11.080.10 - Sterilizing equipment
ICS 11.080.10 Details
Sterilizing equipment
Sterilisiergerate. Desinfiziergerate
Matériel de stérilisation et de désinfection
Sterilizacijska oprema
General Information
e-Library Subscription
Create subscription and get permanent access to documents within 11.080.10 - Sterilizing equipment
Currently subscription includes documents marked with .We are working on making all documents available within the subscription.
This part of IEC 61010 specifies safety requirements for electrical equipment intended for
sterilization, washing, and disinfection of medical materials in the medical, veterinary,
pharmaceutical and laboratory fields, when used under the environmental conditions of 1.4.
Examples of such equipment include the following:
a) STERILIZERS and disinfectors using steam and/or hot water as the sterilant;
b) STERILIZERS and disinfectors using toxic gas, toxic aerosol or toxic vapour as the sterilant;
c) STERILIZERS and disinfectors using hot air or hot inert gas as the sterilant; and
d) WASHER-DISINFECTORS.
- Standard42 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard specifies requirements and the relevant tests for large steam sterilizers primarily used in health care for the sterilization of medical devices and their accessories contained in one or more sterilization modules. The test loads described in this European Standard are selected to represent the majority of loads (i.e. wrapped goods consisting of metal, rubber and porous materials) for the evaluation of general purpose steam sterilizers for medical devices. However, specific loads (e.g. heavy metal objects or long and/or narrow lumen) will require the use of other test loads.
This European Standard applies to steam sterilizers designed to accommodate at least one sterilization module or having a chamber volume of at least 60 l.
Large steam sterilizers can also be used during the commercial production of medical devices.
This European Standard does not specify requirements for equipment intended to use, contain or be exposed to flammable substances or substances which could cause combustion. This European Standard does not specify requirements for equipment intended to process biological waste or human tissues.
This European Standard does not describe a quality management system for the control of all stages of the manufacture of the sterilizer.
NOTE 1 Attention is drawn to the standards for quality management systems e.g. EN ISO 13485.
NOTE 2 Environmental aspects are addressed in Annex A.
- Standard118 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies procedures and test methods used to demonstrate the cleaning efficacy of washer-disinfectors (WD) and their accessories intended to be used for cleaning of reusable medical devices.
NOTE 1 The requirements can be used for washer-disinfectors intended for use with other articles used in the context of medical, dental, laboratory, pharmaceutical and veterinary practice.
NOTE 2 This document does not apply to the activities to be performed by the manufacturers of reusable medical devices.
- Standard68 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard specifies requirements and the relevant tests for large steam sterilizers primarily used in health care for the sterilization of medical devices and their accessories contained in one or more sterilization modules. The test loads described in this European Standard are selected to represent the majority of loads (i.e. wrapped goods consisting of metal, rubber and porous materials) for the evaluation of general purpose steam sterilizers for medical devices. However, specific loads (e.g. heavy metal objects or long and/or narrow lumen) will require the use of other test loads.
This European Standard applies to steam sterilizers designed to accommodate at least one sterilization module or having a chamber volume of at least 60 l.
Large steam sterilizers can also be used during the commercial production of medical devices.
This European Standard does not specify requirements for equipment intended to use, contain or be exposed to flammable substances or substances which could cause combustion. This European Standard does not specify requirements for equipment intended to process biological waste or human tissues.
This European Standard does not describe a quality management system for the control of all stages of the manufacture of the sterilizer.
NOTE 1 Attention is drawn to the standards for quality management systems e.g. EN ISO 13485.
NOTE 2 Environmental aspects are addressed in Annex A.
- Standard118 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies procedures and test methods used to demonstrate the cleaning efficacy of
washer-disinfectors (WD) and their accessories intended to be used for cleaning of reusable medical
devices.
NOTE 1 The requirements can be used for washer-disinfectors intended for use with other articles used in the
context of medical, dental, laboratory, pharmaceutical and veterinary practice.
NOTE 2 This document does not apply to the activities to be performed by the manufacturers of reusable
medical devices.
- Standard68 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies procedures and test methods used to demonstrate the cleaning efficacy of washer-disinfectors (WD) and their accessories intended to be used for cleaning of reusable medical devices. NOTE 1 The requirements can be used for washer-disinfectors intended for use with other articles used in the context of medical, dental, laboratory, pharmaceutical and veterinary practice. NOTE 2 This document does not apply to the activities to be performed by the manufacturers of reusable medical devices.
- Standard59 pagesEnglish languagesale 15% off
- Standard63 pagesFrench languagesale 15% off
This document specifies the particular requirements, including performance criteria for washer-disinfectors (WD) that are intended to be used for cleaning and chemical disinfection of thermolabile endoscopes.
This document also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which can be required to achieve the necessary performance criteria.
The methods, instrumentation and instructions required for type testing, works testing, validation (installation, operational and performance qualification on first installation), routine control and monitoring, and requalification of WD periodically and after essential repairs, are also specified.
