices and porous devices as well as the chemical indicators and biological indicators that are utilized within these devices for testing a specific steam penetration performance of type B cycles and some type S cycles of small steam sterilizers according to EN 13060.
NOTE The hollow and porous devices described in this document are not intended for use as surrogate devices for hollow and porous medical devices used in health care facilities.
a) Chemical indicators used with a porous device specified in this document are designed to demonstrate the adequacy of steam penetration into a porous device in small steam sterilizers (see EN 13060).
This document specifies the requirements for:
— a reference porous device (RPD) as a reference device by which alternative porous indicator systems (APISs) can be shown to be equivalent in performance according to this document, i.e. a textile test pack in which steam penetration is judged by thermometric means;
— an alternative porous chemical indicator system equivalent in performance to the RPD, i.e. an APIS, usually commercially manufactured, of any design.
b) Chemical indicators used with a hollow load device specified in this document are designed to demonstrate the adequacy of steam penetration into a narrow lumen (previously known as hollow load A) in small steam sterilizers (see EN 13060).
This document specifies the requirements for:
— a reference hollow device (RHD) used as a reference device in this document, i.e. a lumened device with attached capsule in which steam penetration is judged by inactivation or survival of a specified biological indicator;
— an alternative hollow device:
— employing the same specific test load as defined for the RHD and a chemical indicator designed specifically for use in the reference hollow test load, i.e. a lumened device with an attached capsule in which steam penetration is judged by visual examination of a chemical indicator;
— equivalent in performance to the RHD, i.e. an alternative hollow device, usually commercially manufactured, of any design.
- Standard48 pagesEnglish languagee-Library read for1 day
This document specifies requirements for low-flow nasal cannulae, used in both home care and hospital environments for the administration of oxygen therapy.
This document does not include requirements to prevent the proliferation of fire within the tubing but does specify a user-detachable connection that can be used to fit a fire-activated oxygen shut-off device.
- Standard17 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for manual wheelchairs intended to carry one person of mass not less than 25 kg and not greater than 250 kg, including
- stand-up manual wheelchairs, and
- manual wheelchairs whose intended use includes showering and/or toileting.
This document does not apply to custom-made manual wheelchairs, manual wheelchairs intended for use in sports, or manual wheelchairs intended only for showering and/or toileting.
This document also specifies requirements and test methods for manual wheelchairs with electrically powered ancillary equipment.
- Standard53 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for electrically powered wheelchairs, with a maximum speed not exceeding 20 km/h, intended to carry one person of mass not less than 25 kg and not greater than 300 kg, including
- electrically powered scooters with three or more wheels,
- manual wheelchairs with an add-on electrically powered drive system,
- handrim-activated power-assisted wheelchairs,
- electrically powered stand-up wheelchairs,
- wheelchairs with a pivot drive wheel unit, and
- push-assist wheelchairs.
This document does not apply to balancing wheelchairs, custom-made electrically powered wheelchairs or electrically powered wheelchairs intended for use in sports.
This document also specifies requirements and test methods for manual wheelchairs with electrically powered ancillary equipment.
- Standard76 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for manual wheelchairs intended to carry one person of mass not less than 25 kg and not greater than 250 kg, including
- stand-up manual wheelchairs, and
- manual wheelchairs whose intended use includes showering and/or toileting.
This document does not apply to custom-made manual wheelchairs, manual wheelchairs intended for use in sports, or manual wheelchairs intended only for showering and/or toileting.
This document also specifies requirements and test methods for manual wheelchairs with electrically powered ancillary equipment.
- Standard53 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for electrically powered wheelchairs, with a maximum speed not exceeding 20 km/h, intended to carry one person of mass not less than 25 kg and not greater than 300 kg, including
- electrically powered scooters with three or more wheels,
- manual wheelchairs with an add-on electrically powered drive system,
- handrim-activated power-assisted wheelchairs,
- electrically powered stand-up wheelchairs,
- wheelchairs with a pivot drive wheel unit, and
- push-assist wheelchairs.
This document does not apply to balancing wheelchairs, custom-made electrically powered wheelchairs or electrically powered wheelchairs intended for use in sports.
