Methods for the calibration of vibration and shock transducers - Part 21: Vibration calibration by comparison to a reference transducer

ISO 16063-21:2003 describes the calibration of rectilinear vibration transducers by comparison. Although it mainly describes calibration using direct comparison to a standard calibrated by primary methods, the methods described can be applied between other levels in the calibration hierarchy. ISO 16063-21:2003 specifies procedures for performing calibrations of rectilinear vibration transducers by comparison in the frequency range from 0,4 Hz to 10 kHz. It is primarily intended for those who are required to meet ISO standardized methods for the measurement of vibration under laboratory conditions, where the uncertainty of measurement is relatively small. It can also be used under field conditions, where the uncertainty of measurement may be relatively large. From knowledge of all significant sources of uncertainty affecting the calibration, the expanded uncertainty can be evaluated using the methods given in ISO 16063-21. It also covers the assessment of uncertainties for calibrations performed using a check standard. This document is applicable to calibrations outside the frequency range from 0,4 Hz to 10 kHz, considering the additional facts described in 4.3. Systems have been demonstrated to cover the range from 0,01 Hz to 30 kHz.

Méthodes pour l'étalonnage des transducteurs de vibrations et de chocs — Partie 21: Étalonnage de vibrations par comparaison à un transducteur de référence

General Information

Status
Published
Publication Date
18-Aug-2003
Current Stage
9093 - International Standard confirmed
Start Date
23-Jan-2025
Completion Date
13-Dec-2025
Ref Project

Relations

Overview - ISO 16063-21:2003 (Vibration calibration by comparison to a reference transducer)

ISO 16063-21:2003 specifies methods for the calibration of rectilinear vibration transducers by comparison to a reference (secondary) transducer. The standard is intended mainly for laboratory calibrations but is also applicable in the field. It covers procedures to determine the magnitude and, optionally, the phase shift of the complex sensitivity of vibration transducers in the nominal frequency range 0.4 Hz to 10 kHz (systems have been demonstrated from 0.01 Hz to 30 kHz). The document emphasizes traceability, documented measurement uncertainty, and practical calibration arrangements (back-to-back or in-line mounting).

Key topics and technical requirements

  • Scope and transducer type: Applies to rectilinear vibration transducers (with or without signal conditioning).
  • Calibration method: Direct comparison to a reference transducer; methods may be applied between other levels in the calibration hierarchy (primary → reference → working/check standards).
  • Frequency & dynamic ranges: Nominal 0.4 Hz–10 kHz (with notes on optional ranges and single-point calibrations); frequency-dependent dynamic ranges and practical limits (noise floor, exciter distortion).
  • Uncertainty of measurement: Mandatory uncertainty budgets per Annex A; examples provided for two system classes (Example 1 - well-controlled lab; Example 2 - less stringent conditions). Representative expanded uncertainties (coverage factor k=2) from the standard:
    • Accelerometers 0.4 Hz–1 kHz: Example 1: 1%, Example 2: 3% (magnitude)
    • Accelerometers 1 kHz–2 kHz: 2% / 5%
    • Accelerometers 2 kHz–10 kHz: 3% / 10%
    • Displacement/velocity transducers 20 Hz–1 kHz: 4% / 6%
    • Phase shift at reference conditions: 1° / 3° (Example 1 / 2)
  • Apparatus & environment: Reference transducer preferably calibrated with amplifier; recommended room temperature and humidity tolerances (Example 1: 23 ±5 °C, humidity ≤75%; Example 2: 23 ±10 °C, humidity ≤90%).
  • Documentation and reporting: Procedures for measurement, expression of results, and reporting are specified; Annexes provide uncertainty calculation examples, definitions of amplitude/phase sign, and nomograms.

Applications and practical value

  • Provides a standardized, traceable method for accelerometer calibration, vibration sensor verification, and transfer of calibration between standards.
  • Used by metrology laboratories, calibration service providers, R&D test labs, and maintenance/quality engineers who require documented measurement traceability and uncertainty evaluation.
  • Applicable for laboratory-grade calibrations where low uncertainty is required and for field calibrations using portable check standards (with expected larger uncertainties).

Related standards

  • ISO 16063 series (relevant parts): Part 1 (Basic concepts), Part 11 (Primary calibration by laser interferometry), Part 12 (Reciprocity), Part 13 (Primary shock calibration), Part 22 (Shock calibration by comparison).
  • Guide to the expression of uncertainty in measurement (GUM) - referenced for uncertainty methods.

Keywords: ISO 16063-21:2003, vibration calibration, reference transducer, calibration by comparison, accelerometer calibration, measurement uncertainty, vibration transducers, rectilinear vibration.

Standard
ISO 16063-21:2003 - Methods for the calibration of vibration and shock transducers
English language
29 pages
sale 15% off
Preview
sale 15% off
Preview

Frequently Asked Questions

ISO 16063-21:2003 is a standard published by the International Organization for Standardization (ISO). Its full title is "Methods for the calibration of vibration and shock transducers - Part 21: Vibration calibration by comparison to a reference transducer". This standard covers: ISO 16063-21:2003 describes the calibration of rectilinear vibration transducers by comparison. Although it mainly describes calibration using direct comparison to a standard calibrated by primary methods, the methods described can be applied between other levels in the calibration hierarchy. ISO 16063-21:2003 specifies procedures for performing calibrations of rectilinear vibration transducers by comparison in the frequency range from 0,4 Hz to 10 kHz. It is primarily intended for those who are required to meet ISO standardized methods for the measurement of vibration under laboratory conditions, where the uncertainty of measurement is relatively small. It can also be used under field conditions, where the uncertainty of measurement may be relatively large. From knowledge of all significant sources of uncertainty affecting the calibration, the expanded uncertainty can be evaluated using the methods given in ISO 16063-21. It also covers the assessment of uncertainties for calibrations performed using a check standard. This document is applicable to calibrations outside the frequency range from 0,4 Hz to 10 kHz, considering the additional facts described in 4.3. Systems have been demonstrated to cover the range from 0,01 Hz to 30 kHz.

