Standard Practice for Evaluation of Performance Characteristics of Air Quality Measurement Methods with Linear Calibration Functions

ABSTRACT
This practice covers procedures for evaluating the performance characteristics of air quality measurement methods with linear calibration functions. The steps involved in the measurement method used shall be described, and the performance characteristics to be evaluated shall be specified and tested under explicitly specified conditions. The performance characteristics for evaluation include bias, calibration function and linearity, instability, lower detection limit, period of unattended operation, selectivity, sensitivity, and upper limit of measurement.
SCOPE
1.1 This practice2 covers procedures for evaluating the following performance characteristics of air quality measurement methods: bias (in part only), calibration function and linearity, instability, lower detection limit, period of unattended operation, selectivity, sensitivity, and upper limit of measurement.  
1.2 The procedures presented in this practice are applicable only to air quality measurement methods with linear continuous calibration functions, and the output variable of which is a defined time average. The linearity may be due to postprocessing of the primary output variable. Additionally, replicate values belonging to the same input state are assumed to be normally distributed. Components required to transform the primary measurement method output into the time averages desired are regarded as an integral part of this measurement method.  
1.3 For surveillance of measurement method stability under routine measurement conditions, it may suffice to check the essential performance characteristics using simplified tests, the degree of simplification acceptable being dependent on the knowledge on the invariance properties of the performance characteristics previously gained by the procedures presented here.  
1.4 There is no fundamental difference between the instrumental (automatic) and the manual (for example, wet-chemical) procedures, as long as the measured value is an average representative for a predefined time interval. Therefore, the procedures presented are applicable to both. Furthermore, they are applicable to measurement methods for ambient, workplace, and indoor atmospheres, as well as emissions.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Aug-2021
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D5280-96(2021) - Standard Practice for Evaluation of Performance Characteristics of Air Quality Measurement Methods with Linear Calibration Functions
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D5280 − 96 (Reapproved 2021)
Standard Practice for
Evaluation of Performance Characteristics of Air Quality
Measurement Methods with Linear Calibration Functions
This standard is issued under the fixed designation D5280; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.5 This standard does not purport to address all of the
2 safety concerns, if any, associated with its use. It is the
1.1 This practice covers procedures for evaluating the
responsibility of the user of this standard to establish appro-
following performance characteristics of air quality measure-
priate safety, health, and environmental practices and deter-
ment methods: bias (in part only), calibration function and
mine the applicability of regulatory limitations prior to use.
linearity,instability,lowerdetectionlimit,periodofunattended
1.6 This international standard was developed in accor-
operation, selectivity, sensitivity, and upper limit of measure-
dance with internationally recognized principles on standard-
ment.
ization established in the Decision on Principles for the
1.2 The procedures presented in this practice are applicable
Development of International Standards, Guides and Recom-
only to air quality measurement methods with linear continu-
mendations issued by the World Trade Organization Technical
ous calibration functions, and the output variable of which is a
Barriers to Trade (TBT) Committee.
definedtimeaverage.Thelinearitymaybeduetopostprocess-
ing of the primary output variable. Additionally, replicate
2. Referenced Documents
values belonging to the same input state are assumed to be
2.1 ASTM Standards:
normally distributed. Components required to transform the
D1356Terminology Relating to Sampling and Analysis of
primary measurement method output into the time averages
Atmospheres
desired are regarded as an integral part of this measurement
E177Practice for Use of the Terms Precision and Bias in
method.
ASTM Test Methods
1.3 For surveillance of measurement method stability under
E456Terminology Relating to Quality and Statistics
routine measurement conditions, it may suffice to check the
2.2 ISO Standard:
essentialperformancecharacteristicsusingsimplifiedtests,the
ISO 6879:1983Air Quality—Performance Characteristics
degree of simplification acceptable being dependent on the
and Related Concepts for Air Quality Measuring Meth-
knowledge on the invariance properties of the performance
ods
characteristics previously gained by the procedures presented
here.
