ISO/DIS 15883-4
(Main)Washer-disinfectors
Washer-disinfectors
Laveurs désinfecteurs
General Information
RELATIONS
Standards Content (sample)
DRAFT INTERNATIONAL STANDARD ISO/DIS 15883-4
ISO/TC 198 Secretariat: ANSI
Voting begins on: Voting terminates on:
2003-07-24 2003-12-24
INTERNATIONAL ORGANIZATION FOR STANDARDIZATION • МЕЖДУНАРОДНАЯ ОРГАНИЗАЦИЯ ПО СТАНДАРТИЗАЦИИ • ORGANISATION INTERNATIONALE DE NORMALISATION
Washer-disinfectors —Part 4:
Requirements and tests for washer-disinfectors employing
chemical disinfection for thermolabile endoscopes
Laveurs désinfecteurs —
Partie 4: Exigences et essais pour les laveurs désinfecteurs destinés à la désinfection chimique des
endoscopes thermolabilesICS 11.080.10
ISO/CEN PARALLEL ENQUIRY
This draft International Standard is a draft European Standard developed within the European Committee
for Standardization (CEN) and processed under the CEN-lead mode of collaboration as defined in the
Vienna Agreement. The document has been transmitted by CEN to ISO for circulation for ISO member
body voting in parallel with CEN enquiry. Comments received from ISO member bodies, including those
from non-CEN members, will be considered by the appropriate CEN technical body. Accordingly, ISO
member bodies who are not CEN members are requested to send a copy of their comments on
this DIS directly to CEN/TC 102 (DIN, Burggrafenstraße, 6 D-10787 Berlin) as well as returning their
vote and comments in the normal way to the ISO Central Secretariat. Should this DIS be accepted,
a final draft, established on the basis of comments received, will be submitted to a parallel two-month
FDIS vote in ISO and formal vote in CEN.In accordance with the provisions of Council Resolution 15/1993 this document is circulated in
the English language only.Conformément aux dispositions de la Résolution du Conseil 15/1993, ce document est distribué
en version anglaise seulement.THIS DOCUMENT IS A DRAFT CIRCULATED FOR COMMENT AND APPROVAL. IT IS THEREFORE SUBJECT TO CHANGE AND MAY NOT BE
REFERRED TO AS AN INTERNATIONAL STANDARD UNTIL PUBLISHED AS SUCH.IN ADDITION TO THEIR EVALUATION AS BEING ACCEPTABLE FOR INDUSTRIAL, TECHNOLOGICAL, COMMERCIAL AND USER PURPOSES, DRAFT
INTERNATIONAL STANDARDS MAY ON OCCASION HAVE TO BE CONSIDERED IN THE LIGHT OF THEIR POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN NATIONAL REGULATIONS.© International Organization for Standardization, 2003
---------------------- Page: 1 ----------------------
ISO/DIS 15883-4
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but shall
not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the
unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
Copyright noticeThis ISO document is a Draft International Standard and is copyright-protected by ISO. Except as permitted
under the applicable laws of the user's country, neither this ISO draft nor any extract from it may be
reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, photocopying,
recording or otherwise, without prior written permission being secured.Requests for permission to reproduce should be addressed to either ISO at the address below or ISO's
member body in the country of the requester.ISO copyright office
Case postale 56 CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Reproduction may be subject to royalty payments or a licensing agreement.
Violators may be prosecuted.