NOTE 1 In addition, Annex A gives guidance on an appropriate division of responsibility for the range of activities covered by this document.
NOTE 2 WD complying with this document can also be used for cleaning and chemical disinfection of other thermolabile re-usable medical devices for which the device manufacturer has recommended and validated this method of disinfection.
WD complying with the requirements of this document are not intended for cleaning and disinfection of medical devices, including endoscopic accessories, which are heat stable and can be disinfected or sterilized by thermal methods (see ISO 15883-1:2006+Amd 1:2014, 4.1.5).
The specified performance requirements of this document do not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
NOTE 3 If it is considered that prion protein might be present, particular care is needed in the choice of cleaning agents and disinfectants to ensure that the chemicals used do not react with the prion protein and/or other protein in a manner that can inhibit its removal or inactivation from the load or washer-disinfector.
NOTE 4 This document can be used by prospective purchasers and manufacturers as the basis of agreement on the specification of the WD, manufacturers of endoscopes, cleaning products, and disinfecting products.
- Standard95 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the particular requirements, including performance criteria for washerdisinfectors
(WD) that are intended to be used for cleaning and chemical disinfection of thermolabile
endoscopes.
This document also specifies the performance requirements for the cleaning and disinfection of the
washer-disinfector and its components and accessories which can be required to achieve the necessary
performance criteria.
The methods, instrumentation and instructions required for type testing, works testing, validation
(installation, operational and performance qualification on first installation), routine control and
monitoring, and requalification of WD periodically and after essential repairs, are also specified.
NOTE 1 In addition, Annex A gives guidance on an appropriate division of responsibility for the range of
activities covered by this document.
NOTE 2 WD complying with this document can also be used for cleaning and chemical disinfection of other
thermolabile re-usable medical devices for which the device manufacturer has recommended and validated this
method of disinfection.
WD complying with the requirements of this document are not intended for cleaning and disinfection
of medical devices, including endoscopic accessories, which are heat stable and can be disinfected or
sterilized by thermal methods (see ISO 15883-1:2006+Amd 1:2014, 4.1.5).
The specified performance requirements of this document do not ensure the inactivation or removal of
the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
NOTE 3 If it is considered that prion protein might be present, particular care is needed in the choice of
cleaning agents and disinfectants to ensure that the chemicals used do not react with the prion protein and/or
other protein in a manner that can inhibit its removal or inactivation from the load or washer-disinfector.
NOTE 4 This document can be used by prospective purchasers and manufacturers as the basis of agreement
on the specification of the WD, manufacturers of endoscopes, cleaning products, and disinfecting products.
- Standard95 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard specifies the performance requirements and test methods for small steam sterilizers and sterilization cycles which are used for medical purposes or for materials that are likely to come into contact with blood or body fluids.
This European Standard applies to automatically controlled small steam sterilizers that generate steam using electrical heaters or use steam that is generated by a system external to the sterilizer.
This European Standard applies to small steam sterilizers used primarily for the sterilization of medical devices with a chamber volume of less than 60 l and unable to accommodate a sterilization module (300 mm × 300 mm × 600 mm).
The requirements concerning the quality management and risk management are addressed by other standards (e.g. EN ISO 13485, EN ISO 14971).
This European Standard does not apply to small steam sterilizers that are used to sterilize liquids or pharmaceutical products.
This European Standard does not specify safety requirements related to risks associated with the zone in which the sterilizer is used (e.g. flammable gases).
This European Standard does not specify requirements for the validation and routine control of sterilization by moist heat.
NOTE Requirements for the validation and routine control of sterilization by moist heat are given in EN ISO 17665-1.
This European Standard does not specify requirements for other sterilization processes that also employ moist heat as part of the process (i.e. formaldehyde, ethylene oxide).
- Standard84 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the particular requirements, including performance criteria for washer-disinfectors (WD) that are intended to be used for cleaning and chemical disinfection of thermolabile endoscopes. This document also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which can be required to achieve the necessary performance criteria. The methods, instrumentation and instructions required for type testing, works testing, validation (installation, operational and performance qualification on first installation), routine control and monitoring, and requalification of WD periodically and after essential repairs, are also specified. NOTE 1 In addition, Annex A gives guidance on an appropriate division of responsibility for the range of activities covered by this document. NOTE 2 WD complying with this document can also be used for cleaning and chemical disinfection of other thermolabile re-usable medical devices for which the device manufacturer has recommended and validated this method of disinfection. WD complying with the requirements of this document are not intended for cleaning and disinfection of medical devices, including endoscopic accessories, which are heat stable and can be disinfected or sterilized by thermal methods (see ISO 15883-1:2006+Amd 1:2014, 4.1.5). The specified performance requirements of this document do not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies. NOTE 3 If it is considered that prion protein might be present, particular care is needed in the choice of cleaning agents and disinfectants to ensure that the chemicals used do not react with the prion protein and/or other protein in a manner that can inhibit its removal or inactivation from the load or washer-disinfector. NOTE 4 This document can be used by prospective purchasers and manufacturers as the basis of agreement on the specification of the WD, manufacturers of endoscopes, cleaning products, and disinfecting products.
