ASTM E1026-13
(Practice)Standard Practice for Using the Fricke Dosimetry System
Standard Practice for Using the Fricke Dosimetry System
SIGNIFICANCE AND USE
4.1 The Fricke dosimetry system provides a reliable means for measurement of absorbed dose to water, based on a process of oxidation of ferrous ions to ferric ions in acidic aqueous solution by ionizing radiation (4). In situations not requiring traceability to national standards, this system can be used for absolute determination of absorbed dose without calibration, as the radiation chemical yield of ferric ions is well characterized (see Appendix X3).
4.2 The dosimeter is an air-saturated solution of ferrous sulfate or ferrous ammonium sulfate that indicates absorbed dose by an increase in optical absorbance at a specified wavelength. A temperature-controlled calibrated spectrophotometer is used to measure the absorbance (ICRU 80).
SCOPE
1.1 This practice covers the procedures for preparation, testing and using the acidic aqueous ferrous ammonium sulfate solution dosimetry system to measure absorbed dose to water when exposed to ionizing radiation. The system consists of a dosimeter and appropriate analytical instrumentation. The system will be referred to as the Fricke dosimetry system. The Fricke dosimetry system may be used as either a reference standard dosimetry system or a routine dosimetry system.
1.2 This practice is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of Practice E2628 for the Fricke dosimetry system. It is intended to be read in conjunction with Practice E2628.
1.3 The practice describes the spectrophotometric analysis procedures for the Fricke dosimetry system.
1.4 This practice applies only to gamma radiation, X-radiation (bremsstrahlung), and high-energy electrons.
1.5 This practice applies provided the following are satisfied:
1.5.1 The absorbed dose range shall be from 20 to 400 Gy (1).2
1.5.2 The absorbed-dose rate does not exceed 106 Gy·s−1 (2).
1.5.3 For radioisotope gamma sources, the initial photon energy is greater than 0.6 MeV. For X-radiation (bremsstrahlung), the initial energy of the electrons used to produce the photons is equal to or greater than 2 MeV. For electron beams, the initial electron energy is greater than 8 MeV. Note 1—The lower energy limits given are appropriate for a cylindrical dosimeter ampoule of 12 mm diameter. Corrections for displacement effects and dose gradient across the ampoule may be required for electron beams (3). The Fricke dosimetry system may be used at lower energies by employing thinner (in the beam direction) dosimeter containers (see ICRU Report 35).
1.5.4 The irradiation temperature of the dosimeter should be within the range of 10 to 60°C.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1026 − 13 AnAmerican National Standard
Standard Practice for
1
Using the Fricke Dosimetry System
This standard is issued under the fixed designation E1026; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
beams (3).TheFrickedosimetrysystemmaybeusedatlowerenergiesby
1. Scope
employingthinner(inthebeamdirection)dosimetercontainers(seeICRU
1.1 This practice covers the procedures for preparation,
Report 35).
testingandusingtheacidicaqueousferrousammoniumsulfate
1.5.4 Theirradiationtemperatureofthedosimetershouldbe
solution dosimetry system to measure absorbed dose to water
within the range of 10 to 60°C.
when exposed to ionizing radiation. The system consists of a
dosimeter and appropriate analytical instrumentation. The
1.6 This standard does not purport to address all of the
system will be referred to as the Fricke dosimetry system. The
safety concerns, if any, associated with its use. It is the
Fricke dosimetry system may be used as either a reference
responsibility of the user of this standard to establish appro-
standard dosimetry system or a routine dosimetry system.
priate safety and health practices and determine the applica-
1.2 This practice is one of a set of standards that provides
bility of regulatory limitations prior to use.
recommendations for properly implementing dosimetry in
radiation processing, and describes a means of achieving
2. Referenced Documents
compliance with the requirements of Practice E2628 for the
3
2.1 ASTM Standards:
Fricke dosimetry system. It is intended to be read in conjunc-
C912Practice for Designing a Process for Cleaning Techni-
tion with Practice E2628.
cal Glasses
1.3 The practice describes the spectrophotometric analysis
E170Terminology Relating to Radiation Measurements and
procedures for the Fricke dosimetry system.
Dosimetry
1.4 This practice applies only to gamma radiation,
E178Practice for Dealing With Outlying Observations
X-radiation (bremsstrahlung), and high-energy electrons.
E275PracticeforDescribingandMeasuringPerformanceof
1.5 This practice applies provided the following are satis- Ultraviolet and Visible Spectrophotometers
fied: E666Practice for CalculatingAbsorbed Dose From Gamma
1.5.1 The absorbed dose range shall be from 20 to 400 Gy
or X Radiation
2
(1).
E668 Practice for Application of Thermoluminescence-
6 −1
1.5.2 The absorbed-dose rate does not exceed 10 Gy·s
Dosimetry (TLD) Systems for Determining Absorbed
(2).
