Needle-based injection systems for medical use -- Requirements and test methods

Systèmes d'injection à aiguille pour usage médical -- Exigences et méthodes d'essai

General Information

Status
Published
Current Stage
5020 - FDIS ballot initiated: 2 months. Proof sent to secretariat
Start Date
06-Oct-2021
Completion Date
06-Oct-2021
Ref Project

Buy Standard

Draft
ISO/FDIS 11608-6:Version 24-jul-2021 - Needle-based injection systems for medical use -- Requirements and test methods
English language
20 pages
sale 15% off
Preview
sale 15% off
Preview
Draft
ISO/FDIS 11608-6 - Needle-based injection systems for medical use -- Requirements and test methods
English language
20 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (sample)

FINAL
INTERNATIONAL ISO/FDIS
DRAFT
STANDARD 11608-6
ISO/TC 84
Needle-based injection systems for
Secretariat: DS
medical use — Requirements and test
Voting begins on:
2021-07-30 methods —
Voting terminates on:
Part 6:
2021-09-24
On-body delivery systems
RECIPIENTS OF THIS DRAFT ARE INVITED TO
SUBMIT, WITH THEIR COMMENTS, NOTIFICATION
OF ANY RELEVANT PATENT RIGHTS OF WHICH
THEY ARE AWARE AND TO PROVIDE SUPPOR TING
DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS
Reference number
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/FDIS 11608-6:2021(E)
LOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT INTERNATIONAL STANDARDS MAY ON
OCCASION HAVE TO BE CONSIDERED IN THE
LIGHT OF THEIR POTENTIAL TO BECOME STAN-
DARDS TO WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS. ISO 2021
---------------------- Page: 1 ----------------------
ISO/FDIS 11608-6:2021(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2021

All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may

be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting

on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address

below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2021 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/FDIS 11608-6:2021(E)
Contents Page

Foreword ........................................................................................................................................................................................................................................iv

Introduction ..................................................................................................................................................................................................................................v

1 Scope ................................................................................................................................................................................................................................. 1

2 Normative references ...................................................................................................................................................................................... 1

3 Terms and definitions ..................................................................................................................................................................................... 1

4 Requirements .......................................................................................................................................................................................................... 3

4.1 General ........................................................................................................................................................................................................... 3

4.2 Risk assessment ..................................................................................................................................................................................... 3

4.3 Usability engineering ........................................................................................................................................................................ 3

4.4 Uncertainty of measurement and conformance with specifications........................................................ 3

4.5 General design requirements ..................................................................................................................................................... 3

4.6 Physical or mechanical requirements and test methods .................................................................................... 3

4.6.1 General...................................................................................................................................................................................... 3

4.6.2 Systems comprising rigid needles .................................................................................................................... 3

4.6.3 Systems comprising a soft cannula(s) ........................................................................................................... 3

4.6.4 Leakage from the OBDS ............................................................................................................................................. 3

4.6.5 Means of attachment .................................................................................................................................................... 4

4.6.6 Occlusion ................................................................................................................................................................................ 4

4.7 Functional performance requirements and test methods ................................................................................. 5

4.7.1 General...................................................................................................................................................................................... 5

4.7.2 Dosing requirements and methods ................................................................................................................. 5

4.7.3 Sharps injury protection ........................................................................................................................................... 6

4.7.4 Automated functions .................................................................................................................................................... 6

4.7.5 Injection depth and needle extension ........................................................................................................... 6

4.8 Biological requirements of the OBDS .................................................................................................................................. 7

4.8.1 Sterility of OBDS ............................................................................................................................................................... 7

4.8.2 Biocompatibility ............................................................................................................................................................... 7

4.9 Medicinal product compatibility ............................................................................................................................................. 7

4.9.1 General...................................................................................................................................................................................... 7

4.9.2 Particulates ........................................................................................................................................................................... 7

4.9.3 Pyrogenicity ......................................................................................................................................................................... 7

4.9.4 Extractable/leachables ............................................................................................................................................... 7

4.10 Electrical safety and software requirements ................................................................................................................ 7

4.10.1 Electrical safety ................................................................................................................................................................. 7

4.10.2 Software................................................................................................................................................................................... 7

5 Inspection .................................................................................................................................................................................................................... 7

6 Information supplied by the manufacturer ............................................................................................................................. 8

Annex A (informative) Test methods for adhesion ................................................................................................................................ 9

Annex B (informative) Dose delivery profiles ..........................................................................................................................................10

Annex C (informative) In vitro methods in relation to needle/cannula displacement .................................16

Bibliography .............................................................................................................................................................................................................................17

© ISO 2021 – All rights reserved iii
---------------------- Page: 3 ----------------------
ISO/FDIS 11608-6:2021(E)
Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards

bodies (ISO member bodies). The work of preparing International Standards is normally carried out

through ISO technical committees. Each member body interested in a subject for which a technical

committee has been established has the right to be represented on that committee. International

organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.

ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of

electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are

described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the

different types of ISO documents should be noted. This document was drafted in accordance with the

editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/ directives).