- Standard81 pagesEnglish languagesale 15% off
- Standard81 pagesEnglish languagesale 15% off
- Standard81 pagesFrench languagesale 15% off
- Standard81 pagesFrench languagesale 15% off
This European Standard specifies the performance requirements and test methods for small steam sterilizers and sterilization cycles which are used for medical purposes or for materials that are likely to come into contact with blood or body fluids.
This European Standard applies to automatically controlled small steam sterilizers that generate steam using electrical heaters or use steam that is generated by a system external to the sterilizer.
This European Standard applies to small steam sterilizers used primarily for the sterilization of medical devices with a chamber volume of less than 60 l and unable to accommodate a sterilization module (300 mm × 300 mm × 600 mm).
The requirements concerning the quality management and risk management are addressed by other standards (e.g. EN ISO 13485, EN ISO 14971).
This European Standard does not apply to small steam sterilizers that are used to sterilize liquids or pharmaceutical products.
This European Standard does not specify safety requirements related to risks associated with the zone in which the sterilizer is used (e.g. flammable gases).
This European Standard does not specify requirements for the validation and routine control of sterilization by moist heat.
NOTE Requirements for the validation and routine control of sterilization by moist heat are given in EN ISO 17665-1.
This European Standard does not specify requirements for other sterilization processes that also employ moist heat as part of the process (i.e. formaldehyde, ethylene oxide).
- Standard84 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-7:2016 specifies the particular requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of reusable items such as the following:
a) bedframes;
b) bedside tables;
c) transport carts;
d) containers;
e) surgical tables;
f) sterilization containers;
g) surgical clogs;
h) wheelchairs, aids for the disabled.
ISO 15883-7:2016 also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which may be necessary in order to achieve the required performance.
Devices identified within the scopes of ISO 15883‑2, ISO 15883‑3, ISO I5883-4, and ISO 15883‑6 do not fall within the scope of this part of ISO 15883.
In addition, the methods are specified, as well as instrumentation and instructions required for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control, and monitoring, as well as requalifications required to be carried out periodically and after essential repairs.
NOTE WDs corresponding to this part of ISO 15883 can also be used for cleaning and chemical disinfection of other thermolabile and reusable devices as recommended by the device manufacturer.
The performance requirements specified in this part of ISO 15883 may not ensure the inactivation or removal of the causative agent(s) (prion proteins) of Transmissible Spongiform Encephalopathies.
- Standard47 pagesEnglish languagesale 10% offe-Library read for1 day
This part of ISO 15883 specifies the particular requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of re-usable items such as:
a) bedframes;
b) bedside tables;
c) transport carts;
d) containers;
e) surgical tables;
f) sterilization containers;
g) surgical clogs;
h) wheelchairs, aids for the disabled.
This Part of ISO 15883 also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which may be necessary in order to achieve the required performance.
Devices identified within the Scopes of ISO 15883-2:2006, ISO 15883-3:2006, ISO I5883-4:2008, and ISO 15883-6:2011 do not fall within the scope of this part of ISO 15883.
In addition, the methods are specified as well as instrumentation and instructions required for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control and monitoring as well as re-validations required to be carried out periodically and after essential repairs.
NOTE WDs corresponding to this part of ISO 15883 can also be used for cleaning and chemical disinfection of other thermolabile and re-usable medical devices as recommended by the device manufacturer.
The performance requirements specified in this part of ISO 15883 may not ensure the inactivation or removal of the causative agent(s) (prion proteins) of Transmissible Spongiform Encephalopathies.
- Standard47 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-6:2011 specifies particular requirements for washer-disinfectors (WDs) intended for use when the level of assurance of disinfection that is necessary can be achieved by cleaning and thermal disinfection (A0 not less than 60) and does not require an independent automated record of critical processes to be kept. It is intended to be used in conjunction with ISO 15883-1, which gives general requirements for WDs.
The range of products on which WDs of this particular type can be used is restricted to devices and equipment which are non-invasive and non-critical (i.e. not penetrating skin or contacting mucosal surfaces).
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
It specifies safety requirements for electrical equipment intended for sterilization, washing, and disinfection of medical materials in the medical, veterinary, pharmaceutical and laboratory fields, when used under the environmental conditions of 1.4. Examples of such equipment include the following: a) sterilizers and disinfectors using steam, and/or hot water as the sterilant; b) sterilizers and disinfectors using toxic gas, toxic aerosol or toxic vapour as the sterilant; c) sterilizers and disinfectors using hot air or hot inert gas as the sterilant; and d) washer disinfectors.
