SIST EN 60601-2-4:2011/A2:2027

English language (20 pages)
Preview
Preview
e-Library read for
×1 day

SIST EN 60601-2-4:2011/A2:2027 is the Slovenian adoption of Amendment 2 to the IEC cardiac defibrillator standard for medical electrical equipment, published by SIST. It updates the particular requirements for the basic safety and essential performance of cardiac defibrillators, including manual defibrillators and automated external defibrillators (AED). The document is used by manufacturers, test laboratories, and conformity assessors who need to verify defibrillator safety, performance, labelling, and electromagnetic compatibility.

What does SIST EN 60601-2-4:2011/A2:2027 specify?

SIST EN 60601-2-4:2011/A2:2027 specifies amendments to the particular requirements for cardiac defibrillators in the IEC 60601 series. It updates the requirements that apply alongside the general standard and the EMC collateral standard, with a focus on safety, charging time, battery performance, alarms, rhythm recognition, and immunity testing.

The document is organized as clause updates to 201.1, 201.2, 201.3, 201.7.2.103, 201.8.5.3, 201.8.5.5.1, 201.12.3, 201.101 to 201.107, and clause 202, with explanatory material in Annex AA and an index of defined terms in Annex BB.

AnnexWhat it covers
Annex AAExplanatory notes for alarm systems, charging time, rhythm recognition detector performance, and railway-environment considerations
Annex BBIndex of defined terms used in this particular standard

What are the key requirements of SIST EN 60601-2-4:2011/A2:2027?

SIST EN 60601-2-4:2011/A2:2027 tightens several performance checks for defibrillators and makes the test conditions more specific. The practical effect is that manufacturers must show that the device still performs correctly under low battery, mains variation, EMC exposure, and worst-case operating modes.

Charging time for manual defibrillators and AEDs

Clause 201.101 gives separate charging-time requirements for frequent-use and infrequent-use manual defibrillators, and for frequent-use and infrequent-use AEDs. The measurements are taken from discharge, disarming, activation of the rhythm recognition detector (RRD), or power-on/programming mode to readiness for discharge at maximum energy, depending on the case. In practice, this means the device must be quick enough to use in a rescue situation even when batteries are depleted or mains voltage is reduced.

Battery capacity at 0 °C

Clause 201.102 requires a new, fully charged battery to support at least 20 defibrillation discharges at 0 °C at the maximum delivered energy. The cycles and timing differ for manual defibrillators and AEDs, and the AED requirement follows the pre-programmed defibrillation sequence. This matters because cold environments and repeated shocks can expose weak battery designs.

Alarm systems and risk management

Clause 201.12.3 links alarm requirements to risk management. When alarms are needed to warn the operator about unsafe patient or equipment states, the device has to follow the relevant alarm standard, and any deviation has to be validated. In practice, manufacturers must justify alarm choices against the use environment and the operator’s training level.

Rhythm recognition detector performance

Clause 201.107 requires manufacturers who specify artifact handling to state sensitivity and specificity, identify the artifact type, and explain how it is measured. The overall RRD performance must still meet the clause requirements, with risk management used to identify hazards from rhythm detection in the presence of artifacts. This is important for AEDs used during movement or CPR-like disturbances.

EMC emissions and immunity

Clause 202 requires compliance with CISPR 11 emissions requirements in all configurations and operating modes, and adds pass/fail criteria for immunity. These include no inappropriate energy therapy, no unsafe failure mode, no unintentional state change, no loss of stored data, and readable therapy-relevant information. For users, this means the defibrillator has to remain safe and predictable in real electromagnetic environments.

Special environments and test conditions

The EMC clause also adds proximity-field immunity and railway-related considerations, and identifies trains, train stations, and similar settings as a special environment with disturbance degree X. Test methods reference the IEC 61000 series, and radiated RF testing uses a simulated patient load on the electrodes. In practice, manufacturers need test plans that reflect where the device will actually be used.