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of

any patent rights identified during the development of the document will be in the Introduction and/or

on the ISO list of patent declarations received (see www .iso .org/ patents).

Any trade name used in this document is information given for the convenience of users and does not

constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and

expressions related to conformity assessment, as well as information about ISO's adherence to the

World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www .iso .org/

iso/ foreword .html.

This document was prepared by Technical Committee ISO/TC 84, Devices for administration of medicinal

products and catheters.
A list of all parts in the ISO 11608 series can be found on the ISO website.

Any feedback or questions on this document should be directed to the user’s national standards body. A

complete listing of these bodies can be found at www .iso .org/ members .html.
iv © ISO 2021 – All rights reserved
---------------------- Page: 4 ----------------------
ISO/FDIS 11608-6:2021(E)
Introduction

The ISO 11608 series has traditionally addressed hand-held needle-based injection systems (NISs)

that are intended for parenteral administration by injection of medicinal products through a needle

to humans. These injections are performed manually, through exertion of force by the user, or

automatically through use of an internal power source through a needle into the patient’s tissue.

NOTE Although technically a device using a soft cannula is not “needle-based”, the cannula is placed by a

needle and can be included in this classification.

The user typically places the hand-held NIS at the injection site and holds the NIS in place until the

injection has completed. The intended use and delivery requirements of some medicinal products can

make manual manipulation and stabilization of a hand-held NIS during the medicinal product delivery

process impractical or impossible, and can result in an incomplete dose, missed dose, or user injury.

For example, it might not be appropriate, practical or possible for users to hold a NIS in place for an

extended period of time required by the volume or viscosity of the medicinal product or required to

preclude patient discomfort.

Delivery systems that are affixed to the body of the user eliminate some of the risks associated with

delivery of medicinal product through a traditional NIS. This document provides a consistent method

for evaluating the unique requirements and risks associated with these systems, herein referred to as

"on-body delivery systems" (OBDS).

Similarly to ISO 11608-1 and ISO 11608-5, this document will tend to specify the results of the design

effort instead of the physical and construction requirements used as the basis for OBDS design, so that

innovation in achieving the intended purposes is not unnecessarily restricted.

NISs governed by the ISO 11608 series are defined as “hand-held” or “on-body” delivery systems (OBDSs).

When hand-held, patients control and stabilize the NIS at the injection site during administration of a

discrete volume. Delivery times for this type of NIS would, therefore, be limited to avoid instability and

the potential for injection site trauma. For NISs with larger delivery volumes or physical properties

requiring a longer time to deliver, OBDS might be more practical. The OBDS would likely exist as either

“body-worn” (directly anchored to the body, e.g. using adhesive) or “patient-worn” (indirectly anchored,

e.g. catheter attached to OBDS contained in a backpack or pocket).

In either configuration, the time or speed employed to deliver a discrete volume would be based upon

patient tolerability or patient convenience rather than clinical relevance (e.g. medication efficacy) as

would be the case with insulin patch pumps or traditional infusion pumps associated with continuous

delivery (e.g. insulin).

This document only addresses the basic safety and performance of the product and manufacturers can

through risk assessments, identify additional requirements due to the unique nature of their specific

system or application.

The sampling plans for inspection selected for this document and outlined in ISO 11608-1 are intended

to verify the design, at a high confidence level. The sampling plan does not replace the more general

manufacturing quality systems, including lot release, which appear in International Standards on

quality systems, e.g. ISO 9001 or ISO 13485.

Guidance on transition periods for implementing the requirements of this document is given in

ISO/TR 19244.
© ISO 2021 – All rights reserved v
---------------------- Page: 5 ----------------------
FINAL DRAFT INTERNATIONAL STANDARD ISO/FDIS 11608-6:2021(E)
Needle-based injection systems for medical use —
Requirements and test methods —
Part 6:
On-body delivery systems
1 Scope

This document specifies requirements and test methods for On-Body Delivery Systems (OBDS) needle-

based injection systems (NISs) for single patient use, intended for subcutaneous, intramuscular

or intradermal delivery of a discrete volume (bolus) of medicinal product, through needles or soft

cannulas, incorporating pre-filled or user-filled, replaceable or non-replaceable containers.

NOTE 1 Although technically a device using a soft cannula is not “needle-based”, the soft cannula is placed by a

needle and can be included in this classification.

NOTE 2 Some requirements and methods are already established and included in other parts of the

ISO 11608 series.

Infusion pumps that are designed for continuous delivery at a specific rate required to achieve and/

or maintain a desired plasma medicinal product concentration are excluded from this document.

However, while this document is not intended to directly apply to these pump products, it does contain

requirements and test methods that can be used to help design and evaluate them.

NOTE 3 They are covered by IEC 60601-2-24 (if electronic) or ISO 28620 (if non-electronic).