- Standard40 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-7:2016 specifies the particular requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of reusable items such as the following: a) bedframes; b) bedside tables; c) transport carts; d) containers; e) surgical tables; f) sterilization containers; g) surgical clogs; h) wheelchairs, aids for the disabled. ISO 15883-7:2016 also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which may be necessary in order to achieve the required performance. Devices identified within the scopes of ISO 15883‑2, ISO 15883‑3, ISO I5883-4, and ISO 15883‑6 do not fall within the scope of this part of ISO 15883. In addition, the methods are specified, as well as instrumentation and instructions required for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control, and monitoring, as well as requalifications required to be carried out periodically and after essential repairs. NOTE WDs corresponding to this part of ISO 15883 can also be used for cleaning and chemical disinfection of other thermolabile and reusable devices as recommended by the device manufacturer. The performance requirements specified in this part of ISO 15883 may not ensure the inactivation or removal of the causative agent(s) (prion proteins) of Transmissible Spongiform Encephalopathies.
- Standard30 pagesEnglish languagesale 15% off
- Standard30 pagesEnglish languagesale 15% off
- Standard32 pagesFrench languagesale 15% off
ISO 15883-6:2011 specifies particular requirements for washer-disinfectors (WDs) intended for use when the level of assurance of disinfection that is necessary can be achieved by cleaning and thermal disinfection (A0 not less than 60) and does not require an independent automated record of critical processes to be kept. It is intended to be used in conjunction with ISO 15883-1, which gives general requirements for WDs.
The range of products on which WDs of this particular type can be used is restricted to devices and equipment which are non-invasive and non-critical (i.e. not penetrating skin or contacting mucosal surfaces).
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
2013-08-02 EMA: Draft for // vote received in ISO/CS (see notification of 2013-08-02 in dataservice).
2012-02-14 EMA: WI allocated to WG 8 as per TC secretary request.
2012-01-06 EMA: Draft for // ENQ received in ISO/CS (see notification of 2011-12-06 in dataservice).
- Amendment16 pagesEnglish languagesale 10% offe-Library read for1 day
EN 1422 specifies the requirements and the relevant tests for automatically controlled sterilizers employing ethylene oxide (EO) gas as the sterilant, either as a pure gas or a mixture with other gases, being used for the sterilization of medical devices and their accessories. This European Standard specifies requirements for ethylene oxide sterilizers (EO-sterilizers) working at super or sub-atmospheric pressure for: - the performance and design of sterilizers to ensure that the process is capable of sterilizing medical devices; - the equipment and controls of these sterilizers necessary for the validation and routine control of the sterilization processes. The test loads described in this European Standard are selected to represent a number of loads for the evaluation of the performance of EO sterilizers for medical devices. However, specific loads may require the use of other test loads.
- Standard49 pagesEnglish languagesale 10% offe-Library read for1 day
This part of ISO 15883 specifies general performance requirements for washer-disinfectors (WD) and their accessories that are intended to be used for cleaning and disinfection of re-usable medical devices and other articles used in the context of medical, dental, pharmaceutical and veterinary practice. It specifies performance requirements for cleaning and disinfection as well as for the accessories which can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and re-validation, periodically and after essential repairs, are also specified. The requirements for washer-disinfectors intended to process specific loads are specified in subsequent parts of this standard. For washer-disinfectors intended to process loads of two or more different types the requirements of all relevant parts of this standard apply. This part of ISO 15883 does not specify requirements intended for machines for use for laundry or general catering purposes. This part of ISO 15883 does not include requirements for machines which are intended to sterilize the load, or which are designated as "sterilizers", these are specified in other standards e.g. EN 285. The specified performance requirements of this standard may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies. This part of ISO 15883 may be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of compliance with the requirements of this part of ISO 15883 may also be employed by users to demonstrate continued compliance of the installed WD throughout its working life. Guidance on a routine test programme is given in Annex A.
- Amendment16 pagesEnglish languagesale 10% offe-Library read for1 day
EN 14180 specifies requirements and tests for LTSF sterilizers, which use a mixture of low temperature steam and formaldehyde as sterilizing agent, and which are working below ambient pressure only. These sterilizers are primarily used for the sterilization of heat labile medical devices in health care facilities. This European Standard specifies minimum requirements: - for the performance and design of sterilizers to ensure that the process is capable of sterilizing medical devices; - for the equipment and controls of these sterilizers necessary for the validation and routine control of the sterilization processes.
- Standard61 pagesEnglish languagesale 10% offe-Library read for1 day
- Standard6 pagesEnglish languagesale 15% off
- Standard6 pagesFrench languagesale 15% off
This European Standard specifies the requirements and the relevant tests for automatically controlled sterilizers employing ethylene oxide (EO) gas as the sterilant, either as a pure gas or a mixture with other gases, being used for the sterilization of medical devices and their accessories.
This European Standard specifies requirements for ethylene oxide sterilizers (EO-sterilizers) working at super or sub-atmospheric pressure for:
- the performance and design of sterilizers to ensure that the process is capable of sterilizing medical devices;
- the equipment and controls of these sterilizers necessary for the validation and routine control of the sterilization processes.