ISO 16063-21:2003 describes the calibration of rectilinear vibration transducers by comparison. Although it mainly describes calibration using direct comparison to a standard calibrated by primary methods, the methods described can be applied between other levels in the calibration hierarchy. ISO 16063-21:2003 specifies procedures for performing calibrations of rectilinear vibration transducers by comparison in the frequency range from 0,4 Hz to 10 kHz. It is primarily intended for those who are required to meet ISO standardized methods for the measurement of vibration under laboratory conditions, where the uncertainty of measurement is relatively small. It can also be used under field conditions, where the uncertainty of measurement may be relatively large. From knowledge of all significant sources of uncertainty affecting the calibration, the expanded uncertainty can be evaluated using the methods given in ISO 16063-21. It also covers the assessment of uncertainties for calibrations performed using a check standard. This document is applicable to calibrations outside the frequency range from 0,4 Hz to 10 kHz, considering the additional facts described in 4.3. Systems have been demonstrated to cover the range from 0,01 Hz to 30 kHz.

ISO 16063-21:2003 is classified under the following ICS (International Classification for Standards) categories: 17.160 - Vibrations, shock and vibration measurements. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 16063-21:2003 has the following relationships with other standards: It is inter standard links to ISO 7376:2009, ISO 16063-21:2003/Amd 2:2024, ISO 16063-21:2003/Amd 1:2016, ISO 5347-3:1993. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase ISO 16063-21:2003 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.

Standards Content (Sample)


INTERNATIONAL ISO
STANDARD 16063-21
First edition
2003-08-15
Methods for the calibration of vibration
and shock transducers —
Part 21:
Vibration calibration by comparison to a
reference transducer
Méthodes pour l'étalonnage des transducteurs de vibrations et de
chocs —
Partie 21: Étalonnage de vibrations par comparaison à un transducteur
de référence
Reference number
©
ISO 2003
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.

©  ISO 2003
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2003 — All rights reserved

Contents Page
Foreword. iv
Introduction . v
1 Scope. 1
2 Normative references . 1
3 Uncertainty of measurement. 2
4 Requirements for apparatus and environmental conditions. 3
4.1 General. 3
4.2 Environmental conditions. 3
4.3 Reference transducer . 3
4.4 Vibration generation equipment. 4
4.5 Voltage measuring instrumentation. 6
4.6 Distortion measuring instrumentation. 6
4.7 Oscilloscope. 7
4.8 Phase shift measuring instrumentation . 7
5 Calibration. 7
5.1 Preferred amplitudes and frequencies . 7
5.2 Measurement requirements . 7
5.3 Calibration procedure. 8
6 Expression of results. 8
7 Reporting the calibration results. 9
Annex A (normative) Expression of uncertainty of measurement in calibration . 11
Annex B (normative) Definitions of amplitude sign and phase shift between mechanical motion
and vibration transducer electrical output. 20
Annex C (informative) Nomogram for conversion between acceleration, velocity and
displacement . 22
Annex D (informative) Example of uncertainty calculation. 24
Bibliography . 29

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 16063-21 was prepared by Technical Committee ISO/TC 108, Mechanical vibration and shock,
Subcommittee SC 3, Use and calibration of vibration and shock measuring instruments.
This first edition of ISO 16063-21 cancels and replaces ISO 5347-3:1993, which has been technically revised.
ISO 16063 consists of the following parts, under the general title Methods for the calibration of vibration and
shock transducers:
— Part 1: Basic concepts
— Part 11: Primary vibration calibration by laser interferometry
— Part 12: Primary vibration calibration by the reciprocity method
— Part 13: Primary shock calibration using laser interferometry
— Part 21: Vibration calibration by comparison to a reference transducer
— Part 22: Shock calibration by comparison to an accelerometer, velocity or force transducer
iv © ISO 2003 — All rights reserved

Introduction
The ISO 16063 series of standards is concerned with methods for the calibration of vibration and shock
transducers under both standard laboratory conditions and in the field.
As such, the intended user group of this part of ISO 16063 is wide, ranging from metrologists in mechanical
vibration to technicians evaluating the vibration characteristics of a machine or structure, or human exposure
to vibration. The key to the application of this part of ISO 16063 is in the careful detailed specification and
evaluation of measurement uncertainty, i.e. the error budget and computation of expanded uncertainty
associated with the measurement of vibration.
This part of ISO 16063 is particularly intended for those engaged in vibration measurements requiring
traceability to primary national or international standards through a secondary, reference, working or check
standard (portable calibrator intended for field use) as defined in the International vocabulary of basic and
general terms in metrology (VIM). The specifications for the instrumentation and the procedures given are
intended to be used for calibration of rectilinear vibration transducers (with or without signal conditioning) to
obtain the magnitude and (optionally) phase shift of the complex sensitivity at frequencies in the range of
0,4 Hz to 10 kHz.
INTERNATIONAL STANDARD ISO 16063-21:2003(E)

