Water quality — Gas-chromatographic determination of a number of monocyclic aromatic hydrocarbons, naphthalene and several chlorinated compounds using purge-and-trap and thermal desorption

ISO 15680:2003 specifies a general method for the determination of volatile organic compounds (VOCs) in water by purge-and-trap isolation and gas chromatography (GC). Annexes A, B and C provide examples of analytes that can be determined using ISO 15680:2003. They range from difluorodichloromethane (R-12) up to trichlorobenzene, including all non-polar organic compounds of intermediate volatility. Detection is preferably carried out by mass spectrometry in the electron impact mode (EI), but other detectors may be applied as well. The limit of detection largely depends on the detector in use and the operational parameters. Typically detection limits as low as 10 ng/l can be achieved. The working range typically is up to 100 micrograms per litre. ISO 15680:2003 is applicable to drinking water, ground water, surface water, seawater and to (diluted) waste water.

Qualité de l'eau — Dosage par chromatographie en phase gazeuse d'un certain nombre d'hydrocarbures aromatiques monocycliques, du naphtalène et de divers composés chlorés par dégazage, piégeage et désorption thermique

L'ISO 15680:2003 spécifie une méthode générale de dosage des composés organiques volatils (COV) dans l'eau par dégazage et piégeage suivis d'une chromatographie en phase gazeuse. Les Annexes A, B et C donnent des exemples d'analytes pouvant être dosés par l'ISO 15680:2003. Ces éléments vont du difluorodichlorométhane (R12) au trichlorobenzène, en passant par tous les composés organiques non polaires de volatilité intermédiaire. La détection se fait de préférence par spectrométrie de masse en mode impact électronique (EI) mais il est aussi possible d'utiliser d'autres détecteurs. La limite de détection dépend beaucoup du type de détecteur utilisé et des paramètres opérationnels. En règle générale, il est possible d'obtenir des limites inférieures de détection de 10 ng/l. La plage de travail typique va jusqu'à 100 microgrammes/l. L'ISO 15680:2003 peut être appliquée à l'eau potable, à l'eau souterraine, à l'eau de surface, à l'eau de mer et aux eaux résiduaires (diluées).

General Information

Status
Published
Publication Date
03-Nov-2003
Current Stage
9093 - International Standard confirmed
Completion Date
12-Dec-2019
Ref Project

Relations

Effective Date
06-Jun-2022

Buy Standard

Standard
ISO 15680:2003 - Water quality -- Gas-chromatographic determination of a number of monocyclic aromatic hydrocarbons, naphthalene and several chlorinated compounds using purge-and-trap and thermal desorption
English language
33 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 15680:2003 - Qualité de l'eau -- Dosage par chromatographie en phase gazeuse d'un certain nombre d'hydrocarbures aromatiques monocycliques, du naphtalene et de divers composés chlorés par dégazage, piégeage et désorption thermique
French language
34 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 15680
First edition
2003-11-01


Water quality — Gas-chromatographic
determination of a number of monocyclic
aromatic hydrocarbons, naphthalene and
several chlorinated compounds using
purge-and-trap and thermal desorption
Qualité de l'eau — Dosage par chromatographie en phase gazeuse
d'un certain nombre d'hydrocarbures aromatiques monocycliques, du
naphtalène et de divers composés chlorés par dégazage, piégeage et
désorption thermique




Reference number
ISO 15680:2003(E)
©
ISO 2003

---------------------- Page: 1 ----------------------
ISO 15680:2003(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


©  ISO 2003
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2003 — All rights reserved

---------------------- Page: 2 ----------------------
ISO 15680:2003(E)
Contents Page
Foreword. iv
1 Scope. 1
2 Normative references. 1
3 Terms and definitions. 2
4 Principle. 2
5 Interferences. 3
6 Reagents. 4
7 Apparatus. 7
8 Sample collection, preservation and preparation . 8
9 Analytical procedure. 9
10 Calibration. 11
11 Calculation. 12
12 Expression of results. 12
13 Precision data. 12
14 Test report. 12
Annex A (informative) Application of purge-and-trap concentration to the GC analysis of volatile
compounds in water — Example 1: Validation study in the UK. 14
Annex B (informative) Application of purge-and-trap concentration to the GC-analysis of volatile
compounds in water — Example 2: Data provided by DIN. 19
Annex C (informative) Application of purge-and-trap concentration to the GC-analysis of volatile
compounds in water — Example 3: Validation study in the Netherlands. 22
Annex D (normative) Criteria for the GC-MS identification of target compounds. 25
Annex E (informative) Procedures for the cleaning of glassware and the preparation of
contaminant-free water. 28
Annex F (informative) Preparation of standard solutions of volatile organic compounds . 30
Annex G (informative) Determination of the (absolute) recovery of substances analysed by
purge-and-trap concentration. 32
Bibliography . 33

© ISO 2003 — All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 15680:2003(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 15680 was prepared by Technical Committee ISO/TC 147, Water quality, Subcommittee SC 2, Physical,
chemical and biochemical methods.

iv © ISO 2003 — All rights reserved

---------------------- Page: 4 ----------------------
INTERNATIONAL STANDARD ISO 15680:2003(E)

Water quality — Gas-chromatographic determination of a
number of monocyclic aromatic hydrocarbons, naphthalene
and several chlorinated compounds using purge-and-trap and
thermal desorption
WARNING — Persons using this International Standard should be familiar with normal laboratory
practice. This International Standard does not purport to address all of the safety problems, if any,
associated with its use. It is the responsibility of the user to establish appropriate safety and health
practices and to ensure compliance with any national regulatory conditions.
1 Scope
This International Standard specifies a general method for the determination of volatile organic compounds
(VOCs) in water by purge-and-trap isolation and gas chromatography (GC). Annexes A, B and C provide
examples of analytes that can be determined using this International Standard. They range from
difluorodichloromethane (R-12) up to trichlorobenzene, including all non-polar organic compounds of
intermediate volatility.
Detection is preferably carried out by mass spectrometry in the electron impact mode (EI), but other detectors
may be applied as well.
The limit of detection largely depends on the detector in use and the operational parameters. Typically
1)
detection limits as low as 10 ng/l can be achieved. The working range typically is up to 100 µg/l.
This International Standard is applicable to drinking water, ground water, surface water, seawater and to
(diluted) waste water.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 3696, Water for analytical laboratory use — Specifications and test methods
ISO 5667-3, Water quality — Sampling — Part 3: Guidance on the preservation and handling of water
samples
ISO 8466-1, Water quality — Calibration and evaluation of analytical methods and estimation of performance
characteristics — Part 1: Statistical evaluation of the linear calibration function

