E10.07 - Radiation Dosimetry for Radiation Effects on Materials and Devices
Radiation Dosimetry for Radiation Effects on Materials and Devices
General Information
SIGNIFICANCE AND USE
4.1 Flash X-ray facilities provide intense bremsstrahlung radiation environments, usually in a single sub-microsecond pulse, which often fluctuates in amplitude, shape, and spectrum from shot to shot. Therefore, appropriate dosimetry must be fielded on every exposure to characterize the environment, see ICRU Report 34. These intense bremsstrahlung sources have a variety of applications which include the following:
(1) Studies of the effects of X-rays and gamma rays on materials.
(2) Studies of the effects of radiation on electronic devices such as transistors, diodes, and capacitors.
(3) Computer code validation studies.
4.2 This guide is written to assist the experimenter in selecting the needed dosimetry systems for use at pulsed X-ray facilities. This guide also provides a brief summary on how to use each of the dosimetry systems. Other guides (see Section 2) provide more detailed information on selected dosimetry systems in radiation environments and should be consulted after an initial decision is made on the appropriate dosimetry system to use. There are many key parameters which describe a flash X-ray source, such as dose, dose rate, spectrum, pulse width, etc., such that typically no single dosimetry system can measure all the parameters simultaneously. However, it is frequently the case that not all key parameters must be measured in a given experiment.
SCOPE
1.1 This guide provides assistance in selecting and using dosimetry systems in flash X-ray experiments. Both dose and dose rate techniques are described.
1.2 Operating characteristics of flash X-ray sources are given, with emphasis on the spectrum of the photon output.
1.3 Assistance is provided to relate the measured dose to the response of a device under test (DUT). The device is assumed to be a semiconductor electronic part or system.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide19 pagesEnglish language
- Guide19 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Electronic circuits used in many space, military, and nuclear power systems may be exposed to various levels and time profiles of neutron radiation. It is essential for the design and fabrication of such circuits that test methods be available that can determine the vulnerability or hardness (measure of survivability) of components to be used in them. A determination of hardness is often necessary for the short term (≈100 μs) as well as long term (permanent damage) following exposure. See Practice E722.
SCOPE
1.1 This guide defines the requirements and procedures for testing silicon discrete semiconductor devices and integrated circuits for rapid annealing effects from displacement damage resulting from neutron radiation. This test will produce degradation of the electrical properties of the irradiated devices and should be considered a destructive test. Rapid annealing of displacement damage is usually associated with bipolar technologies.
1.1.1 Heavy ion beams can also be used to characterize displacement damage annealing (1),2 but ion beams have significant complications in the interpretation of the resulting device behavior due to the associated ionizing dose. The use of pulsed ion beams as a source of displacement damage is not within the scope of this standard.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide7 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Electronic devices are typically tested for device response to gamma radiation in pure gamma-ray fields. Testing electronic device response against neutrons is more complex since there is invariably a gamma-ray component in addition to the neutron field. The gamma-ray response of the electronic device is typically subtracted from the overall response to find the device response to neutrons. This approach to the testing requires a determination of the gamma-ray exposure in the mixed field. To enhance the neutron effects, the radiation field is sometimes selected to have as large a neutron component as possible.
4.2 CaF2(Mn) TLDs are often used to monitor the gamma-ray dose in mixed neutron/gamma radiation fields. Since the dosimeters are exposed along with the device under test to the mixed field, their response must be corrected for neutrons. In a field rich in neutrons, the uncertainty in the interpretation of the TLD response grows. In fields with relatively few neutrons, the total TLD response may be used to make a correction for gamma response of the device under test. Under this condition, the relative uncertainty in the TLD neutron response is not likely to drive the overall uncertainty in the correction to the electronic device response.
4.3 This practice gives a means of estimating the response of CaF2(Mn) TLDs to neutrons. This neutron response is then subtracted from the measured response to determine the TLD response due to gamma rays. The procedure has relatively high uncertainty because the neutron response of CaF2(Mn) TLDs may vary depending on the source of the material, and this procedure is a generic calculation applicable to CaF2(Mn) TLDs independent of their manufacturer/source. The neutron response given in this practice is a summary of CaF2(Mn) TLD responses reported in the literature. The associated uncertainty envelops the range of results reported and includes the variety of CaF2(Mn) TLDs used as well as the uncertainties in the det...
SCOPE
1.1 This practice describes a procedure for correcting a CaF2(Mn) thermoluminescence dosimeter (TLD) reading for its response to neutrons during the irradiation. The neutron response may be subtracted from the total TLD response to give the gamma-ray response. In fields with a large neutron contribution to the total response, this procedure may result in large uncertainties.
1.2 More precise experimental techniques may be applied if the uncertainty derived from this practice is larger than the level that the user can accept. These more precise techniques are not discussed here. The references in Section 8 describe some of these techniques.
1.3 This practice does not discuss effects on the TLD reading from neutron interactions with the material surrounding the TLD and used to ensure a charged particle equilibrium. These effects will depend on the isotopic composition of the surrounding material and its thickness, and on the incident neutron spectrum (1).2
1.4 The values stated in SI units are to be regarded as standard.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard6 pagesEnglish language
- Standard6 pagesEnglish language
SIGNIFICANCE AND USE
3.1 Because of the wide variety of materials being used in neutron-activation measurements, this guide is presented with the objective of bringing improved uniformity to the specific field of interest here: hardness testing of electronics primarily in critical assembly reactor environments.
Note 2: Some of the techniques discussed are useful for 14-MeV dosimetry. See Test Method E496 for activation detector materials suitable for 14-MeV neutron effects testing.
Note 3: The materials recommended in this guide are suitable for 252Cf or other weak source effects testing provided the fluence is sufficient to generate countable activities.
3.2 This guide is organized into two overlapping subjects: the criteria used for sensor selection, and the procedures used to ensure the proper determination of activities for determination of neutron spectra. See Terminology E170 and Test Methods E181. Determination of neutron spectra with activation sensor data is discussed in Guides E721 and E944.
SCOPE
1.1 This guide covers the selection and use of neutron-activation detector materials to be employed in neutron spectra adjustment techniques used for radiation-hardness testing of electronic semiconductor devices. Sensors are described that have been used at many radiation hardness-testing facilities, and comments are offered in table footnotes concerning the appropriateness of each reaction as judged by its cross-section accuracy, ease of use as a sensor, and by past successful application. This guide also discusses the fluence-uniformity, neutron self-shielding, and fluence-depression corrections that need to be considered in choosing the sensor thickness, the sensor covers, and the sensor locations. These considerations are relevant for the determination of neutron spectra from assemblies such as TRIGA- and Godiva-type reactors and from Californium irradiators. This guide may also be applicable to other broad energy distribution sources up to 20 MeV.
Note 1: For definitions on terminology used in this guide, see Terminology E170.
1.2 This guide also covers the measurement of the gamma-ray or beta-ray emission rates from the activation foils and other sensors as well as the calculation of the absolute specific activities of these foils. The principal measurement technique is high-resolution gamma-ray spectrometry. The activities are used in the determination of the energy-fluence spectrum of the neutron source. See Guide E721.
1.3 Details of measurement and analysis are covered as follows:
1.3.1 Corrections involved in measuring the sensor activities include those for finite sensor size and thickness in the calibration of the gamma-ray detector, for pulse-height analyzer deadtime and pulse-pileup losses, and for background radioactivity.
1.3.2 The primary method for detector calibration that uses secondary standard gamma-ray emitting sources is considered in this guide and in Test Methods E181. In addition, an alternative method in which the sensors are activated in the known spectrum of a benchmark neutron field is discussed in Guide E1018.
1.3.3 A data analysis method is presented which accounts for the following: detector efficiency; background subtraction; irradiation, waiting, and counting times; fission yields and gamma-ray branching ratios; and self-absorption of gamma rays and neutrons in the sensors.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decis...
- Guide13 pagesEnglish language
- Guide13 pagesEnglish language
SIGNIFICANCE AND USE
4.1 It is important to know the energy spectrum of the particular neutron source employed in radiation-hardness testing of electronic devices in order to relate radiation effects with device performance degradation.
4.2 This guide describes the factors which must be considered when the spectrum adjustment methodology is chosen and implemented. Although the selection of sensors (foils) and the determination of responses (activities) is discussed in Guide E720, the experiment should not be divorced from the analysis. In fact, it is advantageous for the analyst conducting the spectrum determination to be closely involved with the design of the experiment to ensure that the data obtained will provide the most accurate spectrum possible. These data include the following: (1) measured responses such as the activities of foils exposed in the environment and their uncertainties, (2) response functions such as reaction cross sections along with appropriate correlations and uncertainties, (3) the geometry and materials in the test environment, and (4) a trial function or prior spectrum and its uncertainties obtained from a transport calculation or from previous experience.
SCOPE
1.1 This guide covers procedures for determining the energy-differential fluence spectra of neutrons used in radiation-hardness testing of electronic semiconductor devices. The types of neutron sources specifically covered by this guide are fission or degraded energy fission sources used in either a steady-state or pulse mode.
1.2 This guide provides guidance and criteria that can be applied during the process of choosing the spectrum adjustment methodology that is best suited to the available data and relevant for the environment being investigated.
1.3 This guide is to be used in conjunction with Guide E720 to characterize neutron spectra and is used in conjunction with Practice E722 to characterize damage-related parameters normally associated with radiation-hardness testing of electronic semiconductor devices.
Note 1: Although Guide E720 only discusses activation foil sensors, any energy-dependent neutron-responding sensor for which a response function is known may be used (1).2
Note 2: For terminology used in this guide, see Terminology E170.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide12 pagesEnglish language
- Guide12 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Refer to Practice E261 for a general discussion of the measurement of fast-neutron fluence rates with threshold detectors.
5.5.1 Fig. 5 (2) shows how the neutron energy depends upon the angle of scattering in the laboratory coordinate system when the incident deuteron has an energy of 150 keV and is incident on a thick and a thin tritiated target. For thick targets, the incident deuteron loses energy as it penetrates the target and produces neutrons of lower energy. A thick target is defined as a target thick enough to completely stop the incident deuteron. The two curves in Fig. 5, for both thick and thin targets, come from different sources. The dashed line calculations come from Ref (3); the solid curve calculations come from Ref (4); and the measured data come from Ref (5). The dash-dot curve and the right-hand axis give the difference between the calculated neutron energies for thin and thick targets. Computer codes are available to assist in calculating the expected thick and thin target yield and neutron spectrum for various incident deuteron energies (6).
FIG. 5 Dependence of 3H(d,n)4He Neutron Energy on Angle (2)
5.6 The Q-value for the primary 3H(d,n)4He reaction is +17.59 MeV. When the incident deuteron energy exceeds 3.71 MeV and 4.92 MeV, the break-up reactions 3H(d,np)3H and 3H(d,2n)3He, respectively, become energetically possible. Thus, at high deuteron energies (>3.71 MeV) this reaction is no longer monoenergetic. Monoenergetic neutron beams with energies from about 14.8 to 20.4 MeV can be produced by this reaction at forward laboratory angles (7).
5.7 It is recommended that the dosimetry sensors be fielded in the exact positions where the dosimetry results are wanted. There are a number of factors that can affect the monochromaticity or energy spread of the neutron beam (7, 8). These factors include the energy regulation of the incident deuteron energy, energy loss in retaining windows if a gas target is used or energy loss within t...
SCOPE
1.1 This test method covers a general procedure for the measurement of the fast-neutron fluence rate produced by neutron generators utilizing the 3H(d,n)4He reaction. Neutrons so produced are usually referred to as 14-MeV neutrons, but range in energy depending on a number of factors. This test method does not adequately cover fusion sources where the velocity of the plasma may be an important consideration.
1.2 This test method uses threshold activation reactions to determine the average energy of the neutrons and the neutron fluence at that energy. At least three activities, chosen from an appropriate set of dosimetry reactions, are required to characterize the average energy and fluence. The required activities are typically measured by gamma-ray spectroscopy.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard14 pagesEnglish language
SIGNIFICANCE AND USE
4.1 A radiation-hardness assurance program requires a methodology for relating radiation-induced changes in materials exposed to a variety of particle species over a wide range of energies, including those encountered in spacecraft and in terrestrial environments as well as those produced by particle accelerators and nuclear fission and fusion reactors.
4.2 A major source of radiation damage in electronic and photonic devices and materials is the displacement of atoms from their normal lattice site. An appropriate exposure parameter for such damage is the damage energy calculated from NIEL by means of Eq 2. Other analogous measures, which may be used to characterize the irradiation history that is relevant to displacement damage, are damage energy per atom or per unit mass (displacement kerma, when the primary particles are neutral), and displacements per atom (dpa). See Terminology E170 for definitions of those quantities.
4.3 Each of the quantities mentioned in the previous paragraph should convey similar information, but in a different format. In each case the value of the derived exposure parameter depends on approximate nuclear, atomic, and lattice models, and on measured or calculated cross sections. If consistent comparisons are to be made of irradiation effects caused by different particle species and energies, it is essential that these approximations be consistently applied.
4.4 No correspondence should be assumed to exist between damage energy as calculated from NIEL and a particular change in a material property or device parameter. Instead, the damage energy should be used as a parameter which describes the exposure. It may be a useful correlation variate, even when different particle species and energies are included. NIEL should not be reported as a measure of damage, however, unless its correlation with a particular damage modality has been demonstrated in that material or device.
4.5 NIEL is a construct that depends on a model of the par...
SCOPE
1.1 This practice describes a procedure for characterizing particle irradiations of materials in terms of non-ionizing energy loss (NIEL). NIEL is used in published literature to characterize both charged and neutral particle irradiations.
