Analysis of natural gas — Silicon content of biomethane — Part 1: Determination of total silicon by atomic emission spectroscopy (AES)

Analyse du gaz naturel — Teneur en silicium du biométhane — Partie 1: Détermination de la teneur totale en silicium par spectrométrie d’émission atomique (SEA)

General Information

Status
Not Published
Current Stage
6000 - International Standard under publication
Completion Date
01-Apr-2023
Ref Project

Relations

Buy Standard

Draft
ISO/FDIS 2613-1 - Analysis of natural gas — Silicon content of biomethane — Part 1: Determination of total silicon by atomic emission spectroscopy (AES) Released:20. 01. 2023
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview
Draft
REDLINE ISO/FDIS 2613-1 - Analysis of natural gas — Silicon content of biomethane — Part 1: Determination of total silicon by atomic emission spectroscopy (AES) Released:20. 01. 2023
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview
Draft
ISO/FDIS 2613-1 - Analyse du gaz naturel — Teneur en silicium du biométhane — Partie 1: Détermination de la teneur totale en silicium par spectrométrie d’émission atomique (SEA) Released:9. 02. 2023
French language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

FINAL
INTERNATIONAL ISO/FDIS
DRAFT
STANDARD 2613-1
ISO/TC 193/SC 1
Analysis of natural gas — Silicon
Secretariat: NEN
content of biomethane —
Voting begins on:
2023-02-03
Part 1:
Voting terminates on:
Determination of total silicon by
2023-03-31
atomic emission spectroscopy (AES)
Analyse du gaz naturel — Teneur en silicium du biométhane —
Partie 1: Dosage de la teneur totale en silicium par spectrométrie
d'émission atomique (AES)
ISO/CEN PARALLEL PROCESSING
RECIPIENTS OF THIS DRAFT ARE INVITED TO
SUBMIT, WITH THEIR COMMENTS, NOTIFICATION
OF ANY RELEVANT PATENT RIGHTS OF WHICH
THEY ARE AWARE AND TO PROVIDE SUPPOR TING
DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS
Reference number
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/FDIS 2613-1:2023(E)
LOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT INTERNATIONAL STANDARDS MAY ON
OCCASION HAVE TO BE CONSIDERED IN THE
LIGHT OF THEIR POTENTIAL TO BECOME STAN-
DARDS TO WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS. © ISO 2023
---------------------- Page: 1 ----------------------
ISO/FDIS 2613-1:2023(E)
FINAL
INTERNATIONAL ISO/FDIS
DRAFT
STANDARD 2613-1
ISO/TC 193/SC 1
Analysis of natural gas — Silicon
Secretariat: NEN
content of biomethane —
Voting begins on:
Part 1:
Voting terminates on:
Determination of total silicon by
atomic emission spectroscopy (AES)
Analyse du gaz naturel — Teneur en silicium du biométhane —
Partie 1: Dosage de la teneur totale en silicium par spectrométrie
d'émission atomique (AES)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2023
ISO/CEN PARALLEL PROCESSING

All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may

be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on

the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below

or ISO’s member body in the country of the requester.
RECIPIENTS OF THIS DRAFT ARE INVITED TO
ISO copyright office
SUBMIT, WITH THEIR COMMENTS, NOTIFICATION
OF ANY RELEVANT PATENT RIGHTS OF WHICH
CP 401 • Ch. de Blandonnet 8
THEY ARE AWARE AND TO PROVIDE SUPPOR TING
CH-1214 Vernier, Geneva
DOCUMENTATION.
Phone: +41 22 749 01 11
IN ADDITION TO THEIR EVALUATION AS
Reference number
Email: copyright@iso.org
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/FDIS 2613-1:2023(E)
Website: www.iso.org
LOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT INTERNATIONAL STANDARDS MAY ON
Published in Switzerland
OCCASION HAVE TO BE CONSIDERED IN THE
LIGHT OF THEIR POTENTIAL TO BECOME STAN-
DARDS TO WHICH REFERENCE MAY BE MADE IN
© ISO 2023 – All rights reserved
NATIONAL REGULATIONS. © ISO 2023
---------------------- Page: 2 ----------------------
ISO/FDIS 2613-1:2023(E)
Contents Page

Foreword ........................................................................................................................................................................................................................................iv

Introduction .................................................................................................................................................................................................................................v

1 Scope ................................................................................................................................................................................................................................. 1

2 Normative references ..................................................................................................................................................................................... 1

3 Terms and definitions .................................................................................................................................................................................... 1

4 Principle ........................................................................................................................................................................................................................ 2

5 Reagents and labware .................................................................................................................................................................................... 2

6 Apparatus .................................................................................................................................................................................................................... 6

6.1 Sampling and derivatization equipment. ........................................................................................................................ 6

6.2 MWP/ICP-AES instrument. ........................................................................................................................................... ............... 7

6.3 Analytical balance, capable of weighing to the nearest 0,01 mg. .............................................................. 7

7 Sampling ....................................................................................................................................................................................................................... 7

8 Derivatization ......................................................................................................................................................................................................... 9

9 Analytical procedure ...................................................................................................................................................................................... 9

9.1 Set-up of the equipment .................................................................................................................................................................. 9

9.2 Calibration line .................................................................................................................................................................................... 10

9.3 Analysis of unknown and QC samples ............................................................................................................................ 10

10 Calculation ...............................................................................................................................................................................................................10

11 Expression of results ....................................................................................................................................................................................11

12 Precision of the method ............................................................................................................................................................................11

13 Measurement uncertainty ......................................................................................................................................................................11

14 Test report ...............................................................................................................................................................................................................12

Bibliography .............................................................................................................................................................................................................................13

iii
© ISO 2023 – All rights reserved
---------------------- Page: 3 ----------------------
ISO/FDIS 2613-1:2023(E)
Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards

bodies (ISO member bodies). The work of preparing International Standards is normally carried out

through ISO technical committees. Each member body interested in a subject for which a technical

committee has been established has the right to be represented on that committee. International

organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.

ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of

electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are

described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the

different types of ISO documents should be noted. This document was drafted in accordance with the

editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of

any patent rights identified during the development of the document will be in the Introduction and/or

on the ISO list of patent declarations received (see www.iso.org/patents).

Any trade name used in this document is information given for the convenience of users and does not

constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and

expressions related to conformity assessment, as well as information about ISO's adherence to

the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see

www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 193, Natural gas, Subcommittee SC 1,

Analysis of natural gas, in collaboration with the European Committee for Standardization (CEN)

Technical Committee CEN/TC 408, Biomethane for use in transport and injection in natural gas pipelines,

in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna Agreement).

Any feedback or questions on this document should be directed to the user’s national standards body. A

complete listing of these bodies can be found at www.iso.org/members.html.
© ISO 2023 – All rights reserved
---------------------- Page: 4 ----------------------
ISO/FDIS 2613-1:2023(E)
Introduction

This document describes a method for the measurement of the total concentration of silicon in

biomethane, biogas and similar gaseous matrices when used in the natural gas grids and when using it

as a transport fuel. The method is based on using a liquid impinger to accumulate the silicon from a gas

sample, followed by instrumental analysis.

Due to the extensive usage of siloxane compounds, their volatility and great affinity to apolar

environments, siloxanes are considered as one of the most important impurities in biogas. They are

undesired because of their potential for abrasive SiO formation as combustion product that can

damage engines and appliances. Furthermore, some of these compounds present a health risk.

For the purpose of this document, silicon species measured is quoted as total silicon. Silicon measured

is from organosilicon species that are trapped from the gas phase in liquid media and derivatized into

analytical form of hexafluorosilicate (SiF ) ions which remain present in solution when analysed.

© ISO 2023 – All rights reserved
---------------------- Page: 5 ----------------------
FINAL DRAFT INTERNATIONAL STANDARD ISO/FDIS 2613-1:2023(E)
Analysis of natural gas — Silicon content of biomethane —
Part 1:
Determination of total silicon by atomic emission
spectroscopy (AES)
1 Scope

This document is applicable to the measurement of the total silicon content in gaseous matrices such

as biomethane and biogas. Silicon is present in a gas phase contained predominantly in siloxane

compounds, trimethylsilane and trimethylsilanol. The analytical form of the silicon measured in liquid

phase after conducted sampling and derivatization procedure is soluble hexafluorosilicate anion stable

in slightly acidified media. Total silicon is expressed as a mass of silicon in the volume of the analysed

gas.

This document is applicable to stated gaseous matrices with silicon concentrations up to 5 mg/m ,

and it is prevalently intended for the biomethane matrices with Si mass concentration of 0,1 mg/m to

0,5 mg/m .

With adaptation to ensure appropriate absorption efficiency, it can be used for higher concentrations.

3 3

The detection limit of the method is estimated as 0,05 mg/m based on a gas sample volume of 0,020 m .

All compounds present in the gas phase are volatile at the absorption and derivatization temperature

and gaseous organosilicon species are trapped in absorbance media and derivatized into analytical

silicon that is measured by this method. The concentration of the silicon is measured in diluted

derivatization media using atomic emission spectrometry upon atomisation/ionisation in microwave

or inductively coupled plasma.

