ASTM E1311-14(2018)
(Practice)Standard Practice for Minimum Detectable Temperature Difference for Thermal Imaging Systems
Standard Practice for Minimum Detectable Temperature Difference for Thermal Imaging Systems
SIGNIFICANCE AND USE
5.1 This practice gives a measure of a thermal imaging system's effectiveness for detecting a small spot within a large background. Thus, it relates to the detection of small material defects such as voids, pits, cracks, inclusions, and occlusions.
5.2 MDTD values provide estimates of detection capability that may be used to compare one system with another. (Lower MDTD values indicate better detection capability.)
5.3 Due to the partially subjective nature of the procedure, repeatability and reproducibility are apt to be poor and MDTD differences less than 0.2 °C are considered to be insignificant.
Note 2: Values obtained under idealized laboratory conditions may or may not correlate directly with service performance.
SCOPE
1.1 This practice covers the determination of the minimum detectable temperature difference (MDTD) capability of a compound observer-thermal imaging system as a function of the angle subtended by the target.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1311 − 14 (Reapproved 2018)
Standard Practice for
Minimum Detectable Temperature Difference for Thermal
Imaging Systems
This standard is issued under the fixed designation E1311; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 3.1.2.1 Discussion—The size of the field of view is custom-
arily expressed in units of degrees.
1.1 This practice covers the determination of the minimum
detectable temperature difference (MDTD) capability of a 3.1.3 See also Terminology E1316.
compound observer-thermal imaging system as a function of
4. Summary of Practice
the angle subtended by the target.
4.1 A standard circular target is used in conjunction with a
1.2 The values stated in SI units are to be regarded as
differentialblackbodythatcanestablishoneblackbodyisother-
standard. No other units of measurement are included in this
mal temperature for the target and another blackbody isother-
standard.
mal temperature for the background by which the target is
1.3 This standard does not purport to address all of the
framed. The target, at an undisclosed orientation, is imaged
safety concerns, if any, associated with its use. It is the
onto the monochrome video monitor of a thermal imaging
responsibility of the user of this standard to establish appro-
system whence the image may be viewed by an observer. The
priate safety, health, and environmental practices and deter-
temperature difference between the target and the background,
mine the applicability of regulatory limitations prior to use.
initially zero, is increased incrementally until the observer, in a
1.4 This international standard was developed in accor-
limited duration, can just distinguish the target. This critical
dance with internationally recognized principles on standard-
temperature difference is the MDTD.
ization established in the Decision on Principles for the
Development of International Standards, Guides and Recom- NOTE 1—Observers must have good eyesight and be familiar with
viewing thermal imagery.
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
4.2 The temperature distributions of each target and its
background are measured remotely at the critical temperature
2. Referenced Documents
difference that defines the MDTD.
2.1 ASTM Standards:
4.3 The background temperature and the angular subtense
E1316 Terminology for Nondestructive Examinations
for each target are specified together with the measured value
3. Terminology of MDTD. The (fixed) field of view included by the back-
ground is also specified.
3.1 Definitions:
3.1.1 differential blackbody—an apparatus for establishing 4.4 The probability of detection is specified together with
two parallel isothermal planar zones of different temperatures,
the reported value of MDTD.
and with effective emissivities of 1.0.
5. Significance and Use
3.1.2 field of view (FOV)—the shape and angular dimen-
5.1 This practice gives a measure of a thermal imaging
sions of the cone or the pyramid that define the object space
system’s effectiveness for detecting a small spot within a large
imaged by the system; for example, rectangular, 4-deg wide by
background. Thus, it relates to the detection of small material
3-deg high.
defects such as voids, pits, cracks, inclusions, and occlusions.
This practice is under the jurisdiction of ASTM Committee E07 on Nonde-
5.2 MDTD values provide estimates of detection capability
structive Testing and is the direct responsibility of Subcommittee E07.10 on
that may be used to compare one system with another. (Lower
Specialized NDT Methods.
MDTD values indicate better detection capability.)
Current edition approved Nov. 1, 2018. Published December 2018. Originally
approved in 1989. Last previous edition approved in 2014 as E1311 – 14. DOI:
5.3 Due to the partially subjective nature of the procedure,
10.1520/E1311-14R18.
repeatability and reproducibility are apt to be poor and MDTD
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
differences less than 0.2 °C are considered to be insignificant.
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. NOTE 2—Values obtained under idealized laboratory conditions may or
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1311 − 14 (2018)
may not correlate directly with service performance.
7.4 Adjust the monochrome video monitor controls so that
the presence of noise is barely perceivable by the observer.
