ISO/DIS 81060-3
(Main)Non-invasive sphygmomanometers
Non-invasive sphygmomanometers
Sphygmomanomètres non invasifs
General Information
Standards Content (sample)
DRAFT INTERNATIONAL STANDARD
ISO/DIS 81060-3
ISO/TC 121/SC 3 Secretariat: ANSI
Voting begins on: Voting terminates on:
2019-12-13 2020-03-06
Non-invasive sphygmomanometers —
Part 3:
Clinical investigation of continuous automated
measurement type
Sphygmomanomètres non invasifs —
Partie 3: Validation clinique pour type à mesurage automatique continu
ICS: 11.040.10
THIS DOCUMENT IS A DRAFT CIRCULATED
This document is circulated as received from the committee secretariat.
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
ISO/CEN PARALLEL PROCESSING
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 81060-3:2019(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION. ISO 2019
---------------------- Page: 1 ----------------------
ISO/DIS 81060-3:2019(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2019
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2019 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/DIS 81060-3:2019(E)
Contents Page
Foreword ........................................................................................................................................................................................................................................iv
Introduction ..................................................................................................................................................................................................................................v
1 Scope ................................................................................................................................................................................................................................. 1
2 Normative references ...................................................................................................................................................................................... 1
3 Terms and definitions ..................................................................................................................................................................................... 2
4 Requirements specific to the reference methods................................................................................................................ 3
4.1 * Invasive reference method ...................................................................................................................................................... 3
4.1.1 Reference invasive blood pressure monitoring equipment ................................................ 3
4.1.2 Subject requirements ................................................................................................................................................... 3
4.1.3 Blood pressure distribution ................................................................................................................................. 5
4.1.4 * Arterial reference site ............................................................................................................................................. 5
4.2 Auscultatory reference method ............................................................................................................................................... 6
4.2.1 Reference sphygmomanometer ..................................................................................................................... 6
4.2.2 Subject requirements ................................................................................................................................................... 6
4.2.3 Blood pressure distribution ................................................................................................................................ 7
4.2.4 * Observer preparation .............................................................................................................................................. 7
4.2.5 * Reference readings ................................................................................................................................................. 8
4.3 Lateral difference .................................................................................................................................................................................. 9
4.3.1 General...................................................................................................................................................................................... 9
4.3.2 Sequential procedure ................................................................................................................................................ 9
4.3.3 Simultaneous procedure .....................................................................................................................................10
4.3.4 Application of lateral difference ......................................................................................................................10
5 General requirements for the clinical investigation ....................................................................................................10
5.1 Good clinical practice .....................................................................................................................................................................10
5.2 General ........................................................................................................................................................................................................10
5.3 Reference methods ..........................................................................................................................................................................11
6 Methods for clinical investigation ...................................................................................................................................................11
6.1 * Method for the accuracy of blood pressure determination ..................................................................11
6.1.1 Method utilizing the invasive reference method .................. ............................................................11
6.1.2 Method utilizing the auscultatory reference method ..................................................................12
6.1.3 Acceptance criteria .....................................................................................................................................................13
6.2 * Method for blood pressure changes .............................................................................................................................14
6.2.1 Change evaluation interval ............................................................................................................................14
6.2.2 Method utilizing the invasive reference method .................. ............................................................15
6.2.3 Method utilizing the auscultatory reference method ..................................................................17
6.2.4 Acceptance criteria .....................................................................................................................................................19
6.3 * Method for stability ......................................................................................................................................................................19
6.3.1 General...................................................................................................................................................................................19
6.3.2 Procedure .........................................................................................................................................................................19
6.3.3 Data analysis and acceptance criteria ........................................................................................................21
Annex A (informative) Rationale and guidance .....................................................................................................................................22
Annex B (informative) Guidance for the application of this document .........................................................................31
Annex C (informative) Terminology — alphabetized index of defined terms ........................................................33
Bibliography .............................................................................................................................................................................................................................34
© ISO 2019 – All rights reserved iii---------------------- Page: 3 ----------------------
ISO/DIS 81060-3:2019(E)
Foreword
ISO (the International Organization for Standardization) and IEC (the International Electrotechnical
Commission) form the specialized system for worldwide standardization. National bodies that are
members of ISO or IEC participate in the development of International Standards through technical
committees established by the respective organization to deal with particular fields of technical
activity. ISO and IEC technical committees collaborate in fields of mutual interest. Other international
organizations, governmental and non-governmental, in liaison with ISO and IEC, also take part in the
work. In the field of information technology, ISO and IEC have established a joint technical committee,
ISO/IEC JTC 1.The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for
the different types of document should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/ directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject
of patent rights. ISO and IEC shall not be held responsible for identifying any or all such patent
rights. Details of any patent rights identified during the development of the document will be in the
Introduction and/or on the ISO list of patent declarations received (see www .iso .org/ patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO's adherence to the WTO principles in the Technical
Barriers to Trade (TBT) see the following URL: Foreword - Supplementary information
ISO/IEC 81060-3 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory
equipment, Subcommittee SC 3, Respiratory devices and related equipment used for patient care,
in collaboration with Technical Committee IEC/TC 62, Electrical equipment in medical practice,
Subcommittee 62D, Electromedical equipment, in accordance with ISO/IEC mode of cooperation 5.
iv © ISO 2019 – All rights reserved---------------------- Page: 4 ----------------------
ISO/DIS 81060-3:2019(E)
Introduction
The number of continuously measuring non-invasive automated sphygmomanometers has increased
significantly in the last 10 years. This standard is intended to provide the necessary requirements
for clinical investigation to ensure that the essential performance of these sphygmomanometers is at an
adequate level, similar to those standards on intermittent automated non-invasive sphygmomanometer.