The test loads described in this European Standard are selected to represent a number of loads for the evaluation of the performance of EO sterilizers for medical devices. However, specific loads may require the use of other test loads.
This European Standard does not specify those tests which are necessary to determine the probability of a processed product being sterile, nor the routine quality control tests required prior to release of sterile product. These topics are addressed in prEN ISO 11135:2012.
This European Standard does not specify requirements for occupational safety associated with the design and operation of EO sterilization facilities.
NOTE 1 For further information on safety, see examples in the Bibliography. National or regional regulations can exist.
This European Standard does not cover sterilizers which employ the injection of EO or mixtures containing EO directly into packages or into a flexible chamber.
NOTE 2 See EN ISO 14937.
This European Standard is not intended as a checklist for suitability of an existing EO sterilizer when assessing compliance with prEN ISO 11135:2012. This standard is not intended to be applied retrospectively.
This European Standard does not cover analytical methods for determining levels of residual EO and/or its reaction products.
NOTE 3 For further information see ISO 10993 7.
- Standard49 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard specifies requirements and tests for LTSF sterilizers, which use a mixture of low temperature steam and formaldehyde as sterilizing agent, and which are working below ambient pressure only.
These sterilizers are primarily used for the sterilization of heat labile medical devices in health care facilities.
This European Standard specifies minimum requirements:
- for the performance and design of sterilizers to ensure that the process is capable of sterilizing medical devices;
- for the equipment and controls of these sterilizers necessary for the validation and routine control of the sterilization processes.
- Standard61 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-6:2011 specifies particular requirements for washer-disinfectors (WDs) intended for use when the level of assurance of disinfection that is necessary can be achieved by cleaning and thermal disinfection (A0 not less than 60) and does not require an independent automated record of critical processes to be kept. It is intended to be used in conjunction with ISO 15883-1, which gives general requirements for WDs. The range of products on which WDs of this particular type can be used is restricted to devices and equipment which are non-invasive and non-critical (i.e. not penetrating skin or contacting mucosal surfaces).
- Standard7 pagesEnglish languagesale 15% off
- Standard7 pagesFrench languagesale 15% off
- Standard12 pagesRussian languagesale 15% off
This part of ISO 15883 specifies particular requirements for washer-disinfectors (WD) that are intended for
use for the cleaning and thermal disinfection, in a single operating cycle, of re-usable medical devices such as
surgical instruments, anaesthetic equipment, bowls, dishes and receivers, utensils and glassware.
NOTE 1 Thermal disinfection can be achieved by rinsing the load with hot water, exposure to steam or combination of
the two.
The requirements specified in this part of ISO 15883 are applicable in conjunction with the general
requirements specified in ISO 15883-1.
The specified performance requirements of this part of ISO 15883 may not ensure the inactivation or removal
of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
NOTE 2 If it is considered that prion protein can be present, particular care is needed in the choice of disinfectants and
cleaning agents to ensure that the chemicals used do not react with the prion protein in a manner that may inhibit its
removal or inactivation.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
This part of ISO 15883 specifies general performance requirements for washer-disinfectors (WD) and their accessories that are intended to be used for cleaning and disinfection of re-usable medical devices and other articles used in the context of medical, dental, pharmaceutical and veterinary practice. It specifies performance requirements for cleaning and disinfection as well as for the accessories which can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and re-validation, periodically and after essential repairs, are also specified. The requirements for washer-disinfectors intended to process specific loads are specified in subsequent parts of this standard. For washer-disinfectors intended to process loads of two or more different types the requirements of all relevant parts of this standard apply. This part of ISO 15883 does not specify requirements intended for machines for use for laundry or general catering purposes. This part of ISO 15883 does not include requirements for machines which are intended to sterilize the load, or which are designated as "sterilizers", these are specified in other standards e.g. EN 285. The specified performance requirements of this standard may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies. This part of ISO 15883 may be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of compliance with the requirements of this part of ISO 15883 may also be employed by users to demonstrate continued compliance of the installed WD throughout its working life. Guidance on a routine test programme is given in Annex A.
- Standard84 pagesEnglish languagesale 10% offe-Library read for1 day
This part of ISO 15883 specifies particular requirements for washer-disinfectors (WD) that are intended to be
used for emptying, flushing, cleaning and thermal disinfection of containers used to hold human waste for
disposal by one operating cycle.
This part of ISO 15883 is to be applied in conjunction with ISO 15883-1.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-3:2006 specifies particular requirements for washer-disinfectors (WD) that are intended to be used for emptying, flushing, cleaning and thermal disinfection of containers used to hold human waste for disposal by one operating cycle.
ISO 15883-3:2006 is to be applied in conjunction with ISO 15883-1.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-1:2006 specifies general performance requirements for washer-disinfectors (WD) and their accessories that are intended to be used for cleaning and disinfection of re-usable medical devices and other articles used in the context of medical, dental, pharmaceutical and veterinary practice. It specifies performance requirements for cleaning and disinfection as well as for the accessories which can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and re-validation, periodically and after essential repairs, are also specified.
