ABSTRACT
This specification covers reusable blood sedimentation tubes suitable for determining sedimentation rates and the volume of packed red blood cells. The tubes shall be fabricated from borosilicate glass, Type I, Class B, or sodalime glass, Type II. A resistance to centrifugal force test shall be performed on the tubes.
SCOPE
1.1 This specification covers reusable blood sedimentation tubes suitable for determining sedimentation rates and the volume of packed red blood cells.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E961 −97 (Reapproved 2013)
Standard Specification for
1
Blood Sedimentation Tube, Wintrobe, Glass, Reusable
This standard is issued under the fixed designation E961; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 4. Classification
1.1 This specification covers reusable blood sedimentation
4.1 This specification covers a tube that is intended to be
tubes suitable for determining sedimentation rates and the
used until it is no longer considered a functional device for the
volume of packed red blood cells.
purpose intended.
1.2 The values stated in SI units are to be regarded as
5. Materials
standard. No other units of measurement are included in this
standard.
5.1 Glass—The tubes made to this specification shall be
1.3 This standard does not purport to address all of the
fabricated from borosilicate glass, Type I, Class B, or soda-
safety concerns, if any, associated with its use. It is the
lime glass, Type II, in accordance with Specification E438.
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
6. Dimensions and Graduations
bility of regulatory limitations prior to use.
6.1 Dimensions—The tube shall be made of tubing with an
outside diameter (O.D.) of 7.0 to 8.0 mm with an inside
2. Referenced Documents
diameter (I.D.) of 2.9 to 3.3 mm. The uniformity of the bore
2
2.1 ASTM Standards:
shall be 60.1 mm throughout the tube. The tube shall be 110
E438 Specification for Glasses in Laboratory Apparatus
to 117 mm long and have a graduated scale of 105 6 0.25 mm
E920 Specification for Commercially Packaged Laboratory
from the inside bottom of the tube. The tube shall be legibly
Apparatus
marked with the manufacturer’s or vendor’s name or mark and
E921 Specification for Export Packaged Laboratory Appa-
possess a frosted area for marking purposes.
ratus
E1133 Practice for Performance Testing of Packaged Labo-
6.2 Graduation Scale—The tube shall be graduated 105 6
ratory Apparatus for United States Government Procure-
0.25mmin1-mmdivisionsandnumberedevery1cmwithtwo
ments
sets of numerals. One set of graduation numerals shall be from
E1157 Specification for Sampling and Testing of Reusable
0 to 9 cm down the left side of the graduation scale and the
Laboratory Glassware
other set of g (20 to 25°C) for 15 min. Remove tube from the
solution and thoroughly rinse in tap water followed by distilled
3. Terminology
water. Dry the tube by rubbing vigorously, 5 to 10 strokes, with
a laboratory cloth or tissue. This appearance of the markings
3.1 Definitions of Terms Specific to This Standard:
should be the same as before the test, when judged by eye
3.1.1 reusable—Capable of being used again.
under normal room lighting.
3.1.2 Wintrobe—The surname of the individual responsible
3