What terms does SIST EN 60601-2-4:2011/A2:2027 define?

Alarm system - a system used to alert the operator to hazardous or unsatisfactory conditions of the equipment or patient state.

Home healthcare environment - the environment defined in IEC 60601-1-11 for medical electrical equipment used outside conventional clinical settings.

Electromagnetic environment (EM environment) - the surrounding electromagnetic conditions that can affect the defibrillator’s operation.

Electrostatic discharge (ESD) - a sudden transfer of static electricity that can disturb the device and must be covered by immunity testing.

Risk management - the process used to identify hazards, choose risk controls, and judge whether the remaining risk is acceptable.

Rhythm recognition detector (RRD) - the function that analyses a patient rhythm to decide whether shock delivery is appropriate.

Essential performance - the performance necessary to avoid unacceptable risk, especially during therapy-related operation.

Who uses SIST EN 60601-2-4:2011/A2:2027?

SIST EN 60601-2-4:2011/A2:2027 is used by defibrillator manufacturers, especially makers of frequent-use and infrequent-use manual defibrillators and AEDs. Test laboratories use it to design charge-time, battery, and EMC tests. Conformity assessment bodies and regulatory teams use it to check whether a product’s safety, labelling, and performance claims match the standard.

It is also relevant to hospitals, emergency-care providers, and responsible organizations that select, procure, or maintain defibrillators for professional healthcare facilities, home healthcare use, or special environments such as railway settings.

What changed in SIST EN 60601-2-4:2011/A2:2027 from the previous edition?

SIST EN 60601-2-4:2011/A2:2027 is Amendment 2 to IEC 60601-2-4:2010. The amendment aligns the standard with IEC 60601-1:2005+A1:2012+A2:2020 and IEC 60601-1-2:2014+A1:2020, and updates requirements for:

  • normative references
  • disposable defibrillator electrode labelling
  • maximum mains voltage treatment
  • defibrillation protection
  • alarm systems
  • charging time for manual defibrillators and AEDs
  • battery capacity requirements
  • RRD requirements
  • EMC emissions and immunity
  • railway-environment considerations

Which standards are used with SIST EN 60601-2-4:2011/A2:2027?

The amendment uses IEC 60601-1:2005+A1:2012+A2:2020 as the general basic-safety and essential-performance standard, and IEC 60601-1-2:2014+A1:2020 for EMC requirements and tests. It also uses IEC 60601-1-11:2015+A1:2020 for the home healthcare environment, IEC 60601-1-8:2006+A1:2012+A2:2020 for alarm systems, and ISO 15223-1:2021 for symbols and labelling.

For EMC work, the standard points to CISPR 11, CISPR 16-2-3, and IEC 61000-4-2, -4-3, -4-4, -4-5, and -4-8, with IEC 61000-2-5 used for the railway disturbance context.

What does the SIST EN 60601-2-4:2011/A2:2027 document contain?

The document contains the amended clause text for safety, performance, labelling, and EMC, plus detailed test conditions and measurement approaches. It includes timing tests for charging readiness, battery endurance demonstrations at 0 °C, and the specific operating states used for testing manual defibrillators and AEDs.

It also contains EMC test conditions for radiated RF fields, electrostatic discharge, fast transients, surges, conducted RF disturbances, and proximity fields from wireless communications equipment. Annex AA explains the intent of several requirements, including the railway-environment treatment. Annex BB updates the index of defined terms used in the particular standard.

