ASTM E1695-20
(Test Method)Standard Test Method for Measurement of Computed Tomography (CT) System Performance
Standard Test Method for Measurement of Computed Tomography (CT) System Performance
SIGNIFICANCE AND USE
4.1 The major factors affecting the quality of a CT image are total image unsharpness (UTimage), contrast (Δµ), and random noise (σ). Geometrical and detector unsharpness limit the spatial resolution of a CT system, that is, its ability to image fine structural detail in an object. Random noise and contrast response limit the contrast sensitivity of a CT system, that is, its ability to detect the presence or absence of features in an object. Spatial resolution and contrast sensitivity may be measured in various ways. In this test method, spatial resolution is quantified in terms of the modulation transfer function (MTF), and contrast sensitivity is quantified in terms of the contrast discrimination function (CDF). The relationship between contrast sensitivity and spatial resolution describing the resolving and detecting capabilities is given by the contrast-detail-diagram (CDD metric, see also Guide E1441 and Practice E1570). This test method allows the purchaser or the provider of CT systems or services, or both, to measure and specify spatial resolution and contrast sensitivity and is a measure for system stability over time and performance acceptability.
SCOPE
1.1 This test method provides instruction for determining the spatial resolution and contrast sensitivity in X-ray and γ-ray computed tomography (CT) volumes. The determination is based on examination of the CT volume of a uniform cylinder of material. The spatial resolution measurement (Modulation Transfer Function) is derived from an image analysis of the sharpness at the edges of the reconstructed cylinder slices. The contrast sensitivity measurement (Contrast Discrimination Function) is derived from an image analysis of the contrast and the statistical noise at the center of the cylinder slices.
1.2 This test method is more quantitative and less susceptible to interpretation than alternative approaches because the required cylinder is easy to fabricate and the analysis easy to perform.
1.3 This test method is not to predict the detectability of specific object features or flaws in a specific application. This is subject of IQI and RQI standards and standard practices.
1.4 This method tests and describes overall CT system performance. Performance tests of systems components such as X-ray tubes, gamma sources, and detectors are covered by separate documents, namely Guide E1000, Practice E2737, and Practice E2002; c.f. 2.1, which should be consulted for further system analysis.
1.5 Units—The values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are provided for information only and are not considered standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1695 − 20
Standard Test Method for
Measurement of Computed Tomography (CT) System
1
Performance
This standard is issued under the fixed designation E1695; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.
1. Scope 1.7 This international standard was developed in accor-
dance with internationally recognized principles on standard-
1.1 This test method provides instruction for determining
ization established in the Decision on Principles for the
thespatialresolutionandcontrastsensitivityinX-rayand γ-ray
Development of International Standards, Guides and Recom-
computed tomography (CT) volumes. The determination is
mendations issued by the World Trade Organization Technical
based on examination of the CT volume of a uniform cylinder
Barriers to Trade (TBT) Committee.
of material. The spatial resolution measurement (Modulation
Transfer Function) is derived from an image analysis of the
2. Referenced Documents
sharpness at the edges of the reconstructed cylinder slices. The
2
2.1 ASTM Standards:
contrast sensitivity measurement (Contrast Discrimination
E177 Practice for Use of the Terms Precision and Bias in
Function) is derived from an image analysis of the contrast and
ASTM Test Methods
the statistical noise at the center of the cylinder slices.
E691 Practice for Conducting an Interlaboratory Study to
1.2 This test method is more quantitative and less suscep-
Determine the Precision of a Test Method
tible to interpretation than alternative approaches because the
E1000 Guide for Radioscopy
required cylinder is easy to fabricate and the analysis easy to
E1316 Terminology for Nondestructive Examinations
perform.
E1441 Guide for Computed Tomography (CT)
E1570 Practice for Fan Beam Computed Tomographic (CT)
1.3 This test method is not to predict the detectability of
Examination
specific object features or flaws in a specific application. This
E2002 Practice for Determining Total Image Unsharpness
is subject of IQI and RQI standards and standard practices.
and Basic Spatial Resolution in Radiography and Radios-
1.4 This method tests and describes overall CT system
copy
performance. Performance tests of systems components such
E2737 Practice for Digital Detector Array Performance
as X-ray tubes, gamma sources, and detectors are covered by
Evaluation and Long-Term Stability
separatedocuments,namelyGuideE1000,PracticeE2737,and
3
2.2 ISO Standard:
Practice E2002; c.f. 2.1, which should be consulted for further
15708 NDT – Radiation Methods – Computed Tomography
system analysis.