In this document, the following print types are used:— requirements, conformance with which can be verified, and definitions: roman type;
— notes and examples: smaller roman type;— defined terms and test methods: italic type;
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any combination
of the conditions is true.The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC Directives,
Part 2. For the purposes of this standard, the auxiliary verb:— “shall” means that conformance with a requirement or a test is mandatory for conformance with
this standard;— “should” means that conformance with a requirement or a test is recommended but is not mandatory
for conformance with this standard;— “may” is used to describe permission (e.g. a permissible way to achieve conformance with a
requirement or test);— "can" is used to describe a possibility or capability; and
— "must" is used to express an external constraint.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title indicates
that there is guidance or rationale related to that item in Annex A.© ISO 2019 – All rights reserved v
---------------------- Page: 5 ----------------------
DRAFT INTERNATIONAL STANDARD ISO/DIS 81060-3:2019(E)
Non-invasive sphygmomanometers —
Part 3:
Clinical investigation of continuous automated
measurement type
1 Scope
This document specifies the requirements and methods for the clinical investigation of continuous non-
invasive automated sphygmomanometers used for the measurement of the blood pressure of a subject.
Since this document covers both trending devices and absolute accuracy devices and focuses solely on
requirements for the clinical investigation, representation of output is not covered by this document.
NOTE 1 IEC 62366-1 provides requirements on the application of usability engineering to medical devices
which can be used to clarify for the intended user whether the shown data concerns absolute accurate values or
trending values.The requirements and methods for the clinical investigation of continuous non-invasive automated
sphygmomanometers provided in this document are applicable to any subject population, and any
condition of use of the continuous non-invasive automated sphygmomanometers.NOTE 2 Subject populations can, for example, be represented by age or weight ranges.
NOTE 3 Conditions of use can, for example, refer to ambulatory blood pressure monitoring, stress testing blood
pressure monitoring and blood pressure monitors for the home healthcare environment or self-measurement as
well as use in professional healthcare facility or the emergency medical service environment (EMS).
This document specifies additional disclosure requirements for the accompanying documents of
continuous non-invasive automated sphygmomanometers that have undergone clinical investigation
according to this document.This document is not applicable to the clinical investigation of a non-automated sphygmomanometer as
given in ISO 81060-1, the clinical investigation of an intermittent automated sphygmomanometer as given
in ISO 81060-2, an intermittent automated non-invasive sphygmomanometer as given in IEC 80601-2-30
or invasive blood pressure monitoring equipment as given in IEC 60601-2-34.2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
NOTE Informative references are listed in the Bibliography.ISO 14155, Clinical investigation of medical devices for human subjects — Good clinical practice
ISO 16142-1:2016, Medical devices — Recognized essential principles of safety and performance of medical
devices — Part 1: General essential principles and additional specific essential principles for all non-IVD
medical devices and guidance on the selection of standardsISO 81060-1:2007, Non-invasive sphygmomanometers — Part 1: Requirements and test methods for non-
automated measurement type© ISO 2019 – All rights reserved 1
---------------------- Page: 6 ----------------------
ISO/DIS 81060-3:2019(E)
ISO 81060-2:2018, Non-invasive sphygmomanometers — Part 2: Clinical investigation of intermittent
automated measurement typeIEC 60601-1:2005, Medical electrical equipment — Part 1: General requirements for basic safety and
essential performance +Amendment 1:2012IEC 60601-1-11:2015, Medical electrical equipment — Part 1-11: General requirements for basic safety and
essential performance — Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environmentIEC 60601-2-34:2011, Medical electrical equipment — Part 2-34: Particular requirements for the safety,
including essential performance of invasive blood pressure monitoring equipmentIEC 80601-2-30:2018, Medical electrical equipment — Part 2-30: Particular requirements for basic safety
and essential performance of automated non-invasive sphygmomanometers3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 14155, ISO 81060-1:2007,
ISO 81060-2:2018, IEC 60601-1:2005+AMD1: 2012, IEC 60601-2-34:2011, IEC 80601-2-30:2018 and the
following apply.ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http:// www .electropedia .org/— ISO Online browsing platform: available at https:// www .iso .org/ obp
Note 1 to entry For convenience, the sources of all defined terms used in this document are given in Annex C.
3.1change evaluation interval
time interval for which a continuous non-invasive automated sphygmomanometer is demonstrated to be
able to track changes in blood pressure3.2
continuous non-invasive automated sphygmomanometer
device able to estimate blood pressure from the pulse wave of each heart cycle without arterial
punctureNote 1 to entry: While the continuous non-invasive automated sphygmomanometer is able to
estimate the blood pressure from the pulse wave of each heart cycle, this does not mean the device
needs to use data from the pulse wave at each heart cycle. Not using data from a pulse wave at a specific
heart cycle can be useful, for example to omit data from premature ventricular contractions.Note 2
to entry: The manufacturer may choose the form and manner of the data display or output.
