Standard Test Method for Performance Validation of Thermomechanical Analyzers

SIGNIFICANCE AND USE
5.1 This test method may be used to determine and validate the performance of a particular thermomechanical analyzer apparatus.  
5.2 This test method may be used to determine and validate the performance of a particular method based upon thermomechanical analyzer temperature or length change measurements.  
5.3 This test method may be used to determine the repeatability of a particular apparatus, operator, or laboratory.  
5.4 This test method may be used for specification and regulatory compliance purposes.
SCOPE
1.1 This test method provides procedures for validating temperature and length change measurements of thermomechanical analyzers (TMA) and analytical methods based upon the measurement of temperature and length change. Performance parameters include temperature repeatability, linearity and bias; and dimension change repeatability, detection limit, quantitation limit, linearity and bias.  
1.2 Validation of apparatus performance and analytical methods is a necessary requirement for quality initiatives. Results may also be used for legal purposes.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
31-Mar-2013
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E2918-13 - Standard Test Method for Performance Validation of Thermomechanical Analyzers
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E2918 − 13
Standard Test Method for
1
Performance Validation of Thermomechanical Analyzers
This standard is issued under the fixed designation E2918; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 3. Terminology
1.1 This test method provides procedures for validating 3.1 Technical terms used in this test method are defined in
temperature and length change measurements of thermome-
Terminologies E473, E1142, and E2161, including terms
chanical analyzers (TMA) and analytical methods based upon
analyte, bow, Celsius, coeffıcient of linear thermal expansion,
the measurement of temperature and length change. Perfor-
detection limit, linearity, quantification limit, relative standard
mance parameters include temperature repeatability, linearity
deviation, repeatability, standard deviation,
and bias; and dimension change repeatability, detection limit,
thermodilatometry, thermomechanical analysis, and valida-
quantitation limit, linearity and bias.
tion.
1.2 Validation of apparatus performance and analytical
4. Summary of Test Method
methods is a necessary requirement for quality initiatives.
Results may also be used for legal purposes.
4.1 Temperature and time are the primary independent
parameters and length change is the primary dependent experi-
1.3 The values stated in SI units are to be regarded as
mental parameters provided by thermomechanical analysis.
standard. No other units of measurement are included in this
standard.
4.2 Temperature,ameasuredvalue,isvalidatedbyperform-
1.4 This standard does not purport to address all of the ing a measurement of the penetration in sharply melting
safety concerns, if any, associated with its use. It is the materials at three (or more) different known melting tempera-
responsibility of the user of this standard to establish appro- tures.
priate safety and health practices and determine the applica-
4.3 Length change, a measured value, is validated by
bility of regulatory limitations prior to use.
performing a measurement of the linear thermal expansion for
three (or more) test materials.
2. Referenced Documents
4.4 Validation of a thermomechanical test method based
2
2.1 ASTM Standards:
upon length change may be performed using the test specimen
E473 Terminology Relating to Thermal Analysis and Rhe-
as the analyte.
ology
4.5 The length change of three (or more) specimens (nomi-
E1142 Terminology Relating to Thermophysical Properties
E1363 Test Method for Temperature Calibration of Thermo- nally representing the maximum, midpoint and minimum of
the range of the test method) are measured in triplicate (or
mechanical Analyzers
E1970 PracticeforStatisticalTreatmentofThermoanalytical more). A fourth blank specimen, containing no analyte, is
measured in triplicate (or more).
Data
E2113 Test Method for Length Change Calibration of Ther-
NOTE1—Repeatabilityisdeterminedbyperformingasufficientnumber
momechanical Analyzers
of determinations to calculate valid estimates of the standard deviation or
E2161 Terminology Relating to Performance Validation in
relative standard deviation of the measurement.
Thermal Analysis
4.5.1 Temperature and length change measurement linearity
and bias are determined from the linear regression correlation
of the results from measurements of the three (or more)
1
ThistestmethodisunderthejurisdictionofASTMCommitteeE37onThermal
specimens.
Measurements and is the direct responsibility of Subcommittee E37.10 on
4.5.2 Length change detection limit and quantitation limit
Fundamental, Statistical and Mechanical Properties.
are determined from the standard deviation of the blank
Current edition approved April 1, 2013. Published May 2013. DOI: 10.1520/
E2918-13.
determination with no analyte present.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
4.5.3 Temperature and length change repeatability are de-
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
termined from the repeatability measurements of three (or
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. more) specimens.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E2918 − 13
4.5.4 Length change validation is independent of the tem- 6.1.8 A means of sustaining an environment around the
perature validation. The respective validations need not in- specimen of a dry, inert gas at a purge rate of 10 to 50 mL/min
volve consistent ranges. 6 5 mL/min.
NOTE 5—
...

Questions, Comments and Discussion