1) The value given is an indication of the limit of detection. It is calculated as 3 times the standard deviation of a series of
measurements of 10 replicate samples under conditions of repeatability.
© ISO 2003 — All rights reserved 1

---------------------- Page: 5 ----------------------
ISO 15680:2003(E)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
volatile organic compound
VOC
organic compound, generally non-polar, with boiling point between approximately −30 °C and 220 °C
3.2
target compound
selected component whose presence or absence is determined
NOTE This definition can also apply to a derivative of the original compound which is formed during an intentional
derivatization procedure.
3.3
standard compound
target compound with the highest possible purity that can be used as a reference during the analysis and free
of impurities having any influence on its mass spectrum
3.4
retention-time standard
compound that is added to the sample (or to the sample extract) and to the external standard solution (3.6)
and whose retention time is used to calculate the relative retention times of the target compounds
NOTE The retention-time standard may be identical to the internal standard(s).
3.5
relative retention time
ratio between the retention time of the target compound and the retention time of the retention-time standard
3.6
external standard solution
solution of a known concentration of the target compounds
3.7
lowest concentration for identification
lowest concentration of the target compound which, if present in the sample, still can be identified using the
identification criterion that the selected diagnostic ion with the lowest intensity is still present in the mass
spectrum with a signal-to-noise ratio higher than 3:1
NOTE This concentration strongly depends on the sensitivity of the instrument and on the performance
characteristics of the analytical method.
3.8
diagnostic ion
ion selected from the mass spectrum of the target compound with the highest possible specificity
NOTE For the selection of diagnostic ions, see D.5.
4 Principle
A fixed volume of sample is purged with a fixed volume of an inert gas to strip out the volatile components
which are subsequently trapped. This trapping can be either:
a) on a packed adsorbent trap (preferably combined with or followed by a cryofocusing system), or
b) directly on a capillary cold-trap.
2 © ISO 2003 — All rights reserved

---------------------- Page: 6 ----------------------
ISO 15680:2003(E)
After completion of the purge process, the trap is heated to desorb the volatile components which are swept
by the GC carrier gas on to a capillary GC column. This transfer to the GC column can be done in an on-line
or in an off-line set-up. To achieve narrow injection bandwidths, the use of a cryofocusing system is
recommended when the trapping is done on a packed adsorbent trap as in a) or the transfer is done through
an injector-splitter set at approximately 20:1 if the sensitivity of the analytical system allows this.
The components are separated by GC utilizing temperature programming, and are detected by the use of a
mass spectrometer. Data are acquired in the full-scan mode or at a sufficient number of specific fragments to
enable matching against those of the standards. A compound is regarded to be present when the criteria of
Annex D are met. Quantification is carried out using selected characteristic fragments for each determinand.
5 Interferences
5.1 General
In principle, any purgeable compound which elutes at the same chromatographic retention time and produces
a mass spectrum identical, or very similar, to any determinand under investigation will interfere. In practice,
this is unlikely as the spectra of most of the determinands are characteristic. With retention-time data and the
availability of the spectrum over a wide range of masses, the possibility of misidentification is quite small. Co-
eluting peaks with ions with non-specific m/z values might cause interference, but quantification ions can be
chosen to preclude this.
Contamination introduced during the analytical procedure is monitored by the determination of blanks (9.4).
5.2 Interferences in the sampling process
VOCs are amenable to evaporation or degassing during the sampling process, transportation, storage and
preparation of the samples. This can result in measured concentrations which are too low. VOCs can also
diffuse into the samples from the ambient air of the laboratory or from air in the refrigerator where samples are
stored. This results in concentrations which are too high.
5.3 Interferences due to the purge gas and the GC gas
Insufficient purity of the purge gas or the GC carrier gas can cause interferences.
5.4 Interferences in the purge-and-trap process
One of the main sources of contamination during sample transportation is contaminated laboratory air in the
purge vessel or sample container. Therefore, the laboratory should be free of solvents and concentrated
standard solutions.
Laboratory clothing is also a potential source of contamination, particularly of highly volatile halogenated
hydrocarbons.
To avoid interferences, all materials (tubing, seals, valves, etc.) should be made from stainless steel or glass.
The use of plastics material should be avoided. All glassware directly in contact with the sample or purged
compounds should be cleaned thoroughly (see Annex E). There is an especially high risk of entrainment after
the measurement of highly polluted samples.
Purge vessels incorporating a glass frit are liable to cause cross-contamination (see also 7.1).
Purging of water samples containing surfactants can result in formation of foam which might be in direct
contact with the adsorbent. If this occurs, the purge procedure shall be stopped immediately.
© ISO 2003 — All rights reserved 3

---------------------- Page: 7 ----------------------
ISO 15680:2003(E)
5.5 Interferences in the thermal desorption process
During thermal desorption, substances can degrade.
The transfer lines between the adsorption trap and the gas chromatography injection system should not have
any “cold” points which act as adsorbents, as this results in a loss of VOCs.
When using a cryofocusing system and if the adsorbents are not completely dried after the purge process, the
capillaries can block with ice. This results in incomplete desorption, and evaluation of the analytical procedure
will be impossible.
The adsorbents used in the purge-and-trap systems are subject to ageing (contamination, thermal stress)
which can cause changes in the trapping capacity and in the blank values.
5.6 Interferences in automatic samplers
Samples in autosamplers intended for subsequent analysis shall be protected from light (e.g. by use of brown
glass vials).
Special care should be taken for autosamplers with respect to the remarks made under 5.4.
6 Reagents
Use reagents of sufficient purity that do not give rise to interfering peaks in the gas chromatographic analysis.
Check freshly prepared standard solutions against previously prepared standard solutions to ensure for
standard integrity. This should be checked with each batch of material by analysing procedural blank solutions
with each batch of samples. Use solvents of high quality that do not contain interfering compounds and
analytical reagent grade materials, as far as available. Reagents may contaminate by contact with air and/or
other materials, particularly plastics, or by degradation caused by reaction with light. Reagents should be
stored in all-glass containers or other vessels found to be suitable, and kept in the dark, if necessary.
6.1 Water, used for blank determination, dilution of samples and for the preparation of calibration solutions.
Water should be known to be free from contaminants (see Annex E). It should show negligible interferences in
comparison with the smallest concentration to be determined, in accordance with ISO 3696.
A sufficient amount of water from the same batch should be available to complete each batch of analyses,
including all preparations.
6.2 Methanol, CH OH, used as solvent, and for the preparation of standard stock solutions.
3
Other solvents that are readily soluble in water and do not interfere with the analytical process can be used as
well. This includes N,N-dimethylformamide (DMF, C H NO), dimethyl sulfoxide (DMSO, C H SO) and
3 7 2 6
acetone (C H O).
3 6
6.3 Sodium thiosulfate pentahydrate, Na S O ⋅5H O.
2 2 3 2
If necessary, add sodium thiosulfate to samples to remove remaining oxidants like chlorine or ozone. Other
non-interfering substances may be used for the same purpose (e.g. sodium sulfite).
NOTE Already formed intermediate oxidation products like halogenated acetic acids can still form trihalomethanes
regardless of the preservation described in this clause.
6.4 Sodium hydrogensulfate, NaHSO .
4
Other suitable diluted acids or acid salts may be used as well.
4 © ISO 2003 — All rights reserved

