ASTM E1047-85(2005)
(Specification)Standard Specification for Blood Sedimentation Tube, Wintrobe, Glass, Disposable
Standard Specification for Blood Sedimentation Tube, Wintrobe, Glass, Disposable
ABSTRACT
This specification covers disposable blood sedimentation tubes suitable for determining sedimentation rates and the volume of packed red blood cells. This specification covers a tube intended for one-time use only. The tubes shall be fabricated borosilicate glass, Type I, Class B, or from soda lime glass, Type II. The tubes shall be tested for their marking permanency and resistance to centrifugal test.
SCOPE
1.1 This specification covers disposable blood sedimentation tubes suitable for determining sedimentation rates and the volume of packed red blood cells.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 The following precautionary statement pertains only to the test method portion, Section 7 of this specification. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1047 – 85 (Reapproved 2005)
Standard Specification for
Blood Sedimentation Tube, Wintrobe, Glass, Disposable
This standard is issued under the fixed designation E1047; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 5. Materials
1.1 This specification covers disposable blood sedimenta- 5.1 Glass—The tubes made to this specification shall be
tion tubes suitable for determining sedimentation rates and the fabricatedborosilicateglass,TypeI,ClassB,orfromsodalime
volume of packed red blood cells. glass, Type II, in accordance with Specification E438.
1.2 The values stated in SI units are to be regarded as
6. Physical Requirements
standard. No other units of measurement are included in this
standard. 6.1 Design—The tube shall be made of tubing with a
minimum wall thickness of 0.6 mm. The tube shall have a
1.3 The following precautionary statement pertains only to
the test method portion, Section 7 of this specification. This cylindrical body terminating in a round bottom on the outside
standard does not purport to address all of the safety concerns, and a flat bottom on the inside. The tube shall have a uniform
bore and a fire-polished top.
if any, associated with its use. It is the responsibility of the user
of this standard to establish appropriate safety and health 6.2 Workmanship—The tube shall be free as possible from
visible defects which detract from its serviceability. The tube
practices and determine the applicability of regulatory limita-
tions prior to use. shall be well annealed so there will be no ring strain evident
when tested under a polariscope.
2. Referenced Documents
6.3 Dimensions—The tube shall be made of tubing with an
2.1 ASTM Standards: outside diameter (O.D.) of 5.0 6 0.5 mm and an inside
E438 Specification for Glasses in Laboratory Apparatus diameter (I.D.) of 2.9 to 3.5 mm. The uniformity of the bore
shall be 60.15 mm throughout the tube. The tube shall be 110
3. Terminology
to 117 mm long and have a graduated scale of 100 mm.
3.1 Definitions of Terms Specific To This Standard: Manufacturers have the option to place additional graduation
3.1.1 disposable—in accordance with this specification and
lines on the tube to a maximum of 105 mm. Maximum overall
the expected product performance expressed in this standard, tolerance on the total graduation scale shall be 61.0 mm. The
those tubes which are to be used one time only.Any institution
tube shall be legibly marked with the maker’s or vendor’s
or individual who reuses a disposable tube must bear full name or mark and possess a wide color band at the top of the
responsibility for its safety and effectiveness.
tube for marking purposes.
3.1.2 Wintrobe —surname of the individual responsible for 6.4 Graduation Scale—The tube shall be graduated 100 to
the design of the reusableWintrobe tube and the method of use
105 6 1.0 mm in 1 mm divisions and numbered every
for both reusable and disposable tubes. centimetre with two sets of numerals. One set of graduation
numerals shall be from 0 to 9 cm down the left side of the
4. Classification
graduation scale and the other set of graduation numerals shall
4.1 This specification covers a tube that is intended to be
befrom1to10cm up the right side of the graduation scale as
usedonetimeonly.Itisnottobeconfusedwithareusabletube
shown in Fig. 1.
that is described in other published standards.
6.5 Graduation Lines—The thickness of the graduation
lines may vary in uniformity but shall not exceed 0.4 mm in
thickness. They shall lie at right angles to the axis of the tube
This specification is under the juri
...