ASTM E961-97(2021)
(Specification)Standard Specification for Blood Sedimentation Tube, Wintrobe, Glass, Reusable
Standard Specification for Blood Sedimentation Tube, Wintrobe, Glass, Reusable
ABSTRACT
This specification covers reusable blood sedimentation tubes suitable for determining sedimentation rates and the volume of packed red blood cells. The tubes shall be fabricated from borosilicate glass, Type I, Class B, or sodalime glass, Type II. A resistance to centrifugal force test shall be performed on the tubes.
SCOPE
1.1 This specification covers reusable blood sedimentation tubes suitable for determining sedimentation rates and the volume of packed red blood cells.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E961 −97 (Reapproved 2021)
Standard Specification for
Blood Sedimentation Tube, Wintrobe, Glass, Reusable
This standard is issued under the fixed designation E961; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 3. Terminology
3.1 Definitions of Terms Specific to This Standard:
1.1 This specification covers reusable blood sedimentation
3.1.1 reusable—capable of being used again.
tubes suitable for determining sedimentation rates and the
volume of packed red blood cells. 3.1.2 Wintrobe—the surname of the individual responsible
for the design of the Wintrobe tube and the method of use.
1.2 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this
4. Classification
standard.
4.1 This specification covers a tube that is intended to be
1.3 This standard does not purport to address all of the
used until it is no longer considered a functional device for the
safety concerns, if any, associated with its use. It is the
purpose intended.
responsibility of the user of this standard to establish appro-
5. Materials
priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
5.1 Glass—The tubes made to this specification shall be
1.4 This international standard was developed in accor- fabricated from borosilicate glass, Type I, Class B, or soda-
lime glass, Type II, in accordance with Specification E438.
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
6. Dimensions and Graduations
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
6.1 Dimensions—The tube shall be made of tubing with an
Barriers to Trade (TBT) Committee. outside diameter (O.D.) of 7.0 to 8.0 mm with an inside
diameter (I.D.) of 2.9 to 3.3 mm. The uniformity of the bore
shall be 60.1 mm throughout the tube. The tube shall be 110
2. Referenced Documents
to 117 mm long and have a graduated scale of 105 6 0.25 mm
2.1 ASTM Standards:
from the inside bottom of the tube. The tube shall be legibly
E438 Specification for Glasses in Laboratory Apparatus
marked with the manufacturer’s or vendor’s name or mark and
E920 Specification for Commercially Packaged Laboratory
possess a frosted area for marking purposes.
Apparatus
6.2 Graduation Scale—The tube shall be graduated 105 6
E921 Specification for Export Packaged Laboratory Appa-
0.25mmin1-mmdivisionsandnumberedevery1cmwithtwo
ratus
sets of numerals. One set of graduation numerals shall be from
E1133 Practice for Performance Testing of Packaged Labo-
0 to 9 cm down the left side of the graduation scale and the
ratory Apparatus for United States Government Procure-
other set of g (20 to 25 °C) for 15 min. Remove tube from the
ments
solution and thoroughly rinse in tap water followed by distilled
E1157 Specification for Samplin
...
Questions, Comments and Discussion