The requirements for washer-disinfectors intended to process specific loads are specified in subsequent parts of ISO 15883. For washer-disinfectors intended to process loads of two or more different types the requirements of all relevant parts of this standard apply.
ISO 15883-1:2006 does not specify requirements intended for machines for use for laundry or general catering purposes.
ISO 15883-1:2006 does not include requirements for machines which are intended to sterilize the load, or which are designated as "sterilizers", these are specified in other standards e.g. EN 285.
The specified performance requirements of this standard may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
ISO 15883-1:2006 may be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of compliance with the requirements of ISO 15883-1:2006 may also be employed by users to demonstrate continued compliance of the installed WD throughout its working life. Guidance on a routine test programme also is given.
- Standard84 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-2:2006 specifies particular requirements for washer disinfectors (WD) that are intended for use for the cleaning and thermal disinfection, in a single operating cycle, of re-usable medical devices such as surgical instruments, anaesthetic equipment, bowls, dishes and receivers, utensils and glassware.
The requirements in ISO 15883-2:2006 apply in addition to the general requirements specified in ISO 15883-1.
The specified performance requirements of ISO 15883-2:2006 may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-1:2006 specifies general performance requirements for washer-disinfectors (WD) and their accessories that are intended to be used for cleaning and disinfection of re-usable medical devices and other articles used in the context of medical, dental, pharmaceutical and veterinary practice. It specifies performance requirements for cleaning and disinfection as well as for the accessories which can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and re-validation, periodically and after essential repairs, are also specified. The requirements for washer-disinfectors intended to process specific loads are specified in subsequent parts of ISO 15883. For washer-disinfectors intended to process loads of two or more different types the requirements of all relevant parts of this standard apply. ISO 15883-1:2006 does not specify requirements intended for machines for use for laundry or general catering purposes. ISO 15883-1:2006 does not include requirements for machines which are intended to sterilize the load, or which are designated as "sterilizers", these are specified in other standards e.g. EN 285. The specified performance requirements of this standard may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies. ISO 15883-1:2006 may be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of compliance with the requirements of ISO 15883-1:2006 may also be employed by users to demonstrate continued compliance of the installed WD throughout its working life. Guidance on a routine test programme also is given.
- Standard70 pagesEnglish languagesale 15% off
- Standard71 pagesFrench languagesale 15% off
ISO 15883-3:2006 specifies particular requirements for washer-disinfectors (WD) that are intended to be used for emptying, flushing, cleaning and thermal disinfection of containers used to hold human waste for disposal by one operating cycle. ISO 15883-3:2006 is to be applied in conjunction with ISO 15883-1.
- Standard11 pagesEnglish languagesale 15% off
- Standard11 pagesFrench languagesale 15% off
ISO 15883-2:2006 specifies particular requirements for washer disinfectors (WD) that are intended for use for the cleaning and thermal disinfection, in a single operating cycle, of re-usable medical devices such as surgical instruments, anaesthetic equipment, bowls, dishes and receivers, utensils and glassware. The requirements in ISO 15883-2:2006 apply in addition to the general requirements specified in ISO 15883-1. The specified performance requirements of ISO 15883-2:2006 may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
- Standard10 pagesEnglish languagesale 15% off
- Standard11 pagesFrench languagesale 15% off
- Standard12 pagesRussian languagesale 15% off
This European Standard specifies performance criteria for steam sterilizer and autoclave used for the destruction and prevention of release of microorganisms used in biotechnological processes. This European Standard applies if the intended use of steam sterilizers or autoclaves includes hazardous or potentially hazardous microorganisms used in biotechological processes or if exposure of the worker or the environment to such microorganisms is restricted for reasons of safety.
- Standard13 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-7:2016 specifies the particular requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of reusable items such as the following:
a) bedframes;
b) bedside tables;
c) transport carts;
d) containers;
e) surgical tables;
f) sterilization containers;
g) surgical clogs;
h) wheelchairs, aids for the disabled.
ISO 15883-7:2016 also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which may be necessary in order to achieve the required performance.
Devices identified within the scopes of ISO 15883‑2, ISO 15883‑3, ISO I5883-4, and ISO 15883‑6 do not fall within the scope of this part of ISO 15883.
In addition, the methods are specified, as well as instrumentation and instructions required for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control, and monitoring, as well as requalifications required to be carried out periodically and after essential repairs.
NOTE WDs corresponding to this part of ISO 15883 can also be used for cleaning and chemical disinfection of other thermolabile and reusable devices as recommended by the device manufacturer.
The performance requirements specified in this part of ISO 15883 may not ensure the inactivation or removal of the causative agent(s) (prion proteins) of Transmissible Spongiform Encephalopathies.
- Draft42 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the performance requirements and test methods for small steam sterilizers and sterilization cycles which are used for medical purposes or for materials that are likely to come into contact with blood or body fluids.
This document applies to automatically controlled small steam sterilizers that generate steam using electrical heaters or use steam that is generated by a system external to the sterilizer.
