ASTM E578-83(1998)
(Test Method)Standard Test Method for Linearity of Fluorescence Measuring Systems
Standard Test Method for Linearity of Fluorescence Measuring Systems
SCOPE
1.1 This test method covers a procedure for evaluating the limits of the linearity of response with fluorescence intensity of fluorescence-measuring systems under operating conditions. Particular attention is given to slit widths, filters, and sample containers. This test method can be used to test the overall linearity under a wide variety of instrumental and sampling conditions. The results obtained apply only to the tested combination of slit width and filters, and the size, type and illumination of the sample cuvette, all of which must be stated in the report. The sources of nonlinearity may be the measuring electronics, excessive absorption of either the exciting or emitted radiation, or both, and the sample handling technique, particularly at low concentrations.
1.2 This test method has been applied to fluorescence-measuring systems utilizing continuous and low-energy excitation sources (for example, an excitation source of 450-W electrical input or less). There is no assurance that extremely intense illumination will not cause photodecomposition of the compounds suggested in this test method. For this reason it is recommended that this test method not be indiscriminately employed with high-intensity light sources. It is not a test method to determine the linearity of response of other materials. If this test method is extended to employ other chemical substances, the principles within can be applied, but new material parameters, such as the concentration range of linearity, must be established. The user should be aware of the possibility that these other substances may undergo decomposition, or adsorption onto containers.
1.3 This test method is applicable to 10-mm pathlength cuvette formats and instruments covering a wavelength range within 190 to 900 nm. The use of other sample formats has not been established with this test method.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
Designation: E 578 – 83 (Reapproved 1998)
Standard Test Method for
Linearity of Fluorescence Measuring Systems
This standard is issued under the fixed designation E 578; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope can be employed (Note 1). The standard used to determine
linearity should be stated in the report. The fluorescence of the
1.1 This test method covers a procedure for evaluating the
test solution is measured in the measuring system with the
limits of linearity of fluorescence-measuring systems under
cuvettes, slits, or filters that are to be employed in projected
operating conditions. Particular attention is given to slit widths,
use.
filters, and sample containers. This test method can be used to
test the overall linearity under a wide variety of instrumental
NOTE 1—A substitute standard should have the following properties:
and sampling conditions. The results obtained apply only to the
(1) It should have a large quantum yield at very high dilution; (2) it should
be stable to the exciting radiation during spectral measurements; (3) its
tested combination of slit width and filters, and the size, type
fluorescence and its absorption spectra overlap should be small; (4) its
and illumination of the sample cuvette, all of which must be
quantum yield should not be strongly concentration dependent; and (5)it
stated in the report. The sources of nonlinearity may be the
should have a broad emission spectrum, so that little error is introduced
measuring electronics, excessive absorption of either the ex-
when wide slits are used.
citing or emitted radiation, or both, and the sample handling
2.2 Upper Limit of Linearity—The fluorescence intensity of
technique, particularly at low concentrations.
a series of standard solutions is measured, the resultant
1.2 This test method has been applied to fluorescence-
instrument readings are plotted against concentration on log-
measuring systems utilizing continuous and low-energy exci-
log graph paper, and a smooth curve is drawn through the data
tation sources (for example, an excitation source of 150-W
points. The point (concentration) at which the upper end of the
electrical input or less). There is no assurance that extremely
curve deviates by more than 5 % of the signal from the straight
intense illumination will not cause photodecomposition of the
2 line (defined by the center region of the curve) is taken as the
compounds suggested in this test method. For this reason it is
upper limit of linearity. The limit is expressed in micrograms
recommended that this test method not be indiscriminately
per milliliter of quinine sulfate dihydrate.
employed with high-intensity light sources. It is not a test
method to determine the linearity of response of other materi-
NOTE 2—Absorption of the exciting radiation at high solute concentra-
als. If this test method is extended to employ other chemical tions is dependent on instrument geometry, and can result in fluorescence
signal nonlinearity.
substances, the user should be aware of the possibility that
these other substances may undergo decomposition, or adsorp-
2.3 Lower Limit of Linearity—The lower limit of linearity is
tion onto containers.
taken as the point (concentration) at which the lower end of the
1.3 This standard does not purport to address all of the
curve deviates from the straight line defined by the central
safety problems, if any, associated with its use. It is the
portion of the curve by more than twice the average percent
responsibility of the user of this standard to establish appro-
deviation of the points that determine the straight line.
priate safety and health practices and determine the applica-
3. Significance and Use
bility of regulatory limitations prior to use.
3.1 The range of concentration of a fluorescing substance in
2. Summary of Test Method
solution over which the fluorescence varies linearly with the
2.1 This procedure is used for testing the linearity of
concentration is the range most useful for quantitative analysis.
fluorescence-measuring systems by using solutions of quinine
This range is affected by properties of the solution under
sulfate dihydrate in sulfuric acid as standard test solutions.
analysis and by features of the measuring system. This test
Other stable solutions which may be more suitable to the user
method provides a means of testing the performance of a
fluorescence measuring system and of determining the concen-
tration range over which the system is suitable for making a
This test method is under the jurisdiction of ASTM Committee E-13 on
Molecular Spectroscopy and is the direct responsibility of Subcommittee E13.06 on
given quantitative analysis.
Molecular Luminescence.
3.2 This test method is not meant for comparing the
Current edition approved Aug. 26, 1983. Published October 1983. Originally
published as E 578 – 76. Last previous edition E 578 – 76.
Lukasiewicz, R. J., and Fitzgerald, J. M., Analytical Chemistry, ANCHA, Vol
45, 1973, p. 511. Gill, J. E., Photochemistry and Photobiology, PHCBA, Vol 9, 1969, p. 313.
Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
E 578
performance of different fluorescence measuring instruments. standard solutions, ending with the stock solution described in
5.1.
4. Apparatus
7. Calculation of Results and Data P
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.