Unless specified otherwise, all volumes and concentrations refer to standard reference conditions

(temperature, 273 K, and pressure, 101,325 kPa).

NOTE When using appropriate dilution factors, the method can also be applied for silicon concentrations

above 5 mg/m .
2 Normative references

The following documents are referred to in the text in such a way that some or all of their content

constitutes requirements of this document. For dated references, only the edition cited applies. For

undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 3696, Water for analytical laboratory use — Specification and test methods

ISO 6143, Gas analysis — Comparison methods for determining and checking the composition of calibration

gas mixtures
ISO 14532, Natural gas — Vocabulary
ISO 10715, Natural gas — Gas sampling
ISO 14912, Gas analysis — Conversion of gas mixture composition data
3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 14532 and the following apply.

© ISO 2023 – All rights reserved
---------------------- Page: 6 ----------------------
ISO/FDIS 2613-1:2023(E)

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
siloxane
functional group where two silicon atoms are connected via an oxygen atom

Note 1 to entry: Depending on the substrate used to produce biogas and the process used for purification,

biomethane can contain siloxanes. During combustion, siloxanes can be oxidized to silicon dioxide, an abrasive

compound harmful for mechanical moving parts in e.g. engines and turbines.
3.2
atomic emission spectroscopy
AES

method of chemical analysis that uses the intensity of light emitted from a flame, plasma, arc, or spark

at a particular wavelength to determine the quantity of an element in a sample
4 Principle

WARNING — Persons using this document should be familiar with normal laboratory practice.

This standard does not purport to address all of the safety problems, if any, associated with its

use. It is the responsibility of the user to establish appropriate safety and health practices and to

ensure compliance with any national regulatory conditions.

A methane matrix gas sample (e.g. biomethane, biogas, natural gas and blends thereof) containing

siloxane compounds is passed through liquid absorbent (nitric acid) in serially connected gas bubblers/

impingers to collect the silicon-containing compounds. After sampling of an adequate gas volume,

content of sampling vessels (gas bubblers) is subjected to derivatization by adding hydroxide solution

and hydrofluoric acid in order to obtain silicon in analytical from, hexafluorosilicate (SiF ) anion.

The derivatized sample is analysed for silicon content using an ICP/MWP atomic emission spectrometer

at selected characteristic silicon emission wavelengths using a multipoint calibration using a straight

line obtained from analysing a series of standard silicon solutions.
5 Reagents and labware

To carry out the method, the following reagents shall be of a recognized analytical grade and only

ISO 3696 grade 1 water. If it is visually determined that the reagents have changed their appearance

(colour, consistency, turbidity) they shall be discarded, and fresh ones shall be used.

5.1 Absorber media.

5.1.1 Nitric acid (HNO ), ρ = 1,41 g/ml; 65 % HNO (mass fraction) – for trace elemental analysis.

3 20 3

CAUTION — This chemical is especially dangerous if used outside specialized laboratory

conditions. Tests have been performed in which other non-oxo mineral acids (HCl) have been

used, but they have been shown to be inadequate for the absorption of siloxanes from the

gas phase. Special precautions are to be taken when handling this chemical in lab and field

conditions.
© ISO 2023 – All rights reserved
---------------------- Page: 7 ----------------------
ISO/FDIS 2613-1:2023(E)
5.2 Derivatization media.

5.2.1 Sodium hydroxide pellets, for the preparation of 8 mol/l – 10 mol/l hydroxide solution.

Accurately weigh an appropriate amount of sodium hydroxide pellets and dissolve these in an

appropriate amount of reagent water (5.3). As an example for 100 ml of 10 mol/l sodium hydroxide

solutions, weigh 40 g of sodium hydroxide pellets and dissolve in 100 ml water.

Potassium hydroxide can also be used, but sodium hydroxide is preferred due to operation safety.

WARNING — Reaction of dissolving sodium hydroxide in water is highly exothermic! Heat will be

released and care should be taken when handling the reaction. Add pellets slowly to the water

and cool the dissolution vessel until the dissolution is complete.
5.2.2 Hydrofluoric acid (HF), ρ = 1,16 g/ml; 48 % HF (mass fraction).
(20 ˚C)

WARNING — Hydrofluoric acid is a very toxic acid and penetrates the skin and tissues deeply

if not treated immediately. Injury occurs in two stages: firstly, by hydration that induces tissue

necrosis; and secondly, by penetration of fluoride ions deep into the tissue and thereby reacting

with calcium. Boric acid and/or other complexing reagents and appropriate treatment agents

should be administered immediately. Consult appropriate safety literature for determining the

proper protective eyewear, clothing and gloves to use when handling hydrofluoric acid. Always

have appropriate treatment materials readily available prior to working with this acid.

CAUTION — This chemical is especially dangerous if used outside specialized laboratory

conditions. Tests have been performed in which other fluoride donor derivatization reagents

(NaF) have been used, but they have been shown to be inadequate for the derivatization of

absorbed siloxanes from the gas phase. Special precautions are to be taken when handling this

chemical in lab and field conditions
5.3 Water, complying with grade 1 of ISO 3696.
5.4 Pure siloxane compounds.
Linear siloxanes Molecular formula Cyclic siloxanes Molecular formula
Hexamethyldisiloxane – L2 C H OSi Hexamethylcyclotrisiloxane - D3 C H O Si
6 18 2 6 18 3 3
Octamethyltrisiloxane – L3 C H O Si Octamethylcyclotetrasiloxane C H O Si
8 24 2 3 8 24 4 4
- D4
Decamethyltetrasiloxane C H O Si Decamethylcyclopentasiloxane C H O Si
10 30 3 4 10 30 5 5
– L4 - D5
Dodecamethylpentasiloxane C H O Si Dodecamethylcyclohexasilox- C H O Si
12 36 4 5 12 36 6 6
– L5 ane - D6
— L2;
— L3;
— L4;
— L5;

Use at least one representative of chain and one representative of cyclic siloxane compounds for the

purpose of performing initial and regular quality control of the method validity.

5.5 pH colour-fixed indicator strips, pH range from 0 to 14, or, alternatively, a pH meter with HF

resistant electrode.
© ISO 2023 – All rights reserved
---------------------- Page: 8 ----------------------
ISO/FDIS 2613-1:2023(E)
5.6 Calibration solutions.
5.6.1 General

The following procedure for the preparation of standard and calibration solutions of silicon is adjusted

to the lower range of silicon concentration in gas sample. If higher concentrations of silicon shall be

measured, adjust the concentrations of the working standard and calibration solutions accordingly.

When determining silicon in aqueous samples, only plastic, PTFE or quartz labware shall be used from

time of sample collection to completion of analysis.
5.6.2 Certified ICP-Si stock standard solution.

Example of certified Si standard solution is water solution (only trace level of HF is acceptable) with Si

mass concentration of 10 000 μg/ml and relative expanded uncertainty (coverage factor k = 2)0,5 %.

This concentration is used in the example of the Si standard solution preparation in 5.6.3.

Certified Si standard solutions of other concentrations can also be used. Adjust the procedure for

preparing standard solution accordingly.

If Si stock standard solution is prepared in-house gravimetrically from salt-containing silicon, apply

required statistical procedure for obtaining accurate concentration accompanied with uncertainty

value.
NOTE References [1][2] provide guidance.
5.6.3 Si standard solution.

The target Si mass concentration ρ(Si) ≈ 100 mg/kg.Weigh empty 50 ml plastic volumetric flask using

analytical balance (6.3). Add around 10 ml of 2 % nitric acid (mass fraction). Accurately pipette 0,5 ml

of stock solution (5.6.2) and add it to the plastic volumetric flask. Dilute with 2 % nitric acid (mass

fraction) to volume. Weigh full plastic volumetric flask and calculate the concentration of silicon.

Store the solution in plastic volumetric flask or similar vessel of silicon free material properly stoppered

at room temperature or refrigerated (~5 °C). The solution is stable for at least two weeks if stored

properly.
5.6.4 Si calibration solutions.

Gravimetrically prepare a minimum of five calibration solutions in accordance with expected silicon

concentration in the collected sample.

As an example proceed as follows for the Si mass concentration range from 10 μg/kg to 200 μg/kg.

Weigh empty 100 ml (or 200 ml) plastic volumetric flasks.

Pipette 10 μl; 20 μl; 50 μl; 75 μl; 100 μl; 150 μl; and 200 μl; respectively of silicon standard solution

(5.6.3) into 100 ml one-mark plastic volumetric flask that was empty-weighted and prefilled with

around 10 ml - 20 ml of 2 % nitric acid (mass fraction). Dilute with 2 % nitric acid (mass fraction) to

volume. Weigh full plastic volumetric flask and calculate the concentration of silicon.

The Si mass concentration in the calibration solutions is 10 μg/kg; 20 μg/kg; 50 μg/kg; 75 μg/kg;

100 μg/kg; 150 μg/kg and 200 μg/kg respectively.

