X and gamma reference radiations for calibrating dosemeters and dose ratemeters and for determining their response as a function of photon energy — Addendum 1: High rate series of filtered X-radiations

Rayonnements X et gamma de référence pour l'étalonnage des dosimètres et débitmètres et pour la détermination de leur réponse en fonction de l'énergie des photons — Additif 1: Rayonnements X filtrés à fort débit

General Information

Status
Withdrawn
Publication Date
30-Jun-1983
Withdrawal Date
30-Jun-1983
Current Stage
9599 - Withdrawal of International Standard
Completion Date
18-Dec-1997
Ref Project

Relations

Buy Standard

Standard
ISO 4037:1979/Add 1:1983 - High rate series of filtered X-radiations
English language
12 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 4037:1979/Add 1:1983 - Rayonnements X filtrés a fort débit
French language
12 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL STANDARD IS0 4037-1979/ADDENDUM 1
Published 1983-07-15
INTERNATIONAL ORGANIZATION FOR STANDARDIZATION. MEXAYHAPOAHAR OPrAHM3Al&'lA no CTAHAAPTM3Al(MM. ORGANISATION INTERNATIONALE DE NORMALISATION
X and y reference radiations for calibrating dosemeters
and dose ratemeters and for determining their response
as a function of photon energy
ADDENDUM 1 : High rate series of filtered X-radiations
*
Addendum 1 to International Standard IS0 4037-1979 was developed by Technical Committee ISO/TC 85, Nuclear energy, and was
circulated to the member bodies in March 1982.
It has been approved by the member bodies of the following countries:
Australia F.R. Romania
Germany,
South Africa, Rep. of
Austria Hungary
Belgium Sweden
Italy
Switzerland
China Japan
Czechoslovakia Korea, Rep. of Turkey
United Kingdom
Egypt, Arab Rep. of Mexico
Finland Netherlands USA
France Poland
No member body expressed disapproval of the document.
1 Scope and field of application 2 Qualities of specified reference radiations
This Addendum specifies a series of lightly filtered X-radiations
The quality of the reference radiations is specified in terms of
intended for calibration procedures in which high exposure
the X-ray tube constant potential, by the first half value layer
rates are required. The series is also applicable to, for example,
(HVL,) and by the homogeneity coefficient. These specific.-
determination of the saturation characteristics of a dosemeter
tiens are listed in table 10.
or for checking scale linearity.
Since the spectra obtained are broad and asymmetric, the
3 Effective energy
series is not suitable for accurate energy response measure-
ments and shall only be used when the exposure rates pro-
duced by the narrow and wide series prove inadequate. If required, the effective energyl), E, of the radiations listed in
1) The effective energy of filtered X-radiation with a given half value layer is defined as the energy of the monoenergetic radiation having the same
half value layer.
w
UDC 535-34/-36 53.089.6 Ref. No. IS0 4037-1979/Add. 1-1983 (FI
R
g
Descriptors : dosemeters, exposure flowmeters, calibrating, electromagnetic radiation, reference materials, test equipment, test results.
8 O International Organization for Standardization, 1983 0
s Printed in Switzerland Price based on 12 pages

---------------------- Page: 1 ----------------------
IS0 4037-1979/Add. 1-1983 (E)
4.3 Fixed filtration
table 10 may be determined from the half value layer, HLV,, (in
millimetres) using the relationship
For tube potentials up to and including 60 kV, the correspon-
ding total filtration (fixed + additional) shall be less than the
HLV, = /n2/p(E)
equivalent of 4 mm of aluminium. An X-ray tube with low in-
herent filtration is required to generate the lower energy radia-
where p(E) (in reciprocal millimetres) is the characteristic linear
tions. At potentials of 100 kV and above the fixed filtration
attenuation coefficient of the material in which the half value
should be adjusted to the equivalent of 4 mm of aluminium.
layer has been specified, for photons of energy E.
The aluminium filter used to supplement the inherent filtration
of the tube shall be placed after the additional copper filter in
order to reduce any fluorescent radiation arising from the
copper. The thickness of aluminium employed shall not be less
4 Filtration
than 0,5 mm.
4.4 Examples of additional filtration required
4.1 Filters
for specified radiation qualities
Filters of aluminium (constant potential < 100 kV) or copper
Table 11 lists, as an example, the additional filtration required
and aluminium (2 100 kV) shall be used.
to produce the radiation qualities specified for particular values
of the fixed filtration. The air layer thickness included in the ad-
4.2 Additional filtration
ditional filtration, is significant in the case of lower energy
radiations. Also included in table 11 are values of the first and
second half value layer, mean photon energy, effective energy
At a given potential, the thickness of the additional filtration
and homogeneity coefficient. The effective energy has been
shall be adjusted so that the measured first half value layer lies
calculated from the first half value layer using the total narrow
within k 10 % of that specified for radiation generated up to
beam attenuation coefficients, O (tot) cm2/g, of Storm and
and including 30 kV and within If: 5 % for the higher energy
Israel [II, together with densities of aluminium and copper of
radiation. The minimum purity of the additional filters and of
2,702 and 8,94 g/cm3 respectively. The measured photon spec-
the absorbers used to determine the half value layer shall be
20-28 apply to the conditions listed in
99,9 % except in the case of aluminium used at and below tra shown in figures
table 11.
20 kV
...