6. Apparatus
7.5 Make the display luminance and the laboratory ambient
6.1 The apparatus consists of the following:
luminance mutually suitable for visual acuity and viewing
6.1.1 Target Plates, containing single or multiple circular
comfort.
targets of area(s) not greater than 5 % of the combined areas of
7.6 Advise the observer that a visible spot will eventually
target and background (that is, FOV area), and with the
appear in the monitor’s display. Instruct him to signal when he
distance from the center of the target to the center of the FOV
can perceive the spot and to cite its orientation relative to the
equal to one third of the height or the diameter of the FOV. See
12 h of a clock; for example, 1 o’clock, 2 o’clock, 3 o’clock,
Fig. 1.
etc. Refrain from further conversation during the procedure
NOTE 3—A target plate may be fabricated by cutting one or more
that could conceivably influence or bias the observer.
circular apertures in a metal plate of high thermal conductivity, such as
aluminum, and coating with black paint of emissivity greater than 0.95. In 7.7 Set∆T (the temperature of the target minus the nominal
this case an aperture would constitute a target, and the coated metal
temperature of the background) equal to zero.
surrounding the target and within the field of view of the thermal imaging
7.8 Increase∆Tin positive increments not exceeding 0.1 °C
system would constitute the target’s background.
every 60 s or until the observer signals. If the identification is
6.1.2 Facility, for mounting target plates and varying the
incorrect, continue as before.
orientation of any given
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1311 − 14 E1311 − 14 (Reapproved 2018)
Standard Practice for
Minimum Detectable Temperature Difference for Thermal
Imaging Systems
This standard is issued under the fixed designation E1311; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*Scope
1.1 This practice covers the determination of the minimum detectable temperature difference (MDTD) capability of a compound
observer-thermal imaging system as a function of the angle subtended by the target.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety problems,concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and
determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
E1316 Terminology for Nondestructive Examinations
3. Terminology
3.1 Definitions:
3.1.1 differential blackbody—an apparatus for establishing two parallel isothermal planar zones of different temperatures, and
with effective emissivities of 1.0.
3.1.2 field of view (FOV)—the shape and angular dimensions of the cone or the pyramid that define the object space imaged
by the system; for example, rectangular, 4-deg wide by 3-deg high.
3.1.2.1 Discussion—
The size of the field of view is customarily expressed in units of degrees.
3.1.3 See also Terminology E1316.
4. Summary of Practice
4.1 A standard circular target is used in conjunction with a differential blackbody that can establish one blackbody isothermal
temperature for the target and another blackbody isothermal temperature for the background by which the target is framed. The
target, at an undisclosed orientation, is imaged onto the monochrome video monitor of a thermal imaging system whence the image
may be viewed by an observer. The temperature difference between the target and the background, initially zero, is increased
incrementally until the observer, in a limited duration, can just distinguish the target. This critical temperature difference is the
MDTD.
NOTE 1—Observers must have good eyesight and be familiar with viewing thermal imagery.
This practice is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.10 on Specialized NDT
Methods.
Current edition approved Oct. 1, 2014Nov. 1, 2018. Published October 2014December 2018. Originally approved in 1989. Last previous edition approved in 20102014
as E1311 - 89 (2010).E1311 – 14. DOI: 10.1520/E1311-14.10.1520/E1311-14R18.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1311 − 14 (2018)
4.2 The temperature distributions of each target and its background are measured remotely at the critical temperature difference
that defines the MDTD.
4.3 The background temperature and the angular subtense for each target are specified together with the measured value of
MDTD. The (fixed) field of view included by the background is also specified.
4.4 The probability of detection is specified together with the reported value of MDTD.
5. Significance and Use
5.1 This practice gives a measure of a thermal imaging system’s effectiveness for detecting a small spot within a large
background. Thus, it relates to the detection of small material defects such as voids, pits, cracks, inclusions, and occlusions.
5.2 MDTD values provide estimates of detection capability that may be used to compare one system with another. (Lower
MDTD values indicate better detection capability.)
5.3 Due to the partially subjective nature of the procedure, repeatability and reproducibility are apt to be poor and MDTD
differences less than 0.2°C0.2 °C are considered to be insignificant.
NOTE 2—Values obtained under idealized laboratory conditions may or may not correlate directly with service performance.
6. Apparatus
6.1 The apparatus consists of the following:
6.1.1 Target Plates, containing single or multiple circular targets of area(s) not greater than 5 % of the combined areas of target
and background (that is, FOV area), and with the distance from the center of the target to the center of the FOV equal to one third
of the height or the diameter of the FOV. See Fig. 1.
NOTE 3—A target plate may be fabricated by cutting one or more circular apertures in a metal plate of high thermal conductivity, such as aluminum,
and coating with black paint of emissivity greater than 0.95. In this case an aperture would constitute a target, and the coated metal surrounding the target
and within the field of view of the thermal imaging system would constitute the target’s background.
6.1.2 Facility, for mounting target plates and varying the orientation of any given target through 360°.
6.1.3 Differential Blackbody, controllable to within 0.1°C0.1 °C and stable over the procedure period to within 0.1°C.0.1 °C.
6.1.4 Infrared Spot Radiometer, calibrated with the aid of a blackbody source to an uncertainty not exceeding 0.1°C.0.1 °C.
7. Procedure
7.1 Mount a target plate and orient the target in correspondence with some integral hour marking on an imaginary clock. Do
not divulge the orientation to the observer.
NOTE 4—Only one observer at a time is to be present during the procedure.
FIG. 1 Schematic Showing 1. Target Plate; 2. FOV; and 3. Target
E1311 − 14 (2018)
7.2 Optimally focus the thermal imaging system directly on the target or on an optical projection of the target.
7.3 Adjust the thermal imaging system for quasi-linear operation.
7.4 Adjust the monochrome video monitor controls so that the presence of noise is barely perceivable by the observer.
7.5 Make the display luminance and the laboratory ambient lum
...