ii ISO 2003 – All rights reserved
---------------------- Page: 2 ----------------------
prEN ISO 15883-4:2003 (E)
Contents Page
Foreword ............................................................................................................................................................. 3
Introduction ........................................................................................................................................................ 4
1 Scope...................................................................................................................................................... 5
2 Normative references............................................................................................................................ 5
3 Definitions.............................................................................................................................................. 5
4 Performance requirements .................................................................................................................. 6
5 Mechanical and process requirements............................................................................................. 14
6 Testing for conformity ........................................................................................................................ 16
7 Documentation and inspection.......................................................................................................... 29
8 Information to be supplied by the manufacturer ............................................................................. 29
9 Marking, labelling and packaging...................................................................................................... 30
10 Information to be requested from the purchaser by the manufacturer......................................... 30
Annex A (normative) Tests for microbial contamination of post-disinfection rinse water....................... 31
Annex B (normative) Microbiological testing of the efficacy of chemical disinfection of the
load ....................................................................................................................................................... 32
Annex C (informative) Summary programme of tests in addition to prEN ISO 15883-1:2003 .................. 36
Annex D (normative) Method for microbiological evaluation of disinfection of water
treatment system................................................................................................................................. 37
Annex E (informative) Microbiological testing of chemical disinfection processes ................................. 43
Annex ZA (informative) Clauses of this European Standard addressing essentialrequirements or other provisions of EU Directives......................................................................... 45
Bibliography ..................................................................................................................................................... 47
---------------------- Page: 3 ----------------------prEN ISO 15883-4:2003 (E)
Foreword
This document (prEN ISO 15883-4:2003) has been prepared by Technical Committee CEN/TC 102,
"Sterilizers for medical purposes", the secretariat of which is held by DIN, in collaboration with
Technical Committee ISO/TC 198 "Sterilization of health care products".This document is currently submitted to the parallel Enquiry.
EN ISO 15883 consists of the following Parts under the general title “Washer-Disinfectors”.
Part 1: General requirements, definitions and tests.Part 2: Requirements and test for washer-disinfectors, employing thermal disinfection, for surgical
instruments, anaesthetic equipment, hollowware, utensils, glass ware etc.Part 3: Requirements and tests for washer-disinfectors employing thermal disinfection for human
waste containers.Part 4: Requirements and tests for washer-disinfectors employing chemical disinfection for thermo-
labile endoscopes.Other fields of application within the scope of CEN/TC 102/WG 8 include laboratory, veterinary,
dental, and pharmaceutical applications. Performance and test requirements for these applications
may be specified in other Parts of this standard (to be produced later).Annexes A, B, and D are normative parts of this European Standard. Annexes C, E and ZA are given
for information only.This document has been prepared under a mandate given to CEN by the European Commission and
the European Free Trade Association, and supports essential requirements of EU Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this
document.In respect of potential adverse effects on the quality of drinking water intended for human
consumption, caused by the product covered by this standard:a) this standard provides no information as to whether the product may be used without restriction in
any of the member states of the EU or EFTA;b) it should be noted that while awaiting the adoption of verifiable European criteria existing national
regulation concerning the use and/or the characteristics of the product remain in force.
---------------------- Page: 4 ----------------------prEN ISO 15883-4:2003 (E)
Introduction
It is recommended that this Introduction be read in conjunction with the Introduction to
prEN ISO 15883-1.This standard is the fourth part of a series specifying the performance of washer-disinfectors and
specifies the particular requirements for performance applicable to washer-disinfectors employing
chemical disinfection for thermo-labile endoscopes.The washer-disinfectors specified in this standard are intended to process devices which can be
immersed in water or aqueous solutions. For some devices this will require that, prior to processing,
relevant parts of the device are protected from immersion in accordance with the device
manufacturer’s operating instructions.This standard also specifies the performance requirements for the cleaning and disinfection of the
washer-disinfector and its components and accessories which may be required to achieve the
necessary performance.The methods, instrumentation and instructions required for type testing, works testing, validation
(installation, operational and performance qualification on first installation), routine control and
monitoring and re-validation, periodically and after essential repairs are also specified.
Requirements for washer-disinfectors for other applications are specified in other Parts of this
standard.Safety requirements for washer-disinfectors are given in IEC 61010-2-045.