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content

constitutes requirements of this document. For dated references, only the edition cited applies. For

undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 11608-1:2021, Needle-based injection systems for medical use - Requirements and test methods - Part 1:

Needle-based injection systems

ISO 11608-3:2021, Needle-based injection systems for medical use - Requirements and test methods -

Part 3: NIS containers and fluid paths

ISO 11608-4, Needle-based injection systems for medical use - Requirements and test methods - Part 4:

Needle-based injection systems containing electronics

ISO 11608-5, Needle-based injection systems for medical use - Requirements and test methods - Part 5:

Automated functions
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at http:// www .electropedia .org/
© ISO 2021 – All rights reserved 1
---------------------- Page: 6 ----------------------
ISO/FDIS 11608-6:2021(E)
3.1
body-worn on-body delivery system
body-worn OBDS
on-body delivery system (3.7) that is directly adhered to the skin
3.2
dose accuracy
measure of the total volume of medicinal product delivered to the patient
3.3
dose delivery time
measure of the total time over which the total dose volume is delivered
3.4
dose delivery profile

plot of the volumetric output of the on-body delivery system (3.7) against unit time throughout the

duration of the delivery
3.5
leakage

escape of medicinal product from the on-body delivery system (3.7) other than from the patient end of

the fluid path during delivery
3.6
needle extension

axial distance from the patient end of the needle or soft cannula tip to the nearest part of the on-body

delivery system body (3.8)

Note 1 to entry: See ISO 11608-5:2021, Figures C.1 to C.7 for 90-degree insertion and less than 90-degree

insertion.
3.7
on-body delivery system
OBDS

delivery system, which is affixed to the body of the user that actively delivers medicinal product, and

includes the medicinal product container and components for administration through a needle or soft

cannula
3.8
on-body delivery system body
OBDS body
defining the point of contact with the patient adjacent to the injection site
3.9
occlusion

blockage or closing of fluid path of on-body delivery system (3.7) during drug administration that is not

part of the intended use
3.10
patient tolerability

level to which pain, discomfort and other effects experienced during use of the on-body delivery system

(3.7) are accepted by patients
3.11
patient-worn on-body delivery system

on-body delivery system (3.7) that is attached over or under patient’s clothes, but not directly adhered to

the skin

Note 1 to entry: This type of injector is attached to the patient through tubing and a catheter.

2 © ISO 2021 – All rights reserved
---------------------- Page: 7 ----------------------
ISO/FDIS 11608-6:2021(E)
4 Requirements
4.1 General
The requirements of ISO 11608-1:2021, 5.1 shall apply.
4.2 Risk assessment
The requirements of ISO 11608-1:2021, 5.3 shall apply.

For OBDS, a Safety Assurance Case (SAC) (for example, developed using recommendations such as those

in AAMI/TIR 38) may be used to fulfil the ISO 14971 requirement for a risk management report.

4.3 Usability engineering
The requirements of ISO 11608-1:2021, 5.4 shall apply.
4.4 Uncertainty of measurement and conformance with specifications
The requirements of ISO 11608-1:2021, 5.5 shall apply.
4.5 General design requirements
Applicable requirements of ISO 11608-1:2021, 5.6 shall apply.
4.6 Physical or mechanical requirements and test methods
4.6.1 General

Unless otherwise specified, the testing shall be performed at standard atmosphere as specified in

ISO 11608-1. For OBDS, manufacturers shall determine the temperature range of the OBDS and drug

during delivery (which may be impacted by body temperature). In use testing, and testing of drug

compatibility should be completed throughout that temperature range.
4.6.2 Systems comprising rigid needles

In addition to the requirements in ISO 11608-3, additional physical and functional evaluations shall be

considered, e.g. flexural fatigue. Risk assessment shall be used to determine appropriate evaluations.

4.6.3 Systems comprising a soft cannula(s)
The requirements of ISO 11608-3 shall apply.
4.6.4 Leakage from the OBDS

The OBDS shall be inspected for leakage when tested in accordance with ISO 11608-1 in at least

worst-case environments and orientations representative of the expected use. Any amount of leakage

observed during testing shall be assessed for its ability to impact OBDS performance, fluid path or

medicinal product sterility, or risk to humans or the environment.

Any allowable leakage shall be addressed by risk assessment and appropriate information shall be

provided to the user in the instructions for use.

If any preconditioning creates the appearance of condensation or any other external evidence of fluid,

the manufacturer shall assess and confirm that this was not medicinal product leakage.

© ISO 2021 – All rights reserved 3
---------------------- Page: 8 ----------------------
ISO/FDIS 11608-6:2021(E)
4.6.5 Means of attachment

The developer of the OBDS shall establish performance criteria to ensure that the means of attachment

to the body is adequate to maintain a reliable medicinal product delivery pathway.

If the means of attachment uses adhesive, the following apply:

— attachment of the OBDS to the adhesive patch and the adhesive to the body are adequate to maintain

reliable medicinal product delivery pathway;

— if odour control, MVTR (moisture vapour transmission rate), water resistance or impermeability,

absorbency or conformity have been identified requirement for the OBDS, the adhesive material

should be tested for these properties.

Conduct adhesion/attachment tests when used as specified in the instructions for use. Where the OBDS

shall be maintained in a specific orientation, visually confirm that the OBDS maintains the required

orientation during testing. If an adhesive is used, see Annex A, which contains suggested test methods

for verifying the functional performance of an adhesive. It is up to the manufacturer to identify suitable

tests from those suggested or to develop their own test.