Relations

Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
27-Oct-2026
Effective Date
17-Nov-2026
Effective Date
17-Nov-2026
Effective Date
17-Nov-2026
Effective Date
01-Jan-2027

Buy Documents

Amendment

SIST EN 60601-2-4:2011/A2:2027

English language (20 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

SIST EN 60601-2-4:2011/A2:2027 is a amendment published by the Slovenian Institute for Standardization (SIST). Its full title is "Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators (IEC 60601-2-4:2010/AMD2:2026)". This standard covers: Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators (IEC 60601-2-4:2010/AMD2:2026)

Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators (IEC 60601-2-4:2010/AMD2:2026)

SIST EN 60601-2-4:2011/A2:2027 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN 60601-2-4:2011/A2:2027 has the following relationships with other standards: It is inter standard links to SIST EN 60601-1-2:2015, SIST EN 61000-4-4:2013, SIST EN 60601-1-2:2015/A1:2021, SIST EN 60601-1-8:2008/A1:2014, SIST EN 61000-4-8:2010, SIST EN 61000-4-5:2014, SIST EN IEC 61000-4-3:2021, SIST EN 60601-1:2007, SIST EN 60601-1:2007/A2:2021, SIST EN 60601-1:2007/A1:2014, SIST EN 1789:2007+A2:2015, SIST EN 1789:2007+A1:2010, SIST EN 13718-1:2008, SIST EN 60601-2-4:2011. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

SIST EN 60601-2-4:2011/A2:2027 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC; Standardization Mandates: M/295. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

SIST EN 60601-2-4:2011/A2:2027 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
01-januar-2027
Medicinska električna oprema - 2-4. del: Posebne zahteve za osnovno varnost in
bistvene lastnosti srčnih defibrilatorjev - Dopolnilo A2 (IEC 60601-2-
4:2010/AMD2:2026)
Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and
essential performance of cardiac defibrillators (IEC 60601-2-4:2010/AMD2:2026)
Medizinische elektrische Geräte - Teil 2-4: Besondere Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale von Defibrillatoren (IEC 60601-2-
4:2010/AMD2:2026)
Appareils électromédicaux - Partie 2-4: Exigences particulières pour la sécurité de base
et les performances essentielles des défibrillateurs cardiaques (IEC 60601-2-
4:2010/AMD2:2026)
Ta slovenski standard je istoveten z: EN 60601-2-4:2011/A2:2026
ICS:
11.040.10 Anestezijska, respiratorna in Anaesthetic, respiratory and
reanimacijska oprema reanimation equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD EN 60601-2-4:2011/A2

NORME EUROPÉENNE
EUROPÄISCHE NORM November 2026
ICS 11.040.10
English Version
Medical electrical equipment - Part 2-4: Particular requirements
for the basic safety and essential performance of cardiac
defibrillators
(IEC 60601-2-4:2010/AMD2:2026)
Appareils électromédicaux - Partie 2-4: Exigences Medizinische elektrische Geräte - Teil 2-4: Besondere
particulières pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles des défibrillateurs cardiaques wesentlichen Leistungsmerkmale von Defibrillatoren
(IEC 60601-2-4:2010/AMD2:2026) (IEC 60601-2-4:2010/AMD2:2026)
This amendment A2 modifies the European Standard EN 60601-2-4:2011; it was approved by CENELEC on 2026-11-02. CENELEC
members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the
status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Türkiye and the United Kingdom.

European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 60601-2-4:2011/A2:2026 E