3. Terminology
1.4 There is no fundamental difference between the instru-
3.1 Definitions:
mental (automatic) and the manual (for example, wet-
3.1.1 For definitions of terms used in this practice, refer to
chemical) procedures, as long as the measured value is an
Terminology D1356.
average representative for a predefined time interval.
3.2 Definitions of Terms Specific to This Standard:
Therefore, the procedures presented are applicable to both.
NOTE 1—The statistical performance characteristics used throughout
Furthermore, they are applicable to measurement methods for
this practice are estimated, by convention, at the confidence level
ambient, workplace, and indoor atmospheres, as well as
1− α=0.95.
emissions.
3.2.1 averaging time, n—predefined time interval length for
which the air quality characteristic is made representative and
∆θ the averaging time.
ThispracticeisunderthejurisdictionofASTMCommitteeD22onAirQuality
and is the direct responsibility of Subcommittee D22.03 on Ambient Atmospheres
and Source Emissions.
Current edition approved Sept. 1, 2021. Published October 2021. Originally For referenced ASTM standards, visit the ASTM website, www.astm.org, or
approved in 1994. Last previous edition approved in 2013 as D5280–96 (2013). contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
DOI: 10.1520/D5280-96R21. Standards volume information, refer to the standard’s Document Summary page on
This practice was adapted from International Standard ISO/DP9169, prepared the ASTM website.
byISO/TC146/SC4/WG4,bythekindpermissionoftheChairmanofISO/TC146 Available from International Organization for Standardization (ISO), 1, ch. de
and the Secretariat of ISO/TC 146/SC 4. la Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http://www.iso.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D5280 − 96 (2021)
3.2.1.1 Discussion—Every measured value obtained is rep- one successively without replacement until the population is
resentative for a defined interval of time, τ, the value of which exhausted, the numbers are said to be drawn in random order.
always lies above a certain minimum due to the intrinsic 3.2.7.1 Discussion—If these numbers have been associated
properties of the measuring procedure applied. In order to in advance with n distinct objects or n distinct operations that
attain mutual comparability of data pertaining to comparable are then rearranged in the order in which the numbers are
objects, a normalization to a common, predefined interval of drawn, the order of the objects or operations is said to be
time is necessary. randomized.
3.2.1.2 Discussion—By convention, this normalization is 3.2.8 reference conditions, n—a specified set of values
(including tolerances) of influence variables delivering repre-
achieved by transformation by means of a simple, linear, and
unweighted averaging process. sentative values of performance characteristics.
3.2.9 variance function, n—avarianceoftheoutputvariable
(a) Series of Discrete Samples:
as a function of the air quality characteristic observed.
cˆ θ ∆θ 5
~ !
?
3.2.10 warm-up time, n—the minimum waiting time for an
K
(1)
instrument to meet predefined values of its performance
cˆ~θ 1 ~k 21! τ τ!
( 0 ?
K
k51
characteristics after activating an instrument stabilized in a
where: nonoperating condition.
3.2.10.1 Discussion—In practice, the warm-up time can be
θ = θ− ∆θ, and
determined by using the performance characteristic that is
Kτ = ∆θ, τ << ∆θ
expected to require the longest interval of time.
(b) Continuous Time Series:
3.2.10.2 Discussion—In the case of the manual procedures,
cˆ θ ∆θ 5
~ !
? run-up time is used correspondingly.
1 θ (2)
dθcˆ θ τ 3.3 Symbols and Abbreviations:
* ~ !
?
∆θ θ
3.3.1 a ,a ,a —coefficientsofthevariancefunctionmodel.
0 1 2
In both cases (a and b), the original sample, described by
3.3.2 b,b —parameters of the estimate function for the
0 1
ĉ(t), is linked to a representative interval of time of length τ
calibration function.
whereas ĉ(∆θ), the result after application of the averaging
3.3.3 C—air quality characteristic.
process,ismaderepresentativefortheintervaloftime ∆θ(just
3.3.4 c—value of C.
preceding θ), the averaging time.