3.3initialization
re-initialization
process of the continuous non-invasive automated sphygmomanometer to determine subject- or condition-
specific parameters needed to estimate the blood pressureNote 1 to entry: In this document, the term initialization is used for the initial initialization; re-initialization is
used for the repeated process during the measurement period.3.4
intermittent automated non-invasive sphygmomanometer
automated sphygmomanometer estimating at intervals systolic blood pressure, diastolic blood pressure or
mean arterial pressure values over a series of cardiac cycles2 © ISO 2019 – All rights reserved
---------------------- Page: 7 ----------------------
ISO/DIS 81060-3:2019(E)
3.5
paired measurement
two measurements of the same blood pressure event within a subject's cardiac cycles, one of which
is recorded with the continuous non-invasive automated sphygmomanometer and the other with the
reference methodNote 1 to entry: An example for a blood pressure event is the occurrence of a systolic or a diastolic blood pressure.
3.6paired values
pair of blood pressure values as a result of a paired measurement
Note 1 to entry: The blood pressure values can be systolic, diastolic or mean blood pressure values.
3.7reference measurement
procedure defined in this standard with established accuracy used for the clinical investigation of a
continuous non-invasive automated sphygmomanometer3.8
reference reading
result of the process of measuring blood pressure using a reference method
Note 1 to entry: The result can be a systolic blood pressure, a diastolic blood pressure or a mean arterial pressure.
4 Requirements specific to the reference methods4.1 * Invasive reference method
4.1.1 Reference invasive blood pressure monitoring equipment
a) Reference invasive blood pressure monitoring equipment:
1) shall conform with the requirements of IEC 60601-2-34; except that
2) the maximum allowable error shall be ± 2 mmHg (± 0,27 kPa).
b) The resonance frequency and damping coefficient of the reference invasive blood pressure
monitoring equipment shall be shall be examined and optimised to meet dynamic requirements. See
[2]Reference .
c) The transducer shall be kept at the level of the heart.
d) Appropriate measures shall be taken to remove air bubbles and clots from the system prior to
taking the reference measurements.NOTE The ability to measure accurately blood pressure can be degraded by the presence of air bubbles
or blood clots in the catheter/transducer system.e) Reference invasive blood pressure monitoring equipment that does not directly output the blood
pressure waveform or beat-to-beat data may be modified to permit such data collection.
4.1.2 Subject requirements4.1.2.1 Number
A clinical investigation shall consist of
a) a minimum of 15 subjects,
© ISO 2019 – All rights reserved 3
---------------------- Page: 8 ----------------------
ISO/DIS 81060-3:2019(E)
b) with the exception of sphygmomanometers intended for use in neonates, infants and children of less
than 3 years of age for which it shall consist of a minimum of 18 subjects.Check conformance by inspection of the clinical investigation report.
4.1.2.2 Gender distribution
a) At least 30 % of the subjects shall be male.
b) At least 30 % of the subjects shall be female.
Check conformance by inspection of the clinical investigation report.
4.1.2.3 * Age distribution
4.1.2.3.1 Sphygmomanometers intended for use in adults and adolescents
For a continuous non-invasive automated sphygmomanometer intended for use in adult or adolescent
subjects, the age of every subject included in the clinical investigation shall be greater than 12 years.
NOTE Minimum total of 15 subjects.Check conformance by inspection of the accompanying document and the clinical investigation report.
4.1.2.3.2 Sphygmomanometers intended for use in children aged between 3 years and 12 years
a) For a continuous non-invasive automated sphygmomanometer intended for use in children aged
between 3 years and 12 years, the age of every subject included in the clinical investigation shall be
between 3 years and 12 years.NOTE Minimum total of 15 subjects.
b) Children are exempt from:
1) the gender distribution requirements of 4.1.2.2; and
2) the blood pressure distribution requirements of 4.1.3.
Check conformance by inspection of the accompanying document and the clinical investigation report.
4.1.2.3.3 Sphygmomanometer intended for use in neonates, infants and children of less than 3
years of agea) A continuous non-invasive automated sphygmomanometer intended for use in neonates, infants and
children of less than 3 years of age, shall be investigated in those subject populations.
b) The following age or weight ranges are required for a neonatal mode clinical investigation:
1) At least 3 subjects shall be less than 1 000 g in weight.2) At least 3 subjects shall be 1 000 g to 2 000 g in weight.
3) At least 3 subjects shall be more than 2 000 g in weight.
4) At least 3 subjects shall be at least 29 days but not yet 1 year of age.
5) At least 3 subjects shall be at least 1 year but not yet 3 years of age.
c) The remaining subjects may be from any of the above age or weight groups in order to complete the
sample size of 18.NOTE 1 Minimum total of 18 subjects.
4 © ISO 2019 – All rights reserved
---------------------- Page: 9 ----------------------
ISO/DIS 81060-3:2019(E)
NOTE 2 A subject can be in more than one category simultaneously.
d) Neonates, infants and children of less than 3 years of age are exempt from:
1) the gender distribution requirements of 4.1.2.2; and
2) the blood pressure distribution requirements of 4.1.3.