This document also specifies requirements and test methods for manual wheelchairs with electrically powered ancillary equipment.
- Standard76 pagesEnglish languagee-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization.
The tests that are specified in this document are type tests, and are to be carried out on samples of a device to show compliance.
This document was designed for bradyarrhythmia pulse generators used with endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of leadless technologies for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The electrical characteristics of the implantable pulse generator or lead are determined either by the appropriate method detailed in this particular standard or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in this particular standard applies.
Any features of an active implantable medical device intended to treat tachyarrhythmias are covered by ISO 14708-6.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
- Standard80 pagesEnglish languagee-Library read for1 day
This document specifies particular requirements for active implantable medical devices intended to deliver a medicinal substance to site-specific locations within the human body, to provide basic assurance of safety for both patients and users. It amends and supplements ISO 14708-1:2014. The requirements of this document take priority over those of ISO 14708-1.
This document is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body.
This document is also applicable to some non-implantable parts and accessories of the devices defined in Clause 3.
The tests that are specified in this document are type tests intended to be carried out on a sample of a device to show compliance and are not intended to be used for the routine testing of manufactured products.
NOTE This document is not intended to apply to non-implantable infusion systems.
- Standard68 pagesEnglish languagee-Library read for1 day
This document specifies requirements for safety and performance of active implantable circulatory support devices, including type tests, animal studies and clinical evaluation requirements.
NOTE The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify main requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses and are not intended to be used for the routine testing of manufactured products.
Included in the scope of this document are:
— ventricular assist devices (VAD), left or right heart support;
— total artificial hearts (TAH);
— biventricular assist devices (biVAD);
— percutaneous assist devices;
— paediatric assist devices.
- Standard79 pagesEnglish languagee-Library read for1 day
This document specifies requirements that are applicable to implantable cardioverter defibrillators and CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this document or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac resynchronization is covered by ISO 14708-2.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
- Standard75 pagesEnglish languagee-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices that are intended to treat hearing impairment via electrical stimulation of the auditory pathways. Devices which treat hearing impairment via means other than electrical stimulation are not covered by this document.
The tests that are specified in this document are type tests and are to be carried out on samples of a device to show compliance.
This document is also applicable to non-implantable parts and accessories of the devices (see NOTE).
The electrical characteristics of the implantable part are determined by either the appropriate method detailed in this document or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document applies.
NOTE A device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, this document specifies those requirements of non-implantable parts and accessories which could affect the safety or performance of the implantable part.
- Standard77 pagesEnglish languagee-Library read for1 day
This document specifies requirements for low-flow nasal cannulae, used in both home care and hospital environments for the administration of oxygen therapy.
This document does not include requirements to prevent the proliferation of fire within the tubing but does specify a user-detachable connection that can be used to fit a fire-activated oxygen shut-off device.
- Standard17 pagesEnglish languagee-Library read for1 day
This document establishes a vocabulary of terms used for anaesthetic and respiratory equipment and supplies, related devices and supply systems.
NOTE 1 To avoid multiple definitions of the same term in different categories, this document attempts to ensure consistency by the inclusion of a ‘general’ category, and by use of domain specifiers and unique pre-coordinated domain-specific term names.
NOTE 2 In addition to terms and definitions used in two of the three official ISO languages (English and French), this document gives the equivalent terms in the German language; these are published under the responsibility of the member body for Germany. However, only the terms and definitions given in the official languages can be considered as ISO terms and definitions.
- Standard72 pagesEnglish languagee-Library read for1 day
This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
- Amendment11 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.
- Amendment11 pagesEnglish languagee-Library read for1 day
IEC 60601-2-75:2017 applies to the basic safety and essential performance of photodynamic therapy and photodynamic diagnosis equipment. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. This document applies to photodynamic therapy and photodynamic diagnosis equipment used for compensation or alleviation of disease, injury or disability. In the case of combined equipment (e.g. equipment additionally provided with a function or an applied part for the target area), such equipment also comply with any particular standard specifying safety requirements for the additional function. This particular standard does not apply to: - light therapy equipment intended for use in photothermal ablation, coagulation, and hyperthermia; - low-level laser therapy equipment not intended for use with a photosensitizer; - illumination equipment intended for use in observation, monitoring, and diagnosis, not intended for use with a photosensitizer. This document is to be used with IEC 60601-1:2005 and its amendment 1.