European foreword
The text of document 62D/2317/FDIS, future edition 3 of IEC 60601-2-4/AMD2, prepared by SC 62D
"Particular medical equipment, software, and systems" of IEC/TC 62 "Medical equipment, software,
and systems" was submitted to the IEC-CENELEC parallel vote and approved by CENELEC as
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2027-11-30
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2029-11-30
document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
Endorsement notice
The text of the International Standard IEC 60601-2-4:2010/AMD2:2026 was approved by CENELEC
as a European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standard indicated:
IEC 60601-1-11:2015 NOTE Approved as EN 60601-1-11:2015 (not modified)
CISPR 16-2-3 NOTE Approved as EN 55016-2-3
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1  Where an International Publication has been modified by common modifications, indicated by (mod),
the relevant EN/HD applies.
NOTE 2  Up-to-date information on the latest versions of the European Standards listed in this annex is available
here: www.cencenelec.eu.
Replace IEC 60601-1-2 and its amendment with:
Publication Year Title EN/HD Year
IEC 60601-1-2 2014 Medical electrical equipment - Part 1-2: EN 60601-1-2 2015
General requirements for basic safety and
essential performance - Collateral
Standard: Electromagnetic disturbances -
Requirements and tests
+ A1 2020 + A1 2021
- - + prAA
Replace IEC 60601-1 and its amendment with:
Publication Year Title EN/HD Year
IEC 60601-1 2005 Medical electrical equipment - Part 1: EN 60601-1 2006
General requirements for basic safety and
essential performance
- - + corrigendum Mar. 2010
+ A1 2012 + A1 2013
- - + A12 2014
+ A2 2020 + A2 2021
- - + A13 2024
Replace ISO 15223-1 and its amendment with:
Publication Year Title EN/HD Year
ISO 15223-1 2021 Medical devices - Symbols to be used with EN ISO 15223-1 2021
information to be supplied by the
manufacturer - Part 1: General
requirements
Add:
Publication Year Title EN/HD Year
IEC 60601-1-8 2006 Medical electrical equipment – Part 1-8: - -
General requirements for basic safety and
essential performance - Collateral
Standard: General requirements, tests and
guidance for alarm systems in medical
electrical equipment and medical electrical
systems
+ A1 2012 + A1 2013
- - + AC 2014
+ A2 2020 - -
IEC/TR 61000-2-5 - Electromagnetic compatibility (EMC) - Part - -
2-5: Environment - Description and
classification of electromagnetic
environments
IEC 61000-4-3 - Electromagnetic compatibility (EMC) - Part EN IEC 61000-4-3 -
4-3 : Testing and measurement techniques
- Radiated, radio-frequency,
electromagnetic field immunity test
IEC 61000-4-4 - Electromagnetic compatibility (EMC) - Part EN 61000-4-4 -
4-4: Testing and measurement techniques
- Electrical fast transient/burst immunity
test
IEC 61000-4-5 2014 Electromagnetic compatibility (EMC) - Part EN 61000-4-5 2014
4-5: Testing and measurement techniques
- Surge immunity test
IEC 61000-4-8 - Electromagnetic compatibility (EMC) - Part EN 61000-4-8 -
4-8: Testing and measurement techniques
- Power frequency magnetic field immunity
test
IEC 60601-2-4 ®
Edition 3.0 2026-09
INTERNATIONAL
STANDARD
AMENDMENT 2
Medical electrical equipment -
Part 2-4: Particular requirements for the basic safety and essential performance
of cardiac defibrillators
ICS 11.040.10  ISBN 978-2-8327-1502-4

IEC 60601-2-4:2010-12/AMD2:2026-09(en)

IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
Medical electrical equipment -
Part 2-4: Particular requirements for the basic safety and essential
performance of cardiac defibrillators