1.2 Although the methods described in this practice apply to any particles and target materials for which displacement cross sections are known (see Practice E521), this practice is intended for use in irradiations in which observed damage effects may be correlated with atomic displacements. This is true of some, but not all, radiation effects in electronic and photonic materials.
1.3 Procedures analogous to this one are used for calculation of displacements per atom (dpa) in charged particle irradiations (see Practice E521) or neutron irradiations (see Practice E693).
1.4 Guidance on calculation of dpa from NIEL is provided.
1.5 Procedures related to this one are used for calculation of 1-MeV equivalent neutron fluence in electronic materials (see Practice E722), but in that practice the concept of damage efficiency, based on correlation of observed damage effects, is included.
1.6 Guidance on conversion of NIEL in silicon to monoenergetic neutron fluence in silicon (see Practice E722), and vice versa, is provided.
1.7 The application of this standard requires knowledge of the particle fluence and energy distribution of particles whose interaction leads to displacement damage.
1.8 The correlation of radiation effects data is beyond the scope of this standard. A comprehensive review (1)2 of displacement damage effects in silicon and their correlation with NIEL provides appropriate guidance that is applicable to semiconductor materials and electronic devices.
1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standa...
- Standard5 pagesEnglish language
SIGNIFICANCE AND USE
5.1 An accurate measure of the total absorbed dose is necessary to ensure the validity of the data taken, to enable comparison to be made of data taken at different facilities, and to verify that components or circuits are tested to the radiation specification applied to the system for which they are to be used.
5.2 The primary value of a calorimetric method for measuring dose is that the results are absolute. They are based only on physical properties of materials, that is, the specific heat of the calorimeter-block material and the Seebeck EMF of the thermocouple used or the temperature coefficient of resistance (α) of the thermistor used, all of which can be established with non-radiation measurements.
5.3 The method permits repeated measurements to be made without requiring entry into the radiation cell between measurements.
SCOPE
1.1 This test method covers a calorimetric measurement of the total absorbed dose delivered by a single pulse of electrons from an electron linear accelerator or a flash X-ray machine (FXR, e-beam mode). The test method is designed for use with pulses of electrons in the energy range from 10 to 50 MeV and is only valid for cases in which both the calorimeter and the test specimen to be irradiated are “thin” compared to the range of these electrons in the materials of which they are constructed.
1.2 The procedure described can be used in those cases in which (1) the dose delivered in a single pulse is 5 Gy(matl)2 [500 rd (matl)] or greater, or (2) multiple pulses of a lower dose can be delivered in a short time compared to the thermal time constant of the calorimeter. The units for the total absorbed dose delivered to a material require the specification of the material and the notation “matl” refers to the active material of the calorimeter. The minimum dose per pulse that can be acceptably monitored depends on the variables of the particular test, including pulse rate, pulse uniformity, and the thermal time constant of the calorimeter.
1.3 A determination of the total dose is made directly for the material of which the calorimeter block is made. The total dose in other materials can be calculated from this measured value using Eq 3 presented in this test method. The need for such calculations and the choice of materials for which calculations are to be made shall be subject to agreement by the parties to the test.
1.4 The values stated in SI units are to be regarded as the standard. The values in parenthesis are provided for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard13 pagesEnglish language
- Standard13 pagesEnglish language
SIGNIFICANCE AND USE
4.1 The absorbed dose is a more meaningful parameter than exposure for use in relating the effects of radiation on materials. It expresses the energy absorbed by the irradiated material per unit mass, whereas exposure is related to the amount of charge produced in air per unit mass. Absorbed dose, as referred to here, implies that the measurement is made under conditions of charged particle (electron) equilibrium (see Appendix X1). In practice, such conditions are not rigorously achievable but, under some circumstances, can be approximated closely.
4.2 Different materials, when exposed to the same radiation field, absorb different amounts of energy. Using the techniques of this standard, charged particle equilibrium must exist in order to relate the absorbed dose in one material to the absorbed dose in another. Also, if the radiation is attenuated by a significant thickness of an absorber, the energy spectrum of the radiation will be changed, and it will be necessary to correct for this.
Note 1: For comprehensive discussions of various dosimetry methods applicable to the radiation types and energies and absorbed dose rate ranges discussed in this method, see ICRU Reports 34 and 80.
SCOPE
1.1 This practice presents a technique for calculating the absorbed dose in a material from knowledge of the radiation field, the composition of the material, (1-5)2,3 and a related measurement. The procedure is applicable for X and gamma radiation provided the energy of the photons fall within the range from 0.01 to 20 MeV.
1.2 A method is given for calculating the absorbed dose in a material from the knowledge of the absorbed dose in another material exposed to the same radiation field. The procedure is restricted to homogeneous materials composed of the elements for which absorption coefficients have been tabulated. All 92 natural elements are tabulated in (2). It also requires some knowledge of the energy spectrum of the radiation field produced by the source under consideration. Generally, the accuracy of this method is limited by the accuracy to which the energy spectrum of the radiation field is known.
1.3 The results of this practice are only valid if charged particle equilibrium exists in the material and at the depth of interest. Thus, this practice is not applicable for determining absorbed dose in the immediate vicinity of boundaries between materials of widely differing atomic numbers. For more information on this topic, see Practice E1249.
1.4 Energy transport computer codes4 exist that are formulated to calculate absorbed dose in materials more precisely than this method. To use these codes, more effort, time, and expense are required. If the situation warrants, such calculations should be used rather than the method described here.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard9 pagesEnglish language
- Standard9 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Division of the Co-60 Hardness Testing into Five Parts:
4.1.1 The equilibrium absorbed dose shall be measured with a dosimeter, such as a TLD, located adjacent to the device under test. Alternatively, a dosimeter may be irradiated in the position of the device before or after irradiation of the device.
4.1.2 This absorbed dose measured by the dosimeter shall be converted to the equilibrium absorbed dose in the material of interest within the critical region within the device under test, for example the SiO2 gate oxide of an MOS device.
4.1.3 A correction for absorbed-dose enhancement effects shall be considered. This correction is dependent upon the photon energy that strikes the device under test.
4.1.4 A correlation should be made between the absorbed dose in the critical region (for example, the gate oxide mentioned in 4.1.2) and some electrically important effect (such as charge trapped at the Si/SiO2 interface as manifested by a shift in threshold voltage).
4.1.5 An extrapolation should then be made from the results of the test to the results that would be expected for the device under test under actual operating conditions.
Note 5: The parts of a test discussed in 4.1.2 and 4.1.3 are the subject of this practice. The subject of 4.1.1 is covered and referenced in other standards such as Practice E668 and ICRU Report 14. The parts of a test discussed in 4.1.4 and 4.1.5 are outside the scope of this practice.
4.2 Low-Energy Components in the Spectrum—Some of the primary Co-60 gamma rays (1.17 and 1.33 MeV) produce lower energy photons by Compton scattering within the Co-60 source structure, within materials that lie between the source and the device under test, and within materials that lie beyond the device but contribute to backscattering. As a result of the complexity of these effects, the photon energy spectrum striking the device usually is not well known. This point is further discussed in Section 5 and Appendix X1. The presence of ...
SCOPE
1.1 This practice covers recommended procedures for the use of dosimeters, such as thermoluminescent dosimeters (TLD's), to determine the absorbed dose in a region of interest within an electronic device irradiated using a Co-60 source. Co-60 sources are commonly used for the absorbed dose testing of silicon electronic devices.
Note 1: This absorbed-dose testing is sometimes called “total dose testing” to distinguish it from “dose rate testing.”
Note 2: The effects of ionizing radiation on some types of electronic devices may depend on both the absorbed dose and the absorbed dose rate; that is, the effects may be different if the device is irradiated to the same absorbed-dose level at different absorbed-dose rates. Absorbed-dose rate effects are not covered in this practice but should be considered in radiation hardness testing.
1.2 The principal potential error for the measurement of absorbed dose in electronic devices arises from non-equilibrium energy deposition effects in the vicinity of material interfaces.
1.3 Information is given about absorbed-dose enhancement effects in the vicinity of material interfaces. The sensitivity of such effects to low energy components in the Co-60 photon energy spectrum is emphasized.
1.4 A brief description is given of typical Co-60 sources with special emphasis on the presence of low energy components in the photon energy spectrum output from such sources.
1.5 Procedures are given for minimizing the low energy components of the photon energy spectrum from Co-60 sources, using filtration. The use of a filter box to achieve such filtration is recommended.
1.6 Information is given on absorbed-dose enhancement effects that are dependent on the device orientation with respect to the Co-60 source.
1.7 The use of spectrum filtration and appropriate device orientation provides a radiation environment whereby the absorbed dose in the sensitive region of an elect...
- Standard17 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Absorbed dose in a material is an important parameter that can be correlated with radiation effects produced in electronic components and devices that are exposed to ionizing radiation. Reasonable estimates of this parameter can be calculated if knowledge of the source radiation field (that is, energy spectrum and particle fluence) is available. Sufficiently detailed information about the radiation field is generally not available. However, measurements of absorbed dose with passive dosimeters in a radiation test facility can provide information from which the absorbed dose in a material of interest can be inferred. Under certain prescribed conditions, TLDs are quite suitable for performing such measurements.
Note 2: For comprehensive discussions of various dosimetry methods applicable to the radiation types and energy and absorbed dose-rate range discussed in this practice, see ICRU Reports 14, 17, 21, and 34.
SCOPE
1.1 This practice covers procedures for the use of thermoluminescence dosimeters (TLDs) to determine the absorbed dose in a material irradiated by ionizing radiation. Although some elements of the procedures have broader application, the specific area of concern is radiation-hardness testing of electronic devices. This practice is applicable to the measurement of absorbed dose in materials irradiated by gamma rays, X rays, and electrons of energies from 12 to 60 MeV. Specific energy limits are covered in appropriate sections describing specific applications of the procedures. The range of absorbed dose covered is approximately from 10−2 to 104 Gy (1 to 106 rad), and the range of absorbed dose rates is approximately from 10−2 to 1010 Gy/s (1 to 1012 rad/s). Absorbed dose and absorbed dose-rate measurements in materials subjected to neutron irradiation are not covered in this practice. (See Practice E2450 for guidance in mixed fields.) Further, the portion of these procedures that deal with electron irradiation are primarily intended for use in parts testing. Testing of devices as a part of more massive components such as electronics boards or boxes may require techniques outside the scope of this practice.
Note 1: The purpose of the upper and lower limits on the energy for electron irradiation is to approach a limiting case where dosimetry is simplified. Specifically, the dosimetry methodology specified requires that the following three limiting conditions be approached: (a) energy loss of the primary electrons is small, (b) secondary electrons are largely stopped within the dosimeter, and (c) bremsstrahlung radiation generated by the primary electrons is largely lost.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard19 pagesEnglish language
- Standard19 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Although Co-60 nuclei only emit monoenergetic gamma rays at 1.17 and 1.33 MeV, the finite thickness of sources, and encapsulation materials and other surrounding structures that are inevitably present in irradiators can contribute a substantial amount of low-energy gamma radiation, principally by Compton scattering (1, 2).3 In radiation-hardness testing of electronic devices this low-energy photon component of the gamma spectrum can introduce significant dosimetry errors for a device under test since the equilibrium absorbed dose as measured by a dosimeter can be quite different from the absorbed dose deposited in the device under test because of absorbed dose enhancement effects (3, 4). Absorbed dose enhancement effects refer to the deviations from equilibrium absorbed dose caused by non-equilibrium electron transport near boundaries between dissimilar materials.
4.2 The ionization chamber technique described in this method provides an easy means for estimating the importance of the low-energy photon component of any given irradiator type and configuration.
4.3 When there is an appreciable low-energy spectral component present in a particular irradiator configuration, special experimental techniques should be used to ensure that dosimetry measurements adequately represent the absorbed dose in the device under test. (See Practice E1249.)
SCOPE
1.1 Low energy components in the photon energy spectrum of Co-60 irradiators lead to absorbed dose enhancement effects in the radiation-hardness testing of silicon electronic devices. These low energy components may lead to errors in determining the absorbed dose in a specific device under test. This method covers procedures for the use of a specialized ionization chamber to determine a figure of merit for the relative importance of such effects. It also gives the design and instructions for assembling this chamber.
1.2 This method is applicable to measurements in Co-60 radiation fields where the range of exposure rates is 7 × 10 −6 to 3 × 10−2 C kg −1 s−1 (approximately 100 R/h to 100 R/s). For guidance in applying this method to radiation fields where the exposure rate is >100 R/s, see Appendix X1.
Note 1: See Terminology E170 for definition of exposure and its units.
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard10 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Refer to Guides E720 and E844 for the selection, irradiation, and quality control of neutron dosimeters.
5.2 Refer to Practice E261 for a general discussion of the determination of fast-neutron fluence and fluence rate with threshold detectors.
5.3 The activation reaction produces 32P, which decays by the emission of a single beta particle in 100 % of the decays, and which emits no gamma rays. The half life of 32P is 14.284 (36)3 days (1) 4 and the maximum beta energy is 1710.66 (21) keV (1).
5.4 Elemental sulfur is readily available in pure form and any trace contaminants present do not produce significant amounts of radioactivity. Natural sulfur, however, is composed of 32S (94.99 % (26)), 34S (4.25 % (24)) (2), and trace amounts of other sulfur isotopes. The presence of these other isotopes leads to several competing reactions that can interfere with the counting of the 1710-keV beta particle. This interference can usually be eliminated by the use of appropriate techniques, as discussed in Section 8.