Testing of adhesion/attachment shall also measure performance of the OBDS under conditions

consistent with the intended use of the product, which may include exposure to typical fluids that can

be encountered during use (e.g. water, personal cleaning products, deodorants, skin lotion, medical

alcohols, perspiration) including the medicinal product, if appropriate. Adhesive tests may be performed

on material that has undergone sterilization or aging if it is anticipated through the risk analysis that

the adhesive property will be adversely affected by such conditioning. Based on risk assessment,

additional evidence might be required to demonstrate the performance and safety of the means of

attachment. This may include use on humans under simulated conditions of actual use (not necessarily

including the actual delivery). This evidence may also need to confirm that the adhesive bond between

the OBDS and the user does not cause unacceptable tissue trauma (as defined by risk assessment) or

create a bond that is too difficult for the user to remove. It is recognized that in some cases, reference to

existing clinical and/or other evidence may be sufficient to demonstrate performance of the means of

attachment.

The factors that affect Medical Adhesive Related Skin Injuries (MARSI) are complex and related to

factors that the OBDS manufacturer can control (adhesive properties, design of patch geometry,

suggested application points, etc.) and factors that the OBDS manufacturer cannot control (patient

[74]

population, removal technique, etc.). The selection of the adhesive should show due consideration

for the risk of MARSI to the end user balanced against the performance requirements of the OBDS,

applying risk control, where practicable.

Extended wear can result in the adhesive developing a stronger bond over time, which could impact the

patient. This should be considered and addressed in the testing.
4.6.6 Occlusion

The potential harm to the patient of a partial or complete occlusion resulting in a reduction or cessation

of delivery, or delivery of a fast bolus upon clearance of the occlusion, shall be determined, and the risk

based on the criticality of the medicinal product shall be addressed in the risk assessment. If required,

appropriate control(s) (design and/or indicator, instruction for use etc.) shall be implemented, which

may include a mechanism for the user to determine the ongoing status of the delivery (i.e., a delivery

indicator).

NOTE 1 Occlusion might lead to OBDS not meeting its primary function such as dose accuracy or delivery

time. The clearance after occlusion might lead to an instantaneous fast injection that might adversely impact the

patient.

NOTE 2 Delivery indication can be done by audible or tactile means or visually by an analogue or digital

indicator.
4 © ISO 2021 – All rights reserved
---------------------- Page: 9 ----------------------
ISO/FDIS 11608-6:2021(E)
4.7 Functional performance requirements and test methods
4.7.1 General

In addition to the conditioning specified in ISO 11608-1, manufacturers shall evaluate if simulating

additional conditions to which the OBDS is subjected as worn before and/or during delivery (e.g.

"normal/anticipated conditions" from ISO 11608-1) when testing primary functions is appropriate.

These additional test conditions shall be based on the risk analysis (e.g. due to the potential for extended

dose delivery time and warming of the OBDS while affixed to the body). Potential conditions to consider

include the following:
— vibration;
— temperature;
— humidity;
— atmospheric pressure;
— light exposure;
— orientation.

Each additional test shall be carried out at conditions that simulate the operation of the OBDS.

NOTE Primary functions can be able to be assessed during the same testing protocol and on one set of

samples.
4.7.2 Dosing requirements and methods
4.7.2.1 General
There are three measurements relevant to the dosing of OBDS:
— dose accuracy;
— dose delivery time;
— dose delivery profile.

At a minimum, dose accuracy is considered a primary function. Risk assessment shall determine

whether dose delivery time is considered a primary function, in accordance with ISO 11608-1.

4.7.2.2 Dose accuracy

The dose accuracy (dose delivered) shall be verified by measuring the total dose delivered. Where

the dose is specified as discontinuous dosing segments, dose accuracy shall be assessed for each dose

segment, including last dose accuracy (for variable dose OBDS) and dose delivery efficiency for user

filled OBDS.

If the OBDS is intended to be paused or stopped by the user (i.e., delivery volume during pause = 0),

then the dose accuracy testing at standard atmosphere conditions shall include this state to ensure that

the accuracy of the dose delivered shall not be adversely affected by any planned interruption (pause/

stop feature on OBDS) of the dose. Based on the risk assessment, the manufacturer shall determine if

assessment of the dose accuracy including the pause/stop feature is required after any additional pre-

conditionings besides standard.

Dose accuracy testing should be performed under conditions that simulate in vivo tissue back pressure,

if this is determined to be relevant through risk assessment.
© ISO 2021 – All rights reserved 5
---------------------- Page: 10 ----------------------
ISO/FDIS 11608-6:2021(E)
4.7.2.3 Dose delivery time

Dose delivery time is a measure of the time over which the total (or each if there are multiple dose

segments) dose is delivered.

Design verification of the required dose delivery time, determined by risk assessment, shall be

performed in accordance with ISO 11608-5.

The dose delivery time shall be verified by measuring the time over which the total dose is delivered.

Where the dose is specified as discontinuous dosing segments, the time of the delivery of each dose

segment shall be measured.