---------------------- Page: 5 ----------------------
prEN ISO 15883-4:2003 (E)
1 Scope
This part of prEN ISO 15883 specifies the particular performance requirements for washer-
disinfectors (WDs) that are intended to be used for cleaning and chemical disinfection of thermo-labile
endoscopes.NOTE 1 WDs complying with this part of prEN ISO 15883 may also be used for cleaning and chemical
disinfection of other thermo-labile re-usable medical devices for which the device manufacturer has
recommended this method of disinfection.WDs complying with the requirements of this Part are not intended for cleaning and disinfection of
medical devices, including endoscopic accessories, which are heat stable and can be disinfected or
sterilized by thermal methods (see prEN ISO 15883-1:2003, 4.1.5).The specified performance requirements of this Part are not intended to ensure the inactivation or
removal of the causative agent(s) (prion protein) of Transmissible Spongiform Encephalopathies.
NOTE 2 Many disinfectants are known to fix protein and this should be considered when prion protein may be
present.2 Normative references
This European Standard incorporates by dated or undated reference, provisions from other
publications. These normative references are cited at the appropriate places in the text and the
publications are listed hereafter. For dated references, subsequent amendments to or revisions of any
of these publications apply to this European Standard only when incorporated in it by amendment or
revision. For undated references the latest edition of the publication referred to applies (including
amendments).prEN ISO 15883-1:2003, Washer-disinfectors — Part 1: General requirements, definitions and tests.
ISO 4064-1, Measurement of water flow in closed conduitsMeters for cold potable water 1:
Specifications.ISO 11731:1998, Water quality — Detection and enumeration of Legionella.
WHO Guidelines for drinking water quality 1998.
3 Definitions
For the purposes of this Part of prEN ISO 15883, the definitions given in prEN ISO 15883-1:2003
apply together with the following:3.1
air break
a physical separation in water supply pipes to prevent back syphonage into the water supply from a
device connected to it (see EN1717)3.2
inoculated carrier
a carrier on which a defined number of test organisms has been deposited [EN 866-1:1997]
1) EN 866-1 is currently under revision (Vienna Agreement), see ISO/CD 11138-1.---------------------- Page: 6 ----------------------
prEN ISO 15883-4:2003 (E)
3.3
leak test
a test intended to establish that the surface covering the device and/or lining a device channel is
intact to the extent necessary to maintain a slight positive pressure3.4
liquid transport systems
those components of the WD used to store, pump or transport water and /or solutions within the WD
excluding pipework before the air break3.5
minimum effective concentration
the concentration of disinfectant which, under the stated conditions of temperature and time, will meet
the required microbicidal performance3.6
self-disinfection cycle
an operating cycle under the control of the automatic controller, for use without any load in the WD,
which is intended to disinfect all liquid transport systems piping, chamber(s), tanks and other
components which come into contact with the water and / or solutions used for cleaning, disinfecting
and rinsing the loadNOTE This does not include single-use, multi-dose, containers used to provide process chemicals for use in
the WD.3.7
thermo-labile device
heat sensitive device
a re-usable device for which the temperature to which it may be exposed without damage precludes
the use of thermal disinfection4 Performance requirements
4.1 General
4.1.1 The requirements of prEN ISO 15883-1:2003 apply with the exception of sub-clauses :
4.3.1 (specification for thermal disinfection); 5.3.2.5 (microbial quality of final rinse water – amended in this Part);
6.4.2 (test for quality of final rinse water – amended in this Part);
6.5.6 (test for chamber venting to prevent pressurisation by steam);
6.7.2 (tests on trolleys for handling loads outside the WD);
6.8.2 (load temperature test – amended in this Part);
6.10 (cleaning efficacy test – amended in this Part);
which have been replaced or modified in this Part.
4.1.2 This standard deals with three aspects of WD performance. It specifies
a) performance requirements for the WD,
---------------------- Page: 7 ----------------------
prEN ISO 15883-4:2003 (E)
b) methods for reference tests used to verify that the WD meets the specified requirements, and
c) methods for tests that may be carried out to verify continued compliance of the installed WD with
the validated process conditions.The attention of WD manufacturers is drawn to clauses 4, 5, 7 and 8 in relation to specified
requirements and Clause 6 for methods of test that may be used to verify that the WD meets the
specified requirements.4.1.3 Each device, including any device channels and/or cavities, shall be processed by the WD as
follows :a) leak testing (when appropriate see 4.2.1);
b) cleaning;
c) rinsing (when appropriate);
d) disinfecting;
e) removing toxic residues (when appropriate);
f) drying (when appropriate see 4.6.1).
