ISO/TC 121/SC 3 - Respiratory devices and related equipment used for patient care
Appareils respiratoires et équipements connexes utilisés pour les soins aux patients
General Information
This document specifies the requirements and methods for the clinical investigation of medical electrical (ME) equipment used to measure the body temperature in indirect measurement mode. This document covers both intermittently and continuously measuring clinical thermometers. NOTE 1 This document does not apply to clinical thermometers measuring the body temperature in direct measurement mode. NOTE 2 For clinical thermometers in direct measurement mode determining the technical accuracy in accordance with ISO 80601-2-56:—1) is considered sufficient. This document is applicable to clinical thermometers with claimed measurement time shorter than 60 seconds (for methods such as oral or rectal measurement), or shorter than 5 minutes (for methods such as axillary measurement), and which are treated as predictive type thermometers and fall under the scope of this document. This document specifies additional disclosure requirements. This document does not apply to the clinical investigation of a screening thermographs for human febrile temperature screening whose laboratory accuracy requirements are described in IEC 80601-2-59. This document does not apply to pulmonary artery catheter for the determination of cardiac output by thermodilution. NOTE 3 ISO 80601-2-56:—1) does include pulmonary artery catheter for the determination of cardiac output by thermodilution.
- Standard25 pagesEnglish languagesale 15% off
- Standard25 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of pulse oximeter equipment intended for use on humans, hereafter referred to as ME equipment. This includes any part necessary for normal use, including the pulse oximeter monitor, pulse oximeter probe, and probe cable extender. These requirements apply to pulse oximeter equipment, including pulse oximeter monitors, pulse oximeter probes and probe cable extenders regardless of their origin (i.e. including remanufactured products). The intended use of pulse oximeter equipment includes, but is not limited to, the estimation of arterial oxygen haemoglobin saturation and pulse rate of patients in professional healthcare institutions as well as patients in the home healthcare environment and the emergency medical services environment. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause says so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 201.11.1.2.2, IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 2 See also IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document can also be applied to ME equipment and their accessories used for compensation or alleviation of disease, injury, or disability. This document is not applicable to pulse oximeter equipment intended for use in laboratory research applications nor to oximeters that require a blood sample from the patient. This document is not applicable to pulse oximeter equipment intended solely for foetal use. This document is not applicable to remote or slave (secondary) equipment that displays SpO2 values that are located outside of the patient environment. NOTE 3 ME equipment that provides selection between diagnostic and monitoring functions is expected to meet the appropriate requirements of this document when configured for that function. This document is applicable to pulse oximeter equipment intended for use under extreme or uncontrolled environmental conditions outside the hospital environment or physician’s office, such as in ambulances and air transport. Additional standards can apply to pulse oximeter equipment for those environments of use. This document is a particular standard in the IEC 60601-1 and ISO and IEC 80601 series of standards.
- Standard150 pagesEnglish languagesale 15% off
- Standard159 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of a humidifier, also hereafter referred to as ME equipment, in combination with its accessories, the combination also hereafter referred to as ME system. This document is also applicable to those accessories intended by their manufacturer to be connected to a humidifier where the characteristics of those accessories can affect the basic safety or essential performance of the humidifier. EXAMPLE 1 Heated breathing tubes (heated-wire breathing tubes) or ME equipment intended to control these heated breathing tubes (heated breathing tube controllers). NOTE 2 Heated breathing tubes and their controllers are ME equipment and are subject to the requirements of IEC 60601‑1. NOTE 3 ISO 5367 specifies other safety and performance requirements for breathing tubes. This document includes requirements for the different medical uses of humidification, such as invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep apnoea therapy, as well as humidification therapy for tracheostomy patients. NOTE 4 A humidifier can be integrated into other equipment. When this is the case, the requirements of the other equipment also apply to the humidifier. EXAMPLE 2 Heated humidifier incorporated into a critical care ventilator where ISO 80601‑2-12 also applies. EXAMPLE 3 Heated humidifier incorporated into a homecare ventilator for dependent patients where ISO 80601‑2‑72 also applies. EXAMPLE 4 Heated humidifier incorporated into sleep apnoea therapy equipment where ISO 80601‑2‑70 also applies. EXAMPLE 5 Heated humidifier incorporated into ventilatory support equipment where either ISO 80601-2-79 or ISO 80601-2-80 also apply. EXAMPLE 6 Heated humidifier incorporated into respiratory high-flow therapy equipment where ISO 80601‑2‑90 also applies. This document also includes requirements for an active HME (heat and moisture exchanger), ME equipment which actively adds heat and moisture to increase the humidity level of the gas delivered from the HME to the patient. This document is not applicable to a passive HME, which returns a portion of the expired moisture and heat of the patient to the respiratory tract during inspiration without adding heat or moisture. NOTE 5 ISO 9360‑1 and ISO 9360‑2 specify safety and performance requirements for a passive HME. NOTE 6 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 7 Additional information can be found in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 4.2. This document does not specify the requirements for cold pass-over or cold bubble-through humidification devices, the requirements for which are given in ISO 20789. This document is not applicable to equipment commonly referred to as “room humidifiers” or humidifiers used in heating, ventilation and air conditioning systems, or humidifiers incorporated into infant incubators to humidify the chamber air (i.e., are not directly connected to the patient). This document is not applicable to nebulizers used for the delivery of a drug to patients. NOTE 8 ISO 27427 specifies the safety and performance requirements for nebulizers.