---------------------- Page: 8 ----------------------
ISO 15680:2003(E)
6.5 Purge gas.
Use a high quality helium or nitrogen gas for purging, free of interfering substances. Impurities can be
eliminated by a purification cartridge, if necessary.
6.6 Standard solutions.
Owing to the high volatility of the gases and the more volatile compounds to be analysed, great care is
required in the preparation of standard solutions; losses may occur in the headspace of the vessel used to
prepare standard solutions. For a detailed description of the preparation of standard stock solutions of volatile
compounds see Annex F. It is advisable, and more appropriate, to use commercially available standard
solutions. Store intermediate standard solutions at about 4 °C and allow them to reach room temperature
before use.
Whilst the following procedures are given as examples, users may wish to prepare their own standard
solutions by an alternative procedure or by diluting commercially available stock solutions (preferably certified),
which are shown to produce equivalent results.
6.6.1 Stock calibration standard solution (2 mg/ml).
Dissolve defined quantities of approximately 200 mg of each VOC in a 100 ml volumetric flask partially filled
with the same solvent (6.2), make up to the mark and mix well. See also Annex F.
6.6.2 Stock internal standard solutions (2 mg/ml).
Dissolve defined quantities of approximately 200 mg of each internal standard compound in a 100 ml
volumetric flask partially filled with the same solvent (6.2), make up to the mark and mix well. See also
Annex F.
At least one internal standard compound should be used for quantitation and additional internal standard
compounds could be used as surrogate standards. Suitable compounds may be selected from Table 1. Use
deuterated standards only for GC-MS. For the indicated internal standard compounds (*) of Table 1, the range
of analytes covered by each of them is given in Annex A as an example (Table A.2).
6.6.3 Spiking solutions.
Prepare spiking solutions from solutions 6.6.1 and 6.6.2 by appropriate dilution in a volumetric flask containing
the same solvent (6.2). As an example, Table 2 gives a dilution scheme in 100 ml of solvent and consecutive
spiking of 5 µl of it to 100 ml water to give spiking solutions 6.6.3.1 to 6.6.3.6. In this example, analyte
concentrations range from 0 µg/l to 5 µg/l in water.
If the desired measuring range differs from that of Table 2, different dilution ratios should be taken or the
spiking volume should be adapted.
NOTE Solution 6.6.3.1 is used as the internal standard solution to be added to each sample (see 9.3).
6.6.4 Calibration solutions.
Add a small volume of the spiking solutions 6.6.3.2. to 6.6.3.6 from Table 2 to the water (6.1) in the purge
vessel (7.1) [or in the sample container (7.3) when an autosampler is used]. Table 2 gives an example of a
5 µl addition to 100 ml of water (with concentrations as indicated in the fourth column). If larger sample
volumes are analysed, add an equivalently larger volume of spiking solution.
Make sure that the content of the organic solvent in the final aqueous calibration standard solution does not
exceed 2 % (volume fraction). If a high percentage of solvent is present, linearity should be checked.
6.6.5 Blank solution.
Reserve a portion of the unspiked water for use as a quality control blank.
© ISO 2003 — All rights reserved 5

---------------------- Page: 9 ----------------------
ISO 15680:2003(E)
Table 1 — Internal standard compounds
CAS-RN Compound Formula
462-06-6 * monofluorobenzene C H F
6 5
3114-55-4 * monochlorobenzene-d C CID
5 6 5
3855-82-1 * 1,4-dichlorobenzene-d C Cl D
4 6 2 4
540-36-3 * 1,4-difluorobenzene C H F
6 4 2
460-00-4 1-bromo-4-fluorobenzene C H BrF
6 4
2037-26-5 toluene-d C D
8 7 8
1868-53-7 dibromofluoromethane CHBr F
2
109-70-6 1-bromo-3-chloropropane C H BrCl
3 6
107-04-0 1-bromo-2-chloroethane C H BrCl
2 4
75-62-7 bromotrichloromethane CBrCl
3
363-72-4 pentafluorobenzene C HF
6 5
1076-43-3 benzene-d C D
6 6 6
17060-07-0 1,2-dichloroethane-d C Cl D
4 2 2 4
20302-26-5 ethylbenzene-ring-d C H D
5 8 5 5
74-97-5 bromochloromethane CH BrCl
2
3017-95-6 2-bromo-1-chloropropane C H BrCl
3 6
110-56-5 1,4-dichlorobutane C H Cl
4 8 2
56004-61-6 o-xylene-d C D
10 8 10
* Range of analytes covered is given in Table A.2
Table 2 — Dilution scheme in 100 ml solvent
ml of 6.6.2 ml of 6.6.1 Concentration (in µg/l) in
Analyte concentration in
Spiking solution (added to (added to calibration solution (5 µl of
spiking solution (in mg/l of
(100 ml of solvent) 100 ml of 100 ml of spiking solution added to
solvent)
a
solvent) solvent) 100 ml of water)
6.6.3.1 5 0 0 0
6.6.3.2 5 1 20 1
6.6.3.3 5 2 40 2
6.6.3.4 5 3 60 3
6.6.3.5 5 4 80 4
6.6.3.6 5 5 100 5
a
The concentration of the internal standard compound in each spiking solution is 100 mg/l.
6 © ISO 2003 — All rights reserved

