FprEN IEC 61846:2024
(Main)Ultrasonics - Therapeutic focused short pressure pulse sources - Characteristics of fields
Ultrasonics - Therapeutic focused short pressure pulse sources - Characteristics of fields
Ultraschall - Therapeutisch fokussierte Kurzdruckimpulsquellen – Feldeigenschaften
Ultrasons - Sources d’impulsions de pression courtes focalisées thérapeutiques - Caractéristiques des champs
Ultrazvok - Terapevtsko usmerjeni viri kratkih tlačnih impulzov - Karakteristike polj
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
oSIST prEN IEC 61846:2023
01-september-2023
Ultrazvok - Terapevtsko usmerjeni viri kratkih tlačnih impulzov - Karakteristike polj
Ultrasonics - Therapeutic focused short pressure pulse sources - Characteristics of fields
Ultraschall - Druckpuls-Lithotripter - Feldcharakterisierung
Ultrasons - Lithotripteurs à ondes de pression - Caractérisation des champs
Ta slovenski standard je istoveten z: prEN IEC 61846:2023
ICS:
11.040.50 Radiografska oprema Radiographic equipment
17.140.50 Elektroakustika Electroacoustics
oSIST prEN IEC 61846:2023 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
oSIST prEN IEC 61846:2023
oSIST prEN IEC 61846:2023
87/836/CDV
COMMITTEE DRAFT FOR VOTE (CDV)
PROJECT NUMBER:
IEC 61846 ED2
DATE OF CIRCULATION: CLOSING DATE FOR VOTING:
2023-07-14 2023-10-06
SUPERSEDES DOCUMENTS:
87/806/CD, 87/821A/CC
IEC TC 87 : ULTRASONICS
SECRETARIAT: SECRETARY:
United Kingdom Mr Petar Luzajic
OF INTEREST TO THE FOLLOWING COMMITTEES: PROPOSED HORIZONTAL STANDARD:
Other TC/SCs are requested to indicate their interest, if any, in this
CDV to the secretary.
FUNCTIONS CONCERNED:
EMC ENVIRONMENT QUALITY ASSURANCE SAFETY
SUBMITTED FOR CENELEC PARALLEL VOTING NOT SUBMITTED FOR CENELEC PARALLEL VOTING
Attention IEC-CENELEC parallel voting
The attention of IEC National Committees, members of
CENELEC, is drawn to the fact that this Committee Draft
for Vote (CDV) is submitted for parallel voting.
The CENELEC members are invited to vote through the
CENELEC online voting system.
This document is still under study and subject to change. It should not be used for reference purposes.
Recipients of this document are invited to submit, with their comments, notification of any relevant patent rights of which they
are aware and to provide supporting documentation.
Recipients of this document are invited to submit, with their comments, notification of any relevant “In So me Countries” clauses
to be included should this proposal proceed. Recipients are reminded that the CDV stage is the final stage for submitting ISC
clauses. (SEE AC/22/2007 OR NEW GUIDANCE DOC).
TITLE:
Ultrasonics – Therapeutic focused short pressure pulse sources – Characteristics of fields
PROPOSED STABILITY DATE: 2027
NOTE FROM TC/SC OFFICERS:
electronic file, to make a copy and to print out the content for the sole purpose of preparing National Committee positions.
You may not copy or "mirror" the file or printed version of the document, or any part of it, for any other purpose without
permission in writing from IEC.