- Standard113 pagesEnglish languagesale 15% off
- Standard124 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of respiratory high-flow therapy equipment, as defined in 201.3.262, hereafter also referred to as ME equipment or ME system, in combination with its accessories: intended for use with patients who can breathe spontaneously; and intended for patients who would benefit from improved alveolar gas exchange; and who would benefit from receiving high-flow humidified respiratory gases, which can include a patient whose upper airway is bypassed. EXAMPLE 1 Patients with Type 1 Respiratory Failure who exhibit a reduction in arterial blood oxygenation. EXAMPLE 2 Patients who would benefit from reduced work of breathing, as needed in Type 2 Respiratory Failure, where arterial carbon dioxide is high. EXAMPLE 3 Patients requiring humidification to improve mucociliary clearance. Respiratory high-flow therapy equipment is utilized in both professional healthcare facilities and the home healthcare environment. This standard specifically addresses respiratory high-flow therapy equipment for acute or infant care, predominantly found in hospitals. A separate document for long term high-flow therapy in the home healthcare environment is expected to be forthcoming. Respiratory high-flow therapy equipment can be: fully integrated ME equipment; or a combination of separate items forming a ME system. This document also applies to other types of respiratory equipment when that equipment includes a respiratory high-flow therapy mode. NOTE 2 This document and ISO 80601-2-12 are applicable to a critical care ventilator with a high-flow therapy mode. NOTE 3 This document and ISO 80601-2-72 are applicable to ventilator for ventilator-dependent patients in the home healthcare environment with a high-flow therapy mode. NOTE 4 This document and ISO 80601-2-13 are applicable to an anaesthetic workstation with a high-flow therapy mode. Respiratory high-flow therapy equipment can be transit-operable. This document is also applicable to those accessories intended by their manufacturer to be connected to the respiratory high-flow therapy equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the respiratory high-flow therapy equipment. EXAMPLE 4 Breathing sets, connectors, humidifier, breathing system filter, external electrical power source, distributed alarm system, high-flow nasal cannula, tracheal tube, tracheostomy tube, face mask and supra-laryngeal airway. NOTE 5 Accessories are assessed with the relevant clauses of this document when configured as part of respiratory high-flow therapy equipment. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in the general standard, 7.2.13 and 8.4.1. NOTE 6 Additional information can be found in the general standard, 4.2. This document does not specify the requirements for: ventilators or accessories for ventilator-dependent patients intended for critical care applications, which are given in ISO 80601‑2‑12; ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13; ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601‑2‑84; ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2‑72; ventilatory support equipment or accessories intended for patients with ventilatory impairment, which are given in ISO 80601‑2‑79; ventilatory support equipment or accessories intende
- Standard109 pagesEnglish languagesale 15% off
- Standard116 pagesFrench languagesale 15% off
This document is applicable to the basic safety and essential performance of oxygen conserving equipment, hereafter referred to as ME equipment, in combination with its accessories intended to conserve supplemental oxygen by delivering gas intermittently and synchronized with the patient's inspiratory cycle, when used in the home healthcare environment. Oxygen conserving equipment is typically used by a lay operator. NOTE 1 Conserving equipment can also be used in professional health care facilities. This document is also applicable to conserving equipment that is incorporated with other equipment. EXAMPLE Conserving equipment combined with a pressure regulator[4], an oxygen concentrator[12] or liquid oxygen equipment[7]. This document is also applicable to those accessories intended by their manufacturer to be connected to conserving equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the conserving equipment. This document is intended to clarify the difference in operation of various conserving equipment models, as well as between the operation of conserving equipment and continuous flow oxygen equipment, by requiring standardized performance testing and labelling. This document is only applicable to active devices (e.g. pneumatically or electrically powered) and is not applicable to non-active devices (e.g. reservoir cannulas). If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 2 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.
- Standard59 pagesEnglish languagesale 15% off
- Standard62 pagesFrench languagesale 15% off
This document is applicable to a transit-operable and non-transit-operable oxygen concentrator. This document is applicable to an oxygen concentrator integrated into or used with other medical devices, ME equipment or ME systems. EXAMPLE 1 An oxygen concentrator with integrated oxygen conserving equipment function or humidifier function. EXAMPLE 2 An oxygen concentrator used with a flowmeter stand. EXAMPLE 3 An oxygen concentrator as part of an anaesthetic system for use in areas with limited logistical supplies of electricity and anaesthetic gases[2]. EXAMPLE 4 An oxygen concentrator with an integrated liquid reservoir function or gas cylinder filling system function. This document is also applicable to those accessories intended by their manufacturer to be connected to an oxygen concentrator, where the characteristics of those accessories can affect the basic safety or essential performance of the oxygen concentrator. NOTE 2 Such accessories can include, but are not limited to, masks, cannulae, extension tubing, humidifiers, carts, carrying cases, external power sources and oxygen conserving equipment. This document does not specify requirements for oxygen concentrators for use with a medical gas pipeline system. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. NOTE 3 See also 4.2 of the general standard.
- Standard68 pagesEnglish languagesale 15% off
- Standard76 pagesFrench languagesale 15% off
This document specifies requirements for masks and accessories, including any connecting element, that are required to connect the patient-connection port of sleep apnoea breathing therapy equipment to a patient for the application of sleep apnoea breathing therapy (e.g. nasal masks, exhaust ports and headgear). This document applies to masks and their accessories used to connect sleep apnoea breathing therapy equipment to the patient. The requirements in this document take priority over the requirements in ISO 18190. This document does not cover oral appliances. NOTE This document has been prepared to address the relevant essential principles[14] and labelling principles[15] of the International Medical Devices Regulators Forum (IMDRF) as indicated in Annex I.
- Standard36 pagesEnglish languagesale 15% off
- Standard37 pagesFrench languagesale 15% off
This document is applicable to the basic safety and essential performance of sleep apnoea breathing therapy equipment, hereafter referred to as ME equipment, intended to alleviate the symptoms of patients who suffer from obstructive sleep apnoea by delivering a therapeutic breathing pressure to the respiratory tract of the patient. Sleep apnoea breathing therapy equipment is intended for use in the home healthcare environment by lay operators as well as in professional healthcare institutions. Sleep apnoea breathing therapy equipment is not considered to utilize a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the therapy settings. This document excludes sleep apnoea breathing therapy equipment intended for use with neonates. This document is applicable to ME equipment or an ME system intended for those patients who are not dependent on artificial ventilation. This document is not applicable to ME equipment or an ME system intended for those patients who are dependent on artificial ventilation such as patients with central sleep apnoea. This document is also applicable to those accessories intended by their manufacturer to be connected to sleep apnoea breathing therapy equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the sleep apnoea breathing therapy equipment. Masks and application accessories intended for use during sleep apnoea breathing therapy are additionally addressed by ISO 17510. Refer to Figure AA.1 for items covered further under this document. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. NOTE 2 See also 4.2 of the general standard. This document does not specify the requirements for: – ventilators or accessories intended for critical care ventilators for ventilator-dependent patients, which are given in ISO 80601‑2‑12. – ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601-2-13. – ventilators or accessories intended for home care ventilators for ventilator-dependent patients, which are given in ISO 80601-2-72. – ventilators or accessories intended for emergency and transport, which are given in ISO 80601-2-84. – ventilators or accessories intended for home-care ventilatory support, which are given in ISO 80601‑2-79 and ISO 80601‑2‑80. – high-frequency ventilators[23], which are given in ISO 80601-2-87. – respiratory high flow equipment, which are given in ISO 80601‑2‑90; NOTE 3 ISO 80601-2-80 ventilatory support equipment can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. – user-powered resuscitators, which are given in ISO 10651-4; – gas-powered emergency resuscitators, which are given in ISO 10651-5; – oxygen therapy constant flow ME equipment; and – cuirass or “iron-lung” ventilation equipment.