SCOPE
1.1 This test method describes procedures for measuring reaction rates and fast-neutron fluences by the activation reaction 32S(n,p)32P.
1.2 This activation reaction is useful for measuring neutrons with energies above approximately 3 MeV.
1.3 With suitable techniques, fission-neutron fluences from about 5 × 108 to 1016 n/cm 2 can be measured.
1.4 Detailed procedures for other fast-neutron detectors are described in Practice E261.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard6 pagesEnglish language
SIGNIFICANCE AND USE
5.1 The neutron test spectrum must be known in order to use a measured device response to predict the device performance in an operational environment (Practice E1854). Typically, neutron spectra are determined using a set of sensors with response functions sensitive over the neutron energy region to which the device under test (DUT) responds (Guide E721). For silicon bipolar devices exposed in reactor neutron spectra, this effective energy range is between 0.01 and 10 MeV. A typical set of activation reactions that lack fission reactions from nuclides such as 235U, 237Np, or 239Pu, will have very poor sensitivity to the spectrum between 0.01 and 2 MeV. For a pool-type reactor spectrum, 70 % of the DUT electronic damage response may lie in this range making its determination of critical importance.
5.2 When dosimeters with a significant response in the 0.01 to 2 MeV energy region, such as fission foils, are unavailable, silicon transistors can provide a dosimeter with the needed response to define the spectrum in this critical energy range. When fission foils are part of the sensor set, the silicon sensor provides confirmation of the spectral shape in this energy region.
5.3 Silicon bipolar transistors, such as type 2N2222A, are inexpensive, smaller than fission foils contained in a boron ball, and sensitive to a part of the neutron spectrum important to the damage of modern silicon electronics. They also can be used directly in arrays to spatially map 1-MeV(Si) equivalent displacement damage fluence. The proper set of steps to take in reading the transistor-gain degradation is described in this test method.
5.4 The energy-dependence of the displacement damage function for silicon is found in Practice E722. The major portion of the response for the silicon transistors will generally be above 100 keV.
SCOPE
1.1 This test method covers the use of 2N2222A silicon bipolar transistors as dosimetry sensors in the determination of neutron energy spectra and as 1-MeV(Si) equivalent displacement damage fluence monitors.
1.2 The neutron displacement in silicon can serve as a neutron spectrum sensor in the range 0.1 to 2.0 MeV and can serve as a substitute when fission foils are not available. It has been applied in the fluence range between 2 × 1012 n/cm 2 to 1 × 1014 n/cm2 and should be useful up to 1 × 1015 n/cm2. This test method details the acquisition and use of 1-MeV(Si) equivalent fluence information for the partial determination of the neutron spectra by using 2N2222A transistors.
1.3 This sensor yields a direct measurement of the silicon 1-MeV equivalent fluence by the transfer technique.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard10 pagesEnglish language
- Standard10 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Overtesting should be done when (a) testing by variables is impractical because of time and cost considerations or because the probability distribution of stress to failure cannot be estimated with sufficient accuracy, and (b) an unrealistically large number of parts would have to be tested at the specification stress for the necessary confidence and survival probability.
SCOPE
1.1 This guide covers the use of overtesting in order to reduce the required number of parts that must be tested to meet a given quality acceptance standard. Overtesting is testing a sample number of parts at a stress level higher than their specification stress in order to reduce the amount of necessary data taking. This guide discusses when and how overtesting may be applied to forming probabilistic estimates for the survival of electronic piece parts subjected to radiation stress. Some knowledge of the probability distribution governing the stress-to-failure of the parts is necessary, although exact knowledge may be replaced by over-conservative estimates of this distribution.
1.2 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
4.1 This practice was written primarily to guide test participants in establishing, identifying, maintaining, and using suitable environments for conducting high quality neutron tests. Its development was motivated, in large measure, because inadequate controls in the neutron-effects-test process have, in some past instances, resulted in exposures that have differed by factors of three or more from irradiation specifications. A radiation test environment generally differs from the environment in which the electronics must operate (the operational environment); therefore, a high quality test requires not only the use of a suitable radiation environment, but also control and compensation for contributions to damage that differ from those in the operational environment. In general, the responsibility for identifying suitable test environments to accomplish test objectives lies with the sponsor/user/tester and test specialist part of the team, with the assistance of an independent validator, if available. The responsibility for the establishment and maintenance of suitable environments lies with the facility operator/dosimetrist and test specialist, again with the possible assistance of an independent validator. Additional guidance on the selection of an irradiation facility is provided in Practice F1190.
4.2 This practice identifies the tasks that must be accomplished to ensure a successful high quality test. It is the overall responsibility of the sponsor or user to ensure that all of the required tasks are complete and conditions are met. Other participants provide appropriate documentation to enable the sponsor or user to make that determination.
4.3 The principal determinants of a properly conducted test are: (1) the radiation test environment shall be well characterized, controlled, and correlated with the specified irradiation levels; (2) damage produced in the electronic materials and devices is caused by the desired, specified component of the environment...
SCOPE
1.1 This practice sets forth requirements to ensure consistency in neutron-induced displacement damage testing of silicon and gallium arsenide electronic piece parts. This requires controls on facility, dosimetry, tester, and communications processes that affect the accuracy and reproducibility of these tests. It provides background information on the technical basis for the requirements and additional recommendations on neutron testing.
1.2 Methods are presented for ensuring and validating consistency in neutron displacement damage testing of electronic parts such as integrated circuits, transistors, and diodes. The issues identified and the controls set forth in this practice address the characterization and suitability of the radiation environments. They generally apply to reactor sources, accelerator-based neutron sources, such as 14-MeV DT sources, and 252Cf sources. Facility and environment characteristics that introduce complications or problems are identified, and recommendations are offered to recognize, minimize or eliminate these problems. This practice may be used by facility users, test personnel, facility operators, and independent process validators to determine the suitability of a specific environment within a facility and of the testing process as a whole. Electrical measurements are addressed in other standards, such as Guide F980. Additional information on conducting irradiations can be found in Practices E798 and F1190. This practice also may be of use to test sponsors (organizations that establish test specifications or otherwise have a vested interest in the performance of electronics in neutron environments).
1.3 Methods for the evaluation and control of undesired contributions to damage are discussed in this practice. References to relevant ASTM standards and technical reports are provided. Processes and methods used to arrive at the appropriate test environments and specification levels f...
- Standard13 pagesEnglish language
- Standard13 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This practice is important in characterizing the radiation hardness of electronic devices irradiated by neutrons. This characterization makes it feasible to predict some changes in operational properties of irradiated semiconductor devices or electronic systems. To facilitate uniformity of the interpretation and evaluation of results of irradiations by sources of different fluence spectra, it is convenient to reduce the incident neutron fluence from a source to a single parameter—an equivalent monoenergetic neutron fluence—applicable to a particular semiconductor material.
5.2 In order to determine an equivalent monoenergetic neutron fluence, it is necessary to evaluate the displacement damage of the particular semiconductor material. Ideally, this quantity is correlated to the degradation of a specific functional performance parameter (such as current gain) of the semiconductor device or system being tested. However, this correlation has not been established unequivocally for all device types and performance parameters since, in many instances, other effects also can be important. Ionization effects produced by the incident neutron fluence or by gamma rays in a mixed neutron fluence, short-term and long-term annealing, and other factors can contribute to observed performance degradation (damage). Thus, caution should be exercised in making a correlation between calculated displacement damage and performance degradation of a given electronic device. The types of devices for which this correlation is applicable, and numerical evaluation of displacement damage are discussed in the annexes.
5.3 The concept of 1-MeV equivalent fluence is widely used in the radiation-hardness testing community. It has merits and disadvantages that have been debated widely (9-12). For these reasons, specifics of a standard application of the 1-MeV equivalent fluence are presented in the annexes.
SCOPE
1.1 This practice covers procedures for characterizing neutron fluence from a source in terms of an equivalent monoenergetic neutron fluence. It is applicable to neutron effects testing, to the development of test specifications, and to the characterization of neutron test environments. The sources may have a broad neutron-energy range, or may be mono-energetic neutron sources with energies up to 20 MeV. This practice is not applicable in cases where the predominant source of displacement damage is from neutrons of energy less than 10 keV. The relevant equivalence is in terms of a specified effect on certain physical properties of materials upon which the source spectrum is incident. In order to achieve this, knowledge of the effects of neutrons as a function of energy on the specific property of the material of interest is required. Sharp variations in the effects with neutron energy may limit the usefulness of this practice in the case of mono-energetic sources.
1.2 This practice is presented in a manner to be of general application to a variety of materials and sources. Correlation between displacements (1-3)2 caused by different particles (electrons, neutrons, protons, and heavy ions) is out of the scope of this practice but is addressed in Practice E3084. In radiation-hardness testing of electronic semiconductor devices, specific materials of interest include silicon and gallium arsenide, and the neutron sources generally are test and research reactors and californium-252 irradiators.
1.3 The technique involved relies on the following factors: (1) a detailed determination of the fluence spectrum of the neutron source, and (2) a knowledge of the degradation (damage) effects of neutrons as a function of energy on specific material properties.
1.4 The detailed determination of the neutron fluence spectrum referred to in 1.3 need not be performed afresh for each test exposure, provided the exposure conditions are repeatable. When the spectrum determination is not repeated, a neutron fluence monit...
- Standard28 pagesEnglish language
- Standard28 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Semiconductor devices can be permanently damaged by neutrons (1, 2)6. The effect of such damage on the performance of an electronic component can be determined by measuring the component’s electrical characteristics before and after exposure to fast neutrons in the neutron fluence range of interest. The resulting data can be utilized in the design of electronic circuits that are tolerant of the degradation exhibited by that component.
5.2 This guide provides a method by which the exposure of silicon and gallium arsenide semiconductor devices to neutron irradiation may be performed in a manner that is repeatable and which will allow comparison to be made of data taken at different facilities.
5.3 For semiconductors other than silicon and gallium arsenide, applicable validated 1-MeV damage functions are not available in codified National standards. In the absence of a validated 1-MeV damage function, the non-ionizing energy loss (NIEL) or the displacement kerma, as a function of incident neutron energy, normalized to the response in the 1 MeV energy region, may be used as an approximation. See Practice E722 for a description of the method used to determine the damage functions in Si and GaAs (3).
SCOPE
1.1 This guide strictly applies only to the exposure of unbiased silicon (Si) or gallium arsenide (GaAs) semiconductor components (integrated circuits, transistors, and diodes) to neutron radiation to determine the permanent damage in the components. Validated 1-MeV displacement damage functions codified in National Standards are not currently available for other semiconductor materials.
1.2 Elements of this guide, with the deviations noted, may also be applicable to the exposure of semiconductors comprised of other materials except that validated 1-MeV displacement damage functions codified in National standards are not currently available.
1.3 Only the conditions of exposure are addressed in this guide. The effects of radiation on the test sample should be determined using appropriate electrical test methods.
1.4 This guide addresses those issues and concerns pertaining to irradiations with neutrons.
1.5 System and subsystem exposures and test methods are not included in this guide.
1.6 The range of interest for neutron fluence in displacement damage semiconductor testing range from approximately 109 to 1016 1-MeV n/cm2.
1.7 This guide does not address neutron-induced single or multiple neutron event effects or transient annealing.
1.8 This guide provides an alternative to Test Method 1017, Neutron Displacement Testing, a component of MIL-STD-883 and MIL-STD-750.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide6 pagesEnglish language
- Guide6 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Electronic circuits used in space, military, and nuclear power systems may be exposed to various levels of ionizing radiation. It is essential for the design and fabrication of such circuits that test methods be available that can determine the vulnerability or hardness (measure of nonvulnerability) of components to be used in such systems.
5.2 Some manufacturers currently are selling semiconductor parts with guaranteed hardness ratings. Use of this guide provides a basis for standardized qualification and acceptance testing.
SCOPE
1.1 This guide presents background and guidelines for establishing an appropriate sequence of tests and data analysis procedures for determining the ionizing radiation (total dose) hardness of microelectronic devices for dose rates below 300 rd(SiO2)/s. These tests and analysis will be appropriate to assist in the determination of the ability of the devices under test to meet specific hardness requirements or to evaluate the parts for use in a range of radiation environments.
1.2 The methods and guidelines presented will be applicable to characterization, qualification, and lot acceptance of silicon-based MOS and bipolar discrete devices and integrated circuits. They will be appropriate for treatment of the effects of electron and photon irradiation.
1.3 This guide provides a framework for choosing a test sequence based on general characteristics of the parts to be tested and the radiation hardness requirements or goals for these parts.
1.4 This guide provides for tradeoffs between minimizing the conservative nature of the testing method and minimizing the required testing effort.
1.5 Determination of an effective and economical hardness test typically will require several kinds of decisions. A partial enumeration of the decisions that typically must be made is as follows:
1.5.1 Determination of the Need to Perform Device Characterization—For some cases it may be more appropriate to adopt some kind of worst case testing scheme that does not require device characterization. For other cases it may be most effective to determine the effect of dose-rate on the radiation sensitivity of a device. As necessary, the appropriate level of detail of such a characterization also must be determined.
1.5.2 Determination of an Effective Strategy for Minimizing the Effects of Irradiation Dose Rate on the Test Result—The results of radiation testing on some types of devices are relatively insensitive to the dose rate of the radiation applied in the test. In contrast, many MOS devices and some bipolar devices have a significant sensitivity to dose rate. Several different strategies for managing the dose rate sensitivity of test results will be discussed.
1.5.3 Choice of an Effective Test Methodology—The selection of effective test methodologies will be discussed.
1.6 Low Dose Requirements—Hardness testing of MOS and bipolar microelectronic devices for the purpose of qualification or lot acceptance is not necessary when the required hardness is 100 rd(SiO2) or lower.