Check conformance by inspection of the accompanying document and the clinical investigation report.
4.1.2.4 * Special subject populationsa) When there is evidence that a certain subject characteristic might affect the performance of a
continuous non-invasive automated sphygmomanometer, and if this is within the intended use of the
device, that population (which is well defined by such subject characteristics) shall be considered a
special subject population.b) The continuous non-invasive automated sphygmomanometer shall be investigated across the range
of the subject characteristics that is within the intended use of the device.c) Unless otherwise justified, each special subject population identified shall be investigated
separately.Check conformance by inspection of the accompanying document and the clinical investigation report.
4.1.3 Blood pressure distributiona) At least 5 % of the reference readings shall have a systolic blood pressure less than or equal to
100 mmHg (13,33 kPa).b) At least 5 % of the reference readings shall have a systolic blood pressure greater than or equal to
160 mmHg (21,33 kPa).c) At least 20 % of the reference readings shall have a systolic blood pressure greater than or equal to
140 mmHg (18,67 kPa).d) At least 5 % of the reference readings shall have a diastolic blood pressure less than or equal to
60 mmHg (8,00 kPa).e) At least 5 % of the reference readings shall have a diastolic blood pressure greater than or equal to
100 mmHg (13,33 kPa).f) At least 20 % of the reference readings shall have a diastolic blood pressure greater than or equal to
85 mmHg (11,33 kPa).Check conformance by inspection of the clinical investigation report.
4.1.4 * Arterial reference site
a) Any reference site may be used for simultaneous comparison of intra-arterial blood pressure readings
and continuous non-invasive automated sphygmomanometer blood pressure determinations, but
b) the instructions for use of the continuous non-invasive automated sphygmomanometer shall disclose
the arterial site used as the reference site.NOTE Different sites can produce different results due to the pressure difference between the central
aorta and other arteries.c) If the opposite limb is used as the reference site, the lateral difference in blood pressure may be
determined and used according to 4.3.© ISO 2019 – All rights reserved 5
---------------------- Page: 10 ----------------------
ISO/DIS 81060-3:2019(E)
d) A continuous non-invasive automated sphygmomanometer claiming to output central or aortic blood
pressure values shall utilize a central or aortic arterial reference site for the clinical investigation.
Check conformance by inspection of the accompanying document.4.2 Auscultatory reference method
4.2.1 Reference sphygmomanometer
a) The auscultatory reference measurement shall be performed on the upper arm.
b) Use a reference sphygmomanometer that conforms with the requirements of ISO 81060-1, except
that the maximum permissible error shall be ± 1 mmHg (0,13 kPa).4.2.2 Subject requirements
4.2.2.1 * Number
a) An auscultatory reference sphygmomanometer validation study shall consist of a minimum of 85
subjects.b) For each parameter (e. g. systolic or diastolic blood pressure) to be validated at least 3 valid paired
measurements shall be carried out for each subject.NOTE This results in a minimum of 255 valid paired values for each parameter (e. g. systolic or diastolic blood
pressure) to be validated.Check conformance by inspection of the clinical investigation report.
4.2.2.2 * Gender distribution
a) At least 30 % of the subjects shall be male.
b) At least 30 % of the subjects shall be female.
Check conformance by inspection of the clinical investigation report.
4.2.2.3 * Age distribution
4.2.2.3.1 Sphygmomanometers intended for use in adults and adolescents
For a continuous non-invasive automated sphygmomanometer intended only for use in adults or adolescent
subjects, the age of every subject included in the clinical investigation shall be greater than 12 years.
NOTE Minimum total of 85 subjects.4.2.2.3.2 Sphygmomanometers intended for use in children aged between 3 years and 12 years
a) For a continuous non-invasive automated sphygmomanometer intended for use in children aged
between 3 years and 12 years, the age of every subject included in the clinical investigation shall be
between 3 years and 12 years.NOTE Minimum total of 85 subjects.
b) Children are exempt from:
1) the gender distribution requirements of 4.2.2.2; and
2) the blood pressure distribution requirements of 4.2.3.
6 © ISO 2019 – All rights reserved
---------------------- Page: 11 ----------------------
ISO/DIS 81060-3:2019(E)
NOTE The auscultatory reference method is not usable for neonates, infants and children of
...PROJET DE NORME INTERNATIONALE
ISO/DIS 81060-3
ISO/TC 121/SC 3 Secrétariat: ANSI
Début de vote: Vote clos le:
2019-12-13 2020-03-06
Sphygmomanomètres non invasifs —
Partie 3:
Investigation clinique pour type à mesurage
automatique continu
Non-invasive sphygmomanometers —
Part 3: Clinical investigation of continuous automated measurement type
ICS: 11.040.10
Il est demandé aux comités membres de consulter les intérêts nationaux
respectifs concernant l’IEC/SC 62D avant de donner leur position sur la
plateforme de e-Balloting.
CE DOCUMENT EST UN PROJET DIFFUSÉ POUR
OBSERVATIONS ET APPROBATION. IL EST DONC
SUSCEPTIBLE DE MODIFICATION ET NE PEUT
Le présent document est distribué tel qu’il est parvenu du secrétariat du comité.