Calculate the uncertainty of the mass concentrations of the calibration solutions. Check what the

contributions are of the combination of shared effects, such as the uncertainty of the concentration

of the stock solution and the calibration of the pipette. If these effects account for more than 40 % of

the uncertainty budget, then calculate the dilution factors and their associated uncertainties. Then the

concentrations of these standards are substantially correlated.
© ISO 2023 – All rights reserved
---------------------- Page: 9 ----------------------
ISO/FDIS 2613-1:2023(E)
5.6.5 Solution for wavelength calibration control.

Perform wavelength check using solution containing assorted elements covering the wavelength range

of the instrumentation used provided by the manufacturer prior to daily calibration of the instrument

for the analysis of silicon. This solution is usually provided as concentrate that needs to be diluted prior

to the analysis in accordance with the manufacturer’s instructions. Wavelength calibration control test

result shows if the optical settings of the instrument are appropriate, and if the readings of the emission

lines for each individual element correspond to the instrumental settings when selecting the analytical

wavelengths for the analyte of interest.

NOTE The solution for wavelength calibration control is usually provided by the manufacturer of the

equipment.
5.7 Quality control.
5.7.1 Blanks.

Three types of blanks are used during the analysis. The calibration blank is used in establishing the

analytical curve, the laboratory reagent blank is used to assess any contamination from the sample

preparation procedure and a rinse blank is used to flush the instrument uptake system and nebulizer

between standards, check solutions, and samples to reduce memory interferences.

5.7.1.1 The calibration blank is prepared by acidifying reagent water to the same concentrations

of the acids as used for the standards; in this case it is 2 % nitric acid (mass fraction). The calibration

blank should be stored in a plastic container as samples.

5.7.1.2 The laboratory reagent blank should contain all the reagents in the same volumes as used

in the processing of the samples. The laboratory reagent blank shall be carried through the same entire

preparation scheme as the samples including sample derivatization. This type of blank should be

prepared at least every time new reagents are used.

5.7.1.3 The rinse blank is prepared by acidifying reagent water to the same concentrations of nitric

acid as used in the calibration blank and stored in a convenient manner.

5.7.1.4 Labware blank – pure methane gas free from silicon used as blank gas sample to test the

cleanliness of labware used.

5.7.2 Instrument performance check i.e. wavelength calibration control sample (5.6.5).

5.7.3 Calibration Control Sample (CC).

A calibration control sample shall be used for initial and periodic verification of calibration standards

or stock standard solutions in order to verify instrument performance. The CC shall be obtained from

an outside source different from the standard stock solutions and prepared in the same acid mixture as

the calibration standards. It can be either ready standard solution obtained from a different supplier,

or at least from a different lot, or it can be prepared gravimetrically using pure (NH ) SiF salt. The

4 2 6

concentration of the silicon in the CC solution should be near to expected concentration of silicon in the

sample or at the middle of calibration range. A fresh solution should be prepared prior to the analysis

and stored in plastic container as samples.
5.7.4 Derivatization control sample (DC).

A derivatization control sample shall be used for initial and periodic verification of the completeness

of the derivatization process. For this purpose pure siloxane compounds are used. For example, L2

and D4 siloxanes represent linear and cyclic siloxanes found in biomethane matrices. Other siloxanes

may be used as well. The DC is prepared by accurately pipetting appropriate amount of siloxane with

previously calculated mass of silicon contained, and adding this amount to the aliquot of nitric acid thus

© ISO 2023 – All rights reserved
---------------------- Page: 10 ----------------------
ISO/FDIS 2613-1:2023(E)

simulating the absorbance procedure. The solution of siloxane(s) is then subjected to derivatization by

adding appropriate amount of hydroxide solution and hydrofluoric sample. The DC should be stored in a

plastic container as sample. Concentration of the silicon in DC shall be within the calibration range and

can be adjusted by dilution if needed.

5.7.5 Reference gas mixture of siloxanes in methane with certified silicon content, over the

3 3
mass concentration range of 0,1 mg/m to 0,5 mg/m .

Certified reference gas mixtures containing different siloxanes and combinations of siloxanes in

methane are available with certified siloxane amount fractions. These amount fractions can be

converted to a total silicon concentration, but it should be noted that the total silicon in the mixture

may differ, i.e. be higher due to siloxane impurities that are present as non-certified siloxanes. Such gas

mixtures are suitable for assessing the recovery of the sampling and derivatization.

ISO 14912 shall be used for the conversion of amount fractions to concentrations, including the

associated uncertainties.
6 Apparatus
6.1 Sampling and derivatization equipment.

An outline of the equipment for the sampling of gas is given in Figure 1. The apparatus consists of a gas

flow meter and an impinger train containing absorbent (concentrated nitric acid) to capture gaseous

siloxanes. A thermometer shall be used if the laboratory has no controlled ambient temperature within

±3 °C. If gas flow meter used is not equipped with the embedded ambient pressure sensor providing

data for normalization to standard reference conditions of 273,15 K and 101,325 kPa, a barometer shall

be used to measure atmospheric pressure during collection of the gas. Using the measured temperature

and pressure, volumes and concentrations shall be converted to appropriate standard reference

conditions.

All tubing, gaskets and seals used to for passing of the sample gas, as well as the impingers and

derivatization vessels and stirring rod shall be made of plastic polymer silicon free.

The sampling and sample derivatization described in this document refers to the laboratory equipment

and conditions. Specialized sampling equipment may be used that allows the absorption of siloxane

from the gaseous medium in the field if the described requirements are met.

Field sampling/absorption and derivatization were not covered by the study during the development of

this standard. In the case of the development of equipment that enables field sampling and derivatization,

it shall be validated in terms of applicability and minimize and avoid any losses. To generate the best

results, it is recommended to perform sampling/absorption and derivatization procedures in the

laboratory.
6.1.1 Gas flo
...

Style Definition: Heading 1: Indent: Left: 0 pt, First line:
ISO/FDIS 2613-1:20222023(E)
0 pt, Tab stops: Not at 21.6 pt
Style Definition: Heading 2: Font: Bold, Tab stops: Not at
Date: 2023-xx
18 pt
Style Definition: Heading 3: Font: Bold
ISO/TC 193/SC 1/WG 25
Style Definition: Heading 4: Font: Bold
Date: 2022-12-28
Style Definition: Heading 5: Font: Bold
Style Definition: Heading 6: Font: Bold
Secretariat: NEN
Style Definition: ANNEX

Analysis of natural gas — Silicon content of biomethane — Part 1: Determination Style Definition: AMEND Terms Heading: Font: Bold

of total silicon by AESatomic emission spectroscopy (AES) Style Definition: AMEND Heading 1 Unnumbered: Font:

Bold

Analyse du gaz naturel — Teneur en silicium du biométhane — Partie 1: Détermination Formatted: Font: Not Bold, Italic

de la teneur totale en silicium par AESspectrométrie d'émission atomique (AES)
Formatted: Font: Not Bold, Italic
---------------------- Page: 1 ----------------------
ISO/FDIS 2613-1:20222023(E)
COPYRIGHT PROTECTED DOCUMENT

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or

utilized otherwise in any form or by any means, electronic or mechanical, including photocopying,

or posting on the internet or an intranet, without prior written permission. Permission can be

requested from either ISO at the address below or ISO’s member body in the country of the

requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
copyright@iso.org
www.iso.org
www.iso.org
ii © ISO 20222023 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/FDIS 2613-1:20222023(E)
Contents Page

Foreword ..................................................................................................................................................................... iv

Introduction v

1 Scope ................................................................................................................................................................ 1

2 Normative references ................................................................................................................................ 1

3 Terms and definitions ................................................................................................................................ 1

4 Principle ......................................................................................................................................................... 2

5 Reagents and labware ................................................................................................................................ 2

5.1 Absorber media............................................................................................................................................ 2

5.2 Derivatization media .................................................................................................................................. 3

5.3 Water, complying with grade 1 of ISO 3696 ........................................................................................ 3

5.4 Pure siloxane compounds: ....................................................................................................................... 3

5.5 pH colour-fixed indicator strips, pH range from 0 - 14, or, alternatively, a pH meter

with HF resistant electrode ...................................................................................................................... 4

5.6 Calibration solutions .................................................................................................................................. 4

5.7 Quality control .............................................................................................................................................. 5

6 Apparatus ....................................................................................................................................................... 6

6.1 Sampling and derivatization equipment ............................................................................................. 6

6.2 MWP/ICP-AES instrument ........................................................................................................................ 7

6.3 Analytical balance accurate to 0,01 mg. ............................................................................................... 7

7 Sampling ......................................................................................................................................................... 7

8 Derivatization ............................................................................................................................................... 9

9 Analytical procedure .................................................................................................................................. 9

9.1 Set-up of the equipment ............................................................................................................................ 9

9.2 Calibration line ........................................................................................................................................... 10

9.3 Analysis of unknown and QC samples ................................................................................................. 10

10 Calculation ................................................................................................................................................... 10

11 Expression of results ................................................................................................................................ 11

12 Precision of the method .......................................................................................................................... 11

13 Measurement uncertainty ...................................................................................................................... 11

14 Test report ................................................................................................................................................... 12

Bibliography.............................................................................................................................................................. 13

1 Scope ................................................................................................................................................................ 1

2 Normative references ................................................................................................................................ 1

