Standard Specification for Pipet, Sahli Hemoglobin

ABSTRACT
This specification covers reusable Sahli hemoglobin pipets. The pipets shall be made of common spirit bore white back tubing. Pipet design shall be straight and of one-piece construction. Pipet delivery tips shall be made with a gradual or concave taper. Pipet dimensions shall be within tolerance limit indicated in this specification. The pipet shall have markings which include graduation line, volumetric designation, identification, and capacity deviation. Testing methods include capacity test, capacity deviation, pigmentation test, and strain free test.
SCOPE
1.1 This specification covers reusable pipets calibrated “to contain” 20 cmm of whole blood and used for hemoglobin determinations.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

General Information

Status
Historical
Publication Date
31-Aug-2016
Current Stage
Ref Project

Buy Standard

Technical specification
ASTM E1045-00(2016) - Standard Specification for Pipet, Sahli Hemoglobin
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview
Technical specification
REDLINE ASTM E1045-00(2016) - Standard Specification for Pipet, Sahli Hemoglobin
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E1045 −00 (Reapproved 2016)
Standard Specification for
1
Pipet, Sahli Hemoglobin
This standard is issued under the fixed designation E1045; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 4.3 Thetopofthepipetshallbegroundtoataperorformed
to a funnel shape according to the dimensions specified in Fig.
1.1 This specification covers reusable pipets calibrated “to
1.
contain” 20 cmm of whole blood and used for hemoglobin
determinations.
5. Markings
1.2 The values stated in SI units are to be regarded as
5.1 Graduation Line—The pipet shall have one graduation
standard. No other units of measurement are included in this
line located 20 cmm from the pipet tip. The graduation line
standard.
shall be located on the clear portion of the tubing and shall
2. Referenced Documents
extend at least two thirds around the pipet and not exceed 0.4
2
mm in width.
2.1 ASTM Standards:
E920Specification for Commercially Packaged Laboratory
5.2 Volumetric Designation—The pipet shall be marked
Apparatus
20CMM on the clear portion of the tubing with the markings
E921Specification for Export Packaged Laboratory Appa-
located approximately 5 mm above the graduation line.
ratus
5.3 Identification—Each pipet shall be marked with the
E1133Practice for Performance Testing of Packaged Labo-
manufacturer’s name or trademark on the white stripe portion
ratory Apparatus for United States Government Procure-
of the pipet. Catalog number markings are optional. All
ments
markings shall be permanently fused onto the pipet. The
E1157Specification for Sampling and Testing of Reusable
markings shall be amber or black in color. When tested in
Laboratory Glassware
accordance to 6.3, the pigmentation shall not discolor. The
appearance of the markings, when viewed by the eye under
3. Materials
normal room lighting, shall be the same before and after
3.1 The pipets shall be made of common spirit bore white
testing.
back tubing or of clear glass with a white stripe applied to the
5.4 Capacity Deviation—SahliHemoglobinpipetsaremade
outer surface of the tubing.
with maximum capacity deviation of 61.0% or 62.0%. The
4. Design
selected capacity deviation shall be marked on the clear or
4.1 Shape—Pipets shall be straight and of one-piece con- whitestripeportionofthepipet.Thecapacityofthepipetshall
be within the selected capacity deviation marked on the pipet
struction. Any cross section of a pipet taken in a plane
perpendicular to the longitudinal axis should be circular. when tested as specified in 6.2.
4.2 Delivery Tips—Delivery tips shall be made with a
6. Testing
gradual or concave taper to a length of 10 to 25 mm. The tip
end shall be ground and tapered with fine abrasive, or fire- 6.1 Capacity Test—The capacity of the pipet shall be
polished.Dimensionsofthedeliverytipshallbeasspecifiedin
determined by means of using distilled water and a weighing
Fig. 1. device with weight sensitivity not less than 0.001 mg.
6.1.1 The pipet shall be thoroughly cleaned, dried, and
1
allowed to adjust to room temperature.
This specification is under the jurisdiction of ASTM Committee E41 on
Laboratory Apparatus and is the direct responsibility of Subcommittee E41.01 on
6.1.2 The pipet shall be weighed and the weight recorded.
Laboratory Ware and Supplies.
6.1.3 The pipet shall be filled to the calibration line with
Current edition approved Sept. 1, 2016. Published September 2016. Originally
distilled water and weighed, and the weight recorded.
approved in 1985. Last previous edition approved in 2010 as E1045–00(2010).
DOI: 10.1520/E1045-00R16.