If the OBDS is intended to be paused or stopped by the user (i.e., delivery volume during pause = 0),

then the testing of the dose delivery time at standard atmosphere conditions shall include this state

to ensure that the time over which the dose is delivered is not adversely affected by any planned

interruption (pause/stop feature on OBDS) of the dose. Based on the risk assessment, the manufacturer

shall determine if assessment of the time of the dose delivered including the pause/stop feature is

required after any additional pre-conditionings besides standard.
4.7.2.4 Dose
...

FINAL
INTERNATIONAL ISO/FDIS
DRAFT
STANDARD 11608-6
ISO/TC 84
Needle-based injection systems for
Secretariat: DS
medical use — Requirements and test
Voting begins on:
2021-10-06 methods —
Voting terminates on:
Part 6:
2021-12-01
On-body delivery systems
RECIPIENTS OF THIS DRAFT ARE INVITED TO
SUBMIT, WITH THEIR COMMENTS, NOTIFICATION
OF ANY RELEVANT PATENT RIGHTS OF WHICH
THEY ARE AWARE AND TO PROVIDE SUPPOR TING
DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS
Reference number
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/FDIS 11608-6:2021(E)
LOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT INTERNATIONAL STANDARDS MAY ON
OCCASION HAVE TO BE CONSIDERED IN THE
LIGHT OF THEIR POTENTIAL TO BECOME STAN-
DARDS TO WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS. © ISO 2021
---------------------- Page: 1 ----------------------
ISO/FDIS 11608-6:2021(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2021

All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may

be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on

the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below

or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
© ISO 2021 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/FDIS 11608-6:2021(E)
Contents Page

Foreword ........................................................................................................................................................................................................................................iv

Introduction .................................................................................................................................................................................................................................v

1 Scope ................................................................................................................................................................................................................................. 1

2 Normative references ..................................................................................................................................................................................... 1

3 Terms and definitions .................................................................................................................................................................................... 1

4 Requirements .......................................................................................................................................................................................................... 3

4.1 General ........................................................................................................................................................................................................... 3

4.2 Risk assessment ..................................................................................................................................................................................... 3

4.3 Usability engineering ........................................................................................................................................... ............................. 3

4.4 Uncertainty of measurement and conformance with specifications ..................................................... 3

4.5 General design requirements .................................................................................................................................................... 3

4.6 Physical or mechanical requirements and test methods .................................................................................. 3

4.6.1 General ........................................................................................................................................................................................ 3

4.6.2 Systems comprising rigid needles ...................................................................................................................... 3

4.6.3 Systems comprising a soft cannula(s) ............................................................................................................. 3

4.6.4 Leakage from the OBDS ........................................................................................................................................... ..... 3

4.6.5 Means of attachment....................................................................................................................................................... 4

4.6.6 Occlusion ................................................................................................................................................................................... 4

4.7 Functional performance requirements and test methods .............................................................................. 5

4.7.1 General ........................................................................................................................................................................................ 5

4.7.2 Dosing requirements and methods ................................................................................................................... 5

4.7.3 Sharps injury protection ............................................................................................................................................. 6

4.7.4 Automated functions ...................................................................................................................................................... 6

4.7.5 Injection depth and needle extension ............................................................................................................. 6

4.8 Biological requirements of the OBDS ................................................................................................................................. 7

4.8.1 Sterility of OBDS ................................................................................................................................................................. 7

4.8.2 Biocompatibility ................................................................................................................................................................. 7

4.9 Medicinal product compatibility ............................................................................................................................................ 7

4.9.1 General ........................................................................................................................................................................................ 7

4.9.2 Particulates ............................................................................................................................................................................. 7

4.9.3 Pyrogenicity ........................................................................................................................................................................... 7

4.9.4 Extractable/leachables ................................................................................................................................................. 7

4.10 Electrical safety and software requirements ............................................................................................................. 7

4.10.1 Electrical safety .................................................................................................................................................................. 7

4.10.2 Software ..................................................................................................................................................................................... 7

5 Inspection .................................................................................................................................................................................................................... 7

6 Information supplied by the manufacturer ............................................................................................................................ 8

Annex A (informative) Test methods for adhesion ............................................................................................................................... 9

Annex B (informative) Dose delivery profiles ........................................................................................................................................10

Annex C (informative) In vitro methods in relation to needle/cannula displacement ...............................16

Bibliography .............................................................................................................................................................................................................................17

iii
© ISO 2021 – All rights reserved
---------------------- Page: 3 ----------------------
ISO/FDIS 11608-6:2021(E)
Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards

bodies (ISO member bodies). The work of preparing International Standards is normally carried out

through ISO technical committees. Each member body interested in a subject for which a technical

committee has been established has the right to be represented on that committee. International

organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.

ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of

electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are

described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the

different types of ISO documents should be noted. This document was drafted in accordance with the

editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of

any patent rights identified during the development of the document will be in the Introduction and/or

on the ISO list of patent declarations received (see www.iso.org/patents).

Any trade name used in this document is information given for the convenience of users and does not

constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and

expressions related to conformity assessment, as well as information about ISO's adherence to

the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see

www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 84, Devices for administration of medicinal

products and catheters.
A list of all parts in the ISO 11608 series can be found on the ISO website.