- Standard32 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for Needle-Based Injection Systems (NISs) for single-patient use intended to deliver discrete volumes (bolus) of medicinal product, which can be delivered through needles or soft cannulas for intradermal, subcutaneous and/or intramuscular delivery, incorporating pre-filled or user-filled, replaceable or non-replaceable containers.
This document applies in cases where the NIS incorporates a prefilled syringe. However, stand-alone prefilled syringes defined by ISO 11040-8 are not covered by this document (see exclusions below).
It is important to note that other functions and characteristics of the prefilled syringe, such as dose accuracy, are subject to the requirements (delivered volume) in ISO 11040-8 and not this document, unless the addition impacts the delivery function (e.g. a mechanism that intends to restrict or stop the plunger movement, which would limit the dose delivered). In that case, the system is completely covered by this document and applicable requirements of the ISO 11608 series.
Excluded from the scope are:
— stand-alone prefilled syringes defined by ISO 11040-8 (with noted exceptions above);
— NISs that provide continuous delivery and require a delivery rate clinically specified in the medicinal product labelling or determined by a physician based on clinical relevance (i.e. medication efficacy) as would be the case with insulin patch pumps or traditional infusion pumps (e.g. IEC 60601-2-24, ISO 28620) associated with continuous delivery of medicinal products (e.g. insulin);
— NISs with containers that can be refilled multiple times;
— requirements relating to methods or equipment associated with user filling of containers unless they are dedicated accessories (a component necessary for primary function, whether included in the original kitted product or not);
— NISs intended for dental use;
— NISs intended for different routes of administration (e.g. intravenous, intrathecal, intraocular).
NOTE These products that are excluded might benefit from elements in this document but might not completely fulfil the basic safety and effectiveness of such products.
- Standard81 pagesEnglish languagee-Library read for1 day
This document specifies a test method and the minimum requirements for fungicidal or yeasticidal activity of chemical disinfectant and antiseptic products that form a homogeneous, physically stable preparation when diluted with hard water, or - in the case of ready-to-use products - with water. Products can only be tested at a concentration of 80 % or less (97 % with a modified method for special cases) as some dilution is always produced by adding the test organisms and interfering substance.
This document applies to products that are used in the medical area in the fields of hygienic handrub, hygienic handwash, surgical handrub, surgical handwash, instrument disinfection by immersion, and surface disinfection by wiping, spraying, flooding or other means.
This document applies to areas and situations where disinfection or antisepsis is medically indicated. Such indications occur in patient care, for example:
- in hospitals, in community medical facilities and in dental institutions;
- in clinics of schools, of kindergartens and of nursing homes;
and can occur in the workplace and in the home. It can also include services such as laundries and kitchens supplying products directly for the patients.
NOTE 1 The method described is intended to determine the activity of commercial formulations or active substances under the conditions in which they are used.
NOTE 2 This method corresponds to a phase 2 step 1 test.
EN 14885 specifies in detail the relationship of the various tests to one another and to “use recommendations”.
- Standard57 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for hoists and body-support units intended for the transfer of persons with disabilities. The document applies to the following products classified in ISO 9999:—[1].
— 12 36 03 Mobile hoists for transferring a person in sitting position with sling seats;
— 12 36 04 Mobile hoists for transferring a person in standing position;
— 12 36 06 Mobile hoists for transferring a person in sitting position with solid seats;
— 12 36 09 Mobile hoists for transferring a person in lying position;
— 12 36 12 Stationary hoists fixed to walls, floor or ceiling;
— 12 36 15 Stationary hoists fixed to, or mounted in or on, another product;
— 12 36 18 Stationary free-standing hoists;
— 12 36 21 Body-support units for hoists.
This document covers different types of mobile and stationary hoists. Some of the requirements and test methods are general and others are only valid for specific product types.
Annexes A, B and C provide general recommendations.