4.1.4 Rinsing shall take place between cleaning and disinfection unless it can be demonstrated
that:a) there is no reaction between incompatible process chemicals being used for each of these
phases, andb) there is no adverse reaction between suspended or residual soiling and the disinfectant.
NOTE 1 When specified by the device manufacturer and / or WD manufacturer pre-treatment or dis-assembly
may be required.NOTE 2 The efficacy of cleaning and disinfection depends on a number of factors which include
a) the nature of the device being processed;b) the extent and nature of the soiling to be removed;
c) the temperature;
d) the mechanical energy (type, output, duration);
e) the detergent system;
f) the nature, volume, concentration and temperature of the cleaning and disinfectant solutions and their ability
to wet the surfaces to be cleaned and disinfected.4.1.5 The value of any process variable which affects the efficacy of the cycle shall be pre-set and
adjustment shall require the use of a key, code or tool. (See also 5.18.8 and 5.18.12 of
prEN ISO 15883-1:2003)4.1.6 When the WD uses two, or more, different process chemicals, means shall be provided to
ensure that connection is made to the correct container of process chemical.NOTE Clear, permanent labelling of the connection system may be deemed to satisfy this requirement with
the prior agreement of the user (see clause 10).---------------------- Page: 8 ----------------------
prEN ISO 15883-4:2003 (E)
4.1.7 The WD manufacturer’s instructions shall recommend that, heat-stable endoscopic
accessories to thermo-labile devices should be thermally disinfected and/or sterilized. (see clause 8
and prEN ISO 15883-1:2003, 4.1.5)4.2 Leak test
4.2.1 The requirements of this clause shall apply only to WDs intended to process endoscopes
which require a test to verify that the device is watertight.NOTE This test is intended to demonstrate that the endoscope will not be damaged by liquid ingress during
the WD operating cycle. It should only be regarded as a test of the integrity of the endoscope when all
parameters of the WD leak test (e.g. pressure, duration, maximum leak accepted) are consistent with those
specified by the endoscope manufacturer.4.2.2 The WD shall either be provided with
a) means to carry out an automatic leak test on the endoscope which shall be completed before the
load comes into contact with water or aqueous solutions, orb) instructions for use that include the requirement to carry out the test manually prior to processing
through the WD.NOTE 1 An alternative method specified by the endoscope manufacturer may be used for determining the
integrity of the endoscope when appropriate.NOTE 2 WDs with an automatic leak test may include a user selectable option to repeat the leak test at the
end of the process and/or independently of a normal process cycle.4.2.3 For WDs having an automatic leak test, the automatic controller shall prevent the continuation
of the operating cycle and operate an audible and visible alarm indicating a leak test failure if a leak is
detected in an endoscope.NOTE 1 Variations in temperature may adversely affect the sensitivity of the leak test and the WD
manufacturer should state the maximum temperature range permitted.NOTE 2 A leak test failure indicates that the device is likely to be damaged by further processing; a
satisfactory leak test does not provide absolute assurance that the device will not be damaged by further
processing.4.2.4 In WDs provided with an automatic leak test:
the systems for connection of the device to the WD shall be designed so that the fittings intended
for the leak test cannot be connected to the irrigation system of the device's channel(s);
the means used to monitor the pressure inside the device (e.g. pressure transducer) shall be
independent from the means used to control the initial pressure (e.g. pressure regulating valve);
the system used to pressurise the device during each leak test shall be provided with means to
prevent over-pressurisation of the device in the event of the failure of the pressure control
system; the extent and duration of pressurisation and the pressure drop or air flow which will be used to
indicate a fail shall be either, in accordance with the device manufacturer’s instructions for the
devices which the WD is intended to process, or, independently verified by the WD manufacturer.
