Infusion equipment for medical use — Part 10: Accessories for fluid lines for use with pressure infusion equipment

ISO 8536-10:2004 applies to sterilized accessories for single use in fluid lines and pressure infusion equipment as specified in ISO 8536-8. ISO 8536-10:2004 includes: Two-way stopcocks (2SC), three-way stopcocks (3SC), four-way stopcocks (4SC) and stopcocks manifold (SM). Designation of a stopcock depends on the number of connections. The number of possible functional positions can be expressed by addition of a complementary note, using a diagonal stroke and a numeral indicating the number of possible stopcock positions, e.g. 3/4-way stopcock for three-way stopcock with four possible positions. Units with injection site (UIS) or check valve (UCV). Stoppers (S) or adapters (A).

Matériel de perfusion à usage médical — Partie 10: Accessoires de tubulures pour utilisation avec des appareils de perfusion sous pression

L'ISO 8536-10:2004 s'applique aux accessoires à usage unique stériles des tubulures et appareils de perfusion sous pression spécifiés dans l'ISO 8536-8. L'ISO 8536-10:2004 traite des éléments suivants: les robinets à deux (2SC), trois (3SC), quatre (4SC) voies et les rampes de robinets (SM). La désignation d'un robinet dépend du nombre de raccordements. Le nombre de positions possibles de fonctionnement peut être exprimé en ajoutant une note complémentaire, à l'aide d'une barre oblique et d'un chiffre indiquant le nombre de positions possibles pour le robinet, par exemple, robinet à 3/4 voies, pour un robinet à trois voies comportant quatre positions possibles. les unités avec site d'injection (UIS) ou avec valve anti-retour (UCV); les bouchons (S) ou les adaptateurs (A).

General Information

Status
Withdrawn
Publication Date
11-Oct-2004
Withdrawal Date
11-Oct-2004
Current Stage
9599 - Withdrawal of International Standard
Completion Date
10-Jun-2015
Ref Project

Relations

Buy Standard

Standard
ISO 8536-10:2004 - Infusion equipment for medical use
English language
10 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 8536-10:2004 - Matériel de perfusion a usage médical
French language
10 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 8536-10
First edition
2004-10-01
Infusion equipment for medical use —
Part 10:
Accessories for fluid lines for use with
pressure infusion equipment
Matériel de perfusion à usage médical —
Partie 10: Accessoires de tubulures pour utilisation avec des appareils
de perfusion sous pression

Reference number
ISO 8536-10:2004(E)
©
ISO 2004

---------------------- Page: 1 ----------------------
ISO 8536-10:2004(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but shall
not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the
unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2004
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56  CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
©
ii ISO 2004 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 8536-10:2004(E)
Contents Page
1 Scope . 1
2 Normative references . 1
3 Designation . 2
4 Materials . 2
5 Physical requirements . 2
5.1 Avoidance of air bubbles . 2
5.2 Particulate contamination . 3
5.3 Tensile strength . 3
5.4 Leakage . 3
5.5 Adapters with female and/or male conical fittings . 3
5.6 Protective caps . 3
5.7 Manipulation of stopcocks . 3
5.8 Unit with injection site . 3
5.9 Unit with check valve . 3
6 Chemical requirements . 3
7 Biological requirements . 4
7.1 Sterility . 4
7.2 Pyrogens . 4
7.3 Haemolysis . 4
8 Packaging . 4
9 Labelling . 4
9.1 Unit container . 4
9.2 Shelf or multi-unit container . 5
Annex A (normative) Physical tests . 6
Annex B (normative) Chemical tests . 8
Annex C (normative) Biological tests . 9
Bibliography . 10
©
ISO 2004 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 8536-10:2004(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International
Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 8536-10 was prepared by Technical Committee ISO/TC 76, Transfusion, infusion and injection equipment
for medical and pharmaceutical use.
ISO 8536 consists of the following parts, under the general title Infusion equipment for medical use:
— Part 1: Infusion glass bottles
— Part 2: Closures for infusion bottles
— Part 3: Aluminium caps for infusion bottles
— Part 4: Infusion sets for single use, gravity feed
— Part 5: Burette infusion sets for single use, gravity feed
— Part 6: Freeze drying closures for infusion bottles
— Part 7: Caps made of aluminium-plastics combinations for infusion bottles
— Part 8: Infusion equipment for use with pressure infusion apparatus
— Part 9: Fluid lines for use with pressure infusion equipment
— Part 10: Accessories for fluid lines for use with pressure infusion equipment
— Part 11: Infusion filters for use with pressure infusion equipment
©
iv ISO 2004 – All rights reserved