© ISO 20222023 – All rights reserved iii
---------------------- Page: 3 ----------------------
ISO/FDIS 2613-1:20222023(E)

3 Terms and definitions ................................................................................................................................ 1

4 Principle ......................................................................................................................................................... 2

5 Reagents and labware ................................................................................................................................ 2

5.1 Absorber media ............................................................................................................................................ 2

5.2 Derivatization media .................................................................................................................................. 3

5.3 Water, complying with grade 1 of ISO 3696 ........................................................................................ 3

5.4 Pure siloxane compounds: ........................................................................................................................ 3

5.5 pH colour-fixed indicator strips, pH range from 0 - 14, or, alternatively, a pH meter

with HF resistant electrode ...................................................................................................................... 4

5.6 Calibration solutions .................................................................................................................................. 4

5.7 Quality control .............................................................................................................................................. 5

6 Apparatus ....................................................................................................................................................... 6

6.1 Sampling and derivatization equipment .............................................................................................. 6

6.2 MWP/ICP-AES instrument ........................................................................................................................ 7

6.3 Analytical balance accurate to 0,01 mg. ............................................................................................... 7

7 Sampling ......................................................................................................................................................... 7

8 Derivatization ............................................................................................................................................... 9

9 Analytical procedure .................................................................................................................................. 9

9.1 Set-up of the equipment ............................................................................................................................ 9

9.2 Calibration line ........................................................................................................................................... 10

9.3 Analysis of unknown and QC samples ................................................................................................. 10

10 Calculation ................................................................................................................................................... 10

11 Expression of results ................................................................................................................................ 11

12 Precision of the method ........................................................................................................................... 11

13 Measurement uncertainty ...................................................................................................................... 11

14 Test report ................................................................................................................................................... 12

Bibliography .............................................................................................................................................................. 13

iv © ISO 20222023 – All rights reserved
---------------------- Page: 4 ----------------------
ISO/FDIS 2613-1:20222023(E)
Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national Formatted: English (United States)

standards bodies (ISO member bodies). The work of preparing documentsInternational Standards is

Formatted: Adjust space between Latin and Asian text,

normally carried out through ISO technical committees. Each member body interested in a subject for

Adjust space between Asian text and numbers

which a technical committee has been established has the right to be represented on that committee.

Formatted: English (United States)

International organizations, governmental and non-governmental, in liaison with ISO, also take part in

the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all

matters of electrotechnical standardization. Formatted: English (United States)

The procedures used to develop this document and those intended for its further maintenance are

described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the Formatted: English (United States)

different types of ISO documents should be noted. This document was drafted in accordance with the

editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives 2 (see

www.iso.org/directives). Formatted: English (United States)

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of

any patent rights identified during the development of the document will be in the Introduction and/or

on the ISO list of patent declarations received (see www.iso.org/patentswww.iso.org/patents). Formatted: English (United States)

Any trade name used in this document is information given for the convenience of users and does not

constitute an endorsement. Formatted: English (United States)

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and

expressions related to conformity assessment, as well as information about ISO's adherence to the

World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see

www.iso.org/iso/foreword.htmlwww.iso.org/iso/foreword.html. Formatted: English (United States)

This document was prepared by Technical Committee ISO/TC 193, Natural gas, Subcommittee SC 1,

Analysis of natural gas, in collaboration with the European Committee for Standardization (CEN)

Technical Committee CEN/TC 408, Biomethane for use in transport and injection in natural gas pipelines,

in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna Agreement).

Any feedback or questions on this document should be directed to the user’s national standards body. A

complete listing of these bodies can be found at
www.iso.org/members.htmlwww.iso.org/members.html.
© ISO 20222023 – All rights reserved v
---------------------- Page: 5 ----------------------
ISO/FDIS 2613-1:20222023(E)
Introduction

This document describes a method for the measurement of the total concentration of silicon in

biomethane, biogas and similar gaseous matrices when used in the natural gas grids and when using it

as a transport fuel. The method is based on using a liquid impinger to accumulate the silicon from a gas

sample, followed by instrumental analysis.

Due to the extensive usage of siloxane compounds, their volatility and great affinity to apolar

environments, siloxanes are considered as one of the most important impurities in biogas. They are

undesired because of their potential for abrasive SiO formation as combustion product that can

damage engines and appliances. Furthermore, some of these compounds present a health risk.

For the purpose of this document, silicon species measured is quoted as total silicon. Silicon measured

is from organosilicon species that are trapped from the gas phase in liquid media and derivatized into

analytical form of hexafluorosilicate (SiF ) ions which remain present in solution when analysed.

vi © ISO 20222023 – All rights reserved
---------------------- Page: 6 ----------------------
FINAL DRAFT INTERNATIONAL STANDARD ISO/FDIS 2613-1:20222023(E)
Formatted: Justified
Analysis of natural gas — Silicon content of biomethane —
Part 1: Determination of total silicon by AESatomic emission
spectroscopy (AES)
1 Scope

This document is applicable to the measurement of the total silicon content in gaseous matrices such as

biomethane and biogas. Silicon is present in a gas phase contained predominantly in siloxane

compounds, trimethylsilane and trimethylsilanol. The analytical form of the silicon measured in liquid

phase after conducted sampling and derivatization procedure is soluble hexafluorosilicate anion stable

in slightly acidified media. Total silicon is expressed as a mass of silicon in the volume of the analysed

gas.

This document is applicable to stated gaseous matrices with silicon concentrations up to 5 mg/m , and

it is prevalently intended for the biomethane matrices with Si mass concentration of 0,1 mg/m to

0,5 mg/m .

With adaptation to ensure appropriate absorption efficiency, it can be used for higher concentrations.

The detection limit of the method is estimated as 0,05 mg/m based on a gas sample volume of

0,020 m . All compounds present in the gas phase are volatile at the absorption and derivatization

temperature and gaseous organosilicon species are trapped in absorbance media and derivatized into

analytical silicon that is measured by this method. The concentration of the silicon is measured in

diluted derivatization media using atomic emission spectrometry upon atomisation/ionisation in

microwave or inductively coupled plasma.

Unless specified otherwise, all volumes and concentrations refer to standard reference conditions

(temperature (, 273 K), and pressure (, 101,325 kPa).)).

NOTE When using appropriate dilution factors, the method can also be applied for silicon concentrations above

5 mg/m .
2 Normative references

The following documents are referred to in the text in such a way that some or all of their content

constitutes requirements of this document. For dated references, only the edition cited applies. For

undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 3696, Water for analytical laboratory use — Specification and test methods

ISO 6143, Gas analysis — Comparison methods for determining and checking the composition of

calibration gas mixtures
ISO 14532, Natural gas — Vocabulary
ISO 10715, Natural gas — Gas sampling
ISO 14912, Gas analysis — Conversion of gas mixture composition data
ISO 3696, Water for analytical laboratory use — Specification and test methods
© ISO 20222023 – All rights reserved 1
---------------------- Page: 7 ----------------------
ISO/FDIS 2613-1:20222023(E)

ISO 6143, Gas analysis — Comparison methods for determining and checking the composition of

calibration gas mixtures
ISO 14532, Natural gas — Vocabulary
ISO 10715, Natural gas — Gas sampling
ISO 14912, Gas analysis — Conversion of gas mixture composition data
3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 14532 and the following

Formatted: Pattern: Clear
apply.
Formatted: Pattern: Clear

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https://www.iso.org/obphttps://www.iso.org/obp Formatted: English (United States)

Formatted: Adjust space between Latin and Asian text,

— IEC Electropedia: available at https://www.electropedia.org/https://www.electropedia.org/

Adjust space between Asian text and numbers, Tab stops:
Not at 19.85 pt + 39.7 pt + 59.55 pt + 79.4 pt + 99.25
pt + 119.05 pt + 138.9 pt + 158.75 pt + 178.6 pt +
3.1
198.45 pt
siloxane
Formatted: Hyperlink, English (United States)
functional group where two silicon atoms are connected via an oxygen atom
Formatted: English (United States)

Note 1 to entry: Depending on the substrate used to produce biogas and the process used for purification,

Formatted: Hyperlink, English (United States)

biomethane can contain siloxanes. During combustion, siloxanes can be oxidized to silicon dioxide, an abrasive

[8]
compound harmful for mechanical moving parts in e.g. engines and turbines .
3.2
AES
atomic emission spectroscopy
AES

method of chemical analysis that uses the intensity of light emitted from a flame, plasma, arc, or spark

at a particular wavelength to determine the quantity of an element in a sample
4 Principle

WARNING — Persons using this document should be familiar with normal laboratory practice. This

standard does not purport to address all of the safety problems, if any, associated with its use. It is the

responsibility of the user to establish appropriate safety and health practices and to ensure compliance

with any national regulatory conditions.

A methane matrix gas sample (e.g.,. biomethane, biogas, natural gas and blends thereof) containing

siloxane compounds is passed through liquid absorbent (nitric acid) in serially connected gas

bubblers/impingers to collect the silicon-containing compounds. After sampling of an adequate gas

volume, content of sampling vessels (gas bubblers) is subjected to derivatization by adding hydroxide

solution and hydrofluoric acid in order to obtain silicon in analytical from, hexafluorosilicate (SiF6 )

anion.