4.2.5 The ability of the automatic leak test to detect a leak shall be verified by testing in accordance
with 6.4.---------------------- Page: 9 ----------------------
prEN ISO 15883-4:2003 (E)
4.3 Cleaning
4.3.1 General
All surfaces (internal and external) of the devices which are required to be disinfected by the WD shall
be cleaned.NOTE Some devices have component parts (e.g. electronic connectors) which their manufacturer
recommends should not be immersed in water or aqueous solutions. These component parts should be
manually cleaned in accordance with the manufacturer’s instructions and then protected from immersion during
processing in the WD (see 5.1.2).Cleaning shall comprise washing with a detergent solution which may, when necessary be preceded
by flushing. Washing shall be followed by rinsing unless the conditions specified in 4.1.4 have been
met.4.3.2 Flushing
When necessary, the WD shall provide means to flush the internal and external surfaces of the
endoscope.NOTE Flushing before washing may be recommended to eliminate soils or to avoid any interaction between
the chemicals used during pre-treatment and those of the WD processing cycle.The flushing water or solution shall be discharged during or after each process cycle and shall not be
re-used.4.3.3 Washing
The detergent solution shall be discharged during or after each process cycle and shall not be reused.
4.3.4 Post-washing rinsingRinsing between cleaning and disinfection shall be used to reduce the concentration of residues
(process chemicals and soiling including microbial contamination) to a level established as not
exceeding that which would impair the efficacy of the chemical disinfectant, except as provided for in
4.1.4.The rinse water quality shall be specified by the WD manufacturer.
NOTE This should be of, at least, potable quality.
4.3.5 Determination of cleaning efficacy
Cleaning shall be deemed to be completed satisfactorily when the chosen test soil(s) are no longer
detectable by the assessment method(s) specified (see 6.10).4.4 Disinfecting
4.4.1 General
The capability of the WD to provide disinfection of the device shall be deemed to have been
established if, when the WD is tested as specified in 6.11.5 under the specified conditions of
disinfectant concentration, temperature and contact time the required microbial reduction factor is
attained.---------------------- Page: 10 ----------------------
prEN ISO 15883-4:2003 (E)
The choice of disinfectant shall ensure that the spectrum of activity is appropriate for the intended use
and that the extent of microbial inactivation during the entire process is sufficient to achieve the
overall microbial reduction required.The efficacy of disinfectants may be seriously impaired by residual soiling, inorganic salts etc
remaining on the device(s) and therefore an effective cleaning prior to disinfection is required.
NOTE 1 Tests conducted on disinfectants should be carried out at the end of the shelf life specified by the
disinfectant manufacturer after the disinfectant has been stored under the worst case storage conditions specified
by the disinfectant manufacturer. (These conditions may be simulated by the use of validated accelerated ageing
when necessary).NOTE 2 Other process chemicals, e.g. detergents may react with and seriously impair the activity of
disinfectants if they are not removed before the disinfection stage.4.4.2 Efficacy of the disinfectant
4.4.2.1 The following tests are based on the use of aqueous solutions of a microbicide. Other
systems based on gaseous microbicides are not excluded; equivalent tests would be required.
4.4.2.2 The efficacy of the disinfectant shall be demonstrated by in vitro tests (see 6.11.1).
4.4.2.3 When tested for the minimum exposure time at the minimum concentration and the
minimum temperature to be used in the WD the disinfectant shall demonstrate:a) at least a log 6 reduction in population of vegetative bacteria including mycobacteria, yeasts and
yeast-like fungi;b) at least a log 4 reduction in population of fungal spores and viruses.
NOTE The reduced requirement in respect of fungal spores and viruses is based on recognition of the
practical difficulty of working with high titres of these organisms to reliably demonstrate higher levels of
inactivation.4.4.2.4 The disinfectant chosen shall also be active against bacterial endospores.