AMENDMENT 2
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence between
any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) IEC draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in
respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which
may be required to implement this document. However, implementers are cautioned that this may not represent
the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC
shall not be held responsible for identifying any or all such patent rights.
Amendment 2 to IEC 60601-2-4:2010 has been prepared by subcommittee 62D: Particular
medical equipment, software, and systems, of IEC technical committee 62: Medical equipment,
software, and systems.
The text of this Amendment is based on the following documents:
Draft Report on voting
62D/2317/FDIS 62D/2327/RVD
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
The language used for the development of this Amendment is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
described in greater detail at www.iec.ch/publications.
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under webstore.iec.ch in the data related to the
specific document. At this date, the document will be
– reconfirmed,
– withdrawn, or
– revised.
___________
INTRODUCTION to Amendment 2
Amendment 2 to the International standard IEC 60601-2-4 has been prepared by IEC
subcommittee 62D: Particular medical equipment, software, and systems, of IEC technical
committee 62: Medical equipment, software, and systems in medical practice. Amendment 2
changes is to align to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 and IEC 60601-1-2:2014 and IEC 60601-1-2:2014/AMD1:2020.
201.1 Scope, object and related standards
Replace the existing footnote 1 with the following new text:
The general standard is IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020,
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
201.2 Normative references
Replace the existing reference to IEC 60601-1-2, modified by Amendment 1, with the following
new reference:
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic
safety and essential performance - Collateral standard: Electromagnetic disturbances -
Requirements and tests
IEC 60601-1-2:2014/AMD1:2020
Replace the existing reference to IEC 60601-1, modified by Amendment 1, with the following
new reference:
IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety
and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1:2005/AMD2:2020
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
Replace the existing reference to ISO 15223-1, modified by Amendment 1, with the following
new reference:
ISO 15223-1:2021, Medical devices - Symbols to be used with information to be supplied by
the manufacturer - Part 1: General requirements
Add the following normative references:
IEC 60601-1-8:2006, Medical electrical equipment - Part 1-8: General requirements for basic
safety and essential performance - Collateral standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
IEC 60601-1-8:2006/AMD1:2012
IEC 60601-1-8:2006/AMD2:2020
IEC TR 61000-2-5, Electromagnetic compatibility (EMC) - Part 2-5: Environment - Description
and classification of electromagnetic environments
IEC 61000-4-3, Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement
techniques - Radiated, radio-frequency, electromagnetic field immunity test
IEC 61000-4-4, Electromagnetic compatibility (EMC) - Part 4-4: Testing and measurement
techniques - Electrical fast transient/burst immunity test
IEC 61000-4-5:2014, Electromagnetic compatibility (EMC) - Part 4-5: Testing and measurement
techniques - Surge immunity test
IEC 61000-4-8, Electromagnetic compatibility (EMC) - Part 4-8: Testing and measurement
techniques - Power frequency magnetic field immunity test
201.3 Terms and definitions
Replace the existing text in the first paragraph with the following updated text:
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, IEC 60601-1-2:2014, and
IEC 60601-1-2:2014/AMD1:2020 apply, except as follows:
201.7.2.103 Disposable defibrillator electrodes
Replace, in the existing item a), modified by Amendment 1, "ISO 15223-1:2007" with
"ISO 15223-1:2021".
201.8.5.3 * MAXIMUM MAINS VOLTAGE
Replace, in the existing last paragraph, modified by Amendment 1, "250 V" with "240 V".
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
201.8.5.5.1 * Defibrillation protection
Replace the existing second sentence that reads “Replacement of the third paragraph ("After
the operation of S, the peak.”) by the following:” with the following:
Replacement of the second-to-last paragraph (“After the operation of S, the peak….”) by the
following:
Add, at the end of the existing Subclause 201.12.2.102, the following new Subclause 201.12.3
and text:
201.12.3 * ALARM SYSTEMS
Replacement:
RISK MANAGEMENT is used to determine RISK CONTROL measures required for reducing the RISK(S)
due to unsatisfactory PATIENT physiological states or unsatisfactory functional states of the
MEDICAL ELECTRICAL EQUIPMENT or MEDICAL ELECTRICAL SYSTEM, to an acceptable level.
Requirements of IEC 60601-1-8:2006, IEC 60601-1-8:2006/AMD1:2012 and
IEC 60601-1-8:2006/AMD2:2020 apply when RISK MANAGEMENT has determined a requirement
to indicate to the OPERATOR of unsatisfactory physiological PATIENT states or unsatisfactory
functional states of the MEDICAL ELECTRICAL EQUIPMENT or MEDICAL ELECTRICAL SYSTEM, or to
warn the OPERATOR of HAZARDS to the PATIENT or OPERATOR due to the MEDICAL ELECTRICAL
EQUIPMENT or MEDICAL ELECTRICAL SYSTEM.
RISK CONTROL measures shall be appropriate for the environment of use and level of OPERATOR