The derivatized sample is analysed for silicon content using an ICP/MWP atomic emission spectrometer

at selected characteristic silicon emission wavelengths using a multipoint calibration using a straight

line obtained from analysing a series of standard silicon solutions.
2 © ISO 20222023 – All rights reserved
---------------------- Page: 8 ----------------------
ISO/FDIS 2613-1:20222023(E)
5 Reagents and labware

To carry out the method, the following reagents shall be of a recognized analytical grade and only

ISO 3696 grade 1 water. If it is visually determined that the reagents have changed their appearance

Formatted: Pattern: Clear

(colour, consistency, turbidity) they shall be discarded, and fresh ones shall be used.

Formatted: Pattern: Clear
5.1 Absorber media.

5.1.1 Nitric acid (HNO ), ρ = 1,41 g/ml; 65 % HNO (mass fraction) – for trace elemental analysis.

3 20 3 Formatted: Font: Not Bold
Formatted: Font: Bold

CAUTION This chemical is especially dangerous if used outside specialized laboratory conditions.

Formatted: p3

Tests have been performed in which other non-oxo mineral acids (HCl) have been used, but they have

been shown to be inadequate for the absorption of siloxanes from the gas phase. Special precautions are

to be taken when handling this chemical in lab and field conditions.
5.2 Derivatization media.

5.2.1 Sodium hydroxide pellets, for the preparation of 8 mol/l – 10 mol/l hydroxide solution.

Accurately weigh an appropriate amount of sodium hydroxide pellets and dissolve these in an

appropriate amount of reagent water (5.3). As an example for 100 ml of 10 mol/l sodium hydroxide

Formatted: Pattern: Clear
solutions, weigh 40 g of sodium hydroxide pellets and dissolve in 100 ml water.

Potassium hydroxide can also be used, but sodium hydroxide is preferred due to operation safety.

WARNING — Reaction of dissolving sodium hydroxide in water is highly exothermic! Heat will be

released and care should be taken when handling the reaction. Add pellets slowly to the water and cool

the dissolution vessel until the dissolution is complete.
5.2.2 Hydrofluoric acid (HF), ρ(20 ˚C) = 1,16 g/ml; 48 % HF (mass fraction).

WARNING — Hydrofluoric acid is a very toxic acid and penetrates the skin and tissues deeply if not

treated immediately. Injury occurs in two stages: firstly, by hydration that induces tissue necrosis; and

secondly, by penetration of fluoride ions deep into the tissue and thereby reacting with calcium. Boric

acid and/or other complexing reagents and appropriate treatment agents should be administered

immediately. Consult appropriate safety literature for determining the proper protective eyewear,

clothing and gloves to use when handling hydrofluoric acid. Always have appropriate treatment

materials readily available prior to working with this acid.

CAUTION This chemical is especially dangerous if used outside specialized laboratory conditions.

Tests have been performed in which other fluoride donor derivatization reagents (NaF) have been used,

but they have been shown to be inadequate for the derivatization of absorbed siloxanes from the gas

phase. Special precautions are to be taken when handling this chemical in lab and field conditions

5.3 Water, complying with grade 1 of ISO 3696. Formatted: Pattern: Clear
Formatted: Pattern: Clear
5.4 Pure siloxane compounds:.
Formatted: Font: Bold
Linear siloxanes Molecular formula Cyclic siloxanes Molecular formula
Formatted: Centered

Hexamethyldisiloxane – L2 C6H18OSi2 Hexamethylcyclotrisiloxane - D3 C6H18O3Si3 Formatted: Centered

Formatted: Centered
Octamethyltrisiloxane – L3 C H O Si Octamethylcyclotetrasiloxane - C H O Si
8 24 2 3 8 24 4 4
Formatted: Centered
Formatted: Centered
Decamethyltetrasiloxane – C10H30O3Si4 Decamethylcyclopentasiloxane - C10H30O5Si5
L4 D5
Formatted: Centered
Formatted: Centered
© ISO 20222023 – All rights reserved 3
---------------------- Page: 9 ----------------------
ISO/FDIS 2613-1:20222023(E)
Dodecamethylpentasiloxane C H O Si Dodecamethylcyclohexasiloxane C H O Si
12 36 4 5 12 36 6 6 Formatted: Centered
– L5 - D6
Formatted: Centered
— L2;
— L3;
— L4;
— L5;

Use at least one representative of chain and one representative of cyclic siloxane compounds for the

purpose of performing initial and regular quality control of the method validity.

5.5 pH colour-fixed indicator strips, pH range from 0 to 14, or, alternatively, a pH meter with HF

resistant electrode.
5.6 Calibration solutions.
5.6.1 General

The following procedure for the preparation of standard and calibration solutions of silicon is adjusted

to the lower range of silicon concentration in gas sample. If higher concentrations of silicon shall be

measured, adjust the concentrations of the working standard and calibration solutions accordingly.

When determining silicon in aqueous samples, only plastic, PTFE or quartz labware shall be used from

time of sample collection to completion of analysis.
5.6.2 Certified ICP-Si stock standard solution.

Example of certified Si standard solution is water solution (only trace level of HF is acceptable) with Si

mass concentration of 10 000 μg/ml and relative expanded uncertainty (coverage factor k = 2)0,5 %. Formatted: Font: Italic

This concentration is used in the example of the Si standard solution preparation in 5.6.3.

Formatted: Pattern: Clear

Certified Si standard solutions of other concentrations can also be used. Adjust the procedure for

preparing standard solution accordingly.

If Si stock standard solution is prepared in-house gravimetrically from salt-containing silicon, apply

required statistical procedure for obtaining accurate concentration accompanied with uncertainty

value.
NOTE References [1][2] provide guidance.
Formatted: Pattern: Clear
Formatted: Pattern: Clear
5.6.3 Si standard solution.

The target Si mass concentration ρ(Si) ≈ 100 mg/kg.Weigh empty 50 ml plastic volumetric flask using Formatted: Font: Italic

analytical balance (6.3). Add around 10 ml of 2 % nitric acid (mass fraction). Accurately pipette 0,5 ml

Formatted: Pattern: Clear

of stock solution (5.6.2) and add it to the plastic volumetric flask. Dilute with 2 % nitric acid (mass

Formatted: Pattern: Clear

fraction) to volume. Weigh full plastic volumetric flask and calculate the concentration of silicon.

Store the solution in plastic volumetric flask or similar vessel of silicon free material properly stoppered

at room temperature or refrigerated (~5 °C). The solution is stable for at least two weeks if stored

properly.
5.6.4 Si calibration solutions.

Gravimetrically prepare a minimum of five calibration solutions in accordance with expected silicon

concentration in the collected sample.

As an example proceed as follows for the Si mass concentration range from 10 μg/kg to 200 μg/kg.

4 © ISO 20222023 – All rights reserved
---------------------- Page: 10 ----------------------
ISO/FDIS 2613-1:20222023(E)
Weigh empty 100 ml (or 200 ml) plastic volumetric flasks.

Pipette 10 μl; 20 μl; 50 μl; 75 μl; 100 μl; 150 μl; and 200 μl; respectively of silicon standard solution

(5.6.3) into 100 ml one-mark plastic volumetric flask that was empty-weighted and prefilled with Formatted: Pattern: Clear

around 10 ml - 20 ml of 2 % nitric acid (mass fraction). Dilute with 2 % nitric acid (mass fraction) to

volume. Weigh full plastic volumetric flask and calculate the concentration of silicon.

The Si mass concentration in the calibration solutions is 10 μg/kg; 20 μg/kg; 50 μg/kg; 75 μg/kg;

100 μg/kg; 150 μg/kg and 200 μg/kg respectively.

Calculate the uncertainty of the mass concentrations of the calibration solutions. Check what the

contributions are of the combination of shared effects, such as the uncertainty of the concentration of

the stock solution and the calibration of the pipette. If these effects account for more than 40 % of the

uncertainty budget, then calculate the dilution factors and their associated uncertainties. Then the

concentrations of these standards are substantially correlated.
5.6.5 Solution for wavelength calibration control.

Perform wavelength check using solution containing assorted elements covering the wavelength range

of the instrumentation used provided by the manufacturer prior to daily calibration of the instrument

for the analysis of silicon. This solution is usually provided as concentrate that needs to be diluted prior

to the analysis in accordance with the manufacturer’s instructions. Wavelength calibration control test

result shows if the optical settings of the instrument are appropriate, and if the readings of the emission

lines for each individual element correspond to the instrumental settings when selecting the analytical

wavelengths for the analyte of interest.

NOTE The solution for wavelength calibration control is usually provided by the manufacturer of the

equipment.
5.7 Quality control.
5.7.1 Blanks.

Three types of blanks are used during the analysis. The calibration blank is used in esta

...