---------------------- Page: 10 ----------------------
ISO 15680:2003(E)
7 Apparatus
Usual laboratory glassware and equipment is not specified, as the actual devices used depend on the specific
application and circumstances. Make sure that all devices are free of interfering compounds. Clean all
glassware, including sample bottles, thoroughly. A standard procedure for cleaning is included in Annex E.
7.1 Purge vessels
A variety of purge vessels are commercially available. The specific type is defined by the purge-and-trap
apparatus in use. There are systems available which allow purging in the sampling vessels. Cleaning of the
purge vessels should be carried out according to Annex E.
7.2 Sample containers
Various sample containers can be used, e.g. screw-cap containers fitted with PTFE-faced silicone discs. For
purge-and-trap systems with an autosampler, use sample containers recommended by the autosampler
manufacturer. Whenever septa are employed, do not re-use them.
7.3 Purge-and-trap apparatus
Purge-and-trap apparatus is commercially available or can be self-constructed. This includes fully automated,
on-line purge-and-trap GC-equipment with an autosampler and the thermal desorption device incorporated in
the instrument, as well as manually operated off-line equipment. All instruments may be used that meet the
requirements and have proven to give reliable results. According to Annexes A, B and C, various instruments
have been used. Other devices may be suitable but should be examined appropriately to ensure satisfactory
performance.
The purge-and-trap apparatus should include:
a) autosampler;
b) purge vessel, heating mantle and temperature control, purge gas supply, flowrate control, timer;
c) condenser and coolant supply or dry-purge system;
d) adsorbent trap;
e) thermal desorption device, temperature control, timer;
f) cold trap, coolant supply, heater, temperature control;
g) GC-MS or a GC with suitable detector(s), GC-auxiliaries, data system.
Various combinations of components a), c), d), e) and f) are possible and don’t all need to be included.
7.4 Adsorbent trap
7.4.1 For purge-and-trap apparatus using intermediate trapping on a packed adsorption column (see
Clause 4), these traps are often home-made or can be obtained in various modifications. As an example,
adsorption columns are made of glass or stainless steel with an internal diameter of 2 mm to 5 mm,
appropriate for use in the apparatus for thermal desorption. Adsorbent traps are packed with a suitable
adsorbent.
© ISO 2003 — All rights reserved 7

---------------------- Page: 11 ----------------------
ISO 15680:2003(E)
2)
Generally a polymer, a carbonaceous or silica adsorbent is used . Typical dimensions of the packing are
diameter 2 mm to 5 mm, length 10 mm to 50 mm, corresponding to at least 90 mg of adsorbent. The
adsorbent is kept in position by inert material such as glass wool plugs or glass screens. This description is an
example; other adsorbent traps can be used as well, provided their performance meets the requirements of
this International Standard.
Prior to their first use, adsorbent traps should be conditioned by heating them above their desorption
temperature for approximately 30 min while passing a gentle stream of inert gas through them. A blank
procedure shall be performed with the conditioned adsorbent trap before using it in routine.
7.4.2 Special requirements for use in off-line purge-and-trap equipment may apply.
In off-line purge-and-trap equipment, the adsorbent traps are not defined by the instrument in use, whereas
most of the other parts under 7.3 a) to f) are. For off-line use, mark the traps on one side to allow desorption in
a back-flush mode. In case of off-line purge-and-trap, for the adsorbent traps use caps of inert material, for
example PTFE, or of metal with screw windings and a PTFE-washer, so that after purging they can be closed
leakproof for storage or transfer to the apparatus for thermal desorption.
7.5 Gas chromatograph-mass spectrometer (GC-MS)
A variety of gas chromatographic columns can be used in purge-and-trap analysis. Examples of suitable
columns are given in Annexes A, B and C.
The mass spectrometer should be capable of operating across the mass range of interest and incorporate a
data system capable of quantifying ions using selected m/z values. See Annexes A, B and C for typical
chromatograms.
Other GC detectors, such as flame ionization detector (FID), electron capture detector (ECD), photo-ionization
detector (PID) or electrolytic conductivity detector (ELCD), can be used, depending on the substances to be
analysed (see 9.7.2).
For operational aspects of the instruments, the manufacturer’s instructions should be followed.
8 Sample collection, preservation and preparation
Collect samples in accordance with ISO 5667-3 in suitable containers, preferably directly into the sample
containers (7.3). It is advisable to take two samples, one to be retained in the event of a repeat analysis being
required. Fill sample containers, avoiding turbulence, until overflowing. Cap sample containers without leaving
a headspace. For samples containing free chlorine or any other strong oxidant, solid sodium thiosulfate
pentahydrate (6.3) or other reducing salt should be added to the container (approximately 100 mg/l).
Additionally, for the preservation of aromatic compounds in surface waters, the pH should be lowered to 2
using sodium hydrogensulfate (6.4). Other appropriate acids are allowed.
Samples shall not be diluted if the concentration exceeds the working range established by the calibration
function, as dilution can cause evaporative losses of the analytes. Preferably extend the calibration function or
[1] [2]
apply (static) headspace analysis, e.g. according to ISO 10301 and/or ISO 11423-1 . Avoid contamination
of the equipment by dirty samples.
The stability of certain determinands is known to be matrix-depend
...

NORME ISO
INTERNATIONALE 15680
Première édition
2003-11-01



Qualité de l'eau — Dosage par
chromatographie en phase gazeuse d'un
certain nombre d'hydrocarbures
aromatiques monocycliques, du
naphtalène et de divers composés
chlorés par dégazage, piégeage et
désorption thermique
Water quality — Gas-chromatographic determination of a number of
monocyclic aromatic hydrocarbons, naphthalene and several
chlorinated compounds using purge-and-trap and thermal desorption




Numéro de référence
ISO 15680:2003(F)
©
ISO 2003

---------------------- Page: 1 ----------------------
ISO 15680:2003(F)
PDF – Exonération de responsabilité
Le présent fichier PDF peut contenir des polices de caractères intégrées. Conformément aux conditions de licence d'Adobe, ce fichier
peut être imprimé ou visualisé, mais ne doit pas être modifié à moins que l'ordinateur employé à cet effet ne bénéficie d'une licence
autorisant l'utilisation de ces polices et que celles-ci y soient installées. Lors du téléchargement de ce fichier, les parties concernées
acceptent de fait la responsabilité de ne pas enfreindre les conditions de licence d'Adobe. Le Secrétariat central de l'ISO décline toute
responsabilité en la matière.
Adobe est une marque déposée d'Adobe Systems Incorporated.
Les détails relatifs aux produits logiciels utilisés pour la création du présent fichier PDF sont disponibles dans la rubrique General Info
du fichier; les paramètres de création PDF ont été optimisés pour l'impression. Toutes les mesures ont été prises pour garantir
l'exploitation de ce fichier par les comités membres de l'ISO. Dans le cas peu probable où surviendrait un problème d'utilisation,
veuillez en informer le Secrétariat central à l'adresse donnée ci-dessous.