1.7 Sources—This guide will cover effects due to device testing using irradiation from photon sources, such as 60Co γ irradiators, 137Cs γ irradiators, and low energy (approximately 10 keV) X-ray sources. Other sources of test radiation such as linacs, Van de Graaff sources, Dymnamitrons, SEMs, and flash X-ray sources occasionally are used but are outside the scope of this guide.
1.8 Displacement damage effects are outside the scope of this guide, as well.
1.9 The values stated in SI units are to be regarded as the standard.
1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide41 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Many modern integrated circuits, power transistors, and other devices experience SEP when exposed to cosmic rays in interplanetary space, in satellite orbits or during a short passage through trapped radiation belts. It is essential to be able to predict the SEP rate for a specific environment in order to establish proper techniques to counter the effects of such upsets in proposed systems. As the technology moves toward higher density ICs, the problem is likely to become even more acute.
5.2 This guide is intended to assist experimenters in performing ground tests to yield data enabling SEP predictions to be made.
SCOPE
1.1 This guide defines the requirements and procedures for testing integrated circuits and other devices for the effects of single event phenomena (SEP) induced by irradiation with heavy ions having an atomic number Z ≥ 2. This description specifically excludes the effects of neutrons, protons, and other lighter particles that may induce SEP via another mechanism. SEP includes any manifestation of upset induced by a single ion strike, including soft errors (one or more simultaneous reversible bit flips), hard errors (irreversible bit flips), latchup (persistent high conducting state), transients induced in combinatorial devices which may introduce a soft error in nearby circuits, power field effect transistor (FET) burn-out and gate rupture. This test may be considered to be destructive because it often involves the removal of device lids prior to irradiation. Bit flips are usually associated with digital devices and latchup is usually confined to bulk complementary metal oxide semiconductor, (CMOS) devices, but heavy ion induced SEP is also observed in combinatorial logic programmable read only memory, (PROMs), and certain linear devices that may respond to a heavy ion induced charge transient. Power transistors may be tested by the procedure called out in Method 1080 of MIL STD 750.
1.2 The procedures described here can be used to simulate and predict SEP arising from the natural space environment, including galactic cosmic rays, planetary trapped ions, and solar flares. The techniques do not, however, simulate heavy ion beam effects proposed for military programs. The end product of the test is a plot of the SEP cross section (the number of upsets per unit fluence) as a function of ion LET (linear energy transfer or ionization deposited along the ion's path through the semiconductor). This data can be combined with the system's heavy ion environment to estimate a system upset rate.
1.3 Although protons can cause SEP, they are not included in this guide. A separate guide addressing proton induced SEP is being considered.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide11 pagesEnglish language
SIGNIFICANCE AND USE
5.1 PN Junction Diode—The steady-state photocurrent of a simple p-n junction diode is a directly measurable quantity that can be directly related to device response over a wide range of ionizing radiation. For more complex devices the junction photocurrent may not be directly related to device response.
5.2 Zener Diode—In this device, the effect of the photocurrent on the Zener voltage rather than the photocurrent itself is usually most important. The device is most appropriately tested while biased in the Zener region. In testing Zener diodes or precision voltage regulators, extra precaution must be taken to make certain the photocurrent generated in the device during irradiations does not cause the voltage across the device to change during the test.
5.3 Bipolar Transistor—As device geometries dictate that photocurrent from the base-collector junction be much greater than current from the base-emitter junction, measurements are usually made only on the collector-base junction with emitter open; however, sometimes, to obtain data for computer-aided circuit analysis, the emitter-base junction photocurrent is also measured.
5.4 Junction Field-Effect Device—A proper photocurrent measurement requires that the source be shorted (dc) to the drain during measurement of the gate-channel photocurrent. In tetrode-connected devices, the two gate-channel junctions should be monitored separately.
5.5 Insulated Gate Field-Effect Device—In this type of device, the true photocurrent is between the substrate and the channel, source, and drain regions. A current which can generate voltage that will turn on the device may be measured by the technique used here, but it is due to induced conductivity in the gate insulator and thus is not a junction photocurrent.
SCOPE
1.1 This test method covers the measurement of steady-state primary photocurrent, Ipp, generated in semiconductor devices when these devices are exposed to ionizing radiation. These procedures are intended for the measurement of photocurrents greater than 10−9 A·s/Gy(Si or Ge), in cases for which the relaxation time of the device being measured is less than 25 % of the pulse width of the ionizing source. The validity of these procedures for ionizing dose rates as great as 108Gy(Si or Ge)/s has been established. The procedures may be used for measurements at dose rates as great as 1010Gy(Si or Ge)/s; however, extra care must be taken. Above 108Gy/s, the package response may dominate the device response for any device. Additional precautions are also required when measuring photocurrents of 10−9 A·s/Gy(Si or Ge) or lower.
1.2 Setup, calibration, and test circuit evaluation procedures are also included in this test method.
1.3 Because of the variability between device types and in the requirements of different applications, the dose rate range over which any specific test is to be conducted is not given in this test method but must be specified separately.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard7 pagesEnglish language
- Standard7 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Electronic circuits used in many space, military and nuclear power systems may be exposed to various levels of ionizing radiation dose. It is essential for the design and fabrication of such circuits that test methods be available that can determine the vulnerability or hardness (measure of nonvulnerability) of components to be used in such systems.
4.2 Manufacturers are currently selling semiconductor parts with guaranteed hardness ratings, and the military specification system is being expanded to cover hardness specification for parts. Therefore test methods and guides are required to standardize qualification testing.
4.3 Use of low energy (≈10 keV) X-ray sources has been examined as an alternative to cobalt-60 for the ionizing radiation effects testing of microelectronic devices (3, 4, 5, 6). The goal of this guide is to provide background information and guidance for such use where appropriate.
Note 3: Cobalt-60—The most commonly used source of ionizing radiation for ionizing radiation (“total dose”) testing is cobalt-60. Gamma rays with energies of 1.17 and 1.33 MeV are the primary ionizing radiation emitted by cobalt-60. In exposures using cobalt-60 sources, test specimens must be enclosed in a lead-aluminum container to minimize dose-enhancement effects caused by low-energy scattered radiation (unless it has been demonstrated that these effects are negligible). For this lead-aluminum container, a minimum of 1.5 mm of lead surrounding an inner shield of 0.7 to 1.0 mm of aluminum is required. (See 8.2.2.2 and Practice E1249.)
4.4 The X-ray tester has proven to be a useful ionizing radiation effects testing tool because:
4.4.1 It offers a relatively high dose rate, in comparison to most cobalt-60 sources, thus offering reduced testing time.
4.4.2 The radiation is of sufficiently low energy that it can be readily collimated. As a result, it is possible to irradiate a single device on a wafer.
4.4.3 Radiation safety issues are more easily...
SCOPE
1.1 This guide covers recommended procedures for the use of X-ray testers (that is, sources with a photon spectrum having ≈10 keV mean photon energy and ≈50 keV maximum energy) in testing semiconductor discrete devices and integrated circuits for effects from ionizing radiation.
1.2 The X-ray tester may be appropriate for investigating the susceptibility of wafer level or delidded microelectronic devices to ionizing radiation effects. It is not appropriate for investigating other radiation-induced effects such as single-event effects (SEE) or effects due to displacement damage.
1.3 This guide focuses on radiation effects in metal oxide semiconductor (MOS) circuit elements, either designed (as in MOS transistors) or parasitic (as in parasitic MOS elements in bipolar transistors).
1.4 Information is given about appropriate comparison of ionizing radiation hardness results obtained with an X-ray tester to those results obtained with cobalt-60 gamma irradiation. Several differences in radiation-induced effects caused by differences in the photon energies of the X-ray and cobalt-60 gamma sources are evaluated. Quantitative estimates of the magnitude of these differences in effects, and other factors that should be considered in setting up test protocols, are presented.
1.5 If a 10-keV X-ray tester is to be used for qualification testing or lot acceptance testing, it is recommended that such tests be supported by cross checking with cobalt-60 gamma irradiations.
1.6 Comparisons of ionizing radiation hardness results obtained with an X-ray tester with results obtained with a LINAC, with protons, etc. are outside the scope of this guide.
1.7 Current understanding of the differences between the physical effects caused by X-ray and cobalt-60 gamma irradiations is used to provide an estimate of the ratio (number-of-holes-cobalt-60)/(number-of-holes-X-ray). Several cases are defined where the differences in the eff...
- Guide18 pagesEnglish language
- Guide18 pagesEnglish language
ABSTRACT
This practice covers procedures for irradiations at accelerator-based neutron sources. The discussion focuses on nearly monoenergetic 14-MeV neutrons from the deuterium-tritium T(d,n) interaction, and broad spectrum neutrons from stopping deuterium beams in thick beryllium or lithium targets. However, most of the recommendations also apply to other types of accelerator-based sources, including spallation neutron sources. The procedures to be considered include methods for characterizing the accelerator beam and target, the irradiated sample, and the neutron flux and spectrum, as well as procedures for recording and reporting irradiation data.
SCOPE
1.1 This practice covers procedures for irradiations at accelerator-based neutron sources. The discussion focuses on two types of sources, namely nearly monoenergetic 14-MeV neutrons from the deuterium-tritium T(d,n) interaction, and broad spectrum neutrons from stopping deuterium beams in thick beryllium or lithium targets. However, most of the recommendations also apply to other types of accelerator-based sources, including spallation neutron sources (1).2 Interest in spallation sources has increased recently due to their development of high-power, high-flux sources for neutron scattering and their proposed use for transmutation of fission reactor waste (2).
1.2 Many of the experiments conducted using such neutron sources are intended to provide a simulation of irradiation in another neutron spectrum, for example, that from a DT fusion reaction. The word simulation is used here in a broad sense to imply an approximation of the relevant neutron irradiation environment. The degree of conformity can range from poor to nearly exact. In general, the intent of these experiments is to establish the fundamental relationships between irradiation or material parameters and the material response. The extrapolation of data from such experiments requires that the differences in neutron spectra be considered.
1.3 The procedures to be considered include methods for characterizing the accelerator beam and target, the irradiated sample, and the neutron flux (fluence rate) and spectrum, as well as procedures for recording and reporting irradiation data.
1.4 Other experimental problems, such as temperature control, are not included.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard13 pagesEnglish language
- Standard13 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Flash X-ray facilities provide intense bremsstrahlung radiation environments, usually in a single sub-microsecond pulse, which often fluctuates in amplitude, shape, and spectrum from shot to shot. Therefore, appropriate dosimetry must be fielded on every exposure to characterize the environment, see ICRU Report 34. These intense bremsstrahlung sources have a variety of applications which include the following:
4.1.1 Studies of the effects of X-rays and gamma rays on materials.
4.1.2 Studies of the effects of radiation on electronic devices such as transistors, diodes, and capacitors.
4.1.3 Computer code validation studies.
4.2 This guide is written to assist the experimenter in selecting the needed dosimetry systems for use at pulsed X-ray facilities. This guide also provides a brief summary of the information on how to use each of the dosimetry systems. Other guides (see Section 2) provide more detailed information on selected dosimetry systems in radiation environments and should be consulted after an initial decision is made on the appropriate dosimetry system to use. There are many key parameters which describe a flash X-ray source, such as dose, dose rate, spectrum, pulse width, etc., such that typically no single dosimetry system can measure all the parameters simultaneously. However, it is frequently the case that not all key parameters must be measured in a given experiment.
SCOPE
1.1 This guide provides assistance in selecting and using dosimetry systems in flash X-ray experiments. Both dose and dose-rate techniques are described.
1.2 Operating characteristics of flash X-ray sources are given, with emphasis on the spectrum of the photon output.
1.3 Assistance is provided to relate the measured dose to the response of a device under test (DUT). The device is assumed to be a semiconductor electronic part or system.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide19 pagesEnglish language
- Guide19 pagesEnglish language
- Guide19 pagesEnglish language
SIGNIFICANCE AND USE
4.1 A radiation-hardness assurance program requires a methodology for relating radiation induced changes in materials exposed to a variety of particle species over a wide range of energies, including those encountered in spacecraft and in terrestrial environments as well as those produced by particle accelerators and nuclear fission and fusion reactors.
4.2 A major source of radiation damage in electronic and photonic devices and materials is the displacement of atoms from their normal lattice site. An appropriate exposure parameter for such damage is the damage energy calculated from NIEL by means of Eq 2. Other analogous measures, which may be used to characterize the irradiation history that is relevant to displacement damage, are damage energy per atom or per unit mass (displacement kerma, when the primary particles are neutral), and displacements per atom (dpa). See Terminology E170 for definitions of those quantities.
4.3 Each of the quantities mentioned in the previous paragraph should convey similar information, but in a different format. In each case the value of the derived exposure parameter depends on approximate nuclear, atomic, and lattice models, an on measured or calculated cross sections. If consistent comparisons are to be made of irradiation effects caused by different particle species and energies, it is essential that these approximations be consistently applied.
4.4 No correspondence should be assumed to exist between damage energy as calculated from NIEL and a particular change in a material property or device parameter. Instead, the damage energy should be used as a parameter which describes the exposure. It may be a useful correlation variate, even when different particle species and energies are included. NIEL should not be reported as a measure of damage, however, unless its correlation with a particular damage modality has been demonstrated in that material or device.
4.5 NIEL is a construct that depends on a model of the part...
SCOPE
1.1 This practice describes a procedure for characterizing particle irradiations of materials in terms of non-ionizing energy loss (NIEL). NIEL is used in published literature to characterize both charged and neutral particle irradiations.