PROJET
NORME ISO/FDIS
FINAL
INTERNATIONALE 2613-1
ISO/TC 193/SC 1
Analyse du gaz naturel — Teneur en
Secrétariat: NEN
silicium du biométhane —
Début de vote:
2023-02-03
Partie 1:
Vote clos le:
Détermination de la teneur totale en
2023-03-31
silicium par spectrométrie d’émission
atomique (SEA)
Analysis of natural gas — Silicon content of biomethane —
Part 1: Determination of total silicon by atomic emission spectroscopy
(AES)
TRAITEMENT PARALLÈLE ISO/CEN
LES DESTINATAIRES DU PRÉSENT PROJET SONT
INVITÉS À PRÉSENTER, AVEC LEURS OBSER-
VATIONS, NOTIFICATION DES DROITS DE PRO-
PRIÉTÉ DONT ILS AURAIENT ÉVENTUELLEMENT
CONNAISSANCE ET À FOURNIR UNE DOCUMEN-
TATION EXPLICATIVE.
OUTRE LE FAIT D’ÊTRE EXAMINÉS POUR
ÉTABLIR S’ILS SONT ACCEPTABLES À DES FINS
INDUSTRIELLES, TECHNOLOGIQUES ET COM-
Numéro de référence
MERCIALES, AINSI QUE DU POINT DE VUE
ISO/FDIS 2613-1:2023(F)
DES UTILISATEURS, LES PROJETS DE NORMES
INTERNATIONALES DOIVENT PARFOIS ÊTRE
CONSIDÉRÉS DU POINT DE VUE DE LEUR POSSI-
BILITÉ DE DEVENIR DES NORMES POUVANT
SERVIR DE RÉFÉRENCE DANS LA RÉGLEMENTA-
TION NATIONALE. © ISO 2023
---------------------- Page: 1 ----------------------
ISO/FDIS 2613-1:2023(F)
PROJET
NORME ISO/FDIS
FINAL
INTERNATIONALE 2613-1
ISO/TC 193/SC 1
Analyse du gaz naturel — Teneur en
Secrétariat: NEN
silicium du biométhane —
Début de vote:
2023-02-03
Partie 1:
Vote clos le:
Détermination de la teneur totale en
2023-03-31
silicium par spectrométrie d’émission
atomique (SEA)
Analysis of natural gas — Silicon content of biomethane —
Part 1: Determination of total silicon by atomic emission spectroscopy
(AES)
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2023
TRAITEMENT PARALLÈLE ISO/CEN

Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette

publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,

LES DESTINATAIRES DU PRÉSENT PROJET SONT

y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut

INVITÉS À PRÉSENTER, AVEC LEURS OBSER-
VATIONS, NOTIFICATION DES DROITS DE PRO-

être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.

PRIÉTÉ DONT ILS AURAIENT ÉVENTUELLEMENT
ISO copyright office
CONNAISSANCE ET À FOURNIR UNE DOCUMEN-
TATION EXPLICATIVE.
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève OUTRE LE FAIT D’ÊTRE EXAMINÉS POUR
ÉTABLIR S’ILS SONT ACCEPTABLES À DES FINS
Tél.: +41 22 749 01 11
INDUSTRIELLES, TECHNOLOGIQUES ET COM-
Numéro de référence
E-mail: copyright@iso.org
MERCIALES, AINSI QUE DU POINT DE VUE
ISO/FDIS 2613-1:2023(F)
Web: www.iso.org
DES UTILISATEURS, LES PROJETS DE NORMES
INTERNATIONALES DOIVENT PARFOIS ÊTRE
Publié en Suisse
CONSIDÉRÉS DU POINT DE VUE DE LEUR POSSI-
BILITÉ DE DEVENIR DES NORMES POUVANT
SERVIR DE RÉFÉRENCE DANS LA RÉGLEMENTA-
© ISO 2023 – Tous droits réservés
TION NATIONALE. © ISO 2023
---------------------- Page: 2 ----------------------
ISO/FDIS 2613-1:2023(F)
Sommaire Page

Avant-propos .............................................................................................................................................................................................................................iv

Introduction .................................................................................................................................................................................................................................v

1 Domaine d'application ...................................................................................................................................................................................1

2 Références normatives ..................................................................................................................................................................................1

3 Termes et définitions ...................................................................................................................................................................................... 2

4 Principe.......................................................................................................................................................................................................................... 2

5 Réactifs et matériel de laboratoire ..................................................................................................................................................2

6 Appareillage .............................................................................................................................................................................................................. 6

6.1 Équipement d'échantillonnage et de dérivatisation .............................................................................................. 6

6.2 Instrument SEA MWP/ICP ........................................................................................................................................................... 7

6.3 Balance analytique d'une précision de 0,01 mg. ....................................................................................................... 7

7 Échantillonnage ....................................................................................................................................................................................................7

8 Dérivatisation .........................................................................................................................................................................................................9

9 Procédure analytique ..................................................................................................................................................................................... 9

9.1 Installation de l'équipement ....................................................................................................................................................... 9

9.2 Courbe d'étalonnage ....................................................................................................................................................................... 10

9.3 Analyse d'échantillons inconnus et de contrôle qualité.................................................................................. 10

10 Calcul .............................................................................................................................................................................................................................11

11 Expression des résultats ..........................................................................................................................................................................11

12 Précision de la méthode ............................................................................................................................................................................11

13 Incertitude de mesure.................................................................................................................................................................................12

14 Rapport d'essai ...................................................................................................................................................................................................12

Bibliographie ...........................................................................................................................................................................................................................13

iii
© ISO 2023 – Tous droits réservés
---------------------- Page: 3 ----------------------
ISO/FDIS 2613-1:2023(F)
Avant-propos

L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes

nationaux de normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est

en général confiée aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude

a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,

gouvernementales et non gouvernementales, en liaison avec l'ISO participent également aux travaux.

L'ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui

concerne la normalisation électrotechnique.

Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont

décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents

critères d'approbation requis pour les différents types de documents ISO. Le présent document

a été rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2

(voir www.iso.org/directives).

L'attention est attirée sur le fait que certains des éléments du présent document peuvent faire l'objet de

droits de propriété intellectuelle. L'ISO ne saurait être tenue pour responsable de ne pas avoir identifié

de tels droits de propriété. Les détails concernant les références aux droits de propriété intellectuelle

identifiés lors de l'élaboration du document sont indiqués dans l'Introduction et/ou dans la liste des

déclarations de brevets reçues par l'ISO (voir www.iso.org/brevets).

Les appellations commerciales éventuellement mentionnées dans le présent document sont données

pour information, par souci de commodité, à l'intention des utilisateurs et ne sauraient constituer un

engagement.

Pour une explication de la nature volontaire des normes, la signification des termes et expressions

spécifiques de l'ISO liés à l'évaluation de la conformité, ou pour toute information au sujet de l'adhésion

de l'ISO aux principes de l'Organisation mondiale du commerce (OMC) concernant les obstacles

techniques au commerce (OTC), voir le lien suivant: www.iso.org/iso/fr/avant-propos.

Le présent document a été élaboré par le comité technique ISO/TC 193, Gaz naturel, sous-comité SC 1,

Analyse du gaz naturel, en collaboration avec le Comité Européen de Normalisation (CEN) Comité Technique

CEN/TC 408, Biométhane pour utilisation dans les transports et injection dans le réseau de gaz naturel,

conformément à l’accord de coopération technique entre l’ISO et le CEN (Accord de Vienne).

Il convient que l'utilisateur adresse tout retour d'information ou toute question concernant le présent

document à l'organisme national de normalisation de son pays. Une liste exhaustive desdits organismes

se trouve à l'adresse www.iso.org/fr/members.html.
© ISO 2023 – Tous droits réservés
---------------------- Page: 4 ----------------------
ISO/FDIS 2613-1:2023(F)
Introduction

Le présent document décrit une méthode pour mesurer la concentration totale de silicium dans le

biométhane, le biogaz ou des matrices gazeuses similaires lorsqu'ils sont injectés dans des réseaux de

gaz naturel ou utilisés comme combustibles pour le transport. Cette méthode est fondée sur l'utilisation

d’un barboteur en solution liquide pour piéger le silicium d’un échantillon de gaz, suivie d'une analyse

instrumentale.

En raison de l'utilisation intensive de composés de siloxane, de la volatilité de ces derniers et de leur

grande affinité avec les milieux apolaires, les siloxanes sont considérés comme l'une des impuretés

majeures du biogaz. Ces composés sont indésirables car leur combustion peut produire du SiO abrasif,

qui peut endommager les moteurs et les appareils. De plus, certains d'entre eux présentent un risque

pour la santé.

Pour les besoins du présent document, les composés au silicium mesurés sont exprimés en tant que

«silicium total». Le silicium mesuré provient de composés organo-siliciés qui sont piégés en phase

gazeuse dans des milieux liquides puis convertis sous forme analytique d’ions hexafluorosilicate (SiF

), toujours présents lors de l'analyse de la solution.
© ISO 2023 – Tous droits réservés
---------------------- Page: 5 ----------------------
PROJET FINAL DE NORME INTERNATIONALE ISO/FDIS 2613-1:2023(F)
Analyse du gaz naturel — Teneur en silicium du
biométhane —
Partie 1:
Détermination de la teneur totale en silicium par
spectrométrie d’émission atomique (SEA)
1 Domaine d'application

Le présent document s'applique à la mesure de la teneur en silicium total dans des matrices

gazeuses telles que le biométhane et le biogaz. Le silicium est présent sous forme gazeuse, et contenu

principalement dans des composés de siloxane, le triméthylsilane et le triméthylsilanol. La forme

analytique du silicium mesurée dans la phase liquide après échantillonnage et dérivatisation, est

un anion hexafluorosilicate soluble stable dans les milieux légèrement acidifiés. Le silicium total est

exprimé sous la forme d'une masse de silicium dans le volume du gaz analysé.