ÊTRE CITÉ COMME NORME INTERNATIONALEAVANT SA PUBLICATION EN TANT QUE TELLE.
OUTRE LE FAIT D’ÊTRE EXAMINÉS POUR
ÉTABLIR S’ILS SONT ACCEPTABLES À DES
FINS INDUSTRIELLES, TECHNOLOGIQUES ET
COMMERCIALES, AINSI QUE DU POINT DE VUE TRAITEMENT PARALLÈLE ISO/CEN
DES UTILISATEURS, LES PROJETS DE NORMES
INTERNATIONALES DOIVENT PARFOIS ÊTRE
CONSIDÉRÉS DU POINT DE VUE DE LEUR
POSSIBILITÉ DE DEVENIR DES NORMES
POUVANT SERVIR DE RÉFÉRENCE DANS LA
RÉGLEMENTATION NATIONALE.
Numéro de référence
LES DESTINATAIRES DU PRÉSENT PROJET
ISO/DIS 81060-3:2019(F)
SONT INVITÉS À PRÉSENTER, AVEC LEURS
OBSERVATIONS, NOTIFICATION DES DROITS
DE PROPRIÉTÉ DONT ILS AURAIENT
ÉVENTUELLEMENT CONNAISSANCE ET À
FOURNIR UNE DOCUMENTATION EXPLICATIVE. ISO 2019
---------------------- Page: 1 ----------------------
ISO/DIS 81060-3:2020(F)
ISO/DIS 81060-3:2019(F)
Sommaire Page
Avant-propos ................................................................................................................................................................. iv
Introduction ..................................................................................................................................................................... v
1 Domaine d’application ................................................................................................................................... 1
2 Références normatives .................................................................................................................................. 2
3 Termes et définitions ..................................................................................................................................... 2
4 Exigences spécifiques aux méthodes de référence .............................................................................. 4
4.1 * Méthode de référence invasive ................................................................................................................ 4
4.1.1 Appareils de surveillance de la pression artérielle prélevée directement de référence ............ 4
4.1.2 Exigences relatives aux sujets ..................................................................................................................... 4
4.1.3 Distribution de la pression artérielle ........................................................................................................ 6
4.1.4 * Site artériel de référence ............................................................................................................................ 6
4.2 Méthode de référence auscultatoire ......................................................................................................... 7
4.2.1 Sphygmomanomètre de référence .............................................................................................................. 7
4.2.2 Exigences relatives aux sujets ..................................................................................................................... 7
4.2.3 Distribution de la pression artérielle ........................................................................................................ 8
4.2.4 * Préparation des observateurs ................................................................................................................. 9
4.2.5 * Lectures de référence ................................................................................................................................... 9
4.3 Différence de latéralité ............................................................................................................................... 10
4.3.1 Généralités ...................................................................................................................................................... 10
4.3.2 Procédure séquentielle ............................................................................................................................... 11
4.3.3 Procédure simultanée ................................................................................................................................. 11
4.3.4 Application de la différence de latéralité ............................................................................................ 12
5 Exigences générales relatives à l’investigation clinique ................................................................. 12
5.1 Bonnes pratiques cliniques ....................................................................................................................... 12
5.2 Généralités ...................................................................................................................................................... 12
5.3 Méthodes de référence ................................................................................................................................ 13
6 Méthodes applicables à l’investigation clinique ................................................................................. 13
6.1 * Méthode permettant de déterminer l’exactitude de la détermination de la pression
artérielle ........................................................................................................................................................... 13
6.1.1 Méthode utilisant la méthode de référence invasive ....................................................................... 13
6.1.2 Méthode utilisant la méthode de référence auscultatoire ............................................................. 14
6.1.3 Critères d’acceptation ................................................................................................................................. 15
6.2 * Méthode permettant de déterminer les variations de la pression artérielle ....................... 16
6.2.1 Intervalle d’évaluation des variations .................................................................................................... 16
6.2.2 Méthode utilisant la méthode de référence invasive ....................................................................... 17
6.2.3 Méthode utilisant la méthode de référence auscultatoire ............................................................. 20
6.2.4 Critères d’acceptation ................................................................................................................................. 22
6.3 * Méthode permettant de déterminer la stabilité ............................................................................. 22
DOCUMENT PROTÉGÉ PAR COPYRIGHT6.3.1 Généralités ...................................................................................................................................................... 22
6.3.2 Procédure ......................................................................................................................................................... 22
© ISO 20196.3.3 Analyse des données et critères d’acceptation .................................................................................. 24
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en oeuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
Annexe A (informative) Justification et lignes directrices ............................................................................ 26
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
Annexe B (informative) Recommandations pour l’application du présent document ...................... 36
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright officeAnnexe C (informative) Terminologie — Index alphabétique des termes définis ............................... 38
Case postale 401 • Ch. de Blandonnet 8CH-1214 Vernier, Geneva
Bibliographie ................................................................................................................................................................ 40
Tél.: +41 22 749 01 11Fax: +41 22 749 09 47
E-mail: copyright@iso.org
Website: www.iso.org
© ISO 2020 – Tous droits réservés
iii
Publié en Suisse
ii © ISO 2019 – Tous droits réservés
---------------------- Page: 2 ----------------------
ISO/DIS 81060-3:2019(F)
Sommaire Page
Avant-propos ................................................................................................................................................................. iv
Introduction ..................................................................................................................................................................... v
1 Domaine d’application ...................................................................................................................................1
