Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging

SIGNIFICANCE AND USE
5.1 This test method describes a test procedure for evaluating the ∆T associated with RF power deposition during an MR procedure, involving a specific frequency of RF irradiation of a passive implant. The method allows characterization of the heating propensity of an implant rather than the prediction of heating during a specific MR procedure in a patient. The results may be used as an input to a computational model for estimating ∆T due to the presence of that implant in a patient. The combination of the test results and the computational model results may then be used to help assess the safety of a patient with the implant during an MR examination.
SCOPE
1.1 This test method covers measurement of radio frequency (RF)-induced heating on or near a passive medical implant within a phantom during magnetic resonance imaging (MRI). The test method does not specify levels of heating considered to be safe to the patient and relies on users to define their own acceptance criteria.  
1.2 This test method does not address other possible safety issues which include, but are not limited to: issues of magnetically-induced displacement, magnetically-induced torque, image artifact, acoustic noise, tissue heating, interaction among devices, and the functionality of the device and the MR system.  
1.3 The amount of RF-induced temperature rise (∆T) for a given incident electric field will depend on the RF frequency, which is dependent on the static magnetic field strength of the MR system. While the focus in this test method is on 1.5 tesla (T) or 3 T MR systems, the ∆T for an implant in MR systems of other static magnetic field strengths or magnet designs can be evaluated by suitable modification of the method described herein.  
1.4 This test method assumes that testing is done on devices that will be entirely inside the body. Testing for devices with other implantation conditions (e.g., external fixation devices, percutaneous needles, catheters or tethered devices such as ablation probes) is beyond the scope of this standard; for such devices, modifications of this test method may be necessary.
Note 1: RF-heating induced by any electrically conductive implanted device may be impacted by the presence of other metallic or otherwise electrically conductive devices present nearby.  
1.5 This test method is written for several possible RF exposure systems, including Volume RF transmit coils. The exposure system needs to be properly characterized, within the stated uncertainties, in term of local background RF exposure for the implants which are tested.  
1.6 The values stated in SI units are to be regarded as standard.  
1.7 A device with deployed dimensions of less than 2 cm in all directions may not need to be tested with respect to RF-induced heating, as it is expected to generate ∆T of less than 2°C over 1 h of exposure at 1.5 T/64-MHz or 3 T/128-MHz frequencies (1, 2)2 and ANSI/AAMI/ISO 14708-3:2017). This condition is not valid when multiple replicas of the device (e.g., multiple anchors) are implanted within 3 cm of the device.
Note 2: The above values were derived from existing data and literature. The 3 cm distance is recommended to avoid any RF coupling with other neighboring devices.  
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Historical
Publication Date
14-Sep-2019
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM F2182-19e1 - Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging
English language
11 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM F2182-19e1 - Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging
English language
11 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
´1
Designation: F2182 − 19
Standard Test Method for
Measurement of Radio Frequency Induced Heating On or
Near Passive Implants During Magnetic Resonance
1
Imaging
This standard is issued under the fixed designation F2182; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1
ε NOTE—Editorially revised throughout in January 2020.
1. Scope exposuresystemneedstobeproperlycharacterized,withinthe
stated uncertainties, in term of local background RF exposure
1.1 This test method covers measurement of radio fre-
for the implants which are tested.
quency (RF)-induced heating on or near a passive medical
implant within a phantom during magnetic resonance imaging 1.6 The values stated in SI units are to be regarded as
(MRI). The test method does not specify levels of heating standard.
consideredtobesafetothepatientandreliesonuserstodefine
1.7 Adevice with deployed dimensions of less than 2 cm in
their own acceptance criteria.
all directions may not need to be tested with respect to
1.2 This test method does not address other possible safety RF-induced heating, as it is expected to generate �T of less
issues which include, but are not limited to: issues of than 2°C over1hof exposure at 1.5 T/64-MHz or 3
2
magnetically-induced displacement, magnetically-induced T/128-MHz frequencies (1, 2) and ANSI/AAMI/ISO 14708-
torque, image artifact, acoustic noise, tissue heating, interac- 3:2017). This condition is not valid when multiple replicas of
tionamongdevices,andthefunctionalityofthedeviceandthe thedevice(e.g.,multipleanchors)areimplantedwithin3cmof
MR system. the device.
1.3 The amount of RF-induced temperature rise (�T) for a
NOTE 2—The above values were derived from existing data and
literature. The 3 cm distance is recommended to avoid any RF coupling
given incident electric field will depend on the RF frequency,
with other neighboring devices.
which is dependent on the static magnetic field strength of the
1.8 This standard does not purport to address all of the
MR system. While the focus in this test method is on 1.5 tesla
safety concerns, if any, associated with its use. It is the
(T) or 3 T MR systems, the �T for an implant in MR systems
responsibility of the user of this standard to establish appro-
of other static magnetic field strengths or magnet designs can
priate safety, health, and environmental practices and deter-
be evaluated by suitable modification of the method described
mine the applicability of regulatory limitations prior to use.
herein.