©  ISO 2003
Droits de reproduction réservés. Sauf prescription différente, aucune partie de cette publication ne peut être reproduite ni utilisée sous
quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les microfilms, sans l'accord écrit
de l'ISO à l'adresse ci-après ou du comité membre de l'ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax. + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse

ii © ISO 2003 — Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 15680:2003(F)
Sommaire Page
Avant-propos. iv
1 Domaine d'application. 1
2 Références normatives. 1
3 Termes et définitions . 2
4 Principe. 3
5 Interférences. 3
6 Réactifs. 4
7 Appareillage. 7
8 Prélèvement, conservation et préparation des échantillons. 9
9 Mode opératoire analytique. 9
10 Étalonnage. 11
11 Calcul. 12
12 Expression des résultats. 13
13 Données de fidélité . 13
14 Rapport d'essai. 13
Annexe A (informative) Application de la concentration par dégazage et piégeage à l'analyse par
chromatographie en phase gazeuse des composés volatils dans l'eau — Exemple 1:
Étude de validation conduite au Royaume-Uni. 14
Annexe B (informative) Application de la concentration par dégazage et piégeage à l'analyse par
chromatographie en phase gazeuse des composés volatils dans l'eau — Exemple 2:
fourni par le DIN . 19
Annexe C (informative) Application de la concentration par dégazage et piégeage à l'analyse par
chromatographie en phase gazeuse des composés volatils dans l'eau — Exemple 3:
Étude de validation conduite aux Pays-Bas. 22
Annexe D (normative) Critères d'identification des composés cibles par CG-MS. 25
Annexe E (informative) Modes opératoires de nettoyage de la verrerie et de préparation de l'eau
exempte de contaminants. 29
Annexe F (informative) Préparation de solutions étalons de composés organiques volatils. 31
Annexe G (informative) Détermination du rendement (absolu) des substances analysées par
dégazage et piégeage . 33
Bibliographie . 34

© ISO 2003 — Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 15680:2003(F)
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux de
normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général confiée
aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire partie du
comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec
la Commission électrotechnique internationale (CEI) en ce qui concerne la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives ISO/CEI,
Partie 2.
La tâche principale des comités techniques est d'élaborer les Normes internationales. Les projets de Normes
internationales adoptés par les comités techniques sont soumis aux comités membres pour vote. Leur
publication comme Normes internationales requiert l'approbation de 75 % au moins des comités membres
votants.
L'attention est appelée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de ne
pas avoir identifié de tels droits de propriété et averti de leur existence.
L'ISO 15680 a été élaborée par le comité technique ISO/TC 147, Qualité de l'eau, sous-comité SC 2,
Méthodes physiques, chimiques et biochimiques.

iv © ISO 2003 — Tous droits réservés

---------------------- Page: 4 ----------------------
NORME INTERNATIONALE ISO 15680:2003(F)

Qualité de l'eau — Dosage par chromatographie en phase
gazeuse d'un certain nombre d'hydrocarbures aromatiques
monocycliques, du naphtalène et de divers composés chlorés
par dégazage, piégeage et désorption thermique
AVERTISSEMENT — Il convient que les personnes utilisant la présente Norme internationale soient
familières avec les pratiques courantes de laboratoire. La présente Norme internationale ne prétend
pas aborder tous les éventuels problèmes de sécurité liés à son utilisation. Il est de la responsabilité
de l'utilisateur d'établir des pratiques de santé et de sécurité appropriées et de s'assurer de la
conformité aux exigences réglementaires nationales.
1 Domaine d'application
La présente Norme internationale spécifie une méthode générale de dosage des composés organiques
volatils (COV) dans l'eau par dégazage et piégeage suivis d'une chromatographie en phase gazeuse. Les
Annexes A, B et C donnent des exemples d'analytes pouvant être dosés par la présente Norme internationale.
Ces éléments vont du difluorodichlorométhane (R12) au trichlorobenzène, en passant par tous les composés
organiques non polaires de volatilité intermédiaire.
La détection se fait de préférence par spectrométrie de masse en mode impact électronique (EI) mais il est
aussi possible d'utiliser d'autres détecteurs.
La limite de détection dépend beaucoup du type de détecteur utilisé et des paramètres opérationnels. En
1)
règle générale, il est possible d'obtenir des limites inférieures de détection de 10 ng/l . La plage de travail
typique va jusqu'à 100 µg/l.
La présente Norme internationale peut être appliquée à l'eau potable, à l'eau souterraine, à l'eau de surface, à
l'eau de mer et aux eaux résiduaires (diluées).
2 Références normatives
Les documents de référence suivants sont indispensables pour l'application du présent document. Pour les
références datées, seule l'édition citée s'applique. Pour les références non datées, la dernière édition du
document de référence s'applique (y compris les éventuels amendements).
ISO 3696, Eau pour laboratoire à usage analytique — Spécification et méthodes d'essai
ISO 5667-3, Qualité de l'eau — Échantillonnage — Partie 3: Guide général pour la conservation et la
manipulation des échantillons d’eau
ISO 8466-1, Qualité de l'eau — Étalonnage et évaluation des méthodes d'analyse et estimation des
caractères de performance — Partie 1: Évaluation statistique de la fonction linéaire d'étalonnage

1) La valeur de limite de détection est donnée à titre indicatif. Elle est calculée comme étant égale à 3 fois l'écart-type
d'une série de mesures effectuées sur 10 répliques dans des conditions de répétabilité.
© ISO 2003 — Tous droits réservés 1