- Standard78 pagesEnglish languagesale 15% off
- Standard83 pagesFrench languagesale 15% off
This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of FUNCTIONAL NIRS EQUIPMENT, as defined in 201.3.205, intended to be used by itself, or as a part of an ME SYSTEM hereinafter referred to as ME EQUIPMENT. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, 7.2.13 and 8.4.1. This document is not applicable to equipment for the measurement of oxygen saturation of the haemoglobin in the micro vessels (capillaries, arterioles and venules), i.e. tissue oximeters; frequency-domain and time-domain equipment for functional near-infrared spectroscopy; equipment for the measurement of changes in the concentration of chromophores other than oxy- and deoxy-haemoglobin; equipment for the measurement of changes in the concentration of oxy- and deoxy-haemoglobin in tissues other than the brain. This document does not specify the requirements for: cerebral tissue oximeter equipment, which are given in ISO 80601-2-85; and pulse oximeter equipment, which are given in ISO 80601-2-61.
- Draft48 pagesEnglish languagesale 15% off
- Standard9 pagesEnglish and French languagesale 15% off
- Draft9 pagesEnglish languagesale 15% off
NOTE 1 There is guidance or rationale for this subclause contained in Clause AA.2. This document applies to the basic safety and essential performance of ventilatory support equipment, as defined in 201.3.302, for ventilatory impairment, as defined in 201.3.300, hereafter also referred to as ME equipment, in combination with its accessories: — intended for use in the home healthcare environment; NOTE 2 In the home healthcare environment, the supply mains driving the ventilatory support equipment is often not reliable. NOTE 3 Such ventilatory support equipment can also be used in professional health care facilities. — intended for use by a lay operator; — intended for use with patients who have ventilatory impairment, the most fragile of these patients, would not likely experience injury with the loss of this artificial ventilation; and — not intended for patients who are dependent on artificial ventilation for their immediate life support. EXAMPLE 1 Patients with mild to moderate chronic obstructive pulmonary disease (COPD). Ventilatory support equipment is not considered to use a physiologic closed-loop control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to the breathing system of ventilatory support equipment for ventilatory impairment, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilatory support equipment for ventilatory impairment. EXAMPLE 2 Breathing sets, connectors, water traps, expiratory valve, humidifier, breathing system filter, external electrical power source, distributed alarm system. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 4 Additional information can be found in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 4.2. NOTE 5 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. This document does not specify the requirements for: — ventilators or accessories for ventilator-dependent patients intended for critical care applications, which are given in ISO 80601‑2‑12; — ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13; — ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601‑2‑84; — ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2‑72; — ventilatory support equipment or accessories intended for ventilatory insufficiency, which are given in ISO 80601‑2‑80; — sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70; — high-frequency jet ventilators (HFJVs), which are given in ISO 80601‑2‑87; — high-frequency oscillatory ventilators (HFOVs); — respiratory high flow equipment, which are given in ISO 80601‑2‑90; NOTE 6 Ventilatory support equipment can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. — user-powered resuscitators, which are given in ISO 10651-4; — gas-powered emergency resuscitators, which are given in ISO 10651-5; — oxygen therapy constant flow ME equipment; and — cuirass or “iron-lung” ventilation equipment.
- Standard107 pagesEnglish languagesale 15% off
- Standard118 pagesFrench languagesale 15% off
NOTE 1 There is guidance or rationale for this subclause contained in Clause AA.2. This document applies to the basic safety and essential performance of ventilatory support equipment, as defined in 201.3.302, for ventilatory insufficiency, as defined in 201.3.302, hereafter also referred to as ME equipment, in combination with its accessories: — intended for use in the home healthcare environment; NOTE 2 In the home healthcare environment, the supply mains driving the ventilatory support equipment is often not reliable. NOTE 3 Such ventilatory support equipment can also be used in professional health care facilities. — intended for use by a lay operator; — intended for use with patients who have ventilatory insufficiency or failure, the most fragile of which would likely experience injury with the loss of this artificial ventilation; — intended for transit-operable use; and — not intended for patients who are dependent on artificial ventilation for their immediate life support. EXAMPLE 1 Patients with moderate to severe chronic obstructive pulmonary disease (COPD), moderate amyotrophic lateral sclerosis (ALS), severe bronchopulmonary dysplasia or muscular dystrophy. Ventilatory support equipment is not considered to use a physiologic closed-loop control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system of ventilatory support equipment for ventilatory insufficiency, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilatory support equipment for ventilatory insufficiency. EXAMPLE 2 Breathing sets, connectors, water traps, expiratory valve, humidifier, breathing system filter, external electrical power source, distributed alarm system. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 4 Additional information can be found in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 4.2. NOTE 5 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. This document does not specify the requirements for: — ventilators or accessories for ventilator-dependent patients intended for critical care applications, which are given in ISO 80601‑2‑12; — ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13; — ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601‑2‑84; — ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2‑72; — ventilatory support equipment or accessories intended for ventilatory impairment, which are given in ISO 80601‑2‑79; — sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70; — high-frequency jet ventilators (HFJVs), which are given in ISO 80601‑2‑87; — high-frequency oscillatory ventilators (HFOVs); — respiratory high flow equipment, which are given in ISO 80601‑2‑90; NOTE 6 Ventilatory support equipment can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. — user-powered resuscitators, which are given in ISO 10651-4; — gas-powered emergency resuscitators, which are given in ISO 10651-5; — oxygen therapy constant flow ME equipment; and — cuirass or “iron-lung” ventilation equipment
- Standard120 pagesEnglish languagesale 15% off
- Standard130 pagesFrench languagesale 15% off
- Standard5 pagesEnglish languagesale 15% off
- Standard5 pagesFrench languagesale 15% off
This document specifies tests for the emissions of particulate matter from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify particles from 0,25 µm diameter to 10 µm diameter that are emitted by the medical device, its parts or accessories into the respirable gas stream. This document establishes acceptance criteria for these tests. This document does not address nanoparticles. Insufficient data exist to establish exposure limits for particles less than 0,25 µm diameter. This document does not address particles larger than 10 µm diameter. These particles are deposited in the nasal cavity. Additional information can be needed for medical devices or accessories that bypass the nose. This is outside the scope of this document but can be required by some authorities having jurisdiction. This document therefore adopts the same approach as the US Environmental Protection Agency (EPA) in setting limits based solely on particle size and not their chemistry. This document addresses potential contamination of the gas stream arising from the gas pathways, which is then conducted to the patient. This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing. This document does not address biological evaluation of the particles that are deliberately released by a nebulizer (i.e. the therapeutic agent). This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories, containing gas pathways that are addressed by this document, include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces, and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 (all parts).
