Standard Test Method for Permanent Gases in C<inf>2</inf> and C<inf>3</inf> Hydrocarbon Products by Gas Chromatography and Pulse Discharge Helium Ionization Detector

SIGNIFICANCE AND USE
5.1 The presence of trace amounts of hydrogen, oxygen, carbon monoxide, and carbon dioxide can have deleterious effects in certain processes using hydrocarbon products as feed stock. This test method is suitable for setting specifications, for use as an internal quality control tool, and for use in development and research work.
SCOPE
1.1 This test method covers the determination of hydrogen, nitrogen, oxygen, methane, carbon monoxide, and carbon dioxide in the parts per billion mole (nmol/mol) to parts per million mole (µmol/mol) range in C2 and C3 hydrocarbons.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For some specific hazard statements, see Annex A1.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Historical
Publication Date
14-Jul-2017
Current Stage
Ref Project

Buy Standard

Standard
ASTM D8098-17 - Standard Test Method for Permanent Gases in C<inf>2</inf> and C<inf>3</inf> Hydrocarbon Products by Gas Chromatography and Pulse Discharge Helium Ionization Detector
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D8098 − 17
Standard Test Method for
Permanent Gases in C and C Hydrocarbon Products by
2 3
Gas Chromatography and Pulse Discharge Helium
1
Ionization Detector
This standard is issued under the fixed designation D8098; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope D7915 Practice for Application of Generalized Extreme
Studentized Deviate (GESD) Technique to Simultane-
1.1 This test method covers the determination of hydrogen,
ously Identify Multiple Outliers in a Data Set
nitrogen, oxygen, methane, carbon monoxide, and carbon
E260 Practice for Packed Column Gas Chromatography
dioxide in the parts per billion mole (nmol/mol) to parts per
F307 Practice for Sampling Pressurized Gas for Gas Analy-
million mole (µmol/mol) range in C and C hydrocarbons.
2 3
sis
3
1.2 The values stated in SI units are to be regarded as
2.2 Other Standards:
standard. No other units of measurement are included in this
G-4 and G-4.1, Compressed Gas Association Booklet on the
standard.
Use of Oxygen
1.3 This standard does not purport to address all of the
3. Terminology
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
3.1 For definitions of terms used in this standard, see
priate safety and health practices and determine the applica-
Terminology D4175.
bility of regulatory limitations prior to use. For some specific
4. Summary of Test Method
hazard statements, see Annex A1.
1.4 This international standard was developed in accor-
4.1 The sample is separated in a gas chromatographic
dance with internationally recognized principles on standard-
system using gas chromatography columns. Hydrogen,
ization established in the Decision on Principles for the
oxygen, nitrogen, methane, carbon monoxide, and carbon
Development of International Standards, Guides and Recom-
dioxide (also known as the permanent gases) are detected on a
mendations issued by the World Trade Organization Technical
pulse discharge detector. The concentration of the gases to be
Barriers to Trade (TBT) Committee.
determined is calculated from the peak areas relative to an
external standard. Helium is the required carrier gas for this
2. Referenced Documents
detector. Ultra-high purity carrier gases and leak-free gas
2
chromatography (GC) systems with getters are essential due to
2.1 ASTM Standards:
the extreme sensitivity of the detector. Argon, if present in the
D2505 Test Method for Ethylene, Other Hydrocarbons, and
sample, may interfere with oxygen determination.
Carbon Dioxide in High-Purity Ethylene by Gas Chroma-
tography
5. Significance and Use
D4175 Terminology Relating to Petroleum Products, Liquid
Fuels, and Lubricants 5.1 The presence of trace amounts of hydrogen, oxygen,
D6300 Practice for Determination of Precision and Bias carbon monoxide, and carbon dioxide can have deleterious
Data for Use in Test Methods for Petroleum Products and effects in certain processes using hydrocarbon products as feed
Lubricants stock. This test method is suitable for setting specifications, for
use as an internal quality control tool, and for use in develop-
ment and research work.
1
This test method is under the jurisdiction of ASTM Committee D02 on
6. Apparatus
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.D0.02 on Ethylene.
6.1 Chromatograph—Any gas chromatograph capable of
Current edition approved July 15, 2017. Published September 2017. DOI:
maintaining the temperatures, pressures, and flows necessary
10.1520/D8098-17.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3
Standards volume information, refer to this standard’s Document Summary page on Available from Compressed Gas Association (CGA), 14501 George Carter
the ASTM website. Way, Suite 103, Chantilly, VA 20151, http://www.cganet.com.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
D8098 − 17
for this analyses. The GC should also be capable of tempera- adsorptive with respect to the components in the sample. The
ture programming of the oven to obtain optimum separation of preferred material of construction is stainless steel. Copper,
the components. A typical configuration is shown in Fig. 1. brass, and other copper-bearing alloys are unacceptable. The
analysis of oxygen and carbon monoxide may benefit from the
6.2 Detectors—(PDHID Pu
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.