PROJET D'ADDITIF IS0 -/DAD 1
ISO/TC Secrétariat
85 DIN
Soumis le Vote clos le
1982-03-18 1982-09-18
INTERNATIONAL ORGANIZATION FOR STANOAROIZATION- MEMflYHAPOflHAR OPrAHMBAUMR Il0 CTAHflAPTM3AL(MM. ORGANISATION INTERNATIONALE DE NORMALISATION
Rayonnements X et y de référence pour l'étalonnage des,
dosimètres et débitmètres et pour la détermination de leur
réponse en fonction de l'énergie des photons
ADDITIF 1 : Rayonnements X filtrés à fort débit de dose
CDU 535-341-36 : 53.089.6
Descripteurs : dosimètre, débitmètre d'exposition, étalonnage, rayonnement électromagnétique, matériel de référence,
matériel d'essai, résultats d'essai.
CE DOCUMENT EST UN PROJET DIFFUSE POUR COMMENTAIRES ET APPROBATION IL EST DONC SUSCEPTIBLE DE MODIFICATION ET NE PEUT ETRE CIrE
COMME NORME INTERNATIONALE AVANT SON ACCEPTATION PAR LE CONSEIL DE L'ISO
OUTRE LE FAIT D ETRE EXAMINES POUR ETABLIR S'ILS SONT ACCEPTABLES A DES FINS INDUSTRIELLES, TECHNOLOGIQUES ET COMMERCIALES, AINSI QUE
DU POINT DE VUE DES UTILISATEURS. LES PROJETS DE NORMES INTERNATIONALES DOIVENT PARFOIS ËTRE CONSIDERES DU POINT DE VUE DE LEUR
UTILISATION POSSIBLE COMME DOCUMENTS POUVANT SERVIR DE REFI%ENCE DANS LA REGLEMENTATION NATIONALE
Organisation internationale de normalisation, 1982

---------------------- Page: 1 ----------------------
PROJET D'ADDITIF IS0 -/DAD 1
Rayonnements X et y de référence pour l'étalonnage des
dosimètres et débitmètres et pour la détermination de leur
lant à 4 mm d'aluminium,"et il est nécessaire d'avoir un tube
1 Objet et domaine d'application
à faible filtration inhérente pour générer les rayonne-
radiogène
Le présent additif spécifie une série de rayonnements X légère- ments à plus faible énergie. Pour des tensions de 100 kV et
ment filtrés, convenant pour les procédures d'étalonnage dans au-dessus, la filtration fixe doit être réglée sur une valeur équi-
lesquelles de forts débits de dose sont nécessaires. Cette série valant à 4 mm d'aluminium. Le filtre en aluminium utilisé pour
de rayonnements est également applicable, par exemple, à la compléter la filtration inhérente du tube doit être placé après le
e
détermination des caractérisitques de saturation d'un dosimè- filtre en cuivre, de manière à réduire tout rayonnement fluores-
tre ou à la vérification de la linéarité.
cent provenant du cuivre. L'épaisseur de l'aluminium employé
ne doit pas être inférieure à 0.5 mm.
Etant donné que les spectres obtenus sont larges et asymétri-
ques, cette série de rayonnements n'est pas satisfaisante pour
2.4 Energie efficace
les mesurages précis de réponse en énergie et ne doit être
employée que si les débits de dose produits par les séries étroi-
Si nécessaire, l'énergie efficace') pour les qualités données
tes et larges se sont révélées inadéquats.
1, peut être déterminée A partir de la couche de
dans le tableau
demi-atténuation (en millimètres), en utilisant la relation
La qualité des rayonnements de référence est spécifiée en fonc-
tion de la tension du tube radiogène, de la couche de demi-
atténuation (CDA,) et du coefficient d'homogénéité. Les spéci-
fications sont données dans le tableau 10.
proques, approprié au matériau dans lequel la couche de demi-
atténuation a été spécifiée.
2 Filtration
2.5 Exemples de filtration additionnelle requise
2.1 Filtres
pour des caractéristiques de rayonnement
Des filtres en aluminium (potentiels du tube < 100 kV) ou en spécifiées
cuivre et aluminium ( > 100 kV) doivent être utilisés.
(I)
Le tableau 11 donne, à titre d'exemple, la filtration additionnelle
requise pour produire les qualités de rayonnement spécifiées
2.2 Filtration additionnelle
pour les valeurs particulières de la filtration fixe. La longueur du
parcours employé dans l'air, incluse dans la filtration addition-
Pour un potentiel donné, l'épaisseur de la filtration addition-
des rayonnements de plus fai-
nelle, est significative dans le cas
nelle doit être réglée de manière que la première couche de
ble énergie. Le tableau 2 comprend égale
...