This document does not apply to devices that transport persons between two levels (floors) of a building.
It does not include methods for the determination of ageing or corrosion of such hoists and units.
It does not include methods to qualify individual units prior to use.
The requirements of this document are formulated with regard to the needs of both the persons being hoisted and the attendant using the hoist.
[1] Under preparation. Stage at the time of publication: ISO/FDIS 9999:2021.
- Standard86 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for Needle-Based Injection Systems (NISs) for single-patient use intended to deliver discrete volumes (bolus) of medicinal product, which can be delivered through needles or soft cannulas for intradermal, subcutaneous and/or intramuscular delivery, incorporating pre-filled or user-filled, replaceable or non-replaceable containers.
This document applies in cases where the NIS incorporates a prefilled syringe. However, stand-alone prefilled syringes defined by ISO 11040-8 are not covered by this document (see exclusions below).
It is important to note that other functions and characteristics of the prefilled syringe, such as dose accuracy, are subject to the requirements (delivered volume) in ISO 11040-8 and not this document, unless the addition impacts the delivery function (e.g. a mechanism that intends to restrict or stop the plunger movement, which would limit the dose delivered). In that case, the system is completely covered by this document and applicable requirements of the ISO 11608 series.
Excluded from the scope are:
— stand-alone prefilled syringes defined by ISO 11040-8 (with noted exceptions above);
— NISs that provide continuous delivery and require a delivery rate clinically specified in the medicinal product labelling or determined by a physician based on clinical relevance (i.e. medication efficacy) as would be the case with insulin patch pumps or traditional infusion pumps (e.g. IEC 60601-2-24, ISO 28620) associated with continuous delivery of medicinal products (e.g. insulin);
— NISs with containers that can be refilled multiple times;
— requirements relating to methods or equipment associated with user filling of containers unless they are dedicated accessories (a component necessary for primary function, whether included in the original kitted product or not);
— NISs intended for dental use;
— NISs intended for different routes of administration (e.g. intravenous, intrathecal, intraocular).
NOTE These products that are excluded might benefit from elements in this document but might not completely fulfil the basic safety and effectiveness of such products.
- Standard81 pagesEnglish languagee-Library read for1 day
This document applies to ocular endotamponades (OE), a group of non-solid surgically invasive medical devices introduced into the vitreous cavity of the eye to flatten and position a detached retina onto the retinal pigment epithelium (RPE), or to tamponade the retina.
With regard to the safety and efficacy of OE, this document specifies requirements for their intended performance, design attributes, pre-clinical and clinical evaluation, sterilization, product packaging, product labelling and the information supplied by the manufacturer.
- Standard23 pagesEnglish languagee-Library read for1 day
This document specifies requirements for the information to be provided by the medical device manufacturer for the processing of critical or semi-critical medical devices (i.e. a medical device that enters normally sterile parts of the human body or a medical device that comes into contact with mucous membranes or non-intact skin) or medical devices that are intended to be sterilized.
This includes information for processing prior to use or reuse of the medical device.
Processing instructions are not defined in this document. Rather, this document specifies requirements to assist manufacturers of medical devices in providing detailed processing instructions that consist of the following activities, where applicable:
a) initial treatment at the point of use;
b) preparation before cleaning;
c) cleaning;
d) disinfection;
e) drying;
f) inspection and maintenance;
g) packaging;
h) sterilization;
i) storage;
j) transportation.
This document excludes processing of the following:
— non-critical medical devices unless they are intended to be sterilized;
— textile devices used in patient draping systems or surgical clothing;
— medical devices specified by the manufacturer for single use only and supplied ready for use.
NOTE See ISO 17664-2:2021, Annex E, for further guidance on the application of the ISO 17664 series to a medical device.
- Standard34 pagesEnglish languagee-Library read for1 day
This document establishes a vocabulary of terms used for anaesthetic and respiratory equipment and supplies, related devices and supply systems.
NOTE 1 To avoid multiple definitions of the same term in different categories, this document attempts to ensure consistency by the inclusion of a ‘general’ category, and by use of domain specifiers and unique pre-coordinated domain-specific term names.