1.2 Although the methods described in this practice apply to any particles and target materials for which displacement cross sections are known (see Practice E521), this practice is intended for use in irradiations in which observed damage effects may be correlated with atomic displacements. This is true of some, but not all, radiation effects in electronic and photonic materials.
1.3 Procedures analogous to this one are used for calculation of displacements per atom (dpa) in charged particle irradiations (see Practice E521) or neutron irradiations (see Practice E693).
1.4 Guidance on calculation of dpa from NIEL is provided.
1.5 Procedures related to this one are used for calculation of 1-MeV equivalent neutron fluence in electronic materials (see Practice E722), but in that practice the concept of damage efficiency, based on correlation of observed damage effects, is included.
1.6 Guidance on conversion of NIEL in silicon to monoenergetic neutron fluence in silicon (see Practice E722), and vice versa, is provided.
1.7 The application of this standard requires knowledge of the particle fluence and energy distribution of particles whose interaction leads to displacement damage.
1.8 The correlation of radiation effects data is beyond the scope of this standard. A comprehensive review (1)2 of displacement damage effects in silicon and their correlation with NIEL provides appropriate guidance that is applicable to semiconductor materials and electronic devices.
- Standard5 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Electronic circuits used in many space, military, and nuclear power systems may be exposed to various levels and time profiles of neutron radiation. It is essential for the design and fabrication of such circuits that test methods be available that can determine the vulnerability or hardness (measure of survivability) of components to be used in them. A determination of hardness is often necessary for the short term (≈100 μs) as well as long term (permanent damage) following exposure. See Practice E722.
SCOPE
1.1 This guide defines the requirements and procedures for testing silicon discrete semiconductor devices and integrated circuits for rapid-annealing effects from displacement damage resulting from neutron radiation. This test will produce degradation of the electrical properties of the irradiated devices and should be considered a destructive test. Rapid annealing of displacement damage is usually associated with bipolar technologies.
1.1.1 Heavy ion beams can also be used to characterize displacement damage annealing (1)2, but ion beams have significant complications in the interpretation of the resulting device behavior due to the associated ionizing dose. The use of pulsed ion beams as a source of displacement damage is not within the scope of this standard.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to consult and establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide7 pagesEnglish language
- Guide7 pagesEnglish language
- Guide7 pagesEnglish language
SIGNIFICANCE AND USE
4.1 It is important to know the energy spectrum of the particular neutron source employed in radiation-hardness testing of electronic devices in order to relate radiation effects with device performance degradation.
4.2 This guide describes the factors which must be considered when the spectrum adjustment methodology is chosen and implemented. Although the selection of sensors (foils) and the determination of responses (activities) is discussed in Guide E720, the experiment should not be divorced from the analysis. In fact, it is advantageous for the analyst conducting the spectrum determination to be closely involved with the design of the experiment to ensure that the data obtained will provide the most accurate spectrum possible. These data include the following : (1) measured responses such as the activities of foils exposed in the environment and their uncertainties, (2) response functions such as reaction cross sections along with appropriate correlations and uncertainties, (3) the geometry and materials in the test environment, and (4) a trial function or prior spectrum and its uncertainties obtained from a transport calculation or from previous experience.
SCOPE
1.1 This guide covers procedures for determining the energy-differential fluence spectra of neutrons used in radiation-hardness testing of electronic semiconductor devices. The types of neutron sources specifically covered by this guide are fission or degraded energy fission sources used in either a steady-state or pulse mode.
1.2 This guide provides guidance and criteria that can be applied during the process of choosing the spectrum adjustment methodology that is best suited to the available data and relevant for the environment being investigated.
1.3 This guide is to be used in conjunction with Guide E720 to characterize neutron spectra and is used in conjunction with Practice E722 to characterize damage-related parameters normally associated with radiation-hardness testing of electronic-semiconductor devices.
Note 1: Although Guide E720 only discusses activation foil sensors, any energy-dependent neutron-responding sensor for which a response function is known may be used (1).2
Note 2: For terminology used in this guide, see Terminology E170.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide12 pagesEnglish language
- Guide12 pagesEnglish language
SIGNIFICANCE AND USE
3.1 Because of the wide variety of materials being used in neutron-activation measurements, this guide is presented with the objective of bringing improved uniformity to the specific field of interest here: hardness testing of electronics primarily in critical assembly reactor environments.
Note 2: Some of the techniques discussed are useful for 14-MeV dosimetry. See Test Method E496 for activation detector materials suitable for 14-MeV neutron effects testing.
Note 3: The materials recommended in this guide are suitable for 252Cf or other weak source effects testing provided the fluence is sufficient to generate countable activities.
3.2 This guide is organized into two overlapping subjects; the criteria used for sensor selection, and the procedures used to ensure the proper determination of activities for determination of neutron spectra. See Terminology E170 and Test Methods E181. Determination of neutron spectra with activation sensor data is discussed in Guides E721 and E944.
SCOPE
1.1 This guide covers the selection and use of neutron-activation detector materials to be employed in neutron spectra adjustment techniques used for radiation-hardness testing of electronic semiconductor devices. Sensors are described that have been used at many radiation hardness-testing facilities, and comments are offered in table footnotes concerning the appropriateness of each reaction as judged by its cross-section accuracy, ease of use as a sensor, and by past successful application. This guide also discusses the fluence-uniformity, neutron self-shielding, and fluence-depression corrections that need to be considered in choosing the sensor thickness, the sensor covers, and the sensor locations. These considerations are relevant for the determination of neutron spectra from assemblies such as TRIGA- and Godiva-type reactors and from Californium irradiators. This guide may also be applicable to other broad energy distribution sources up to 20 MeV.
Note 1: For definitions on terminology used in this guide, see Terminology E170.
1.2 This guide also covers the measurement of the gamma-ray or beta-ray emission rates from the activation foils and other sensors as well as the calculation of the absolute specific activities of these foils. The principal measurement technique is high-resolution gamma-ray spectrometry. The activities are used in the determination of the energy-fluence spectrum of the neutron source. See Guide E721.
1.3 Details of measurement and analysis are covered as follows:
1.3.1 Corrections involved in measuring the sensor activities include those for finite sensor size and thickness in the calibration of the gamma-ray detector, for pulse-height analyzer deadtime and pulse-pileup losses, and for background radioactivity.
1.3.2 The primary method for detector calibration that uses secondary standard gamma-ray emitting sources is considered in this guide and in Test Methods E181. In addition, an alternative method in which the sensors are activated in the known spectrum of a benchmark neutron field is discussed in Guide E1018.
1.3.3 A data analysis method is presented which accounts for the following: detector efficiency; background subtraction; irradiation, waiting, and counting times; fission yields and gamma-ray branching ratios; and self-absorption of gamma rays and neutrons in the sensors.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide12 pagesEnglish language
- Guide12 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Electronic devices are typically tested for survivability to gamma radiation in pure gamma-ray fields. Testing electronic device response against neutrons is more complex since there is invariably a gamma-ray component in addition to the neutron field. The gamma-ray response of the electronic device is typically subtracted from the overall response to find the device response to neutrons. This approach to the testing requires a determination of the gamma-ray exposure in the mixed field. To enhance the neutron effects, the radiation field is sometimes selected to have as large a neutron component as possible.
4.2 CaF2(Mn) TLDs are often used to monitor the gamma-ray dose in mixed neutron/gamma radiation fields. Since the dosimeters are exposed along with the device under test to the mixed field, their response must be corrected for neutrons. In a field rich in neutrons, the uncertainty in the interpretation of the TLD response grows. In fields with relatively few neutrons, the total TLD response may be used to make a correction for gamma response of the device under test. Under this condition, the relative uncertainty in the TLD neutron response is not likely to drive the overall uncertainty in the correction to the electronic device response.
4.3 This practice gives a means of estimating the response of CaF2(Mn) TLDs to neutrons. This neutron response is then subtracted from the measured response to determine the TLD response due to gamma rays. The procedure has relatively high uncertainty because the neutron response of CaF2(Mn) TLDs may vary depending on the source of the material, and this procedure is a generic calculation applicable to CaF2(Mn) TLDs independent of their manufacturer/source. The neutron response given in this practice is a summary of CaF2(Mn) TLD responses reported in the literature. The associated uncertainty envelops the range of results reported, and includes the variety of CaF2(Mn) TLDs used as well as the uncertainties in the dete...
SCOPE
1.1 This practice describes a procedure for correcting a CaF2(Mn) thermoluminescence dosimeter (TLD) reading for its response to neutrons during the irradiation. The neutron response may be subtracted from the total TLD response to give the gamma-ray response. In fields with a large neutron contribution to the total response, this procedure may result in large uncertainties.
1.2 More precise experimental techniques may be applied if the uncertainty derived from this practice is larger than the level that the user can accept. These more precise techniques are not discussed here. The references in Section 8 describe some of these techniques.
1.3 This practice does not discuss effects on the TLD reading from neutron interactions with the material surrounding the TLD and used to ensure a charged particle equilibrium. These effects will depend on the isotopic composition of the surrounding material and its thickness, and on the incident neutron spectrum (1).2
1.4 The values stated in SI units are to be regarded as standard.
- Standard6 pagesEnglish language
- Standard6 pagesEnglish language
SIGNIFICANCE AND USE
5.1 An accurate measure of the total absorbed dose is necessary to ensure the validity of the data taken, to enable comparison to be made of data taken at different facilities, and to verify that components or circuits are tested to the radiation specification applied to the system for which they are to be used.
5.2 The primary value of a calorimetric method for measuring dose is that the results are absolute. They are based only on physical properties of materials, that is, the specific heat of the calorimeter-block material and the Seebeck EMF of the thermocouple used or the temperature coefficient of resistance (α) of the thermistor used, all of which can be established with non-radiation measurements.
5.3 The method permits repeated measurements to be made without requiring entry into the radiation cell between measurements.
SCOPE
1.1 This test method covers a calorimetric measurement of the total absorbed dose delivered in a single pulse of electrons from an electron linear accelerator or a flash X-ray machine (FXR, e-beam mode) used as an ionizing source in radiation-effects testing. The test method is designed for use with pulses of electrons in the energy range from 10 to 50 MeV and is only valid for cases in which both the calorimeter and the test specimen to be irradiated are “thin” compared to the range of these electrons in the materials of which they are constructed.
1.2 The procedure described can be used in those cases in which (1) the dose delivered in a single pulse is 5 Gy(matl)2 [500 rd (matl)] or greater, or (2) multiple pulses of a lower dose can be delivered in a short time compared to the thermal time constant of the calorimeter. The units for the total absorbed dose delivered to a material require the specification of the material and the notation “matl” refers to the active material of the calorimeter. The minimum dose per pulse that can be acceptably monitored depends on the variables of the particular test, including pulse rate, pulse uniformity, and the thermal time constant of the calorimeter.
1.3 A determination of the total dose is made directly for the material of which the calorimeter block is made. The total dose in other materials can be calculated from this measured value by formulas presented in this test method. The need for such calculations and the choice of materials for which calculations are to be made shall be subject to agreement by the parties to the test.
1.4 The values stated in SI units are to be regarded as the standard. The values in parenthesis are provided for information only.
1.5 This standard does not purport to address the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard12 pagesEnglish language
- Standard12 pagesEnglish language
SIGNIFICANCE AND USE
5.1 The neutron test spectrum must be known in order to use a measured device response to predict the device performance in an operational environment (E1854). Typically, neutron spectra are determined by use of a set of sensors with response functions sensitive over the neutron energy region to which the device under test (DUT) responds (E721). For silicon bipolar devices exposed in reactor neutron spectra, this effective energy range is between 0.01 and 10 MeV. A typical set of activation reactions that lack fission reactions from nuclides such as 235U, 237Np, or 239Pu, will have very poor sensitivity to the spectrum between 0.01 and 2 MeV. For a pool-type reactor spectrum, 70 % of the DUT electronic damage response may lie in this range.
5.2 When dosimeters with a significant response in the 10 keV to 2 MeV energy region, such as fission foils, are unavailable, silicon transistors may provide a dosimeter with the needed response to define the spectrum in this critical energy range. When fission foils are part of the sensor set, the silicon sensor provides confirmation of the spectral shape in this energy region.
5.3 Silicon bipolar transistors, such as type 2N2222A, are inexpensive, smaller than fission foils contained in a boron ball, and sensitive to a part of the neutron spectrum important to the damage of modern silicon electronics. They also can be used directly in arrays to map 1 Mev(Si) equivalent displacement damage fluence. The proper set of steps to take in reading the transistor-gain degradation is described in this test method.
5.4 The energy-dependence of the displacement damage function for silicon is found in E722. The major portion of the response for the silicon transistors will generally be above 100 keV.
SCOPE
1.1 This test method covers the use of 2N2222A silicon bipolar transistors as dosimetry sensors in the determination of neutron energy spectra and as 1 Mev(Si) equivalent displacement damage fluence monitors.
1.2 The neutron displacement in silicon can serve as a neutron spectrum sensor in the range 0.1 to 2.0 MeV when fission foils are not available. It has been applied in the fluence range between 2 × 1012 n/cm 2 to 1 × 1014 n/cm2 and should be useful up to 1 × 1015 n/cm2. This test method details the acquisition and use of 1 Mev(Si) equivalent fluence information for the partial determination of the neutron spectra by using 2N2222A transistors.
1.3 This sensor yields a direct measurement of the silicon 1 Mev equivalent fluence by the transfer technique.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.