---------------------- Page: 4 ----------------------
INTERNATIONAL STANDARD ISO 8536-10:2004(E)
Infusion equipment for medical use —
Part 10:
Accessories for fluid lines for use with pressure infusion
equipment
1Scope
This part of ISO 8536 applies to sterilized accessories for single use in fluid lines and pressure infusion
equipment as specified in ISO 8536-8.
This part of ISO 8536 includes:
a) Two-way stopcocks (2SC), three-way stopcocks (3SC), four-way stopcocks (4SC) and stopcocks manifold
(SM);
NOTE Designation of a stopcock depends on the number of connections. The number of possible functional positions
can be expressed by addition of a complementary note, using a diagonal stroke and a numeral indicating the number of
possible stopcock positions, e.g. 3/4-way stopcock for three-way stopcock with four possible positions.
b) units with injection site (UIS) or check valve (UCV);
c) stoppers (S) or adapters (A).
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced document
(including any amendments) applies.
ISO 594-2, Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment —
Part 2: Lock fittings
ISO 7886-1, Sterile hypodermic syringes for single use — Part 1: Syringes for manual use
ISO 8536-4, Infusion equipment for medical use — Part 4: Infusion sets for single use, gravity feed
ISO 8536-8, Infusion equipment for medical use — Part 8: Infusion equipment for use with pressure infusion
apparatus
ISO 15223, Medical devices — Symbols to be used with medical device labels, labelling and information to be
supplied
©
ISO 2004 – All rights reserved 1

---------------------- Page: 5 ----------------------
ISO 8536-10:2004(E)
3 Designation
Designation of two-way stopcock (2SC) for infusions under pressure (P):
Two-way stopcock ISO 8536-10 – 2SC –P
Designation of three-way stopcock (3SC) for infusions under pressure (P):
Three-way stopcock ISO 8536-10 – 3SC –P
Designation of four-way stopcock (4SC) for infusions under pressure (P):
Four-way stopcock ISO 8536-10 – 4SC –P
Designation of stopcock manifold (SM) for infusions under pressure (P):
Stopcock manifold ISO 8536-10 – SM –P
Designation of two-stopcock manifold (2SM) for infusions under pressure (P):
Stopcock manifold ISO 8536-10 – 2SM –P
Designation of three-stopcock manifold (3SM) for infusions under pressure (P):
Stopcock manifold ISO 8536-10 – 3SM –P
Designation of four-stopcock manifold (4SM) for infusions under pressure (P):
Stopcock manifold ISO 8536-10 – 4SM –P
Designation of unit with injection site (UIS) for infusions under pressure (P):
Injection unit ISO 8536-10 – UIS –P
Designat
...

NORME ISO
INTERNATIONALE 8536-10
Première édition
2004-10-01
Matériel de perfusion à usage médical —
Partie 10:
Accessoires de tubulures pour utilisation
avec des appareils de perfusion sous
pression
Infusion equipment for medical use —
Part 10: Accessories for fluid lines for use with pressure infusion
equipment