Methods for the calibration of vibration and shock
transducers —
Part 21:
Vibration calibration by comparison to a reference transducer
1 Scope
This part of ISO 16063 describes the calibration of rectilinear vibration transducers by comparison. Although it
mainly describes calibration using direct comparison to a standard calibrated by primary methods, the
methods described can be applied between other levels in the calibration hierarchy.
This part of ISO 16063 specifies procedures for performing calibrations of rectilinear vibration transducers by
comparison in the frequency range from 0,4 Hz to 10 kHz. It is primarily intended for those who are required to
meet ISO standardized methods for the measurement of vibration under laboratory conditions, where the
uncertainty of measurement is relatively small. It can also be used under field conditions, where the
uncertainty of measurement may be relatively large.
From knowledge of all significant sources of uncertainty affecting the calibration, the expanded uncertainty
can be evaluated using the methods given in this part of ISO 16063. It also covers the assessment of
uncertainties for calibrations performed using a check standard.
Comparison calibrations made in accordance with this part of ISO 16063 need to allow for the environmental
conditions of the reference transducer calibration.
NOTE Transducer calibrations made under extreme environmental conditions are covered by other International
Standards.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 266, Acoustics — Preferred frequencies
ISO 2041:1990, Vibration and shock — Vocabulary
ISO 16063-1:1998, Methods for the calibration of vibration and shock transducers — Part 1: Basic concepts
ISO 16063-11:1999, Methods for the calibration of vibration and shock transducers — Part 11: Primary
vibration calibration by laser interferometry
Guide to the expression of uncertainty in measurement (GUM). BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML,
1)
1) Corrected and reprinted in 1995.
3 Uncertainty of measurement
3.1 All users of this part of ISO 16063 are expected to make uncertainty budgets according to Annex A to
document their level of uncertainty (see example in Annex D).
To help set up systems fulfilling different requirements two examples are given. System requirements for each
are set up and the attainable uncertainty is given. Example 1 is typical for calibrations under well-controlled
laboratory conditions with the requirement to obtain a high accuracy. Example 2 is typical for calibrations
where less than the highest accuracy can be accepted or where calibration conditions are such that only less
narrow tolerances can be maintained. These two examples will be used throughout this part of ISO 16063.
a) Example 1
The reference transducer is calibrated by primary means and documented uncertainty. The calibration
may be transferred to a working standard for practical reasons. The temperature and other conditions are
kept within narrow limits during the comparison calibration as indicated in the appropriate clauses.
b) Example 2
The reference transducer is not calibrated by primary means, but has a traceable calibration, as defined
in VIM (see [2]), with the corresponding uncertainty documented. The calibration may be transferred to a
working standard for practical reasons. The requirements on other parameters and instruments are
indicated in the appropriate clauses.
3.2 For both examples, the minimum calibration requirement for the reference transducer is calibration
under suitable reference conditions (i.e. frequency, amplitude and temperature). Normally the conditions will
be chosen as indicated in ISO 16063-11.
It is applicable for the following parameters:
Frequency range: 20 Hz to 5 000 Hz, optionally 0,4 Hz to 10 000 Hz (see Note 1)
2 2 2 2
Dynamic range: 10 m/s to 1 000 m/s r.m.s., optionally 0,1 m/s to 1 000 m/s
(frequency dependent)
NOTE The indicated frequency ranges are not mandatory and single-point calibrations are also acceptable.
At any given frequency and amplitude of acceleration, velocity or displacement, the dynamic range will be
limited by the noise floor and the amount of distortion produced by the excitation apparatus (if no filtering is
used) or its maximum power. (Techniques are also used to counteract the inherent distortion at large
displacements for spring-controlled exciters by changing the waveform of the input voltage.) Typical maximum
values for electrodynamic vibration exciters designed for the frequency range from 10 Hz to 10 kHz are
2 2
200 m/s to 1000 m/s r.m.s. acceleration, 0,5 m/s to 1 m/s r.m.s. velocity and 5 mm peak displacement. The
lower limits will be set by the noise in the two measurement channels, and by the bandwidth used. Typical
2 2
values used for measurement are 50 m/s to 100 m/s r.m.s. acceleration or 0,1 m/s r.m.s. velocity. For
2 2
calibrators, values between 1 m/s and 10 m/s r.m.s. are normally used. A graph similar to the one shown in
Annex C is useful when considering the ranges covered.
When measurements are performed at the lowest frequencies, the limiting factor is normally displacement. At
2 2
1 Hz, typical values for long-stroke vibrators are 1 m/s to 2 m/s r.m.s. acceleration or 0,1 m/s to 0,3 m/s
r.m.s. velocity.
3.3 The attainable uncertainties (expanded uncertainties calculated using a coverage factor of 2 in
accordance with ISO 16063-1) for the two examples are given in Table 1. In practice, these limits may be
exceeded depending on the uncertainty with which the reference transducer has been calibrated, the
response characteristics of the reference transducer and the transducer to be calibrated, the vibratory
characteristics of the exciter and the instrumentation used in the measurement apparatus. It is the
responsibility of the laboratory or end user to make sure that the reported values of expanded uncertainty are
credible.
2 © ISO 2003 — All rights reserved

Table 1 — Attainable uncertainties of magnitude and phase shift of the complex sensitivity
Parameter Example 1 Example 2
Magnitude
For accelerometers (0,4 Hz to 1 000 Hz) 1 % 3 %
For accelerometers (1 000 Hz to 2 000 Hz) 2 % 5 %
For accelerometers (2 kHz to 10 kHz) 3 % 10 %
For displacement and velocity transducers (20 Hz to 1 000 Hz) 4 % 6 %
a
Phase shift
b
At reference conditions (i.e. the level and frequency at which the 1° 3°
reference transducer was calibrated)
Outside reference conditions 2,5° 5°
a
Phase shift measurement is not mandatory.
b
Recommended reference conditions are as follows (from ISO 16063-11:1999, Clause 2):
 frequency in hertz: 160, 80, 40, 16 or 8 (or angular frequency ω in radians per second: 1000, 500, 250, 100 or 50),
 acceleration in metres per second squared (acceleration amplitude or r.m.s. value): 100, 50, 20, 10, 5, 2 or 1.
NOTE The expanded uncertainties given as examples (e.g. 1 %) are based on concrete uncertainty budgets such
as given in Annex D as an example (resulting expanded uncertainty 0,84 %).