3.2.1.3 Discussion—Theaveragingtime,∆θ,isthereforethe 3.3.5 ĉ—measured value at c.
predefined and, by convention, common time interval length
3.3.6 c—value of C in the i-th sample; this sample may be
i
for which the measured variable ĉ is made representative in a
generated from reference material.
sensethatthesquaredeviationoftheoriginalvalues,attributed
3.3.7 c —normalizationfactorforairqualitycharacteristics;
to time interval lengths τ << ∆θ from ĉ over ∆θ is a minimum.
in this case | c |=1.
3.2.1.4 Discussion—The averaging process can alterna-
3.3.8 ∆c —inaccuracy of C at c.
tively be realized by means of a special sampling technique 1 I
(averaging by sampling).
3.3.9 c¯ —weighted mean, with set of weights ω .
ω k
3.2.2 continuously measuring system, n—asystemreturning
3.3.10 D(b )—drift (see ISO 6879:1983) of the intercept of
acontinuousoutputsignaluponcontinuousinteractionwiththe
the linear calibration function.
air quality characteristic.
3.3.11 D(b )—drift of the slope of the linear calibration
3.2.3 influence variable, n—a variable affecting the interre-
function.
lationship between the (true) values of the air quality charac-
3.3.12 D(ĉ)—drift of the measured value, ĉ,at c.
teristic observed and the corresponding measured values; for
example, variable affecting the slope or the intercept of or the
3.3.13 DEP(b ) —firstordermeasureofdependenceofthe
0 IVi
scatter around the calibration function.
intercept on the influence variable labeled by i.
3.2.4 noncontinuously measuring system, n—a system re-
3.3.14 DEP(b ) —firstordermeasureofdependenceofthe
1 IVi
turning a series of discrete output signals.
slope on the influence variable labeled by i.
3.2.4.1 Discussion—The discretization of the output vari-
3.3.15 DEP(ĉ) —first order measure of dependence of the
IVi
able can be due to sampling in discrete portions or to inner
measured value on the influence variable labeled by i at c.
function characteristics of the system components.
3.3.16 DEP(x) —first order measure of dependence of the
IVi
3.2.5 period of unattended operation, n—the maximum
output signal on the influence variable labeled by i.
admissible interval of time for which the performance charac-
teristicswillremainwithinapredefinedrangewithoutexternal 3.3.17 F—statistic (cf F-test).
servicing, for example, refill, calibration, adjustment.
3.3.18 F —x-quantile of the F-distribution.
x
3.2.6 random variable, n—a variable that may take any of
3.3.19 I —selectivitywithrespecttotheinfluencevariable
IVi
the values of a specified set of values and with which is
labeled by i.
associated a probability distribution.
3.3.20 IV— influence variable labeled by i.
i
3.2.7 randomization, n—if, from a population consisting of
the natural numbers 1 to n, these are drawn at random one by 3.3.21 iv— value of IV.
i i
D5280 − 96 (2021)
3.3.22 ∆ iv—difference of values of IV. 3.3.52 β , β —intercept and slope of the linear calibration
i i 0 1
3.3.23 L—total number of time intervals of the instability function, respectively.
test.
3.3.53 θ—time.
3.3.24 LDL—lower detection limit.
3.3.54 ∆θ—averaging time.
3.3.25 M—total number of samples generated by reference
3.3.55 υ—number of degrees of freedom in the calibration
material within one calibration experiment.
experiment.
3.3.26 N— number of values of the output variable at c.
i i 3.3.56 υ , υ —number of degrees of freedom for the nu-
1 2
merator and denominator of the F-distribution, respectively.