Le présent document s'applique aux matrices gazeuses énoncées, avec des concentrations de silicium

allant jusqu'à 5 mg/m , et concerne principalement les matrices de biométhane avec une teneur en

3 3
silicium comprise entre 0,1 mg/m et 0,5 mg/m .

Avec une adaptation assurant une efficacité d'absorption appropriée, il peut être utilisé pour des

concentrations plus élevées. La limite de détection de cette méthode est estimée à 0,05 mg/m pour

un volume d'échantillon de gaz de 0,020 m . Tous les composés présents dans la phase gazeuse sont

volatils à la température d'absorption et de dérivatisation, et les composés organosiliciés gazeux sont

piégés dans le milieu d'absorption puis dérivés sous forme de silicium analytique qui est mesuré selon

la présente méthode. La concentration de silicium est mesurée dans le milieu de dérivatisation dilué,

par spectrométrie d'émission atomique, après atomisation/ionisation dans un plasma micro-ondes ou

à couplage inductif.

Sauf indication contraire, tous les volumes et concentrations sont établis pour des conditions normales

(temperature, 273 K et pression, 101,325 kPa).

NOTE Avec des facteurs de dilution appropriés, la méthode peut également être appliquée à des

concentrations de silicium supérieures à 5 mg/m .
2 Références normatives

Les documents suivants sont cités dans le texte de sorte qu'ils constituent, pour tout ou partie de leur

contenu, des exigences du présent document. Pour les références datées, seule l'édition citée s'applique.

Pour les références non datées, la dernière édition du document de référence s'applique (y compris les

éventuels amendements).

ISO 3696, Eau pour laboratoire à usage analytique — Spécification et méthodes d'essai

ISO 6143, Analyse des gaz — Méthodes comparatives pour la détermination et la vérification de la

composition des mélanges de gaz pour étalonnage
ISO 14532, Gaz naturel — Vocabulaire
ISO 10715, Gaz naturel — Échantillonnage de gaz

ISO 14912, Analyse des gaz — Conversion des données de composition de mélanges gazeux

© ISO 2023 – Tous droits réservés
---------------------- Page: 6 ----------------------
ISO/FDIS 2613-1:2023(F)
3 Termes et définitions

Pour les besoins du présent document, les termes et définitions donnés dans l'ISO 14532 ainsi que les

suivants s'appliquent.

L'ISO et l'IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en

normalisation, consultables aux adresses suivantes:

— ISO Online browsing platform: disponible à l'adresse https:// www .iso .org/ obp.

— IEC Electropedia: disponible à l'adresse https:// www .electropedia .org/
3.1
siloxanes

groupes fonctionnels composés de deux atomes de silicium liés par un atome d'oxygène

Note 1 à l'article: En fonction du substrat utilisé pour produire le biogaz et du procédé utilisé pour la purification,

le biométhane peut contenir des siloxanes. Lors de la combustion, les siloxanes peuvent être oxydés en dioxyde de

silicium, un composé abrasif dangereux pour les pièces mécaniques en mouvement des moteurs ou des turbines

par exemple.
3.2
spectrométrie d'émission atomique
SEA

méthode d'analyse chimique qui utilise l'intensité de la lumière émise par une flamme, un plasma, un

arc ou une étincelle à une longueur d'ondes donnée, pour déterminer la quantité d'un élément dans un

échantillon
4 Principe

AVERTISSEMENT — Il convient que l'utilisateur de la présente Norme internationale connaisse

bien les pratiques courantes de laboratoire. La présente norme n'a pas pour but de traiter tous

les problèmes de sécurité qui sont, le cas échéant, liés à son utilisation. Il incombe à l'utilisateur

d'établir des pratiques appropriées en matière d'hygiène et de sécurité et de s'assurer de la

conformité à la réglementation nationale en vigueur.

On fait passer un échantillon de matrice gazeuse de méthane (par exemple, biométhane, biogaz, gaz

naturel ou mélanges de ces composés) contenant des composés de siloxane dans un liquide absorbant

(acide nitrique) contenu dans des barboteurs à gaz/barboteurs connectés en série afin de recueillir

les composés contenant du silicium. Après échantillonnage d'un volume adéquat de gaz, on soumet

le contenu des récipients d'échantillonnage (barboteurs à gaz) à une dérivatisation en y ajoutant une

solution d'hydroxyde et de l'acide fluorhydrique afin d'obtenir du silicium sous forme analytique: l'anion

hexafluorosilicate (SiF ).

On analyse la teneur en silicium de l'échantillon dérivé à l'aide d'un spectromètre d'émission atomique

ICP/MWP aux longueurs d'ondes d'émission caractéristiques du silicium, avec un étalonnage multi-

points linéaire obtenu par l'analyse d'une série de solutions étalons de silicium.

5 Réactifs et matériel de laboratoire

Pour appliquer la méthode, seuls les réactifs de qualité analytique reconnue suivants et de l'eau de

qualité 1 selon l'ISO 3696 doivent être utilisés. En cas de changement d'aspect (couleur, consistance,

turbidité) visible à l'œil nu, les réactifs doivent être éliminés et remplacés par de nouveaux réactifs.

5.1 Milieu d'absorption

5.1.1 Acide nitrique (HNO ), ρ = 1,41 g/ml; 65 % HNO (fraction massique) – pour l'analyse des

3 20 3
éléments à l'état de traces.
© ISO 2023 – Tous droits réservés
---------------------- Page: 7 ----------------------
ISO/FDIS 2613-1:2023(F)

ATTENTION — Ce produit chimique est particulièrement dangereux s'il n'est pas utilisé dans

des conditions de laboratoire spécifiques. Des essais ont été réalisés avec d'autres acides

minéraux non oxydants (HCI), mais il a été démontré qu'ils ne conviennent pas pour l'absorption

des siloxanes de la phase gazeuse. Des précautions particulières sont à prendre lors de la

manipulation de ce produit chimique dans des conditions de laboratoire et en conditions réelles.

5.2 Milieu de dérivatisation

5.2.1 Hydroxyde de sodium en pastilles, pour la préparation d'une solution d'hydroxyde

de 8 mol/l à 10 mol/l.

Peser avec précision la quantité appropriée d'hydroxyde de sodium en pastilles et la dissoudre dans la

quantité d'eau ultrapure correspondante (5.3). Par exemple, pour 100 ml de solution d'hydroxyde de

sodium à 10 mol/l, peser 40 g d'hydroxyde de sodium en pastilles et les dissoudre dans 100 ml d'eau.

Il est également possible d'utiliser de l'hydroxyde de potassium, mais l'hydroxyde de sodium est préféré

pour des raisons de sécurité.

AVERTISSEMENT — La réaction liée à la dissolution de l'hydroxyde de sodium dans l'eau est

fortement exothermique! Elle libère de la chaleur et il convient d'agir avec précaution lors des

manipulations. Ajouter les pastilles dans l'eau petit à petit, et refroidir le récipient jusqu'à ce

que la dissolution soit terminée.
5.2.2 Acide fluorhydrique (HF), ρ = 1,16 g/ml; 48 % HF (fraction massique).
(20 °C)

AVERTISSEMENT — L'acide fluorhydrique est un acide très toxique qui pénètre profondément

dans la peau et les tissus s'il n'est pas traité immédiatement. Les lésions se produisent en deux

temps: d'abord par l'hydratation, qui entraîne une nécrose des tissus, puis par la pénétration

d'ions fluorure en profondeur dans les tissus et leur réaction avec le calcium. Il convient

d'appliquer immédiatement de l'acide borique ou d'autres réactifs complexants ainsi que des

réactifs de traitement appropriés. Se référer aux documents de sécurité pour déterminer quels

équipements de sécurité (lunettes, vêtements, gants) porter lors de la manipulation d'acide

fluorhydrique. Toujours avoir le matériel de soin nécessaire à portée de main avant de démarrer

les travaux avec cet acide.

ATTENTION — Ce produit chimique est particulièrement dangereux s'il n'est pas utilisé dans

des conditions de laboratoire spécifiques. Des essais ont été réalisés avec d'autres réactifs de

dérivatisation donneurs de fluorure (NaF), cependant ils semblent être inadéquats pour la

dérivatisation des siloxanes absorbés en phase gazeuse. Il convient de prendre des précautions

particulières lors de la manipulation de ce produit chimique dans des conditions de laboratoire

et en conditions réelles.
5.3 Eau, de qualité 1 conformément à l'ISO 3696.
5.4 Composés de siloxanes purs:
Siloxanes linéaires Formule brute Siloxanes cycliques Formule brute
Hexaméthyldisiloxane - L2 C H OSi Hexaméthylcyclotrisiloxane - D3 C H O Si
6 18 2 6 18 3 3
Octaméthyltrisiloxane- L3 C H O Si Octaméthylcyclotétrasiloxane - D4 C H O Si
8 24 2 3 8 24 4 4
Décaméthyltétrasiloxane - L4 C H O Si Décaméthylcyclopentasiloxane - D5 C H O Si
10 30 3 4 10 30 5 5

Dodécaméthylpentasiloxane - L5 C H O Si Dodécaméthylcyclohexasiloxane - D6 C H O Si

12 36 4 5 12 36 6 6

Utiliser au moins un représentant linéaire et un représentant cycliques des composés de siloxanes afin

de réaliser un contrôle qualité initial et régulier de la validité de la méthode.