4 Requirements for apparatus and environmental conditions
4.1 General
The examples referred to in this clause are those described in Clause 3.
If the recommended specifications listed below are met for each item, the uncertainties given in Clause 3
should be obtainable over the applicable frequency range depending on the uncertainty with which the
reference transducer has been calibrated, and the response characteristics of the reference transducer and
transducer to be calibrated. Other combinations of requirements can, however, lead to the same uncertainty.
Special instrumentation may be required in order to meet the expanded uncertainties given in Clause 3 at
frequencies less than 1 Hz. It is mandatory to document the expanded uncertainty using the methods of
Annex A.
4.2 Environmental conditions
These shall be the following.
Example 1 Example 2
Room temperature (23 ± 5) °C (23 ± 10) °C
Relative humidity 75 % max. 90 % max.

4.3 Reference transducer
This should preferably be calibrated together with the amplifier.
a) Example 1
The transducer shall be calibrated in accordance with suitable primary methods or by comparison against
a transducer calibrated in accordance with suitable primary methods (see ISO 16063-11 or other parts)
with an expanded uncertainty of 0,5 % (magnitude) and 0,5° (phase shift) at selected reference frequency
and acceleration (the uncertainties are those obtained when calculating expanded uncertainties using a
coverage factor of 2). Higher uncertainty values are accepted at high and low frequencies.
b) Example 2
The transducer shall be calibrated by suitable and known methods with traceability to a primary reference
transducer and an uncertainty of less than 2 % (magnitude) and 2° (phase shift) at selected reference
frequency and acceleration (the uncertainties are those obtained when calculating expanded
uncertainties using a coverage factor of 2). Higher uncertainty values are accepted at high and low
frequencies.
The reference transducer may be of the so-called back-to-back type meant for direct mounting of the
transducer to be calibrated on top of it in a so-called back-to-back configuration (see Figure 1). It may also be
a transducer with normal mounting provisions used underneath a fixture in line with the transducer to be
calibrated. It is not recommended to mount the two transducers side by side as rocking motions will often be
present, causing large errors in many circumstances. For calibrators, the reference transducer may be an
integral part of a moving element.
Subclauses 4.4 to 4.8 specify characteristics of apparatus that contribute to the uncertainty of measurement.
4.4 Vibration generation equipment
This shall fulfil the requirements given in Table 2.
Table 2 — Vibration generation equipment
Parameter Unit Example 1 Example 2
Frequency uncertainty % u 0,1 u 0,2
% of reading over the
Frequency stability 0,1 0,2
measurement period
% of reading over the
Acceleration amplitude stability 0,1 0,3
measurement period
Total harmonic distortion at
% u 5 u 10
frequencies > 20 Hz
Total harmonic distortion over the
% u 10 u 20
whole frequency range
u 10 at f u 1 kHz
Transverse, bending and rocking
%
acceleration
u 30 at f > 1 kHz
Hum and noise ( f W 10 Hz) dB below full output W 50 W 40
Hum and noise ( f < 10 Hz) dB below full output W 20 W 10

The hum and noise will only be important when inside the measurement bandwidth used. For every
combination of frequency, acceleration and load that is used during calibration, the magnitude of the
transverse, bending and rocking accelerations, hum and noise shall be consistent with the uncertainties given
in Clause 3.
Static or dynamic base strain introduced to the transducer from the attachment surfaces shall not unduly
influence the calibration result.
All mounting surfaces between any two transducers compared shall have flatness and roughness
specifications suitable for the purpose. If the highest frequency range is used, strict tolerances are necessary.
The surface on which the transducer is to be mounted shall have a roughness value, expressed as the
arithmetical mean deviation Ra of < 1 µm. The flatness shall be such that the surface is contained between
two parallel planes 5 µm apart, over the area corresponding to the maximum mounting surface of any
4 © ISO 2003 — All rights reserved

transducer to be calibrated. The drilled and tapped hole for connecting the transducer shall have a
perpendicularly tolerance to the surface of < 10 µm, i.e. the centreline of the hole shall be contained in a
cylindrical zone with 10 µm diameter and a height equal to the hole depth.
The mounting surface of the vibration exciter should be perpendicular to the direction of motion. Any deviation
from perpendicularity should be taken into account in the uncertainty budget, see Annex A.

Key
1 exciter 5 reference transducer 8 distortion meter for occasional checks
2 amplifiers 6 transducer to be calibrated 9 oscilloscope for visual inspection (optional)
3 power amplifier 7 voltmeter 10 phase meter (optional)
4 frequency generator and indicator
Figure 1 — Example of a measuring system for vibration calibration by comparison to a
reference transducer
NOTE 1 Multisine, sine or random generators can be used in conjunction with frequency analysers. Typically Fast
Fourier Transform (FFT) analysers are used for random and multisine signals and Single-Sine Correlation or Frequency
Response Analysers (FRA) are used for single-sine signals. The distortion is then normally of no importance. Therefore
analysers are normally preferred instead of broadband r.m.s. voltmeters which, although fundamentally more accurate, are
sensitive to distortion and other signals at frequencies differing from the measurement frequency. Measurement of
coherence can be used to estimate whether or not the signal-to-noise ratio and the linearity of the transducers are within
well-defined limits when spectral averaging is used. With random excitation and 64 averages a minimum coherence limit
of 0,98 will ensure that the errors due to signal-to-noise ratio and linearity are less than 0,9 % for a dual channel
measurement. In rare cases, broadband excitation can, however, create unwanted (transverse) vibration or output signals
at a measuring frequency due to non-linear behaviour of shaker or transducer at other frequencies.
NOTE 2 The items in 4.3 and 4.4 may be integrated into a calibrator.
4.5 Voltage measuring instrumentation
Two alternative set-ups are considered.
a) A single voltmeter measuring true r.m.s. at transducer amplifier output is used. The outputs from the
reference transducer and the transducer to be calibrated are measured consecutively and the reference
transducer output at least twice. This equipment shall fulfil the requirements given in Table 3.
Table 3 — Voltage measuring instrumentation — Single voltmeter
Parameter Unit Example 1 Example 2
Frequency range Hz 1 to 10 000 1 to 10 000
% of reading for max.
Maximum deviation from linearity 0,1 0,3
difference in signal levels
Maximum deviation between two
consecutive reference transducer % 0,1 0,3
measurements
NOTE The last row describes the repeatability of the measurement. This includes more than the voltmeter repeatability but is
treated here as a general requirement.