DoseinRadiation-HardnessTestingofElectronicDevices
1.5.3 For radioisotope gamma sources, the initial photon
E925Practice for Monitoring the Calibration of Ultraviolet-
energy is greater than 0.6 MeV. For X-radiation
Visible Spectrophotometers whose Spectral Bandwidth
(bremsstrahlung), the initial energy of the electrons used to
does not Exceed 2 nm
produce the photons is equal to or greater than 2 MeV. For
E958Practice for Estimation of the Spectral Bandwidth of
electron beams, the initial electron energy is greater than 8
Ultraviolet-Visible Spectrophotometers
MeV.
E2628Practice for Dosimetry in Radiation Processing
NOTE1—Thelowerenergylimitsgivenareappropriateforacylindrical 3
2.2 ISO/ASTM Standards:
dosimeter ampoule of 12 mm diameter. Corrections for displacement
ISO/ASTM 51261Practice for Calibration of Routine Do-
effects and dose gradient across the ampoule may be required for electron
simetry Systems for Radiation Processing
ISO/ASTM 51707Guide for Estimating Uncertainties in
Dosimetry for Radiation Processing
1
This practice is under the jurisdiction of ASTM Committee E61 on Radiation
Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry
Systems.
Current edition approved Jan. 1, 2013. Published March 2013. Originally
ε1
3
approved in 1984. Last previous edition approved in 2004 as E1026–04 . DOI: For referenced ASTM and ISO/ASTM standards, visit the ASTM webiste,
10.1520/E1026-13. www.astm.org, or contact ASTM Customer Service at service@astm.org. For
2
The boldface numbers that appear in parentheses refer to a list of references at Annual Book of ASTM Standards volume information, refer to the standard’s
the end of this practice. Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1026 − 13
2.3 ISO/IEC Standard: produced, destroyed, or changed by the mean energy, ε,
¯
ISO/IEC 17025General requirements for the competence of imparted to the matter.
4
testing and calibration laboratories
n x
~ !
G~x! 5 (2)
S D
2.4 Internat
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
´1
Designation: E1026 − 04 E1026 − 13 An American National Standard
Standard Practice for
1
Using the Fricke Reference-Standard Dosimetry System
This standard is issued under the fixed designation E1026; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1
ε NOTE—Equations 3 and 4 were corrected editorially in August 2005.
1. Scope
1.1 This practice covers the procedures for preparation, testing and using the acidic aqueous ferrous ammonium sulfate solution
dosimetry system to measure absorbed dose to water when exposed to ionizing radiation. The system consists of a dosimeter and
appropriate analytical instrumentation. The system will be referred to as the Fricke system. It is classified as a reference-standard
dosimetry system (see ISO/ASTM 51261).dosimetry system. The Fricke dosimetry system may be used as either a reference
standard dosimetry system or a routine dosimetry system.
1.2 This practice is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation
processing, and describes a means of achieving compliance with the requirements of Practice E2628 for the Fricke dosimetry
system. It is intended to be read in conjunction with Practice E2628.
1.3 The practice describes the spectrophotometric analysis procedures for the Fricke dosimeter.dosimetry system.
1.4 This practice applies only to gamma rays, x-raysradiation, X-radiation (bremsstrahlung), and high-energy electrons.
1.5 This practice applies provided the following are satisfied:
2
1.5.1 The absorbed dose range shall be from 20 to 400 Gy (1).
6 −1
1.5.2 The absorbed-dose rate does not exceed 10 Gy·s (2).
1.5.3 For radioisotope gamma-raygamma sources, the initial photon energy is greater than 0.6 MeV. For x-raysX-radiation
(bremsstrahlung), the initial energy of the electrons used to produce the photons is equal to or greater than 2 MeV. For electron
beams, the initial electron energy is greater than 8 MeV (see ICRU Reports 34 and 35). MeV.
NOTE 1—The lower energy limits given are appropriate for a cylindrical dosimeter ampoule of 12-mm outside12 mm diameter. Corrections for dose
gradients across an ampoule of that diameter or less are notdisplacement effects and dose gradient across the ampoule may be required for electron beams
required.(3). The Fricke dosimetry system may be used at lower energies by employing thinner (in the beam direction) dosimeter containers (see ICRU
Report 35).
1.5.4 The irradiation temperature of the dosimeter should be within the range of 10 to 60°C.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
3
2.1 ASTM Standards:
C912 Practice for Designing a Process for Cleaning Technical Glasses
D1193 Specification for Reagent Water
E170 Terminology Relating to Radiation Measurements and Dosimetry
E178 Practice for Dealing With Outlying Observations
E275 Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers
E666 Practice for Calculating Absorbed Dose From Gamma or X Radiation
1
This practice is under the jurisdiction of ASTM Committee E61 on Radiation Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry Systems.
Current edition approved Jan. 1, 2004Jan. 1, 2013. Published February 2004March 2013. Originally approved in 1984. Last previous edition approved in 20032004 as
ε1
E1026 – 03.E1026 – 04 . DOI: 10.1520/E1026-04E01.10.1520/E1026-13.
2
The boldface numbers that appear in parentheses refer to a list of references at the end of this practice.
3
For referenced ASTM and ISO/ASTM standards, visit the ASTM webiste, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book
of ASTM Standards volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1026 − 13
E668 Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining Absorbe
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.