NOTE 2 In addition to terms and definitions used in two of the three official ISO languages (English and French), this document gives the equivalent terms in the German language; these are published under the responsibility of the member body for Germany. However, only the terms and definitions given in the official languages can be considered as ISO terms and definitions.
- Standard72 pagesEnglish languagee-Library read for1 day
This document specifies procedures and test methods used to demonstrate the cleaning efficacy of washer-disinfectors (WD) and their accessories intended to be used for cleaning of reusable medical devices.
NOTE 1 The requirements can be used for washer-disinfectors intended for use with other articles used in the context of medical, dental, laboratory, pharmaceutical and veterinary practice.
NOTE 2 This document does not apply to the activities to be performed by the manufacturers of reusable medical devices.
- Standard68 pagesEnglish languagee-Library read for1 day
Establishes particular requirements for the safety of electromyographs and evoked response equipment as defined in clause 2 of this standard.
- Standard33 pagesEnglish languagee-Library read for1 day
2021-06-25 - IEC Corrected version
- Corrigendum3 pagesEnglish and French languagee-Library read for1 day
This document specifies requirements for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin.
This document covers the control of risks arising from contamination with bacteria and fungi by application of a liquid chemical sterilization process. Risks associated with other microorganisms can be assessed using other methods (see NOTE 1).
This document is not applicable to material of human origin.
This document does not describe methods for the validation of the inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents (see NOTE 2 and NOTE 3).
This document does not describe methods for validation of the inactivation or elimination of protozoa and parasites.
The requirements for validation and routine control described in this document are only applicable to the defined sterilization process of a medical device, which is performed after the manufacturing process, and do not take account of the lethal effects of other bioburden reduction steps (see NOTE 4).
This document does not specify tests to establish the effects of any chosen sterilization process upon the fitness for use of the medical device (see NOTE 5).
This document does not cover the level of residual sterilizing agent within medical devices (see NOTE 6).
Guidance for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin is provided in informative Annex A.
NOTE 1 The prior application of risk management principles to medical devices utilizing animal tissues, as described in ISO 22442-1 is important. ISO 18362 provides information on control of microbial risks during processing of cell-based health-care products.
NOTE 2 Liquid chemical sterilizing agents traditionally employed to sterilize animal tissues in medical devices might not be effective in inactivating the causative agents of TSE such as bovine spongiform encephalopathy (BSE), or scrapie. Satisfactory validation in accordance with this document does not necessarily demonstrate inactivation of infective agents of this type. Risk controls related to sourcing, collection and handling of animal materials are described in ISO 22442-2.
NOTE 3 The validation of the inactivation, elimination, or elimination and inactivation of viruses and TSE agents is described in ISO 22442-3.
NOTE 4 Manufacturing processes for medical devices containing animal tissues frequently include exposure to chemical agents which can significantly reduce the bioburden on the medical device. Following the manufacturing process, a medical device is exposed to a specified sterilization process.
NOTE 5 Such testing is a crucial part of the design and development of a medical device.
NOTE 6 ISO 10993-17 specifies a method to establish allowable limits for residues of sterilizing agents.
NOTE 7 Standards for quality management systems (see ISO 13485) can be used in the control of all stages of manufacture including the sterilization process.
- Standard50 pagesEnglish languagee-Library read for1 day
This document specifies requirements and test methods for hoists and body-support units intended for the transfer of persons with disabilities. The document applies to the following products classified in ISO 9999:—[1].
— 12 36 03 Mobile hoists for transferring a person in sitting position with sling seats;
— 12 36 04 Mobile hoists for transferring a person in standing position;
— 12 36 06 Mobile hoists for transferring a person in sitting position with solid seats;
— 12 36 09 Mobile hoists for transferring a person in lying position;
— 12 36 12 Stationary hoists fixed to walls, floor or ceiling;
— 12 36 15 Stationary hoists fixed to, or mounted in or on, another product;
— 12 36 18 Stationary free-standing hoists;
— 12 36 21 Body-support units for hoists.
This document covers different types of mobile and stationary hoists. Some of the requirements and test methods are general and others are only valid for specific product types.