b) An instrument measuring voltage ratio between transducer amplifier outputs is used. This equipment shall
have the characteristics specified in Table 4.
Table 4 — Voltage measuring instrumentations
Parameter Unit Example 1 Example 2
Frequency range Hz 1 to 10 000 1 to 10 000
Maximum uncertainty % 0,2 0,5
4.6 Distortion measuring instrumentation
Distortion measuring instrumentation (limited use, see Note) capable of measuring total harmonic distortion of
1 % to 10 % shall have the characteristics specified in Table 5.
Table 5 — Distortion measuring instrumentation
Parameter Unit Example 1 Example 2
Frequency range Hz 1 to 50 000 1 to 50 000
Maximum uncertainty % of reading 10 10
NOTE Distortion measurement is only needed for sine calibration and is not included in the standard procedure. It is used to
check the performance of the vibration generating equipment initially and then only with suitable intervals or in case of doubt.
6 © ISO 2003 — All rights reserved

4.7 Oscilloscope
An oscilloscope or similar display may be used for examining the waveforms of the transducer signals.
Its use is strongly recommended but not mandatory.
4.8 Phase shift measuring instrumentation
This equipment shall have the characteristics specified in Table 6.
Table 6 — Phase shift measuring
Parameter Unit Example 1 Example 2
Frequency range Hz 1 to 10 000 1 to 10 000
Maximum uncertainty ° (degree) 0,2 0,5
5 Calibration
5.1 Preferred amplitudes and frequencies
Six frequencies, each with associated acceleration (amplitude or r.m.s. value) and equally covering the
transducer range, should preferably be chosen from the following series.
a) Acceleration (m/s ):
 1, 2, 5, 10 or their multiples of ten.
If broadband signals are used, these values are the total r.m.s. values.
b) Frequency:
 selected from standardized one-third-octave frequency series (see ISO 266).
If broadband signals are used, the desired range should be covered in one or more calibrations.
Values chosen should preferably be the same as those used in the reference transducer calibration. If the
transducer is to be calibrated at frequencies and accelerations other than those at which the reference has
been calibrated, the characteristics of the reference transducer should be assessed at those frequencies and
accelerations. The resulting uncertainty component shall be taken into account in the uncertainty budget (see
Annex A).
5.2 Measurement requirements
When a calibration is to be performed using a new set-up or a new transducer, it is good practice to carry out
the calibration more than once to ensure sufficient repeatability.
It is important to ensure that cable motion and base strain do not appreciably affect the measurement results,
particularly at low frequencies. Altering the attachment of the cable, the mounting of the transducer, or both,
and noting changes in the measurement results or harmonic distortion may be used to evaluate effects due to
these causes. If the measured sensitivity or distortion does not change significantly when compared to the
uncertainty in the calibration, then these influences may be neglected. The mounting conditions of the
transducer should also be repeatable. This can be verified by remounting the transducer several times and
measuring the sensitivity after each successive attachment of the transducer.
If the transducer under test is not being calibrated in combination with an associated signal conditioner or
amplifier, then the gain and frequency response (i.e. magnitude and, if needed, phase shift) of the complex
sensitivity of the signal conditioner or amplifier used with the transducer under test should be determined in a
traceable fashion at all measurement frequencies. The sensitivity and frequency response of the reference
(transducer plus amplifier) shall also be determined in a traceable fashion at all measurement frequencies.
If any variations, significant compared to the desired uncertainty, are found in the above tests, these should be
quantified by making a sufficiently large number of repeated measurements to get a good estimate of the
variance. This shall then be included in the final uncertainty statement. This is especially important if the
measurement is not made at the frequencies and amplitudes at which the reference transducer was calibrated.
5.3 Calibration procedure
The surfaces of the reference transducer (or fixture) and the transducer to be calibrated shall be examined to
verify that they are free from burrs, etc. and that they comply with the manufacturer’s flatness specifications
and the specifications of Clause 4.
Mount the reference transducer (see 4.4) and the transducer to be calibrated back-to-back or in-line on a
fixture on the exciter or on the exciter with integral working reference transducer using the recommended
torque. Below approximately 5 kHz, good fixtures with known characteristics may be used between the
transducers. At higher frequencies, the direct back-to-back configuration or integral working reference
transducer shall be used. An example of a block diagram of a typical laboratory calibration apparatus is shown
in Figure 1. The voltmeter, selector, generator and phase meter are often substituted by a two-channel
instrumentation (e.g. dual-channel analyser with internal generator or voltage ratio meter) with sufficient
accuracy.
Measure the ratio of the two outputs and the relative phase shift, if needed.
Determine the sensitivity at the reference frequency, for accelerometers preferably at 160 Hz (second choice:
2 2
80 Hz), and at the reference acceleration, for accelerometers preferably at 100 m/s (other choices: 10 m/s ,
2 2
20 m/s or 50 m/s ), then determine the sensitivity at other calibration frequencies and accelerations. The
results shall be given in absolute terms and/or as a relative deviation (percentage or decibels) and degrees
deviation from the sensitivity at the reference point.
In the case of stud-mounted transducers, a thin film of light oil, wax or grease should be used between the
mounting surfaces of the transducer(s) and exciter, particularly in the case of calibrations performed at high
frequencies (see ISO 5348 for details).
6 Expression of results
If the reference transducer and the transducer under test respond to the same vibration quantity, calculate the
magnitude S and the phase shift ϕ of the complex sensitivity of the transducer to be calibrated using the
2 2
following formulae:
Magnitude: Phase shift:
X
SS= ϕ=+ϕϕ
21 22,1 1
X
where
S , ϕ are the magnitude and phase shift of the complex sensitivity of the reference transducer;
1 1
X is the output from the reference transducer;
X is the output from the transducer to be calibrated;
ϕ is the phase shift between the outputs from the transducer to be calibrated and the reference
2,1
transducer.
8 © ISO 2003 — All rights reserved