---------------------- Page: 5 ----------------------
ISO 15680:2003(F)
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s'appliquent.
3.1
composé organique volatil
COV
composé organique (généralement non polaire) dont la température d'ébullition est généralement comprise
entre −30 °C et 220 °C
3.2
composé cible
composé sélectionné dont la présence ou l'absence est déterminée
NOTE Cette définition s'applique également aux dérivés du composé d'origine qui se forment pendant une opération
volontaire de dérivation.
3.3
composé étalon
composé cible de pureté la plus élevée possible pouvant servir de référence pendant l'analyse et ne
renfermant aucune impureté ayant une quelconque influence sur son spectre de masse
3.4
étalon de temps de rétention
composé ajouté à l'échantillon (ou à l'extrait d'échantillon) et à la solution d'étalonnage externe (3.6) et dont
le temps de rétention sert à calculer les temps de rétention relatifs des composés cibles
NOTE L'étalon de temps de rétention peut être identique à l'étalon ou aux étalons internes.
3.5
temps de rétention relatif
rapport entre le temps de rétention du composé cible et le temps de rétention de l'étalon de temps de
rétention
3.6
solution d'étalonnage externe
solution de concentration connue des composés cibles
3.7
concentration minimale pour identification
concentration la plus faible du composé cible qui, si ce dernier est présent dans l'échantillon, peut encore être
identifiée à partir du critère d'identification suivant: l'ion de diagnostic sélectionné de plus faible intensité est
toujours présent dans le spectre de masse avec un rapport signal/bruit supérieur à 3:1
NOTE Cette concentration dépend fortement de la sensibilité de l'instrument et des caractéristiques de performance
de la méthode d'analyse.
3.8
ion de diagnostic
ion sélectionné dans le spectre de masse du composé cible qui a la plus grande spécificité
NOTE Pour le choix des ions de diagnostic, voir D.5.
2 © ISO 2003 — Tous droits réservés

---------------------- Page: 6 ----------------------
ISO 15680:2003(F)
4 Principe
Un volume fixe d'échantillon est dégazé avec un volume fixe de gaz inerte pour extraire les composés volatils
qui sont ensuite adsorbés. Le piégeage peut se faire:
a) sur un piège garni de substance adsorbante (couplé ou non avec un système de cryofocalisation), ou
b) directement sur un piège capillaire froid.
Une fois le processus de dégazage terminé, le piège est chauffé pour désorber les composés volatils qui sont
ensuite déplacés par le gaz vecteur vers une colonne capillaire de chromatographie en phase gazeuse. Le
transfert dans la colonne chromatographique peut se faire dans un montage en ligne ou hors ligne. Pour
obtenir une bande d'injection de faible largeur, il est recommandé d'utiliser un système de cryofocalisation
quand le piégeage se fait dans un piège garni de substance adsorbante comme en a) ou quand le transfert
intervient par l'intermédiaire d'un diviseur réglé à un rapport d'environ 20:1 si la sensibilité du système
analytique le permet.
Les composés sont ensuite séparés par chromatographie en phase gazeuse avec programmation de
température et détectés à l'aide d'un spectromètre de masse. Les données doivent être recueillies en mode
de balayage total ou sur un nombre suffisant de fragments spécifiques pour permettre la comparaison avec
les étalons. Un composé est considéré présent lorsque les critères de l'Annexe D sont respectés. La
quantification est effectuée à l'aide des fragments caractéristiques choisis pour chaque analyte.
5 Interférences
5.1 Généralités
En principe il y a interférence pour tout composé pouvant être dégazé qui élue au même temps de rétention
chromatographique que l'élément à doser et qui produit un spectre de masse identique ou presque similaire à
celui de cet élément. En pratique, cette interférence est peu probable dans la mesure où les spectres de la
plupart des éléments à doser sont caractéristiques. La possibilité d'une erreur d'identification est donc très
faible, compte tenu des données disponibles sur les temps de rétention et les spectres sur une large gamme
de masses. Des pics coélués avec des ions de valeur m/z non spécifique peuvent provoquer des interférences
mais les ions de quantification peuvent être choisis de manière à éviter ce phénomène.
La contamination introduite au cours du mode opératoire d'analyse est contrôlée par une détermination des
blancs (9.4).
5.2 Interférences dans le processus d'échantillonnage
Les composés organiques volatils peuvent subir une évaporation ou un dégazage pendant le processus
d'échantillonnage, le transport, le stockage ou la préparation des échantillons. Cela peut entraîner des valeurs
de concentration mesurées trop faibles. Des composés organiques volatils provenant de l'air ambiant du
laboratoire ou du réfrigérateur où sont conservés les échantillons peuvent diffuser dans les échantillons
eux-mêmes. Cela conduit à des concentrations trop élevées.
5.3 Interférences dues au gaz de dégazage et au gaz de chromatographie
Une pureté insuffisante du gaz pour dégazage et du gaz vecteur de chromatographie peut provoquer des
interférences.
5.4 Interférences dans le processus de dégazage et de piégeage
L'une des principales sources de contamination pendant le transport est la contamination par l'air du
laboratoire du récipient pour dégazage ou du récipient de l'échantillon. Il convient donc que le laboratoire soit
exempt de solvants et de solutions étalons concentrées.
© ISO 2003 — Tous droits réservés 3