Annexes A, B and C provide general recommendations.
This document does not apply to devices that transport persons between two levels (floors) of a building.
It does not include methods for the determination of ageing or corrosion of such hoists and units.
It does not include methods to qualify individual units prior to use.
The requirements of this document are formulated with regard to the needs of both the persons being hoisted and the attendant using the hoist.
[1] Under preparation. Stage at the time of publication: ISO/FDIS 9999:2021.
- Standard86 pagesEnglish languagee-Library read for1 day
This document specifies a test method and the minimum requirements for fungicidal or yeasticidal activity of chemical disinfectant and antiseptic products that form a homogeneous, physically stable preparation when diluted with hard water, or - in the case of ready-to-use products - with water. Products can only be tested at a concentration of 80 % or less (97 % with a modified method for special cases) as some dilution is always produced by adding the test organisms and interfering substance.
This document applies to products that are used in the medical area in the fields of hygienic handrub, hygienic handwash, surgical handrub, surgical handwash, instrument disinfection by immersion, and surface disinfection by wiping, spraying, flooding or other means.
This document applies to areas and situations where disinfection or antisepsis is medically indicated. Such indications occur in patient care, for example:
- in hospitals, in community medical facilities and in dental institutions;
- in clinics of schools, of kindergartens and of nursing homes;
and can occur in the workplace and in the home. It can also include services such as laundries and kitchens supplying products directly for the patients.
NOTE 1 The method described is intended to determine the activity of commercial formulations or active substances under the conditions in which they are used.
NOTE 2 This method corresponds to a phase 2 step 1 test.
EN 14885 specifies in detail the relationship of the various tests to one another and to “use recommendations”.
- Standard57 pagesEnglish languagee-Library read for1 day
This document applies to ocular endotamponades (OE), a group of non-solid surgically invasive medical devices introduced into the vitreous cavity of the eye to flatten and position a detached retina onto the retinal pigment epithelium (RPE), or to tamponade the retina.
With regard to the safety and efficacy of OE, this document specifies requirements for their intended performance, design attributes, pre-clinical and clinical evaluation, sterilization, product packaging, product labelling and the information supplied by the manufacturer.
- Standard23 pagesEnglish languagee-Library read for1 day
This document specifies requirements for the information to be provided by the medical device manufacturer for the processing of critical or semi-critical medical devices (i.e. a medical device that enters normally sterile parts of the human body or a medical device that comes into contact with mucous membranes or non-intact skin) or medical devices that are intended to be sterilized.
This includes information for processing prior to use or reuse of the medical device.
Processing instructions are not defined in this document. Rather, this document specifies requirements to assist manufacturers of medical devices in providing detailed processing instructions that consist of the following activities, where applicable:
a) initial treatment at the point of use;
b) preparation before cleaning;
c) cleaning;
d) disinfection;
e) drying;
f) inspection and maintenance;
g) packaging;
h) sterilization;
i) storage;
j) transportation.
This document excludes processing of the following:
— non-critical medical devices unless they are intended to be sterilized;
— textile devices used in patient draping systems or surgical clothing;
— medical devices specified by the manufacturer for single use only and supplied ready for use.
NOTE See ISO 17664-2:2021, Annex E, for further guidance on the application of the ISO 17664 series to a medical device.
- Standard34 pagesEnglish languagee-Library read for1 day
Frequently Asked Questions
A European Standardization Mandate is a formal request from the European Commission to the European Standardization Organizations (CEN, CENELEC, and ETSI) to develop European standards (ENs) in support of EU legislation and policies. Mandates are issued under Regulation (EU) No 1025/2012 and help ensure that products and services meet the essential requirements set out in EU directives and regulations.
M/023 is a European Standardization Mandate titled "Standardization mandate to CEN/CENELEC concerning the development of European standards relating to medical devices". Standardization mandate to CEN/CENELEC concerning the development of European standards relating to medical devices There are 1434 standards developed under this mandate.
Standards developed in response to a mandate and cited in the Official Journal of the European Union become "harmonized standards". Products manufactured in compliance with harmonized standards benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation, facilitating CE marking and market access across the European Economic Area.