training.
Deviations from the alarm standard shall be validated (e.g. usability).
201.101 * Charging time
201.101.1 Requirements for FREQUENT USE, MANUAL DEFIBRILLATORS
Replace the entire existing text of this subclause, including items a) and b), modified by
Amendment 1, with the following new text:
a) The time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not
exceed 15 s under the following conditions:
1) when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the MANUAL DEFIBRILLATOR is operated with batteries depleted by whichever of the
following conditions comes last:
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which the
battery is specified as still useable by the MANUFACTURER.
b) The time from initially switching power on, or from within any OPERATOR programming mode,
to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not exceed
25 s under the following conditions:
1) when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the MANUAL DEFIBRILLATOR is operated with batteries depleted by whichever of the
following conditions comes last:
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which the
battery is specified as still useable by the MANUFACTURER.
Check compliance with a) and b) above by measurement. In the case of INTERNALLY POWERED
ME EQUIPMENT, the test is to start with a new and fully charged battery. In the case of such
ME EQUIPMENT also capable of charging the ENERGY STORAGE DEVICE when connected to the
SUPPLY MAINS or to a separate battery charger, compliance is also checked when the
ME EQUIPMENT is connected to the SUPPLY MAINS or to the battery charger. In cases with a
discharged or missing battery, verify performance is consistent with the markings provided as
required by 201.7.2.102.
201.101.2 Requirements for INFREQUENT USE, MANUAL DEFIBRILLATORS
Replace the entire existing text of this subclause, including items a) and b), with the following
new text:
a) For the time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy, the
following charge time requirements apply:
1) the time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall
not exceed 20 s when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS
VOLTAGE;
2) the time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall
not exceed 25 s when the MANUAL DEFIBRILLATOR is operated with batteries depleted by
whichever of the following conditions comes last:
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
b) For the time from initially switching power on, or from within any OPERATOR programming
mode, to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy, the
following charge time requirements apply:
1) the time from switching power on, or from within any OPERATOR programming mode, to
the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not
exceed 30 s when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS
VOLTAGE;
2) the time from switching power on, or from within any OPERATOR programming mode, to
the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not
exceed 35 s when the MANUAL DEFIBRILLATOR is operated with batteries depleted by
whichever of the following conditions comes last:
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
201.101.3 * Requirements for FREQUENT USE, AUTOMATED EXTERNAL DEFIBRILLATORS
Replace the existing text of this subclause with the following new text:
a) The time from activation of the RHYTHM RECOGNITION DETECTOR to the AUTOMATED EXTERNAL
DEFIBRILLATORS (AED) being ready for discharge at maximum energy shall not exceed 30 s
under the following conditions:
1) when the AED is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the AED is operated with batteries depleted by whichever of the following conditions
comes last:
i) the delivery of charge and discharge cycles until the AED indicates when non-
rechargeable batteries require replacement or rechargeable batteries require
recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
b) * The time from initially switching power on, or from within any OPERATOR programming
mode, to the AED being ready for discharge at maximum energy shall not exceed 40 s under
the following conditions:
1) when the AED is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the AED is operated with batteries depleted by whichever of the following conditions
comes last:
i) the delivery of charge and discharge cycles until the AED indicates when non-
rechargeable batteries require replacement or rechargeable batteries require
recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
Check compliance with a) and b) above by the following test:
A shockable simulated patient rhythm signal as defined by the manufacturer is applied to the
separate monitoring electrodes or the defibrillator electrodes. The visual or audible instructions
given by the defibrillator are followed. The charge time is measured from RRD activation [for
201.101.3 a)] or initial power on or exiting an operator programming mode [for 201.101.3 b)] to
the AED being ready for discharge at maximum energy.
In the case of internally powered ME EQUIPMENT, the test is to start with a new and fully charged
battery. In the case of such ME EQUIPMENT also capable of charging the energy storage device
when connected to the supply mains or to a separate battery charger, compliance is also
checked when the ME EQUIPMENT is connected to the SUPPLY MAINS or to the battery charger. In
cases with a discharged or missing battery, verify the performance is consistent with the
markings provided as required by 201.7.2.102.
201.101.4 * Requirements for INFREQUEN
...