If the two transducers measure different vibration quantities, calculate the sensitivity of the transducer to be
calibrated, using the following formulae.
a) If the magnitude and phase shift of the complex acceleration sensitivity S , ϕ were measured:
a a
Magnitude: Phase:
Sf=π2S 90ϕϕ=− °
va va
Sf=π4 S 180ϕϕ=−°
s a sa
b) If the magnitude and phase shift of the complex velocity sensitivity S , ϕ were measured:
v v
Magnitude: Phase:
Sf=π2S ϕϕ=− 90°
s v sv
where
S , ϕ are the magnitude and phase shift of the complex acceleration sensitivity;
a a
S , ϕ are the magnitude and phase shift of the complex velocity sensitivity;
v v
S , ϕ are the magnitude and phase shift of the complex displacement sensitivity;
s s
f is the frequency of the vibration, in hertz.
The phase shift is undetermined within modulus 180° until definitions of directions for the transducers, motions
and electrical signals are specified (see ISO 16063-1:1998, 3.3, and Annex B for guidance).
For calibrators designed to provide one or more fixed amplitude(s) at one or more frequencies, the sensitivity
of the transducer to be calibrated is obtained by using the specified amplitude A and the measured output X in
the formula
X
S =
A
where A and S must refer to the same vibration quantity.
7 Reporting the calibration results
In addition to the calibration method and instrumentation with calibration due dates, at least the following
conditions and characteristics shall be stated when the calibration results are reported.
a) Ambient conditions, Example 1:
 temperature of the transducer if measured, or
 ambient air temperature and estimated difference from transducer temperature if this is not
measured.
b) Ambient conditions, Example 2:
 estimated transducer temperature.
c) Mounting technique:
 material of mounting surface,
 mounting torque (if stud mounted and optional for Example 2) or adhesive used,
 characteristics of mounting components or adapters (if used),
 oil or grease or wax (if used),
 cable fixing,
 orientation (vertical or horizontal).
d) All amplifier settings (if adjustable) when the transducer is calibrated in combination with a signal
conditioner or amplifier:
 gain,
 cut-off frequencies and slope of filters.
e) Calibration results:
 values of calibration frequencies and vibration amplitudes,
 values of sensitivity (magnitude and phase shift, if measured),
 coherence of the measurement if measured,
 expanded uncertainty of the calibration (calculated in accordance with Annex A).
f) Coverage factor, if different from k = 2 (corresponding to a confidence level of 95 % for a normal
distribution).
10 © ISO 2003 — All rights reserved

Annex A
(normative)
Expression of uncertainty of measurement in calibration
A.1 Calculation of expanded uncertainty of measurement, U
A.1.1 Purpose of U
The uncertainty of measurement in calibration shall be expressed by the "expanded uncertainty" U in
accordance with GUM, based on the approach recommended by the International Committee for Weights and
Measures (CIPM). The purpose of U is to provide an interval y − U to y + U within which the value of Y, the
specific quantity subject to calibration and estimated by y, can be expected to lie with high probability. To
confidently assert that y − U u Y u y + U, the expanded uncertainty U shall be determined as follows.
A.1.2 Corrections
Every effort has to be made to identify each effect that significantly influences the measurement result and to
compensate for such effects by applying the estimated corrections or correction factors.
If an effect influencing the measurement result is appropriately described by a probability distribution
(preferably probability density, see A.1.3) having a significant expected value (in particular for an asymmetrical
distribution), the latter shall be treated as systematic error and compensated by correction.
A.1.3 Standard uncertainty estimation
Each component of uncertainty that contributes to the uncertainty of the measurement shall be represented by
a standard deviation u , termed "standard uncertainty", equal to the positive square root of the variance u .
i i
Some standard uncertainties can be obtained as statistically estimated standard deviations by the statistical
analysis of series of observations (type A evaluation of standard uncertainty). Other standard uncertainties
shall be evaluated as the standard deviation of a probability distribution describing the scientific judgement of
all possible values of the respective quantity (type B evaluation of standard uncertainty). The judgement is
based on all information available about the quantity. In particular, if there is no specific information about the
possible values of a quantity responsible for systematic effects except that these values are within the bounds
b and b , a uniform distribution over the interval [;bb] may be used to represent this information. It has a
− + − +
standard uncertainty b 3 where bb=−b 2 . The expected value is bb+ 2 to be used for
( ) ( )
+− +−
correction in this case.
If an influence quantity can be considered uniformly distributed (rectangular probability density) but is known to
be transformed into the measurement result with a specific non-linear function (e.g. sinusoidal; polynomial of
second or third order), this information shall be taken into account by choosing the associated distribution model.
EXAMPLE The sensitivity S of an accelerometer to sinusoidal accelerations in the nominal measurement direction is
calculated from the output, voltage or charge amplitude xˆ, stimulated by a vibration, acceleration amplitude aˆ, using the
formula Sx=ˆˆ/.a Among the various disturbing effects influencing the measurement result in calibration, there may be a
significant transverse motion component from the vibration exciter, acceleration amplitude aˆ , transformed into an error
T
component e in the output, in conjunction with the accelerometer's transverse sensitivity, S . It is assumed for this
xˆT
T
example that the acceleration to be measured and the transverse acceleration have the same frequency and that there is
no phase angle difference. As the transverse sensitivity is usually sinusoidally dependent on the angle β between the
direction of maximum transverse sensitivity S and the direction of a transverse excitation, the error component can
()
T,max
be expressed by
eS==aˆˆS a cos β
xˆTTTT,maxT,max
If the values of the maximum transverse sensitivity S and the maximum transverse acceleration aˆ are
()T,max ()T,max
known while the angle β is not, it is reasonable to assume a rectangular distribution of β within the interval −π;.π Thus,