2 Références normatives ..................................................................................................................................2
3 Termes et définitions .....................................................................................................................................2
4 Exigences spécifiques aux méthodes de référence ..............................................................................4
4.1 * Méthode de référence invasive.................................................................................................................4
4.1.1 Appareils de surveillance de la pression artérielle prélevée directement de référence ............4
4.1.2 Exigences relatives aux sujets .....................................................................................................................4
4.1.3 Distribution de la pression artérielle ........................................................................................................6
4.1.4 * Site artériel de référence ............................................................................................................................6
4.2 Méthode de référence auscultatoire .........................................................................................................7
4.2.1 Sphygmomanomètre de référence ..............................................................................................................7
4.2.2 Exigences relatives aux sujets .....................................................................................................................7
4.2.3 Distribution de la pression artérielle ........................................................................................................8
4.2.4 * Préparation des observateurs .................................................................................................................9
4.2.5 * Lectures de référence ...................................................................................................................................9
4.3 Différence de latéralité ............................................................................................................................... 10
4.3.1 Généralités ...................................................................................................................................................... 10
4.3.2 Procédure séquentielle ............................................................................................................................... 11
4.3.3 Procédure simultanée ................................................................................................................................. 11
4.3.4 Application de la différence de latéralité............................................................................................. 12
5 Exigences générales relatives à l’investigation clinique ................................................................. 12
5.1 Bonnes pratiques cliniques ....................................................................................................................... 12
5.2 Généralités ...................................................................................................................................................... 12
5.3 Méthodes de référence ................................................................................................................................ 13
6 Méthodes applicables à l’investigation clinique................................................................................. 13
6.1 * Méthode permettant de déterminer l’exactitude de la détermination de la pression
artérielle ........................................................................................................................................................... 13
6.1.1 Méthode utilisant la méthode de référence invasive ....................................................................... 13
6.1.2 Méthode utilisant la méthode de référence auscultatoire ............................................................. 14
6.1.3 Critères d’acceptation ................................................................................................................................. 15
6.2 * Méthode permettant de déterminer les variations de la pression artérielle ....................... 16
6.2.1 Intervalle d’évaluation des variations .................................................................................................... 16
6.2.2 Méthode utilisant la méthode de référence invasive ....................................................................... 17
6.2.3 Méthode utilisant la méthode de référence auscultatoire ............................................................. 20
6.2.4 Critères d’acceptation ................................................................................................................................. 22
6.3 * Méthode permettant de déterminer la stabilité ............................................................................. 22
6.3.1 Généralités ...................................................................................................................................................... 22
6.3.2 Procédure ......................................................................................................................................................... 22
6.3.3 Analyse des données et critères d’acceptation .................................................................................. 24
(informative) Justification et lignes directrices ............................................................................ 26
(informative) Recommandations pour l’application du présent document ...................... 36
(informative) Terminologie — Index alphabétique des termes définis ............................... 38
Bibliographie ................................................................................................................................................................ 40
© ISO 2019 – Tous droits réservésiii
---------------------- Page: 3 ----------------------
ISO/DIS 81060-3:2019(F)
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude a le
droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui
concerne la normalisation électrotechnique.Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).L’attention est attirée sur le fait que certains des éléments du présent document peuvent faire l’objet de
droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable
de ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant les
références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de
l’élaboration du document sont indiqués dans l’Introduction et/ou dans la liste des déclarations de
brevets reçues par l’ISO (voir www.iso.org/brevets).Les appellations commerciales éventuellement mentionnées dans le présent document sont données
pour information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.Pour une explication de la signification des termes et expressions spécifiques de l’ISO liés à l’évaluation
de la conformité, ou pour toute information au sujet de l’adhésion de l’ISO aux principes de
l’Organisation mondiale du commerce (OMC) concernant les obstacles techniques au commerce (OTC),
voir le lien suivant: http://www.iso.org/iso/fr/foreword.html.L’ISO/IEC 81060-3 a été élaborée par le comité technique ISO/TC 121, Matériel d’anesthésie et de
réanimation respiratoire, sous-comité SC 3, Appareils respiratoires et équipements connexes utilisés pour
les soins aux patients en collaboration avec le comité d’études IEC/TC 62, Équipements électriques dans
la pratique médicale, sous-comité 62D, Appareils électromédicaux, conformément au mode de
coopération 5 de l’ISO/IEC.© ISO 2019 – Tous droits réservés
---------------------- Page: 4 ----------------------
ISO/DIS 81060-3:2019(F)
Introduction
Le nombre de sphygmomanomètres automatiques non invasifs à mesurage continu a considérablement
augmenté ces dix dernières années. La présente norme est destinée à fournir les exigences nécessaires à
l’investigation clinique permettant de veiller à ce que le niveau des performances essentielles de ces
sphygmomanomètres soit adéquat, de manière similaire aux normes relatives aux sphygmomanomètres
non invasifs automatiques à usage ponctuel.Dans le présent document, les types de polices suivants sont utilisés :
— exigences dont la conformité peut être vérifiée et définitions : caractères romains ;
— notes et exemples : petits caractères romains ;— termes définis et méthodes d’essai : caractères italiques.