- Standard16 pagesEnglish languagesale 15% off
- Standard16 pagesFrench languagesale 15% off
This document specifies tests for the emissions of volatile organic substances from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify emissions of volatile organic substances that are added to the respirable gas stream by the materials of the gas pathway. This document establishes acceptance criteria for these tests. NOTE Gaseous emission of volatile organic substances includes emissions of volatile organic compounds, semi-volatile organic compounds and very volatile organic compounds. This document addresses potential contamination of the gas stream arising from the gas pathways of medical devices or accessories, which is then conducted to the patient. This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing. This document does not address biological evaluation of the surfaces of gas pathways that are in direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 series. This document is intended to be read in conjunction with ISO 18562-1.
- Standard15 pagesEnglish languagesale 15% off
- Standard16 pagesFrench languagesale 15% off
This document specifies: — the general principles governing the biological evaluation within a risk management process of the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments; — the general categorization of gas pathways based on the nature and duration of their contact with the gas stream; — the evaluation of existing relevant data from all sources; — the identification of gaps in the available data set on the basis of a risk analysis; — the identification of additional data sets necessary to analyse the biological safety of the gas pathway; — the assessment of the biological safety of the gas pathway. This document covers general principles regarding biocompatibility assessment of medical device materials, which make up the gas pathway, in normal use and normal condition. This document does not cover biological hazards arising from mechanical damage. The other parts of ISO 18562 cover specific tests that address potentially hazardous substances that are added to the respirable gas stream and establish acceptance criteria for these substances. This document addresses potential contamination of the gas stream arising from the gas pathways within the medical device, which might then be conducted to the patient. This document applies over the expected lifetime of the medical device when operated according to the instructions for use. This includes degradation arising from exposure to environmental conditions as well as cleaning, disinfection and sterilisation (i.e. processing). It also includes user action or inaction (omission) that leads to an unintended or unexpected outcome (result) (i.e. use error). It does not include conscious/intentional action or inaction that violates the instructions for use and is beyond reasonable risk control by the manufacturer (i.e. abnormal use). This document does not address biological evaluation of the surfaces of medical devices that have direct contact with the patient or user. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving equipment, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing system filters and Y-pieces as well as any breathing accessories intended to be used with such medical devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device is not addressed by ISO 18562 (all parts).
- Standard34 pagesEnglish languagesale 15% off
- Standard35 pagesFrench languagesale 15% off
This document specifies tests for substances leached by liquid water condensing in gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The chemical characterization methods described in this document apply to chemical substances that could leach from the medical device, its parts or accessories into the condensate. This document establishes verifiable acceptance criteria for these tests. The identity and quantity of each chemical released is intended for toxicological risk assessment as described in ISO 18562-1:2024. This document addresses potential contamination of the gas stream arising from the gas pathways, which deliver breathing gas to the patient. This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing. This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device. This document does not address contact with drugs or anaesthetic agents. If a medical device or accessory is intended to be used with anaesthetic agents or drugs, then additional testing can be required. This document is intended to quantify hazardous water-soluble substances that are leached from the medical device, its parts or accessories by condensate and then conveyed by that liquid to the patient.
- Standard16 pagesEnglish languagesale 15% off
- Standard17 pagesFrench languagesale 15% off
- Standard6 pagesEnglish and French languagesale 15% off
- Draft6 pagesEnglish languagesale 15% off
This document applies to the basic safety and essential performance of a critical care ventilator in combination with its accessories, hereafter referred to as ME equipment: ¾ intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility; NOTE 2 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining. NOTE 3 For the purposes of this document, such a critical care ventilator can provide ventilation during transport within a professional healthcare facility (i.e. be a transit-operable ventilator). NOTE 4 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator. ¾ intended to be operated by a healthcare professional operator; and ¾ intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. A critical care ventilator is not considered to use a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator. NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 6 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation. NOTE 7 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining. This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. This document does not specify the requirements for: NOTE 8 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. ¾ ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13; ¾ ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84; ¾ ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72; ¾ ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79 and ISO 80601-2-80; ¾ obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70; ¾ continuous positive airway pressure (CPAP) ME equipment. ¾ high-frequency ventilators, which are given in ISO 80601‑2‑87; NOTE 9 A critical care ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilator-operational modes. ¾ respiratory high-flow therapy equipment, which are given in ISO 80601‑2‑90; NOTE 10 A critical care ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing p
- Standard158 pagesEnglish languagesale 15% off
- Standard171 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of an EMS ventilator in combination with its accessories, hereafter also referred to as ME equipment: ¾ intended for patients who need differing levels of support from artificial ventilation including ventilator-dependent patients; ¾ intended to be operated by a healthcare professional operator; ¾ intended for use in the EMS environment; and ¾ intended for invasive or non-invasive ventilation. NOTE 2 An EMS ventilator can also be used for transport within a professional healthcare facility. An EMS ventilator is not considered to use a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system, or to an EMS ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the EMS ventilator. NOTE 3 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 4 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document does not specify the requirements for the following: NOTE 5 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. ¾ ventilators or accessories intended for ventilator-dependent patients in critical care applications, which are given in ISO 80601-2-12. ¾ ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601-2-72. ¾ ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13. ¾ ventilators or accessories intended for ventilatory support equipment (intended only to augment the ventilation of spontaneously breathing patients), which are given in ISO 80601‑2-79 and ISO 80601-2-80. ¾ obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‐2‐70. ¾ user-powered resuscitators, which are given in ISO 10651‐4. ¾ gas-powered emergency resuscitators, which are given in ISO 10651‐5. ¾ continuous positive airway pressure (CPAP) ME equipment. ¾ high‐frequency jet ventilators (HFJVs), which are given in ISO 80601-2-87. ¾ high‐frequency oscillatory ventilators (HFOVs) REF _Ref69393406 \r \h \* MERGEFORMAT [44] 08D0C9EA79F9BACE118C8200AA004BA90B02000000080000000D0000005F00520065006600360039003300390033003400300036000000 , which are given in ISO 80601-2-87. NOTE 6 An EMS ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilation-modes. ¾ respiratory high-flow therapy equipment, which are given in ISO 80601-2-90. NOTE 7 An EMS ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. ¾ oxygen therapy constant flow ME equipment. ¾ cuirass or “iron‐lung” ventilators.