This document applies to small steam sterilizers used primarily for the sterilization of medical devices with a chamber volume of less than 60 l and unable to accommodate a sterilization module (300 mm × 300 mm × 600 mm).
The requirements concerning the quality management and risk management are addressed by other standards (e.g. EN ISO 13485, EN ISO 14971).
This document does not apply to small steam sterilizers that are used to sterilize liquids or pharmaceutical products.
This document does not specify safety requirements related to risks associated with the zone in which the sterilizer is used (e.g. flammable gases).
This document does not specify requirements for the validation and routine control of sterilization by moist heat.
NOTE Requirements for the validation and routine control of sterilization by moist heat are given in EN ISO 17665.
This document does not specify requirements for other sterilization processes that also employ moist heat as part of the process (i.e. formaldehyde, ethylene oxide).
- Draft116 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements and tests for low temperature hydrogen peroxide sterilizers, which use a vaporized aqueous solution of hydrogen peroxide as the sterilizing agent.
These sterilizers are used for the sterilization of medical devices, particularly thermolabile medical devices.
This document specifies minimum requirements
• for the performance and design of sterilizers intended to deliver a process capable of sterilizing medical devices;
• for the equipment and controls of these sterilizers needed for operation, control and monitoring, and which can be used for validation of the sterilization processes;
• for the test equipment and test procedures used to verify the sterilizer performance specified by this document.
This document does not specify requirements for equipment intended to process liquids, biological waste, or human tissues.
This document does not describe a quality management system for the control of all stages of the manufacture of the sterilizer.
This document does not specify requirements and tests for decontamination systems for use in rooms, enclosures, or environmental spaces.
NOTE 1 Attention is drawn to the standards for quality management, e.g. EN ISO 13485.
NOTE 2 Environmental aspects of this standard are addressed in Annex H.
- Draft125 pagesEnglish languagesale 10% offe-Library read for1 day
1.1 This document specifies requirements and tests for LTSF sterilizers, which use a mixture of low temperature steam and formaldehyde as sterilizing agent, and which are working below ambient pressure only.
These sterilizers are primarily used for the sterilization of heat labile medical devices in health care facilities.
1.2 This document specifies minimum requirements:
- for the performance and design of sterilizers intended to deliver an LTSF-process capable of sterilizing medical devices;
- for the equipment and controls of these sterilizers needed for operation, control and monitoring of the sterilization processes, and which can be used for validation of the sterilization process.
1.3 This document specifies further test equipment and test procedures used to verify conformance of the equipment design and performance specified by this document.
1.4 This document does not specify requirements and tests for decontamination systems for use in rooms, enclosures, or environmental spaces.
- Draft119 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-2:2006 specifies particular requirements for washer disinfectors (WD) that are intended for use for the cleaning and thermal disinfection, in a single operating cycle, of re-usable medical devices such as surgical instruments, anaesthetic equipment, bowls, dishes and receivers, utensils and glassware.
The requirements in ISO 15883-2:2006 apply in addition to the general requirements specified in ISO 15883-1.
The specified performance requirements of ISO 15883-2:2006 may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
- Draft24 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-3:2006 specifies particular requirements for washer-disinfectors (WD) that are intended to be used for emptying, flushing, cleaning and thermal disinfection of containers used to hold human waste for disposal by one operating cycle.
ISO 15883-3:2006 is to be applied in conjunction with ISO 15883-1.
- Draft21 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-7:2016 specifies the particular requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of reusable items such as the following:
a) bedframes;
b) bedside tables;
c) transport carts;
d) containers;
e) surgical tables;
f) sterilization containers;
g) surgical clogs;
h) wheelchairs, aids for the disabled.
ISO 15883-7:2016 also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which may be necessary in order to achieve the required performance.
Devices identified within the scopes of ISO 15883‑2, ISO 15883‑3, ISO I5883-4, and ISO 15883‑6 do not fall within the scope of this part of ISO 15883.
In addition, the methods are specified, as well as instrumentation and instructions required for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control, and monitoring, as well as requalifications required to be carried out periodically and after essential repairs.
NOTE WDs corresponding to this part of ISO 15883 can also be used for cleaning and chemical disinfection of other thermolabile and reusable devices as recommended by the device manufacturer.
The performance requirements specified in this part of ISO 15883 may not ensure the inactivation or removal of the causative agent(s) (prion proteins) of Transmissible Spongiform Encephalopathies.
- Draft42 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the performance requirements and test methods for small steam sterilizers and sterilization cycles which are used for medical purposes or for materials that are likely to come into contact with blood or body fluids.
This document applies to automatically controlled small steam sterilizers that generate steam using electrical heaters or use steam that is generated by a system external to the sterilizer.
This document applies to small steam sterilizers used primarily for the sterilization of medical devices with a chamber volume of less than 60 l and unable to accommodate a sterilization module (300 mm × 300 mm × 600 mm).
