Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications

ISO 80369-6:2016 specifies requirements for small-bore connectors intended to be used for connections in neuraxial applications. Neuraxial applications involve the use of medical devices intended to administer medications to neuraxial sites, wound infiltration anaesthesia delivery, and other regional anaesthesia procedures or to monitor or remove cerebro-spinal fluid for therapeutic or diagnostic purposes.
NOTE 1 Sites for the neuraxial application include the spine, intrathecal or subarachnoid space, ventricles of the brain, and the epi-, extra-, or peri-dural space. Neuraxial application anaesthetics can be administered regionally affecting a large part of the body, such as a limb, and include plexus blocks, such as the branchial plexus blocks or single nerve blocks. Neuraxial application procedures include continuous infusion of wounds with local anaesthetic agents.
NOTE 2 For the purposes of this part of ISO 80369, local anaesthesia injected hypodermically is not considered a neuraxial application.
EXAMPLES Intended administration includes intrathecal chemotherapy, local anaesthetics, radiological contrast agents, antibiotics, analgesics. This part of ISO 80369 specifies dimensions and requirements for the design and functional performance of these small-bore connectors intended to be used with medical devices. This part of ISO 80369 does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in particular International Standards for specific medical devices or accessories.
NOTE 3 Manufacturers are encouraged to incorporate the small-bore connectors specified in this part of ISO 80369 into medical devices, medical systems, or accessories, even if currently not required by the relevant particular medical device standards. It is expected that when the relevant particular medical device standards are revised, requirements for small-bore connectors, as specified in this part of ISO 80369, will be included. Furthermore, it is recognized that standards need to be developed for many medical devices used for neuraxial applications.
NOTE 4 ISO 80369-1:2010, 5.8, specifies alternative methods of compliance with ISO 80369-1:2010, for small-bore connectors intended for use with neuraxial application medical devices or accessories, which do not comply with this part of ISO 80369.

Raccords de petite taille pour liquides et gaz utilisés dans les domaine de la santé - Partie 6: Raccords destinés à des applications en contact avec le système nerveux (neuraxiales)

General Information

Status
Published
Publication Date
09-Feb-2016
Current Stage
PPUB - Publication issued
Start Date
31-Mar-2016
Completion Date
14-Mar-2016
Ref Project

Buy Standard

Standard
ISO 80369-6:2016 - Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications
English language
45 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 80369-6
First edition
2016-03-15
Small bore connectors for liquids and
gases in healthcare applications —
Part 6:
Connectors for neuraxial applications
Raccords destinés à des applications en contact avec le système
nerveux (neuraxiales) —
Partie 6: Raccords destinés à des applications en contact avec le
système nerveux (neuraxiales)
Reference number
ISO 80369-6:2016(E)
©
ISO 2016

---------------------- Page: 1 ----------------------
ISO 80369-6:2016(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2016, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2016 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 80369-6:2016(E)

Contents Page
Foreword .iv
Introduction .v
1 * Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 General requirements . 3
4.1 General requirements for the neuraxial application . 3
4.2 * Material used for small-bore connectors . 4
4.3 Type tes ts . 4
5 Dimensional requirements for neuraxial small-bore connectors.4
6 Performance requirements . 4
6.1 Fluid leakage . 4
6.1.1 Fluid leakage requirement . 4
6.1.2 Leakage by pressure decay . 4
6.1.3 Positive pressure liquid leakage . 4
6.2 Subatmospheric pressure air leakage . 5
6.3 Stress cracking . 5
6.4 Resistance to separation from axial load. 5
6.5 Resistance to separation from unscrewing . 5
6.6 Resistance to overriding .
...