NOTE When tested at the minimum concentration and the minimum temperature to be used in the WD the
disinfectant should reduce the population of bacterial spores by not less than log106 within 5 h exposure. The
disinfectant should be tested against spores of known high resistance to the disinfectant from both aerobic and
anaerobic organisms.4.4.2.5 Demonstration by the disinfectant manufacturer that the disinfectant meets the above
requirements may be made employing methods described in relevant published standards
(e.g. prEN 13727, prEN 14348, AOAC sporicidal test, etc).4.4.2.6 National regulatory requirements may specify approval procedures for disinfectants to be
used in WDs for medical devices. Compliance with these national requirements shall be deemed to
meet the requirements of 4.4.1.4.4.3 Temperature
The temperature of the disinfectant solution throughout the disinfection stage shall be monitored and
controlled within the limits specified by the manufacturer of the disinfectant and be compatible with the
temperature limits for the device(s) to be processed.NOTE This may be achieved by a) controlling the temperature of the disinfectant solution or b) where
appropriate, by operating the WD at ambient temperature with means to prevent operation of the WD when the
ambient temperature is outside the specified temperature range.---------------------- Page: 11 ----------------------
prEN ISO 15883-4:2003 (E)
4.4.4 Process monitoring
The process monitoring of each operating cycle by the automatic controller shall include verification
that the process conditions established by the WD manufacturer as necessary and sufficient for
disinfection to take place (e.g. disinfectant concentration, temperature and contact time) were attained
(see also 5.5).Microbial testing (e.g. with biological indicators) of the disinfection stage on each cycle shall not be
used to meet this requirement.NOTE Confirmation of the concentration of disinfectant may require e.g. measurement of the volume of
disinfectant admitted together with a certificate of conformity for the concentration of the disinfectant supplied,
together with data to support the shelf life, expiry date etc.4.4.5 Disinfectant use
Whenever practicable the disinfectant solution should be discharged at the end of the process cycle.
If the WD is designed to allow the same disinfectant solution to be used on two or more consecutive
operating cycles then care shall be taken to ensure that the activity and safety (e.g. accumulation of
foreign material, device compatibility) of the disinfectant solution is not impaired during its working life.
This shall include the following:a) the WD manufacturer shall recommend the disinfectant(s) to be used;
b) the WD manufacturer shall specify the means which shall be used to ensure that the disinfectant
solution has retained the required anti-microbial activity. These means shall be based on
validation studies, which would normally be carried out by the disinfectant manufacturer, to
determine a suitable parameter, or parameters, which may be monitored to indicate the anti-
microbial activity of the disinfectant. Suitable parameters may include e.g. the concentration of
the active ingredient and adjuvants that may also affect performance;NOTE Minor changes in formulation of the disinfectant may have a significant effect on storage life, anti-
microbial activity etc..c) the WD manufacturer shall recommend to the user the maximum period or number of operating
cycles for which the disinfectant may be used. This shall be based on validated experimental
data.4.5 Final (Post-disinfection) rinsing
4.5.1 The chemical purity of the rinse water used after the disinfection stage shall be of, at least,
that specified for potable water in the WHO Guidelines for drinking water quality 1998 (see also
prEN ISO 15883-1:2003, 5.23).4.5.2 The rinse water shall, when tested, by the methods given in Annex A, be free from aerobic
mesophilic micro-organisms including legionella and mycobacteria at the point of discharge to the WD
chamber and the device(s) to be processed.NOTE 1 The following methods may be suitable for control of the microbial contamination of rinse water. The
rinse water should be: maintained in a dedicated reservoir at a temperature not less than 60°C for the time demonstrated to
achieve disinfection of the incoming supply, or disinfected immediately prior to use, or
filtered to remove suspended particles of a size greater than 0,2 mm, or
---------------------- Page: 12 ----------------------
prEN ISO 15883-4:2003 (E)
sterile, in a closed container, with connection to the WD designed and constructed to provide aseptic
transfer.NOTE 2 These requirements may be met either by appropriate provision within the WD or by specifying the
water treatment which should be provided external to the WD on the incoming water supply (see 4.8.2).
4.5.3 The connection between the water supply that has been treated to remove mic
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.