the influence quantity, i.e. transverse acceleration, with rectangularly distributed angle β leads to a measurement error
component e whose probability density is described by
xˆT
we =
()
xˆT
e
xˆT
bπ−1

b

− xˆT
bS= aˆ
T,max T,max
(often referred to as arcsin distribution). The associated standard uncertainty is
ue =b 2
( )
ˆ
xT
The expected value Ee is zero in this case. This is the best estimate of the error e .
{ }
ˆ ˆ
xT xT
A.1.4 Combined standard uncertainty
The "combined standard uncertainty" u , as the standard uncertainty of the measurement of Y, shall be
c
determined by combination of the individual standard uncertainties (and covariances as appropriate) using the
law of propagation of uncertainty. Accordingly, the combined standard uncertainty is obtained from
N = 1
NN

∂∂ff∂f
uy=+u x2,uxx (A.1)
() ()
 ()
c∑∑ii∑ j
∂∂xx∂x
iij
ii==11j=i+1
This equation is based on a first-order Taylor series approximation of
Y = f (X , X ,.,X ) (A.2)
1 2 N
where Y is the measurand determined from N input quantities X , X , ., X through a functional relationship f.
1 2 N
An estimate of the measurand Y, denoted by y, is obtained from Equation (A.1) using input estimates x , x , .,
1 2
x for the values of the input quantities. Thus the output estimate, which is the result of measurement, is given
N
by
y = f (x , x ,.,x ) (A.3)
1 2 N
In Equation (A.1), the symbols ∂ f / ∂ x are often referred to as sensitivity coefficients c . They are equal to the
i i
partial derivatives ∂ f / ∂ X evaluated at X = x. The symbol u(x ,x) designates the estimated covariance
i i i i j
associated with x and x .
i j
For the case where no significant correlations are present, Equation (A.1) is reduced to
N

∂ f
uy = u x (A.4)
() ()

c ∑ i
∂ x
i
i = 1
NOTE The first-order Taylor series approximation of Equation (A.2) resulting in Equation (A.1) is only applicable if the
model function f is sufficiently linear with respect to the variation of the input estimates x within the ranges characterized
i
by the uncertainties u(x ). This is not the case in the example given in A.1.3 if the angle β is considered to be an input
i
quantity X . To overcome this obstacle, which similarly exists with other influence quantities acting in measurements within
i
calibrations of vibration and shock transducers, an adequate model has been introduced (see reference [4]). To briefly
12 © ISO 2003 — All rights reserved

specify this model for the example above, a factor (1 −ex/ ˆ) with ex/ ˆ << 1 is introduced, as an input quantity X , into
ˆ ˆ
xT xT i
the functional relationship used for calculating the measurand. Equation (A.2), specially tailored to the example, is reduced
to three input quantities
Y = f (X , X , X )
1 2 3
where
Y is the measurand (sensitivity S);
X is the accelerometer output (voltage or charge amplitude xˆ);
X is the acceleration amplitude;
ˆ
X =−1 ex .
( )
3Txˆ
Thus, the relationship
X
YX=
X
can be established. The first Taylor series approximation can be used now, leading to the relative combined standard
uncertainty
22 2
   
u(y) ux( ) u()x ux()
c1 2 3
=+ +
   

yx x x
12 3
   
if there are no significant correlations.
Using the symbols introduced in the Example, the latter relationship can be written as follows:
22 2
uS() u(xˆˆ) u(a) u(e /xˆ)
    
ˆ
cTx
=+ +
    
Sxˆˆa 1
    
where
ˆˆ
ux()/x is the relative standard uncertainty of the output voltage amplitude measurement;
ˆˆ
ua() a the relative standard uncertainty in acceleration amplitude measurement;
ˆˆ
ue( x) =u (e ) x , with ue( ) =b 2 as explained in the Example.
xxˆˆTT xˆT
Accordingly, further factors whose deviations from the value 1 are similarly expressed by the relative error component of
the respective quantity (e.g. voltage, acceleration or the sensitivity as a whole) might be introduced as input quantities
(X , X , .), allowing the variety of uncertainty sources to be taken into account separately.
4 5
A.1.5 Expanded uncertainty
The "expanded uncertainty" U shall be determined by multiplying u by a coverage factor k:
c
U = ku
c
where a value of k = 2 should preferably be used. If it can be assumed that the possible values of the
calibration result are approximately normally distributed with approximate standard deviation u , the unknown
c
value can be asserted to lie in the interval defined by U with a level of confidence, or probability, of
approximately 95 %.
A.1.6 Reporting the result
When reporting the result of the measurement y, the expanded uncertainty and the value of the coverage
factor k used, if different from k = 2, shall be stated. In addition, the approximate coverage probability or level
of confidence of the interval may be stated.
A.2 Calculation of expanded uncertainties at reference conditions
A.2.1 Calculation of the relative expanded uncertainty U (S) for the sensitivity magnitude
rel
The relative expanded uncertainty of measurement of the magnitude of the complex sensitivity, U (S), for
rel
each of the applied frequencies, accelerations and amplifier gain settings (if an amplifier is part of the
calibrated transducer) is calculated from the following formulae:
US() =ku ()S
rel c,rel
with the coverage factor k = 2 (see A.1.5);
N − 1 N
uS() 1 ∂∂ff∂f
c
uS()== u (S)+2 u(x,x)
c,rel iij
∑∑∑
SS∂∂x x∂x
iij