- Standard10 pagesEnglish language
- Standard10 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Although Co-60 nuclei only emit monoenergetic gamma rays at 1.17 and 1.33 MeV, the finite thickness of sources, and encapsulation materials and other surrounding structures that are inevitably present in irradiators can contribute a substantial amount of low-energy gamma radiation, principally by Compton scattering (1, 2).3 In radiation-hardness testing of electronic devices this low-energy photon component of the gamma spectrum can introduce significant dosimetry errors for a device under test since the equilibrium absorbed dose as measured by a dosimeter can be quite different from the absorbed dose deposited in the device under test because of absorbed dose enhancement effects (3, 4). Absorbed dose enhancement effects refer to the deviations from equilibrium absorbed dose caused by non-equilibrium electron transport near boundaries between dissimilar materials.
4.2 The ionization chamber technique described in this method provides an easy means for estimating the importance of the low-energy photon component of any given irradiator type and configuration.
4.3 When there is an appreciable low-energy spectral component present in a particular irradiator configuration, special experimental techniques should be used to ensure that dosimetry measurements adequately represent the absorbed dose in the device under test. (See Practice E1249.)
SCOPE
1.1 Low energy components in the photon energy spectrum of Co-60 irradiators lead to absorbed dose enhancement effects in the radiation-hardness testing of silicon electronic devices. These low energy components may lead to errors in determining the absorbed dose in a specific device under test. This method covers procedures for the use of a specialized ionization chamber to determine a figure of merit for the relative importance of such effects. It also gives the design and instructions for assembling this chamber.
1.2 This method is applicable to measurements in Co-60 radiation fields where the range of exposure rates is 7 × 10 −6 to 3 × 10−2 C kg −1 s−1 (approximately 100 R/h to 100 R/s). For guidance in applying this method to radiation fields where the exposure rate is >100 R/s, see Appendix X1.
Note 1: See Terminology E170 for definition of exposure and its units.
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard10 pagesEnglish language
- Standard10 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Refer to Guides E720 and E844 for the selection, irradiation, and quality control of neutron dosimeters.
5.2 Refer to Practice E261 for a general discussion of the determination of fast-neutron fluence and fluence rate with threshold detectors.
5.3 The activation reaction produces 32P, which decays by the emission of a single beta particle in 100 % of the decays, and which emits no gamma rays. The half life of 32P is 14.284 (36)3 days (1) 4 and the maximum beta energy is 1710.66 (21) keV (1).
5.4 Elemental sulfur is readily available in pure form and any trace contaminants present do not produce significant amounts of radioactivity. Natural sulfur, however, is composed of 32S (94.99 % (26)), 34S (4.25 % (24)) (2), and trace amounts of other sulfur isotopes. The presence of these other isotopes leads to several competing reactions that can interfere with the counting of the 1710-keV beta particle. This interference can usually be eliminated by the use of appropriate techniques, as discussed in Section 8.
SCOPE
1.1 This test method describes procedures for measuring reaction rates and fast-neutron fluences by the activation reaction 32S(n,p)32P.
1.2 This activation reaction is useful for measuring neutrons with energies above approximately 3 MeV.
1.3 With suitable techniques, fission-neutron fluences from about 5 × 108 to 1016 n/cm 2 can be measured.
1.4 Detailed procedures for other fast-neutron detectors are described in Practice E261.
1.5 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard6 pagesEnglish language
- Standard6 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Division of the Co-60 Hardness Testing into Five Parts:
4.1.1 The equilibrium absorbed dose shall be measured with a dosimeter, such as a TLD, located adjacent to the device under test. Alternatively, a dosimeter may be irradiated in the position of the device before or after irradiation of the device.
4.1.2 This absorbed dose measured by the dosimeter shall be converted to the equilibrium absorbed dose in the material of interest within the critical region within the device under test, for example the SiO2 gate oxide of an MOS device.
4.1.3 A correction for absorbed-dose enhancement effects shall be considered. This correction is dependent upon the photon energy that strikes the device under test.
4.1.4 A correlation should be made between the absorbed dose in the critical region (for example, the gate oxide mentioned in 4.1.2) and some electrically important effect (such as charge trapped at the Si/SiO2 interface as manifested by a shift in threshold voltage).
4.1.5 An extrapolation should then be made from the results of the test to the results that would be expected for the device under test under actual operating conditions.
Note 5: The parts of a test discussed in 4.1.2 and 4.1.3 are the subject of this practice. The subject of 4.1.1 is covered and referenced in other standards such as Practice E668 and ICRU Report 14. The parts of a test discussed in 4.1.4 and 4.1.5 are outside the scope of this practice.
4.2 Low-Energy Components in the Spectrum—Some of the primary Co-60 gamma rays (1.17 and 1.33 MeV) produce lower energy photons by Compton scattering within the Co-60 source structure, within materials that lie between the source and the device under test, and within materials that lie beyond the device but contribute to backscattering. As a result of the complexity of these effects, the photon energy spectrum striking the device usually is not well known. This point is further discussed in Section 5 and Appendix X1. The presence of ...
SCOPE
1.1 This practice covers recommended procedures for the use of dosimeters, such as thermoluminescent dosimeters (TLD's), to determine the absorbed dose in a region of interest within an electronic device irradiated using a Co-60 source. Co-60 sources are commonly used for the absorbed dose testing of silicon electronic devices.
Note 1: This absorbed-dose testing is sometimes called “total dose testing” to distinguish it from “dose rate testing.”
Note 2: The effects of ionizing radiation on some types of electronic devices may depend on both the absorbed dose and the absorbed dose rate; that is, the effects may be different if the device is irradiated to the same absorbed-dose level at different absorbed-dose rates. Absorbed-dose rate effects are not covered in this practice but should be considered in radiation hardness testing.
1.2 The principal potential error for the measurement of absorbed dose in electronic devices arises from non-equilibrium energy deposition effects in the vicinity of material interfaces.
1.3 Information is given about absorbed-dose enhancement effects in the vicinity of material interfaces. The sensitivity of such effects to low energy components in the Co-60 photon energy spectrum is emphasized.
1.4 A brief description is given of typical Co-60 sources with special emphasis on the presence of low energy components in the photon energy spectrum output from such sources.
1.5 Procedures are given for minimizing the low energy components of the photon energy spectrum from Co-60 sources, using filtration. The use of a filter box to achieve such filtration is recommended.
1.6 Information is given on absorbed-dose enhancement effects that are dependent on the device orientation with respect to the Co-60 source.
1.7 The use of spectrum filtration and appropriate device orientation provides a radiation environment whereby the absorbed dose in the sensitive region of an elect...
- Standard17 pagesEnglish language
- Standard17 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This practice is important in characterizing the radiation hardness of electronic devices irradiated by neutrons. This characterization makes it feasible to predict some changes in operational properties of irradiated semiconductor devices or electronic systems. To facilitate uniformity of the interpretation and evaluation of results of irradiations by sources of different fluence spectra, it is convenient to reduce the incident neutron fluence from a source to a single parameter—an equivalent monoenergetic neutron fluence—applicable to a particular semiconductor material.
5.2 In order to determine an equivalent monoenergetic neutron fluence, it is necessary to evaluate the displacement damage of the particular semiconductor material. Ideally, this quantity is correlated to the degradation of a specific functional performance parameter (such as current gain) of the semiconductor device or system being tested. However, this correlation has not been established unequivocally for all device types and performance parameters since, in many instances, other effects also can be important. Ionization effects produced by the incident neutron fluence or by gamma rays in a mixed neutron fluence, short-term and long-term annealing, and other factors can contribute to observed performance degradation (damage). Thus, caution should be exercised in making a correlation between calculated displacement damage and performance degradation of a given electronic device. The types of devices for which this correlation is applicable, and numerical evaluation of displacement damage are discussed in the annexes.
5.3 The concept of 1-MeV equivalent fluence is widely used in the radiation-hardness testing community. It has merits and disadvantages that have been debated widely (9-12). For these reasons, specifics of a standard application of the 1-MeV equivalent fluence are presented in the annexes.
SCOPE
1.1 This practice covers procedures for characterizing neutron fluence from a source in terms of an equivalent monoenergetic neutron fluence. It is applicable to neutron effects testing, to the development of test specifications, and to the characterization of neutron test environments. The sources may have a broad neutron-energy range, or may be mono-energetic neutron sources with energies up to 20 MeV. This practice is not applicable in cases where the predominant source of displacement damage is from neutrons of energy less than 10 keV. The relevant equivalence is in terms of a specified effect on certain physical properties of materials upon which the source spectrum is incident. In order to achieve this, knowledge of the effects of neutrons as a function of energy on the specific property of the material of interest is required. Sharp variations in the effects with neutron energy may limit the usefulness of this practice in the case of mono-energetic sources.
1.2 This practice is presented in a manner to be of general application to a variety of materials and sources. Correlation between displacements (1-3)2 caused by different particles (electrons, neutrons, protons, and heavy ions) is beyond the scope of this practice. In radiation-hardness testing of electronic semiconductor devices, specific materials of interest include silicon and gallium arsenide, and the neutron sources generally are test and research reactors and californium-252 irradiators.
1.3 The technique involved relies on the following factors: (1) a detailed determination of the fluence spectrum of the neutron source, and (2) a knowledge of the degradation (damage) effects of neutrons as a function of energy on specific material properties.
1.4 The detailed determination of the neutron fluence spectrum referred to in 1.3 need not be performed afresh for each test exposure, provided the exposure conditions are repeatable. When the spectrum determination is not repeated, a neutron fluence monitor shall be used for each test expo...
- Standard27 pagesEnglish language
- Standard27 pagesEnglish language
SIGNIFICANCE AND USE
4.1 The absorbed dose is a more meaningful parameter than exposure for use in relating the effects of radiation on materials. It expresses the energy absorbed by the irradiated material per unit mass, whereas exposure is related to the amount of charge produced in air per unit mass. Absorbed dose, as referred to here, implies that the measurement is made under conditions of charged particle (electron) equilibrium (see Appendix X1). In practice, such conditions are not rigorously achievable but, under some circumstances, can be approximated closely.
4.2 Different materials, when exposed to the same radiation field, absorb different amounts of energy. Using the techniques of this standard, charged particle equilibrium must exist in order to relate the absorbed dose in one material to the absorbed dose in another. Also, if the radiation is attenuated by a significant thickness of an absorber, the energy spectrum of the radiation will be changed, and it will be necessary to correct for this.Note 1—For comprehensive discussions of various dosimetry methods applicable to the radiation types and energies and absorbed dose rate ranges discussed in this method, see ICRU Reports 34 and 80.
SCOPE
1.1 This practice presents a technique for calculating the absorbed dose in a material from knowledge of the radiation field, the composition of the material, (1-5)2,3 and a related measurement. The procedure is applicable for X and gamma radiation provided the energy of the photons fall within the range from 0.01 to 20 MeV.
1.2 A method is given for calculating the absorbed dose in a material from the knowledge of the absorbed dose in another material exposed to the same radiation field. The procedure is restricted to homogeneous materials composed of the elements for which absorption coefficients have been tabulated. All 92 natural elements are tabulated in (2). It also requires some knowledge of the energy spectrum of the radiation field produced by the source under consideration. Generally, the accuracy of this method is limited by the accuracy to which the energy spectrum of the radiation field is known.
1.3 The results of this practice are only valid if charged particle equilibrium exists in the material and at the depth of interest. Thus, this practice is not applicable for determining absorbed dose in the immediate vicinity of boundaries between materials of widely differing atomic numbers. For more information on this topic, see Practice E1249.
1.4 Energy transport computer codes4 exist that are formulated to calculate absorbed dose in materials more precisely than this method. To use these codes, more effort, time, and expense are required. If the situation warrants, such calculations should be used rather than the method described here.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard9 pagesEnglish language
- Standard9 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Refer to Practice E261 for a general discussion of the measurement of fast-neutron fluence rates with threshold detectors.
FIG. 1 Variation of 0 Degree 3H(d,n)4He Differential Cross Section with Incident Deuteron Energy (1)
FIG. 2 Variation of 0 Degree 3H(d,n)4He Differential Cross Section with Incident Deuteron Energy (1)
5.2 Refer to Test Method E265 for a general discussion of the measurement of fast-neutron fluence rates by radioactivation of sulfur-32.
5.3 Reactions used for the activity measurements can be chosen to provide a convenient means for determining the absolute fluence rates of 14-MeV neutrons obtained with 3H(d,n)4He neutron generators over a range of irradiation times from seconds to approximately 100 days. High purity threshold sensors referenced in this test method are readily available.
5.4 The neutron-energy spectrum must be known in order to measure fast-neutron fluence using a single threshold detector. Neutrons produced by bombarding a tritiated target with deuterons are commonly referred to as 14-MeV neutrons; however, they can have a range of energies depending on: (1) the angle of neutron emission with respect to the deuteron beam, (2) the kinetic energy of the deuterons, and (3) the target thickness. In most available neutron generators of the Cockroft-Walton type, a thick target is used to obtain high-neutron yields. As deuterons penetrate through the surface and move into the bulk of the thick target, they lose energy, and interactions occurring deeper within the target produce neutrons with correspondingly lower energy.
5.5 Wide variations in neutron energy are not generally encountered in commercially available neutron generators of the Cockroft-Walton type. Figs. 1 and 2 (1)6 show the variation of the zero degree 3H(d,n)4He neutron production cross section with energy, and clearly indicate that maximum neutron yield is obtained with deuterons having energies near the 107 keV resonance. Since most generators are d...
SCOPE
1.1 This test method covers a general procedure for the measurement of the fast-neutron fluence rate produced by neutron generators utilizing the 3H(d,n) 4He reaction. Neutrons so produced are usually referred to as 14-MeV neutrons, but range in energy depending on a number of factors. This test method does not adequately cover fusion sources where the velocity of the plasma may be an important consideration.