The requirements concerning the quality management and risk management are addressed by other standards (e.g. EN ISO 13485, EN ISO 14971).
This document does not apply to small steam sterilizers that are used to sterilize liquids or pharmaceutical products.
This document does not specify safety requirements related to risks associated with the zone in which the sterilizer is used (e.g. flammable gases).
This document does not specify requirements for the validation and routine control of sterilization by moist heat.
NOTE Requirements for the validation and routine control of sterilization by moist heat are given in EN ISO 17665.
This document does not specify requirements for other sterilization processes that also employ moist heat as part of the process (i.e. formaldehyde, ethylene oxide).
- Draft116 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-2:2006 specifies particular requirements for washer disinfectors (WD) that are intended for use for the cleaning and thermal disinfection, in a single operating cycle, of re-usable medical devices such as surgical instruments, anaesthetic equipment, bowls, dishes and receivers, utensils and glassware.
The requirements in ISO 15883-2:2006 apply in addition to the general requirements specified in ISO 15883-1.
The specified performance requirements of ISO 15883-2:2006 may not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
- Draft24 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 15883-3:2006 specifies particular requirements for washer-disinfectors (WD) that are intended to be used for emptying, flushing, cleaning and thermal disinfection of containers used to hold human waste for disposal by one operating cycle.
ISO 15883-3:2006 is to be applied in conjunction with ISO 15883-1.
- Draft21 pagesEnglish languagesale 10% offe-Library read for1 day
1.1 This document specifies requirements and tests for LTSF sterilizers, which use a mixture of low temperature steam and formaldehyde as sterilizing agent, and which are working below ambient pressure only.
These sterilizers are primarily used for the sterilization of heat labile medical devices in health care facilities.
1.2 This document specifies minimum requirements:
- for the performance and design of sterilizers intended to deliver an LTSF-process capable of sterilizing medical devices;
- for the equipment and controls of these sterilizers needed for operation, control and monitoring of the sterilization processes, and which can be used for validation of the sterilization process.
1.3 This document specifies further test equipment and test procedures used to verify conformance of the equipment design and performance specified by this document.
1.4 This document does not specify requirements and tests for decontamination systems for use in rooms, enclosures, or environmental spaces.
- Draft119 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard specifies general performance requirements for washer- disinfectors and their accessories that are intended to be used for cleaning and disinfection of re-usable medical devices and other articles used in the context of medical, dental, pharmaceutical and veterinary practice. It specifies performance requirements for cleaning and disinfection as well as the accessories which may be required to achieve the necessary performance.
- Draft90 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements and tests for low temperature hydrogen peroxide sterilizers, which use a vaporized aqueous solution of hydrogen peroxide as the sterilizing agent.
These sterilizers are used for the sterilization of medical devices, particularly thermolabile medical devices.
This document specifies minimum requirements
• for the performance and design of sterilizers intended to deliver a process capable of sterilizing medical devices;
• for the equipment and controls of these sterilizers needed for operation, control and monitoring, and which can be used for validation of the sterilization processes;
• for the test equipment and test procedures used to verify the sterilizer performance specified by this document.
This document does not specify requirements for equipment intended to process liquids, biological waste, or human tissues.
This document does not describe a quality management system for the control of all stages of the manufacture of the sterilizer.
This document does not specify requirements and tests for decontamination systems for use in rooms, enclosures, or environmental spaces.
NOTE 1 Attention is drawn to the standards for quality management, e.g. EN ISO 13485.
NOTE 2 Environmental aspects of this standard are addressed in Annex H.
- Draft125 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard specifies requirements and the relevant tests for large steam sterilizers primarily used in health care for the sterilization of medical devices and their accessories contained in one or more sterilization modules. The test loads described in this European Standard are selected to represent the majority of loads (i.e. wrapped goods consisting of metal, rubber and porous materials) for the evaluation of general purpose steam sterilizers for medical devices. However, specific loads (e.g. heavy metal objects or long and/or narrow lumen) will require the use of other test loads.
This European Standard applies to steam sterilizers designed to accommodate at least one sterilization module or having a chamber volume of at least 60 l.
Large steam sterilizers can also be used during the commercial production of medical devices.
This European Standard does not specify requirements for equipment intended to use, contain or be exposed to flammable substances or substances which could cause combustion. This European Standard does not specify requirements for equipment intended to process biological waste or human tissues.
This European Standard does not describe a quality management system for the control of all stages of the manufacture of the sterilizer.
NOTE 1 Attention is drawn to the standards for quality management systems e.g. EN ISO 13485.
NOTE 2 Environmental aspects are addressed in Annex A.
- Draft4 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard specifies general performance requirements for washer- disinfectors and their accessories that are intended to be used for cleaning and disinfection of re-usable medical devices and other articles used in the context of medical, dental, pharmaceutical and veterinary practice. It specifies performance requirements for cleaning and disinfection as well as the accessories which may be required to achieve the necessary performance.
- Draft90 pagesEnglish languagesale 10% offe-Library read for1 day