Dans la présente norme, la conjonction « ou » est utilisée comme « ou inclusif » ; une affirmation est
donc vraie si une combinaison quelconque des conditions est vraie.Les formes verbales utilisées dans la présente norme sont conformes à l’usage décrit dans l’Annexe H
des directives ISO/IEC, Partie 2. Pour les besoins de la présente norme, la forme verbale :
— « doit » signifie que la conformité à une exigence ou à un essai est obligatoire pour la conformité à
la présente norme ;— « il convient/il est recommandé » signifie que la conformité à une exigence ou à un essai est
recommandée, mais n’est pas obligatoire pour la conformité à la présente norme ;— « peut » est utilisée pour décrire une permission (par exemple un moyen admissible pour satisfaire
à une exigence ou un essai) ;— « peut » est également utilisée pour décrire une possibilité ou une capacité ; et
— « il faut » est utilisée pour exprimer une contrainte externe.Lorsqu’un astérisque (*) est utilisé comme premier caractère devant un titre ou au début d’un alinéa ou
d’un titre de tableau, il indique l’existence d’une recommandation ou d’une justification à consulter dans
l’Annexe A.© ISO 2019 – Tous droits réservés
---------------------- Page: 5 ----------------------
PROJET DE NORME INTERNATIONALE ISO/DIS 81060-3:2019(F)
Sphygmomanomètres non invasifs — Partie 3 : Investigation
clinique pour type à mesurage automatique continu
1 Domaine d’application
Le présent document spécifie les exigences et les méthodes relatives à l’investigation clinique des
sphygmomanomètres automatiques non invasifs à mesurage continu servant à mesurer la pression
artérielle d’un sujet.Comme le présent document traite à la fois des dispositifs de mesurage de tendance et des dispositifs à
exactitude absolue, et comme il se concentre uniquement sur les exigences de l’investigation clinique, la
représentation des données de sortie n’est pas couverte par le présent document.NOTE 1 L’IEC 62366-1 fournit des exigences concernant l’application de l’ingénierie de l’aptitude à l’utilisation
aux dispositifs médicaux qui peuvent être utilisés pour clarifier, auprès de l’utilisateur cible, si les données
indiquées concernent des valeurs à exactitude absolue ou des valeurs de tendance.
Les exigences et méthodes d’investigation clinique des sphygmomanomètres automatiques non invasifs à
mesurage continu données dans le présent document s’appliquent à toutes les populations de sujets et
toutes les conditions d’utilisation des sphygmomanomètres automatiques non invasifs à mesurage
continu.NOTE 2 Les populations de sujets peuvent par exemple être représentées par tranches d’âge ou gammes de
poids.NOTE 3 Les conditions d’utilisation peuvent par exemple faire référence à la surveillance ambulatoire de la
pression artérielle, à la surveillance de la pression artérielle lors d’épreuves d’effort, à la surveillance de la pression
artérielle en environnement des soins à domicile ou à l’auto-surveillance, et à la surveillance de la pression
artérielle dans un établissement de soins professionnel ou dans l’environnement des services médicaux d’urgence
(SMU).Le présent document spécifie des exigences supplémentaires de divulgation d’informations pour les
documents d’accompagnement des sphygmomanomètres automatiques non invasifs à mesurage continu
ayant subi une investigation clinique conformément au présent document.Le présent document n’est pas applicable à l’investigation clinique des sphygmomanomètres non
automatiques tels que définis dans l’ISO 81060-1, ou à l’investigation clinique des sphygmomanomètres
automatiques à usage ponctuel tels que définis dans l’ISO 81060-2, des sphygmomanomètres non invasifs
automatiques à usage intermittent tels que définis dans l’IEC 80601-2-30 ou les appareils de surveillance
de la pression artérielle prélevée directement tels que définis dans l’IEC 60601-2-34.
© ISO 2019 – Tous droits réservés---------------------- Page: 6 ----------------------
ISO/DIS 81060-3:2019(F)
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur
contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique.
Pour les références non datées, la dernière édition du document de référence s’applique (y compris les
éventuels amendements).NOTE La Bibliographie fournit une liste de références informatives.
ISO 14155, Investigation clinique des dispositifs médicaux pour sujets humains — Bonnes pratiques
cliniques.ISO 16142-1:2016, Dispositifs médicaux — Lignes directrices pour le choix des normes correspondant aux
principes essentiels reconnus de sécurité et de performance des dispositifs médicaux — Partie 1 : Principes
essentiels généraux et principes essentiels spécifiques supplémentaires pour tous les dispositifs médicaux
non DIV et directives sur le choix des normes.ISO 81060-1:2007, Sphygmomanomètres non invasifs — Partie 1 : Exigences et méthodes d’essai pour
type à mesurage non automatique.ISO 81060-2:2018, Sphygmomanomètres non invasifs — Partie 2 : Investigation clinique pour type
ponctuel à mesurage automatique.IEC 60601-1:2005, Appareils électromédicaux — Partie 1 : Exigences générales pour la sécurité de base et
les performances essentielles +Amendement 1:2012.IEC 60601-1-11:2015, Appareils électromédicaux — Partie 1-11 : Exigences générales pour la sécurité de
base et les performances essentielles — Norme collatérale : Exigences pour les appareils électromédicaux
et les systèmes électromédicaux utilisés dans l’environnement des soins à domicile.