Any feedback or questions on this document should be directed to the user’s national standards body. A

complete listing of these bodies can be found at www.iso.org/members.html.
© ISO 2021 – All rights reserved
---------------------- Page: 4 ----------------------
ISO/FDIS 11608-6:2021(E)
Introduction

The ISO 11608 series has traditionally addressed hand-held needle-based injection systems (NISs)

that are intended for parenteral administration by injection of medicinal products through a needle

to humans. These injections are performed manually, through exertion of force by the user, or

automatically through use of an internal power source through a needle into the patient’s tissue.

NOTE Although technically a device using a soft cannula is not “needle-based”, the cannula is placed by a

needle and can be included in this classification.

The user typically places the hand-held NIS at the injection site and holds the NIS in place until the

injection has completed. The intended use and delivery requirements of some medicinal products can

make manual manipulation and stabilization of a hand-held NIS during the medicinal product delivery

process impractical or impossible, and can result in an incomplete dose, missed dose, or user injury.

For example, it might not be appropriate, practical or possible for users to hold a NIS in place for an

extended period of time required by the volume or viscosity of the medicinal product or required to

preclude patient discomfort.

Delivery systems that are affixed to the body of the user eliminate some of the risks associated with

delivery of medicinal product through a traditional NIS. This document provides a consistent method

for evaluating the unique requirements and risks associated with these systems, herein referred to as

"on-body delivery systems" (OBDS).

Similarly to ISO 11608-1 and ISO 11608-5, this document will tend to specify the results of the design

effort instead of the physical and construction requirements used as the basis for OBDS design, so that

innovation in achieving the intended purposes is not unnecessarily restricted.

NISs governed by the ISO 11608 series are defined as “hand-held” or “on-body” delivery systems (OBDSs).

When hand-held, patients control and stabilize the NIS at the injection site during administration of a

discrete volume. Delivery times for this type of NIS would, therefore, be limited to avoid instability and

the potential for injection site trauma. For NISs with larger delivery volumes or physical properties

requiring a longer time to deliver, OBDS might be more practical. The OBDS would likely exist as either

“body-worn” (directly anchored to the body, e.g. using adhesive) or “patient-worn” (indirectly anchored,

e.g. catheter attached to OBDS contained in a backpack or pocket).

In either configuration, the time or speed employed to deliver a discrete volume would be based upon

patient tolerability or patient convenience rather than clinical relevance (e.g. medication efficacy) as

would be the case with insulin patch pumps or traditional infusion pumps associated with continuous

delivery (e.g. insulin).

This document only addresses the basic safety and performance of the product and manufacturers can

through risk assessments, identify additional requirements due to the unique nature of their specific

system or application.

The sampling plans for inspection selected for this document and outlined in ISO 11608-1 are intended

to verify the design, at a high confidence level. The sampling plan does not replace the more general

manufacturing quality systems, including lot release, which appear in International Standards on

quality systems, e.g. ISO 9001 or ISO 13485.
© ISO 2021 – All rights reserved
---------------------- Page: 5 ----------------------
FINAL DRAFT INTERNATIONAL STANDARD ISO/FDIS 11608-6:2021(E)
Needle-based injection systems for medical use —
Requirements and test methods —
Part 6:
On-body delivery systems
1 Scope

This document specifies requirements and test methods for On-Body Delivery Systems (OBDS) needle-

based injection systems (NISs) for single patient use, intended for subcutaneous, intramuscular

or intradermal delivery of a discrete volume (bolus) of medicinal product, through needles or soft

cannulas, incorporating pre-filled or user-filled, replaceable or non-replaceable containers.

NOTE 1 Although technically a device using a soft cannula is not “needle-based”, the soft cannula is placed by a

needle and can be included in this classification.

NOTE 2 Some requirements and methods are already established and included in other parts of the

ISO 11608 series.

Infusion pumps that are designed for continuous delivery at a specific rate required to achieve and/

or maintain a desired plasma medicinal product concentration are excluded from this document.

However, while this document is not intended to directly apply to these pump products, it does contain

requirements and test methods that can be used to help design and evaluate them.

NOTE 3 They are covered by IEC 60601-2-24 (if electronic) or ISO 28620 (if non-electronic).

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content

constitutes requirements of this document. For dated references, only the edition cited applies. For

undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 11608-1:2021, Needle-based injection systems for medical use - Requirements and test methods - Part

1: Needle-based injection systems

ISO 11608-3:2021, Needle-based injection systems for medical use - Requirements and test methods - Part

3: NIS containers and fluid paths

ISO 11608-4, Needle-based injection systems for medical use - Requirements and test methods - Part 4:

Needle-based injection systems containing electronics

ISO 11608-5, Needle-based injection systems for medical use - Requirements and test methods - Part 5:

Automated functions
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at http:// www .electropedia .org/
© ISO 2021 – All rights reserved
---------------------- Page: 6 ----------------------
ISO/FDIS 11608-6:2021(E)
3.1
body-worn on-body delivery system
body-worn OBDS
on-body delivery system (3.7) that is directly adhered to the skin
3.2
dose accuracy
measure of the total volume of medicinal product delivered to the patient
3.3
dose delivery time
measure of the total time over which the total dose volume is delivered
3.4
dose delivery profile

plot of the volumetric output of the on-body delivery system (3.7) against unit time throughout the

duration of the delivery
3.5
leakage

escape of medicinal product from the on-body delivery system (3.7) other than from the patient end of

the fluid path during delivery
3.6
needle extension

axial distance from the patient end of the needle or soft cannula tip to the nearest part of the on-body

delivery system body (3.8)

Note 1 to entry: See ISO 11608-5:2021, Figures C.1 to C.7 for 90-degree insertion and less than 90-degree

insertion.
3.7
on-body delivery system
OBDS

delivery system, which is affixed to the body of the user that actively delivers medicinal product, and

includes the medicinal product container and components for administration through a needle or soft

cannula
3.8
on-body delivery system body
OBDS body
defining the point of contact with the patient adjacent to the injection site
3.9
occlusion

blockage or closing of fluid path of on-body delivery system (3.7) during drug administration that is not

part of the intended use
3.10
patient tolerability

level to which pain, discomfort and other effects experienced during use of the on-body delivery system

(3.7) are accepted by patients
3.11
patient-worn on-body delivery system

on-body delivery system (3.7) that is attached over or under patient’s clothes, but not directly adhered to

the skin

Note 1 to entry: This type of injector is attached to the patient through tubing and a catheter.

© ISO 2021 – All rights reserved
---------------------- Page: 7 ----------------------
ISO/FDIS 11608-6:2021(E)
4 Requirements
4.1 General
The requirements of ISO 11608-1:2021, 5.1 shall apply.
4.2 Risk assessment
The requirements of ISO 11608-1:2021, 5.3 shall apply.

For OBDS, a Safety Assurance Case (SAC) (for example, developed using recommendations such as those

in AAMI/TIR 38) may be used to fulfil the ISO 14971 requirement for a risk management report.

4.3 Usability engineering
The requirements of ISO 11608-1:2021, 5.4 shall apply.
4.4 Uncertainty of measurement and conformance with specifications
The requirements of ISO 11608-1:2021, 5.5 shall apply.
4.5 General design requirements
Applicable requirements of ISO 11608-1:2021, 5.6 shall apply.
4.6 Physical or mechanical requirements and test methods
4.6.1 General

Unless otherwise specified, the testing shall be performed at standard atmosphere as specified in

ISO 11608-1. For OBDS, manufacturers shall determine the temperature range of the OBDS and drug

during delivery (which may be impacted by body temperature). In use testing, and testing of drug

compatibility should be completed throughout that temperature range.
4.6.2 Systems comprising rigid needles

In addition to the requirements in ISO 11608-3, additional physical and functional evaluations shall be

considered, e.g. flexural fatigue. Risk assessment shall be used to determine appropriate evaluations.

4.6.3 Systems comprising a soft cannula(s)
The requirements of ISO 11608-3 shall apply.
4.6.4 Leakage from the OBDS

The OBDS shall be inspected for leakage when tested in accordance with ISO 11608-1 in at least

worst-case environments and orientations representative of the expected use. Any amount of leakage

observed during testing shall be assessed for its ability to impact OBDS performance, fluid path or

medicinal product sterility, or risk to humans or the environment.

Any allowable leakage shall be addressed by risk assessment and appropriate information shall be

provided to the user in the instructions for use.

If any preconditioning creates the appearance of condensation or any other external evidence of fluid,

the manufacturer shall assess and confirm that this was not medicinal product leakage.

© ISO 2021 – All rights reserved
---------------------- Page: 8 ----------------------
ISO/FDIS 11608-6:2021(E)
4.6.5 Means of attachment

The developer of the OBDS shall establish performance criteria to ensure that the means of attachment

to the body is adequate to maintain a reliable medicinal product delivery pathway.

If the means of attachment uses adhesive, the following apply:

— attachment of the OBDS to the adhesive patch and the adhesive to the body are adequate to maintain

reliable medicinal product delivery pathway;

— if odour control, MVTR (moisture vapour transmission rate), water resistance or impermeability,

absorbency or conformity have been identified requirement for the OBDS, the adhesive material

should be tested for these properties.

Conduct adhesion/attachment tests when used as specified in the instructions for use. Where the OBDS

shall be maintained in a specific orientation, visually confirm that the OBDS maintains the required

orientation during testing. If an adhesive is used, see Annex A, which contains suggested test methods

for verifying the functional performance of an adhesive. It is up to the manufacturer to identify suitable

tests from those suggested or to develop their own test.

Testing of adhesion/attachment shall also measure performance of the OBDS under conditions

consistent with the intended use of the product, which may include exposure to typical fluids that can

be encountered during use (e.g. water, personal cleaning products, deodorants, skin lotion, medical

alcohols, perspiration) including the medicinal product, if appropriate. Adhesive tests may be performed

on material that has undergone sterilization or aging if it is anticipated through the risk analysis that

the adhesive property will be adversely affected by such conditioning. Based on risk assessment,

additional evidence might be required to demonstrate the performance and safety of the means of

attachment. This may include use on humans under simulated conditions of actual use (not necessarily

including the actual delivery). This evidence may also need to confirm that the adhesive bond between

the OBDS and the user does not cause unacceptable tissue trauma (as defined by risk assessment) or

create a bond that is too difficult for the user to remove. It is recognized that in some cases, reference to

existing clinical and/or other evidence may be sufficient to demonstrate performance of the means of

attachment.