- Standard138 pagesEnglish languagesale 15% off
- Standard148 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of a ventilator in combination with its accessories, hereafter referred to as ME equipment: — intended for use in the home healthcare environment; — intended for use by a lay operator; and — intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system or to a ventilator where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator.
- Standard142 pagesEnglish languagesale 15% off
- Standard153 pagesFrench languagesale 15% off
This document specifies requirements for user-powered resuscitators intended for use with all age groups and which are intended to provide lung ventilation to patients whose breathing is inadequate. User-powered resuscitators are designated according to ideal body mass range. Example user-powered resuscitators include: — self-inflating bag resuscitators intended to be squeezed by the user’s hand and refilled by elastic recoil; and NOTE 1 Self-inflating bag resuscitators are generally transit-operable and can be used in a wide range of environmental and emergency situations. — flow-inflating bag resuscitators intended to be squeezed by the user’s hand and refilled by a flow from a medical gas source. This document is also applicable to those accessories that are intended for use with resuscitators where the characteristics of those accessories can affect the safety of the user-powered resuscitator. Examples of such accessories include face masks, PEEP valves, capnometric indicators, manometers, metronomes, flow restrictors, filters, gas refill valves, oxygen gas mixers, connectors, electronic feedback devices, electronic sensors and transmission of data to other equipment. This document is also applicable to point-of-use packaging. This document does not specify the requirements for: — gas-powered emergency resuscitators, which are given in ISO 10651-5; — electrically-powered resuscitators; — gas powered resuscitators for professional healthcare facilities; and — anaesthetic reservoir bags, which are given in ISO 5362. NOTE 2 This document has been prepared to address the relevant essential principles[24] and labelling[25] guidances of the International Medical Devices Regulators Forum (IMDRF) as indicated in Annex D. NOTE 3 This document has been prepared to address the relevant essential principles of safety and performance of ISO 16142-1:2016 as indicated in Annex E. NOTE 4 This document has been prepared to address the relevant general safety and performance requirements of European regulation (EU) 2017/745[23] as indicated in Annex F.
- Standard66 pagesEnglish languagesale 15% off
- Standard80 pagesFrench languagesale 15% off
This document specifies the requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers used for the measurement of the blood pressure of a patient. This document does not cover usability aspects such as the form and manner of the data display or output. This document does not specify a numerical threshold on the minimum output period. A continuous automated non-invasive sphygmomanometer providing blood pressure parameters (e.g., systolic blood pressure, diastolic blood pressure or mean arterial pressure) with an output period considerably larger than 30 s is not typically considered a continuous automated non-invasive sphygmomanometer. This document covers both trending continuous automated non-invasive sphygmomanometers and absolute accuracy continuous automated non-invasive sphygmomanometers and focuses solely on requirements for the clinical investigation. Representation of output is not covered by this document. NOTE 1 IEC 62366-1 provides requirements on the application of usability engineering to medical devices. The usability engineering process can be used to clarify for the intended user whether the displayed data concerns absolute accurate values or trending values. The requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers provided in this document are applicable to any subject population, and any condition of use of the continuous automated non-invasive sphygmomanometers. NOTE 2 Subject populations can, for example, be represented by age or weight ranges. NOTE 3 This document does not provide a method to assess the effect of artefacts during the clinical investigation (e.g. motion artefacts induced by the movement of the subject or the movement of the platform supporting the subject). This document specifies additional disclosure requirements for the accompanying documents of continuous automated non-invasive sphygmomanometers that have undergone clinical investigation according to this document. This document is not applicable to: — the clinical investigation of a non-automated sphygmomanometer as given in ISO 81060-1, — the clinical investigation of an intermittent automated non-invasive sphygmomanometer as given in ISO 81060-2, — an automated non-invasive sphygmomanometer as given in IEC 80601-2-30, or — invasive blood pressure monitoring equipment as given in IEC 60601‑2‑34.