1.2 This test method uses threshold activation reactions to determine the average energy of the neutrons and the neutron fluence at that energy. At least three activities, chosen from an appropriate set of dosimetry reactions, are required to characterize the average energy and fluence. The required activities are typically measured by gamma ray spectroscopy.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard14 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Refer to Practice E261 for a general discussion of the measurement of fast-neutron fluence rates with threshold detectors.
5.2 Refer to Test Method E265 for a general discussion of the measurement of fast-neutron fluence rates by radioactivation of sulfur-32.
5.3 Reactions used for the activity measurements can be chosen to provide a convenient means for determining the absolute fluence rates of 14-MeV neutrons obtained with 3H(d,n)4He neutron generators over a range of irradiation times from seconds to approximately 100 days. High purity threshold sensors referenced in this test method are readily available.
5.4 The neutron-energy spectrum must be known in order to measure fast-neutron fluence using a single threshold detector. Neutrons produced by bombarding a tritiated target with deuterons are commonly referred to as 14-MeV neutrons; however, they can have a range of energies depending on: (1) the angle of neutron emission with respect to the deuteron beam, (2) the kinetic energy of the deuterons, and (3) the target thickness. In most available neutron generators of the Cockroft-Walton type, a thick target is used to obtain high-neutron yields. As deuterons penetrate through the surface and move into the bulk of the thick target, they lose energy, and interactions occurring deeper within the target produce neutrons with correspondingly lower energy.
5.5 Wide variations in neutron energy are not generally encountered in commercially available neutron generators of the Cockroft-Walton type. Figs. 1 and 2 (1)6 show the variation of the zero degree 3H(d,n)4He neutron production cross section with energy, and clearly indicate that maximum neutron yield is obtained with deuterons having energies near the 107 keV resonance. Since most generators are designed for high yield, the deuteron energy is typically about 200 keV, giving a range of neutron energies from approximately 14 to 15 MeV. The differential center-of-mass cross section is typically paramet...
SCOPE
1.1 This test method covers a general procedure for the measurement of the fast-neutron fluence rate produced by neutron generators utilizing the 3H(d,n) 4He reaction. Neutrons so produced are usually referred to as 14-MeV neutrons, but range in energy depending on a number of factors. This test method does not adequately cover fusion sources where the velocity of the plasma may be an important consideration.
1.2 This test method uses threshold activation reactions to determine the average energy of the neutrons and the neutron fluence at that energy. At least three activities, chosen from an appropriate set of dosimetry reactions, are required to characterize the average energy and fluence. The required activities are typically measured by gamma ray spectroscopy.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard13 pagesEnglish language
- Standard13 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Flash X-ray facilities provide intense bremsstrahlung radiation environments, usually in a single sub-microsecond pulse, which often fluctuates in amplitude, shape, and spectrum from shot to shot. Therefore, appropriate dosimetry must be fielded on every exposure to characterize the environment, see ICRU Report 34. These intense bremsstrahlung sources have a variety of applications which include the following:
4.1.1 Generation of X-ray and gamma-ray environments similar to that from a nuclear weapon burst.
4.1.2 Studies of the effects of X-rays and gamma rays on materials.
4.1.3 Studies of the effects of radiation on electronic devices such as transistors, diodes, and capacitors.
4.1.4 Vulnerability and survivability testing of military systems and components.
4.1.5 Computer code validation studies.
4.2 This guide is written to assist the experimenter in selecting the needed dosimetry systems (not all radiation parameters must be measured in a given experiment) for use at pulsed X-ray facilities. This guide also provides a brief summary of the information on how to use each of the dosimetry systems. Other guides (see Section 2) provide more detailed information on selected dosimetry systems in radiation environments and should be consulted after an initial decision is made on the appropriate dosimetry system to use. There are many key parameters which describe a flash X-ray source, such as dose, dose rate, spectrum, pulse width, etc., such that typically no single dosimetry system can measure all the parameters simultaneously.
SCOPE
1.1 This guide provides assistance in selecting and using dosimetry systems in flash X-ray experiments. Both dose and dose-rate techniques are described.
1.2 Operating characteristics of flash X-ray sources are given, with emphasis on the spectrum of the photon output.
1.3 Assistance is provided to relate the measured dose to the response of a device under test (DUT). The device is assumed to be a semiconductor electronic part or system.
- Guide19 pagesEnglish language
- Guide19 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Flash X-ray facilities provide intense bremsstrahlung radiation environments, usually in a single sub-microsecond pulse, which often fluctuates in amplitude, shape, and spectrum from shot to shot. Therefore, appropriate dosimetry must be fielded on every exposure to characterize the environment, see ICRU Report 34. These intense bremsstrahlung sources have a variety of applications which include the following:
4.1.1 Generation of X-ray and gamma-ray environments similar to that from a nuclear weapon burst.
4.1.2 Studies of the effects of X-rays and gamma rays on materials.
4.1.3 Studies of the effects of radiation on electronic devices such as transistors, diodes, and capacitors.
4.1.4 Vulnerability and survivability testing of military systems and components.
4.1.5 Computer code validation studies.
4.2 This guide is written to assist the experimenter in selecting the needed dosimetry systems (not all radiation parameters must be measured in a given experiment) for use at pulsed X-ray facilities. This guide also provides a brief summary of the information on how to use each of the dosimetry systems. Other guides (see Section 2) provide more detailed information on selected dosimetry systems in radiation environments and should be consulted after an initial decision is made on the appropriate dosimetry system to use. There are many key parameters which describe a flash X-ray source, such as dose, dose rate, spectrum, pulse width, etc., such that typically no single dosimetry system can measure all the parameters simultaneously.
SCOPE
1.1 This guide provides assistance in selecting and using dosimetry systems in flash X-ray experiments. Both dose and dose-rate techniques are described.
1.2 Operating characteristics of flash X-ray sources are given, with emphasis on the spectrum of the photon output.
1.3 Assistance is provided to relate the measured dose to the response of a device under test (DUT). The device is assumed to be a semiconductor electronic part or system.
- Guide19 pagesEnglish language
- Guide19 pagesEnglish language
SIGNIFICANCE AND USE
4.1 This practice was written primarily to guide test participants in establishing, identifying, maintaining, and using suitable environments for conducting high quality neutron tests. Its development was motivated, in large measure, because inadequate controls in the neutron-effects-test process have in some past instances resulted in exposures that have differed by factors of three or more from irradiation specifications. A radiation test environment generally differs from the environment in which the electronics must operate (the operational environment); therefore, a high quality test requires not only the use of a suitable radiation environment, but also control and compensation for contributions to damage that differ from those in the operational environment. In general, the responsibility for identifying suitable test environments to accomplish test objectives lies with the sponsor/user/tester and test specialist part of the team, with the assistance of an independent validator, if available. The responsibility for the establishment and maintenance of suitable environments lies with the facility operator/dosimetrist and test specialist, again with the possible assistance of an independent validator. Additional guidance on the selection of an irradiation facility is provided in Practice F1190.
4.2 This practice identifies the tasks that must be accomplished to ensure a successful high quality test. It is the overall responsibility of the sponsor or user to ensure that all of the required tasks are complete and conditions are met. Other participants provide appropriate documentation to enable the sponsor or user to make that determination.
4.3 The principal determinants of a properly conducted test are: (1) the radiation test environment shall be well characterized, controlled, and correlated with the specified irradiation levels; (2) damage produced in the electronic materials and devices is caused by the desired, specified component of the environment a...
SCOPE
1.1 This practice sets forth requirements to ensure consistency in neutron-induced displacement damage testing of silicon and gallium arsenide electronic piece parts. This requires controls on facility, dosimetry, tester, and communications processes that affect the accuracy and reproducibility of these tests. It provides background information on the technical basis for the requirements and additional recommendations on neutron testing.
1.2 Methods are presented for ensuring and validating consistency in neutron displacement damage testing of electronic parts such as integrated circuits, transistors, and diodes. The issues identified and the controls set forth in this practice address the characterization and suitability of the radiation environments. They generally apply to reactor sources, accelerator-based neutron sources, such as 14-MeV DT sources, and 252Cf sources. Facility and environment characteristics that introduce complications or problems are identified, and recommendations are offered to recognize, minimize or eliminate these problems. This practice may be used by facility users, test personnel, facility operators, and independent process validators to determine the suitability of a specific environment within a facility and of the testing process as a whole. Electrical measurements are addressed in other standards, such as Guide F980. Additional information on conducting irradiations can be found in Practices E798 and F1190. This practice also may be of use to test sponsors (organizations that establish test specifications or otherwise have a vested interest in the performance of electronics in neutron environments).
1.3 Methods for the evaluation and control of undesired contributions to damage are discussed in this practice. References to relevant ASTM standards and technical reports are provided. Processes and methods used to arrive at the appropriate test environments and specification levels f...
- Standard13 pagesEnglish language
- Standard13 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Refer to Guides E720 and E844 for the selection, irradiation, and quality control of neutron dosimeters.
5.2 Refer to Practice E261 for a general discussion of the determination of fast-neutron fluence and fluence rate with threshold detectors.
5.3 The activation reaction produces 32P, which decays by the emission of a single beta particle in 100 % of the decays, and which emits no gamma rays. The half life of 32P is 14.262 (14)3 days (1) 4 and the maximum beta energy is 1710 keV (2).
5.4 Elemental sulfur is readily available in pure form and any trace contaminants present do not produce significant amounts of radioactivity. Natural sulfur, however, is composed of 32S (95.02 % (9)), 34S (4.21 % (8)) (1), and trace amounts of other sulfur isotopes. The presence of these other isotopes leads to several competing reactions that can interfere with the counting of the 1710-keV beta particle. This interference can usually be eliminated by the use of appropriate techniques, as discussed in Section 8.
SCOPE
1.1 This test method describes procedures for measuring reaction rates and fast-neutron fluences by the activation reaction 32S(n,p)32P.
1.2 This activation reaction is useful for measuring neutrons with energies above approximately 3 MeV.
1.3 With suitable techniques, fission-neutron fluences from about 5 × 108 to 1016 n/cm 2 can be measured.
1.4 Detailed procedures for other fast-neutron detectors are described in Practice E261.
1.5 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard6 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Absorbed dose in a material is an important parameter that can be correlated with radiation effects produced in electronic components and devices that are exposed to ionizing radiation. Reasonable estimates of this parameter can be calculated if knowledge of the source radiation field (that is, energy spectrum and particle fluence) is available. Sufficiently detailed information about the radiation field is generally not available. However, measurements of absorbed dose with passive dosimeters in a radiation test facility can provide information from which the absorbed dose in a material of interest can be inferred. Under certain prescribed conditions, TLDs are quite suitable for performing such measurements.Note 2—For comprehensive discussions of various dosimetry methods applicable to the radiation types and energy and absorbed dose-rate range discussed in this practice, see ICRU Reports 14, 17, 21, and 34.
SCOPE
1.1 This practice covers procedures for the use of thermoluminescence dosimeters (TLDs) to determine the absorbed dose in a material irradiated by ionizing radiation. Although some elements of the procedures have broader application, the specific area of concern is radiation-hardness testing of electronic devices. This practice is applicable to the measurement of absorbed dose in materials irradiated by gamma rays, X rays, and electrons of energies from 12 to 60 MeV. Specific energy limits are covered in appropriate sections describing specific applications of the procedures. The range of absorbed dose covered is approximately from 10−2 to 104 Gy (1 to 106 rad), and the range of absorbed dose rates is approximately from 10−2 to 1010 Gy/s (1 to 1012 rad/s). Absorbed dose and absorbed dose-rate measurements in materials subjected to neutron irradiation are not covered in this practice. (See Practice E2450 for guidance in mixed fields.) Further, the portion of these procedures that deal with electron irradiation are primarily intended for use in parts testing. Testing of devices as a part of more massive components such as electronics boards or boxes may require techniques outside the scope of this practice.Note 1—The purpose of the upper and lower limits on the energy for electron irradiation is to approach a limiting case where dosimetry is simplified. Specifically, the dosimetry methodology specified requires that the following three limiting conditions be approached: (a) energy loss of the primary electrons is small, (b) secondary electrons are largely stopped within the dosimeter, and (c) bremsstrahlung radiation generated by the primary electrons is largely lost.
1.2 This standard dose not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard19 pagesEnglish language
- Standard19 pagesEnglish language
SIGNIFICANCE AND USE
It is important to know the energy spectrum of the particular neutron source employed in radiation-hardness testing of electronic devices in order to relate radiation effects with device performance degradation.
This guide describes the factors which must be considered when the spectrum adjustment methodology is chosen and implemented. Although the selection of sensors (foils) and the determination of responses (activities) is discussed in Guide E720, the experiment should not be divorced from the analysis. In fact, it is advantageous for the analyst conducting the spectrum determination to be closely involved with the design of the experiment to ensure that the data obtained will provide the most accurate spectrum possible. These data include the following : (1) measured responses such as the activities of foils exposed in the environment and their uncertainties, (2) response functions such as reaction cross sections along with appropriate correlations and uncertainties, (3) the geometry and materials in the test environment, and (4) a trial function or prior spectrum and its uncertainties obtained from a transport calculation or from previous experience.
SCOPE
1.1 This guide covers procedures for determining the energy-differential fluence spectra of neutrons used in radiation-hardness testing of electronic semiconductor devices. The types of neutron sources specifically covered by this guide are fission or degraded energy fission sources used in either a steady-state or pulse mode.
1.2 This guide provides guidance and criteria that can be applied during the process of choosing the spectrum adjustment methodology that is best suited to the available data and relevant for the environment being investigated.