3.3.27 P ,p —estimate of the slope of the regression
|ll u
function of the output variable on time at c = c , c = c ,
|ll u 3.3.57 ω = ω(c)—continuous weighing factor gained by
respectively.
modeling s.
i
3.3.28 R—reproducibility.
3.3.58 ω —weighing factor at c .
1 1
3.3.29 r—repeatability.
4. Requirements and Provisions
3.3.30 RES — resolution at C = c.
c
4.1 Description of the Steps of the Measurement Methods
3.3.31 ŝ—estimate of the smoothed standard deviation of X
Under Test—Describe all steps of the measurement method
at c.
used, such as sampling, analysis, postprocessing, and calibra-
3.3.32 ŝ —smoothedestimateofthevarianceof X(repeated
tion. Fig. 1 illustrates schematically the steps to be followed in
measurements) at c. making a measurement or performing a series of calibration
experiments in order to determine the performance character-
3.3.33 s —normalization factor for the standard deviation;
istics.
the magnitude of s equals to 1.
NOTE 2—Under certain conditions it may be suitable to test only one
3.3.34 s ,s —estimate of the standard deviation of insta-
b b
0 1
step or a selected group of steps of the measurement method. Under other
bility(seeISO6879:1983)oftheinterceptandtheslopeofthe
conditions it may not be possible to include all the steps of the
linear calibration function.
measurement method. However, include as many steps as possible.
3.3.35 sc—estimate of the standard deviation of instability
4.2 Specification of Performance Characteristics to Be
at c.
Tested—Specify the performance characteristics of the mea-
3.3.36 s— estimate of the standard deviation of repeated x
surement method in order of their relevance for the final
i
at c ; j repetition index.
ij i
3.3.37 ŝ—smoothed estimate of the standard deviation of
i
“repeated” x at c; j repetition index.
ij i
3.3.38 s — estimate of the repeatability standard deviation.
r
3.3.39 s —estimate of the standard deviation of the experi-
ĉx
mentally determined calibration function (in units of the air
quality characteristic).
3.3.40 s — estimate of the standard deviation of the
xc
experimentally determined calibration function (in units of the
output variable).
3.3.41 t —q-quantileofthet-distributionwithυdegreesof
υ;q
freedom.
3.3.42 TC—test characteristic of Grubbs’ outlier test.
3.3.43 X—output variable.
3.3.44 x—value of X.
3.3.45 x—estimate of x.
3.3.46 x— estimate of output signal at c .
i i
3.3.47 x¯ —mean of the set of output signals at c.
i i
3.3.48 x —output signal at c with the highest absolute
i,extr i
distance from x¯ .
i
3.3.49 x —j-th output signal at c.
ij i
3.3.50 x ,x —output signal after i time intervals at the
l;i' u;i
lower and upper value of the air quality characteristic of
NOTE 1— ____Measurement Branch.
reference material.
NOTE2—___ Calibration Branch.
3.3.51 x¯ —weightedmeanofthewholesetofoutputsignals
ω FIG. 1 Schematic of the Procedures of Measurement and of
within the calibration experiment. Evaluation for Performance Characteristics
D5280 − 96 (2021)
assessment of accuracy. The descriptors of the calibration time, filling time, or accumulation time, depending on the
function, for example, intercept, β , and slope, β , as well as measurement method.
0 1
their qualifying performance characteristics are vital. Those
5.2 FunctionalandStatisticalPerformanceCharacteristics:
performance characteristics for which prior knowledge is
5.2.1 The performance characteristics to be determined are:
available, and those pertaining to influence variables covered
5.2.1.1 Performancecharacteristicsrelatedtothecalibration
by randomization are of lesser importance and need not be
function and its stability under reference conditions, and
determined.
5.2.1.2 Performance characteristics related to the depen-
4.3 Test Conditions—Perform the tests under explicitly dence of the calibration function on influence variables.
stated conditions representative of the operational measure-
5.2.2 Determine a linear calibration function by its slope
ments. When testing for performance characte
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.