5.5 Bandelettes indicatrices de pH, plages de pH allant de 0 à 14, ou, le cas échéant, un pH-mètre

avec une électrode résistante à l'HF.
© ISO 2023 – Tous droits réservés
---------------------- Page: 8 ----------------------
ISO/FDIS 2613-1:2023(F)
5.6 Solutions d'étalonnage
5.6.1 Généralités

Le mode opératoire suivant, pour la préparation de solutions étalons et de solutions d'étalonnage du

silicium, est ajusté à la concentration de silicium la plus basse dans un échantillon gazeux. Dans le cas

où des concentrations de silicium plus élevées doivent être mesurées, régler les concentrations des

solutions étalons de travail et d'étalonnage en conséquence.

Pour déterminer la concentration de silicium dans des échantillons aqueux, seul du matériel de

laboratoire en plastique, en PTFE ou en quartz doit être utilisé, à partir du moment où l'échantillon est

recueilli et jusqu'à la fin de l'analyse.
5.6.2 Solution étalon mère de Silicium certifiée pour l’ICP

Un exemple de solution étalon de Si certifiée peut être une solution avec une concentration massique en

Si de 10 000 μg/ml dans de l’eau (seul du HF a l’état de traces est acceptable) et une incertitude élargie

relative (facteur d’élargissement k = 2) de 0,5 %. Cette concentration est utilisée dans l’exemple de la

préparation de la solution étalon de si du 5.6.3.

Des solutions étalons de Si certifiées avec d'autres concentrations peuvent aussi être utilisées. Ajuster

le mode opératoire pour la préparation de solution étalon en conséquence.

Si la solution étalon mère de Si est préparée, en interne, par pesée de sels de silicium, appliquer les

méthodes statistiques requises afin d'obtenir une concentration précise, accompagnée d'une valeur

d'incertitude.
NOTE Les références [1][2] fournissent des recommandations.
5.6.3 Solution étalon de Si
Concentration masique en Si visée ρ(Si) ≈ 100 mg/kg

Peser une fiole jaugée de 50 ml, vide et en plastique, à l'aide d'une balance analytique (6.3). Ajouter

environ 10 ml d'acide nitrique 2 % (fraction massique). Pipetter avec précision 0,5 ml de solution

étalon (5.6.2) et l'ajouter à la fiole jaugée. Diluer au volume avec de l'acide nitrique 2 % (fraction

massique). Peser la fiole jaugée remplie et calculer la concentration de silicium.

Conserver la solution dans une fiole jaugée ou un récipient similaire sans silicone, correctement fermé,

à température ambiante ou au frais (~5 °C). La solution reste stable pendant deux semaines environ, si

elle est correctement conservée.
5.6.4 Solutions d'étalonnage de Si

Préparer, par pesée, un minimum de cinq solutions d'étalonnage, en fonction de la concentration de

silicium attendue dans les échantillons prélevés.

Par exemple, procéder comme suit pour une plage de concentrations massiques de Si de 10 μg / kg

à 200 μg / kg:
Peser des fioles jaugées de 100 ml (ou 200 ml), vides et en plastique.

Pipetter respectivement 10 μl, 20 μl, 50 μl, 75 μl, 100 μl, 150 μl et 200 μl de solution étalon de

silicium (5.6.3) dans une fiole jaugée en plastique à un trait de 100 ml, qui a été pesée vide, et préremplie

avec environ 10 ml à 20 ml d'acide nitrique 2 % (fraction massique). Diluer au volume avec de l'acide

nitrique 2 % (fraction massique). Peser la fiole jaugée remplie et calculer la concentration de silicium.

Les concentrations massiques de silicium des solutions d'étalonnage sont respectivement 10 μg/kg;

20 μg/kg; 50 μg/kg; 75 μg/kg; 100 μg/kg; 150 μg/kg et 200 μg/kg.
© ISO 2023 – Tous droits réservés
---------------------- Page: 9 ----------------------
ISO/FDIS 2613-1:2023(F)

Calculer l’incertitude sur les concentrations massiques des solutions d’étalonnage. Vérifiez quelles sont

les contributions de la combinaison des effets partagés, telles que l'incertitude de la concentration

de la solution mère ou de l'étalonnage de la pipette. Si ces effets contribuent pour plus de 40 % du

budget d’incertitude, alors calculer les facteurs de dilution et leurs incertitudes associées. Ainsi les

concentrations de ces solutions étalon sont substantiellement corrélées.
5.6.5 Solution pour le contrôle d'étalonnage de la longueur d'ondes

Procéder à la vérification de la longueur d'ondes avant l'étalonnage quotidien de l'instrument

pour l'analyse en silicium, à l'aide d'une solution qui contient les différents éléments couvrant la

plage de longueur d'ondes de l'instrumentation utilisée, fournie par le fabricant,. Cette solution est

habituellement fournie sous la forme d'un concentré qui doit être dilué avant l'analyse, conformément

aux instructions du fabricant. Le résultat de l'essai du contrôle d'étalonnage de la longueur d'ondes

montre si les paramètres optiques de l'instrument sont appropriés, et si le relevé des raies d'émission de

chaque élément correspond aux paramètres de l'instrument lorsque les longueurs d'ondes analytiques

sont sélectionnées pour l'analyte d'intérêt.

NOTE La solution pour le contrôle d'étalonnage est habituellement fournie par le fabricant de l'équipement.

5.7 Contrôle qualité
5.7.1 Blancs

Trois types de blancs sont utilisés lors de l'analyse. Le blanc d'étalonnage est utilisé pour établir la

courbe analytique, le blanc de réactif de laboratoire est utilisé pour repérer toute contamination issue

du mode opératoire de la préparation d'échantillon et un blanc de rinçage est utilisé pour rincer les

instruments et le nébuliseur entre les étalons, les solutions de contrôle et les échantillons pour réduire

les interférences de mémoire.

5.7.1.1 Le blanc d'étalonnage est préparé en acidifiant une eau ultrapure aux mêmes concentrations

que celles des acides utilisés pour les étalons; dans ce cas il s'agit d'acide nitrique 2 % (fraction

massique). Il convient de conserver le blanc d'étalonnage dans un récipient en plastique comme les

échantillons.

5.7.1.2 Il convient que le blanc de réactif de laboratoire contienne tous les réactifs dans les mêmes

volumes que ceux utilisés dans le traitement des échantillons. Le blanc de réactif de laboratoire doit

être appliqué tout au long d'un même mode opératoire de préparation en tant qu'échantillon, y compris

la dérivatisation d'échantillon. Il convient de préparer ce type de blanc au moins à chaque utilisation

d'un nouveau réactif.

5.7.1.3 Le blanc de rinçage est préparé en acidifiant de l'eau ultrapure aux mêmes concentrations

d'acide nitrique que celles utilisées pour le blanc d'étalonnage et conservé de manière pratique.

5.7.1.4 Blanc d'ustensiles de laboratoire – méthane gazeux pur sans silicium, utilisé en tant que

blanc d’échantillon de gaz pour vérifier la propreté du matériel utilisé.

5.7.2 Contrôle de performances des instruments, à savoir échantillon de contrôle de

l'étalonnage de la longueur d'ondes (voir 5.6.5)
5.7.3 Échantillon de contrôle d'étalonnage (CC)

Un échantillon de contrôle d'étalonnage doit être utilisé pour des vérifications initiales et périodiques

des étalons de référence ou des solutions étalons mères afin de vérifier la performance des instruments.

Le CC doit être obtenu via une source externe différente de celle des solutions mères étalons et préparé

dans le même mélange d'acides que les solutions d’étalonnage. Il peut s'agir d'une solution étalon déjà

prête provenant d'un fournisseur différent, ou au moins d'un lot différent, ou bien il peut être préparé

© ISO 2023 – Tous droits réservés
---------------------- Page: 10 ----------------------
ISO/FDIS 2613-1:2023(F)

par pesée à l'aide de sel (NH ) SiF pur. Il convient d'obtenir une concentration de silicium dans la

4 2 6

solution CC proche de celle attendue dans l'échantillon ou au milieu du domaine d'étalonnage. Il convient

de préparer une nouvelle solution avant l'analyse et de la conserver dans un récipient en plastique,

comme les échantillons.
5.7.4 Échantillon de contrôle de la dérivatisation (DC)

Un échantillon de contrôle de la dérivatisation doit être utilisé pour des vérifications initiales et

périodiques de complétude du processus de dérivatisation. À cet effet, des composés de siloxane pur

sont utilisés. Par exemple,
...