IEC 60601-2-34:2011, Appareils électromédicaux — Partie 2-34 : Exigences particulières pour la sécurité
de base et les performances essentielles des appareils de surveillance de la pression artérielle prélevée
directement.IEC 80601-2-30:2018, Appareils électromédicaux — Partie 2-30 : Exigences particulières pour la sécurité
de base et les performances essentielles des sphygmomanomètres non invasifs automatiques.
3 Termes et définitionsPour les besoins du présent document, les termes et définitions de l’ISO 14155, l’ISO 81060-1:2007,
l’ISO 81060-2:2018, l’IEC 60601-1:2005+AMD1:2012, l’IEC 60601-2-34:2011, l’IEC 80601-2-30:2018
ainsi que les suivants s’appliquent.L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en
normalisation, consultables aux adresses suivantes :— IEC Electropedia : disponible à l’adresse http://www.electropedia.org/ ;
— ISO Online browsing platform : disponible à l’adresse https://www.iso.org/obp.
Note 1 à l’article : Pour plus de commodité, les sources de tous les termes définis utilisés dans le présent
document sont données dans l’Annex C.© ISO 2019 – Tous droits réservés
---------------------- Page: 7 ----------------------
ISO/DIS 81060-3:2019(F)
3.1
intervalle d’évaluation des variations
intervalle de temps durant lequel il a été démontré qu’un sphygmomanomètre automatique non invasif à
mesurage continu est capable de suivre les variations de la pression artérielle3.2
sphygmomanomètre automatique non invasif à mesurage continu
dispositif capable d’estimer la pression artérielle à partir de l’onde de pouls de chaque cycle cardiaque
sans ponction artérielleNote 1 à l’article : Même si les sphygmomanomètres automatiques non invasifs à mesurage continu sont capables
d’estimer la pression artérielle à partir de l’onde de pouls de chaque cycle cardiaque, cela ne signifie pas que ces
dispositifs ont besoin d’utiliser les données de l’onde de pouls de chacun des cycles cardiaques. Il peut être utile de
ne pas utiliser les données d’une onde de pouls d’un cycle cardiaque particulier, par exemple pour ne pas tenir
compte des données de l’extrasystole ventriculaire.Note 2 à l’article : Le fabricant peut décider de l’apparence des résultats ou des données affichées, ainsi que la
façon dont ils sont présentés.3.3
initialisation
ré-initialisation
processus mis en œuvre par le sphygmomanomètre automatique non invasif à mesurage continu pour
déterminer les paramètres spécifiques à un sujet, ou à un trouble, nécessaires à l’estimation de la
pression artérielleNote 1 à l’article : Dans le présent document, le terme « initialisation » est utilisé pour indiquer l’« initialisation »
initiale ; la « ré-initialisation » concerne le processus utilisé de manière répétée durant la période de mesurage.
3.4sphygmomanomètre automatique non invasif à usage ponctuel
sphygmomanomètre automatique permettant d’estimer, à intervalles réguliers, les valeurs de la pression
artérielle systolique, la pression artérielle diastolique ou la pression artérielle moyenne sur une série de
cycles cardiaques3.5
mesurage apparié
deux mesurages d’un même événement de pression artérielle au cours d’un cycle cardiaque d’un sujet,
l’un des mesurages étant enregistré avec le sphygmomanomètre automatique non invasif à mesurage
continu et l’autre en suivant la méthode de référenceNote 1 à l’article : Un événement de pression artérielle peut être par exemple la survenue d’une pression artérielle
diastolique ou systolique.3.6
valeurs appariées
paire de valeurs de pression artérielle résultant d’un mesurage apparié
Note 1 à l’article : Les valeurs de pression artérielle peuvent être des valeurs de pression artérielle moyenne,
systolique ou diastolique.3.7
mesurage de référence
procédure définie dans la présente norme et présentant une exactitude établie utilisée pour
l’investigation clinique d’un sphygmomanomètre automatique non invasif à mesurage continu
© ISO 2019 – Tous droits réservés---------------------- Page: 8 ----------------------
ISO/DIS 81060-3:2019(F)
3.8
lecture de référence
résultat du processus de mesure de la pression artérielle à l’aide de la méthode de référence
Note 1 à l’article : Le résultat peut être une pression artérielle systolique, une pression artérielle diastolique ou une
pression artérielle moyenne.4 Exigences spécifiques aux méthodes de référence
4.1 * Méthode de référence invasive
4.1.1 Appareils de surveillance de la pression artérielle prélevée directement de référence
a) Les appareils de surveillance de la pression artérielle prélevée directement de référence :
1) doivent respecter les exigences de l’IEC 60601-2-34 ; toute...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.