ii==11ji+
N − 1
N

∂∂ff∂f
=+uS() 2 u (
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

ISO 16063-21:2003는 진동 및 충격 변환기의 교정을 위한 표준 문서로, 주로 직선 진동 변환기의 비교 교정 방법을 설명하고 있습니다. 이 표준은 기본 방법에 의해 교정된 기준과의 직접 비교를 통해 교정을 수행하는 방법을 다루고 있으며, 교정 계층의 다른 수준에도 적용할 수 있는 방법들을 포함하고 있습니다. 특히, 이 표준은 0.4 Hz에서 10 kHz까지의 주파수 범위에서 직선 진동 변환기의 교정을 수행하는 절차를 규정하고 있습니다. ISO 16063-21:2003의 주요 강점은 대부분의 실험실 환경에서 진동 측정의 ISO 표준화된 방법을 준수해야 하는 사용자를 위한 것입니다. 측정의 불확실성이 상대적으로 작을 때, 이 표준은 신뢰할 수 있는 교정 결과를 제공하여, 실험실 사용할 상관된 교수방법을 보장합니다. 뿐만 아니라, 현장 조건에서도 적용될 수 있는 유연성을 제공하여, 측정의 불확실성이 상대적으로 큰 상황에서도 활용 가능성이 있습니다. 문서에서는 모든 중요한 불확실성 소스를 고려하여 교정의 확장된 불확실성을 평가하는 방법을 제시하고 있으며, 비교 기준을 사용한 교정의 불확실성 평가도 포함되어 있어 실질적으로 교정의 정확도를 높이는 데 기여하고 있습니다. 또한, 주파수 범위가 0.4 Hz에서 10 kHz를 초과하는 교정에 대한 적용 가능성도 제시하므로, 0.01 Hz에서 30 kHz 범위의 시스템 교정을 위한 유용한 지침을 제공합니다. 결론적으로, ISO 16063-21:2003은 진동 변환기의 교정 방법에 대한 포괄적인 지침을 제공하여, 다양한 환경에서의 신뢰성이 높은 측정 결과를 보장하고, 해당 분야에서 ISO 표준을 준수할 수 있도록 돕는 중요한 역할을 하고 있습니다.

ISO 16063-21:2003は、振動および衝撃トランスデューサの校正方法に関する基準であり、特に直線的振動トランスデューサの校正に焦点を当てています。この基準は、主に一次方法によって校正された標準との直接比較を用いた校正手法を詳細に説明しており、校正階層内の他のレベルにおいても適用可能です。このように、ISO 16063-21:2003は、その包括的なアプローチにより、振動測定を行う研究者や技術者にとって非常に便利なリソースです。 特に注目すべきは、周波数範囲が0.4 Hzから10 kHzまでの直線的振動トランスデューサの校正手続きを具体的に示している点です。これにより、ISO標準化された方法での測定が要求されるラボ環境下でも、測定の不確かさが相対的に小さい状況での使用が推奨されます。また、現場条件下でも使用できる柔軟性を持ち合わせており、不確かさが大きくなる可能性がある状況にも対応できることは、この基準の大きな強みです。 さらに、ISO 16063-21:2003は、校正に影響を与えるすべての重要な不確かさの要因を考慮し、拡張不確かさの評価方法も提供しています。チェック標準を使用した校正に関する不確かさの評価を含むことで、より信頼性の高い結果を期待できます。加えて、4.3節に記載された追加の事実を考慮することで、0.4 Hzから10 kHz以外の周波数範囲においても校正が適用できる点は、さまざまなアプリケーションにおける実用性を高めています。校正システムが0.01 Hzから30 kHzの範囲をカバーできることも、ISO 16063-21:2003の適用性をさらに広げており、極めて価値のある基準として位置付けられています。

ISO 16063-21:2003 serves as a critical standard in the field of vibration calibration, specifically for rectilinear vibration transducers. Its primary focus on calibration by comparison is a notable strength, allowing for robust and accurate measurement of vibration according to ISO standards. The document delineates a clear scope for calibrations within the frequency range of 0.4 Hz to 10 kHz, making it highly relevant for laboratory environments where precision is paramount. A key strength of ISO 16063-21:2003 is its adaptability for different levels within the calibration hierarchy. Although it primarily emphasizes comparisons to standards calibrated by primary methods, it effectively accommodates various calibration scenarios. This flexibility is crucial for laboratories adhering to strict ISO standardized methods while also providing a pathway for field applications where measurement uncertainty is greater. Moreover, the standard provides comprehensive procedures for evaluating the expanded uncertainty associated with the calibration process. This emphasis on uncertainty assessment adds significant value to practitioners, offering a methodical approach to identify and mitigate potential sources of error. Additionally, the guidelines for using check standards further enhance the usability of ISO 16063-21:2003 across diverse calibration contexts. The inclusion of specifications for calibrations extending beyond the core frequency range is another noteworthy aspect. By addressing potential calibrations from 0.01 Hz to 30 kHz with appropriate considerations, the standard broadens its applicability and ensures comprehensive support for a wide range of vibration measurement scenarios. Overall, ISO 16063-21:2003 stands out as a significant contribution to the calibration of vibration and shock transducers, effectively meeting the demands of both laboratory and field conditions while maintaining rigorous adherence to ISO standards. Its focus on comparison methods, uncertainty evaluation, and adaptability to varied calibration contexts ensures its relevance and strength in the field of vibration measurement.