– Part 1: Terminology, Part 2: Principles, Equipment and
1.5 Units—The values stated in SI units are to be regarded
Samples, Part 3: Operation and Interpretation, Part 4:
as standard. The values given in parentheses after SI units are
Qualification
provided for information only and are not considered standard.
3. Terminology
1.6 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the 3.1 Definitions—The definitions of terms relating to
responsibility of the user of this standard to establish appro- Gamma- and X-Radiology, which appear in Terminology
priate safety, health, and environmental practices and deter- E1316 and Guide E1441, shall apply to the terms used in this
mine the applicability of regulatory limitations prior to use. test method.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
1
This test method is under the jurisdiction of ASTM Committee E07 on contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Nondestructive Testing and is the direct responsibility of Subcommittee E07.01 on Standards volume information, refer to the standard’s Document Summary page on
Radiology (X and Gamma) Method. the ASTM website.
3
Current edition approved June 1, 2020. Published August 2020. Originally Available from International Organization for Standardization (ISO), ISO
approved in 1995. Last previous edition approved in 2013 as E1695 – 95(2013). Central Secretariat, BIBC II, Chemin de Blandonnet 8, CP 401, 1214 Vernier,
DOI: 10.1520/E1695-20. Geneva, Switzerland, http://www.iso.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1695 − 20
3.2 Definitions of Terms Specific to This Standard: 6. Procedure of Measurement
3.2.1 examination object, n—a part or specimen being
6.1 The phantom shall be mounted on the CT system with
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1695 − 95 (Reapproved 2013) E1695 − 20
Standard Test Method for
Measurement of Computed Tomography (CT) System
1
Performance
This standard is issued under the fixed designation E1695; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.
1. Scope
1.1 This test method provides instruction for determining the spatial resolution and contrast sensitivity in X-ray and γ-ray
computed tomography (CT) images.volumes. The determination is based on examination of the CT imagevolume of a uniform
diskcylinder of material. The spatial resolution measurement (Modulation Transfer Function) is derived from an image analysis
of the sharpness at the edgeedges of the disk. reconstructed cylinder slices. The contrast sensitivity measurement (Contrast
Discrimination Function) is derived from an image analysis of the contrast and the statistical noise at the center of the disk.cylinder
slices.
1.2 This test method is more quantitative and less susceptible to interpretation than alternative approaches because the required
diskcylinder is easy to fabricate and the analysis is immune to cupping artifacts. This test method may not yield meaningful results
if the disk image occupies less than a significant fraction of the field of view.easy to perform.
1.3 This test method may also be used to evaluate other performance parameters. Among those characteristics of a CT system that
are detectable with this test method are: the mid-frequency enhancement of the reconstruction kernel, the presence (or absence)
of detector crosstalk, the undersampling of views, and the clipping of unphysical (that is, negative) CT numbers (see Air Force
Technical Report WL-TR-94-4021is not to predict the detectability of specific object features or flaws in a specific application. This
). It is highly likely that other characteristics as well can be detected with this test method.is subject of IQI and RQI standards and
standard practices.
1.4 This method tests and describes overall CT system performance. Performance tests of systems components such as X-ray
tubes, gamma sources, and detectors are covered by separate documents, namely Guide E1000, Practice E2737, and Practice
E2002; c.f. 2.1, which should be consulted for further system analysis.
1.5 Units—The values stated in SI units are to be regarded as the standard. Inch-pound standard. The values given in parentheses
after SI units are provided for information only.only and are not considered standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1
This test method is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.01 on Radiology
(X and Gamma) Method.
Current edition approved June 1, 2013June 1, 2020. Published June 2013August 2020. Originally approved in 1995. Last previous edition approved in 20062013 as
ε1
E1695 - 95E1695 – 95(2006)(2013). . DOI: 10.1520/E1695-95R13.10.1520/E1695-20.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1695 − 20
2. Referenced Documents
2
2.1 ASTM Standards:
E177 Practice for Use of the Terms Precision and Bias in ASTM Test Methods
E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
E1000 Guide for Radioscopy
E1316 Terminology for Nondestructive Examinations
E1441 Guide for Computed Tomography (CT)
E1570 Practice for Fan Beam Computed Tomographic (CT) Examination
E2002 Practice for Determining Total Image Unsharpness and Basic Spatial Resolution in Radiography and Radioscop
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.