- Standard36 pagesEnglish languagesale 15% off
- Standard40 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of an infant cardiorespiratory monitor, as defined in 3.10, hereafter also referred to as ME equipment, in combination with its accessories: — intended for use in the home healthcare environment; — intended for use by a lay operator; — intended to monitor cardiorespiratory parameters in sleeping or resting children under three years of age; and — intended for transit-operable use. NOTE An infant cardiorespiratory monitor can also be used in professional health care facilities. This document is also applicable to those accessories intended by their manufacturer to be connected to the infant cardiorespiratory monitor, where the characteristics of those accessories can affect the basic safety or essential performance of the infant cardiorespiratory monitor. EXAMPLE probes, cables distributed alarm system
- Standard57 pagesEnglish languagesale 15% off
- Standard60 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of a high-frequency ventilator (HFV) in combination with its accessories, hereafter referred to as ME equipment: intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility; NOTE 1 For the purposes of this document, such an environment is referred to as a critical care environment. High-frequency ventilators for this environment are considered life-sustaining. NOTE 2 For the purposes of this document, such a high-frequency ventilator can provide transport within a professional healthcare facility (i.e., be a transit-operable ventilator). NOTE 3 A high-frequency ventilator intended for use in transport within a professional healthcare facility is not considered as a ventilator intended for the emergency medical services environment. intended to be operated by a healthcare professional operator; intended for those patients who need differing levels of support from artificial ventilation including ventilator-dependent patients; and capable of providing more than 150 inflations/min. There are three principal designations of HFV: high-frequency percussive ventilation [HFPV, with a typical HFV frequency of (60 to 1 000) HFV inflations/min]; high-frequency jet ventilation [HFJV, with a typical HFV frequency of (100 to 1 500) HFV inflations/min]; and high-frequency oscillatory ventilation [HFOV, with a typical HFV frequency of (180 to 1200) HFV inflations/min and typically having an active expiratory phase]. Additionally, HFV designations can be combined together or with ventilation at rates less than 150 inflations/min. * A high-frequency ventilator is not considered a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to an HFV breathing system, or to a high-frequency ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the high-frequency ventilator. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of IEC 60601-1:2005. NOTE 4 Additional information can be found in 4.2 of IEC 60601-1:2005+AMD1:2012. This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. This document does not specify the requirements for: non-high-frequency ventilators or accessories which provide conventional ventilation for use in critical care environments, which are given in ISO 80601-2-12 [23];. NOTE 5 An HFV can incorporate conventional critical care ventilator operational modes, in which case ISO 80601-2-12 is applicable to those modes. ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601-2-13 [24]; ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84, the replacement for ISO 10651-3 [13]; NOTE 6 An HFV can incorporate EMS ventilator capability. ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72 [26]; ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-
- Standard136 pagesEnglish languagesale 15% off
- Standard136 pagesEnglish languagesale 15% off
- Standard144 pagesFrench languagesale 15% off
- Standard144 pagesFrench languagesale 15% off
This document applies to basic safety and essential performance of cerebral tissue oximeter equipment, that employs light at multiple wavelengths to derive a quantitative measure of oxygen saturation of haemoglobin within the volume of tissue sampled under the probe attached to the head. The cerebral tissue oximeter equipment can be based on continuous light, frequency domain or time domain technologies. This document applies to ME equipment used in a hospital environment as well as when used outside the hospital environment, such as in ambulances and air transport. Additional standards may apply to ME equipment for those environments of use. NOTE 1 Cerebral tissue oximeters are sometimes referred to as near infrared spectroscopy equipment in medical literature. Not included within the scope of this document are: invasive tissue or vascular oximeters; oximeters that require a blood sample from the patient; equipment measuring dissolved oxygen; ME equipment, or part thereof, that measures path-length-dependent haemoglobin change. The requirements for functional near-infrared spectroscopy equipment are found in ISO 80601-2-71[4]; ME equipment, or part thereof, that measures arterial saturation based on pulsatile changes in tissue optical properties (SpO2). The requirements for pulse oximeter equipment are found in ISO 80601‑2‑61[3]; ME equipment, or any part thereof, that claims to monitor tissue in parts of the body other than the head. This document also applies to cerebral tissue oximeter equipment, including cerebral tissue oximeter monitors, cerebral tissue oximeter probes and probe cable extenders, that have been remanufactured. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 201.11 and in 201.7.2.13 and 201.8.4.1 of the general standard. NOTE 2 See also 4.2 of the general standard. This document can also be applied to ME equipment and their accessories used for compensation or alleviation of disease, injury or disability. This document is not applicable to remote or slave (secondary) equipment that displays StO2 values that are located outside of the patient environment. NOTE 3 ME equipment that provides selection between diagnostic and monitoring functions is expected to meet the requirements of the appropriate document when configured for that function.
- Standard107 pagesEnglish languagesale 15% off
- Standard107 pagesEnglish languagesale 15% off
- Standard128 pagesFrench languagesale 15% off
- Standard128 pagesFrench languagesale 15% off
- Draft4 pagesEnglish languagesale 15% off
- Standard58 pagesEnglish languagesale 15% off
- Standard58 pagesEnglish and French languagesale 15% off
- Amendment with corrigendum56 pagesFrench languagesale 15% off
- Draft7 pagesEnglish languagesale 15% off
This document specifies requirements for the repeatability and reproducibility of non-invasive blood pressure (NIBP) simulators intended to test automated sphygmomanometers utilizing the oscillometric non-continuous method only. In addition, the pulse rate set on the NIBP simulator is tested. This document is not intended to relate the signals, generated by the NIBP simulator, to the oscillometric signal recorded in a cuff attached to a human. It does not intend to test the interaction between the NIBP simulator and the tested automated sphygmomanometer (e.g. the agreement of the set values of the NIBP simulator and the displayed values of the tested automated sphygmomanometer or the properties of the cuff and tubing, such as design or elastic properties). NOTE 1 These parameters can be tested separately in a clinical investigation or by using different special test setups. This document does not check whether or not the NIBP simulator is able to test the accuracy of the absolute blood pressure value of oscillometric automated sphygmomanometers. NOTE 2 Usually this is tested by a clinical investigation according ISO 81060-2 or other protocols. This document is applicable to NIBP simulators testing automated sphygmomanometers for adults, children and neonates at the upper arm, thigh etc. and automated sphygmomanometers measuring at the wrist.
- Technical specification11 pagesEnglish languagesale 15% off
- Technical specification11 pagesEnglish languagesale 15% off
- Standard3 pagesEnglish languagesale 15% off
- Standard3 pagesFrench languagesale 15% off
This part of the 80601 International Standard applies to the basic safety and essential performance of electroencephalographs as defined in 201.3.204, hereafter also referred to as me equipment or me system. This document is applicable to electroencephalographs intended for use in professional healthcare facilities, the emergency medical services environment or the home healthcare environment. This document does not cover requirements for other equipment used in electroencephalography such as: - phono-photic stimulators; - EEG data storage and retrieval; - ME EQUIPMENT particularly intended for monitoring during electroconvulsive therapy. If a clause or subclause is specifically intended to be applicable to me equipment only, or to me systems only, the title or content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to me equipment and to me systems, as follows. The clause or subclause applies to me equipment, as default. For me equipment with the corresponding safety measure or function not completely integrated into the me equipment but instead implemented in an me system, the me equipment manufacturer specifies in the accompanying documents which functionality and safety requirements are provided by the me system to comply with this document. The me system is verified accordingly. Hazards inherent in the intended physiological function of me equipment or me systems within the scope of this document are not covered by specific requirements in this document. NOTE See also 4.2 of the general standard.