---------------------- Page: 7 ----------------------
ISO 15680:2003(F)
Les vêtements du personnel de laboratoire sont également une source potentielle de contamination,
notamment des hydrocarbures halogénés hautement volatils.
Pour éviter les interférences, il convient que tous les matériels (tubes, joints, robinetterie, etc.) soient
fabriqués en acier inoxydable ou en verre. Il est recommandé d'éviter les matériaux plastiques. Il y a lieu de
nettoyer soigneusement toute la verrerie directement en contact avec l'échantillon ou les composés dégazés
(voir Annexe E). Le risque d'entraînement est élevé, notamment après la mesure d'échantillons fortement
pollués.
Les récipients pour dégazage comportant un verre fritté sont susceptibles de provoquer des contaminations
croisées (voir aussi 7.1).
Le dégazage des échantillons d'eau contenant des agents tensio-actifs peut provoquer la formation de
mousse qui peut entrer en contact direct avec l'adsorbant. Dans ce cas, le dégazage doit être arrêté
immédiatement.
5.5 Interférences dans le processus de désorption thermique
Les substances peuvent se dégrader pendant le processus de désorption thermique.
Il est recommandé que les conduites de transfert entre le piège adsorbant et les systèmes d'injection pour
chromatographie en phase gazeuse n'aient aucun point «froid» susceptible de jouer le rôle d'adsorbant et
d'entraîner la perte de composés organiques volatils.
Si un système de cryofocalisation est utilisé et si les adsorbants ne sont pas totalement séchés après le
dégazage, les capillaires peuvent être complètement obstrués par la glace. Ce phénomène entraîne une
désorption incomplète et l'impossibilité d'évaluer le mode opératoire d'analyse.
Les adsorbants utilisés dans les systèmes de dégazage et de piégeage sont sujets à un vieillissement
(contamination, contrainte thermique) qui peut provoquer des modifications de la capacité de piégeage et des
valeurs à blanc.
5.6 Interférences dans les échantillonneurs automatiques
Les échantillons provenant d'échantillonneurs automatiques et destinés à des analyses ultérieures doivent
être protégés de la lumière (flacons en verre brun).
Il convient de respecter scrupuleusement les remarques de 5.4 pour les échantillonneurs automatiques.
6 Réactifs
Utiliser des réactifs de pureté suffisante qui ne donnent pas lieu à des pics interférents au cours de l'analyse
par chromatographie en phase gazeuse. Contrôler les solutions étalons préparées extemporanément par
comparaison avec des solutions étalons préparées antérieurement pour vérifier leur intégrité. Il convient de
faire ce contrôle sur chaque lot de matériau en analysant des solutions à blanc avec chaque lot d'échantillons.
Utiliser si possible des solvants de haute qualité qui ne contiennent pas de composés interférents et des
réactifs de qualité analytique. Les réactifs peuvent contaminer par contact avec l'air et/ou d'autres matériaux,
notamment les plastiques, ou par dégradation à la suite d'une réaction avec la lumière. Il convient de
conserver les réactifs, à l'abri de la lumière si nécessaire, dans des récipients en verre ou en matériau jugé
approprié.
6.1 Eau, utilisée pour les déterminations à blanc, la dilution des échantillons et la préparation des solutions
d'étalonnage.
Il est recommandé que l'eau soit connue pour être exempte de contaminants (voir Annexe E). Il convient
qu'elle donne des interférences négligeables comparées à la concentration la plus faible à déterminer (voir
l'ISO 3696).
4 © ISO 2003 — Tous droits réservés

---------------------- Page: 8 ----------------------
ISO 15680:2003(F)
Il est recommandé de disposer d'une quantité suffisante d'eau du même lot pour pouvoir terminer chaque
série d'analyses, y compris toutes les préparations.
6.2 Méthanol, CH OH, utilisé comme solvant et pour la préparation des solutions mères étalons.
3
D'autres solvants facilement solubles dans l'eau et qui n'interfèrent pas avec le processus analytique peuvent
également être utilisés. Il s'agit notamment du N,N-diméthylformamide (DMF, C H NO), du diméthylsulfoxyde
3 7
(DMSO, C H SO) et de l'acétone (C H O).
2 6 3 6
6.3 Thiosulfate de sodium pentahydraté, Na S O , 5H O.
2 2 3 2
Ajouter, si nécessaire, du thiosulfate de sodium aux échantillons pour éliminer les oxydants restants tels que
le chlore ou l'ozone. D'autres substances non interférentes (par exemple du sulfite de sodium) peuvent
également être utilisées dans ce but.
NOTE Les produits d'oxydation intermédiaires déjà formés comme les acides acétiques halogénés peuvent toujours
former des trihalométhanes quelle que soit la méthode de préservation décrite dans le présent article.
6.4 Hydrogénosulfate de sodium, NaHSO .
4
D'autres acides ou sels acides dilués convenables peuvent également être utilisés.
6.5 Gaz pour dégazage.
Utiliser pour le dégazage de l'hélium ou de l'azote de haute qualité, exempt de substances interférentes. Les
impuretés peuvent, si nécessaire, être éliminées à l'aide d'une cartouche de purification.
6.6 Solutions étalons.
Compte tenu de la nature extrêmement volatile des gaz et des composés les plus volatils analysés, la
préparation des solutions étalons doit être effectuée avec le plus grand soin. Des pertes peuvent se produire
dans l'espace de tête du récipient utilisé pour préparer les solutions étalons. Pour une description détaillée de
la préparation des solutions mères étalons de composés volatils, voir l'Annexe F. Il est conseillé et plus
approprié d'utiliser des solutions étalons disponibles dans le commerce. Conserver les solutions étalons
intermédiaires à environ 4 °C et les laisser revenir à température ambiante avant l'emploi.
Le mode opératoire qui suit est donné à titre d'exemple, mais les utilisateurs peuvent souhaiter préparer leurs
propres solutions étalons par une autre méthode ou en diluant des solutions mères disponibles dans le
commerce (de préférence certifiées), dans la mesure où ces dernières donnent des résultats équivalents.
6.6.1 Solution mère étalon pour l'étalonnage (2 mg/ml).
Dissoudre des quantités définies d'environ 200 mg de chaque composé organique volatil dans une fiole
jaugée de 100 ml remplie partiellement du même solvant (6.2), compléter jusqu'au volume et bien
homogénéiser. Voir aussi l'Annexe F.
6.6.2 Solution mère d'étalon interne (2 mg/ml).
Dissoudre des quantités définies d'environ 200 mg de chaque composé étalon interne dans une fiole jaugée
de 100 ml remplie partiellement du même solvant (6.2), compléter jusqu'au volume et bien homogénéiser.
Voir aussi l'Annexe F.
Il convient d'utiliser au moins un composé étalon interne pour l'analyse quantitative et il est possible d'utiliser
des composés étalons internes supplémentaires comme étalons traceurs. Des composés appropriés peuvent
être choisis à partir du Tableau 1. Utiliser des étalons deutérés seulement pour l'analyse par chromatographie
en phase gazeuse suivie d'une spectrométrie de masse. Pour les composés étalons internes indiqués (*)
dans le Tableau 1, la gamme des éléments à analyser correspondants est donnée à titre d'exemple en
Annexe A (Tableau A.2).
© ISO 2003 — Tous droits réservés 5