oSIST prEN IEC 61846:2023
oSIST prEN IEC 61846:2023
87/836/CDV – 2 – IEC CDV 61846 © IEC:2023
1 CONTENTS
3 FOREWORD . 4
4 INTRODUCTION . 6
5 1 Scope . 7
6 2 Normative references . 7
7 3 Terms and definitions . 8
8 3.1 8
9 4 List of symbols . 16
10 5 Conditions of measurement . 17
11 6 Test equipment . 17
12 6.1 Test chamber . 17
13 6.2 Hydrophone . 18
14 6.2.1 Hydrophone for pressure pulse measurements . 18
15 6.2.2 Hydrophones for quality assurance . 18
16 6.3 Voltage measurement . 19
17 6.3.1 Oscilloscope or transient recorder . 19
18 6.3.2 Pressure-pulse-waveform recording . 19
19 7 Measurement procedure . 19
20 7.1 Spatial measurements . 20
21 7.1.1 Beam plots of peak-compressional acoustic pressure . 20
22 7.1.2 Beam plots of peak-rarefactional acoustic pressure . 20
23 7.1.3 Focus position . 20
24 7.1.4 Focal −n dB width . 21
25 7.1.5 Focal −n dB extent . 21
26 7.1.6 Focal −n dB cross sectional area . 21
27 7.1.7 Focal −n dB volume . 21
28 7.2 Temporal measurements . 21
29 7.3 Acoustic energy measurements . 21
30 7.3.1 Pulse-pressure-squared integral . 21
31 7.3.2 Derived pulse-intensity integral . 21
32 7.3.3 Derived focal acoustic pulse energy. 22
33 7.3.4 Derived acoustic pulse energy . 22
34 7.3.5 Average positive pressure . 22
35 7.3.6 Momentum . 22
36 Annex A (informative) Short Pressure Pulse lithotripsy . 24
37 A.1 Background . 24
38 A.2 Percutaneous continuous wave systems and alternative burst wave
39 applications . 24
40 A.3 Exclusions . 24
41 A.4 Extracorporeally induced lithotripsy . 24
42 Annex B (informative) Types of pressure wave transducers . 26
43 B.1 Introduction . 26
44 B.1.1 Spark discharge. 26
45 B.1.2 Piezoelectric . 26
46 B.1.3 Electromagnetic . 26
47 B.2 Positioning systems . 27
oSIST prEN IEC 61846:2023
IEC CDV 61846 © IEC:2023 – 3 – 87/836/CDV
48 Annex C (informative) Field measurement . 28
49 C.1 Measurement probes and hydrophones . 28
50 C.2 Test chamber . 30
51 C.3 Degassing procedures . 30
52 C.4 Acoustic pulse energy . 30
53 Annex D (informative) Bibliography . 34
oSIST prEN IEC 61846:2023
87/836/CDV – 4 – IEC CDV 61846 © IEC:2023
58 INTERNATIONAL ELECTROTECHNICAL COMMISSION
59 ____________
61 ULTRASONICS – THERAPEUTIC FOCUSED SHORT PRESSUREPULSE
62 SOURCES – CHARACTERISTICS OF FIELDS
66 FOREWORD
67 1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
68 all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
69 co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
70 in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
71 Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their
72 preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
73 may participate in this preparatory work. International, governmental and non-governmental organizations liaising
74 with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
75 Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
76 2) The formal decisions or agreements of IEC on technical matters express, as nearly as possibl e, an international
77 consensus of opinion on the relevant subjects since each technical committee has representation from all
78 interested IEC National Committees.
79 3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
80 Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
81 Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
82 misinterpretation by any end user.
83 4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
84 transparently to the maximum extent possible in their national and regional publications. Any divergence between
85 any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
86 5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
87 assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
88 services carried out by independent certification bodies.
89 6) All users should ensure that they have the latest edition of this publication.
90 7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
91 members of its technical committees and IEC National Committees for any personal injury, property damage or
92 other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
93 expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
94 Publications.
95 8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
96 indispensable for the correct application of this publication.
97 9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of patent
98 rights. IEC shall not be held responsible for identifying any or all such patent ri ghts.
99 International Standard IEC 61846 has been prepared by IEC technical committee 87:
100 Ultrasonics.
101 This second edition cancels and replaces the first edition published in [1998], This edition
102 constitutes a technical revision.
103 This edition includes the following significant technical changes with respect to the previous
104 edition:
105 a) Change of title: Old standard: PRESSURE PULSE LITHOTRIPTERS is changed to
106 “THERAPEUTIC FOCUSED SHORT PRESSURE-PULSE SOURCES” in order to take into
107 account the development in the relevant technical and biomedical applications of such
108 sources, which were originally used only for (kidney) lithotripsy, while recent applications
109 include a wide range for the treatment of e.g. stone diseases, orthopaedic pain, tissue,
110 cardiac and brain diseases.
111 The term “focused” was added to differentiate this standard from IEC 63045 “Non -focusing
112 short pressure pulse sources including ballistic pressure pulse sources” [26].
oSIST prEN IEC 61846:2023
IEC CDV 61846 © IEC:2023 – 5 – 87/836/CDV
113 The term “short” was added to align the nomenclature to IEC 63045 and differenti ate this
114 standard from standards in the HIFU / HITU fields.
115 b) Scope and elsewhere in the document: The term “lithotripsy” is changed to “therapy” in
116 order to account for the wide range of applications beyond stone diseases.
117 c) 3. Definitions: The “-6 dB” parameter definitions are replaced by “-n dB” to avoid
118 misconceptions in the significance and use of these parameters and to account for newer
119 findings in literature.
120 Additional “n MPa” parameters are introduced for the same reasons.
121 The definitions of “derived” parameters was aligned to those in recently published
122 standards, e.g. IEC 62127-1.
123 New definitions were added which describe parameters appearing in newer relevant
124 literature, e.g. “momentum”, “average positive acoustic pressure”, “ca
...