- Standard75 pagesEnglish and French languagesale 15% off
- Standard2 pagesEnglish languagesale 15% off
- Standard2 pagesFrench languagesale 15% off
This document specifies the requirements and methods for the clinical investigation of me equipment used for the intermittent non-invasive automated estimation of the arterial blood pressure by utilizing a cuff. This document is applicable to all sphygmomanometers that sense or display pulsations, flow or sounds for the estimation, display or recording of blood pressure. These sphygmomanometers need not have automatic cuff inflation. This document covers sphygmomanometers intended for use in all patient populations (e.g. all age and weight ranges), and all conditions of use (e.g. ambulatory blood pressure monitoring, stress testing blood pressure monitoring and blood pressure monitors for the home healthcare environment for self-measurement as well as use in a professional healthcare facility). EXAMPLE Automated sphygmomanometer as given in IEC 80601-2-30 undergoing clinical investigation according to this document. This document specifies additional disclosure requirements for the accompanying documents of sphygmomanometers that have passed a clinical investigation according to this document. This document is not applicable to clinical investigations of non-automated sphygmomanometers as given in ISO 81060-1 or invasive blood pressure monitoring equipment as given in IEC 60601-2-34. This document is not applicable to clinical investigations of a set of cuffs that are not of same materials and construction. Each type of cuff set is required to be evaluated separately according to this document.
- Standard36 pagesEnglish languagesale 15% off
- Standard36 pagesEnglish languagesale 15% off
- Standard38 pagesFrench languagesale 15% off
- Standard38 pagesFrench languagesale 15% off
This document specifies requirements for so-called "cold bubble-through" or "cold pass-over" humidifying equipment, hereafter referred to as a passive humidifier. Figure 1 and Figure 2 illustrate these passive humidifiers.
- Standard45 pagesEnglish languagesale 15% off
- Standard45 pagesEnglish languagesale 15% off
- Standard47 pagesFrench languagesale 15% off
- Standard47 pagesFrench languagesale 15% off
- Standard47 pagesFrench languagesale 15% off
This document considers and identifies criteria about the intended patient, intended use environment, and intended operator across the spectrum of the types of ventilation-related equipment as listed below: — gas-powered resuscitator as specified in ISO 10651-5[1] [1]; — operator-powered resuscitator as specified in ISO 10651-4[2]; — ventilator for critical care as specified in ISO 80601-2-12[3] [2]; — ventilator for emergency medical services environment as specified in ISO 80601-2-84[4] [3], the future replacement for ISO 10651-3[5]; NOTE 1 ISO 80601‐2‐84 updates the content of ISO 10651‐3 and harmonizes it with IEC 60601-1:2005+AMD1:2012[6] and IEC 60601-1-12:2014[7]. — ventilator for ventilatory impairment in the home healthcare environment as specified in ISO 80601‑2‑79[8]; — ventilator for ventilatory insufficiency in the home healthcare environment as specified in ISO 80601‑2‑80[9]; — ventilator for ventilator-dependent patients in the home healthcare environment as specified in ISO 80601-2-72[10]; — sleep apnoea breathing therapy equipment as specified in ISO 80601-2-70[11]. NOTE 2 Sleep apnoea breathing therapy equipment is not considered to be an artificial ventilator. It is included in this discussion to highlight the differences, which indicate why sleep apnoea breathing therapy equipment is not considered a ventilator. This document is intended to provide guidance that can assist manufacturers, authorities having jurisdiction and users in the development, selection and application of different types of ventilatory equipment based on the intended patient, intended use environment and intended operator. [1] Numbers in square brackets refer to the Bibliography. [2] Under preparation. Stage at the time of publication: ISO/FDIS 80601-2-12:2018. [3] Under preparation. Stage at the time of publication: ISO/DIS 80601-2-84:2018.
- Technical report40 pagesEnglish languagesale 15% off
- Technical report40 pagesEnglish languagesale 15% off
IEC 80601-2-30:2018 applies to the basic safety and essential performance of automated sphygmomanometers, hereafter referred to as ME equipment, which by means of an inflatable cuff, are used for non continuous indirect estimation of the blood pressure without arterial puncture. This document specifies requirements for the basic safety and essential performance for this ME equipment and its accessories, including the requirements for the accuracy of a determination. This document covers automatic electrically-powered ME equipment used for the intermittent, indirect estimation of the blood pressure without arterial puncture, including blood pressure monitors for the home healthcare environment. Requirements for indirect estimation of the blood pressure without arterial puncture ME equipment with an electrically-powered pressure transducer and/or displays used in conjunction with a stethoscope or other manual methods for determining blood pressure (non-automated sphygmomanometers) are specified in document ISO 81060-1. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 201.11 and 201.105.3.3, as well as 7.2.13 and 8.4.1 of IEC 60601-1:2005. This second edition cancels and replaces the first edition published in 2009 and Amendment 1:2013. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) alignment with IEC 60601-1:2005/AMD1:2012 and IEC 60601-1-8:2006/AMD1:2012 , and with IEC 60601-1-2:2014 and IEC 60601-1-11:2015; b) referencing IEC 60601-1-10:2007 and IEC 60601-1-12; c) changing an operator-accessible cuff-sphygmomanometer connector from not compatible with the ISO 594 series to compatible with the ISO 80369 series; d) added additional requirements for public self-use sphygmomanometers; e) added a list of primary operating functions.
- Standard104 pagesEnglish and French languagesale 15% off
IEC 80601-2-49:2018 applies to basic safety and essential performance requirements of multifunction patient monitors, hereafter referred to as ME equipment or medical electrical systems. This particular standard applies to multifunction patient monitors intended for use in professional healthcare facilities as well as in the emergency medical service environment or the home healthcare environment. The scope of this document is restricted to ME equipment or medical electrical systems intended for connection to a single patient that has two or more physiological monitoring units. For purposes of this document, a pregnant mother and her fetus(es) are considered a single patient. This document does not specify requirements for individual physiological monitoring units such as ECG, invasive pressure and pulse oximetry. The particular standards related to these physiological monitoring units specify requirements from the perspective of stand-alone ME equipment. This particular standard addresses the additional requirements related to multifunction patient monitors. Multifunction patient monitors can be integrated into other ME equipment or medical electrical systems. When this is the case, other relevant standards also apply. This document does not apply to implantable parts of multifunction patient monitors. This first edition cancels and replaces the second edition of IEC 60601-2-49, published in 2011. This edition constitutes a technical revision to align with the current edition and Amendment to IEC 60601-1, new versions of collateral standards and amendments thereto. Major changes are in Clause 208 because many of the former requirements are now addressed by IEC 60601-1-8.
- Standard69 pagesEnglish and French languagesale 15% off
IEC 80601-2-59:2017 applies to the basic safety and essential performance of screening thermographs intended to be used for the individual non-invasive febrile temperature screening of a human under controlled environmental conditions, hereafter referred to as ME equipment. This document sets laboratory characterization test limits for the screening thermograph. This edition includes the following significant technical changes with respect to the previous edition: a) updates of the normative references and the bibliography; b) expansion of the applicability to pandemic infectious diseases in general.
