ASTM E1421-99(2021)
(Practice)Standard Practice for Describing and Measuring Performance of Fourier Transform Mid-Infrared (FT-MIR) Spectrometers: Level Zero and Level One Tests
Standard Practice for Describing and Measuring Performance of Fourier Transform Mid-Infrared (FT-MIR) Spectrometers: Level Zero and Level One Tests
SIGNIFICANCE AND USE
4.1 This practice permits an analyst to compare the general performance of an instrument on any given day with the prior performance of an instrument. This practice is not necessarily meant for comparison of different instruments with each other even if the instruments are of the same type and model. This practice is not meant for comparison of the performance of one instrument operated under differing conditions.
SCOPE
1.1 This practice describes two levels of tests to measure the performance of laboratory Fourier transform mid-infrared (FT-MIR) spectrometers equipped with a standard sample holder used for transmission measurements.
1.2 This practice is not directly applicable to Fourier transform infrared (FT-IR) spectrometers equipped with various specialized sampling accessories such as flow cells or reflectance optics, nor to Fourier transform near-infrared (FT-NIR) spectrometers, nor to FT-IR spectrometers run in step scan mode.
1.2.1 If the specialized sampling accessory can be removed and replaced with a standard transmission sample holder, then this practice can be used. However, the user should recognize that the performance measured may not reflect that which is achieved when the specialized accessory is in use.
1.2.2 If the specialized sampling accessory cannot be removed, then it may be possible to employ a modified version of this practice to measure spectrometer performance. The user is referred to Guide E1866 for a discussion of how these tests may be modified.
1.2.3 Spectrometer performance tests for FT-NIR spectrometers are described in Practice E1944.
1.2.4 Performance tests for dispersive MIR instruments are described in Practice E932.
1.2.5 For FT-IR spectrometers run in a step scan mode, variations on this practice and information provided by the instrument vendor should be used.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3.1 Exception—Informational inch-pound units are provided in 5.4.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1421 − 99 (Reapproved 2021)
Standard Practice for
Describing and Measuring Performance of Fourier
Transform Mid-Infrared (FT-MIR) Spectrometers: Level Zero
1
and Level One Tests
This standard is issued under the fixed designation E1421; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.4 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.1 Thispracticedescribestwolevelsofteststomeasurethe
responsibility of the user of this standard to establish appro-
performance of laboratory Fourier transform mid-infrared
priate safety, health, and environmental practices and deter-
(FT-MIR) spectrometers equipped with a standard sample
mine the applicability of regulatory limitations prior to use.
holder used for transmission measurements.
1.5 This international standard was developed in accor-
1.2 This practice is not directly applicable to Fourier trans-
dance with internationally recognized principles on standard-
form infrared (FT-IR) spectrometers equipped with various
ization established in the Decision on Principles for the
specialized sampling accessories such as flow cells or reflec-
Development of International Standards, Guides and Recom-
tance optics, nor to Fourier transform near-infrared (FT-NIR)
mendations issued by the World Trade Organization Technical
spectrometers, nor to FT-IR spectrometers run in step scan
Barriers to Trade (TBT) Committee.
mode.
1.2.1 If the specialized sampling accessory can be removed
2. Referenced Documents
and replaced with a standard transmission sample holder, then
2
2.1 ASTM Standards:
this practice can be used. However, the user should recognize
E131Terminology Relating to Molecular Spectroscopy
that the performance measured may not reflect that which is
E932PracticeforDescribingandMeasuringPerformanceof
achieved when the specialized accessory is in use.
Dispersive Infrared Spectrometers
1.2.2 If the specialized sampling accessory cannot be
E1866Guide for Establishing Spectrophotometer Perfor-
removed,thenitmaybepossibletoemployamodifiedversion
mance Tests
ofthispracticetomeasurespectrometerperformance.Theuser
E1944Practice for Describing and Measuring Performance
is referred to Guide E1866 for a discussion of how these tests
of Laboratory Fourier Transform Near-Infrared (FT-NIR)
may be modified.
Spectrometers: Level Zero and Level One Tests
1.2.3 SpectrometerperformancetestsforFT-NIRspectrom-
eters are described in Practice E1944.
3. Terminology
1.2.4 Performance tests for dispersive MIR instruments are
described in Practice E932. 3.1 Definitions—For definitions of terms used in this
practice, refer to Terminology E131. All identifications of
1.2.5 For FT-IR spectrometers run in a step scan mode,
variations on this practice and information provided by the spectral regions and absorption band positions are given in
−1
wavenumbers (cm ), and spectral energy, transmittance, and
instrument vendor should be used.
absorbancearesignifiedinequationsbytheletters E, T,and A,
1.3 The values stated in SI units are to be regarded as
respectively. The ratio of two transmittance or absorbance
standard. No other units of measurement are included in this
values, and the ratio of energy levels at two different wave-
standard.
numbers are signified by the letter R. A subscripted number
1.3.1 Exception—Informational inch-pound units are pro-
signifies a spectral position in wavenumbers (for example,
vided in 5.4.
−1
A , the absorbance at 3082 cm ).
3082
1
This practice is under the jurisdiction ofASTM Committee E13 on Molecular
Spectroscopy and Separation Science and is the direct responsibility of Subcom-
2
mittee E13.03 on Infrared and Near Infrared Spectroscopy. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Sept. 1, 2021. Published September 2021. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
ɛ1
approved in 1991. Last previous edition approved in 2015 as E1421–99(2015) . Standards volume information, refer to the standard’s Document Summary page on
DOI: 10.1520/E1421-99R21. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1421 − 99 (2021)
3.1.1 level one (1) test, n—a simple series of measurements affects apparent transmitt
...









Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.