1.3 This guide is to be used in conjunction with Guide E720 to characterize neutron spectra and is used in conjunction with Practice E722 to characterize damage-related parameters normally associated with radiation-hardness testing of electronic-semiconductor devices.
Note 1—Although Guide E720 only discusses activation foil sensors, any energy-dependent neutron-responding sensor for which a response function is known may be used (1).
Note 2—For terminology used in this guide, see Terminology E170.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide11 pagesEnglish language
- Guide11 pagesEnglish language
SCOPE
1.1 This practice describes a procedure for measuring gamma-ray absorbed dose in CaF2(Mn) thermoluminescence dosimeters (TLDs) exposed to mixed neutron-photon environments during irradiation of materials and devices. The practice has broad application, but is primarily intended for use in the radiation-hardness testing of electronics. The practice is applicable to the measurement of absorbed dose from gamma radiation present in fields used for neutron testing.
1.2 This practice describes a procedure for correcting for the neutron response of a CaF2(Mn) TLD. The neutron response may be subtracted from the total response to give the gamma-ray response. In fields with a large neutron contribution to the total response, this procedure may result in large uncertainties.
1.3 More precise experimental techniques may be applied if the uncertainty derived from this practice is larger than the user can accept. These techniques are not discussed here. The references in Section 8 describe some of these techniques.
1.4 This practice does not discuss effects on the TLD reading of neutron interactions with material surrounding the TLD to ensure charged particle equilibrium. These effects depend on the surrounding material and its thickness, and on the neutron spectrum (1).
- Standard6 pagesEnglish language
- Standard6 pagesEnglish language
SIGNIFICANCE AND USE
Because of the wide variety of materials being used in neutron-activation measurements, this guide is presented with the objective of bringing improved uniformity to the specific field of interest here: hardness testing of electronics primarily in critical assembly reactor environments.
Note 2—Some of the techniques discussed are useful for 14-MeV dosimetry. See Test Method E496 for activation detector materials suitable for 14-MeV neutron effects testing.
Note 3—The materials recommended in this guide are suitable for 252Cf or other weak source effects testing provided the fluence is sufficient to generate countable activities.
This guide is organized into two overlapping subjects; the criteria used for sensor selection, and the procedures used to ensure the proper determination of activities for determination of neutron spectra. See Terminology E170 and General Methods E181. Determination of neutron spectra with activation sensor data is discussed in Guides E721 and E944.
SCOPE
1.1 This guide covers the selection and use of neutron-activation detector materials to be employed in neutron spectra adjustment techniques used for radiation-hardness testing of electronic semiconductor devices. Sensors are described that have been used at many radiation hardness-testing facilities, and comments are offered in table footnotes concerning the appropriateness of each reaction as judged by its cross-section accuracy, ease of use as a sensor, and by past successful application. This guide also discusses the fluence-uniformity, neutron self-shielding, and fluence-depression corrections that need to be considered in choosing the sensor thickness, the sensor covers, and the sensor locations. These considerations are relevant for the determination of neutron spectra from assemblies such as TRIGA- and Godiva-type reactors and from Californium irradiators. This guide may also be applicable to other broad energy distribution sources up to 20 MeV.
Note 1—For definitions on terminology used in this guide, see Terminology E170.
1.2 This guide also covers the measurement of the gamma-ray or beta-ray emission rates from the activation foils and other sensors as well as the calculation of the absolute specific activities of these foils. The principal measurement technique is high-resolution gamma-ray spectrometry. The activities are used in the determination of the energy-fluence spectrum of the neutron source. See Guide E721.
1.3 Details of measurement and analysis are covered as follows:
1.3.1 Corrections involved in measuring the sensor activities include those for finite sensor size and thickness in the calibration of the gamma-ray detector, for pulse-height analyzer deadtime and pulse-pileup losses, and for background radioactivity.
1.3.2 The primary method for detector calibration that uses secondary standard gamma-ray emitting sources is considered in this guide and in General Methods E181. In addition, an alternative method in which the sensors are activated in the known spectrum of a benchmark neutron field is discussed in Guide E1018.
1.3.3 A data analysis method is presented which accounts for the following: detector efficiency; background subtraction; irradiation, waiting, and counting times; fission yields and gamma-ray branching ratios; and self-absorption of gamma rays and neutrons in the sensors.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide12 pagesEnglish language
- Guide12 pagesEnglish language
SIGNIFICANCE AND USE
An accurate measure of the dose is necessary to ensure the validity of the data taken, to enable comparison to be made of data taken at different facilities, and to verify that components or circuits are tested to the radiation specification applied to the system for which they are to be used.
The primary value of a calorimetric method for measuring dose is that the results are absolute. They are based only on physical properties of materials, that is, the specific heat of the calorimeter-block material and the Seebeck EMF of the thermocouple used or the temperature coefficient of resistance (α) of the thermistor used, all of which can be established with non-radiation measurements.
The method permits repeated measurements to be made without requiring entry into the radiation cell between measurements.
SCOPE
1.1 This test method covers a calorimetric measurement of the total dose delivered in a single pulse of electrons from an electron linear accelerator or a flash X-ray machine (FXR, e-beam mode) used as an ionizing source in radiation-effects testing. The test method is designed for use with pulses of electrons in the energy range from 10 to 50 MeV and is only valid for cases in which both the calorimeter and the test specimen to be irradiated are“thin” compared to the range of these electrons in the materials of which they are constructed.
1.2 The procedure described can be used in those cases in which (1) the dose delivered in a single pulse is 5 Gy (matl) (500 rd (matl)) or greater, or (2) multiple pulses of a lower dose can be delivered in a short time compared to the thermal time constant of the calorimeter. Matl refers to the material of the calorimeter. The minimum dose per pulse that can be acceptably monitored depends on the variables of the particular test, including pulse rate, pulse uniformity, and the thermal time constant of the calorimeter.
1.3 A determination of the total dose is made directly for the material of which the calorimeter block is made. The total dose in other materials can be calculated from this measured value by formulas presented in this test method. The need for such calculations and the choice of materials for which calculations are to be made shall be subject to agreement by the parties to the test.
1.4 The values stated in SI units are to be regarded as the standard. The values in parenthesis are provided for information only.
1.5 This standard does not purport to address the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard12 pagesEnglish language
- Standard12 pagesEnglish language
SIGNIFICANCE AND USE
Although Co-60 nuclei only emit monoenergetic gamma rays at 1.17 and 1.33 MeV, the finite thickness of sources, and encapsulation materials and other surrounding structures that are inevitably present in irradiators can contribute a substantial amount of low-energy gamma radiation, principally by Compton scattering (1, 2). In radiation-hardness testing of electronic devices this low-energy photon component of the gamma spectrum can introduce significant dosimetry errors for a device under test since the equilibrium absorbed dose as measured by a dosimeter can be quite different from the absorbed dose deposited in the device under test because of absorbed dose enhancement effects (3, 4). Absorbed dose enhancement effects refer to the deviations from equilibrium absorbed dose caused by non-equilibrium electron transport near boundaries between dissimilar materials.
The ionization chamber technique described in this method provides an easy means for estimating the importance of the low-energy photon component of any given irradiator type and configuration.
When there is an appreciable low-energy spectral component present in a particular irradiator configuration, special experimental techniques should be used to ensure that dosimetry measurements adequately represent the absorbed dose in the device under test. (See Practice E1249.)
SCOPE
1.1 Low energy components in the photon energy spectrum of Co-60 irradiators lead to absorbed dose enhancement effects in the radiation-hardness testing of silicon electronic devices. These low energy components may lead to errors in determining the absorbed dose in a specific device under test. This method covers procedures for the use of a specialized ionization chamber to determine a figure of merit for the relative importance of such effects. It also gives the design and instructions for assembling this chamber.
1.2 This method is applicable to measurements in Co-60 radiation fields where the range of exposure rates is 7 × 10 −6 to 3 × 10−2 C kg −1 s−1 (approximately 100 R/h to 100 R/s). For guidance in applying this method to radiation fields where the exposure rate is >100 R/s, see Appendix X1.
Note 1—See Terminology E170 for definition of exposure and its units.
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard10 pagesEnglish language
- Standard10 pagesEnglish language
SIGNIFICANCE AND USE
Division of the Co-60 Hardness Testing into Five Parts:
The equilibrium absorbed dose shall be measured with a dosimeter, such as a TLD, located adjacent to the device under test. Alternatively, a dosimeter may be irradiated in the position of the device before or after irradiation of the device.
This absorbed dose measured by the dosimeter shall be converted to the equilibrium absorbed dose in the material of interest within the critical region within the device under test, for example the SiO 2 gate oxide of an MOS device.
A correction for absorbed-dose enhancement effects shall be considered. This correction is dependent upon the photon energy that strikes the device under test.
A correlation should be made between the absorbed dose in the critical region (for example, the gate oxide mentioned in 4.1.2) and some electrically important effect (such as charge trapped at the Si/SiO2 interface as manifested by a shift in threshold voltage).
An extrapolation should then be made from the results of the test to the results that would be expected for the device under test under actual operating conditions.
SCOPE
1.1 This practice covers recommended procedures for the use of dosimeters, such as thermoluminescent dosimeters (TLD's), to determine the absorbed dose in a region of interest within an electronic device irradiated using a Co-60 source. Co-60 sources are commonly used for the absorbed dose testing of silicon electronic devices.
Note 1—This absorbed-dose testing is sometimes called “total dose testing” to distinguish it from “dose rate testing.”
Note 2—The effects of ionizing radiation on some types of electronic devices may depend on both the absorbed dose and the absorbed dose rate; that is, the effects may be different if the device is irradiated to the same absorbed-dose level at different absorbed-dose rates. Absorbed-dose rate effects are not covered in this practice but should be considered in radiation hardness testing.
1.2 The principal potential error for the measurement of absorbed dose in electronic devices arises from non-equilibrium energy deposition effects in the vicinity of material interfaces.
1.3 Information is given about absorbed-dose enhancement effects in the vicinity of material interfaces. The sensitivity of such effects to low energy components in the Co-60 photon energy spectrum is emphasized.
1.4 A brief description is given of typical Co-60 sources with special emphasis on the presence of low energy components in the photon energy spectrum output from such sources.
1.5 Procedures are given for minimizing the low energy components of the photon energy spectrum from Co-60 sources, using filtration. The use of a filter box to achieve such filtration is recommended.
1.6 Information is given on absorbed-dose enhancement effects that are dependent on the device orientation with respect to the Co-60 source.
1.7 The use of spectrum filtration and appropriate device orientation provides a radiation environment whereby the absorbed dose in the sensitive region of an electronic device can be calculated within defined error limits without detailed knowledge of either the device structure or of the photon energy spectrum of the source, and hence, without knowing the details of the absorbed-dose enhancement effects.
1.8 The recommendations of this practice are primarily applicable to piece-part testing of electronic devices. Electronic circuit board and electronic system testing may introduce problems that are not adequately treated by the methods recommended here.
1.9 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard17 pagesEnglish language
- Standard17 pagesEnglish language
SIGNIFICANCE AND USE
The neutron spectrum in a test (simulation) environment must be known in order to use a measured device response in the test environment to predict the device performance in an operational environment (see Practice E1854). Typically, neutron spectra are determined by use of a set of sensors that have response functions that are sensitive over the neutron energy region to which the device under test (DUT) responds (see Guide E721). In particular, for silicon bipolar devices exposed in reactor neutron spectra, this effective energy range is between 0.01 and 10 MeV. A typical set of activation reactions that lack fission reactions from nuclides such as 235U, 237Np, or 239Pu, will have very poor sensitivity to the spectrum between 0.01 and 2 MeV. For a pool-type reactor spectrum, 70 % of the DUT electronic damage response may lie in this range. Often, fission foils are not included in the sensor set for spectrum determinations because their use must be licensed, and they require special handling for health physics considerations. The silicon transistors provide the needed response to define the spectrum in this critical range.
If fission foils are a part of the sensor set, the silicon sensor provides confirmation of the spectrum shape.
Bipolar transistors, such as type 2N2222A, are inexpensive, are smaller than fission foils contained in a boron ball, and are sensitive to a part of the neutron spectrum important to the damage of modern silicon electronics. They also can be used directly in arrays to map 1-MeV(Si) equivalent fluence. The proper set of steps to take in reading the transistor-gain degradation is the primary subject of this test method.
SCOPE
1.1 This test method covers the use of 2N2222A silicon bipolar transistors as dosimetry sensors in the determination of neutron energy spectra, and as silicon 1-MeV(Si) equivalent displacement damage fluence monitors.
1.2 The neutron displacement damage is especially valuable as a neutron spectrum sensor in the range 0.1 to 2.0 MeV when fission foils are not available. It has been applied in the fluence range between 2 × 10 12 n/cm2 and 1 × 1014 n/cm2 and should be useful up to 1015 n/cm2. This test method details the steps for the acquisition and use of silicon 1-MeV equivalent fluence information (in a manner similar to the use of activation foil data) for the determination of neutron spectra.
1.3 In addition, this sensor can provide important confirmation of neutron spectra determined with other sensors, and yields a direct measurement of the silicon 1-MeV fluence by the transfer technique.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.
- Standard10 pagesEnglish language
- Standard10 pagesEnglish language
Frequently Asked Questions
E10.07 is a Technical Committee within ASTM International. It is named "Radiation Dosimetry for Radiation Effects on Materials and Devices". This committee has published 101 standards.
E10.07 develops ASTM standards in the area of Information technology. Currently, there are 101 published standards from this technical committee.
ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.
A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.