- Standard75 pagesEnglish and French languagesale 15% off
ISO/TR 13154:2017 provides general guidelines for the deployment, implementation and operation of a screening thermograph intended to be used for non-invasive febrile temperature screening of individuals under indoor environmental conditions to prevent the spread of infection. NOTE The equipment standard for screening thermographs is found in IEC 80601?2-59.
- Technical report23 pagesEnglish languagesale 15% off
- Technical report23 pagesEnglish languagesale 15% off
ISO 80601-2-56:2017 applies to the basic safety and essential performance of a clinical thermometer in combination with its accessories, hereafter referred to as me equipment. This document specifies the general and technical requirements for electrical clinical thermometers. This document applies to all electrical clinical thermometers that are used for measuring the body temperature of patients. Clinical thermometers can be equipped with interfaces to accommodate secondary indicators, printing equipment, and other auxiliary equipment to create me systems. This document does not apply to auxiliary equipment. Me equipment that measures a body temperature is inside the scope of this document. ISO 80601-2-56:2017 does not specify the requirements for screening thermographs intended to be used for the individual non-invasive human febrile temperature screening of groups of individual humans under indoor environmental conditions, which are given in IEC 80601‑2‑59[4]. If a clause or subclause is specifically intended to be applicable to me equipment only, or to me systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to me equipment and to me systems, as relevant. Hazards inherent in the intended physiological function of me equipment or me systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+A1:2012, 7.2.13 and 8.4.1. NOTE Additional information can be found in IEC 60601?1:2005+A1:2012, 4.2.
- Standard49 pagesEnglish languagesale 15% off
- Standard49 pagesEnglish languagesale 15% off
- Standard51 pagesFrench languagesale 15% off
- Standard51 pagesFrench languagesale 15% off
IEC 60601-1-12:2014 constitutes a collateral standard to IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance hereafter referred to as the general standard. Medical practice is increasingly using medical electrical equipment and medical electrical systems for monitoring, treatment or diagnosis of patients in the emergency medical services environment. The safety of medical electrical equipment in this uncontrolled, rough environment is a cause for concern. This collateral standard was developed with contributions from clinicians, engineers and regulators. The terminology, requirements, general recommendations and guidance of this collateral standard are intended to be useful for manufacturers of medical electrical equipment and medical electrical systems and for technical committees responsible for the development of particular standards. This International Standard applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems, hereafter referred to as ME equipment and ME systems, which are intended, as indicated in the instructions for use by their manufacturer, for use in the EMS environment (Emergency Medical Services environment). The object of this collateral standard is to provide general requirements for ME equipment and ME systems carried to the scene of an emergency and used there, as well as in transport, in situations where the ambient conditions differ from indoor conditions. The object of this collateral standard is to specify general requirements that are in addition to those of the general standard and to serve as the basis for particular standards.
- Standard100 pagesEnglish languagesale 15% off
ISO 23747:2015 specifies requirements for a peak expiratory flow meter (pefm) intended for the assessment of pulmonary function in spontaneously breathing humans. ISO 23747:2015 covers all medical devices that measure peak expiratory flowrate in spontaneously breathing humans either as part of an integrated lung function medical device or as a stand-alone medical device. Planning and design of products applying to this International Standard are to consider the environmental impact from the product during its life cycle. Environmental aspects are addressed in Annex E. NOTE Additional aspects of environmental impact are addressed in ISO 14971.
- Standard26 pagesEnglish languagesale 15% off
- Standard26 pagesEnglish languagesale 15% off
- Standard26 pagesFrench languagesale 15% off
- Standard26 pagesFrench languagesale 15% off
IEC 60601-1-11:2015 applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems for use in the home healthcare environment. It applies regardless of whether the medical electrical equipment or medical electrical system is intended for use by a lay operator or by trained healthcare personnel. The home healthcare environment includes: - the dwelling place in which a patient lives; - other places where patients are present both indoors and outdoors, excluding professional healthcare facility environments where operators with medical training are continually available when patients are present. This second edition cancels and replaces the first edition of IEC 60601-1-11, published in 2010, and constitutes a technical revision. The most significant changes with respect to the previous edition include the following modifications: - correction of test method for relative humidity control at temperatures above 35 °C; - redrafting of subclauses that altered instead of adding to the general standard or other collateral standards; and - harmonizing with the changes to the amendments to the general standard and other collateral standards.
- Standard100 pagesEnglish languagesale 15% off
- Standard109 pagesEnglish and French languagesale 15% off
- Standard10 pagesEnglish languagesale 15% off
- Corrigendum60 pagesEnglish languagesale 15% off
ISO 26782:2009 specifies requirements for spirometers intended for the assessment of pulmonary function in humans weighing more than 10 kg. ISO 26782:2009 applies to spirometers that measure timed forced expired volumes, either as part of an integrated lung function device or as a stand-alone device, irrespective of the measuring method employed.
- Standard27 pagesEnglish languagesale 15% off
- Standard27 pagesFrench languagesale 15% off
- Standard35 pagesRussian languagesale 15% off
Specifies requirements for the development (analysis, design, verification and validation) of a physiologic closed-loop controller (PCLC) as part of a physiologic closed-loop control system (PCLCS) in medical electrical equipment and medical electrical systems to control a physiologic variable.
- Standard80 pagesEnglish languagesale 15% off
Frequently Asked Questions
ISO/TC 121/SC 3 is a Subcommittee within the International Organization for Standardization (ISO). It is named "Respiratory devices and related equipment used for patient care". This committee has published 118 standards.
ISO/TC 121/SC 3 develops ISO standards. Currently, there are 118 published standards from this subcommittee.
The International Organization for Standardization (ISO) is an independent, non-governmental international organization that develops and publishes international standards. Founded in 1947 and headquartered in Geneva, Switzerland, ISO brings together experts from 170+ member countries to share knowledge and develop voluntary, consensus-based standards that support innovation and provide solutions to global challenges.
A Subcommittee (SC) in ISO operates under a Technical Committee and focuses on a specific subset of the TC's scope. Subcommittees develop standards and technical specifications in their specialized area, reporting to their parent Technical Committee. They may also have working groups for detailed technical work.