Standard Practice for Minimizing Dosimetry Errors in Radiation Hardness Testing of Silicon Electronic Devices Using Co-60 Sources

SIGNIFICANCE AND USE
4.1 Division of the Co-60 Hardness Testing into Five Parts:  
4.1.1 The equilibrium absorbed dose shall be measured with a dosimeter, such as a TLD, located adjacent to the device under test. Alternatively, a dosimeter may be irradiated in the position of the device before or after irradiation of the device.  
4.1.2 This absorbed dose measured by the dosimeter shall be converted to the equilibrium absorbed dose in the material of interest within the critical region within the device under test, for example the SiO2 gate oxide of an MOS device.  
4.1.3 A correction for absorbed-dose enhancement effects shall be considered. This correction is dependent upon the photon energy that strikes the device under test.  
4.1.4 A correlation should be made between the absorbed dose in the critical region (for example, the gate oxide mentioned in 4.1.2) and some electrically important effect (such as charge trapped at the Si/SiO2 interface as manifested by a shift in threshold voltage).  
4.1.5 An extrapolation should then be made from the results of the test to the results that would be expected for the device under test under actual operating conditions.
Note 5: The parts of a test discussed in 4.1.2 and 4.1.3 are the subject of this practice. The subject of 4.1.1 is covered and referenced in other standards such as Practice E668 and ICRU Report 14. The parts of a test discussed in 4.1.4 and 4.1.5 are outside the scope of this practice.  
4.2 Low-Energy Components in the Spectrum—Some of the primary Co-60 gamma rays (1.17 and 1.33 MeV) produce lower energy photons by Compton scattering within the Co-60 source structure, within materials that lie between the source and the device under test, and within materials that lie beyond the device but contribute to backscattering. As a result of the complexity of these effects, the photon energy spectrum striking the device usually is not well known. This point is further discussed in Section 5 and Appendix X1. The presence of ...
SCOPE
1.1 This practice covers recommended procedures for the use of dosimeters, such as thermoluminescent dosimeters (TLD's), to determine the absorbed dose in a region of interest within an electronic device irradiated using a Co-60 source. Co-60 sources are commonly used for the absorbed dose testing of silicon electronic devices.  
Note 1: This absorbed-dose testing is sometimes called “total dose testing” to distinguish it from “dose rate testing.”  
Note 2: The effects of ionizing radiation on some types of electronic devices may depend on both the absorbed dose and the absorbed dose rate; that is, the effects may be different if the device is irradiated to the same absorbed-dose level at different absorbed-dose rates. Absorbed-dose rate effects are not covered in this practice but should be considered in radiation hardness testing.  
1.2 The principal potential error for the measurement of absorbed dose in electronic devices arises from non-equilibrium energy deposition effects in the vicinity of material interfaces.  
1.3 Information is given about absorbed-dose enhancement effects in the vicinity of material interfaces. The sensitivity of such effects to low energy components in the Co-60 photon energy spectrum is emphasized.  
1.4 A brief description is given of typical Co-60 sources with special emphasis on the presence of low energy components in the photon energy spectrum output from such sources.  
1.5 Procedures are given for minimizing the low energy components of the photon energy spectrum from Co-60 sources, using filtration. The use of a filter box to achieve such filtration is recommended.  
1.6 Information is given on absorbed-dose enhancement effects that are dependent on the device orientation with respect to the Co-60 source.  
1.7 The use of spectrum filtration and appropriate device orientation provides a radiation environment whereby the absorbed dose in the sensitive region of an elect...

General Information

Status
Published
Publication Date
31-Jan-2021
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1249-15(2021) - Standard Practice for Minimizing Dosimetry Errors in Radiation Hardness Testing of Silicon Electronic Devices Using Co-60 Sources
English language
17 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1249 − 15 (Reapproved 2021)
Standard Practice for
Minimizing Dosimetry Errors in Radiation Hardness Testing
1
of Silicon Electronic Devices Using Co-60 Sources
This standard is issued under the fixed designation E1249; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.
1. Scope 1.7 The use of spectrum filtration and appropriate device
orientation provides a radiation environment whereby the
1.1 This practice covers recommended procedures for the
absorbed dose in the sensitive region of an electronic device
use of dosimeters, such as thermoluminescent dosimeters
can be calculated within defined error limits without detailed
(TLD’s),todetermine the absorbed dose in a regionofinterest
knowledge of either the device structure or of the photon
within an electronic device irradiated using a Co-60 source.
energyspectrumofthesource,andhence,withoutknowingthe
Co-60 sources are commonly used for the absorbed dose
details of the absorbed-dose enhancement effects.
testing of silicon electronic devices.
1.8 The recommendations of this practice are primarily
NOTE 1—This absorbed-dose testing is sometimes called “total dose
applicable to piece-part testing of electronic devices. Elec-
testing” to distinguish it from “dose rate testing.”
tronic circuit board and electronic system testing may intro-
NOTE 2—The effects of ionizing radiation on some types of electronic
devicesmaydependonboththeabsorbeddoseandtheabsorbeddoserate;
duce problems that are not adequately treated by the methods
that is, the effects may be different if the device is irradiated to the same
recommended here.
absorbed-dose level at different absorbed-dose rates. Absorbed-dose rate
1.9 This standard does not purport to address all of the
effects are not covered in this practice but should be considered in
radiation hardness testing. safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
1.2 The principal potential error for the measurement of
priate safety, health, and environmental practices and deter-
absorbed dose in electronic devices arises from non-
mine the applicability of regulatory limitations prior to use.
equilibriumenergydepositioneffectsinthevicinityofmaterial
1.10 This international standard was developed in accor-
interfaces.
dance with internationally recognized principles on standard-
1.3 Information is given about absorbed-dose enhancement
ization established in the Decision on Principles for the
effects in the vicinity of material interfaces. The sensitivity of
Development of International Standards, Guides and Recom-
such effects to low energy components in the Co-60 photon
mendations issued by the World Trade Organization Technical
energy spectrum is emphasized.
Barriers to Trade (TBT) Committee.
1.4 A brief description is given of typical Co-60 sources
with special emphasis on the presence of low energy compo-
2. Referenced Documents
nentsinthephotonenergy spectrum output from suchsources. 2
2.1 ASTM Standards:
1.5 Procedures are given for minimizing the low energy E170Terminology Relating to Radiation Measurements and
components of the photon energy spectrum from Co-60 Dosimetry
sources, using filtration.The use of a filter box to achieve such E666Practice for CalculatingAbsorbed Dose From Gamma
filtration is recommended. or X Radiation
E668Practice for Application of Thermoluminescence-
1.6 Information is given on absorbed-dose enhancement
Dosimetry (TLD) Systems for Determining Absorbed
effectsthataredependentonthedeviceorientationwithrespect
DoseinRadiation-HardnessTestingofElectronicDevices
to the Co-60 source.
E1250Test Method forApplication of Ionization Chambers
to Assess the Low Energy Gamma Component of
1
This practice is under the jurisdiction of ASTM Committee E10 on Nuclear
Technology and Applicationsand is the direct responsibility of Subcommittee
2
E10.07 on Radiation Dosimetry for Radiation Effects on Materials and Devices. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Feb. 1, 2021. Published February 2021. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1988. Last previous edition approved in 2015 as E1249–15. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.