1.9 This international standard was developed in accor-
1.4 Thistestmethodassumesthattestingisdoneondevices
dance with internationally recognized principles on standard-
that will be entirely inside the body. Testing for devices with
ization established in the Decision on Principles for the
other implantation conditions (e.g., external fixation devices,
Development of International Standards, Guides and Recom-
percutaneous needles, catheters or tethered devices such as
mendations issued by the World Trade Organization Technical
ablation probes) is beyond the scope of this standard; for such
Barriers to Trade (TBT) Committee.
devices, modifications of this test method may be necessary.
2. Referenced Documents
NOTE 1—RF-heating induced by any electrically conductive implanted
device may be impacted by the presence of other metallic or otherwise
3
2.1 ASTM Standards:
electrically conductive devices present nearby.
B348Specification for Titanium and Titanium Alloy Bars
1.5 This test method is written for several possible RF
and Billets
exposure systems, including Volume RF transmit coils. The
1 2
ThistestmethodisunderthejurisdictionofASTMCommitteeF04onMedical The boldface numbers in parentheses refer to a list of references at the end of
andSurgicalMaterialsandDevicesandisthedirectresponsibilityofSubcommittee this standard.
3
F04.15 on Material Test Methods. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Sept. 15, 2019. Published October 2019. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2002. Last previous edition approved in 2011 as F2182–11a. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/F2182-19E01. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
´1
F2182 − 19
F2052Test Method for Measurement of Magnetically In- Local background SAR can alternatively be derived from
duced Displacem
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
´1
Designation: F2182 − 19 F2182 − 19
Standard Test Method for
Measurement of Radio Frequency Induced Heating On or
Near Passive Implants During Magnetic Resonance
1
Imaging
This standard is issued under the fixed designation F2182; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1
ε NOTE—Editorially revised throughout in January 2020.
1. Scope
1.1 This test method covers measurement of radio frequency (RF)-induced heating on or near a passive medical implant within
a phantom during magnetic resonance imaging (MRI). The test method does not specify levels of heating considered to be safe
to the patient and relies on users to define their own acceptance criteria.
1.2 This test method does not address other possible safety issues which include, but are not limited toto: issues of magnetically
induced-displacement,magnetically-induced displacement, magnetically-induced torque, image artifact, acoustic noise, tissue
heating, interaction among devices, and the functionality of the device and the MR system.
1.3 The amount of RF-induced temperature rise (T) for a given incident electric field will depend on the RF frequency, which
is dependent on the static magnetic field strength of the MR system. While the focus in this test method is on 1.5 tesla (T) or 3
T MR systems, theT for an implant in MR systems of other static magnetic field strengths or magnet designs can be evaluated
by suitable modification of the method described herein.
1.4 This test method assumes that testing is done on devices that will be entirely inside the body. Testing for devices with other
implantation conditions (for example, (e.g., external fixation devices, percutaneous needles, catheters or tethered devices such as
ablation probes) is beyond the scope of this standard; for such devices, modifications of this test method may be necessary.
NOTE 1—RF-heating induced by any electrically conductive implanted device may be impacted by the presence of other metallic or otherwise
electrically conductive devices present nearby.
1.5 This test method is written for several possible RF exposure systems, including Volume RF transmit coils. The exposure
system needs to be properly characterized, within the stated uncertainties, in term of local background RF exposure for the implants
which are tested.
1.6 The values stated in SI units are to be regarded as standard.
1.7 A device with deployed dimensions of less than 2 cm in all directions doesmay not need to be tested with respect to
RF-induced heating, as it is expected to generateT of less than 2°C over 1 hourh of exposure at 1.5 T andT/64-MHz or 3
2
TT/128-MHz frequencies ((1, 2) and ANSI/AAMI/ISO 14708-3:2017). This condition is not valid when multiple replicas of the
device (for example, (e.g., multiple anchors) are implanted within 3 cm of the device.
NOTE 2—The above values were derived from existing data and literature. The 3 cm distance is recommended to avoid any RF coupling with other
neighboring devices.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.9 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1
This test method is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.15 on Material Test Methods.
Current edition approved Sept. 15, 2019. Published October 2019. Originally approved in 2002. Last previous edition approved in 2011 as F2182 – 11a. DOI:
10.1520/F2182-19.10.1520/F2182-19E01.
2
The boldface numbers in parentheses refer to a list of references at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
´1
F2182 − 19
2. Referenced Documents
3
2.1 ASTM
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.