Numéro de référence
ISO 8536-10:2004(F)
©
ISO 2004

---------------------- Page: 1 ----------------------
ISO 8536-10:2004(F)
PDF — Exonération de responsabilité
Le présent fichier PDF peut contenir des polices de caractères intégrées. Conformément aux conditions de licence d'Adobe, ce fichier
peut être imprimé ou visualisé, mais ne doit pas être modifié à moins que l'ordinateur employé à cet effet ne bénéficie d'une licence
autorisant l'utilisation de ces polices et que celles-ci y soient installées. Lors du téléchargement de ce fichier, les parties concernées
acceptent de fait la responsabilité de ne pas enfreindre les conditions de licence d'Adobe. Le Secrétariat central de l'ISO décline toute
responsabilité en la matière.
Adobe est une marque déposée d'Adobe Systems Incorporated.
Les détails relatifs aux produits logiciels utilisés pour la création du présent fichier PDF sont disponibles dans la rubrique General Info du
fichier; les paramètres de création PDF ont été optimisés pour l'impression. Toutes les mesures ont été prises pour garantir l'exploitation
de ce fichier par les comités membres de l'ISO. Dans le cas peu probable où surviendrait un problème d'utilisation, veuillez en informer
le Secrétariat central à l'adresse donnée ci-dessous.
© ISO 2004
Droits de reproduction réservés. Sauf prescription différente, aucune partie de cette publication ne peut être reproduite ni utilisée sous
quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les microfilms, sans l'accord écrit
de l'ISO à l'adresse ci-après ou du comité membre de l'ISO dans le pays du demandeur.
ISO copyright office
Case postale 56  CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse
©
ii ISO 2004 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 8536-10:2004(F)
Sommaire Page
1 Domaine d'application . 1
2Références normatives . 1
3Désignation . 2
4Matériaux . 2
5 Exigences physiques . 2
5.1 Absence de bulles d'air . 2
5.2 Contamination particulaire . 2
5.3 Résistance à la rupture . 3
5.4 Étanchéité . 3
5.5 Adaptateurs à raccords coniques mâle et/ou femelle . 3
5.6 Protecteurs . 3
5.7 Manipulation des robinets . 3
5.8 Unité équipée d'un site d'injection . 3
5.9 Unité à valve anti-retour . 3
6 Exigences chimiques . 3
7 Exigences biologiques . 3
7.1 Stérilité . 3
7.2 Pyrogènes . 4
7.3 Hémolyse . 4
8 Emballage . 4
9 Étiquetage . 4
9.1 Emballage primaire . 4
9.2 Emballage secondaire . 4
Annexe A (normative) Essais physiques . 6
Annexe B (normative) Essais chimiques . 8
Annexe C (normative) Essais biologiques . 9
Bibliographie . 10
©
ISO 2004 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 8536-10:2004(F)
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux de
normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général confiée
aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire partie du
comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec la
Commission électrotechnique internationale (CEI) en ce qui concerne la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives ISO/CEI,
Partie 2.
La tâche principale des comités techniques est d'élaborer les Normes internationales. Les projets de Normes
internationales adoptés par les comités techniques sont soumis aux comités membres pour vote. Leur
publication comme Normes internationales requiert l'approbation de 75 % au moins des comités membres
votants.
L'attention est appelée sur le fait que certains des éléments du présent document peuvent faire l'objet de droits
de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de ne pas avoir
identifié de tels droits de propriété et averti de leur existence.
L'ISO 8536-10 a été élaborée par le comité technique ISO/TC 76, Appareils de transfusion, de perfusion et
d'injection à usage médical et pharmaceutique.
L'ISO 8536 comprend les parties suivantes, présentées sous le titre général Matériel de perfusion à usage
médical:
— Partie 1: Flacons en verre pour perfusion
— Partie 2: Bouchons pour flacons de perfusion
— Partie 3: Capsules en aluminium pour flacons de perfusion
— Partie 4: Appareils de perfusion non réutilisables, à alimentation par gravité
— Partie 5: Appareils non réutilisables de perfusion à burette, à alimentation par gravité
— Partie 6: Bouchons à lyophilisation pour flacons de perfusion
— Partie 7: Capsules en combinaison aluminium-plastique pour flacons de perfusion
— Partie 8: Matériel de perfusion pour utilisation avec des appareils de perfusion sous pression
— Partie 9: Tubulures pour utilisation avec des appareils de perfusion sous pression
— Partie 10: Accessoires de tubulures pour utilisation avec des appareils de perfusion sous pression
— Partie 11: Filtres de perfusion pour utilisation avec des appareils de perfusion sous pression
©
iv ISO 2004 – Tous droits réservés