The factors that affect Medical Adhesive Related Skin Injuries (MARSI) are complex and related to

factors that the OBDS manufacturer can control (adhesive properties, design of patch geometry,

suggested application points, etc.) and factors that the OBDS manufacturer cannot control (patient

[73]

population, removal technique, etc.). The selection of the adhesive should show due consideration

for the risk of MARSI to the end user balanced against the performance requirements of the OBDS,

applying risk control, where practicable.

Extended wear can result in the adhesive developing a stronger bond over time, which could impact the

patient. This should be considered and addressed in the testing.
4.6.6 Occlusion

The potential harm to the patient of a partial or complete occlusion resulting in a reduction or cessation

of delivery, or delivery of a fast bolus upon clearance of the occlusion, shall be determined, and the risk

based on the criticality of the medicinal product shall be addressed in the risk assessment. If required,

appropriate control(s) (design and/or indicator, instruction for use etc.) shall be implemented, which

may include a mechanism for the user to determine the ongoing status of the delivery (i.e., a delivery

indicator).

NOTE 1 Occlusion might lead to OBDS not meeting its primary function such as dose accuracy or delivery

time. The clearance after occlusion might lead to an instantaneous fast injection that might adversely impact the

patient.

NOTE 2 Delivery indication can be done by audible or tactile means or visually by an analogue or digital

indicator.
© ISO 2021 – All rights reserved
---------------------- Page: 9 ----------------------
ISO/FDIS 11608-6:2021(E)
4.7 Functional performance requirements and test methods
4.7.1 General

In addition to the conditioning specified in ISO 11608-1, manufacturers shall evaluate if simulating

additional conditions to which the OBDS is subjected as worn before and/or during delivery (e.g.

"normal/anticipated conditions" from ISO 11608-1) when testing primary functions is appropriate.

These additional test conditions shall be based on the risk analysis (e.g. due to the potential for extended

dose delivery time and warming of the OBDS while affixed to the body). Potential conditions to consider

include the following:
— vibration;
— temperature;
— humidity;
— atmospheric pressure;
— light exposure;
— orientation.

Each additional test shall be carried out at conditions that simulate the operation of the OBDS.

NOTE Primary functions can be able to be assessed during the same testing protocol and on one set of

samples.
4.7.2 Dosing requirements and methods
4.7.2.1 General
There are three measurements relevant to the dosing of OBDS:
— dose accuracy;
— dose delivery time;
— dose delivery profile.

At a minimum, dose accuracy is considered a primary function. Risk assessment shall determine

whether dose delivery time is considered a primary function, in accordance with ISO 11608-1.

4.7.2.2 Dose accuracy

The dose accuracy (dose delivered) shall be verified by measuring the total dose delivered. Where

the dose is specified as discontinuous dosing segments, dose accuracy shall be assessed for each dose

segment, including last dose accuracy (for variable dose OBDS) and dose delivery efficiency for user

filled OBDS.

If the OBDS is intended to be paused or stopped by the user (i.e., delivery volume during pause = 0),

then the dose accuracy testing at standard atmosphere conditions shall include this state to ensure that

the accuracy of the dose delivered shall not be adversely affected by any planned interruption (pause/

stop feature on OBDS) of the dose. Based on the risk assessment, the manufacturer shall determine if

assessment of the dose accuracy including the pause/stop feature is required after any additional pre-

conditionings besides standard.

Dose accuracy testing should be performed under conditions that simulate in vivo tissue back pressure,

if this is determined to be relevant through risk assessment.
© ISO 2021 – All rights reserved
---------------------- Page: 10 ----------------------
ISO/FDIS 11608-6:2021(E)
4.7.2.3 Dose delivery time

Dose delivery time is a measure of the time over which the total (or each if there are multiple dose

segments) dose is delivered.

Design verification of the required dose delivery time, determined by risk assessment, shall be

performed in accordance with ISO 11608-5.

The dose delivery time shall be verified by measuring the time over which the total dose is delivered.

Where the dose is specified as discontinuous dosing segments, the time of the delivery of each dose

segment shall be measured.

If the OBDS is intended to be paused or stopped by the user (i.e., delivery volume during pause = 0),

then the testing of the dose delivery time at standard atmosphere conditions shall include this state

to ensure that the time over which the dose is delivered is not adversely affected by any planned

interruption (pause/stop feature on OBDS) of the dose. Based on the risk assessment, the manufacturer

shall determine if assessment of the time of the dose delivered including the pause/stop feature is

required after any additional pre-conditionings besides standard.
4.7.2.4 Dose delivery profile
[6]

In instances where the dose delivery profile is determined to be clinically relevant , the manu

...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.