---------------------- Page: 9 ----------------------
ISO 15680:2003(F)
Tableau 1 — Composés étalons internes
Numéro CAS Composé Formule
462-06-6 * monofluorobenzène C H F
6 5
3114-55-4 * monochlorobenzène-d C CID
5 7 5
3855-82-1 * 1,4-dichlorobenzène-d C Cl D
4 6 2 4
540-36-3 * 1,4-difluorobenzène C H F
6 4 2
460-00-4 1-bromo-4-fluorobenzène C H BrF
6 4
2037-26-5 toluène-d C D
8 7 8
1868-53-7 dibromofluorométhane CHBr F
2
109-70-6 1-bromo-3-chloropropane C H BrCl
3 6
107-04-0 1-bromo-2-chloroéthane C H BrCl
2 4
75-62-7 bromotrichlorométhane CBrCl
3
363-72-4 pentafluorobenzène C HF
6 5
1076-43-3 benzène-d C D
6 6 6
17060-07-0 1,2-dichloroéthane-d C Cl D
4 2 2 4
20302-26-5 éthylbenzène-noyau-d C H D
5 8 5 5
74-97-5 bromochlorométhane CH BrCl
2
3017-95-6 2-bromo-1-chloropropane C H BrCl
3 6
110-56-5 1,4-dichlorobutane C H Cl
4 8 2
56004-61-6 o-xylène-d C D
10 8 10
* La gamme des éléments à analyser correspondants est donnée dans le
Tableau A.2.
6.6.3 Solutions de dopage.
Préparer des solutions de dopage à partir des solutions 6.6.1 et 6.6.2 par dilution appropriée dans une fiole
jaugée contenant le même solvant (6.2). Le Tableau 2 donne, à titre d'exemple, un schéma de dilution dans
100 ml de solvant suivi du dopage de 5 µl de ce solvant dans 100 ml d'eau pour donner les solutions de
dopage 6.6.3.1 à 6.6.3.6. Dans cet exemple, la concentration de l'élément à analyser va de 0 µg/l à 5 µg/l
dans l'eau.
Si la gamme de mesure désirée diffère de celle du Tableau 2, il convient d'adopter des rapports de dilution
différents ou d'adapter le volume de dopage en conséquence.
NOTE La solution 6.6.3.1 est utilisée comme solution étalon interne à ajouter à chaque échantillon (voir 9.3).
Tableau 2 — Schéma de dilution dans 100 ml de solvant
ml de 6.6.2 ml de 6.6.1 Concentration de Concentration (en µg/l) dans
Solution de dopage (ajouté à (ajouté à l'élément à analyser dans la solution d'étalonnage (5 µl
(100 ml de solvant) 100 ml of 100 ml of la solution de dopage de solution de dopage ajoutée
a
solvant) solvant) (en mg/l de solvant) à 100 ml d'eau)
6.6.3.1 5 0 0 0
6.6.3.2 5 1 20 1
6.6.3.3 5 2 40 2
6.6.3.4 5 3 60 3
6.6.3.5 5 4 80 4
6.6.3.6 5 5 100 5
a
La concentration du composé étalon interne de chaque solution de dopage est de 100 mg/l.
6 © ISO 2003 — Tous droits réservés

---------------------- Page: 10 ----------------------
ISO 15680:2003(F)
6.6.4 Solutions d'étalonnage.
Ajouter un petit volume des solutions de dopage de 6.6.3.2 à 6.6.3.6 du Tableau 2 à l'eau (6.1) dans le
récipient pour dégazage (7.1) [ou dans le récipient contenant l'échantillon (7.3) en cas d'utilisation d'un
échantillonneur automatique]. Le Tableau 2 donne l'exemple de l'ajout de 5 µl dans 100 ml d'eau (aux
concentrations indiquées dans la quatrième colonne). Pour analyser des volumes d'échantillon supérieurs,
ajouter un volume supérieur équivalent de solution de dopage.
S'assurer que la teneur en solvant organique de la solution aqueuse d'étalonnage finale ne dépasse pas 2 %
(fraction volumique). Si le solvant est présent en pourcentage supérieur, il convient de vérifier la linéarité.
6.6.5 Solution à blanc.
Réserver une partie de l'eau non dopée pour servir de solution à blanc de contrôle qualité.
7 Appareillage
Il n'est pas spécifié de verrerie ou de matériel courant de laboratoire dans la mesure où le matériel véritable
utilisé dépend de l'application et des circonstances particulières. Vérifier que tous les dispositifs sont exempts
de composés interférents. Nettoyer soigneusement toute la verrerie, y compris les flacons d'échantillonnage.
Un mode opératoire de nettoyage est indiqué en Annexe E.
7.1 Récipients pour dégazage
Toute une variété de récipients pour dégazage est disponible dans le commerce. Le type particulier à adopter
dépend du type d'appareil utilisé pour le dégazage et le piégeage. Certains systèmes permettent le dégazage
dans le récipient d'échantillonnage. Il convient de nettoyer les récipients pour dégazage de la manière
indiquée en Annexe E.
7.2 Récipients d'échantillons
Divers récipients peuvent être utilisés pour contenir les échantillons, par exemple des récipients à bouchon
vissé à joint d'étanchéité en silicone revêtu de PTFE. Pour les systèmes de dégazage et de piégeage avec
échantillonneur automatique, utiliser les récipients d'échantillons préconisés par le fabricant d'échantillonneurs
automatiques. Lorsque des flacons à septum sont utilisés, ceux-ci ne doivent pas être réutilisés.
7.3 Appareillage de dégazage et de piégeage
L'appareillage de dégazage et de piégeage peut être acheté dans le commerce ou construit. Il comprend
l'équipement complètement automatisé de chromatographie en phase gazeuse avec dégazage et piégeage
en ligne, échantillonneur automatique et dispositif de désorption thermique incorporé, ainsi qu'un équipement
manuel hors ligne. Tous les instruments qui respectent les critères et ont prouvé qu'ils donnaient des résultats
fiables peuvent être utilisés. Selon les Annexes A, B et C, divers instruments ont été utilisés. D'autres
dispositifs peuvent être appropriés mais il convient de les vérifier pour s'assurer qu'ils fonctionnent de manière
satisfaisante.
Il convient que l'appareillage de dégazage et de piégeage comporte les éléments suivants:
a) un échantillonneur automatique;
b) un récipient de dégazage, une enveloppe chauffante et un thermostat, une alimentation en gaz pour
dégazage, un régulateur de débit, un chronomètre;
c) un condenseur et une alimentation en réfrigérant ou un système de gaz de purge sec;
d) un piège adsorbant;
© ISO 2003 — Tous droits réservés 7

---------------------- Page: 11 ----------------------
ISO 15680:2003(F)
e) un dispositif de désorption thermique, un thermostat, un chronomètre;
f) un piège froid, une alimentation en réfrigérant, un dispositif de chauffage, un thermostat;
g) un chromatographe en phase gazeuse couplé à un spectromè
...