Standard Guide for Measuring Securement of Balloon Expandable Vascular Stent Mounted on Delivery System

SIGNIFICANCE AND USE
The securement of the endovascular stent on the balloon is a critical parameter to ensure that the stent is safely delivered to or from the treatment site.
This guide is intended for use by researchers and manufacturers for the development and selection of pre-test treatments, tests and test endpoints to measure stent securement (displacement distances and dislodgment forces).
This guide may be used to investigate which practical combinations of in vitro tests best characterize clinical scenarios.
This guide should be used with discretion in choosing securement tests and evaluating results due to the myriad possible combinations of clinical conditions, failure modes, and stent delivery system designs.
This guide may be of use for developing a test for meeting parts 2 and 3 of the requirements of EN 14299, Section 7.3.4.4 on Trackability.
This guide may be of use for developing a test to meet section VII-C-8 of CDRH Guidance document.
SCOPE
1.1 This guide provides guidance for the design and development of pre-test treatments, tests, and test endpoints to measure stent securement of pre-mounted, unsheathed, balloon-expandable stent delivery systems. This guide is intended to aid investigators in the design, development, and in vitro characterization of pre-mounted, unsheathed, balloon-expandable stent delivery systems.
1.2 This guide covers the laboratory determination of the shear force required to displace or dislodge a balloon-expandable endovascular stent mounted on a delivery system. The guide proposes a set of options to consider when testing stent securement. The options cover pre-test treatments, possible stent securement tests, and relevant test endpoints. An example test apparatus is given in .
1.3 This guide covers in vitro bench testing characterization only. Measured levels of securement and product design/process differentiation may be particularly influenced by selections of pre-test treatments, securement test type (for example, stent gripping method), and test endpoint. In vivo characteristics may also differ from in vitro results.
1.4 This guide does not cover all possible pre-test treatments, stent securement tests, or test endpoints. It is intended to provide a starting point from which to select and investigate securement test options.
1.5 This guide does not specify a method for mounting the stent onto the delivery system.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.

General Information

Status
Historical
Publication Date
14-Jul-2007
Current Stage
Ref Project

Relations

Buy Standard

Guide
ASTM F2394-07 - Standard Guide for Measuring Securement of Balloon Expandable Vascular Stent Mounted on Delivery System
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: F2394 − 07
StandardGuide for
Measuring Securement of Balloon Expandable Vascular
1
Stent Mounted on Delivery System
This standard is issued under the fixed designation F2394; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2
2.1 ASTM Standards:
1.1 This guide provides guidance for the design and devel-
E1169 Practice for Conducting Ruggedness Tests
opment of pre-test treatments, tests, and test endpoints to
E1488 GuideforStatisticalProcedurestoUseinDeveloping
measure stent securement of pre-mounted, unsheathed,
and Applying Test Methods
balloon-expandable stent delivery systems. This guide is in-
2.2 Other Documents:
tended to aid investigators in the design, development, and in
ISO 10555-1 Sterile Sterile Sterile, Single-use Intravascular
vitro characterization of pre-mounted, unsheathed, balloon-
3
Catheters—Part 1: General Requirements
expandable stent delivery systems.
Quality System Regulation, Part VII Dept. Health and Hu-
1.2 This guide covers the laboratory determination of the
man Services, Food and Drug Administration, 21 CFR
shear force required to displace or dislodge a balloon-
Part 820 Medical Devices; Current Good Manufacturing
4
expandable endovascular stent mounted on a delivery system.
Practice; Final Rule. Federal Register, October 7, 1996
The guide proposes a set of options to consider when testing
EN 14299 Non Active Surgical Implants—Particular Re-
stent securement. The options cover pre-test treatments, pos-
quirements for Cardiac and Vascular Implants—Specific
5
sible stent securement tests, and relevant test endpoints. An
Requirements For Arterial Stents, May 2004
example test apparatus is given in 7.1.
CDRH Guidance, Non-Clinical Tests and Recommended
Labeling for Intravascular Stents andAssociated Delivery
1.3 This guide covers in vitro bench testing characterization
6
Systems, January 13, 2005
only. Measured levels of securement and product design/
7
MAUDE Database
process differentiation may be particularly influenced by selec-
tions of pre-test treatments, securement test type (for example,
3. Terminology
stent gripping method), and test endpoint. In vivo characteris-
3.1 Definitions:
tics may also differ from in vitro results.
3.1.1 balloon expandable stent, n—a stent that is expanded
1.4 This guide does not cover all possible pre-test
at the treatment site by a balloon catheter. The stent material is
treatments, stent securement tests, or test endpoints. It is
plastically deformed by the balloon expansion such that the
intended to provide a starting point from which to select and
stent remains expanded after deflation of the balloon.
investigate securement test options.
3.1.2 crimp, v—tosecurethestentonthedeliverysystemby
radially compressing and plastically deforming the stent onto
1.5 This guide does not specify a method for mounting the
the balloon.
stent onto the delivery system.
1.6 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the 2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
responsibility of the user of this standard to establish appro-
Standards volume information, refer to the standard’s Document Summary page on
priate safety and health practices and determine the applica-
the ASTM website.
bility of regulatory requirements prior to use.
3
Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
4th Floor, New York, NY 10036, http://www.ansi.org.
4
AvailablefromU.S.GovernmentPrintingOfficeSuperintendentofDocuments,
732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http://
1
This guide is under the jurisdiction of ASTM Committee F04 on Medical and www.access.gpo.gov.
5
Surgical Materials and Devices and is the direct responsibility of Subcommittee Available from British Standards Institute (BSI), 389 Chiswick High Rd.,
F04.30 on Cardiovascular Standards. London W4 4AL, U.K., http://www.bsi-global.com.
6
Current edition approved July 15, 2007. Published August 2007. Originally Available from Food and Drug Administration (FDA), 5600 Fishers Ln.,
approved in 2004. Last previous edition approved in 2004 as F2394 – 04. DOI: Rockville, MD 20857,. Http://www.fda.gov/cdrh/ode/guidance/1545.pdf.
7
10.1520/F2394-07. Http://www.fda.gov/cdrh/maude.html.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
F2394 − 0
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.