6.1.4 Therecordedweightofthecleananddrypipetshallbe
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
subtractedfromtherecordedweightofthedistilledwater-filled
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
pipet providing the observed volumetric capacity (V)ofthe
Standards volume information, refer to the standard’s Document Summary page on c
the ASTM website. pipet in grams.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1045−00 (2016)
FIG. 1 Pipet, Sahli Hemoglobin
6.1.5 The observed volumetric capacity (V) shall then be Cr O 2H O),1000mLofwater,and1500mLofsulfuricacid
t 2 7 2
corrected to actual volumetric capacity at 20°C (V ), deter- (H SO , ACS Reagent—95 to 98%). Immerse the pipets in
c 2 4
mined by: thechromicacidsolution.Letstandatroomtemperature(20to
25°C) for 15 min. Remove the pipets from the solution and
V
t
V 5 (1)
thoroughly r
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1045 − 00 (Reapproved 2010) E1045 − 00 (Reapproved 2016)
Standard Specification for
1
Pipet, Sahli Hemoglobin
This standard is issued under the fixed designation E1045; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This specification covers reusable pipets calibrated “to contain” 20 cmm of whole blood and used for hemoglobin
determinations.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
2. Referenced Documents
2
2.1 ASTM Standards:
E920 Specification for Commercially Packaged Laboratory Apparatus
E921 Specification for Export Packaged Laboratory Apparatus
E1133 Practice for Performance Testing of Packaged Laboratory Apparatus for United States Government Procurements
E1157 Specification for Sampling and Testing of Reusable Laboratory Glassware
3. Materials
3.1 The pipets shall be made of common spirit bore white back tubing or of clear glass with a white stripe applied to the outer
surface of the tubing.
4. Design
4.1 Shape—Pipets shall be straight and of one-piece construction. Any cross section of a pipet taken in a plane perpendicular
to the longitudinal axis should be circular.
4.2 Delivery Tips—Delivery tips shall be made with a gradual or concave taper to a length of 10 to 25 mm. The tip end shall
be ground and tapered with fine abrasive, or fire-polished. Dimensions of the delivery tip shall be as specified in Fig. 1.
4.3 The top of the pipet shall be ground to a taper or formed to a funnel shape according to the dimensions specified in Fig.
1.
5. Markings
5.1 Graduation Line—The pipet shall have one graduation line located 20 cmm from the pipet tip. The graduation line shall be
located on the clear portion of the tubing and shall extend at least two thirds around the pipet and not exceed 0.4 mm in width.
5.2 Volumetric Designation—The pipet shall be marked 20CMM on the clear portion of the tubing with the markings located
approximately 5 mm above the graduation line.
5.3 Identification—Each pipet shall be marked with the manufacturer’s name or trademark on the white stripe portion of the
pipet. Catalog number markings are optional. All markings shall be permanently fused onto the pipet. The markings shall be amber
or black in color. When tested in accordance to 6.3, the pigmentation shall not discolor. The appearance of the markings, when
viewed by the eye under normal room lighting, shall be the same before and after testing.
5.4 Capacity Deviation—Sahli Hemoglobin pipets are made with maximum capacity deviation of 61.0 % or 62.0 %. The
selected capacity deviation shall be marked on the clear or white stripe portion of the pipet. The capacity of the pipet shall be within
the selected capacity deviation marked on the pipet when tested as specified in 6.2.
1
This specification is under the jurisdiction of ASTM Committee E41 on Laboratory Apparatus and is the direct responsibility of Subcommittee E41.01 on
ApparatusLaboratory Ware and Supplies.
Current edition approved July 1, 2010Sept. 1, 2016. Published July 2010September 2016. Originally approved in 1985. Last previous edition approved in 20052010 as
E1045 – 00 (2005).(2010). DOI: 10.1520/E1045-00R10.10.1520/E1045-00R16.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1045 − 00 (2016)
FIG. 1 Pipet, Sahli Hemoglobin
6. Testing
6.1 Capacity Test—The capacity of the pipet shall be determined by means of using distilled water and a weighing device with
weight sensitivity not less than 0.001 mg.
6.1.1 The pipet shall be thoroughly cleaned, dried, and allowed to adjust to room temperature.
6.1.2 The pipet shall be weighed and the weight recorded.
6.1.3 The pipet shall be filled to the calibration line with distilled water and weighed, and the weight recorded.
6.1.4 The recorded weight of the clean and dry pipet shall be subtracted from the recorded weight of the distilled water-filled
pipet providing the observ
...