---------------------- Page: 4 ----------------------
NORME INTERNATIONALE ISO 8536-10:2004(F)
Matériel de perfusion à usage médical —
Partie 10:
Accessoires de tubulures pour utilisation avec des appareils de
perfusion sous pression
1 Domaine d'application
La présente partie de l'ISO 8536 s'applique aux accessoires à usage unique stériles des tubulures et appareils
de perfusion sous pression spécifiés dans l'ISO 8536-8.
La présente partie de l'ISO 8536 traite des éléments suivants:
a) les robinets à deux (2SC), trois (3SC), quatre (4SC) voies et les rampes de robinets (SM);
NOTE La désignation d'un robinet dépend du nombre de raccordements. Le nombre de positions possibles de
fonctionnement peut être exprimé en ajoutant une note complémentaire, à l'aide d'une barre oblique et d'un chiffre
indiquant le nombre de positions possibles pour le robinet, par exemple, robinet à 3/4 voies, pour un robinet à trois voies
comportant quatre positions possibles.
b) les unités avec site d'injection (UIS) ou avec valve anti-retour (UCV);
c) les bouchons (S) ou les adaptateurs (A).
2Références normatives
Les documents de référence suivants sont indispensables pour l'application du présent document. Pour les
références datées, seule l'édition citée s'applique. Pour les références non datées, la dernière édition du
document de référence s'applique (y compris les éventuels amendements).
ISO 594-2, Assemblages coniques à 6 % (Luer) des seringues et aiguilles et de certains autres appareils à
usage médical — Partie 2: Assemblages à verrouillage
ISO 7886-1, Seringues hypodermiques stériles, non réutilisables — Partie1: Seringues pour utilisation
manuelle
ISO 8536-4, Matériel de perfusion à usage médical — Partie 4: Appareils de perfusion non réutilisables, à
alimentation par gravité
ISO 8536-8, Matériel de perfusion à usage médical — Partie 8: Matériel de perfusion pour utilisation avec des
appareils de perfusion sous pression
ISO 15223, Dispositifs médicaux — Symboles à utiliser avec les étiquettes, l'étiquetage et les informations à
fournir relatifs aux dispositifs médicaux
©
ISO 2004 – Tous droits réservés 1

---------------------- Page: 5 ----------------------
ISO 8536-10:2004(F)
3Désignation
Désignation d'un robinet à deux voies (2SC) pour perfusions sous pression (P):
Robinet à 2 voies ISO 8536-10 – 2SC – P
Désignation d'un robinet à trois voies (3SC) pour perfusions sous pression (P):
Robinet à 3 voies ISO 8536-10 – 3SC – P
Désignation d'un robinet à quatre voies (4SC) pour perfusions sous pression (P):
Robinet à 4 voies ISO 8536-10 – 4SC – P
Désignation d'une rampe de robinets (SM) pour perfusions sous pression (P):
Rampe de robinets ISO 8536-10 – SM – P
Désignation d'une rampe à deux robinets (2SM) pour perfusions sous pression (P):
Rampe de robinets ISO 8536-10 – 2SM – P
Désignation d'une rampe à trois robinets (3SM) pour perfusions sous pression (P):
Rampe de robinets ISO 8536-10 – 3SM – P
Désignation d'une rampe à quatre robinets (4SM) pour perfusions sous pression (P):
Rampe de robinets ISO 8536-10 – 4SM – P
Désignation d'une unité équipée d'un site d'injection (UIS) pour perfusions sous pression (P):
Unité d’injection ISO 8536-10 – UIS – P
Désignation d'une unité avec valve anti-retour (UCV) pour perfusions sous pression (P):
Unité d’injection ISO 85
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.